Teledyne 200Eh Em Users Manual API 200EH/EM Operation
200EHEM to the manual 1fdf21b1-0289-439d-97c3-00b08cb545fa
2015-02-03
: Teledyne Teledyne-200Eh-Em-Users-Manual-464565 teledyne-200eh-em-users-manual-464565 teledyne pdf
Open the PDF directly: View PDF .
Page Count: 370
Download | |
Open PDF In Browser | View PDF |
INSTRUCTION MANUAL MODEL 200EH/EM NITROGEN OXIDES ANALYZER © TELEDYNE ADVANCED POLLUTION INSTRUMENTATION 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 USA Toll-free Phone: Phone: Fax: Email: Website: Copyright 2007-2010 Teledyne Advanced Pollution Instrumentation 800-324-5190 858-657-9800 858-657-9816 api-sales@teledyne.com http://www.teledyne-api.com/ 04521 Rev. C DCN 5731 14 May 2010 Teledyne API - Model 200EH/EM Operation Manual Safety Messages SAFETY MESSAGES Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol. These symbols are found in the manual and inside the instrument. The definition of these symbols is described below: GENERAL SAFETY HAZARD: Refer to the instructions for details on the specific hazard. CAUTION: Hot Surface Warning. CAUTION: Electrical Shock Hazard. TECHNICIAN SYMBOL: All operations marked with this symbol are to be performed by qualified maintenance personnel only. CAUTION The analyzer should only be used for the purpose and in the manner described in this manual. If you use the analyzer in a manner other than that for which it was intended, unpredictable behavior could ensue with possible hazardous consequences. NOTE Technical Assistance regarding the use and maintenance of the Model 200EH/EM NOx Analyzer or any other Teledyne Instruments product can be obtained by: Contacting Teledyne Instruments’ Customer Service Department at 800-324-5190 or Via the internet at http://www.teledyne-api.com/inquiries.asp i 04521C (DCN5731) This page intentionally left blank. ii 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Table of Contents TABLE OF CONTENTS 1. M200EH/EM Documentation.................................................................................................................................................... 1 1.1. Using This Manual............................................................................................................................................................ 1 2. Specifications, Approvals and Warranty................................................................................................................................... 3 2.1. M200EH/EM Operating Specifications ............................................................................................................................. 3 2.2. CE Mark Compliance ....................................................................................................................................................... 4 2.3. Warranty........................................................................................................................................................................... 4 3. Getting Started ......................................................................................................................................................................... 7 3.1. Unpacking and Initial Setup.............................................................................................................................................. 7 3.1.1. M200EH/EM Layout.................................................................................................................................................. 8 3.1.2. Electrical Connections ............................................................................................................................................ 10 3.1.2.1. Power Connection........................................................................................................................................... 10 3.1.2.2. Analog Output Connections ............................................................................................................................ 11 3.1.2.3. Connecting the Status Outputs ....................................................................................................................... 12 3.1.2.4. Connecting the Control Inputs......................................................................................................................... 13 3.1.2.5. Connecting the Serial Ports ............................................................................................................................ 14 3.1.2.6. Connecting to a LAN or the Internet................................................................................................................ 14 3.1.2.7. Connecting to a Multidrop Network ................................................................................................................. 14 3.1.3. Pneumatic Connections.......................................................................................................................................... 15 3.1.3.1. Calibration Gases ........................................................................................................................................... 16 3.1.3.2. Pneumatic Connections to M200EH/EM Basic Configuration: ........................................................................ 18 3.1.3.3. Connections with Internal Valve Options Installed .......................................................................................... 19 3.2. Initial Operation .............................................................................................................................................................. 20 3.2.1. Startup .................................................................................................................................................................... 20 3.2.2. Warm-Up ................................................................................................................................................................ 22 3.2.3. Warning Messages ................................................................................................................................................. 22 3.2.4. Functional Check .................................................................................................................................................... 24 3.3. Calibration ...................................................................................................................................................................... 25 3.3.1. Basic NOx Calibration Procedure............................................................................................................................ 25 3.3.2. Basic O2 Sensor Calibration Procedure.................................................................................................................. 28 3.3.2.1. O2 Calibration Setup ....................................................................................................................................... 28 3.3.2.2. O2 Calibration Method .................................................................................................................................... 28 3.3.3. Interferences for NOX Measurements ..................................................................................................................... 31 4. Frequently Asked Questions & Glossary................................................................................................................................ 33 4.1. Frequently Asked Questions .......................................................................................................................................... 33 4.2. Glossary ......................................................................................................................................................................... 34 5. Optional Hardware and Software ........................................................................................................................................... 37 5.1. External Pumps (OPT 10) .............................................................................................................................................. 37 5.2. Rack Mount Kits (OPTs 20-23)....................................................................................................................................... 37 5.3. Carrying Strap Handle (OPT 29) .................................................................................................................................... 38 5.4. Current Loop Analog Outputs (OPT 41) ......................................................................................................................... 39 5.4.1. Converting Current Loop Analog Outputs to Standard Voltage Outputs................................................................. 39 5.5. Particulate Filter Kit (OPT 42A) ...................................................................................................................................... 40 5.6. Ozone Supply Filter (OPT 49) ........................................................................................................................................ 40 5.7. Calibration Valve Options ............................................................................................................................................... 40 5.7.1. Zero/Span Valves (OPT 50) ................................................................................................................................... 40 5.7.2. Second Range span Valve (OPT 52)...................................................................................................................... 42 5.8. Oxygen Sensor (OPT 65) ............................................................................................................................................... 45 5.8.1. Theory of Operation................................................................................................................................................ 45 5.8.1.1. Paramagnetic measurement of O2 ..................................................................................................................45 5.8.1.2. Operation Within the M200EH/EM Analyzer ................................................................................................... 46 5.8.1.3. Pneumatic Operation of the O2 Sensor ........................................................................................................... 46 5.8.2. Zero Air Scrubber (OPT 64B) ................................................................................................................................. 48 5.8.3. Zero Air Scrubber Maintenance Kit (OPT 43) ......................................................................................................... 48 5.8.4. M200EH/EM Expendables Kit (OPT 42)................................................................................................................. 48 5.8.5. M200EH/EM Spare Parts Kit (OPT 43)................................................................................................................... 48 5.9. Communication Options ................................................................................................................................................. 48 5.9.1. RS232 Modem Cables (OPTs 60 and 60A)............................................................................................................ 48 5.9.2. RS-232 Multidrop (OPT 62) .................................................................................................................................... 49 5.9.3. Ethernet (OPT 63) .................................................................................................................................................. 49 5.10. Sample Gas Conditioners (OPTs 86 & 88)................................................................................................................... 50 iii 04521C (DCN5731) 5.11. Alarm Relay Option (OPT 67)....................................................................................................................................... 51 5.12. Special Software Features ........................................................................................................................................... 53 5.12.1. Maintenance Mode Switch.................................................................................................................................... 53 5.12.2. Second Language Switch ..................................................................................................................................... 53 5.12.3. Dilution Ratio Option............................................................................................................................................. 53 5.13. Additional Manual (OPT 70) ......................................................................................................................................... 53 5.14. Extended Warranty (OPTs 92 & 93) ............................................................................................................................. 54 6. Operating Instructions ............................................................................................................................................................ 55 6.1. Overview of Operating Modes ........................................................................................................................................ 55 6.2. Sample Mode ................................................................................................................................................................. 57 6.2.1. Test Functions ........................................................................................................................................................ 57 6.2.2. Warning Messages ................................................................................................................................................. 59 6.3. Calibration Mode ............................................................................................................................................................ 60 6.3.1. Calibration Functions .............................................................................................................................................. 60 6.4. SETUP MODE................................................................................................................................................................ 61 6.5. SETUP CFG: Viewing the Analyzer’s Configuration Information ............................................................................... 62 6.6. SETUP ACAL: Automatic Calibration......................................................................................................................... 62 6.7. SETUP DAS - Using the Data Acquisition System (iDAS) ........................................................................................ 63 6.7.1. iDAS Structure ........................................................................................................................................................ 64 6.7.1.1. iDAS Channels................................................................................................................................................ 64 6.7.1.2. iDAS Parameters ............................................................................................................................................ 65 6.7.1.3. iDAS Triggering Events................................................................................................................................... 65 6.7.2. Default iDAS Channels ........................................................................................................................................... 66 6.7.2.1. Viewing iDAS Data and Settings..................................................................................................................... 68 6.7.2.2. Editing iDAS Data Channels ........................................................................................................................... 69 6.7.2.3. Trigger Events................................................................................................................................................. 70 6.7.2.4. Editing iDAS Parameters ................................................................................................................................ 71 6.7.2.5. Sample Period and Report Period .................................................................................................................. 73 6.7.2.6. Number of Records......................................................................................................................................... 75 6.7.2.7. RS-232 Report Function ................................................................................................................................. 76 6.7.2.8. Compact Report.............................................................................................................................................. 76 6.7.2.9. Starting Date ................................................................................................................................................... 76 6.7.2.10. Disabling/Enabling Data Channels................................................................................................................ 77 6.7.2.11. HOLDOFF Feature ....................................................................................................................................... 77 6.7.3. Remote iDAS Configuration.................................................................................................................................... 78 6.8. SETUP RNGE: Range Units and Dilution Configuration............................................................................................ 80 6.8.1. Range Units............................................................................................................................................................ 80 6.8.2. Dilution Ratio .......................................................................................................................................................... 81 6.9. SETUP PASS: Password Feature ............................................................................................................................. 82 6.10. SETUP CLK: Setting the Internal Time-of-Day Clock .............................................................................................. 84 6.11. SETUP MORE COMM: Setting Up the Analyser’s Communication Ports ........................................................... 86 6.11.1. Analyzer ID ........................................................................................................................................................... 86 6.11.2. COM Port Default Settings ................................................................................................................................... 87 6.11.3. RS-232 COM Port Cable Connections ................................................................................................................. 87 6.11.4. RS-485 Configuration of COM2............................................................................................................................ 89 6.11.5. DTE and DCE Communication ............................................................................................................................. 90 6.11.6. Ethernet Card Configuration ................................................................................................................................. 91 6.11.6.1. Ethernet Card COM2 Communication Modes and Baud Rate ...................................................................... 91 6.11.6.2. Configuring the Ethernet Interface Option using DHCP ................................................................................ 91 6.11.6.3. Manually Configuring the Network IP Addresses .......................................................................................... 94 6.11.6.4. Changing the Analyzer’s HOSTNAME.......................................................................................................... 96 6.11.7. Multidrop RS-232 Set Up...................................................................................................................................... 97 6.11.8. COM Port Communication Modes ........................................................................................................................ 99 6.11.9. COM Port Baud Rate.......................................................................................................................................... 101 6.11.10. COM Port Testing ............................................................................................................................................. 102 6.12. SETUP MORE VARS: Internal Variables (VARS) ............................................................................................. 103 6.12.1. Setting the Gas Measurement Mode .................................................................................................................. 105 6.13. SETUP MORE DIAG: Diagnostics MENU ........................................................................................................ 106 6.13.1. Accessing the Diagnostic Features..................................................................................................................... 107 6.13.2. Signal I/O............................................................................................................................................................ 108 6.13.3. Analog Output Step Test .................................................................................................................................... 109 6.13.4. ANALOG OUTPUTS and Reporting Ranges...................................................................................................... 110 6.13.4.1. Analog Output Signals Available on the M200EH/EM................................................................................. 110 6.13.4.2. Physical Range versus Analog Output Reporting Ranges.......................................................................... 111 6.13.5. ANALOG I/O CONFIGURATION ........................................................................................................................ 113 iv 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Table of Contents 6.13.5.1. The Analog I/O Configuration Submenu. .................................................................................................... 113 6.13.5.2. Analog Output Signal Type and Range Selection ....................................................................................... 115 6.13.5.3. Turning the Analog Output Over-Range Feature ON/OFF.......................................................................... 116 6.13.5.4. Adding a Recorder Offset to an Analog Output........................................................................................... 117 6.13.5.5. Assigning an iDAS parameter to an Analog Output Channel...................................................................... 118 Reporting Gas Concentrations via the M200EH/EM Analog Output Channels ........................................118 6.13.5.6. Setting the Reporting Range Scale for an Analog Output........................................................................... 121 6.13.5.7. Setting Data Update Rate for an Analog Output ......................................................................................... 123 6.13.5.8. Turning an Analog Output On or Off ........................................................................................................... 124 6.13.6. ANALOG OUTPUT CALIBRATION .................................................................................................................... 125 6.13.6.1. Automatic Analog Output Calibration .......................................................................................................... 126 6.13.6.2. Manual Calibration of Analog Output configured for Voltage Ranges ......................................................... 127 6.13.6.3. Manual Calibration of Analog Outputs configured for Current Loop Ranges .............................................. 128 6.13.6.4. AIN Calibration............................................................................................................................................ 131 6.13.7. OTHER DIAG MENU FUNCTIONS .................................................................................................................... 132 6.13.7.1. Display Sequence Configuration................................................................................................................. 132 6.13.7.2. Optic Test ................................................................................................................................................... 135 6.13.7.3. Electrical Test ............................................................................................................................................. 136 6.13.7.4. Ozone Generator Override ......................................................................................................................... 137 6.13.7.5. Flow Calibration .......................................................................................................................................... 138 6.14. SETUP – ALRM: Using the optional Gas Concentration Alarms (OPT 67) ................................................................ 139 6.15. REMOTE OPERATION OF THE ANALYZER ............................................................................................................ 140 6.15.1. Remote Operation Using the External Digital I/O ............................................................................................... 140 6.15.1.1. Status Outputs ............................................................................................................................................ 140 6.15.1.2. Control Inputs.............................................................................................................................................. 141 6.15.2. Remote Operation Using the External Serial I/O ................................................................................................ 142 6.15.2.1. Terminal Operating Modes ......................................................................................................................... 142 6.15.2.2. Help Commands in Terminal Mode............................................................................................................. 143 6.15.2.3. Command Syntax ....................................................................................................................................... 143 6.15.2.4. Data Types.................................................................................................................................................. 144 6.15.2.5. Status Reporting ......................................................................................................................................... 145 6.15.2.6. Remote Access by Modem ......................................................................................................................... 145 6.15.2.7. COM Port Password Security ..................................................................................................................... 147 6.15.2.8. APICOM Remote Control Program ............................................................................................................. 148 6.15.3. Additional Communications Documentation ....................................................................................................... 148 6.15.4. Using the M200EH/EM with a Hessen Protocol Network.................................................................................... 149 6.15.4.1. General Overview of Hessen Protocol ........................................................................................................ 149 6.15.4.2. Hessen COMM Port Configuration.............................................................................................................. 149 6.15.4.3. Selecting a Hessen Protocol Type .............................................................................................................. 150 6.15.4.4. Setting The Hessen Protocol Response Mode ........................................................................................... 151 6.15.4.5. Hessen Protocol Gas ID ............................................................................................................................. 152 6.15.4.6. Setting Hessen Protocol Status Flags......................................................................................................... 153 6.15.4.7. Instrument ID Code..................................................................................................................................... 154 7. Calibration Procedures......................................................................................................................................................... 155 7.1. Calibration Preparations ............................................................................................................................................... 155 7.1.1. Required Equipment, Supplies, and Expendables................................................................................................ 155 7.1.2. Zero Air................................................................................................................................................................. 155 7.1.3. Span Calibration Gas Standards & Traceability.................................................................................................... 156 7.1.3.1. Traceability ................................................................................................................................................... 156 7.1.4. Data Recording Devices ....................................................................................................................................... 157 7.1.5. NO2 Conversion Efficiency ................................................................................................................................... 157 7.1.5.1. Determining / Updating the NO2 Converter Efficiency................................................................................... 157 7.2. Manual Calibration ....................................................................................................................................................... 160 7.3. Calibration Checks ....................................................................................................................................................... 163 7.4. Manual Calibration with Zero/Span Valves................................................................................................................... 164 7.5. Calibration Checks with Zero/Span Valves................................................................................................................... 167 7.6. Calibration With Remote Contact Closures .................................................................................................................. 168 7.7. Automatic Calibration (AutoCal) ................................................................................................................................... 169 7.8. Calibration Quality Analysis.......................................................................................................................................... 172 8. EPA Protocol Calibration...................................................................................................................................................... 173 9. Instrument Maintenance....................................................................................................................................................... 175 9.1. Maintenance Schedule ................................................................................................................................................. 175 9.2. Predictive Diagnostics .................................................................................................................................................. 