St Jude Medical SJMRF Rf Implantable Medical Device User Manual

St. Jude Medical Rf Implantable Medical Device

User Manual

Current™/Current RFDR Model 2107–36/30RF DR Model 2207–36/30VR Model 1107–36/30RF VR Model 1207–36/30Implantable Cardioverter-DefibrillatorPromote™/Promote RFModel 3107-36/30RF Model 3207-36/30Cardiac Resynchronization Device,Implantable Cardioverter-DefibrillatorUser’s ManualDRAFT
CAUTIONFederal (USA) law restricts this device to sale by oron the order of a physician.© 2007 St. Jude Medical Cardiac Rhythm Management Division.All Rights Reserved.Unless otherwise noted, ® or ™ indicates that the name is a trademark of,or licensed to, St. Jude Medical, Inc. or its subsidiaries.DRAFT
Current™ and Promote™ Devices User’s Manual 1Device DescriptionThis manual describes the following St. Jude Medical® pulse generators:The pulse generator, along with compatible, commercially available leads, con-stitutes the implantable portion of the ICD and CRT-D systems. The lead sys-tems are implanted using either transvenous or transthoracic techniques. The St. Jude Medical Merlin™ Patient Care System (PCS) with software version 6.0 (or greater), a Merlin Antenna (for devices with RF commnication), and a telem-etry wand constitute the external portion of the ICD and CRT-D systems.Indications and UsageThe Current™ and Promote™ pulse generators are intended to provide ventric-ular antitachycardia pacing and ventricular defibrillation for automated treat-ment of life-threatening ventricular arrhythmias.Name Model # Description Delivered Energy (approx.)Current DR 2107-36 Dual-chamber ICD  36 JCurrent DR 2107-30 Dual-chamber ICD  30 JCurrent VR 1107-36 Single-chamber ICD  36 JCurrent VR 1107-30 Single-chamber ICD 30 JCurrent RF DR 2207-36 Dual-chamber ICD with RF telemetry 36 JCurrent RF DR 2207-30 Dual-chamber ICD with RF telemetry 30 JCurrent RF VR 1207-36 Single-chamber ICD with RF telemetry 36 JCurrent RF VR 1207-30 Single-chamber ICD with RF telemetry 30 JPromote 3107-36 CRT-D 36 JPromote 3107-30 CRT-D 30 JPromote RF 3207-36 CRT-D with RF telemetry 36 JPromote RF 3207-30 CRT-D with RF telemetry 30 JTable 1.  Current™ and Promote™ pulse-generator descriptionsDRAFT
2 ContraindicationsAF suppression pacing is indicated for suppression of paroxysmal or persistent atrial fibrillation in patients with the above ICD indication and sinus node dys-function.In patients indicated for an ICD, the Promote pulse generators are also intended:• to provide a reduction of the symptoms of moderate to severe heart failure (NYHA Functional Class III or IV) in those patients who remain symptomatic despite stable, optimal medical therapy (as defined in the clinical trials sec-tion included in the Merlin PCS on-screen help) and have a left ventricular ejection fraction less than or equal to 35% and a prolonged QRS duration• to maintain synchrony of the left and right ventricles in patients who have undergone an AV nodal ablation for chronic (permanent) atrial fibrillation and have NYHA Class II or III heart failureContraindicationsContraindications for use of the pulse generator system include ventricular tachyarrhythmias resulting from transient or correctable factors such as drug toxicity, electrolyte imbalance, or acute myocardial infarction.Warnings and PrecautionsResuscitation Availability. Do not perform device testing unless an external defibrillator and medical personnel skilled in cardiopulmonary resuscitation (CPR) are readily available.Lead system. Do not use another manufacturer’s lead system without demon-strated compatibility as undersensing cardiac activity and failure to deliver nec-essary therapy may result.Avoiding shock during handling. Disable tachyarrhythmia therapy (Enable/Disable Tachy Therapy) or program tachyarrhythmia therapies Off during surgi-cal implant and explant or post-mortem procedures as well as when discon-necting leads as the device can deliver a serious shock if you touch the defibrillation terminals while the device is charged.Additional pacemaker implanted. These devices provide bradycardia pacing. If another pacemaker is used, it should have a bipolar pacing reset mode and be programmed for bipolar pacing to minimize the possibility of the output pulses being detected by the device.Modifying the device. This device has been tested for compliance to FCC reg-ulations. Changes or modifications of any kind not expressly approved by St. Jude Medical Inc. could void the user’s authority to operate this device.DRAFT
Current™ and Promote™ Devices User’s Manual 3Suboptimal radio frequency (RF) communication. The Merlin PCS indicates the quality of the RF communication by the telemetry strength indicator LEDs on both the Merlin PCS and the Merlin Antenna. Below is a list of potential causes to suboptimal radio communication:Sterilization, Storage and HandlingResterilization. Do not resterilize and re-implant explanted pulse generators.Use before date. Do not implant the device after the “use before” date because the battery may have reduced longevity.If package is damaged. Do not use the device or accessories if the packaging is wet, punctured, opened or damaged because the integrity of the sterile packaging may be compromised. Return the device to St. Jude Medical.Device storage. Store the device in a clean area, away from