177 9.3. Maintenance Procedures.............................................................................................................................................. 177 9.3.1. Changing the Sample Particulate Filter ................................................................................................................ 178 v 04521C (DCN5731) 9.3.2. Changing the O3 Dryer Particulate Filter............................................................................................................... 179 9.3.3. Maintaining the External Sample Pump................................................................................................................ 180 9.3.3.1. Rebuilding the Pump..................................................................................................................................... 180 9.3.3.2. Changing the Inline Exhaust Scrubber.......................................................................................................... 180 9.3.4. Changing the Pump and IZS Dust Filters ............................................................................................................. 180 9.3.5. Changing the External Zero Air Scrubber ............................................................................................................. 182 9.3.6. Changing the NO2 converter................................................................................................................................. 183 9.3.7. Cleaning the Reaction Cell ................................................................................................................................... 184 9.3.8. Changing Critical Flow Orifices............................................................................................................................. 185 9.3.9. Checking for Light Leaks ...................................................................................................................................... 186 10. Theory of Operation ........................................................................................................................................................... 189 10.1. Measurement Principle............................................................................................................................................... 189 10.1.1. Chemiluminescence ........................................................................................................................................... 189 10.1.2. NOX and NO2 Determination ............................................................................................................................... 190 10.2. Chemiluminescence Detection ................................................................................................................................... 191 10.2.1. The Photo Multiplier Tube................................................................................................................................... 191 10.2.2. Optical Filter ....................................................................................................................................................... 192 10.2.3. Auto Zero............................................................................................................................................................ 192 10.2.4. Measurement Interferences................................................................................................................................ 193 10.2.4.1. Direct Interference ...................................................................................................................................... 193 10.2.4.2. Third Body Quenching ................................................................................................................................ 193 10.2.4.3. Light Leaks.................................................................................................................................................. 194 10.3. Pneumatic Operation.................................................................................................................................................. 195 10.3.1. Pump and Exhaust Manifold............................................................................................................................... 195 10.3.2. Sample Gas Flow ............................................................................................................................................... 196 10.3.2.1. NO/NOx and AutoZero cycles ..................................................................................................................... 196 10.3.3. Flow Rate Control - Critical Flow Orifices ........................................................................................................... 197 10.3.3.1. Critical Flow Orifice ..................................................................................................................................... 199 10.3.4. Sample Particulate Filter..................................................................................................................................... 201 10.3.5. Ozone Gas Air Flow............................................................................................................................................ 201 10.3.6. O3 Generator...................................................................................................................................................... 202 10.3.7. Perma Pure® Dryer ............................................................................................................................................. 203 10.3.8. Ozone Supply Air Filter....................................................................................................................................... 205 10.3.9. Ozone Scrubber ................................................................................................................................................. 205 10.3.10. Pneumatic Sensors........................................................................................................................................... 206 10.3.10.1. Vacuum Manifold ...................................................................................................................................... 206 10.3.10.2. Sample Pressure Sensor .......................................................................................................................... 206 10.3.10.3. Vacuum Pressure Sensor ......................................................................................................................... 207 10.3.10.4. O3 Supply Air Flow Sensor........................................................................................................................ 207 10.3.11. Dilution Manifold ............................................................................................................................................... 207 10.4. Electronic Operation ................................................................................................................................................... 209 10.4.1. CPU .................................................................................................................................................................... 210 10.4.1.1. Disk On Chip............................................................................................................................................... 211 10.4.1.2. Flash Chip................................................................................................................................................... 211 10.4.2. Sensor Module, Reaction Cell ............................................................................................................................ 211 10.4.2.1. Reaction Cell Heating Circuit ...................................................................................................................... 211 10.4.3. Photo Multiplier Tube (PMT)............................................................................................................................... 212 10.4.4. PMT Cooling System. ......................................................................................................................................... 213 10.4.4.1. TEC Control Board...................................................................................................................................... 213 10.4.5. PMT Preamplifier ................................................................................................................................................ 214 10.4.6. Pneumatic Sensor Board.................................................................................................................................... 215 10.4.7. Relay Board........................................................................................................................................................ 215 10.4.7.1. Relay PCA Location and Layout ................................................................................................................. 215 10.4.7.2. Heater Control............................................................................................................................................. 215 10.4.7.3. Thermocouple Inputs and Configuration Jumper (JP5)............................................................................... 216 10.4.7.4. Valve Control .............................................................................................................................................. 217 10.4.8. Status LEDs & Watch Dog Circuitry.................................................................................................................... 218 10.4.8.1. Watchdog Indicator (D1) ............................................................................................................................. 218 10.4.9. Motherboard ....................................................................................................................................................... 219 10.4.9.1. A to D Conversion....................................................................................................................................... 219 10.4.9.2. Sensor Inputs.............................................................................................................................................. 219 10.4.9.3. Thermistor Interface.................................................................................................................................... 220 10.4.10. Analog Outputs ................................................................................................................................................. 220 10.4.11. External Digital I/O............................................................................................................................................ 221 10.4.12. I2C Data Bus ..................................................................................................................................................... 221 vi 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Table of Contents 10.4.13. Power-up Circuit ............................................................................................................................................... 221 10.5. Power Distribution &Circuit Breaker ........................................................................................................................... 222 10.6. Communications Interface.......................................................................................................................................... 223 10.6.1. Front Panel Interface .......................................................................................................................................... 224 10.6.1.1. Analyzer Status LED’s ................................................................................................................................ 224 10.6.1.2. Keyboard .................................................................................................................................................... 224 10.6.1.3. Display ........................................................................................................................................................ 225 10.6.1.4. Keyboard/Display Interface Electronics ...................................................................................................... 225 10.7. Software Operation .................................................................................................................................................... 227 10.7.1. Adaptive Filter..................................................................................................................................................... 228 10.7.2. Calibration - Slope and Offset............................................................................................................................. 228 10.7.3. Temperature/Pressure Compensation (TPC) ..................................................................................................... 229 10.7.4. NO2 Converter Efficiency Compensation ............................................................................................................ 230 10.7.5. Internal Data Acquisition System (iDAS) ............................................................................................................ 230 11. Troubleshooting & Repair .................................................................................................................................................. 231 11.1. General Troubleshooting ............................................................................................................................................ 231 11.1.1. Warning Messages ............................................................................................................................................. 232 11.1.2. Fault Diagnosis with Test Functions ................................................................................................................... 232 11.1.3. Using the Diagnostic Signal I/O Function ........................................................................................................... 233 11.1.4. Status LED’s ....................................................................................................................................................... 235 11.1.4.1. Motherboard Status Indicator (Watchdog) .................................................................................................. 235 11.1.4.2. CPU Status Indicator .................................................................................................................................. 235 11.1.4.3. Relay Board and Status LEDs .................................................................................................................... 235 11.2. Gas Flow Problems .................................................................................................................................................... 238 11.2.1. M200EH Internal Gas Flow Diagrams ................................................................................................................ 238 11.2.2. M200EM Internal Gas Flow Diagrams ................................................................................................................ 241 11.2.3. Zero or Low Flow Problems ................................................................................................................................ 243 11.2.3.1. Sample Flow is Zero or Low........................................................................................................................ 243 11.2.3.2. Ozone Flow is Zero or Low ......................................................................................................................... 244 11.2.4. High Flow............................................................................................................................................................ 245 11.2.5. Sample Flow is Zero or Low But Analyzer Reports Correct Flow ....................................................................... 245 11.3. Calibration Problems .................................................................................................................................................. 246 11.3.1. Negative Concentrations .................................................................................................................................... 246 11.3.2. No Response...................................................................................................................................................... 246 11.3.3. Unstable Zero and Span..................................................................................................................................... 247 11.3.4. Inability to Span - No SPAN Key ........................................................................................................................ 247 11.3.5. Inability to Zero - No ZERO Key ......................................................................................................................... 248 11.3.6. Non-Linear Response......................................................................................................................................... 248 11.3.7. Discrepancy Between Analog Output and Display ............................................................................................. 249 11.3.8. Discrepancy between NO and NOX slopes......................................................................................................... 249 11.4. Other Performance Problems..................................................................................................................................... 249 11.4.1. Excessive noise .................................................................................................................................................. 249 11.4.2. Slow Response................................................................................................................................................... 249 11.4.3. Auto-zero Warnings ............................................................................................................................................ 250 11.5. Subsystem Checkout.................................................................................................................................................. 251 11.5.1. Simple Vacuum Leak and Pump Check ............................................................................................................. 251 11.5.2. Detailed Pressure Leak Check ........................................................................................................................... 251 11.5.3. Performing a Sample Flow Check ...................................................................................................................... 252 11.5.4. AC Power Configuration ..................................................................................................................................... 253 11.5.4.1. AC configuration – Internal Pump (JP7)...................................................................................................... 254 11.5.4.2. AC Configuration – Standard Heaters (JP2) ............................................................................................... 255 11.5.4.3. AC Configuration –Heaters for Option Packages (JP6) .............................................................................. 256 11.5.5. DC Power Supply Test Points ............................................................................................................................ 257 11.5.6. I2C Bus ............................................................................................................................................................... 257 11.5.7. Keyboard / Display Interface............................................................................................................................... 258 11.5.8. Genreal Relay Board Diagnostic ........................................................................................................................ 258 11.5.9. Motherboard ....................................................................................................................................................... 259 11.5.9.1. A/D functions............................................................................................................................................... 259 11.5.9.2. Analog Output Voltages .............................................................................................................................. 259 11.5.9.3. Status Outputs ............................................................................................................................................ 260 11.5.9.4. Control Inputs.............................................................................................................................................. 261 11.5.10. CPU .................................................................................................................................................................. 261 11.5.11. RS-232 Communication.................................................................................................................................... 262 11.5.11.1. General RS-232 Troubleshooting ............................................................................................................. 262 11.5.11.2. Modem or Terminal Operation .................................................................................................................. 262 vii 04521C (DCN5731) 11.5.12. PMT Sensor...................................................................................................................................................... 263 11.5.13. PMT Preamplifier Board ................................................................................................................................... 263 11.5.14. High Voltage Power Supply .............................................................................................................................. 263 11.5.15. Pneumatic Sensor Assembly ............................................................................................................................ 264 11.5.15.1. Reaction Cell Pressure ............................................................................................................................. 264 11.5.15.2. Sample Pressure ...................................................................................................................................... 264 11.5.15.3. Ozone Flow............................................................................................................................................... 265 11.5.16. NO2 Converter .................................................................................................................................................. 265 11.5.17. O3 Generator .................................................................................................................................................... 266 11.5.18. Box Temperature .............................................................................................................................................. 266 11.5.19. PMT Temperature............................................................................................................................................. 266 11.6. Repair Procedures ..................................................................................................................................................... 267 11.6.1. Disk-on-Chip Replacement................................................................................................................................. 267 11.6.2. Flash Chip Replacement or Upgrade.................................................................................................................. 268 11.6.3. O3 Generator Replacement ................................................................................................................................ 268 11.6.4. Sample and Ozone Dryer Replacement ............................................................................................................. 269 11.6.5. PMT Sensor Hardware Calibration ..................................................................................................................... 270 11.6.6. Replacing the PMT, HVPS or TEC ..................................................................................................................... 271 11.7. Removing / Replacing the Relay PCA from the Instrument ........................................................................................ 274 11.8. Technical Assistance.................................................................................................................................................. 275 12. A Primer on Electro-Static Discharge................................................................................................................................. 277 12.1. How Static Charges are Created................................................................................................................................ 277 12.2. How Electro-Static Charges Cause Damage.............................................................................................................. 278 12.3. Common Myths About ESD Damage ......................................................................................................................... 279 12.4. Basic Principles of Static Control................................................................................................................................ 279 12.4.1. General Rules..................................................................................................................................................... 280 12.4.2. Basic anti-ESD Procedures for Analyzer Repair and Maintenance .................................................................... 281 12.4.2.1. Working at the Instrument Rack.................................................................................................................. 281 12.4.2.2. Working at an Anti-ESD Work Bench.......................................................................................................... 281 12.4.2.3. Transferring Components from Rack to Bench and Back ........................................................................... 282 12.4.2.4. Opening Shipments from Teledyne Instruments Customer Service............................................................ 282 12.4.2.5. Packing Components for Return to Teledyne Instruments Customer Service. ........................................... 283 LIST OF FIGURES Figure 3-1: Figure 3-2: Figure 3-3: Figure 3-4: Figure 3-5: Figure 3-6: Figure 3-7: Figure 3-8: Figure 3-9: Figure 3-10: Figure 3-11: Figure 3-12: Figure 3-13: Figure 3-14: Figure 5-1: Figure 5-2: Figure 5-3: Figure 5-4: Figure 5-5: Figure 5-6: Figure 5-7: Figure 5-8: Figure 5-9: Figure 5-10: M200EH/EM Layout.......................................................................................................................8 M200EH/EM Rear Panel Layout....................................................................................................9 M200EH/EM Front Panel Layout ...................................................................................................9 Analog Output Connector ............................................................................................................11 Status Output Connector .............................................................................................................12 Control Input Connector...............................................................................................................13 M200EH Internal Pneumatic Block Diagram - Standard Configuration.......................................15 M200EM Internal Pneumatic Block Diagram - Standard Configuration ......................................16 Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator.............................18 Pneumatic Connections–Basic Configuration–Using Bottled Span Gas.....................................18 Pneumatic Connections–With Zero/Span Valve Option (50) ......................................................19 Pneumatic Connections–With 2-Span point Option (52) –Using Bottled Span Gas ...................20 Front Panel Display During Startup Sequence............................................................................21 O2 Sensor Calibration Set Up ......................................................................................................28 M200EH/EM with Carrying Strap Handle and Rack Mount Brackets..........................................38 Current Loop Option Installed on the Motherboard .....................................................................39 M200EH – Internal Pneumatics with Zero-Span Valve Option 50...............................................41 M200EM – Internal Pneumatics with Zero-Span Valve Option 50 ..............................................41 M200EH – Internal Pneumatics with Second Span Point Valve Option 52.................................44 M200EM – Internal Pneumatics with Second Span Point Valve Option 52 ................................45 Oxygen Sensor - Principle of Operation ......................................................................................46 M200EH – Internal Pneumatics with O2 Sensor Option 65 .........................................................47 M200EM – Internal Pneumatics with O2 Sensor Option 65.........................................................47 M200EH/EM Multidrop Card........................................................................................................49 viii 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Figure 5-11: Figure 5-12: Figure 5-13: Figure 6-6-1: Figure 6-6-2: Figure 6-6-3: Figure 6-6-4: Figure 6-6-5: Figure 6-6-6: Figure 6-6-7: Figure 6-6-8: Figure 6-6-9: Figure 6-6-10: Figure 6-6-11: Figure 6-6-12: Figure 6-6-13: Figure 6-6-14: Figure 6-6-15: Figure 6-6-16: Figure 6-6-17: Table 6-29: Figure 6-6-18: Figure 6-6-19: Figure 6-6-20: Figure 7-1: Figure 7-2: Figure 7-3: Figure7-4: Figure 9-1: Figure 9-2: Figure 9-3: Figure 9-4: Figure 9-5: Figure 9-6: Figure 10-10-1: Figure 10-10-2: Figure 10-10-3: Figure 10-10-4: Figure 10-10-5: Figure 10-10-6: Figure 10-10-7: Figure 10-10-8: Figure 10-10-9: Figure 10-10-10: Figure 10-10-11: Figure 10-10-12: Figure 10-10-13: Figure 10-10-14: Figure 10-10-15: Figure 10-10-16: Figure 10-10-17: Figure 10-10-18: Figure 10-10-19: Figure 10-10-20: Figure 10-10-21: Figure 10-10-22: Figure 10-10-23: Figure 10-10-24: List of Figures M200EH/EM Ethernet Card .........................................................................................................50 M200EH/EM Rear Panel with Ethernet Installed.........................................................................50 Alarm Relay Output Pin Assignments..........................................................................................52 Front Panel Display......................................................................................................................55 Viewing M200EH/EM TEST Functions ........................................................................................58 Viewing and Clearing M200EH/EM WARNING Messages .........................................................59 APICOM Graphical User Interface for Configuring the iDAS ......................................................78 iDAS Configuration Through a Terminal Emulation Program......................................................79 Back Panel connector Pin-Outs for COM1 & COM2 in RS-232 mode. .......................................87 CPU connector Pin-Outs for COM1 & COM2 in RS-232 mode...................................................88 CPU card Locations of RS-232/486 Switches, Connectors and Jumpers...................................89 Back Panel connector Pin-Outs for COM2 in RS-485 mode.......................................................90 CPU connector Pin-Outs for COM2 in RS-485 mode..................................................................90 Location of JP2 on RS232-Multidrop PCA (option 62) ...............................................................97 RS232-Multidrop PCA Host/Analyzer Interconnect Diagram ......................................................98 Analog Output Connector Key .................................................................................................. 110 Setup for Calibrating Analog Outputs ....................................................................................... 127 Setup for Calibrating Current Outputs ...................................................................................... 129 Alternative Setup for Calibrating Current Outputs .................................................................... 129 Status Output ConnectorTable 6-29: Status Output Pin Assignments..................................... 140 Status Output Pin Assignments ................................................................................................ 141 Control Inputs with local 5 V power supply ............................................................................... 142 Control Inputs with external 5 V power supply ......................................................................... 142 APICOM Remote Control Program Interface ........................................................................... 148 Gas Supply Setup for Determination of NO2 Conversion Efficiency......................................... 157 Pneumatic Connections–With Zero/Span Valve Option (50) ................................................... 160 Pneumatic Connections–With 2-Span point Option (52) –Using Bottled Span Gas ................ 160 Pneumatic Connections–With Zero/Span Valve Option (50) ................................................... 164 Sample Particulate Filter Assembly .......................................................................................... 178 Particle Filter on O3 Supply Air Dryer ....................................................................................... 179 Zero Air Scrubber Assembly..................................................................................................... 182 NO2 Converter Assembly.......................................................................................................... 183 Reaction Cell Assembly............................................................................................................ 184 Critical Flow Orifice Assembly .................................................................................................. 186 M200EH/EM Sensitivity Spectrum............................................................................................ 190 NO2 Conversion Principle ......................................................................................................... 191 Reaction Cell with PMT Tube ................................................................................................... 192 Reaction Cell During the AutoZero Cycle................................................................................. 193 External Pump Pack ................................................................................................................. 195 Location of Gas Flow Control Assemblies for M200EH............................................................ 197 Location of Gas Flow Control Assemblies for M200EM ........................................................... 198 Location of Gas Flow Control Assemblies for M200EH with O2 sensor Option 65 .................. 198 Location of Gas Flow Control Assemblies for M200EH with Second Span Point Option 52 ... 199 Flow Control Assembly & Critical Flow Orifice ......................................................................... 200 Ozone Generator Principle ....................................................................................................... 202 Semi-Permeable Membrane Drying Process ........................................................................... 203 M200EH/EM Perma Pure® Dryer.............................................................................................. 204 Vacuum Manifold ...................................................................................................................... 206 Dilution Manifold ....................................................................................................................... 208 M200EH/EM Electronic Block Diagram .................................................................................... 209 M200EH/EM CPU Board Annotated......................................................................................... 210 PMT Housing Assembly ........................................................................................................... 212 Basic PMT Design .................................................................................................................... 212 PMT Cooling System ................................................................................................................ 213 PMT Preamp Block Diagram .................................................................................................... 214 Heater Control Loop Block Diagram. ........................................................................................ 215 Thermocouple Configuration Jumper (JP5) Pin-Outs............................................................... 216 Status LED Locations – Relay PCA......................................................................................... 218 ix 04521C (DCN5731) Figure 10-10-25: Power Distribution Block Diagram ........................................................................................... 222 Figure 10-10-26: Interface Block Diagram ........................................................................................................... 223 Figure 10-10-27: M200EH/EM Front Panel Layout .............................................................................................. 224 Figure 10-10-28: Keyboard and Display Interface Block Diagram ....................................................................... 225 Figure 10-10-29: Schematic of Basic Software Operation ................................................................................... 227 Figure 11-1: Viewing and Clearing Warning Messages ................................................................................ 232 Figure 11-2: Switching Signal I/O Functions ................................................................................................. 234 Figure 11-3: Motherboard Watchdog Status Indicator .................................................................................. 235 Figure 11-4: Relay Board PCA...................................................................................................................... 236 Figure 11-5: M200EH – Basic Internal Gas Flow.......................................................................................... 238 Figure 11-6: M200EH – Internal Gas Flow With OPT 50 .............................................................................. 239 Figure 11-7: M200EH – Internal Gas Flow With OPT 52 .............................................................................. 239 Figure 11-8: M200EH – Internal Gas Flow With OPT 65 .............................................................................. 240 Figure 11-9: M200EH – Internal Gas Flow With OPT 50 + OPT 65 ............................................................. 240 Figure 11-10: M200EM – Basic Internal Gas Flow ......................................................................................... 241 Figure 11-11: M200EM – Internal Gas Flow With OPT 50 ............................................................................. 241 Figure 11-12: M200EM – Internal Gas Flow With OPT 52 ............................................................................. 242 Figure 11-13: M200EM – Internal Gas Flow With OPT 65 ............................................................................. 242 Figure 11-14: M200EM – Internal Gas Flow With OPT 50 + OPT 65............................................................. 243 Figure 11-15: Location of AC power Configuration Jumpers .......................................................................... 253 Figure 11-16: Pump AC Power Jumpers (JP7)............................................................................................... 254 Figure 11-17: Typical Set Up of AC Heater Jumper Set (JP2) ....................................................................... 255 Figure 11-18: Typical Set Up of AC Heater Jumper Set (JP2) ....................................................................... 256 Figure 11-19: Typical Set Up of Status Output Test ....................................................................................... 260 Figure 11-20: Pressure / Flow Sensor Assembly............................................................................................ 264 Figure 11-22: Pre-Amplifier Board Layout....................................................................................................... 270 Figure 11-22: M200EH/EM Sensor Assembly ................................................................................................ 272 Figure 11-23: Relay PCA with AC Relay Retainer In Place............................................................................ 274 Figure 11-24: Relay PCA Mounting Screw Locations .................................................................................... 274 Figure 12-1: Triboelectric Charging ....................................................................................................................... 277 Figure 12-2: Basic anti-ESD Work Station ............................................................................................................ 280 LIST OF TABLES Table 2-1: Table 3-1: Table 3-2: Table 3-3: Table 3-4: Table 3-5: Table 3-6: Table 3-7: Table 3-8: Table 5-1: Table 5-2: Table 5-3: Table 5-4: Table 5-5: Table 5-6 Table 6-1: Table 6-2: Table 6-3: Table 6-4: Table 6-5: Model 200EH/EM Basic Unit Specifications ..................................................................................3 Analog Output Data Type Default Settings..................................................................................11 Analog Output Pin-Outs...............................................................................................................11 Status Output Signals ..................................................................................................................12 Control Input Signals ...................................................................................................................13 Inlet / Outlet Connector Nomenclature ........................................................................................15 NIST-SRM's Available for Traceability of NOx Calibration Gases ................................................17 Front Panel Display During System Warm-Up ............................................................................22 Possible Warning Messages at Start-Up .....................................................................................23 Zero/Span Valve States...............................................................................................................42 Two-Point Span Valve Operating States .....................................................................................43 Contents of Zero Air Scrubber Maintenance Kit ..........................................................................48 Dryer and NH3 Removal Options.................................................................................................51 Alarm Relay Output Assignments................................................................................................51 Concentration Alarm Relay Output Operation .............................................................................52 Analyzer Operating modes ..........................................................................................................56 Test Functions Defined................................................................................................................57 List of Warning Messages Revision F.0 ......................................................................................59 Primary Setup Mode Features and Functions .............................................................................61 Secondary Setup Mode Features and Functions ........................................................................61 x 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Table 6-6: Table 6-7: Table 6-8: Table 6-9: Table 6-10: Table 6-11: Table 6-12: Table 6-13: Table 6-14: Table 6-15: Table 6-16: Table 6-17: Table 6-18: Table 6-19: Table 6-20: Table 6-21: Table 6-22: Table 6-23: Table 6-24: Table 6-25: Table 6-26: Table 6-27: Table 6-28: Table 6-30: Table 6-31: Table 6-32: Table 6-33: Table 6-34: Table 6-28: Table 6-35: Table 6-36: Table 7-1: Table 7-2: Table 7-3: Table 7-4: Table 7-5: Table 9-1: Table 9-2: Table 10-1: Table 10-2: Table 10-3: Table 10-4: Table 10-5: Table 10-6: Table 11-1: Table 11-2: Table 11-3: Table 11-4: Table 11-5: Table 11-6: Table 11-7: Table 11-8: Table 11-9: Table 11-10: Table 11-11: Table 12-1: Table 12-2: List of Tables Front Panel LED Status Indicators for iDAS................................................................................63 iDAS Data Channel Properties ....................................................................................................64 iDAS Data Parameter Functions..................................................................................................65 M200EH/EM Default iDAS Configuration ....................................................................................67 Password Levels..........................................................................................................................82 Ethernet Status Indicators ...........................................................................................................91 LAN/Internet Configuration Properties.........................................................................................92 Internet Configuration Keypad Functions ....................................................................................96 COMM Port Communication modes ............................................................................................99 Variable Names (VARS) ........................................................................................................... 103 M200EH/EM Diagnostic (DIAG) Functions............................................................................... 106 Analog Output Voltage Ranges with Over-Range Active ......................................................... 110 Analog Output Pin Assignments ............................................................................................... 110 Analog Output Current Loop Range ......................................................................................... 111 Example of Analog Output configuration for M200EH/EM ....................................................... 111 DIAG - Analog I/O Functions .................................................................................................... 113 Analog Output Data Type Default Settings............................................................................... 118 Analog Output iDAS Parameters Related to Gas Concentration Data..................................... 119 Voltage Tolerances for Analog Output Calibration ................................................................... 127 Current Loop Output Calibration with Resistor ......................................................................... 130 M200EH/EM Available Concentration Display Values ............................................................. 132 M200EH/EM Concentration Display Default Values................................................................. 133 Concentration Alarm Default Settings....................................................................................... 139 Control Input Pin Assignments ................................................................................................. 141 Terminal Mode Software Commands ....................................................................................... 143 Command Types....................................................................................................................... 143 Serial Interface Documents ...................................................................................................... 148 RS-232 Communication Parameters for Hessen Protocol ....................................................... 149 M200EH/EM Hessen Protocol Response Modes..................................................................... 151 M200EH/EM Hessen GAS ID List ............................................................................................ 152 Default Hessen Status Bit Assignments ................................................................................... 153 NIST-SRM's Available for Traceability of NOx Calibration Gases ............................................. 156 AutoCal Modes ......................................................................................................................... 169 AutoCal Attribute Setup Parameters......................................................................................... 169 Example Auto-Cal Sequence.................................................................................................... 170 Calibration Data Quality Evaluation .......................................................................................... 172 M200EH/EM Preventive Maintenance Schedule...................................................................... 176 Predictive Uses for Test Functions ........................................................................................... 177 List of Interferents ..................................................................................................................... 194 M200EH/EM Valve Cycle Phases ............................................................................................ 196 M200EH/EM Critical Flow Orifice Diameters and Gas Flow Rates .......................................... 200 Thermocouple Configuration Jumper (JP5) Pin-Outs............................................................... 216 Typical Thermocouple Settings For M200E Series Analyzers ................................................. 217 Front Panel Status LED’s ......................................................................................................... 224 Test Functions - Possible Causes for Out-Of-Range Values ................................................... 233 Relay Board Status LEDs ......................................................................................................... 237 AC Power Configuration for Internal Pumps (JP7) ................................................................... 254 Power Configuration for Standard AC Heaters (JP2) ............................................................... 255 Power Configuration for Optional AC Heaters (JP6) ................................................................ 256 DC Power Test Point and Wiring Color Code........................................................................... 257 DC Power Supply Acceptable Levels ....................................................................................... 257 Relay Board Control Devices.................................................................................................... 258 Analog Output Test Function - Nominal Values ....................................................................... 259 Status Outputs Pin Assignments ............................................................................................. 260 Example of HVPS Power Supply Outputs ................................................................................ 263 Static Generation Voltages for Typical Activities ...................................................................... 277 Sensitivity of Electronic Devices to Damage by ESD ............................................................... 278 xi 04521C (DCN5731) LIST OF APPENDICES APPENDIX A - VERSION SPECIFIC SOFTWARE DOCUMENTATION APPENDIX A-1: M200EH/EM Software Menu Trees, Revision F.0 APPENDIX A-2: Setup Variables For Serial I/O, Revision F.0 APPENDIX A-3: Warnings and Test Functions, Revision F.0 APPENDIX A-4: M200EH/EM Signal I/O Definitions, Revision F.0 APPENDIX A-5: M200EH/EM iDAS Functions, Revision F.0 APPENDIX A-6: Terminal Command Designators, Revision F.0 APPENDIX B - M200EH/EM SPARE PARTS LIST APPENDIX C - REPAIR QUESTIONNAIRE - M200EH/EM APPENDIX D - ELECTRONIC SCHEMATICS xii 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual M200EH/EM Documentation 1. M200EH/EM DOCUMENTATION Thank you for purchasing the Model 200EH/EM Nitrogen Oxides Analyzer! The documentation (part number 04521) for this instrument is available in several different formats: Printed format, or; Electronic format on a CD-ROM. The electronic manual is in Adobe® Systems Inc. “Portable Document Format” (PDF). The Adobe® Acrobat Reader® software, which is necessary to view these files, can be downloaded for free from the internet at http://www.adobe.com/. The electronic version of the manual has many advantages: Keyword and phrase search feature Figures, tables and internet addresses are linked so that clicking on the item will display the associated feature or open the website. A list of chapters and sections as well as thumbnails of each page are displayed to the left of the text. Entries in the table of contents are linked to the corresponding locations in the manual. Ability to print sections (or all) of the manual Additional documentation for the Model 200EH/EM Nitrogen Oxides Analyzer is available from Teledyne Instruments’ website at http://www.teledyne-api.com/manuals/ APICOM software manual, part number 03945 Multi-drop manual, part number 02179 DAS manual, part number 02837. 1.1. USING THIS MANUAL This manual has the following data structures: 1.0 Table of Contents: Outlines the contents of the manual in the order the information is presented. This is a good overview of the topics covered in the manual. There is also a list of appendices, figures and tables. In the electronic version of the manual, clicking on a any of these table entries automatically views that section. 2.0 Specifications and Warranty A list of the analyzer’s performance specifications, a description of the conditions and configuration under which EPA equivalency was approved and Teledyne Instruments’ warranty statement. 3.0 Getting Started Concise instructions for setting up, installing and running your analyzer for the first time. 4.0 FAQ & Glossary: Answers to the most frequently asked questions about operating the analyzer and a glossary of acronyms and technical terms. 5.0 Optional Hardware & Software A description of optional equipment to add functionality to your analyzer. 1 04521C (DCN5731) M200EH/EM Documentation Teledyne API - Model 200EH/EM Operation Manual 6.0 Operation Instructions Step by step instructions for operating the analyzer. 7.0 Calibration Procedures General information and step by step instructions for calibrating your analyzer. 8.0 EPA Protocol Calibration Because there is no single, standard method for EPA equivalency in application where high concentrations of NOx are measured, no specific EPA calibration/validation method is included in this manual. 9.0 Instrument Maintenance Description of preventative maintenance procedures that should be regularly performed on you instrument to assure good operating condition. This includes information on using the iDAS to predict possible component failures before they happen. 10.0 Theory of Operation An in-depth look at the various principals by which your analyzer operates as well as a description of how the various electronic, mechanical and pneumatic components of the instrument work and interact with each other. A close reading of this section is invaluable for understanding the instrument’s operation. 11.0 Troubleshooting & Repair This section includes pointers and instructions for diagnosing problems with the instrument, such as excessive noise or drift, as well as instructions on performing repairs of the instrument’s major subsystems. 12.0 Electro-static Discharge Primer This section describes how static electricity occurs; why it is a significant concern and; how to avoid it and avoid allowing ESD to affect the reliable and accurate operation of your analyzer. Appendices For easier access and better updating, some information has been separated out of the manual and placed in a series of appendices at the end of this manual. These include version-specific software menu trees, warning messages, definitions of iDAS & serial I/O variables as well as spare part listings, repair questionnaire, interconnect drawing, detailed pneumatic and electronic schematics. NOTE Throughout this manual, words printed in capital, bold letters, such as SETUP or ENTR represent messages as they appear on the analyzer’s front panel display. NOTE The flowcharts in this manual contain typical representations of the analyzer’s display during the various operations being described. These representations are not intended to be exact and may differ slightly from the actual display of your instrument. USER NOTES: 2 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Specifications, Approvals and Warranty 2. SPECIFICATIONS, APPROVALS AND WARRANTY 2.1. M200EH/EM OPERATING SPECIFICATIONS Table 2-1: Model 200EH/EM Basic Unit Specifications Min/Max Range (Physical Analog Output) 200EH: Min: 0-5 ppm; Max: 0-5000 ppm 200EM: Min: 0-1 ppm; Max: 0-200 ppm Measurement Units ppm, mg/m3 (user selectable) Zero Noise <20 ppb (RMS) Span Noise <0.2% of reading above 20 ppm Lower Detectable Limit 40 ppb (2x noise as per USEPA) Zero Drift (24 hours) <20 ppb (at constant temperature and voltage.) Zero Drift (7 days) <20 ppb (at constant temperature and voltage.) Span Drift (7 Days) <1% of reading (at constant temperature and voltage.) Linearity 1% of full scale Precision 0.5% of reading Lag Time 20 s Rise/Fall Time Gas Flow Rates 95% in <60 s (~10 s in NO only or NOX only modes) 200EH: -40 cm³/min sample gas through NO2 converter & sensor module -250 cm3/min ± 10% though bypass manifold; -290 cm³/min total flow 200EM: -250 cm³/min sample gas through NO2 converter & sensor module O2 Sensor option adds 80 cm³/min to total flow though M200EH/EM when installed; Temperature Range 5 - 40 C operating range Humidity Range 0-95% RH non-condensing Dimensions H x W x D 18 cm x 43 cm x 61 cm (7" x 17" x 23.6") Weight, Analyzer 18 kg (40 lbs) without options Weight, Ext Pump Pack 7 kg (16 lbs) AC Power Rating 100 V, 50/60 Hz (3.25A); 115 V, 60 Hz (3.0 A); 220 - 240 V, 50/60 Hz (2.5 A) Power, Ext Pump 100 V, 50/60 Hz (3.25A); 115 V, 60 Hz (3.0 A); 220 - 240 V, 50/60 Hz (2.5 A) Environmental Installation category (over-voltage category) II; Pollution degree 2 Analog Outputs 4 user configurable outputs Analog Output Ranges All Outputs: 0.1 V, 1 V, 5 V or 10 V Three outputs convertible to 4-20 mA isolated current loop. All Ranges with 5% under/over-range Analog Output Resolution 1 part in 4096 of selected full-scale voltage (12 bit) Status Outputs 8 Status outputs from opto-isolators, 7 defined, 1 spare Control Inputs 6 Control inputs, 4 defined, 2 spare Alarm outputs 2 relay alarms outputs (Optional equipment) with user settable alarm limits - 1 Form C: SPDT; 3 Amp @ 125 VAC Serial I/O 1x RS-232; 1x RS-485 or RS-232 (configurable) Communication speed: 300 - 115200 baud (user selectable) Certifications CE: EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section 15.107 Class A, ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97) Class A. 3 04521C (DCN5731) Specifications, Approvals and Warranty Teledyne API - Model 200EH/EM Operation Manual 2.2. CE MARK COMPLIANCE Emissions Compliance The Teledyne-Advanced Pollution Instrumentation Nitrogen Oxides Analyzers M200EH/EM, M200EH/EMH and M200EH/EMM were tested and found to be fully compliant with: EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section 15.107 Class A, ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97) Class A. Tested on January 02-06, 2003 at CKC Laboratories, Inc., Report Number CE03-005. Safety Compliance The Teledyne-Advanced Pollution Instrumentation Nitrogen Oxides Analyzers M200EH/EM, M200EH/EMH and M200EH/EMM were tested and found to be fully compliant with: IEC 61010-1:90 + A1:92 + A2:95, Tested on January 27-20, 2003. 2.3. WARRANTY Warranty Policy (02024D) Prior to shipment, T-API equipment is thoroughly inspected and tested. Should equipment failure occur, T-API assures its customers that prompt service and support will be available. COVERAGE After the warranty period and throughout the equipment lifetime, T-API stands ready to provide on-site or in-plant service at reasonable rates similar to those of other manufacturers in the industry. All maintenance and the first level of field troubleshooting is to be performed by the customer. NON-API MANUFACTURED EQUIPMENT Equipment provided but not manufactured by T-API is warranted and will be repaired to the extent and according to the current terms and conditions of the respective equipment manufacturers warranty. GENERAL During the warranty period, T-API warrants each Product manufactured by T-API to be free from defects in material and workmanship under normal use and service. Expendable parts are excluded. If a Product fails to conform to its specifications within the warranty period, API shall correct such defect by, in API's discretion, repairing or replacing such defective Product or refunding the purchase price of such Product. The warranties set forth in this section shall be of no force or effect with respect to any Product: (i) that has been altered or subjected to misuse, negligence or accident, or (ii) that has been used in any manner other than in accordance with the instruction provided by T-API, or (iii) not properly maintained. 4 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Specifications, Approvals and Warranty THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES THEREFORE ARE EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESSED OR IMPLIED. THE REMEDIES SET FORTH IN THIS SECTION ARE THE EXCLUSIVE REMEDIES FOR BREACH OF ANY WARRANTY CONTAINED HEREIN. API SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF OR RELATED TO THIS AGREEMENT OF T-API'S PERFORMANCE HEREUNDER, WHETHER FOR BREACH OF WARRANTY OR OTHERWISE Terms and Conditions All units or components returned to Teledyne Instruments should be properly packed for handling and returned freight prepaid to the nearest designated Service Center. After the repair, the equipment will be returned, freight prepaid. USER NOTES: 5 04521C (DCN5731) 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Getting Started 3. GETTING STARTED 3.1. UNPACKING AND INITIAL SETUP CAUTION THE M200EH/EM WEIGHS ABOUT 17 KG (40 POUNDS) WITHOUT OPTIONS INSTALLED. TO AVOID PERSONAL INJURY, WE RECOMMEND TO USE TWO PERSONS TO LIFT AND CARRY THE ANALYZER. 1. Inspect the received packages for external shipping damage. If damaged, please advise the shipper first, then Teledyne Instruments. 2. Included with your analyzer is a printed record of the final performance characterization performed on your instrument at the factory. This record, titled Final Test and Validation Data Sheet (P/N 04413) is an important quality assurance and calibration record for this instrument. It should be placed in the quality records file for this instrument. 3. Carefully remove the top cover of the analyzer and check for internal shipping damage. Remove the set-screw located in the top, center of the Front panel. Remove the 2 screws fastening the top cover to the unit (one per side towards the rear). Slide the cover backwards until it clears the analyzer’s front bezel. Lift the cover straight up. NOTE Printed circuit assemblies (PCAs) are sensitive to electro-static discharges too small to be felt by the human nervous system. Failure to use ESD protection when working with electronic assemblies will void the instrument warranty. See Chapter 12 for more information on preventing ESD damage. CAUTION NEVER DISCONNECT ELECTRONIC CIRCUIT BOARDS, WIRING HARNESSES OR ELECTRONIC SUBASSEMBLIES WHILE THE UNIT IS UNDER POWER. 