magnets, kits con-taining magnets, and sources of electromagnetic interference (See ”Environ-mental and Medical Therapy Hazards” on page 5) to avoid device damage. Store the device between 10° and 45°C because temperatures outside this range may damage the device. Temperature Equilibration. After cold storage, allow the device to reach room temperature before charging the capacitors, programming, or implanting the device because cold temperature may affect initial device function.Possible Causes SolutionsThe Merlin Antenna orientation/location is suboptimal. Move or reorient the Merlin Antenna slightly. Make sure that the front of the Merlin Antenna faces the implantable device.People or objects interfere with the communication between the Merlin Antenna and the device. Make sure that the space between the Merlin Antenna and the device is free from interfering objects/people.The Merlin Antenna is too far away from the device.Move the Merlin Antenna closer to the device. Someone is holding the Merlin Antenna. Place the Merlin Antenna on a flat surface. Do not hold the Merlin Antenna.Other products in the vicinity are causing electromagnetic interference (EMI).Power off or remove equipment that could cause EMI.The Merlin Antenna cable is wound around the Merlin Antenna.Make sure the Merlin Antenna cable is not wound around the Merlin Antenna.Table 2.  Possible causes and solutions for suboptimal RF communicationDRAFT
4 Warnings and PrecautionsFollow-up TestingEnsure that an external defibrillator and medical personnel skilled in cardiopul-monary resuscitation (CPR) are present during post-implant device testing should the patient require external rescue.Be aware that the changes in the patient’s condition, drug regimen, and other factors may change the defibrillation threshold (DFT), which may result in non-conversion of the arrhythmia. Successful conversion of ventricular fibrillation or ventricular tachycardia during arrhythmia conversion testing is no assurance that conversion will occur post-operatively.Implantation and Device ProgrammingDo not position a magnet over the device as that suspends detection and treat-ment (unless the device has been programmed to ignore the magnet).Replace the device when the battery voltage reaches 2.45 V. Implant the pulse generator no deeper than 5 cm to ensure reliable data trans-mission. For patient comfort, do not implant the pulse generator within 1.25 cm of bone unless you cannot avoid it.Program the device parameters as specified in the Merlin PCS on-screen help.The results of the DAVID Study demonstrated that, for patients with standard indications for ICD therapy, no indication for cardiac pacing and an EF < 40%, dual-chamber pacing offers no clinical advantage over backup VVI pacing and may be associated with worsening heart failure.1 When programming the device to dual-chamber pacing modes, give particular attention to setting the pacing parameters (such as the A-V delay) to promote intrinsic conduction and minimize the amount of ventricular pacing.Pulse Generator Explant and DisposalInterrogate the device and turn all therapies off before explanting, cleaning or shipping the device to prevent unwanted shocks.Return all explanted pulse generators and leads to St. Jude Medical.Never incinerate the device because of the potential for explosion. Explant the device before cremation.1. Wilkoff BL, Cook JR, Epstein AE, Greene L, Hallstrom AP, Hsia H, Kutalek SP, Sharma A. Dual-Chamber Pacing or Ventricular Backup Pacing in Patients With an Implantable Defibrillator: The Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA. December 25, 2002; Vol 288, No. 24:3115-3123.DRAFT
Current™ and Promote™ Devices User’s Manual 5Environmental and Medical Therapy HazardsInstruct patients to avoid devices which generate a strong electric or magnetic interference (EMI). EMI could cause device malfunction or damage, resulting in non-detection or delivery of unneeded therapy. Moving away from the source or turning it off will usually allow the pulse generator to return to its normal mode of operation.Hospital and Medical EnvironmentsElectrosurgical cautery. Electrosurgical cautery could induce ventricular arrhythmias and/or fibrillation, or may cause device malfunction or damage. If electrocautery is necessary, keep the current path and groundplate as far away from the pulse generator and leads as possible.External defibrillation. External defibrillation may damage the pulse genera-tor or may result in temporary and/or permanent myocardial damage at the electrode-tissue interface as well as temporarily or permanently elevated pacing capture thresholds. Minimize current flowing through the pulse generator and lead system by following these precautions when using external defibrillation on a patient with a pulse generator:• Position defibrillation paddles as far from the pulse generator as possible (minimum of 13 cm)• Use the lowest clinically appropriate energy output• Confirm pulse generator function following any external defibrillationHigh radiation sources. Do not direct high radiation sources such as cobalt 60 or gamma radiation at the pulse generator. If a patient requires radiation therapy in the vicinity of the pulse generator, place lead shielding over the device to prevent radiation damage and confirm its function