4. Inspect the interior of the instrument to make sure all circuit boards and other components are in good shape and properly seated. 5. Check the connectors of the various internal wiring harnesses and pneumatic hoses to make sure they are firmly and properly seated. 6. Verify that all of the optional hardware ordered with the unit has been installed. These are checked on the paperwork (Form 04490) accompanying the analyzer. 7 04521C (DCN5731) Getting Started Teledyne API - Model 200EH/EM Operation Manual VENTILATION CLEARANCE: Whether the analyzer is set up on a bench or installed into an instrument rack, be sure to leave sufficient ventilation clearance. AREA MINIMUM REQUIRED CLEARANCE Back of the instrument 10 cm / 4 inches Sides of the instrument 2.5 cm / 1 inch Above and below the instrument. 2.5 cm / 1 inch Various rack mount kits are available for this analyzer. See Chapter 5 of this manual for more information. 3.1.1. M200EH/EM LAYOUT Figure 3-1 shows a top-down view of the analyzer. The shown configuration includes the Ethernet board, IZS option, zero-air scrubber and an additional sample dryer. See Chapter 5 for optional equipment. Figure 3-2 shows the rear panel configuration with optional zero-air scrubber mounted to it and two optional fittings for the IZS option. Figure 3-3, finally shows the front panel layout of the analyzer. Figure 3-1: M200EH/EM Layout 8 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Getting Started AL1 ALARM OUT AL2 AL3 AL4 NO C NC NO C NC NO C NC NO C NC Figure 3-2: M200EH/EM Rear Panel Layout Figure 3-3: M200EH/EM Front Panel Layout 9 04521C (DCN5731) Getting Started Teledyne API - Model 200EH/EM Operation Manual 3.1.2. ELECTRICAL CONNECTIONS Refer to Figure 3-2 for the location of the rear panel electrical and pneumatic connections. 3.1.2.1. Power Connection Attach the power cord to the analyzer and plug it into a power outlet capable of carrying at least 10 A current at your AC voltage and that it is equipped with a functioning earth ground. CAUTION CHECK THE VOLTAGE AND FREQUENCY LABEL ON THE REAR PANEL OF THE INSTRUMENT (SEE FIGURE 3-2 FOR COMPATIBILITY WITH THE LOCAL POWER BEFORE PLUGGING THE M200EH/EM INTO LINE POWER. Do not plug in the power cord if the voltage or frequency is incorrect. CAUTION POWER CONNECTION MUST HAVE FUNCTIONING GROUND CONNECTION. DO NOT DEFEAT THE GROUND WIRE ON POWER PLUG. TURN OFF ANALYZER POWER BEFORE DISCONNECTING OR CONNECTING ELECTRICAL SUBASSEMBLIES. DO NOT OPERATE WITH COVER OFF. The M200EH/EM analyzer can be configured for both 100-130 V and 210-240 V at either 50 or 60 Hz., To avoid damage to your analyzer, make sure that the AC power voltage matches the voltage indicated on the rear panel serial number label and that the frequency is between 47 and 63 Hz. 10 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Getting Started 3.1.2.2. Analog Output Connections The M200H/EM is equipped with four analog output channels accessible through a connector on the back panel of the instrument. Each of these outputs may be set to reflect the value of any of the instrument’s iDAS data types. (see Table A-6 of M200EH/EM Appendix A – P/N 05147). The following table lists the default settings for each of these channels. To change these settings, see Sections 6.13.4 Table 3-1: PARAMETER DATA TYPE 1 Analog Output Data Type Default Settings CHANNEL DEFAULT SETTING A1 A2 NXCNC1 NOCNC1 3 A3 A4 N2CNC1 NXCNC2 2 RANGE 0 - 5 VDC REC OFS 0 mVDC AUTO CAL. ON CALIBRATED NO OUTPUT ON SCALE 100 ppm UPDATE 5 sec 1 See Table A-6 of M200EH/EM Appendix A for definitions of these iDAS data types 2 Optional current loop outputs are available for analog output channels A1-A3. 3 On analyzers with O2 sensor options installed, iDAS parameter O2CONC is assigned to output A4. To access these signals attach a strip chart recorder and/or data-logger to the appropriate contacts of the analog output connecter on the rear panel of the analyzer. A1 + - Figure 3-4: Table 3-2: ANALOG OUT A2 A3 + + - A4 + - Analog Output Connector Analog Output Pin-Outs PIN ANALOG OUTPUT VOLTAGE SIGNAL CURRENT SIGNAL 1 A1 V Out I Out + 2 3 A2 Ground I Out - V Out I Out + 4 Ground I Out - 5 V Out I Out + Ground I Out - V Out I Out + Ground I Out - A3 6 7 8 A4 11 04521C (DCN5731) Getting Started Teledyne API - Model 200EH/EM Operation Manual 3.1.2.3. Connecting the Status Outputs If you wish to utilize the analyzer’s status outputs to interface with a device that accepts logic-level digital inputs, such as programmable logic controller (PLC) chips, you can access them through a 12 pin connector on the analyzer’s rear panel labeled STATUS. Figure 3-5: 8 + D FOR PINS 1-8 7 EMITTER BUS 6 LOW SPAN 5 DIAGNOSTIC MODE 4 SPAN CAL 3 ZERO CAL 2 HIGH RANGE SYSTEM OK 1 CONC VALID STATUS Status Output Connector NOTE Most PLC’s have internal provisions for limiting the current the input will draw. When connecting to a unit that does not have this feature, external resistors must be used to limit the current through the individual transistor outputs to ≤50mA (120 Ω for 5V supply). Table 3-3: Status Output Signals PIN # STATUS 1 SYSTEM OK ON if no faults are present. CONDITION (ON = CONDUCTING) 2 CONC VALID ON if concentration measurement (NO, NO2 or NOx) is valid. OFF any time the hold-off feature is active. 3 HIGH RANGE ON if unit is in high range of the Auto Range Mode. 4 ZERO CAL ON whenever the instrument is in ZERO point calibration mode. 5 SPAN CAL ON whenever the instrument is in SPAN point calibration mode. 6 DIAG MODE 7 LOW SPAN CAL 8 Unused D EMITTER BUS ON whenever the instrument is in diagnostic mode. ON when in low span calibration (optional equipment necessary) The emitters of the transistors on pins 1-8 are tied together. Unused + DC POWER + 5 VDC, 300 mA (combined rating with Control Output, if used). Digital Ground The ground level from the analyzer’s internal DC power supplies 12 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Getting Started 3.1.2.4. Connecting the Control Inputs If you wish to use the analyzer to remotely activate the zero and span calibration modes, several digital control inputs are provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear panel. There are two methods for energizing the control inputs. The internal +5V available from the pin labeled “+” is the most convenient method. However, if full isolation is required, an external 5 VDC power supply should be used. CONTROL IN E F U + A ZERO CAL B C Figure 3-6: Table 3-4: STATUS DEFINITION E F U + 5 VDC Power Supply + External Power Connections Local Power Connections INPUT # D LOW SPAN D SPAN CAL C LOW SPAN B SPAN CAL ZERO CAL A CONTROL IN Control Input Connector Control Input Signals ON CONDITION A REMOTE ZERO CAL The analyzer is placed in Zero Calibration mode. The mode field of the display will read ZERO CAL R. B REMOTE SPAN CAL The analyzer is placed in Span Calibration mode. The mode field of the display will read SPAN CAL R. C REMOTE LO SPAN CAL The analyzer is placed in low span calibration mode as part of performing a low span (midpoint) calibration. The mode field of the display will read LO CAL R. D REMOTE RANGE HI The analyzer is placed into high range when configured for dual ranges.. E SPARE F SPARE U + Digital Ground The ground level from the analyzer’s internal DC power supplies (same as chassis ground). External Power input Input pin for +5 VDC required to activate pins A - F. 5 VDC output Internally generated 5V DC power. To activate inputs A - F, place a jumper between this pin and the “U” pin. The maximum amperage through this port is 300 mA (combined with the analog output supply, if used). 13 04521C (DCN5731) Getting Started Teledyne API - Model 200EH/EM Operation Manual 3.1.2.5. Connecting the Serial Ports If you wish to utilize either of the analyzer’s two serial interfaces, refer to Section 6.11 and 6.15 of this manual for instructions on configuration and usage. 3.1.2.6. Connecting to a LAN or the Internet If your unit has a Teledyne Instruments Ethernet card (Option 63), plug one end of a 7’ CAT5 cable into the appropriate place on the back of the analyzer (see Figure 5-11 in Section 5.9.3) and the other end into any nearby Ethernet access port. NOTE: The M200EH/EM firmware supports dynamic IP addressing or DHCP. If your network also supports DHCP, the analyzer will automatically configure its LAN connection appropriately, If your network does not support DHCP, see Section 6.11.6.3 for instructions on manually configuring the LAN connection. 3.1.2.7. Connecting to a Multidrop Network If your unit has a Teledyne Instruments RS-232 multidrop card (Option 62), see Section 6.11.7 for instructions on setting it up. CAUTION To prevent dust from getting into the analyzer, it was shipped with small plugs inserted into each of the pneumatic fittings on the rear panel. Make sure that all dust plugs are removed before attaching exhaust and supply gas lines. 14 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Getting Started 3.1.3. PNEUMATIC CONNECTIONS NOTE To prevent dust from getting into the analyzer, it was shipped with small plugs inserted into each of the pneumatic fittings on the rear panel. Make sure that all dust plugs are removed before attaching exhaust and supply gas lines. Table 3-5: REAR PANEL LABEL SAMPLE EXHAUST Inlet / Outlet Connector Nomenclature FUNCTION Connects the sample gas to the analyzer. When operating the analyzer without zero span option, this is also the inlet for any calibration gases. Connects the exhaust of the analyzer with the external vacuum pump. SPAN On Units with zero/span valve or IZS option installed, this port connects the external calibration gas to the analyzer. ZERO AIR On Units with zero/span valve or IZS option installed, this port connects the zero air gas or the zero air cartridge to the analyzer. Figure 3-7 and 3.8 show the internal pneumatic flow of the standard configuration of the M200EH and M200EM respectively. Figure 3-7: M200EH Internal Pneumatic Block Diagram - Standard Configuration 15 04521C (DCN5731) Getting Started Figure 3-8: Teledyne API - Model 200EH/EM Operation Manual M200EM Internal Pneumatic Block Diagram - Standard Configuration Note: Pneumatic Diagrams do not reflect the physical layout of the instrument. The most significant differences between the M200EH and M200 EM versions in regards to pneumatic flow are: A bypass line leading directly from the bypass manifold to the exhaust manifold is present on the M200EH, but not in the M200EM. The Diameter of the critical flow orifice controlling the gas flow into the sample chamber is smaller and therefore the flow rate of sample gas through the instrument is lower. See Section 10.3.2 for more information on these differences See Chapter 5 for or information on the pneumatic flow though instruments with one of the M200Eh/Em’available options installed. 3.1.3.1. Calibration Gases Zero Air Zero air or zero calibration gas is defined as a gas that is similar in chemical composition to the measured medium but without the gas to be measured by the analyzer, in this case NO and NO2. If your analyzer is equipped with an external zero air scrubber option, it is capable of creating zero air from ambient air. 16 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Getting Started If your application is not a measurement in ambient air, the zero calibration gas should be matched to the matrix of the measured medium. Pure nitrogen could be used as a zero gas for applications where NOX is measured in nitrogen. Special considerations apply if measuring NOX in a matrix that does not contain oxygen, see Section 10.3.11 for more information. span gas A gas specifically mixed to match the chemical composition of the type of gas being measured at near full scale of the desired measurement range. In this case, NOX, NO and NO2 measurements made with the M200EH/EM, it is recommended that you use a span gas with an NO concentration equal to 80% of the measurement range for your application. EXAMPLE: If the application is to measure between 0 ppm and 500 ppm, an appropriate span gas concentration would be 400 ppm NOx. Even though NO gas in nitrogen could be used as a span gas, the matrix of the balance gas is different and may cause interference problems or yield incorrect calibrations. The same applies to gases that contain high concentrations of other compounds (for example, CO2 or H2O). The span gas should match all concentrations of all gases of the measured medium as closely as possible. Cylinders of calibrated NO gas traceable to NIST-standard reference materials specifications (also referred to as EPA protocol calibration gases) are commercially available. Table 3-6: NIST-SRM's Available for Traceability of NOx Calibration Gases NIST-SRM4 TYPE NOMINAL CONCENTRATION 2627a 2628a 2629a Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 5 ppm 10 ppm 20 ppm 1683b 1684b 1685b 1686b 1687b Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 50 ppm 100 ppm 250 ppm 5000 ppm 1000 ppm 2630 2631a 2635 2636a Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 1500 ppm 3000 ppm 800 ppm 2000 ppm 2656 2660a Oxides of Nitrogen (NOx) in Air Oxides of Nitrogen (NOx) in Air 2500 ppm 100 ppm NOTE: If a dynamic dilution system such as the Teledyne Instruments model 700 is used to dilute high concentration gas standards to low, ambient concentrations, Make sure that: The NO concentration of the reference gas matches the dilution range of the calibrator. Choose a NO gas concentration that is in the middle of the dilution system’s range. 17 04521C (DCN5731) Getting Started Teledyne API - Model 200EH/EM Operation Manual 3.1.3.2. Pneumatic Connections to M200EH/EM Basic Configuration: Figures 3-7 and 3-8 show the most common configurations for gas supply and exhaust lines to the Model 200EH/EM Analyzer. Please refer to Figure 3-2 for the locations of pneumatic connections on the rear panel and Table 3-5 for nomenclature. NOTE Sample and calibration gases should only come into contact with PTFE (Teflon) or glass or materials. They should not come in contact with FEP or stainless steel materials. VENT here if input Source of MODEL 700 Gas Dilution Calibrator is pressurized SAMPLE GAS Removed during calibration NOx Gas (High Concentration) SAMPLE MODEL 701 Zero Gas Generator VENT EXHAUST MODEL 200EH/EM PUMP Figure 3-9: Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator MODEL 701 Zero Gas Generator 3-way Valve Source of SAMPLE GAS Removed during calibration VENT here if input is pressurized NOX Gas (High Concentration) SAMPLE EXHAUST Manual Control Valve VENT Figure 3-10: MODEL 200EH/EM PUMP Pneumatic Connections–Basic Configuration–Using Bottled Span Gas 18 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Getting Started 1. Attach a 1/4" exhaust line between the external pump exhaust port of the analyzer. 2. Attach an additional 1/4" exhaust port of the pump. CAUTION The exhaust from the analyzer needs to be vented outside the shelter or immediate area surrounding the instrument and conform to all safety requirements using a maximum of 10 meters of 1/4” PTFE tubing. 3. Attach a sample inlet line to the sample inlet port. Ideally, the pressure of the sample gas should be equal to ambient atmospheric pressure. NOTE Maximum pressure of any gas at the sample inlet should not exceed 1.5 in-Hg above ambient pressure and ideally should equal ambient atmospheric pressure. In applications where the sample gas is received from a pressurized manifold, a vent must be provided to equalize the sample gas with ambient atmospheric pressure before it enters the analyzer. The vented gas needs to be routed outside the immediate area or shelter surrounding the instrument. 4. Once the appropriate pneumatic connections have been made, check all pneumatic fittings for leaks using a procedure similar to that defined in Section 11.5.1. 3.1.3.3. Connections with Internal Valve Options Installed If your analyzer is equiped with either the zero/span valve option (50) or the 2-span point valve option(52), the pneumatic connections should be made as shown in figures 3-11 & 3-12: VENT here if input VENT at HIGH Span Concentration Calibrated NO MODEL 700 Gas Dilution Calibrator MODEL 701 Zero Gas Generator is pressurized Source of SAMPLE Gas PUMP Sample Exhaust Span Point External Zero Air Scrubber Figure 3-11: Filter Zero Air MODEL 200EH/EM Pneumatic Connections–With Zero/Span Valve Option (50) 19 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual On/Off Valves Source of SAMPLE Gas VENT at LOW Span Concentration VENT here if input is pressurized PUMP VENT Calibrated NO at HIGH Span Concentration Calibrated NO Getting Started Sample Exhaust High Span Point Low Span Point External Zero Air Scrubber Figure 3-12: Filter Zero Air MODEL 200EH/EM Pneumatic Connections–With 2-Span point Option (52) –Using Bottled Span Gas Once the appropriate pneumatic connections have been made, check all pneumatic fittings for leaks using the procedures defined in Section 11.5. 3.2. INITIAL OPERATION CAUTION! If the presence of ozone is detected at any time, call Teledyne API Customer Service as soon as possible: 800-324-5190 or email: api-customerservice@teledyne.com If you are unfamiliar with the theory of operation of the M200EH/EM analyzer, we recommend that you read Chapter 10 before proceeding. For information on navigating the analyzer’s software menus, see the menu trees described in Appendix A-1. 3.2.1. STARTUP After electrical and pneumatic connections are made, turn on the instrument and supply power to the external pump. The exhaust and PMT cooler fans should start. The display should immediately display a single, horizontal dash in the upper left corner of the display. This will last approximately 30 seconds while the CPU loads the operating system. Once the CPU has completed this activity, it will begin loading the analyzer firmware and configuration data. During this process, a string of messages will appear on the analyzer’s front panel display as shown in Figure 3-13. The analyzer should automatically switch to SAMPLE mode after completing the boot-up sequence and start monitoring NOX, NO, NO2 gases. 20 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual SELECT START OR REMOTE : System waits 3 seconds then automatically begins its initialization routine. No action required. 3 START . CHECKING FLASH STATUS : CHECKING FIRMWARE STATUS 1 : System is checking the format of the instrument’s flash memory chip. 1 STARTING INSTRUMENT CODE : 1 STARTING INSTRUMENT W/FLASH : 1 Getting Started System is checking the firmware stored in the instrument’s memory If at this point, **FLASH FORMAT INVALID** appears, contact Teledyne Instruments customer service The instrument is loading configuration and calibration data from the flash chip The instrument is loading the analyzer firmware. M100E NOX ANALYZER BOOT PROGRESS [XXXXX 50%_ _ _ _ _] SOFTWARE REVISION X.X BOOT PROGRESS [XXXXXXXX 75% _ _] SAMPLE TEST SYSTEM RESET CAL NOX=XXX.X CLR SETUP The revision level of the firmware installed in your analyzer is briefly displayed The startup process may hesitate at this point if the Ethernet option is installed, DHCP mode is turned on and the instrument is not connected to a functioning network. Firmware fully booted Press CLR to clear initial warning messages. (see Section 3.2.3) Figure 3-13: Front Panel Display During Startup Sequence 21 04521C (DCN5731) Getting Started Teledyne API - Model 200EH/EM Operation Manual 3.2.2. WARM-UP The M200EH/EM requires about 45 minutes warm-up time before reliable NOX, NO and NO2 measurements can be taken. During that time, various portions of the instrument’s front panel will behave as follows. Table 3-7: NAME Concentration Field Mode Field COLOR Front Panel Display During System Warm-Up BEHAVIOR SIGNIFICANCE N/A Switches between NOX, NO and NO2 This is normal operation. N/A Displays blinking “SAMPLE” Instrument is in sample mode but is still in the process of warming up (hold-off period is active). Sample Green On Unit is operating in sample mode, front panel display is continuously updated. Cal Yellow Off The instrument’s calibration is not enabled. Fault Red Blinking The analyzer is warming up and out of specification for a fault-free reading. STATUS LEDs See Figure 3-3 for locations 3.2.3. WARNING MESSAGES During warm-up, internal temperatures and other parameters may be outside of specified limits. The software will suppress most warning conditions for 30 minutes after power up. SAMPLE TEST HVPS WARNING CAL SAMPLE CLR SAMPLE MSG HVPS WARNING CAL MSG TEST deactivates warning messages SETUP RANGE=200.0 PPM < TST TST > CAL TEST MSG NOX = 0.0 NO = 0.0 CLR SETUP NOX = 0.0 CLR NOTE: If the warning message persists after several attempts to clear it, the message may indicate a real problem and not an artifact of the warm-up period SETUP MSG activates warning messages.keys replaced with TEST key Press CLR to clear the current message. If more than one warning is active, the next message will take its place Once the last warning has been cleared, the analyzer returns to SAMPLE mode 22 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Getting Started The following table includes a brief description of the various warning messages that may appear after the warm-up time. If warning messages persist after 30 minutes, investigate their cause using the troubleshooting guidelines in Chapter 11. To view and clear warning messages, use the following key strokes: Table 3-8: Possible Warning Messages at Start-Up MESSAGE DEFINITION ANALOG CAL WARNING The instrument’s A/D circuitry or one of its analog outputs is not calibrated. AZERO WRN XXX.X MV The reading taken during the auto-zero cycle is outside of specified limits. The value XXX.X indicates the auto-zero reading at the time of the warning. BOX TEMP WARNING The temperature inside the M200EH/EM chassis is outside the specified limits. CANNOT DYN SPAN Remote span calibration failed while the dynamic span feature was ON CANNOT DYN ZERO Remote zero calibration failed while the dynamic zero feature was ON. CONFIG INITIALIZED Configuration was reset to factory defaults or was erased. CONV TEMP WARNING DATA INITIALIZED FRONT PANEL WARN HVPS WARNING NO2 converter temperature is outside of specified limits. iDAS data and settings were erased. Firmware is unable to communicate with the front panel. High voltage power supply for the PMT is outside of specified limits. IZS TEMP WARNING On units with IZS options installed: The permeation tube temperature is outside of specified limits (if installed). MANIFOLD TEMP WARN Bypass manifold temperature is outside of warning limits. O2 CELL TEMP WARN O2 sensor cell temperature is outside of warning limits (if installed). OZONE FLOW WARNING OZONE GEN OFF Ozone flow is outside of specified limits. Ozone generator is off, which is intentional for the warm-up period. This is the only warning message that automatically clears itself after warm up. PMT TEMP WARNING PMT temperature is outside of specified limits. RCELL PRESS WARN Reaction cell pressure is outside of specified limits. RCELL TEMP WARNING Reaction cell temperature is outside of specified limits. REAR BOARD NOT DET The CPU is unable to communicate with the motherboard. RELAY BOARD WARN The firmware is unable to communicate with the relay board. SAMPLE FLOW WARN The flow rate of the sample gas is outside the specified limits. SYSTEM RESET This message appears every time the analyzer was powered up. 23 04521C (DCN5731) Getting Started Teledyne API - Model 200EH/EM Operation Manual 3.2.4. FUNCTIONAL CHECK After the analyzer’s components have warmed up for at least 30 minutes, verify that the software properly supports any hardware options that were installed. Check to make sure that the analyzer is functioning within allowable operating parameters. Appendix A and C include a list of test functions viewable from the analyzer’s front panel as well as their expected values. These functions are also useful tools for diagnosing performance problems with your analyzer (Chapter 11). The enclosed Final Test and Validation Data Sheet (part number 04490) lists these values before the instrument left the factory. To view the current values of these test functions press the keys: SAMPLE A1:NXCNC1=100 PPM < TST TST > CAL Toggle keys to scroll through list of functions 1 default settings for user selectable reporting range settings. 2 Only appears if O2 sensor option is installed. NOX = XXX SETUP A1:NXCNC1=100 PPM1 A2:N0CNC1=100 PPM1 A3:N2CNC1=25 PPM1 A4:NXCNC2=100%1 NOX STB SAMP FLOW 0ZONE FLOW PMT NORM PMT AZERO Refer to HVPS Section 6.2.1 RCELL TEMP for definitions BOX TEMP PMT TEMP of these test MF TEMP functions. O2 CELL TEMP2 MOLY TEMP RCEL SAMP NOX SLOPE NOX OFFSET NO SLOPE NO OFFSET O2 SLOPE2 O2 OFFSET2 TIME 24 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Getting Started 3.3. CALIBRATION 3.3.1. BASIC NOX CALIBRATION PROCEDURE The initial calibration should be carried out using the same reporting range set up as used during the analyzer’s factory calibration. This will allow you to compare your calibration results to the factory calibration as listed on the Final Test and Validation Data Sheet. The following procedure assumes that the instrument does not have any of the available zero/span (Z/S) or TwoPoint Span valve options installed. Chapter 7 contains instructions for calibrating instruments with these options. If both available iDAS parameters for a specific gas type are being reported via the instruments analog outputs e.g. NXCNC1 and NXCNC2, separate calibrations should be carried out for each parameter. Use the LOW button when calibrating for NXCNC1 Use the HIGH button when calibrating for NXCNC2. See Section 6.13.3 & 6.13.4 for more information on analog output reporting ranges STEP 1 - Set Units: To select the concentration units of measure press: SAMPLE < TST TST > SETUP X.X A1:NXCNC1=100PPM NOX=XXX.X CAL SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X UNIT Press this button to select the concentration units of measure: PPM RANGE CONTROL MENU DIL SETUP X.X PPB EXIT EXIT CONC UNITS: PPM ENTR EXIT PPM or MGM STEP 2 - Dilution Ratio: 25 04521C (DCN5731) Getting Started Teledyne API - Model 200EH/EM Operation Manual If the dilution ration option is enabled on your M200EH/EM and your application involves diluting the sample gas before it enters the analyzer, set the dilution ration as follows: SAMPLE < TST TST > A1:NXCNC1=100PPM CAL NOX=XXX.X SETUP SETUP X.X SETUP X.X UNIT PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE RANGE CONTROL MENU DIL EXIT EXIT SETUP X.X 0 0 IL FACTOR:1.0 Gain 0 0 .0 ENTR EXIT Toggle these buttons to select the dilution ratio factor EXIT ignores the new setting and returns to the previous display. ENTR accepts the new setting and returns to the previous display.. STEP 3 – Set NOx and NO span gas concentrations : Set the expected NO and NOx span gas concentration. These should be 80% of range of concentration values likely to be encountered in this application. The default factory setting is 100 ppm. If one of the configurable analog outputs is to be set to transmit concentration values, use 80% of the reporting range set for that output (see Section 6.13.4.5) If you supply NO span gas to the analyzer as well as NOx, the values for expected NO and NOx span gas concentrations need to be identical. SAMPLE A1:NXCNC1=100PPM < TST TST > SAMPLE NOX NOX=XXX.X CAL SETUP GAS TO CAL:NOX O2 ENTR EXIT SAMPLE RANGE TO CAL:LOW LOW HIGH M-P CAL ENTR EXIT A1:NXCNC1 =100PPM NOX=X.XXX ZERO SPAN CONC M-P CAL NOX CONCENTRATION MENU NO CONV M-P CAL 0 The NOX & NO span concentration values automatically default to 80.0 Conc. If this is not the the concentration of the span gas being used, toggle these buttons to set the correct concentration of the NO X and NO calibration gases. EXIT EXIT NOX SPAN CONC:80.0 Conc 0 8 0 .0 ENTR EXIT EXIT ignores the new setting and returns to the previous display. ENTR accepts the new setting and returns to the CONCENTRATION MENU. If using NO span gas in addition to NOX repeat last step. 26 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Getting Started STEP 4 – Zero/Span Calibration : To perform the zero/span calibration procedure: SAMPLE Analyzer continues to cycle through NOx, NO, and NO2 measurements throughout this procedure. A1:NXCNC1=100PPM < TST TST > NOX=XXX.X CAL SETUP Toggle TST> button until ... SAMPLE NOX STB= XXX.X PPM < TST TST > Set the Display to show the NOX STB test function. This function calculates the stability of the NO/NOx measurement NOX=XXX.X CAL SETUP Allow zero gas to enter the sample port at the rear of the analyzer. Wait until NOX STB falls below 0.5 ppm. This may take several minutes. SAMPLE NOX STB= XXX.X PPM < TST TST > SAMPLE NOX NOX=XXX.X CAL SETUP GAS TO CAL:NOX O2 ENTR EXIT SAMPLE RANGE TO CAL:LOW LOW HIGH M-P CAL ENTR EXIT NOX STB= XXX.X PPM M-P CAL NOX=XXX.X ZERO CONC NOX STB= XXX.X PPM ENTR EXIT NOX=X.XXX CONC EXIT Allow span gas to enter the sample port at the rear of the analyzer. Press ENTR to changes the OFFSET & SLOPE values for both the NO and NOx measurements. Press EXIT to leave the calibration unchanged and return to the previous menu. Wait until NOX STB falls below 0.5 ppm. This may take several minutes. SAMPLE NOX STB= XXX.X PPM < TST TST > SAMPLE NOX You may see both keys. If either the ZERO or SPAN buttons fail to appear see Section 11 for troubleshooting tips. SETUP GAS TO CAL:NOX O2 ENTR EXIT SAMPLE The SPAN key now appears during the transition from zero to span. CAL RANGE TO CAL:LOW LOW HIGH M-P CAL ENTR EXIT NOX STB= XXX.X PPM ZERO SPAN CONC M-P CAL NOX STB= XXX.X PPM ENTR M-P CAL NOX=XXX.X CONC NOX STB= XXX.X PPM ENTR CONC NOX=X.XXX EXIT NOX=X.XXX EXIT NOX=X.XXX EXIT Press ENTR to changes the OFFSET & SLOPE values for both the NO and NOx measurements. Press EXIT to leave the calibration unchanged and return to the previous menu. EXIT at this point returns to the SAMPLE menu. 27 04521C (DCN5731) Getting Started Teledyne API - Model 200EH/EM Operation Manual 3.3.2. BASIC O2 SENSOR CALIBRATION PROCEDURE If your instrument has an O2 sensor option installed that should be calibrated as well. 3.3.2.1. O2 Calibration Setup The pneumatic connections for calibrating are as follows: VENT here if input Source of is pressurized Removed during calibration at HIGH Span Concentration at 20.8% Span Concentration 3-way Valve Calibrated O2 Calibrated N2 SAMPLE GAS SAMPLE EXHAUST Manual Control Valve VENT Figure 3-14: MODEL 200EH/EM PUMP O2 Sensor Calibration Set Up O2 SENSOR ZERO GAS: Teledyne Instruments’ recommends using pure N2 when calibration the zero point of your O2 sensor option. O2 SENSOR SPAN GAS: Teledyne Instruments’ recommends using 21% O2 in N2 when calibration the span point of your O2 sensor option. 3.3.2.2. O2 Calibration Method STEP 1 – Set O2 span gas concentration : Set the expected O2 span gas concentration. This should be equal to the percent concentration of the O2 span gas of the selected reporting range (default factory setting = 20.8%; the approximate O2 content of ambient air). 28 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Getting Started STEP 2 – activate O2 sensor stability function To change the stability test function from NOx concentration to the O2 sensor output, press: SAMPLE A1:NXCNC1=100PPM < TST TST > SETUP X.X NOX=XXX.X CAL SETUP COMM 0) DAS_HOLD_OFF=15.0 Minutes JUMP EDIT PRNT EXIT PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X SETUP X.X EXIT Continue pressing NEXT until ... SECONDARY SETUP MENU VARS DIAG ALRM EXIT SETUP X.X 2) STABIL_GAS=NOX JUMP SETUP X.X 8 1 EDIT PRNT EXIT ENTER PASSWORD:818 8 ENTR EXIT SETUP X.X NO NO2 SETUP X.X NO Press EXIT 3 times to return to SAMPLE menu NO2 STABIL_GAS:NOX NOX O2 ENTR EXIT STABIL_GAS:O2 NOX O2 ENTR EXIT NOTE Use the same procedure to reset the STB test function to NOx when the O2 calibration procedure is complete. 