after treatment.Lithotripsy. Lithotripsy may permanently damage the pulse generator. Avoid it unless the therapy site is not near the pulse generator and leads.Diathermy. Avoid diathermy, even if the device is programmed off, as it may damage tissue around the implanted electrodes or may permanently damage the pulse generator.Magnetic resonance imaging (MRI). MRI for patients with implantable pulse generators has been contraindicated by MRI manufacturers. Clinicians should carefully weigh the decision to use MRI with ICD/CRT-D patients. MRI may cause device malfunction or injury to the patient.Ultrasound therapy. Diagnostic and therapeutic ultrasound treatment is not known to affect the function of the pulse generator.Transcutaneous Electrical Nerve Stimulation (TENS). TENS may interfere with device function. To reduce interference, place the TENS electrodes close to one another and as far from the device/lead system as possible. Monitor cardiac activity during TENS use.DRAFT
6 Warnings and PrecautionsRF Operating Frequencies. Nearby equipment emitting strong magnetic fields can interfere with RF communication, even if the other equipment com-plies with CISPR emission requirements. The operating characteristics are as fol-lows:MICS band: 402-405 MHz. The effective radiated power is below the limits as specified in:• Europe: EN ETSI 301 839-2• USA: FCC 47 CFR Part 95; 95.601-95.673 Subpart E, 95.1201-95.1219.• FCC ID: RIASJMRF.Radiofrequency ablation. RF ablation in a patient with a pulse generator may cause device malfunction or damage.Minimize RF ablation risks by:• Programming all tachyarrhythmia therapies off• Avoiding direct contact between the ablation catheter and the implanted lead or pulse generatorWARNINGThis transmitter is authorized by rule under the Medical Implant Communications Service (part 95 of the FCC Rules) and must not cause harmful interference to stations operating in the 400.150 - 406.000 MHz band in the Meteo-rological Aids (that is, transmitters and receivers used to communicate weather data), the Meteorological Satellite, or the Earth Exploration Satellite Services and must accept interference that may be caused by such aids, including interference that may cause undesired operation. This transmitter shall be used only in accordance with the FCC Rules governing the Medical Implant Commu-nications Service. Analog and digital voice communications are prohibited. Although this transmitter has been approved by the Federal Communications Commission, there is no guarantee that it will not receive inter-ference or that any particular transmission from this transmitter will be free from inter-ference. DRAFT
Current™ and Promote™ Devices User’s Manual 7• Positioning the groundplate so that the current pathway does not pass near the pulse generator system, i.e., place the groundplate under the patient’s buttocks or legs• Having external defibrillation equipment available.Home and Occupational EnvironmentsHigh-voltage power transmission lines. High-voltage power transmission lines may generate enough EMI to interfere with pulse generator operation if approached too closely.Communication equipment. Communication equipment such as microwave transmitters or high-power amateur transmitters may generate enough EMI to interfere with pulse generator operation if approached too closely.Home appliances. Home appliances in good working order and properly grounded do not usually produce enough EMI to interfere with pulse generator operation. There are reports of pulse generator disturbances caused by electric hand tools or electric razors used directly over the pulse generator implant site.Industrial equipment. A variety of industrial equipment produce EMI of suffi-cient field strength and modulation characteristics to interfere with proper operation of the pulse generator. These include, but are not limited to: arc welders; induction furnaces; very large or defective electric motors; and internal combustion engines with poorly shielded ignition systems.Electronic Article Surveillance (EAS)Advise patients that the Electronic Article Surveillance/Anti-theft (EAS) systems such as those at the point of sale and entrances/exits of stores, libraries, banks, etc., emit signals that may interact with the device. It is very unlikely that these systems will interact with their device significantly. However, to minimize the possibility of interaction, advise patients to simply walk through these areas at a normal pace and avoid lingering near or leaning on these systems.Metal DetectorsAdvise patients that metal detector security systems such as those found in air-ports and government buildings emit signals that may interact with ICDs and CRT-Ds. It is very unlikely that these systems will interact with their device sig-nificantly. To minimize the possibility of interaction, advise patients to simply walk through these areas at a normal pace and avoid lingering. Even so, the ICD and CRT-D systems contain metal that may set off the airport security sys-tem alarm. If the alarm does sound, the patient should present security person-nel with their patient identification card. If security personnel perform a search with a handheld wand, the patient should ask that they perform the search DRAFT