29 04521C (DCN5731) Getting Started Teledyne API - Model 200EH/EM Operation Manual STEP 4 – O2 ZERO/SPAN CALIBRATION : To perform the zero/span calibration procedure: 30 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Getting Started The Model 200EH/EM analyzer is now ready for operation. NOTE Once you have completed the above set-up procedures, please fill out the quality questionnaire that was shipped with your unit and return it to Teledyne Instruments. This information is vital to our efforts in continuously improving our service and our products. Thank you. 3.3.3. INTERFERENCES FOR NOX MEASUREMENTS The chemiluminescence method for detecting NOX is subject to interference from a number of sources including water vapor (H2O), ammonia (NH3), sulfur dioxide (SO2) and carbon dioxide (CO2) but the Model 200EH/EM has been designed to reject most of these interferences. Section 10.2.4 contains more detailed information on interferences. Ammonia is the most common interferent, which is converted to NO in the analyzer’s NO2 converter and creates a NOX signal artifact. If the Model 200EH/EM is installed in an environment with high ammonia, steps should be taken to remove the interferent from the sample gas before it enters the reaction cell. Teledyne Instruments offers a sample gas conditioning option to remove ammonia and water vapor (Section 5.10). Carbon dioxide diminishes the NOX signal when present in high concentrations. If the analyzer is used in an application with excess CO2, contact Teledyne Instruments customer service for possible solutions. Excess water vapor can be removed with one of the dryer options described in Section 5.10 In ambient air applications, SO2 interference is usually negligible. Interferences are discussed in more detail in Section 10.2.4. USER NOTES: 31 04521C (DCN5731) 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Frequently Asked Questions & Glossary 4. FREQUENTLY ASKED QUESTIONS & GLOSSARY 4.1. FREQUENTLY ASKED QUESTIONS The following list contains some of the most commonly asked questions relating to the Model 200EH/EM NOx Analyzer. Q: Why is the ZERO or SPAN key not displayed during calibration? A: The M200EH/EM disables certain keys whenever the chosen value is out of range for that particular parameter. In this case, the expected span or zero value is too different from the actually measured value and the instrument does not allow to span or zero to that point. If, for example, the span set point is 80 ppm and the measurement response is only .5 ppm, the SPAN button will not appear to prevent the user from spanning to an out-of-range response curve. Chapter 11 describes this in detail. Q: Why does the ENTR key sometimes disappear on the front panel display? A: Sometimes the ENTR key will disappear if you select a setting that is invalid or out of the allowable range for that parameter, such as trying to set the 24-hour clock to 25:00:00 or a M200EH reporting range to less than more than 5000 ppm (200 ppm for a M200EM). Once you adjust the setting to an allowable value, the ENTR key will re-appear. Q: Why does the analyzer not respond to span gas? A: There are several reasons why this can happen. Section 11.3.2 has some possible answers to this question. Q: Can I automate the calibration of my analyzer? A: Any analyzer with zero/span valve options can be automatically calibrated using the instrument’s AutoCal feature. Q: What do I do if the concentration on the instrument's front panel display does not match the value recorded or displayed on my data logger even if both instruments are properly calibrated? A: This most commonly occurs for one of the following reasons: (1) a difference in circuit ground between the analyzer and the data logger or a wiring problem; (2) a scale problem with the input to the data logger. The analog outputs of the analyzer can be manually calibrated to compensate for either or both of these effects, see Section 6.13.4; analog outputs are not calibrated, which can happen after a firmware upgrade (Section 6.13.5). Q: How do I measure the sample flow? A: Sample flow is measured by attaching a calibrated flow meter to the sample inlet port when the instrument is operating. For the M200EH in its basic configuration, the sample flow should be 290 cm³/min 10%. For the M200Em in its basic configuration, the sample flow should be 250 cm³/min 10%. See Table 10-3 for more detailed information bout gas flow rates. Chapter 11 includes detailed instructions on performing a check of the sample gas flow. Q: How often do I need to change the particulate filter? A: Once per week. Table 9-1 contains a maintenance schedule listing the most important, regular maintenance tasks. Highly polluted sample air may require more frequent changes. Q: How long does the sample pump last? 33 04521C (DCN5731) Frequently Asked Questions & Glossary Teledyne API - Model 200EH/EM Operation Manual A: The sample pump should last one to two years and the pump head should be replaced when necessary. Use the RCEL pressure indicator on the front panel to see if the pump needs replacement. If this value goes above 10 in-Hg-A, on average, the pump head needs to be rebuild. Q: Do I need a strip chart recorder or external data logger? A: No, the M200EH/EM is equipped with a very powerful internal data acquisition system (iDAS). Section 6.7 describes the setup and operation in detail. Q: Why does my RS-232 serial connection not work? A: There are many possible reasons: 1) the wrong cable, please use the provided or a generic “straight-through” cable (do not use a “null-modem” type cable), 2) The DCE/DTE switch on the back of the analyzer is not set properly; make sure that both green and red lights are on, 3) the baud rate of the analyzer’s COM port does not match that of the serial port of your computer/data logger. See Section 11.5.11 more trouble-shooting information. 4.2. GLOSSARY Acronym – A short form or abbreviation for a longer term. Often artificially made up of the first letters of the phrase’s words. APICOM – Name of a remote control program offered by Teledyne-API to its customers ASSY - Acronym for Assembly. cm3 – metric abbreviation for cubic centimeter. Same as the obsolete abbreviation “cc”. Chemical formulas used in this document: CO2 – carbon dioxide H2O – water vapor HNO3 – nitric acid NOX – nitrogen oxides, here defined as the sum of NO and NO2 NO – nitric oxide NO2 – nitrogen dioxide NOy – nitrogen oxides, often called odd nitrogen, the sum of NO, NO2 (NOX) plus other compounds such as HNO3. Definitions vary widely and may include nitrate (NO3-), PAN, N2O and other compounds. NH3 – ammonia O2 - molecular oxygen O3 - ozone SO2 – sulfur dioxide DAS - Acronym for Data Acquisition System, the old acronym of iDAS DIAG - Acronym for diagnostics, the diagnostic menu or settings of the analyzer DHCP: acronym for dynamic host configuration protocol. A protocol used by LAN or Internet servers to automatically set up the interface protocols between themselves and any other addressable device connected to the network. 34 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Frequently Asked Questions & Glossary DOC – Acronym for Disk On Chip, the analyzer’s central storage area for analyzer operating system, firmware, and data. This is a solid state device without mechanical, moving parts that acts as a computer hard disk drive under DOS with disk drive label “C”. DOC chips come with 8 mb space in the E-series analyzer standard configuration but are available in larger sizes DOS - Disk Operating System. The E-series analyzers use DR DOS EEPROM - also referred to as a FLASH chip. FEP - Acronym for Fluorinated Ethylene Propylene polymer, one of the polymers that du Pont markets as Teflon® (along with PFA and PTFE). FLASH - flash memory is non-volatile, solid-state memory. I2C bus – read: I-square-C bus. A serial, clocked serial bus for communication between individual analyzer components IC – Acronym for Integrated Circuit, a modern, semi-conductor circuit that can contain many basic components such as resistors, transistors, capacitors etc in a miniaturized package used in electronic assemblies. iDAS - Acronym for Internal Data Acquisition System, previously referred to as DAS. LAN - acronym for local area network LED - Acronym for Light Emitting Diode. PCA - Acronym for Printed Circuit Assembly, this is the PCB with electronic components installed and ready to use PCB - Acronym for printed circuit board, the bare circuit board without components PLC – Acronym for programmable logic controller, a device that is used to control instruments based on a logic level signal coming from the analyzer PFA – Acronym for Per-Fluoro-Alkoxy, an inert polymer. One of the polymers that du Pont markets as Teflon® (along with FEP and PTFE). PTFE – Acronym for Poly-Tetra-Fluoro-Ethylene, a very inert polymer material used to handle gases that may react on other surfaces. One of the polymers that du Pont markets as Teflon® (along with FEP and PFA). PVC – Acronym for Poly Vinyl Chloride, a polymer used for downstream tubing in the M200EH/EM. RS-232 - An electronic communication protocol of a serial communications port RS-485 - An electronic communication protocol of a serial communications port TCP/IP - Acronym for Transfer Control Protocol / Internet Protocol, the standard communications protocol for Ethernet devices and the Internet VARS - Acronym for variables, the variables menu or settings of the analyzer USER NOTES: 35 04521C (DCN5731) 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Optional Hardware and Software 5. OPTIONAL HARDWARE AND SOFTWARE This section includes a descriptions of the hardware and software options available for the Model 200EH/EM Nitrogen Oxides Analyzer. For assistance with ordering these options please contact the sales department of Teledyne - Advanced Pollution Instruments at: TOLL-FREE: TEL: FAX: E-MAIL: WEB SITE: 800-324-5190 +1 858-657-9800 +1 858-657-9816 apisales@teledyne.com http://www.teledyne-api.com/ 5.1. EXTERNAL PUMPS (OPT 10) The M200EH/EM comes equipped with an external pump specified upon ordering. Whereas the analyzer can be re-configured for other voltages, operation at other than the original voltage/frequency may require a different external pump. A variety of external pumps are available for the M200EH/EM series analyzers. The range of available pump options meets all typical AC power supply standards while exhibiting the same pneumatic performance. TELEDYNE INSTRUMENTS PART NO. DESCRIPTION 009810300 External pump for 115 VAC / 60 Hz power supply 009810400 External pump for 230 VAC / 50 Hz power supply 009810500 External pump for 110 VAC / 50 Hz power supply 009810600 External pump for 100 VAC / 50 Hz power supply 009810700 External pump for 220-240 VAC / 50-60 Hz power supply 5.2. RACK MOUNT KITS (OPTS 20-23) There are several options for mounting the analyzer in standard 19” racks. The slides are three-part extensions, one mounts to the rack, one mounts to the analyzer chassis and the middle part remains on the rack slide when the analyzer is taken out. The analyzer locks into place when fully extended and cannot be pulled out without pushing two buttons, one on each side. The rack mount brackets for the analyzer requires that you have a support structure in your rack to support the weight of the analyzer. The brackets cannot carry the full weight of an analyzer and are meant only to fix the analyzer to the front of a rack and to prevent it from sliding out of the rack through user intervention or vibration. OPTION NUMBER OPT 20A OPT 20B OPT 21 OPT 23 DESCRIPTION Rack mount brackets with 26 in. chassis slides. Rack mount brackets with 24 in. chassis slides. Rack mount brackets only Rack mount for external pump (no slides). 37 04521C (DCN5731) Optional Hardware and Software Teledyne API - Model 200EH/EM Operation Manual 5.3. CARRYING STRAP HANDLE (OPT 29) The chassis of the M200EH/EM analyzer allows to attach a strap handle for carrying the instrument (Figure 5-1). The handle is located on the right side and pulls out to accommodate a hand for transport. When pushed in, the handle is nearly flush with the chassis, only protruding out about 9 mm (3/8”). Figure 5-1: M200EH/EM with Carrying Strap Handle and Rack Mount Brackets NOTE: Installing the strap handle prevents the use of the rack mount slides, although the rack mount brackets, Option 21, can still be used. CAUTION THE M200EH/EM WEIGHS ABOUT 17 KG (40 POUNDS) WITHOUT OPTIONS INSTALLED. TO AVOID PERSONAL INJURY: WE RECOMMEND TWO PERSONS LIFT AND CARRY THE ANALYZER. MAKE SURE TO DISCONNECT ALL CABLES AND TUBING FROM THE ANALYZER BEFORE CARRYING IT. 38 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Optional Hardware and Software 5.4. CURRENT LOOP ANALOG OUTPUTS (OPT 41) This option adds isolated, voltage-to-current conversion circuitry to the analyzer’s analog outputs. This option may be ordered separately for the first three of the analog outputs and can be installed at the factory or added later. Call Teledyne Instruments sales for pricing and availability. The current loop option can be configured for any output range between 0 and 20 mA (for example 0-20, 2-20 or 4-20 mA). Information on calibrating or adjusting these outputs can be found in Section 6.13.5.3. Analog Output A2 J19 J 23 Voltage Output Shunts installed Voltage Output Shunts installed Current Loop Option Installed on J21 (Analog Output A2) Figure 5-2: Current Loop Option Installed on the Motherboard 5.4.1. CONVERTING CURRENT LOOP ANALOG OUTPUTS TO STANDARD VOLTAGE OUTPUTS. NOTE Servicing or handling of circuit components requires electrostatic discharge protection, i.e. ESD grounding straps, mats and containers. Failure to use ESD protection when working with electronic assemblies will void the instrument warranty. See Chapter 12 for more information on preventing ESD damage. To convert an output configured for current loop operation to the standard 0 to 5 VDC output operation: 5. Turn off power to the analyzer. 6. If a recording device was connected to the output being modified, disconnect it. 7. Remove the top cover Remove the set screw located in the top, center of the rear panel Remove the screws fastening the top cover to the unit (four per side). Lift the cover straight up. 8. Disconnect the current loop option PCA from the appropriate connector on the motherboard. 39 04521C (DCN5731) Optional Hardware and Software Teledyne API - Model 200EH/EM Operation Manual 9. Place a shunt between the leftmost two pins of the connector (see Figure 5-1). 6 spare shunts (P/N CN0000132) were shipped with the instrument attached to JP1 on the back of the instruments keyboard and display PCA 10. Reattach the top case to the analyzer. 11. The analyzer is now ready to have a voltage-sensing, recording device attached to that output 5.5. PARTICULATE FILTER KIT (OPT 42A) This option includes a one-year supply of 50 replacement, Teflon membrane, particulate filters, 47 mm in diameter, 1 micrometer pore size. Operation of the particulate filter and weekly maintenance are mandatory on the M200EH/EM. 5.6. OZONE SUPPLY FILTER (OPT 49) This filter removes very fine particulates as well as nitric acid, sulfates, nitrates and other compounds from the ozone supply air stream. The filter is pneumatically connected between ozone generator and reaction cell. The corona discharge ozone generator produces small amounts of the above mentioned compounds, which may deposit on the reaction cell walls and – over time – cause drift and non-linear response. Having this filter in line minimizes these problems and maximizes the reaction cell cleaning intervals. The filter and its maintenance is shown in detail in Section9.3.3. Refer to Figure 3-1 for location of this optional filter. 5.7. CALIBRATION VALVE OPTIONS 5.7.1. ZERO/SPAN VALVES (OPT 50) The Model 200EH/EM NOx analyzer can be equipped with a zero/span valve option for controlling the flow of calibration gases generated from external sources. This option contains two solenoid valves located inside the analyzer that allow the user to switch either zero, span or sample gas to the instrument’s sensor. The user can control these valves from the front panel keyboard either manually or by activating the instrument’s CAL or AutoCal features (Section 7.7). The valves may also be opened and closed remotely through the serial ports (Section 6.11) or through the external, digital control inputs (Section 6.15). This option also includes a two-stage, external zero air scrubber assembly that removes all NO and NO2 from the zero air source (ambient air). The scrubber is filled with 50% Purafil Chemisorbant® (for conversion of NO to NO2) and 50% activated charcoal (for removal of NO2). This assembly also includes a small particle filter to prevent scrubber particles to enter the analyzer as well as two more rear panel fittings so each gas can enter the analyzer separately. Figure 5-3 & 5-4 shows the internal, pneumatic layouts the zero/span valve option installed for a Model 200EH and M200EM respectively. 40 04521C (DCN5731) Optional Hardware and Software EXHAUST MANIFOLD NOX Exhaust Scrubber 2-Stage NOX Scrubber O3 FLOW SENSOR Teledyne API - Model 200EH/EM Operation Manual Filter Figure 5-3: M200EH – Internal Pneumatics with Zero-Span Valve Option 50 Figure 5-4: M200EM – Internal Pneumatics with Zero-Span Valve Option 50 41 04521C (DCN5731) Optional Hardware and Software Teledyne API - Model 200EH/EM Operation Manual Table 5-1: MODE Zero/Span Valve States VALVE CONDITION Sample/Cal Open to sample gas inlet Zero/Span Open to zero air inlet ZERO CALIBRATION Sample/Cal Open to zero/span inlet (activated) Zero/Span Open to zero air inlet SPAN CALIBRATION Sample/Cal Open to zero/span inlet (activated) SAMPLE Zero/Span Open to span gas inlet / IZS gas (activated) The state of the zero/span valves can also be controlled: Manually from the analyzer’s front panel by using the SIGNAL I/O controls located under the DIAG Menu (Section 6.13.1), By activating the instrument’s AutoCal feature (Section 7.7), Remotely by using the external digital control inputs (Section 6.15.1.2) or Ethernet option. Remotely through the RS-232/485 serial I/O ports (Section 6-11). All supply lines should be vented outside of the analyzer’s enclosure. In order to prevent back-diffusion and pressure drop effects, these vent lines should be between 2 and 10 meters in length. 5.7.2. SECOND RANGE SPAN VALVE (OPT 52) This option includes a special set of valves that allows two separate NOx mixtures to enter the analyzer from two independent sources. Typically these two gas mixtures will come from two, separate, pressurized bottles of certified calibration gas: one mixed to produce a NO, NO2 or NOx concentration equal to the expected span calibration value for the application and the other mixed to produce a concentration at or near the midpoint of the intended measurement range. Individual gas inlets, labeled HIGH SPAN and LOW SPAN are provided at the back on the analyzer. The valves allow the user to switch between the two sources via keys on the front panel or from a remote location by way of either the analyzer’s digital control inputs or by sending commands over it’s serial I/O port(s). NOTE The analyzer’s software only allows the SLOPE and OFFSET to be calculated when sample is being routed through the HIGH SPAN inlet. The LOW SPAN gas is for midpoint reference checks only. The state of the optional valves can be controlled: Manually from the analyzer’s front panel by using the SIGNAL I/O controls located under the DIAG Menu (Section 6-13.1), By activating the instrument’s CAL or AutoCal features (Section 7.7), Remotely by using the external digital control inputs (Section 6.15.1.2) or Ethernet option. Remotely through the RS-232/485 serial I/O ports (Section 6.11). 42 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Optional Hardware and Software Table 5-2: Two-Point Span Valve Operating States MODE SAMPLE ZERO CAL HIGH SPAN CAL Low Span Check VALVE CONDITION Sample/Cal Open to SAMPLE inlet Zero Gas Valve Closed to ZERO AIR inlet High Span Valve Closed to HIGH SPAN inlet Low Span Valve Closed to LOW SPAN inlet Sample/Cal Closed to SAMPLE inlet Zero Gas Valve Open to ZERO AIR inlet High Span Valve Closed to HIGH SPAN inlet Low Span Valve Closed to LOW SPAN inlet Sample/Cal Closed to SAMPLE inlet Zero Gas Valve Closed to ZERO AIR inlet High Span Valve Open to HIGH SPAN inlet Low Span Valve Closed to LOW SPAN inlet Sample/Cal Closed to SAMPLE inlet Zero Gas Valve Closed to ZERO AIR inlet High Span Valve Closed to HIGH SPAN inlet Low Span Valve Open to LOW SPAN inlet 43 04521C (DCN5731) Optional Hardware and Software Figure 5-5: Teledyne API - Model 200EH/EM Operation Manual M200EH – Internal Pneumatics with Second Span Point Valve Option 52 44 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Optional Hardware and Software Orifice Dia. 0.003" HIGH SPAN AIR INLET LOW SPAN AIR INLET ZERO AIR INLET Span/Cal Valve NO/NOX VALVE BYPASS MANIFOLD VACUUM PRESSURE SENSOR NO2 Converter High Span Valve SAMPLE PRESSURE SENSOR AUTOZERO VALVE Low Span Valve O3 Purifier Zero Gas Valve Orifice Dia. 0.007" GAS INPUT MANIFOLD EXHAUST GAS OUTLET O3 GENERATOR O3 Scrubber REACTION CELL PMT EXHAUST MANIFOLD NOX Exhaust Scrubber O3 FLOW SENSOR SAMPLE GAS INLET FLOW PRESSURE SENSOR PCA Orifice Dia. 0.004" PUMP Filter Figure 5-6: PERMAPURE DRYER INSTRUMENT CHASSIS M200EM – Internal Pneumatics with Second Span Point Valve Option 52 5.8. OXYGEN SENSOR (OPT 65) 5.8.1. THEORY OF OPERATION 5.8.1.1. Paramagnetic measurement of O2 The oxygen sensor used in the M200EH/EM analyzer utilizes the fact that oxygen is attracted into strong magnetic field, most other gases are not, to obtain fast, accurate oxygen measurements. The sensor’s core is made up of two nitrogen filled glass spheres, which are mounted on a rotating suspension within a magnetic field (Figure 5-7). A mirror is mounted centrally on the suspension and light is shone onto the mirror that reflects the light onto a pair of photocells. The signal generated by the photocells is passed to a feedback loop, which outputs a current to a wire winding (in effect, a small DC electric motor) mounted on the suspended mirror. 45 04521C (DCN5731) Optional Hardware and Software Teledyne API - Model 200EH/EM Operation Manual Oxygen from the sample stream is attracted into the magnetic field displacing the nitrogen filled spheres and causing the suspended mirror to rotate. This changes the amount of light reflected onto the photocells and therefore the output levels of the photocells. The feedback loop increases the amount of current fed into the winding in order to move the mirror back into its original position. The more O2 present, the more the mirror moves and the more current is fed into the winding by the feedback control loop. A sensor measures the amount of current generated by the feedback control loop which is directly proportional to the concentration of oxygen within the sample gas mixture (see Figure 5-7). Figure 5-7: Oxygen Sensor - Principle of Operation 5.8.1.2. Operation Within the M200EH/EM Analyzer The oxygen sensor option is transparently integrated into the core analyzer operation. All functions can be viewed or accessed through the front panel, just like the functions for NOX. The O2 concentration is displayed in the upper right-hand corner, alternating with NOX, NO and NO2 concentrations. Test functions for O2 slope and offset are viewable from the front panel along with the analyzer’s other test functions. O2 sensor calibration is performed via the front panel CAL function and is performed in a nearly identical manner as the standard NOX/NO calibration. See Chapter 0 for more details. Stability of the O2 sensor can be viewed (see 3.3.2.1) The O2 concentration range is 0-100% (user selectable) with 0.1% precision and accuracy and is available to be output via one of the instrument’s four user selectable analog outputs (see Section 6.13.4). The temperature of the O2 sensor is maintained at a constant 50° C by means of a PID loop and can be viewed on the front panel as test function O2 TEMP. The O2 sensor assembly itself does not have any serviceable parts and is enclosed in an insulated canister. 5.8.1.3. Pneumatic Operation of the O2 Sensor Pneumatically, the O2 sensor is connected to the bypass manifold and draws a flow of about 80 cm³/min in addition to the normal sample flow rate (See Table 10.-3 for nominal sample inlet gas flow rates) and is separately controlled with its own critical flow orifice located inside the vacuum manifold. Figures 15-8 shows the internal pneumatics of the M200EH with the O2 Sensor installed. Figures 15-9 shows the internal pneumatics of the M200EM with the O2 Sensor installed. 46 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Optional Hardware and Software Figure 5-8: M200EH – Internal Pneumatics with O2 Sensor Option 65 Figure 5-9: M200EM – Internal Pneumatics with O2 Sensor Option 65 47 04521C (DCN5731) Optional Hardware and Software Teledyne API - Model 200EH/EM Operation Manual 5.8.2. ZERO AIR SCRUBBER (OPT 64B) This kit includes a zero air scrubber cartridge, which can be used to produce and supply zero air to the analyzer’s ZERO inlet port. The cartridge mounts to the outside rear panel by means of two rubberized clips and contains two chemicals, 50% volume of Purafil Chemisorbant to convert NO to NO2, followed 50% volume of charcoal to absorb NO2. The zero air scrubber exit contains a particle filter that retains any dust coming from the cartridge and connects with a 0.25” PVC tubing to the ZERO inlet port. The chemicals need to be exchanged periodically (use Option 43) to prevent saturation and break-through of NOX into the zero air stream. This kit is recommended if no other zero air source is available and if the analyzer is equipped with the zero/span valve option (Section 5.7.1). The kit is included in the IZS option but not in the zero/span valve option. 5.8.3. ZERO AIR SCRUBBER MAINTENANCE KIT (OPT 43) This kit includes the items needed to refurbish the external zero air scrubber. Table 5-3: Contents of Zero Air Scrubber Maintenance Kit TELEDYNE INSTRUMENTS PART DESCRIPTION NO. 005960000 Activated charcoal refill ® 059700000 Purafil Chemisorbant refill 1 FL0000001 Sintered filter for critical orifice port FL0000003 Replacement particulate filter for zero air inlet fitting 1 OR0000001 O-Ring (qty:2) for critical orifice port 1 These items are required for units with IZS option only. They are used for rebuilding the IZS-exhaust critical flow orifice on the analyzer’s exhaust manifold. 5.8.4. M200EH/EM EXPENDABLES KIT (OPT 42) This kit includes a recommended set of expendables for one year of operation of the M200EH/EM. See Appendix B for a detailed listing of the contents. 5.8.5. M200EH/EM SPARE PARTS KIT (OPT 43) This kit includes a recommended set of spare parts for 2-3 years of operation of the M200EH/EM. It includes items such as the orifice holder, a spare PMT and other items that are recommended as backups to minimize down-time in case of component failures. See Appendix B for a detailed listing of the contents. 5.9. COMMUNICATION OPTIONS 5.9.1. RS232 MODEM CABLES (OPTS 60 AND 60A) The analyzer can have come standard with a shielded, straight-through DB-9F to DB-9F cable of about 1.8 m length, which should fit most computers of recent build. This cable can be ordered as Option 60. Option 60A consists of a shielded, straight-through serial cable of about 1.8 m length to connect the analyzer’s COM1 port to a computer, a code activated switch or any other communications device that is equipped with a DB-25 female connector. The cable is terminated with one DB-9 female connector and one DB-25 male connector. The DB-9 connector fits the analyzer’s COM1 port. 48 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Optional Hardware and Software 5.9.2. RS-232 MULTIDROP (OPT 62) The multidrop option is used with any of the RS-232 serial ports to enable communications of up to eight analyzers with the host computer over a chain of RS-232 cables via the instruments COM1 Port. It is subject to the distance limitations of the RS 232 standard. The option consists of a small printed circuit assembly, which is plugs into to the analyzer’s CPU card (see Figure 5-10) and is connected to the RS-232 and COM2 DB9 connectors on the instrument’s back panel via a cable to the motherboard. One option 62 is required for each analyzer along with one 6’ straight-through, DB9 male DB9 Female cable (P/N WR0000101). This option can be installed in conjunction with the Ethernet option (Option 63) allowing the instrument to communicate on both types of networks simultaneously. For more information on using and setting up this option see Section 6.11.7) Rear Panel CPU Card (as seen from inside) Multidrop Card Figure 5-10: M200EH/EM Multidrop Card 5.9.3. ETHERNET (OPT 63) The Ethernet option allows the analyzer to be connected to any Ethernet local area network (LAN) running TCP/IP. The local area network must have routers capable of operating at 10BaseT. If Internet access is available through the LAN, this option also allows communication with the instrument over the public Internet. When installed, this option is electronically connected to the instrument’s COM2 serial port making that port no longer available for RS-232/RS-485 communications through the COM2 connector on the rear panel. The option consists of a Teledyne Instruments designed Ethernet card (see Figure 5-11), which is mechanically attached to the instrument’s rear panel (see Figure 5-12). A 7-foot long CAT-5 network cable, terminated at both ends with standard RJ-45 connectors, is included as well. Maximum communication speed is limited by the RS232 port to 115.2 kBaud. 49 04521C (DCN5731) Optional Hardware and Software Teledyne API - Model 200EH/EM Operation Manual Figure 5-11: Ethernet Card M200EH/EM Ethernet Card CPU Card Rear Panel (as seen from inside) Female RJ-45 Connector LNK LED ACT LED TxD LED RxD LED RE-232 Connector To Motherboard Interior View Figure 5-12: Exterior View M200EH/EM Rear Panel with Ethernet Installed This option can be installed in conjunction with the RS-2323 multidrop (option 62) allowing the instrument to communicate on both types of networks simultaneously. For more information on using and setting up this option see Section 6.11.6) 5.10. SAMPLE GAS CONDITIONERS (OPTS 86 & 88) Several permeation devices using Nafion® permeation gas exchange tubes are available for applications with high moisture and/or moderate levels of NH3 in the sample gas. This type of sample conditioner is part of the standard M200EH/EM equipment to remove H2O and NH3 from the ozone generator supply gas stream but can be purchased for the sample gas stream as well. All gas conditioners remove water vapor to a dew point of about –20° C (~600 ppm H2O) and effectively remove concentrations of ammonia up to about 1 ppm. More information about these dryers and their performance is available at http://www.permapure.com/. The following options include the hardware required to install the dryers. 50 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Table 5-4: Optional Hardware and Software Dryer and NH3 Removal Options OPTION NUMBER DESCRIPTION STANDARD EQUIPMENT Single gas conditioner (dryer / NH3 removal) for ozone generator supply gas stream only. Includes mounting bracket for two dryers (Option 86 mounts on the back). OPT 86 Single gas conditioner (dryer / NH3 removal) for sample gas stream only. Mounts on the back of the existing dryer bracket. Converts analyzer to dual-conditioner instrument. OPT 88 Single combination gas conditioner (dryer / NH3 removal) for both the sample gas and ozone supply air. Replaces the standard dryer for O3 air and comes with mounting bracket. The combination conditioner is a low-cost option for drying both the sample gas and ozone supply air with one dryer. However, this dryer can only be used in applications where both sample and calibration gases (after dilution) are at or near ambient and constant concentrations of oxygen (about 21%), because the ozone generator needs a high and constant amount of oxygen to generate ozone properly. Stack applications or industrial applications in which the sample gas has a significantly reduced or highly variable concentration of oxygen need to use the separate dryer option 86. The combination conditioner needs to be specified upon ordering the analyzer. 5.11. ALARM RELAY OPTION (OPT 67) The M200EH/EM can be equipped with a set of 2 concentration alarms. Each alarm can be independently enabled or disabled as well as programmed with its own, individual alarm limit point (see Section 6.14 for details on programming the alarms). The status of each alarm is available via a set of alarm relay outputs located on the lower right hand corner of the analyzer’s rear panel (see Figure 3-2). While there are four relay outputs on the back of the analyzer, only Two of the outputs correspond to the instrument’s two concentration alarms. Table 5-5: RELAY NAME ASSIGNED ALARM 1 Alarm Relay Output Assignments AL1 AL2 AL3 AL4 ST_SYSTEM_OK21 CONCENTRATION ALARM 1 CONCENTRATION ALARM 2 SPARE ST_SYSTEM OK2 is a second system OK status alam available on some analyzers. 51 04521C (DCN5731) Optional Hardware and Software Teledyne API - Model 200EH/EM Operation Manual ALARM OUT AL2 AL3 AL1 AL4 NO C NC NO C NC NO C NC NO C NC ST_SYSTEM_OK2 (Optional Alert) CONCENTRATION ALARM 1 Figure 5-13: CONCENTRATION ALARM 2 SPARE Alarm Relay Output Pin Assignments Each of the two concentration relay outputs has 3-pin connections that allow the relay to be connected for either normally open or normally closed operation. Table 5-6 describes how to connect the alarm relays. Table 5-6 RELAY FUNCTION Concentration Alarm 1 AL2 Active N O C N C COMMENTS Gas concentration level is above the trigger limit set for CONC_ALARM_1 iDAS Trigger CONCW1 ACTIVATED Inactive Gas concentration level is below the trigger limit set for CONC_ALARM_1 Concentration Alarm 2 Gas concentration level is above the trigger limit set for CONC_ALARM_2 Active iDAS Trigger CONCW2 ACTIVATED CONC ALARM2 WARN appears on Analyzer Display Concentration Alarm 2 Inactive 1 RELAY PIN 1 STATE CONC ALARM1 WARN appears on Analyzer Display Concentration Alarm 1 AL3 Concentration Alarm Relay Output Operation Gas concentration level is below the trigger limit set for CONC_ALARM_2 NO = Normally Open operation. C = Common NC = Normally Closed operation. 52 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Optional Hardware and Software 5.12. SPECIAL SOFTWARE FEATURES 5.12.1. MAINTENANCE MODE SWITCH Teledyne Instruments analyzers are equipped with a switch that places the instrument in maintenance mode. When present, the switch is accessed by opening the hinged front panel and is located on the rearward facing side of the display/keyboard driver PCA; on the left side; near the particulate filter. When in maintenance mode the instrument ignores all commands received via the COMM ports that alter the operation state of the instrument This includes all calibration commands, diagnostic menu commands and the reset instrument command. The instrument continues to measure concentration and send data when requested. This feature is of particular use for instruments connected to multidrop or Hessen protocol networks. 5.12.2. SECOND LANGUAGE SWITCH Teledyne Instruments analyzers are equipped with a switch that activates an alternate set of display messages in a language other than the instrument’s default language. This switch is accessed by opening the hinged front panel and is located on the rearward facing side of the display/keyboard driver PCA; on the right side. To activate this feature, the instrument must also have a specially programmed Disk on Chip containing the second language. Contact Teledyne Instruments Customer Service personnel for more information. 5.12.3. DILUTION RATIO OPTION The dilution ration feature is a software option that is designed for applications where the sample gas is diluted before being analyzed by the Model 200EH/EM. Typically this occurs in continuous emission monitoring (CEM) applications where the quality of gas in a smoke stack is being tested and the sampling method used to remove the gas from the stack dilutes the gas. Once the degree of dilution is known, this feature allows the user to add an appropriate scaling factor to the analyzer’s NO, NO2 and NOx concentration calculations so that the measurement range and concentration values displayed on the instrument’s front panel display and reported via the instruments various outputs reflect the undiluted values. Contact Teledyne Instruments Customer Service personnel for information on activating this feature. Instructions for using the dilution ratio option can be found in Section 6.8.1. 5.13. ADDITIONAL MANUAL (OPT 70) Additional copies of the printed user’s manual can be purchased from the factory. Please specify the serial number of your analyzer so that we can match the manual version. This operators manual is also available on CD. The electronic document is stored in Adobe Systems Inc. Portable Document Format (PDF) and is viewable with Adobe Acrobat Reader® software, which can be downloaded for free at http://www.adobe.com/ The electronic version of this manual can also be downloaded for free at http://www.teledyne-api.com/manuals/. Note that the online version is optimized for fast download and may not print with the same quality as the manual on CD. 53 04521C (DCN5731) Optional Hardware and Software Teledyne API - Model 200EH/EM Operation Manual 5.14. EXTENDED WARRANTY (OPTS 92 & 93) Two options are available for extending the standard manufacturer’s warranty (Section 2.2). Both options have to be specified upon ordering the analyzer. OPTION NUMBER DESCRIPTION OPT 92 Extends warranty to cover a two (2) year period from the date of purchase. OPT 93 Extends warranty to cover a five (5) year period from the date of purchase. USER NOTES: 54 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6. OPERATING INSTRUCTIONS To assist in navigating the analyzer’s software, a series of menu trees can be found in Appendix A of this manual. NOTE The flow charts appearing in this section contain typical representations of the analyzer’s display during the various operations being described. These representations may differ slightly from the actual display of your instrument. The ENTR key may disappear if you select a setting that is invalid or out of the allowable range for that parameter, such as trying to set the 24-hour clock to 25:00:00. Once you adjust the setting to an allowable value, the ENTR key will re-appear. 6.1. OVERVIEW OF OPERATING MODES The M200EH/EM software has a variety of operating modes. Most commonly, the analyzer will be operating in SAMPLE mode. In this mode, a continuous read-out of the NO, NO2 and NOx concentrations are displayed on the front panel and are available to be output as analog signals from the analyzer’s rear panel terminals. Also, calibrations can be performed, and TEST functions and WARNING messages can be examined. The second most important operating mode is SETUP mode. This mode is used for performing certain configuration operations, such as for the iDAS system, configuring the reporting ranges, or the serial (RS232/RS-485/Ethernet) communication channels. The SET UP mode is also used for performing various diagnostic tests during troubleshooting. Mode Field SAMPLE CAL Figure 6-6-1: NOX=050.1 SETUP Front Panel Display The mode field of the front panel display indicates to the user which operating mode the unit is currently running. 55 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual Besides SAMPLE and SETUP, other modes the analyzer can be operated in are: Table 6-1: MODE Analyzer Operating modes EXPLANATION SAMPLE Sampling normally, flashing text indicates adaptive filter is on. M-P CAL This is the basic calibration mode of the instrument and is activated by pressing the CAL key. SETUP X.#2 SETUP mode is being used to configure the analyzer. The gas measurement will continue during this process. SAMPLE A Indicates that unit is in SAMPLE mode and AUTOCAL feature is activated. ZERO CAL M1 Unit is performing ZERO calibration procedure initiated manually by the user. ZERO CAL A1 Unit is performing ZERO calibration procedure initiated automatically by the AUTOCAL feature. ZERO CAL R1 Unit is performing ZERO calibration procedure initiated remotely through the COM ports or digital control inputs. LO CAL A Unit is performing LOW SPAN (midpoint) calibration initiated automatically by the analyzer’s AUTOCAL feature. LO CAL R Unit is performing LOW SPAN (midpoint) calibration initiated remotely through the COM ports or digital control inputs. SPAN CAL M1 Unit is performing SPAN calibration initiated manually by the user. SPAN CAL A1 Unit is performing SPAN calibration initiated automatically by the analyzer’s AUTOCAL feature. SPAN CAL R1 Unit is performing SPAN calibration initiated remotely through the COM ports or digital control inputs. DIAG One of the analyzer’s diagnostic modes is active (Section 6.13). 1 Only Appears on units with Z/S valve or IZS options. The revision of the analyzer firmware is displayed following the word SETUP, e.g., SETUP F.0. 2 The third operating mode is the CAL mode, which allows calibration of the analyzer in various ways. Because of its importance, this mode is described separately in Chapter 7. 56 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.2. SAMPLE MODE This is the analyzer’s standard operating mode. In this mode, the instrument is analyzing NO and NOX and calculating NO2 concentrations. 6.2.1. TEST FUNCTIONS A series of test functions is available at the front panel while the analyzer is in SAMPLE mode. These parameters provide information about the present operating status of the instrument and are useful during troubleshooting (Section 11.1.2). They can also be recorded in one of the iDAS channels (Section 6.7) for data analysis or output on one of the configurable analog outputs. Table 6-2: DISPLAY PARAMETER UNITS Test Functions Defined DESCRIPTION A1:NXCNC1=100 PPM Analog output range configuration A2:N0CNC1=100 PPM These functions show the default settings for the enabled analog output channels. See section 6.13.4 for more information. A3:N2CNC1=25 PPM A4:NXCNC2=100% The stability is a standard deviation of the NOX concentration over 25 samples, each recorded every 10 seconds. A low NOX STB value indicates low variability in NOX. NOX STB STABILITY PPM, MGM SAMP FLW SAMPLE FLOW cm³/min (cc/m) OZONE FL OZONE cm³/min (cc/m) PMT PMT Signal MV The raw output voltage of the PMT. NORM PMT NORMALIZED PMT Signal MV The output voltage of the PMT after normalization for auto-zero offset and temperature/pressure compensation (if activated). AZERO AUTO-ZERO MV The PMT signal with zero NOX, which is usually slightly different from 0 V. This offset is subtracted from the PMT signal and adjusts for variations in the zero signal. HVPS HVPS V The PMT high voltage power supply. RCELL TEMP REACTION CELL TEMP C The current temperature of the reaction cell. BOX TEMP BOX TEMPERATURE C The ambient temperature of the inside of the analyzer case. PMT TEMP PMT TEMPERATURE C The current temperature of the PMT. IZS TEMP IZS TEMPERATURE 1 C The current temperature of the internal zero/span option. Only appears when IZS option is enabled. MOLY TEMP CONV TEMPERATURE RCEL REACTION CELL PRESSURE in-Hg-A The current gas pressure of the reaction cell as measured at the vacuum manifold. This is the vacuum pressure created by the external pump. SAMP SAMPLE PRESSURE in-Hg-A The current pressure of the sample gas as it enters the reaction cell, measured between the NO/NOx and Auto-Zero valves. NOX SLOPE NOx SLOPE -- The slope of the current NOx calibration as calculated from a linear fit during the analyzer’s last zero/span calibration. NOX OFFS NOx OFFSET MV The offset of the current NOx calibration as calculated from a linear fit during the analyzer’s last zero/span calibration. NO SLOPE NO SLOPE -- The slope of the current NO calibration as calculated from a linear fit during the analyzer’s last zero/span calibration. NO OFFS NO OFFSET MV The offset of the current NO calibration as calculated from a linear fit during the analyzer’s last zero/span calibration. C The flow rate of the sample gas through the reaction cell. This value is not measured but calculated from the sample pressure. Flow rate of the O3 gas stream as measured with a flow meter The current temperature of the NO2 converter. NO2 NO2 concentration PPM, MGM The current NO2 concentration in the chosen unit. NOX NOx concentration PPM, MGM The current NOx concentration in the chosen unit. PPM, MGM The current NO concentration in the chosen unit. NO NO concentration TEST TEST SIGNAL MV TIME CLOCK TIME hh:mm:ss 2 Signal of a user-defined test function on output channel A4. The current day time for iDAS records and calibration events. 57 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual SAMPLE A1:NXCNC1=100 PPM < TST TST > CAL 1 NOX = XXX SETUP 1 Toggle keys to scroll through list of functions 1 Default settings for user selectable reporting range settings. 2 Only appears if O2 sensor option is installed. Figure 6-6-2: A1:NXCNC1=100 PPM 1 A2:N0CNC1=100 PPM 1 A3:N2CNC1=25 PPM 1 A4:NXCNC2=100% NOX STB SAMP FLOW 0ZONE FLOW PMT NORM PMT AZERO Refer to HVPS Section RCELL TEMP BOX TEMP 6.2.1 for PMT TEMP definitions MF TEMP of these 2 O2 CELL TEMP test MOLY TEMP functions. RCEL SAMP NOX SLOPE NOX OFFSET NO SLOPE NO OFFSET O2 SLOPE2 2 O2 OFFSET TIME Viewing M200EH/EM TEST Functions NOTE A value of “XXXX” displayed for any of the TEST functions indicates an out-of-range reading or the analyzer’s inability to calculate it. All pressure measurements are represented in terms of absolute pressure. Absolute, atmospheric pressure is 29.92 in-Hg-A at sea level. It decreases about 1 in-Hg per 300 m gain in altitude. A variety of factors such as air conditioning and passing storms can cause changes in the absolute atmospheric pressure. 58 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.2.2. WARNING MESSAGES The most common instrument failures will be reported as a warning on the analyzer’s front panel and through the COM ports. Section 11.1.2 explains how to use these messages to troubleshoot problems. Section 0 shows how to view and clear warning messages. Table 6-3: List of Warning Messages Revision F.0 MESSAGE ANALOG CAL WARNING AZERO WRN XXX.X MV BOX TEMP WARNING CANNOT DYN SPAN CANNOT DYN ZERO CONFIG INITIALIZED CONV TEMP WARNING DATA INITIALIZED HVPS WARNING IZS TEMP WARNING OZONE FLOW WARNING OZONE GEN OFF PMT TEMP WARNING RCELL PRESS WARN RCELL TEMP WARNING REAR BOARD NOT DET RELAY BOARD WARN SAMPLE FLOW WARN SYSTEM RESET MEANING The instruments analog-to-digital converter (A/D) circuitry or one of the analog outputs are not calibrated. The reading taken during the Auto-zero cycle is outside the specified limits. The value shown here as “XXX.X” indicates the actual auto-zero reading at the time of the warning. The temperature inside the M200EH/EM chassis is outside the specified limits. Remote span calibration failed while the dynamic span feature was ON. Remote zero calibration failed while the dynamic zero feature was ON. Configuration storage was reset to factory configuration or was erased. NO2 converter temperature is outside of specified limits. iDAS data storage was erased. High voltage power supply for the PMT is outside of specified limits. On units with IZS option installed: The IZS temperature is outside of specified limits. Ozone flow is outside of specified limits. Ozone generator is off. This is the only warning message that automatically clears itself when the ozone generator is turned on. PMT temperature is outside of specified limits. Reaction cell pressure is outside of specified limits. Reaction cell temperature is outside of specified limits. The firmware is unable to communicate with the motherboard. The firmware is unable to communicate with the relay board. The flow rate of the sample gas is outside the specified limits. The computer rebooted or was powered up. To view and clear warning messages SAMPLE TEST deactivates warning messages TEST A1:NXCNC1=100PPM CAL MSG A1:NXCNC1=100PPM SAMPLE MSG < TST TST > CAL HVPS WARNING SAMPLE NOTE: If the warning message persists after several attempts to clear it, the message may indicate a real problem and not an artifact of the warm-up period Figure 6-6-3: TEST CAL Make sure warning messages are not due to real problems. MSG NOX=XXX.X CLR SETUP NO=XXX.X CLR SETUP NO2=XXX.X CLR SETUP MSG activates warning messages. keys replaced with TEST key All Warning messages are hidden, but MSG button appears Press CLR to clear the current message. If more than one warning is active, the next message will take its place Once the last warning has been cleared, the analyzer returns to SAMPLE mode Viewing and Clearing M200EH/EM WARNING Messages 59 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.3. CALIBRATION MODE 6.3.1. CALIBRATION FUNCTIONS Pressing the CAL key switches the M200EH/EM into calibration mode. In this mode, the user can calibrate the instrument with the use of calibrated zero or span gases. If the instrument includes either the zero/span valve option or IZS option, the display will also include CALZ and CALS keys. Pressing either of these keys also puts the instrument into multipoint calibration mode. The CALZ key is used to initiate a calibration of the zero point. The CALS key is used to calibrate the span point of the analyzer. It is recommended that this span calibration is performed at 90% of full scale of the analyzer’s currently selected reporting range. Because of their critical importance and complexity, calibration operations are described in detail in other sections of the manual: Chapter 7 details basic calibration and calibration check operations. For more information concerning the zero/span, zero/span/shutoff valve options, see Section 5.7. 60 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.4. SETUP MODE The SETUP mode contains a variety of choices that are used to configure the analyzer’s hardware and software features, perform diagnostic procedures, gather information on the instruments performance and configure or access data from the internal data acquisition system (iDAS). The areas access under the Setup mode are: Table 6-4: Primary Setup Mode Features and Functions MODE OR FEATURE KEYPAD LABEL Analyzer Configuration CFG DESCRIPTION MANUAL SECTION Lists key hardware and software configuration information 6.5 Used to set up an operate the AutoCal feature. Only appears if the analyzer has one of the internal valve options installed 7.7 Used to set up the iDAS system and view recorded data 6.7 6.8 Auto Cal Feature ACAL Internal Data Acquisition (iDAS) DAS Analog Output Reporting Range Configuration RNGE Used to set the units of measure for the display and set the dilution ratio on instruments with that option active. Calibration Password Security PASS Turns the password feature ON/OFF 6.9 Internal Clock Configuration CLK Used to Set or adjust the instrument’s internal clock 6.10 Advanced SETUP features MORE This button accesses the instruments secondary setup menu See Table 6-5 Table 6-5: Secondary Setup Mode Features and Functions 1 MODE OR FEATURE KEYPAD LABEL External Communication Channel Configuration COMM Used to set up and operate the analyzer’s various external I/O channels including RS-232; RS 485, modem communication and/or Ethernet access. System Status Variables VARS Used to view various variables related to the instruments current operational status System Diagnostic Features and Analog Output Configuration DIAG Alarm Limit Configuration1 ALRM DESCRIPTION Used to access a variety of functions that are used to configure, test or diagnose problems with a variety of the analyzer’s basic systems. MANUAL SECTION 6.11 & 6.15 6.12 6.13 Most notably, the menus used to configure the output signals generated by the instruments Analog outputs are located here. Used to turn the instrument’s two alarms on and off as well as set the trigger limits for each. 6.14 Only present if the optional alarm relay outputs (Option 67) are installed. NOTE Any changes made to a variable during one of the following procedures is not acknowledged by the instrument until the ENTR Key is pressed If the EXIT key is pressed before the ENTR key, the analyzer will beep, alerting the user that the newly entered value has not been accepted. 61 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.5. SETUP CFG: VIEWING THE ANALYZER’S CONFIGURATION INFORMATION Pressing the CFG key displays the instrument configuration information. This display lists the analyzer model, serial number, firmware revision, software library revision, CPU type and other information. Use this information to identify the software and hardware when contacting customer service. Special instrument or software features or installed options may also be listed here. SAMPLE A1:NXCNC1=100PPM < TST TST > CAL Press NEXT of PREV to move back and forth through the following list of Configuration information: MODEL NAME SERIAL NUMBER SOFTWARE REVISION LIBRARY REVISION iCHIP SOFTWARE REVISION1 HESSEN PROTOCOL REVISION1 ACTIVE SPECIAL SOFTWARE OPTIONS1 CPU TYPE DATE FACTORY CONFIGURATION SAVED SAMPLE NOX=XXX.X SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SAMPLE NEXT EXIT M100E NOX ANALYZER PREV EXIT Press EXIT at any time to return to the SAMPLE display Press EXIT at any time to return to SETUP menu 1 Only appears if relevant option of Feature is active. 6.6. SETUP ACAL: AUTOMATIC CALIBRATION Instruments with one of the internal valve options installed can be set to automatically run calibration procedures and calibration checks. These automatic procedures are programmed using the submenus and functions found under the ACAL menu. A menu tree showing the ACAL menu’s entire structure can be found in Appendix A-1 of this manual. Instructions for using the ACAL feature are located in the Section 7.7 of this manual along with all other information related to calibrating the M200EH/EM analyzer. 62 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.7. SETUP DAS - USING THE DATA ACQUISITION SYSTEM (iDAS) The M200EH/EM analyzer contains a flexible and powerful, internal data acquisition system (iDAS) that enables the analyzer to store concentration and calibration data as well as a host of diagnostic parameters. The iDAS of the M200EH/EM can store up to about one million data points, which can, cover days, weeks or months of valuable measurements. The data are stored in non-volatile memory and are retained even when the instrument is powered off. Data are stored in plain text format for easy retrieval and use in common data analysis programs (such as spreadsheet-type programs). NOTE: Please be aware that all stored data will be erased if the analyzer’s disk-on-chip, CPU board or configuration is replaced/reset. The iDAS is designed to be flexible. Users have full control over the type, length and reporting time of the data. The iDAS permits users to access stored data through the instrument’s front panel or its communication ports. Teledyne Instruments also offers APICOM, a program that provides a visual interface for configuration and data retrieval of the iDAS or using a remote computer. Additionally, the analyzer’s four analog output channels can be programmed to carry data related to any of the available iDAS parameters. The principal use of the iDAS is logging data for trend analysis and predictive diagnostics, which can assist in identifying possible problems before they affect the functionality of the analyzer. The secondary use is for data analysis, documentation and archival in electronic format. IDAS STATUS The green SAMPLE LED on the instrument front panel, which indicates the analyzer status, also indicates certain aspects of the iDAS status: Table 6-6: Front Panel LED Status Indicators for iDAS LED STATE IDAS STATUS Off System is in calibration mode. Data logging can be enabled or disabled for this mode. Calibration data are typically stored at the end of calibration periods, concentration data are typically not sampled, diagnostic data should be collected. Blinking Instrument is in hold-off mode, a short period after the system exits calibrations. IDAS channels can be enabled or disabled for this period. Concentration data are typically disabled whereas diagnostic should be collected. On Sampling normally. The iDAS can be disabled only by disabling or deleting its individual data channels. 63 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.7.1. IDAS STRUCTURE The iDAS is designed around the feature of a “record”, an automatically stored single data point. (e.g. concentration, PMT signal level, etc.). Records are organized into data channels which are defined by properties that characterize the: Type of date recorded (e.g. concentration, PMT signal level, etc.); Trigger event that causes the record to be made (e.g. every minute, upon exiting calibration mode, etc.); How many records to be stored, as well as; How the information is to be stored (e.g. average over 1 hour, individual points, minimum value over last 5 minutes, etc.). The configuration of each iDAS channel is stored in the analyzer’s memory as a script, which can be edited from the front panel or downloaded, edited and uploaded to the instrument in form of a string of plain-text lines through the communication ports. 6.7.1.1. iDAS Channels The key to the flexibility of the iDAS is its ability to store a large number of combinations of triggering events and data parameters in the form of data channels. Users may create up to 20 data channels. For each channel one triggering event is selected and one or all of the M200EH/EM’s 25 data parameters are allowed. The number of parameters and channels is limited by available memory. The properties that define the structure of an iDAS data channel are: Table 6-7: iDAS Data Channel Properties PROPERTY DEFAULT SETTING RANGE The name of the data channel. “NONE” Up to 6 letters or digits1. TRIGGERING EVENT The event that triggers the data channel to measure and store the datum ATIMER Any available event (see Appendix A-5). NUMBER AND LIST OF PARAMETERS A User-configurable list of data types to be recorded in any given channel. 1 - PMTDET Any available parameter (see Appendix A-5). 000:01:00 000:00:01 to 366:23:59 (Days:Hours:Minutes) 100 1 to 1 million, limited by available storage space. OFF OFF or ON ON OFF or ON OFF OFF or ON NAME REPORT PERIOD NUMBER OF RECORDS RS-232 REPORT CHANNEL ENABLED CAL HOLD OFF DESCRIPTION The amount of time between each channel data point. The number of reports that will be stored in the data file. Once the limit is exceeded, the oldest data is over-written. Enables the analyzer to automatically report channel values to the RS-232 ports. Enables or disables the channel. Allows a channel to be temporarily turned off without deleting it. Disables sampling of data parameters while 2 instrument is in calibration mode . 1 More with APICOM, but only the first six are displayed on the front panel). 2 When enabled records are not recorded until the DAS HOLD OFF period is passed after calibration mode. DAS HOLD OFF set in the VARS menu (see Section 6.12.) 64 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.7.1.2. iDAS Parameters Data parameters are types of data that may be measured by the analyzers instrumentality concentrations of measured gases, temperatures of heated zones,, pressures and flows of the pneumatic subsystem as well as calibration data such as slope and offset for each gas. For each Teledyne Instruments analyzer model, the list of available data parameters is different, fully defined and not customizable (see Appendix A.5 for a list of M200EH/EM parameters). Most data parameters have associated measurement units, such as mV, ppm, cm³/min, etc., although some parameters have no units. The only units that can be changed are those of the concentration readings according to the SETUP-RANGE settings. NOTE The iDAS does not keep track of the unit of each concentration value and iDAS data files may contain concentrations in multiple units if the unit was changed during data acquisition. Each data parameter has user-configurable functions that define how the data are recorded. Table 6-8: iDAS Data Parameter Functions FUNCTION PARAMETER SAMPLE MODE EFFECT Instrument-specific parameter name. INST: Records instantaneous reading. AVG: Records average reading during reporting interval. MIN: Records minimum (instantaneous) reading during reporting interval. MAX: Records maximum (instantaneous) reading during reporting interval. SDEV: Records the standard deviation of the data points recorded during the reporting interval. PRECISION STORE NUM. SAMPLES Decimal precision of parameter value(0-4). OFF: stores only the average (default). ON: stores the average and the number of samples in each average for a parameter. This property is only useful when the AVG sample mode is used. Note that the number of samples is the same for all parameters in one channel and needs to be specified only for one of the parameters. 6.7.1.3. iDAS Triggering Events Triggering events define when and how the iDAS records a measurement of any given data channel. Triggering events are firmware-specific and are listed in Appendix A-5. The most common triggering events are: ATIMER: Sampling occurs at regular intervals specified by an automatic timer. Trending information is often stored via such intervals, as either individual datum or averaged. EXITZR, EXITSP, SLPCHG (exit zero, exit span, slope change): Sampling at the end of an irregularly occurring event such as calibration or when the slope changes. These events create individual data points. Zero and slope values can be used to monitor response drift and to document when the instrument was calibrated. WARNINGS: Some data may be useful when stored if one of several warning messages appears. This is helpful for trouble-shooting by monitoring when a particular warning occurred. 65 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.7.2. DEFAULT IDAS CHANNELS The M200EH/EM is configured with a basic iDAS configuration, which is enabled by default. New data channels are also enabled by default but each channel may be turned off for later or occasional use. Note that iDAS operation is suspended while its configuration is edited through the front panel. To prevent such data loss, it is recommended to use the APICOM graphical user interface for iDAS changes. A set of default data channels has been included in the analyzer’s software for logging nitrogen oxides concentrations, calibration and predictive diagnostic data. They are: CONC: Samples NOX, NO and NO2 concentration at one minute intervals and stores an average every hour with a time and date stamp along with the number of (1-minute) samples within each average(for statistical evaluation). Readings during calibration and calibration hold off are not included in the data. By default, the last 800 hourly averages are stored. CALDAT: Every time a zero or span calibration is performed CALDAT logs concentration, slope and offset values for NOX and NO with a time and date stamp. The NOX stability (to evaluate calibration stability) as well as the converter efficiency (for reference) are also stored. This data channel will store data from the last 200 calibrations and can be used to document analyzer calibration. The slope and offset data can be used to detect trends in (instrument response. CALCHECK: This channel logs concentrations and the stability each time a zero or span check (not calibration) is finished. This allows the user to track the quality of zero and span responses over time and assist in evaluating the quality of zero and span gases and the analyzer’s noise specifications. The last 200 data points are retained. DIAG: Daily averages of temperature zones, flow and pressure data as well as some other diagnostic parameters (HVPS, AZERO). These data are useful for predictive diagnostics and maintenance of the M200EH/EM. The last 1100 daily averages are stored to cover more than four years of analyzer performance. HIRES: Records one minute, instantaneous data of all active parameters in the M200EH/EM. Shortterm trends as well as signal noise levels can be detected and documented. Readings during calibration and the calibration hold off period are included in the averages. The last 1500 data points are stored, which covers a little more than one day of continuous data acquisition. This data channel is disabled by default but may be turned on when needed such as for trouble-shooting problems with the analyzer. The default data channels can be used as they are, or they can be customized from the front panel or through APICOM to fit a specific application. The Teledyne Instruments website contains this default and other sample iDAS scripts for free download. We recommend that the user backs up any iDAS configuration and its data before altering it. NOTE Teledyne-API recommends downloading and storing existing data and the iDAS configurations regularly for permanent documentation and future data analysis. Sending an iDAS configuration to the analyzer through its COM ports will replace the existing configuration and will delete all stored data. 66 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions Table 6-9: M200EH/EM Default iDAS Configuration PARAMETERS CHANNELS with PROPERTIES Name: CONC Event: ATIMER Sample Period: 000:00:01 Report Period: 000:01:00 Number of Records: 800 RS-232 report: OFF Channel enabled: ON DAS HOLDOFF: ON Name: CALDAT Event: SLPCHG Number of Records: 200 RS-232 report: OFF Channel enabled: ON DAS HOLDOFF: OFF Name: CALCHECK Event: EXITMP Number of Records: 200 RS-232 report: OFF Channel enabled: ON DAS HOLDOFF: OFF Name: CALCHECK Event: EXITMP Number of Records: 200 RS-232 report: OFF Channel enabled: ON DAS HOLDOFF: OFF Name: HIRES Event: ATIMER Sample Period: 000:00:01 Report Period: 000:00:01 Number of Records: 1500 RS-232 report: OFF Channel enabled: OFF DAS HOLDOFF: OFF NAME MODE EVENT PRECISION NUM SAMPLES NOXCNC1 AVG -- 4 ON NOCNC1 AVG -- 4 OFF N2CNC1 AVG -- 4 OFF STABIL AVG -- 4 OM NXZSC1 -- SLPCHG 4 OFF NOXSLP1 NOXOFFS1 NOZSC1 ---- SLPCHG SLPCHG SLPCHG 4 4 4 OFF OFF OFF NOSLP1 NOOFFS1 N2ZSC1 CNVEF1 STABIL ------ SLPCHG SLPCHG SLPCHG SLPCHG SLPCHG 4 4 4 4 4 OFF OFF OFF OFF OFF NXZSC1 -- EXITMP 4 OFF NOZSC1 -- EXITMP 4 OFF N2ZSC1 -- EXITMP 4 OFF STABIL -- EXITMP 4 OFF SMPFLW O3FLOW AVG AVG --- 2 2 OFF OFF RCPRESS SMPPRES RCTEMP PMTTMP CNVTMP BOXTMP AVG AVG AVG AVG AVG AVG ------- 2 2 2 2 2 2 OFF OFF OFF OFF OFF OFF HVPS AZERO AVG AVG --- 2 2 OFF OFF NOXCNC1 AVG -- 4 OFF NOCNC1 N2CNC1 STABIL SMPFLW O3FLOW RCPRESS SMPPRES AVG AVG AVG AVG AVG AVG AVG -------- 4 4 4 2 2 2 2 OFF OFF OFF OFF OFF OFF OFF RCTEMP PMTTMP CNVTMP BOXTMP HVPS AZERO REFGND AVG AVG AVG AVG AVG AVG AVG ------- 2 2 2 2 1 2 1 OFF OFF OFF OFF OFF OFF OFF REF4096 AVG 1 OFF 67 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.7.2.1. Viewing iDAS Data and Settings IDAS data and settings can be viewed on the front panel through the following keystroke sequence. VIEW KEYPAD FUNCTIONS SAMPLE A1:NXCNC1=100PPM < TST TST > CAL EXIT will return to the main SAMPLE Display. SETUP X.X SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X KEY FUNCTION Moves to the previous Parameter NX10 Moves the view forward 10 data points/channels NEXT Moves to the next data point/channel PREV Moves to the previous data point/channel PV10 Moves the view back 10 data points/channels NOX=XXX.X EXIT DATA ACQUISITION VIEW EDIT EXIT Keys only appear as needed SETUP X.X NEXT SETUP X.X PREV NEXT CONC : DATA AVAILABLE VIEW EXIT SETUP X.X 287:10:00 PV10 PREV NEXT NX10 cc/m EXIT EXIT PV10 PREV SETUP X.X PRM> DIAG: DATA AVAILABLE SETUP X.X Default setting for HIRES is DISABLED. NEXT NX10 CALDAT: DATA AVAILABLE SETUP X.X SETUP X.X NXCNC1: XXX.X PPM 00:00::00 PMTDET=0000.0000 m EXIT HIRES: NO DATA AVAILABLE EXIT 68 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.7.2.2. Editing iDAS Data Channels IDAS configuration is most conveniently done through the APICOM remote control program. The following list of key strokes shows how to edit using the front panel. SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL EXIT will return to the previous SAMPLE display. SETUP X.X SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X EXIT DATA ACQUISITION VIEW EDIT SETUP X.X 8 1 EXIT ENTER DAS PASS: 818 8 ENTR EXIT Edit Data Channel Menu Moves the display up & down the list of Data Channels Inserts a new Data Channel into the list BEFORE the Channel currently being displayed Moves the display between the PROPERTIES for this data channel. SETUP X.X 0) CONC: PREV NEXT INS ATIMER, DEL EDIT 8, PRNT 800 EXIT Exits to the Main Data Acquisition Menu Exports the configuration of all data channels to RS-232 interface. Deletes The Data Channel currently being displayed SETUP X.X NAME:CONC EDIT PRNT Allows to edit the channel name, see next key sequence. EXIT EXITS returns to the previous Menu Reports the configuration of current data channels to the RS-232 ports. When editing the data channels, the top line of the display indicates some of the configuration parameters. For example, the display line: 0) CONC : ATIMER, 4, 800 Translates to the following configuration: Channel No.: 0 NAME: CONC TRIGGER EVENT: ATIMER PARAMETERS: Four parameters are included in this channel EVENT: This channel is set up to record 800 data points. 