8 Warnings and Precautionsquickly, stressing that they should avoid holding the wand over the device for a prolonged period.Cellular PhonesThe pulse generator has been tested for compatibility with handheld wireless transmitters in accordance with the requirements of AAMI PC69. This testing covered the operating frequencies (450 MHz - 3 GHz) and pulsed modulation techniques of all of the digital cellular phone technologies in worldwide use today. Based on the results of this testing, the pulse generator should not be affected by the normal operation of cellular phones.Potential Adverse EventsPossible adverse events (in alphabetical order) associated with the system, include, but are not limited to the following:• Acceleration of arrhythmias (caused by device)• Air embolism• Allergic reaction• Bleeding• Cardiac tamponade• Chronic nerve damage• Death•Erosion• Exacerbation of heart failure• Excessive fibrotic tissue growth• Extracardiac stimulation (phrenic nerve, diaphragm, chest wall)• Extrusion• Fluid accumulation• Formation of hematomas or cysts• Inappropriate shocks•Infection• Keloid formation• Lead abrasion and discontinuity• Lead migration/dislodgment• Myocardial damage• Pneumothorax• Shunting current or insulating myocardium during defibrillation with inter-nal or external paddles• Potential mortality due to inability to defibrillate or pace• ThromboemboliDRAFT
Current™ and Promote™ Devices User’s Manual 9• Venous occlusion• Venous or cardiac perforation.Patients susceptible to frequent shocks despite antiarrhythmic medical manage-ment may develop psychological intolerance to an ICD or CRT-D system that may include the following:• Dependency• Depression• Fear of premature battery depletion• Fear of shocking while conscious• Fear of losing shock capability • Imagined shocking (phantom shock).Clinician Use InformationPhysician TrainingPhysicians should be familiar with sterile pulse generator implant procedure and with follow-up evaluation and management of patients with an ICD or CRT-D (or should refer the patient to such a physician).Directions for UsePulse generator operating characteristics should be verified at the time of implantation and recorded in the patient file. Complete the Patient Registration Form and return it to St. Jude Medical as it provides necessary information for warranty purposes and patient tracking.Copies of this user’s manual can be obtained by contacting your St. Jude Medi-cal representative.Maintaining Device EffectivenessDevice storageFOR SINGLE USE ONLY. Do not resterilize and re-implant explanted pulse gen-erators.St. Jude Medical has sterilized the pulse generator with ethylene oxide prior to shipment. Contact St. Jude Medical if resterilization is necessary.Do not implant the device when:• It has been dropped on a hard surface because this could have damaged pulse generator components.DRAFT
10 Clinician Use Information• The sterility indicator within the inner package is purple, because it might not have been sterilized.• Its storage package has been pierced or altered, because this could have rendered it non-sterile.• It has been stored or transported outside the environmental temperature limits.Storage limits: 10° to 45°C.Transportation limits: -20° to 60°C.An electrical reset condition may occur at temperatures below -20°C.After cold storage, allow the device to reach room temperature before charging the capacitors, programming, or implanting the device because cold temperature may affect initial device function.• Its “use before” date has expired, because this can adversely affect pulse generator longevity or device sterility.Do not resterilize the pulse generator using an autoclave, gamma-irradiation, organic cleaning agents (e.g., alcohol, acetone, etc.), or ultrasonic cleaners.Sterilization InstructionsContact St. Jude Medical if resterilization is necessary.Radiopaque IdentificationEach pulse generator has an x-ray absorptive marker for non-invasive identifica-tion. The two-letter model code is visible on a radiograph.Figure 3.  Current™ and Promote™ devices x-ray markerFigure 4.  Current RF and Promote RF devices x-ray markerDRAFT
Current™ and Promote™ Devices User’s Manual 11Package ContentsThe pulse generator is supplied in a sterile tray for introduction into the operat-ing field. The tray contains:• One pulse generator (with all tachyarrhythmia therapies off) with pre-installed setscrews• Torque driver.The outer box contains:• Literature.Technical ServiceSt. Jude Medical Cardiac Rhythm Management Division maintains 24-hour phone lines for technical questions and product support:• by phone for bradycardia or tachycardia devices- 1 818 364 1506 or- 1 800 722 3774 (toll-free within North America)• by fax for bradycardia devices- 1 818 362 7182 or- 1 800 756 7223 (toll-free within North America)• by fax for tachycardia devices- 1 408 522 6755 or- 1 866 739 0040 (toll-free within North America).For additional assistance, call your local St. Jude Medical representative.Additional InformationFor additional information on this device, see the Merlin PCS on-screen help.DRAFT
12 Additional InformationDRAFT
DRAFT
May 2007Art 60012839/001Cardiac Rhythm Management Division15900 Valley View CourtSylmar, CA 91342 USA+1 818 362 6822701 E. Evelyn AvenueSunnyvale, CA 94086 USA+1 408 738 4883Veddestavägen 19SE-175 84 JärfällaSweden+46 8 474 4000DRAFT

Navigation menu