69 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual To edit the name of a data channel, follow the above key sequence and then press: FROM THE PREVIOUS KEY SEQUENCE … SETUP X.X EDIT SETUP X.X C NAME:CONC O PRINT EXIT NAME:CONC N C - - ENTR EXIT ENTR accepts the new string and returns to the previous menu. EXIT ignores the new string and returns to the previous menu. Press each key repeatedly to cycle through the available character set: 0-9, A-Z, space ’ ~ ! # $ % ^ & * ( ) - _ = +[ ] { } < >\ | ; : , . / ? 6.7.2.3. Trigger Events To edit the list of data parameters associated with a specific data channel, press: From the DATA ACQUISITION menu (see Section 6.7.2.2) Edit Data Channel Menu SETUP X.X 0) CONC: PREV NEXT SETUP X.X PRNT 800 EXIT EXITS to the Main Data Acquisition menu PRINT EXIT EVENT:ATIMER SET> EDIT SETUP X.X DEL EDIT 8, NAME:CONC SET> EDIT SETUP X.X EDIT 8, 800 EXIT PRNT PRINT Exits to the main Data Acquisition menu EXIT Press SET> key until… SETUP X.X EDIT SETUP X.X YES PARAMETERS: 8 PRINT EXIT EDIT PARAMS (DELETE DATA) NO returns to the previous menu and retains all data. NO Edit Data Parameter Menu Moves the display between available Parameters Inserts a new Parameter before the currently displayed Parameter SETUP X.X PREV NEXT 0) PARAM=DETREP, MODE=INST INS DEL EDIT Deletes the Parameter currently displayed. EXIT Exits to the main Data Acquisition menu Use to configure the functions for this Parameter. 71 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual To configure the parameters for a specific data parameter, press: FROM THE EDIT DATA PARAMETER MENU (see previous section) SETUP X.X 0) PARAM=NXCNC!, MODE=AVG PREV NEXT SETUP X.X INS DEL EDIT EXIT PARAMETERS: NOCNC1 EXIT SET> EDIT SETUP X.X PARAMETER: NXCNC1 PREV NEXT ENTR EXIT Cycle through list of available Parameters. SETUP X.X SAMPLE MODE: INST EXIT EDIT SETUP X.X INST AVG SAMPLE MODE: INST MIN MAX EXIT Press the key for the desired mode SETUP X.X PRECISION:4 EDIT EXIT ENTR accepts the new setting and returns to the previous menu. EXIT ignores the new setting and returns to the previous menu. SETUP X.X PRECISION: 4 1 EXIT Set for 0-4 SETUP X.X STORE NUM. SAMPLES: OFF EDIT DEL EDIT 8, PRNT PRINT 8500 EXIT Exits to the main Data Acquisition menu. EXIT Press SET> key until you reach REPORT PERIOD (OR SAMPLE PERIOD) … SETUP X.X EDIT SETUP X.X Set the number of days between reports (0-366). Press keys to set hours between reports in the format : HH:MM (max: 23:59). This is a 24 hour clock . PM hours are 13 thru 23, midnight is 00:00. Example 2:15 PM = 14:15 0 0 SETUP X.X 0 REPORT PERIOD:000:01:00 1 PRINT EXIT REPORT PERIODD:DAYS:0 0 ENTR EXIT REPORT PERIODD:TIME:01:01 0 0 ENTR EXIT IIf at any time an illegal entry is selected (e.g., days > 366) the ENTR key will disappear from the display. ENTR accepts the new string and returns to the previous menu. EXIT ignores the new string and returns to the previous menu. 74 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.7.2.6. Number of Records The number of data records in the M200EH/EM is limited to a cumulative one million data points in all channels (one megabyte of space on the disk-on-chip). However, the actual number of records is also limited by the total number of parameters and channels and other settings in the iDAS configuration. Every additional data channel, parameter, number of samples setting etc. will reduce the maximum amount of data points somewhat. In general, however, the maximum data capacity is divided amongst all channels (max: 20) and parameters (max: 50 per channel). The iDAS will check the amount of available data space and prevent the user from specifying too many records at any given point. If, for example, the iDAS memory space can accommodate 375 more data records, the ENTR key will disappear when trying to specify more than that number of records. This check for memory space may also make an upload of an iDAS configuration with APICOM or a Terminal program fail, if the combined number of records would be exceeded. In this case, it is suggested to either try from the front panel what the maximum number of records can be or use trial-and-error in designing the iDAS script or calculate the number of records using the DAS or APICOM manuals. To set the number of records for one channel from the front panel, follow the instruction shown in section 6.7.2.2 then press. Edit Data Channel DATA ACQUISITION menu From the Menu (see Section 6.7.2.2) SETUP X.X 0) CONC: PREV NEXT INS SETUP X.X EDIT PRINT EXIT Press SET> key until… SETUP X.X EDIT SETUP X.X YES PRINT EXIT EDIT RECOPRDS (DELET DATA) NO returns to the previous menu. NO SETUP X.X 0 NUMBER OF RECORDS:000 0 REPORT PERIODD:DAYS:0 0 0 0 ENTR EXIT ENTR accepts the new setting and returns to the previous menu. EXIT ignores the new setting and returns to the previous menu. 75 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.7.2.7. RS-232 Report Function The M200EH/EM iDAS can automatically report data to the communications ports, where they can be captured with a terminal emulation program or simply viewed by the user. To enable automatic COM port reporting, follow the instruction shown in section 6.7.2.2 then press: From the DATA ACQUISITION menu (see Section 6.7.2.2) Edit Data Channel Menu SETUP X.X PREV NEXT 0) CONC: INS SETUP X.X EDIT PRINT EXIT Press SET> key until… SETUP X.X EDIT SETUP X.X Toggle key to turn reporting ON or OFF OFF RS-232 REPORT: OFF PRINT EXIT RS-232 REPORT: OFF ENTR EXIT ENTR accepts the new setting and returns to the previous menu. EXIT ignores the new setting and returns to the previous menu. 6.7.2.8. Compact Report When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of reporting each parameter in one channel on a separate line, up to five parameters are reported in one line, instead. For example, channel DIAG would report its record in two lines (10 parameters) instead of 10 lines. Individual lines carry the same time stamp and are labeled in sequence. 6.7.2.9. Starting Date This option allows to specify a starting date for any given channel in case the user wants to start data acquisition only after a certain time and date. If the Starting Date is in the past, the iDAS ignores this setting. 76 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.7.2.10. Disabling/Enabling Data Channels Data channels can be temporarily disabled, which can reduce the read/write wear on the disk-on-chip. The HIRES channel of the M200EH/EM, for example, is disabled by default. To disable a data channel, follow the instruction shown in section 6.7.2.2 then press: From the DATA ACQUISITION menu (see Section 6.7.2.2) Edit Data Channel Menu SETUP X.X 0) CONC: PREV NEXT INS SETUP X.X EDIT PRINT EXIT Press SET> key until… SETUP X.X EDIT PRINT EXIT CHANNEL ENABLE:ON OFF ENTR EXIT ENTR accepts the new setting and returns to the previous menu. EXIT ignores the new setting and returns to the previous menu. 6.7.2.11. HOLDOFF Feature The iDAS HOLDOFF feature allows to prevent data collection during calibrations and during the DAS_HOLDOFF period enabled and specified in the VARS (Section 6.12). To enable or disable the HOLDOFF for any one iDAS channel, follow the instruction shown in section 6.7.2.2 then press: From the DATA ACQUISITION menu (see Section 6.7.2.2) Edit Data Channel Menu SETUP X.X 0) CONC: PREV NEXT SETUP X.X EDIT PRINT EXIT Press SET> key until… SETUP X.X CAL HOLD OFF:ON SET> EDIT SETUP X.X Toggle key to turn HOLDOFF ON or OFF ON PRINT EXIT CAL HOLD OFF:ON ENTR EXIT ENTR accepts the new setting and returns to the previous menu. EXIT ignores the new setting and returns to the previous menu. 77 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.7.3. REMOTE IDAS CONFIGURATION Editing channels, parameters and triggering events as described in 6.7 is much more conveniently done in one step through the APICOM remote control program using the graphical interface shown in Figure 6-6-4. Refer to Section 6.15 for details on remote access to the M200EH/EM analyzer. Figure 6-6-4: APICOM Graphical User Interface for Configuring the iDAS Once an iDAS configuration is edited (which can be done offline and without interrupting DAS data collection), it is conveniently uploaded to the instrument and can be stored on a computer for later review, alteration or documentation and archival. Refer to the APICOM manual for details on these procedures. The APICOM user manual (Teledyne Instruments part number 039450000) is included in the APICOM installation file, which can be downloaded at http://www.teledyne-api.com/software/apicom/. Although Teledyne Instruments recommends the use of APICOM, the iDAS can also be accessed and configured through a terminal emulation program such as HyperTerminal (Figure 6-6-5). However, all configuration commands must be created following a strict syntax or be pasted in from of a text file, which was edited offline and then uploaded through a specific transfer procedure. 78 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Figure 6-6-5: Operating Instructions iDAS Configuration Through a Terminal Emulation Program Both procedures are best started by downloading the default iDAS configuration, getting familiar with its command structure and syntax conventions, and then altering a copy of the original file offline before uploading the new configuration. CAUTION Whereas the editing, adding and deleting of iDAS channels and parameters of one channel through the front-panel keyboard can be done without affecting the other channels, uploading an iDAS configuration script to the analyzer through its communication ports will erase all data, parameters and channels by replacing them with the new iDAS configuration. It is advised to download and backup all data and the original iDAS configuration before attempting any iDAS changes. 79 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.8. SETUP RNGE: RANGE UNITS AND DILUTION CONFIGURATION This Menu is used to set the units of measure to be associated with the analyzer’s reporting ranges (see Section 6.13.3.2. for more information on reporting ranges vs. physical ranges) and for instruments with the sample gas dilution option operating, to set the dilution ratio. 6.8.1. RANGE UNITS The M200EH/EM can display concentrations in parts per million (106 mols per mol, PPM) or milligrams per cubic meter (mg/m3, MG). Changing units affects all of the display, COM port and iDAS values for all reporting ranges regardless of the analyzer’s range mode. To change the concentration units: SAMPLE A1:NXCNC1= 100.0 PPM NOX=XXX.X < TST TST > CAL SETUP X.X SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X. UNIT RANGE CONTROL MENU DIL SETUP X.X Select the preferred concentration unit. EXIT EXIT CONC UNITS: PPM PPM MGM SETUP X.X EXIT returns to the main menu. ENTER EXIT CONC UNITS: MGM PPM MGM ENTER EXIT ENTR accepts the new unit, EXIT returns to the SETUP menu. Conversion factors from volumetric to mass units used in the M200EH/EM: NO: ppm x 1.34 = mg/m3 NO2: ppm x 2.05 = mg/m3 Concentrations displayed in mg/m3 and µg/m3 use 0° C and 760 Torr as standard temperature and pressure (STP). Consult your local regulations for the STP used by your agency. EPA protocol applications, for example, use 25° C as the reference temperature. Changing the units may cause a bias in the measurements if standard temperature and pressure other than 0C and 760 Torr are used. This problem can be avoided by recalibrating the analyzer after any change from a volumetric to a mass unit or vice versa. CAUTION In order to avoid a reference temperature bias, the analyzer must be recalibrated after every change in reporting units. 80 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.8.2. DILUTION RATIO The dilution ratio is a software option that allows the user to compensate for any dilution of the sample gas before it enters the sample inlet. 12. The SPAN value entered during calibration is the maximum expected concentration of the undiluted calibration gas 13. The span gas should be either supplied through the same dilution inlet system as the sample gas or be supplied at an appropriately lower actual concentration. For example, with a dilution set to 100, a 1 ppm gas can be used to calibrate a 100 ppm sample gas if the span gas is not routed through the dilution system. On the other hand, if a 100 ppm span gas is used, it needs to pass through the same dilution steps as the sample gas. 14. Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluent and 1 part of sample gas): The analyzer will multiply the measured gas concentrations with this dilution factor and displays the result. SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP PRIMARY SETUP MENU SETUP C.3 CFG DAS RNGE PASS CLK MORE DIL only appears if the dilution ratio option has been installed Toggle these keys to set the dilution factor. This is the number by which the analyzer will multiply the NO, NO2 and NOx concentrations of the gas passing through the reaction cell. SETUP C.3 UNIT RANGE CONTROL MENU DIL EXIT SETUP C.3 0 0 0 EXIT ignores the new setting. DIL FACTOR: 1.0 GAIN 0 SETUP C.3 0 EXIT 1 .0 ENTR EXIT ENTR accepts the new setting. DIL FACTOR: 20.0 GAIN 2 0 .0 ENTR EXIT The analyzer multiplies the measured gas concentrations with this dilution factor and displays the result. Calibrate the analyzer. Once the above settings have been entered, the instrument needs to be recalibrated using one of the methods discussed in Chapter 0. 81 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.9. SETUP PASS: PASSWORD FEATURE The M200EH/EM provides password protection of the calibration and setup functions to prevent unauthorized adjustments. When the passwords have been enabled in the PASS menu item, the system will prompt the user for a password anytime a password-protected function is requested. There are three levels of password protection, which correspond to operator, maintenance, and configuration functions. Each level allows access to all of the functions in the previous level. Table 6-10: Password Levels PASSWORD LEVEL MENU ACCESS ALLOWED No password Operator TEST, MSG, CLR 101 Maintenance CAL, CALZ, CALS 818 Configuration SETUP, VARS, DIAG To enable or disable passwords, press the following keystroke sequence: SAMPLE < TST TST > SETUP X.X A1:NXCNC1=100PPM NOX=XXX.X CAL SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X Toggle this button to enable, disable password feasture OFF SETUP X.X ON EXIT PASSWORD ENABLE: OFF ENTR EXIT PASSWORD ENABLE: ON ENTR EXIT Example: If all passwords are enabled, the following keypad sequence would be required to enter the SETUP menu: 82 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual SAMPLE A1:NXCNC1=100PPM Operating Instructions NOX=XXX.X < TST TST > CAL prompts for password number SAMPLE Press individual keys to set numbers SAMPLE 0 8 SETUP ENTER SETUP PASS: 0 0 0 ENTR EXIT ENTER SETUP PASS: 0 1 SETUP X.X 8 ENTR EXIT Example: this password enables the SETUP mode PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE EXIT Note that the instrument still prompts for a password when entering the VARS and DIAG menus, even if passwords are disabled, but it displays the default password (818) upon entering these menus. The user only has to press ENTR to access the password-protected menus but does not have to enter the required number code. 83 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.10. SETUP CLK: SETTING THE INTERNAL TIME-OF-DAY CLOCK The M200EH/EM has a built-in clock for the AutoCal timer, Time TEST function, and time stamps on COM port messages and iDAS data entries. To set the time-of-day, press: SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP X.X SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X Enter Current Time-of-Day SETUP X.X SETUP X.X3 1 2 :0 0 TIME-OF-DAY CLOCK TIME DATE EXIT SETUP X.X TIME: 12:00 1 2 :0 0 0 1 ENTR EXIT JAN 0 1 ENTR EXIT 0 2 ENTR EXIT DATE: 01-JAN-02 0 2 ENTR EXIT TIME-OF-DAY CLOCK TIME DATE SETUP X.X JAN Enter Current Date-of-Year DATE: 01-JAN-02 SETUP X.X TIME: 12:00 SETUP X.X EXIT EXIT PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE EXIT EXIT returns to the main SAMPLE display 84 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions In order to compensate for CPU clocks which run fast or slow, there is a variable to speed up or slow down the clock by a fixed amount every day. To change this variable, press: SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP SETUPX.X PREV NEXT JUMP PRIMARY SETUP MENU SETUP X.X CFG DAS RNGE PASS CLK MORE EDIT PRNT EXIT Continue to press NEXT until … EXIT SETUP X.X 7) CLOCK_ADJ=0 Sec/Day SECONDARY SETUP MENU SETUP X.X PREV COMM VARS DIAG JUMP SAMPLE ENTER SETUP PASS : 818 1 8 EDIT PRNT EXIT EXIT SETUP X.X 8 1 ) MEASURE_MODE=NOX-NO + 0 CLOCK_ADJ:0 Sec/Day 0 ENTR EXIT ENTR EXIT Enter sign and number of seconds per day the clock gains (-) or loses (+). SETUP X.X 0 ) DAS_HOLD_OFF=15.0 Minutes SETUP X.X NEXT JUMP 7) CLOCK_ADJ=0 Sec/Day EDIT PRNT EXIT PREV NEXT JUMP EDIT PRNT EXIT 3x EXIT returns to the main SAMPLE display 85 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.11. SETUP MORE COMM: SETTING UP THE ANALYSER’S COMMUNICATION PORTS The M200EH/EM is equipped with two serial communication ports located on the rear panel (see Figure 3-2). Both ports operate similarly and give the user the ability to communicate with, issue commands to, and receive data from the analyzer through an external computer system or terminal. By default, both ports operate on the RS-232 protocol. The COM1 port can also be configured to operate in single or RS-232 multidrop mode (option 62; See Section 5.9.2 and 6.11.7). The COM2 port, can be configured for standard RS-232 operation, half-duplex RS-485 communication or for access via an LAN by installing the Teledyne Instruments Ethernet interface card (option 63; see Section 5.9.3 and 6.11.6). A code-activated switch (CAS), can also be used on either port to connect typically between 2 and 16 send/receive instruments (host computer(s) printers, dataloggers, analyzers, monitors, calibrators, etc.) into one communications hub. Contact Teledyne Instruments sales for more information on CAS systems. 6.11.1. ANALYZER ID Each type of Teledyne Instruments analyzer is configured with a default ID code. The default ID code for all M200EH/EM analyzers is 200. The ID number is only important if more than one analyzer is connected to the same communications channel such as when several analyzers are on the same Ethernet LAN (see Section 6.11.6); in a RS-232 multidrop chain (see Section 6.11.7) or operating over a RS-485 network (see Section 6.11.4). If two analyzers of the same model type are used on one channel, the ID codes of one or both of the instruments needs to be changed so that they are unique to the instruments. To edit the instrument’s ID code, press: SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP X.X SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE EXIT SETUP X.X ID Toggle these keys to cycle through the available character set: 0-9 INET COMMUNICATIONS MENU COM1 SETUP X. 0 2 EXIT ENTR key accepts the new settings MACHINE ID: 200 ID 0 0 ENTR EXIT EXIT key ignores the new settings The ID can be any 4 digit number and can also be used to identify analyzers in any number of ways (e.g. location numbers, company asset number, etc.) 86 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.11.2. COM PORT DEFAULT SETTINGS As received from the factory, the analyzer is set up to emulate a DCE or modem, with pin 3 of the DB-9 connector designated for receiving data and pin 2 designated for sending data. COM1: RS-232 (fixed), DB-9 male connector. o Baud rate: 19200 bits per second (baud). o Data Bits: 8 data bits with 1 stop bit. o Parity: None. COM2: RS-232 (configurable), DB-9 female connector. o Baud rate: 115000 bits per second (baud). o Data Bits: 8 data bits with 1 stop bit. o Parity: None. CAUTION Cables that appear to be compatible because of matching connectors may incorporate internal wiring that make the link inoperable. Check cables acquired from sources other than Teledyne Instruments for pin assignments before using. 6.11.3. RS-232 COM PORT CABLE CONNECTIONS In its default configuration, the M200EH/EM analyzer has two available RS-232 Com ports accessible via 2 DB-9 connectors on the back panel of the instrument. The COM1 connector is a male DB-9 connector and the COM2 is a female DB9 connector. Female DB-9 (COM2) Male DB-9 (RS-232) (As seen from outside analyzer) (As seen from outside analyzer) TXD TXD GND RXD 1 2 6 3 7 4 8 5 GND RXD 1 9 6 CTS RTS 2 3 7 4 8 5 9 CTS RTS (DTE mode) (DTE mode) RXD GND TXD 1 2 6 3 7 4 8 5 9 RTS CTS (DCE mode) Figure 6-6-6: Back Panel connector Pin-Outs for COM1 & COM2 in RS-232 mode. 87 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual The signals from these two connectors are routed from the motherboard via a wiring harness to two 10-pin connectors on the CPU card, CN3 (COM1) and CN4 (COM2). CN3 & CN4 (Located on CPU card) CTS RTS RXD 2 4 6 8 10 1 3 5 7 9 TXD GND (As seen from inside analyzer) Figure 6-6-7: CPU connector Pin-Outs for COM1 & COM2 in RS-232 mode. Teledyne Instruments offers two mating cables, one of which should be applicable for your use. Part number WR000077, a DB-9 female to DB-9 female cable, 6 feet long. Allows connection of COM1 with the serial port of most personal computers. Also available as Option 60 (see Section 5.9.1). Part number WR000024, a DB-9 female to DB-25 male cable. Allows connection to the most common styles of modems (e.g. Hayes-compatible) and code activated switches. Both cables are configured with straight-through wiring and should require no additional adapters. To assist in properly connecting the serial ports to either a computer or a modem, there are activity indicators just above the RS-232 port. Once a cable is connected between the analyzer and a computer or modem, both the red and green LEDs should be on. If the lights for COM 1 are not lit, use small switch on the rear panel to switch it between DTE and DCE modes (see 16.10.5). If both LEDs are still not illuminated, check the cable for proper wiring. 88 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.11.4. RS-485 CONFIGURATION OF COM2 As delivered from the factory, COM2 is configured for RS-232 communications. This port can be re-configured for operation as a non-isolated, half-duplex RS-485 port capable of supporting up to 32 instruments with a maximum distance between the host and the furthest instrument being 4000 feet. If you require full-duplex or isolated operation, please contact Teledyne Instruments Customer Service. To reconfigure COM2 as an RS-285 port set switch 6 of SW1 to the ON position(see Figure 6-8). The RS-485 port can be configured with or without a 150 Ω termination resistor. To include the resistor, install jumper at position JP3 on the CPU board (see Figure 6-8). To configure COM2 as an unterminated RS-485 port leave JP3 open. CN4 JP3 COM2 – RS-232 CN3 COM1 – RS-232 CN5 COM2 – RS-485 SW1 Pin 6 Figure 6-6-8: CPU card Locations of RS-232/486 Switches, Connectors and Jumpers 89 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual When COM2 is configured for RS-485 operation the port uses the same female DB-9 connector on the back of the instrument as when Com2 is configured for RS-232 operation, however, the pin assignments are different. Female DB-9 (COM2) (As seen from outside analyzer) RX/TXGND RX/TX+ 1 2 6 3 7 4 8 5 9 (RS-485) Figure 6-6-9: Back Panel connector Pin-Outs for COM2 in RS-485 mode. The signal from this connector is routed from the motherboard via a wiring harness to a 6-pin connector on the CPU card, CN5. CN5 (Located on CPU card) RX/TXGND RX/TX+ 2 4 6 1 3 5 (As seen from inside analyzer) Figure 6-6-10: CPU connector Pin-Outs for COM2 in RS-485 mode. 6.11.5. DTE AND DCE COMMUNICATION RS-232 was developed for allowing communications between data terminal equipment (DTE) and data communication equipment (DCE). Basic terminals always fall into the DTE category whereas modems are always considered DCE devices. The difference between the two is the pin assignment of the Data Receive and Data Transmit functions. DTE devices receive data on pin 2 and transmit data on pin 3. DCE devices receive data on pin 3 and transmit data on pin 2. To allow the analyzer to be used with terminals (DTE), modems (DCE) and computers (which can be either), a switch mounted below the serial ports on the rear panel allows the user to set the configuration of COM1 for one of these two modes. This switch exchanges the receive and transmit lines on COM1 emulating a cross-over or null-modem cable. The switch has no effect on COM2. 90 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.11.6. ETHERNET CARD CONFIGURATION When equipped with the optional Ethernet interface, the analyzer can be connected to any standard 10BaseT Ethernet network via low-cost network hubs, switches or routers. The interface operates as a standard TCP/IP device on port 3000. This allows a remote computer to connect through the internet to the analyzer using APICOM, terminal emulators or other programs. The firmware on board the Ethernet card automatically sets the communication modes and baud rate (115 200 kBaud ) for the COM2 port. Once the Ethernet option is installed and activated, the COM2 submenu is replaced by a new submenu, INET. This submenu is used to manage and configure the Ethernet interface with your LAN or Internet Server(s). The card has four LEDs that are visible on the rear panel of the analyzer, indicating its current operating status. Table 6-11: Ethernet Status Indicators LED FUNCTION LNK (green) ON when connection to the LAN is valid. ACT (yellow) Flickers on any activity on the LAN. TxD (green) Flickers when the RS-232 port is transmitting data. RxD (yellow) Flickers when the RS-232 port is receiving data. 6.11.6.1. Ethernet Card COM2 Communication Modes and Baud Rate The firmware on board the Ethernet card automatically sets the communication modes for the COM2 port. The baud rate is also automatically set at 115 200 kBaud. 6.11.6.2. Configuring the Ethernet Interface Option using DHCP The Ethernet option for you M200EH/EM uses Dynamic Host Configuration Protocol (DHCP) to automatically configure its interface with your LAN. This requires your network servers also be running DHCP. The analyzer will do this the first time you turn the instrument on after it has been physically connected to your network. Once the instrument is connected and turned on it will appear as an active device on your network without any extra set up steps or lengthy procedures. Should you need to, the Ethernet configuration properties are viewable via the analyzer’s front panel See Table 6-12. 91 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual Table 6-12: LAN/Internet Configuration Properties PROPERTY DEFAULT STATE DESCRIPTION This displays whether the DHCP is turned ON or OFF. DHCP STATUS On Editable INSTRUMENT IP ADDRESS Configured by DHCP EDIT key disabled when DHCP is ON This string of four packets of 1 to 3 numbers each (e.g. 192.168.76.55.) is the address of the analyzer itself. Configured by DHCP EDIT key disabled when DHCP is ON A string of numbers very similar to the Instrument IP address (e.g. 192.168.76.1.)that is the address of the computer used by your LAN to access the Internet. GATEWAY IP ADDRESS Also a string of four packets of 1 to 3 numbers each (e.g. 255.255.252.0) that defines that identifies the LAN the device is connected to. SUBNET MASK TCP PORT1 HOST NAME 1 Configured by DHCP 3000 M200EH (EM) EDIT key disabled when DHCP is ON All addressable devices and computers on a LAN must have the same subnet mask. Any transmissions sent devices with different assumed to be outside of the LAN and are routed through gateway computer onto the Internet. Editable This number defines the terminal control port by which the instrument is addressed by terminal emulation software, such as Internet or Teledyne Instruments’ APICOM. Editable The name by which your analyzer will appear when addressed from other computers on the LAN or via the Internet. While the default setting for all Teledyne Instruments analyzers is the model number, the host name may be changed to fit customer needs. Do not change the setting for this property unless instructed to by Teledyne Instruments Customer Service personnel. 92 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions NOTE It is a good idea to check these settings the first time you power up your analyzer after it has been physically connected to the LAN/Internet to make sure that the DHCP has successfully downloaded the appropriate information from you network server(s). If the gateway IP, instrument IP and the subnet mask are all zeroes (e.g. “0.0.0.0”), the DCHP was not successful. You may have to manually configure the analyzer’s Ethernet properties. See your network administrator. To view the above properties, press: SAMPLE NOX=XXX.X SETUP X.X A1:NXCNC1=100PPM DHCP: ON SET> SETUP X.X SETUP X.X PRIMARY SETUP MENU SETUP X.X COMMUNICATIONS MENU TCP PORT: 3000 SET> SETUP X.X SETUP X.X SECONDARY SETUP MENU EDIT EDIT HOSTNAME: M200EH EDIT Do not alter unless directed to by Teledyne Instruments Customer Service personnel 93 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.11.6.3. Manually Configuring the Network IP Addresses There are several circumstances when you may need to manually configure the interface settings of the analyzer’s Ethernet card. The INET sub-menu may also be used to edit the Ethernet card’s configuration properties Your LAN is not running a DHCP software package, The DHCP software is unable to initialize the analyzer’s interface; You wish to program the interface with a specific set of IP addresses that may not be the ones automatically chosen by DHCP. Editing the Ethernet Interface properties is a two step process. STEP 1: Turn DHCP OFF: While DHCP is turned ON, the ability to manually set INSTRUMENT IP, GATEWAY IP and SUBNET MASK is disabled SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP X.X SETUP X.X ID INET EXIT 1 EXIT EXIT OFF Continue with editing of Ethernet interface properties (see Step 2, below). ENTR EXIT DHCP: ON EXIT DHCP: ON ON SETUP X.X COMMUNICATIONS MENU 8 EDIT SETUP X.X SECONDARY SETUP MENU COM1 ENTER SETUP PASS : 818 SETUP X.X PRIMARY SETUP MENU COMM VARS DIAG 8 SETUP CFG DAS RNGE PASS CLK MORE SETUP X.X SAMPLE ENTR EXIT DHCP: ON ENTR EXIT ENTR accept new settings EXIT ignores new settings 94 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions STEP 2: Configure the INSTRUMENT IP, GATEWAY IP and SUBNET MASK addresses by pressing: Internet Configuration Keypad Functions From Step 1 above) SETUP X.X DHCP: OFF SET> EDIT SETUP X.X EXIT FUNCTION [0] Press this key to cycle through the range of numerals and available characters (“0 – 9” & “ . ”) Moves the cursor one character left or right. DEL Deletes a character at the cursor location. ENTR Accepts the new setting and returns to the previous menu. EXIT Ignores the new setting and returns to the previous menu. Some keys only appear as needed. INST IP: 000.000.000.000 EDIT KEY EXIT SETUP X.X Cursor location is indicated by brackets INST IP: [0] 00.000.000 DEL [0] ENTR EXIT SETUP X.X GATEWAY IP: 000.000.000.000 EDIT EXIT SETUP X.X GATEWAY IP: [0] 00.000.000 DEL [?] ENTR EXIT SETUP X.X SUBNET MASK:255.255.255.0 EDIT EXIT SETUP X.X SUBNET MASK:[2]55.255.255.0 SETUP X.X TCP PORT 3000 EDIT DEL [?] ENTR EXIT EXIT The PORT number needs to remain at 3000. Do not change this setting unless instructed to by Teledyne Instruments Customer Service personnel. SETUP X.X SETUP X.X INITIALIZING INET 0% … INITIALIZING INET 100% INITIALIZATI0N SUCCEEDED SETUP X.X ID INET SETUP X.X INITIALIZATION FAILED Contact your IT Network Administrator COMMUNICATIONS MENU COM1 EXIT 95 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.11.6.4. Changing the Analyzer’s HOSTNAME The HOSTNAME is the name by which the analyzer appears on your network. The default name for all Teledyne Instruments Model 200EH/EME analyzers is M????. To change this name (particularly if you have more than one M200EH/EM analyzer on your network), press. SAMPLE A1:NXCNC1=100PPM < TST TST > CAL SET> SETUP CFG DAS RNGE PASS CLK MORE SECONDARY SETUP MENU COMMUNICATIONS MENU INET EDIT EXIT EXIT SETUP X.X ID HOSTNAME: 200E UNTIL … PRIMARY SETUP MENU SETUP X.X DHCP: ON SETUP X.X NOX=XXX.X COM1 HOSTNAME: [M]200E INS DEL [?] ENTR EXIT EXIT Use these keys (See Table 6-19) to edit HOSTNAME SAMPLE ENTER SETUP PASS : 818 SETUP X.X 8 1 8 ENTR HOSTNAME: 200E-FIELD1 EXIT Moves the cursor one character to the right. INS Inserts a character before the cursor location. DEL Deletes a character at the cursor location. [?] Press this key to cycle through the range of numerals and characters available for insertion. 0-9, A-Z, space ’ ~ ! # $ % ^ & * ( ) - _ = +[ ] { } < >\ | ; : , . / ? ENTR Accepts the new setting and returns to the previous menu. EXIT Ignores the new setting and returns to the previous menu. Some keys only appear as needed. 96 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.11.7. MULTIDROP RS-232 SET UP The RS-232 multidrop consists of a printed circuit assembly that plugs onto the CN3, CN4, and CN5 connectors of the CPU card (see Figure 6-11) and the cabling to connect it to the analyzer’s motherboard. This PCA includes all circuitry required to enable your analyzer for multidrop operation. It converts the instrument’s COM1 port to multidrop configuration allowing up to eight analyzers to be connected the same I/O port of the host computer. Because both of the DB9 connectors on the analyzer’s back panel are needed to construct the multidrop chain, COM2 is no longer available for separate RS-232 or RS-485 operation, however, with the addition of an Ethernet Option (option 63, see Sections 5.9.3 and 6.11.6) the COM2 port is available for communication over a 10BaseT LAN. JP2 Rear Panel CPU Card (as seen from inside) Cable to Ethernet Card Multidrop PCA Cable to Motherboard Figure 6-6-11: Location of JP2 on RS232-Multidrop PCA (option 62) Each analyzer in the multidrop chain must have: One Teledyne Instruments option 62 installed. One 6’ straight-through, DB9 male DB9 Female cable (Teledyne Instruments P/N WR0000101) is required for each analyzer. To set up the network, for each analyzer: 1. Turn the analyzer on and change its ID code (see Section 6.11.1) to a unique 4-digit number. 2. Remove the top cover (see Section 3.1) of the analyzer and locate JP2 on the multidrop PCA (see Figure 6-11) 3. Make sure that the jumpers are in place connection pins 9 10 and 11 12. 4. If the analyzer is to be the last instrument on the chain, make sure a jumper is in place connecting pins 21 22. 97 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 5. If you are adding an analyzer to the end of an already existing chain, don’t forget to remove JP2, pins 21 22 on the multidrop PCA on the analyzer that was previous the last instrument in the chain. 6. Close the instrument. 7. Using straight-through, DB9 male DB9 Female cables, interconnect the host and the analyzers as shown in Figure 6-12. NOTE: Teledyne Instruments recommends setting up the first link, between the Host and the first analyzer and testing it before setting up the rest of the chain. KEY: Host Female DB9 RS-232 port Male DB9 Analyzer Analyzer Analyzer Last Analyzer COM2 COM2 COM2 COM2 RS-232 RS-232 RS-232 RS-232 Make Sure Jumper between JP2 pins 21 22 is installed. Figure 6-6-12: RS232-Multidrop PCA Host/Analyzer Interconnect Diagram 98 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.11.8. COM PORT COMMUNICATION MODES Each of the analyzer’s serial ports can be configured to operate in a number of different modes, which are listed in the following table. Each COM port needs to be configured independently. Table 6-14: COMM Port Communication modes MODE1 ID DESCRIPTION 1 Quiet mode suppresses any feedback from the analyzer (iDAS reports, and warning messages) to the remote device and is typically used when the port is communicating with a computer program such as APICOM. Such feedback is still available but a command must be issued to receive them. COMPUTER 2 Computer mode inhibits echoing of typed characters and is used when the port is communicating with a computer program, such as APICOM. SECURITY 4 When enabled, the serial port requires a password before it will respond. The only command that is active is the help screen (? CR). HESSEN PROTOCOL 16 QUIET The Hessen communications protocol is used in some European countries. Teledyne Instruments part number 02252 contains more information on this protocol. When turned on this mode switches the COMM port settings from E, 7, 1 2048 No parity; 8 data bits; 1 stop bit to Even parity; 7 data bits; 1 stop bit RS-485 1024 Configures the COM2 Port for RS-485 communication. RS-485 mode has precedence over multidrop mode if both are enabled. MULTIDROP PROTOCOL 32 Multidrop protocol allows a multi-instrument configuration on a single communications channel. Multidrop requires the use of instrument IDs. ENABLE MODEM 64 Enables to send a modem initialization string at power-up. Asserts certain lines in the RS-232 port to enable the modem to communicate. ERROR CHECKING2 128 Fixes certain types of parity errors at certain Hessen protocol installations. XON/XOFF HANDSHAKE2 256 Disables XON/XOFF data flow control also known as software handshaking. HARDWARE HANDSHAKE 8 HARDWARE FIFO2 512 COMMAND PROMPT 4096 Enables CTS/RTS style hardwired transmission handshaking. This style of data transmission handshaking is commonly used with modems or terminal emulation protocols as well as by Teledyne Instrument’s APICOM software. Improves data transfer rate when on of the COMM ports. Enables a command prompt when in terminal mode. 1 Modes are listed in the order in which they appear in the SETUP MORE COMM COM[1 OR 2] MODE menu 2 The default sting for this feature is ON. Do not disable unless instructed to by Teledyne Instruments Customer Service personnel. 99 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual Press the following keys to select a communication mode for a one of the COMM Ports, such as the following example where HESSEN PROTOCOL mode is enabled: SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP PRIMARY SETUP MENU SETUP X.X CFG DAS RNGE PASS CLK MORE SECONDARY SETUP MENU SETUP X.X COMM VARS DIAG Select which COM port to configure SETUP X.X ID The sum of the mode IDs of the selected modes is displayed here ALRM EXIT EXIT returns to the previous menu COMMUNICATIONS MENU INET COM1 SETUP X.X SET> EXIT EXIT COM1 MODE:0 EDIT SETUP X.X EXIT COM1 QUIET MODE: OFF NEXT OFF ENTR EXIT Continue pressing next until … SETUP X.X Use PREV and NEXT keys to move between available modes. A mode is enabled by toggling the ON/OFF key. PREV NEXT SETUP X.X COM1 HESSEN PROTOCOL : OFF OFF ENTR EXIT COM1 HESSEN PROTOCOL : ON PREV NEXT ON ENTR EXIT ENTR key accepts the new settings EXIT key ignores the new settings Continue pressing the NEXT and PREV keys to select any other modes you which to enable or disable 100 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.11.9. COM PORT BAUD RATE To select the baud rate of one of the COM Ports, press: SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP PRIMARY SETUP MENU SETUP X.X CFG DAS RNGE PASS CLK MORE EXIT SECONDARY SETUP MENU SETUP X.X COMM VARS DIAG Select which COM port to configure. SETUP X.X COMMUNICATIONS MENU ID COM1 INET SETUP X.X Press SET> until you reach COM1 BAUD RATE EXIT SET> EXIT returns to the previous menu EXIT COM1 MODE:0 EDIT EXIT EXAMPLE Use PREV and NEXT keys to move between available baud rates. 300 1200 4800 9600 19200 38400 57600 115200 SETUP X.X COM1 BAUD RATE:19200 EDIT SETUP X.X PREV NEXT SETUP X.X NEXT ON EXIT EXIT key ignores the new setting COM1 BAUD RATE:19200 ENTR EXIT ENTR key accepts the new setting COM1 BAUD RATE:9600 ENTR EXIT 101 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.11.10. COM PORT TESTING The serial ports can be tested for correct connection and output in the COMM menu. This test sends a string of 256 ‘w’ characters to the selected COM port. While the test is running, the red LED on the rear panel of the analyzer should flicker. To initiate the test press the following key sequence. SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP X.X SETUP SETUP X.X SET> SETUP X.X EXIT SETUP X.X SETUP X.X EDIT PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X COM1 MODE:0 COM1 : TEST PORT TEST EXIT COMMUNICATIONS MENU COM1 EXIT SETUP X.X CAL SETUP X.X SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X EXIT SECONDARY SETUP MENU COMM VARS DIAG SETUP X.X EXIT EXIT ignores the new setting. ENTER VARS PASS: 818 ENTR accepts the new setting. 8 1 8 SETUP X.X ENTR EXIT 0 ) DAS_HOLD_OFF=15.0 Minutes SETUP X.X NEXT JUMP SETUP X.X 1 5 .0 ENTR EXIT Toggle this keys to change setting EDIT PRNT EXIT See Section 6.12.1. for information on setting the MEASRUE MODE 2 ) STABIL_GAS=NOX PREV NEXT JUMP EDIT PRNT EXIT SETUP X.X NO SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes 1 ) MEASURE_MODE=NOX-NO NEXT JUMP SETUP X.X EDIT PRNT EXIT NO2 2 ) STABIL GAS =NOX NOX O2 ENTR EXIT 3 ) TPC_ENABLE=ON PREV NEXT JUMP EDIT PRNT EXIT Choose Gas SETUP X.X 3 ) TPC_ENABLE=ON ON SETUP X.X ENTR EXIT 4 ) DYN_ZERO=ON PREV NEXT JUMP EDIT PRNT EXIT Toggle this keys to change setting SETUP X.X 4 ) DYN_ZERO=ON ON SETUP X.X ENTR EXIT 5) DYN_SPAN=ON PREV NEXT JUMP EDIT PRNT EXIT Toggle this keys to change setting SETUP X.X 5 ) DYN_SPAN=ON ON SETUP X.X ENTR EXIT Toggle this keys to change setting 6) CONC_PRECUISION : 1 PREV NEXT JUMP EDIT PRNT EXIT SETUP X.X AUTO 6) CONC_PRECUISION : 3 0 1 2 3 4 ENTR EXIT Toggle these keys to change setting SETUP X.X 7) CLOCK_ADJ=0 Sec/Day PREV NEXT JUMP EDIT PRNT EXIT SETUP X.X + 0 0 7) CLOCK_ADJ=0 Sec/Day ENTR EXIT Toggle this keys to change setting 104 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.12.1. SETTING THE GAS MEASUREMENT MODE In its standard operating mode the M200EH/EM measures NO, NO2 and NOx. It can be set to measure only NO or Only NOX s can be set to measure NO or NOx. To select one of these three measurement modes, press: SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP X.X SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X SECONDARY SETUP MENU COMM VARS DIAG SAMPLE 8 1 8 ENTR EXIT 0 ) DAS_HOLD_OFF=15 minutes NEXT JUMP SETUP X.X SETUP X.X EDIT PRNT EXIT 1 ) MEASURE_MODE=NOX-NO PREV NEXT JUMP Press the PREV and NEXT buttons to move back and forth between gas modes EXIT ENTER SETUP PASS : 818 SETUP X.X NOX-NO mode is the default mode for the M200EH/EM EXIT EDIT PRNT EXIT MEASURE MODE: NOX-NO PREV ENTR EXIT SETUP X.X NEXT ENTR accepts the new setting. MEASURE MODE: NOX PREV NEXT SETUP X.X EXIT ignores the new setting. ENTR EXIT MEASURE MODE: NO ENTR EXIT 105 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.13. SETUP MORE DIAG: DIAGNOSTICS MENU A series of diagnostic tools is grouped together under the SETUP-MORE-DIAG menu. These parameters are dependent on firmware revision. These tools can be used in a variety of troubleshooting and diagnostic procedures and are referred to in many places of the maintenance and trouble-shooting sections. An overview of the entire DIAG menu can be found in menu tree A-6 of Appendix A.1. Table 6-16: M200EH/EM Diagnostic (DIAG) Functions FRONT PANEL MODE INDICATOR SECTION DIAG I/O 6.13.2 ANALOG I/O: When entered, the analyzer performs an analog output step test. This can be used to calibrate a chart recorder or to test the analog output accuracy. DIAG AOUT 6.13.3 ANALOG I/O CONFIGURATION: This submenu allows the user to configure the analyzer’s four analog output channels, including choosing what parameter will be output on each channel. Instructions that appear here allow adjustment and calibration the voltage signals associated with each output as well as calibration of the analog to digital converter circuitry on the motherboard. DIAG AIO 6.13.4, through 6.13.6 DISPLAY SEQUENCE CONFIGURATION: Allows the user to program which concentration values are displayed in the . DIAG DISP 6.13.7.1 OPTIC TEST: When activated, the analyzer performs an optic test, which turns on an LED located inside the sensor module near the PMT (Fig. 10-15). This diagnostic tests the response of the PMT without having to supply span gas. DIAG OPTIC 6.13.7.2 ELECTRICAL TEST: When activated, the analyzer performs an electric test, which generates a current intended to simulate the PMT output to verify the signal handling and conditioning of the PMT preamp board. DIAG ELEC 6.13.7.3 DIAG OZONE 6.13.7.4 DIAG FCAL 6.13.7.5 DIAGNOSTIC FUNCTION AND MEANING SIGNAL I/O: Allows observation of all digital and analog signals in the instrument. Allows certain digital signals such as valves and heaters to be toggled ON and OFF. OZONE GEN OVERRIDE: Allows the user to manually turn the O3 generator on or off. This setting is retained when exiting DIAG. FLOW CALIBRATION: This function is used to calibrate the gas flow output signals of sample gas and ozone supply. These settings are retained when exiting DIAG. 106 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.13.1. ACCESSING THE DIAGNOSTIC FEATURES To access the DIAG functions press the following keys: SAMPLE A1:NXCNC1=100PPM NOX=XXX.X DIAG SETUP PREV < TST TST > CAL EXIT returns to the main SAMPLE display SETUP X.X EXIT returns to the PRIMARY SETUP MENU SETUP X.X CFG DAS RNGE PASS CLK MORE EXIT SETUP X.X 8 1 DIAG ENTR EXIT PREV PREV EXIT ANALOG OUTPUT NEXT NEXT ENTR EXIT ENTR EXIT ENTR EXIT OPTIC TEST NEXT ELECTRICAL TEST NEXT DIAG ENTR DIAG PREV ENTR DISPLAY SEQUENCE CONFIG. DIAG ENTER DIAG PASS: 818 SIGNAL I / O NEXT PREV EXIT 8 PREV NEXT DIAG SECONDARY SETUP MENU COMM VARS DIAG From this point forward, EXIT returns to the SECONDARY SETUP MENU DIAG PRIMARY SETUP MENU ANALOG I / O CONFIGURATION ENTR EXIT OZONE GEN OVERRIDE NEXT DIAG ENTR EXIT FLOW CALIBRATION EXIT PREV NEXT ENTR EXIT 107 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.13.2. SIGNAL I/O The signal I/O diagnostic mode allows to review and change the digital and analog input/output functions of the analyzer. See Appendix A-4 for a complete list of the parameters available for review under this menu. NOTE Any changes of signal I/O settings will remain in effect only until the signal I/O menu is exited. Exceptions are the ozone generator override and the flow sensor calibration, which remain as entered when exiting. To enter the signal I/O test mode, press: SAMPLE A1:NXCNC1=100PPM NOX=XXX.X SETUP < TST TST > CAL SETUP X.X PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE EXIT returns to the main SAMPLE display SETUP X.X EXIT SECONDARY SETUP MENU COMM VARS DIAG SETUP X.X 8 1 EXIT ENTER DIAG PASS: 818 8 ENTR EXIT SIGNAL I / O DIAG PREV NEXT JUMP DIAG I / O ENTR EXIT Test Signals Displayed Here PREV NEXT JUMP PRNT EXIT Use the NEXT & PREV keys to move between signal types. Use the JUMP key to go directly to a specific signal See Appendix A-4 for a complete list of available SIGNALS EXAMPLE DIAG I / O 0 JUMP TO: 5 5 ENTR EXIT DIAG I / O CAL_LED = ON PREV NEXT JUMP ON PRNT EXIT Enter 05 to Jump to Signal 5: (CAL_LED) Exit to return to the DIAG menu Pressing the PRNT key will send a formatted printout to the serial port and can be captured with a computer or other output device. 108 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.13.3. ANALOG OUTPUT STEP TEST This test can be used to check the accuracy and proper operation of the analog outputs. The test forces all four analog output channels to produce signals ranging from 0% to 100% of the full scale range in 20% increments. This test is useful to verify the operation of the data logging/recording devices attached to the analyzer. To begin the Analog Output Step Test press: SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP PRIMARY SETUP MENU SETUP X.X CFG DAS RNGE PASS CLK MORE SETUP X.X EXIT SECONDARY SETUP MENU COMM VARS DIAG SETUP X.X 8 1 EXIT ENTER DIAG PASS: 818 8 DIAG ENTR EXIT SIGNAL I / O NEXT ENTR DIAG ANALOG OUTPUT PREV NEXT DIAG AOUT EXIT ENTR [0%] EXIT Performs analog output step test. 0% - 100% EXIT Exit-Exit returns to the DIAG menu ANALOG OUTPUT 0% DIAG AOUT EXIT ANALOG OUTPUT Pressing the key under “0%” while performing the test will pause the test at that level. Brackets will appear around the value: example: [20%] Pressing the same key again will resume the test. 109 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.13.4. ANALOG OUTPUTS AND REPORTING RANGES 6.13.4.1. Analog Output Signals Available on the M200EH/EM The analyzer has four analog output signals, accessible through a connector on the rear panel. ANALOG OUT A1 + A2 - + - + A3 - A4 + - 0-20 mA current loop output available for these channels only Figure 6-6-13: Analog Output Connector Key The signal levels of each output can be independently configured as follows. An over-range feature is available that allows each range to be usable from -5% to + 5% of its nominal scale: Table 6-17: Analog Output Voltage Ranges with Over-Range Active RANGE MINIMUM OUTPUT MAXIMUM OUTPUT 0-0.1 V -5 mV +105 mV 0-1 V -0.05 V +1.05 V 0-5 V -0.25 V +5.25 V 0-10 V -0.5 V +10.5 V The default offset for all ranges is 0 VDC. Pin assignments for the ANALOG output connector at the rear panel of the instrument: Table 6-18: Analog Output Pin Assignments PIN 1 2 3 4 5 6 7 8 ANALOG OUTPUT A1 A2 A3 A4 VOLTAGE SIGNAL CURRENT SIGNAL V Out I Out + Ground I Out - V Out I Out + Ground I Out - V Out I Out + Ground I Out - V Out I Out + Ground I Out - 110 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions Additionally A1, A2 andA3 may be equipped with optional 0-20 mA current loop drivers. A4 is not available for the current loop option. Table 6-19: Analog Output Current Loop Range RANGE MINIMUM OUTPUT MAXIMUM OUTPUT 0-20 mA 0 mA 20 mA These are the physical limits of the current loop modules, typical applications use 2-20 or 4-20 mA for the lower and upper limits. Please specify desired range when ordering this option. The default offset for all ranges is 0 mA. All of these outputs can be configured output signals representing any of the iDAS parameters available on this model (See Table A-6 of Appendix A.5 for a complete list). The ability to select any one of the M200EH/EM’s 40+ iDAS data types coupled with the ability to select from a variety of signal ranges and scales makes the analog outputs of the M200EH/EM extremely flexible. Table 6-20: Example of Analog Output configuration for M200EH/EM OUTPUT IDAS PARAMETER ASSIGNED SIGNAL SCALE A1 NXCNC1 0-5 V A2 N2CNC2 4-20 mA A3 PMTDET 0-1V A4 O2CONC 0-10 V 1 With current loop option installed 6.13.4.2. Physical Range versus Analog Output Reporting Ranges The entire measurement range of the analyzer is quite large, 0 – 5,000 ppm for the M200Eh and 0-200 PPM for the M200EM, but many applications use only a small part of the analyzer’s full measurement range. This creates two performance challenges: 1. The width of the analyzer’s physical range can create data resolution problems for most analog recording devices. For example, in an application where a M200Eh is being used to measure an expected concentration of typically less than 200 ppm NOx, the full scale of expected values is only 4% of the instrument’s full 5000 ppm measurement range. Unmodified, the corresponding output signal would also be recorded across only 4% of the range of the recording device. The M200EH/EM solves this problem by allowing the user to select a scaled reporting range for the analog outputs that only includes that portion of the physical range relevant to the specific application. Only the reporting range of the analog outputs is scaled, the physical range of the analyzer and the readings displayed on the front panel remain unaltered. 2. Applications where low concentrations of NO, NO2 and NOx are measured require greater sensitivity and resolution than typically necessary for measurements of higher concentrations. The M200EH/EM solves this issue by using two hardware physical ranges that cover the instruments entire measurement range The analyzer’s software automatically selects which physical range is in effect based on the analog output reporting range selected by the user: 111 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual FOR THE M200EM: Low range spans 0 to 20 ppm NOX (20 ppm = 5 V); High range spans 0-200 ppm NOX (200 ppm = 5 V). If the high end of the selected reporting range is 20 ppm. The low physical range is selected. If the high end of the selected reporting range is > 20 ppm. The high physical range is selected. FOR THE M200EH: Low range spans 0 to 500 ppm NOX (500 ppm = 5 V); High range spans 0-5000 ppm NOX (5000 ppm = 5 V). If the high end of the selected reporting range is 500 ppm. The low physical range is selected. If the high end of the selected reporting range is > 500 ppm. The high physical range is selected. Once properly calibrated, the analyzer’s front panel will accurately report concentrations along the entire span of its 0 and 20,000 ppb physical range regardless of which reporting range has been selected for the analog outputs and which physical range is being used by the instruments software. Both reporting ranges need to be calibrated independently to the same span gas concentrations in order to allow switching back and forth between high and low ranges. 112 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.13.5. ANALOG I/O CONFIGURATION 6.13.5.1. The Analog I/O Configuration Submenu. Table 6-21 lists the analog I/O functions that are available in the M200EH/EM. Table 6-21: DIAG - Analog I/O Functions SUB MENU FUNCTION AOUTS CALIBRATED: Shows the status of the analog output calibration (YES/NO) and initiates a calibration of all analog output channels. DATA_OUT_1: Configures the A1 analog output: RANGE1: Selects the signal type (voltage or current loop) and full scale value of the output. OVERRANGE: Turns the ± 5% over-range feature ON/OFF for this output channel. REC_OFS1: Sets a voltage offset (not available when RANGE is set to CURRent loop. AUTO_CAL1: Sets the channel for automatic or manual calibration CALIBRATED1: Performs the same calibration as AOUT CALIBRATED, but on this one channel only. OUTOUT: Turns the output channel ON/OFF. A signal. Equal to the low end of the output scale (zero point) is still output by the analyzer, but no data is sent. DATA: Allows the user to select which iDAS parameter to be output. SCALE: Sets the top end of the reporting range scale for this channel. The analyzer automatically chooses the units of measure appropriate for the iDAS parameter chosen (e.g. ppm for concentration parameters; in-Hg-A for pressure measurements, etc.) UPDATE: Sets the time interval at which the analyzer updates the data being output on the channel. DATA_OUT_2 Same as forDATA_OUT_1 but for analog channel 2 (NO) DATA_OUT_3 Same as for DATA_OUT_1 but for analog channel 3 (NO2) DATA_OUT_4 Same as for DATA_OUT_1 but for analog channel 3 (NO2) TEST OUTPUT Same as for DATA_OUT_1 but for analog channel 4 (TEST) AIN CALIBRATED 1 Shows the calibration status (YES/NO) and initiates a calibration of the analog to digital converter circuit on the motherboard. Changes to RANGE or REC_OFS require recalibration of this output. To configure the analyzer’s four analog outputs, set the electronic signal type of each channel and calibrate the outputs. This consists of: 1. Selecting an output type (voltage or current, if an optional current output driver has been installed) and the signal level that matches the input requirements of the recording device attached to the channel. 2. Determine if the over-range feature is needed and turn it on or off accordingly. 3. If a Voltage scale is in use, a bipolar recorder offset may be added to the signal if required (Section 6.13.4.4). 4. Choose an iDAS parameter to be output on the channel. 5. Set the reporting range scale for the data type chosen. 6. Set the update rate for the channel. 113 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 7. Calibrating the output channel. This can be done automatically or manually for each channel (see Sections 6.13.5). To access the analog I/O configuration sub menu, press: SAMPLE A1:NXCNC1=100PPM < TST TST > NOX=XXX.X CAL SETUP DIAG AIO A OUTS CALIBRATED: NO SETUP X.X CAL EXIT PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE EXIT DIAG AIO DATA_OUT_1: 5V, NXCNC1, NOCAL EDIT SETUP X.X COMM EXIT SECONDARY SETUP MENU VARS DIAG ALRM EXIT DIAG AIO DATA_OUT_2: 5V, NXCNC1, NOCAL EDIT SETUP X.X 8 1 8 ENTR EXIT DIAG AIO DIAG ENTR DATA_OUT_4: 5V, NXCNC1, NOCAL EDIT Continue pressing NEXT until ... AIO Configuration Submenu DIAG AIO ANALOG I/O CONFIGURATION ENTR EXIT EXIT DIAG AIO PREV NEXT DATA_OUT_3: 5V, NXCNC1, NOCAL EDIT SIGNAL I/O NEXT DIAG EXIT ENTER PASSWORD:818 EXIT AIN CALIBRATED: NO CAL EXIT EXIT 114 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.13.5.2. Analog Output Signal Type and Range Selection To select an output signal type (DC Voltage or current) and level for one output channel press: From the AIO CONFIGURATION SUBMENU (See Section 6.13.4.1) DIAG ANALOG I/O CONFIGURATION PREV NEXT DIAG AIO ENTR EXIT AOUTS CALIBRATED: NO SET> CAL EXIT Continue pressing SET> until you reach the output to be configured DIAG AIO DATA_OUT_3: 5V, NXCNC1, NOCAL EDIT These keys set the signal level and type of the selected channel DIAG AIO 0.1V EXIT DATA_OUT_3: RANGE: 5V 1V 5V 10V CURR ENTR EXIT Pressing ENTR records the new setting and returns to the previous menu. Pressing EXIT ignores the new setting and returns to the previous menu. 115 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.13.5.3. Turning the Analog Output Over-Range Feature ON/OFF In its default configuration a ± 5% over-range is available on each of the M200EH/EM’s analog output channels. This over-range can be disabled if your recording device is sensitive to excess voltage or current. NOTE: Instruments with current range options installed on one or more of the outputs often are delivered from the factory with the over-range feature turned OFF on those channels. To Turn the over-range feature on or off, press: From the AIO CONFIGURATION SUBMENU (See Section 6.13.4.1) DIAG ANALOG I/O CONFIGURATION PREV NEXT DIAG AIO SET> ENTR EXIT AOUTS CALIBRATED: NO CAL EXIT Continue pressing SET> until you reach the output to be configured DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL EDIT DIAG AIO DATA_OUT_2 RANGE: 5V SET> EDIT DIAG AIO DIAG AIO ON DIAG AIO OFF EXIT DATA_OUT_2 OVERRANGE: ON EDIT Toggle this key to turn the Over-Range feature ON/OFF EXIT EXIT DATA_OUT_2 OVERRANGE: ON ENTR EXIT DATA_OUT_2 OVERRANGE: OFF ENTR EXIT 116 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.13.5.4. Adding a Recorder Offset to an Analog Output Some analog signal recorders require that the zero signal is significantly different from the baseline of the recorder in order to record slightly negative readings from noise around the zero point. This can be achieved in the M200EH/EM by defining a zero offset, a small voltage (e.g., 10% of span). To add a zero offset to a specific analog output channel, press: From the AIO CONFIGURATION SUBMENU (See Section 6.13.4.1) DIAG ANALOG I/O CONFIGURATION PREV NEXT DIAG AIO ENTR EXIT AOUTS CALIBRATED: NO SET> CAL EXIT Continue pressing SET> until you reach the output to be configured DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL EDIT DIAG AIO EXIT DATA_OUT_2 OUTPUT: 5V SET> EDIT EXIT Continue pressing SET> until ... DIAG AIO DATA_OUT_2 REC OFS: 0 mV EDIT Toggle these keys to set ther value of the desired offset. DIAG AIO + EXIT DATA_OUT_2 REC OFS: 0 mV 0 0 0 0 ENTR EXIT EXAMPLE DIAG AIO – DIAG AIO DATA_OUT_2 REC OFS: -10 mV 0 0 1 0 ENTR EXIT DATA_OUT_2 REC OFS: -10 mV EDIT EXIT 117 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.13.5.5. Assigning an iDAS parameter to an Analog Output Channel The M200EH/EM analog output channels can be assigned to output data from any of the 40+ available iDAS parameters (see Table A-6 of Appendix A.5). The default settings for the four output channels are: Table 6-22: Analog Output Data Type Default Settings PARAMETER DATA TYPE 1 CHANNEL DEFAULT SETTING A1 A2 NXCNC1 NOCNC1 A3 A43 N2CNC1 NXCNC2 2 RANGE 0 - 5 VDC REC OFS 0 mVDC AUTO CAL. ON CALIBRATED NO OUTPUT ON SCALE 100 ppm UPDATE 5 sec 1 See Table A-6 of M200EH/EM Appendix A for definitions of these iDAS data types 2 Optional current loop outputs are available for analog output channels A1-A3. 3 On analyzers with O2 sensor options installed, iDAS parameter O2CONC is assigned to output A4. REPORTING GAS CONCENTRATIONS VIA THE M200EH/EM ANALOG OUTPUT CHANNELS While the iDAS parameters available for output over via the analog channels A1 thru A4 include a vide variety internal temperatures, gas flows and pressures as well as certain key internal voltage levels, most of the iDAS parameters are related to gas concentration levels. Two parameters exist for each gas type measured by the M200EH/EM. They are generally referred to as range 1 and range 2 (e.g. NXCNC1 and NXCNC2; NOCNC1 and NOCNC2; etc.). These take the place of the high and low concentration ranges of previous versions of the analyzer software. Concentrations for each range are computed using separate slopes and offsets which are also stored via separate iDAS parameters. NOTE If an analog output channel is set to report a gas concentration (e.g. NXCNC1; NOx concentration; Range 1) it is generally a good idea to use 80% of the reporting range for that channel for the span point calibration. If both available parameters for a specific gas type are being reported (e.g. NXCNC1 and NXCNC2) separate Calibrations should be carried out for each parameter. 118 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions The available gas concentration iDAS parameters for output via the M200EH/EM analog output channels are: Table 6-23: Analog Output iDAS Parameters Related to Gas Concentration Data REPORTING RANGE PARAMETER NAME1 DESCRIPTION NXCNC1 Concentration NXSLP1 Slope NXOFS1 Offset NXZSC1 Concentration during calibration, prior to computing new slope and offset NXCNC2 Concentration NXSLP2 Slope NXOFS2 Offset NXZSC2 Concentration during calibration, prior to computing new slope and offset NOCNC1 Concentration NOSLP1 Slope NOOFS1 Offset NOZSC1 Concentration during calibration, prior to computing new slope and offset NOCNC2 Concentration NOSLP2 Slope NOOFS2 Offset NOZSC2 Concentration during calibration, prior to computing new slope and offset NO2 Range 12 (LOW) N2CNC1 Concentration - Computed with data from NOx Range 1 and NO Range 1 N2ZSC1 Concentration during calibration, prior to computing new slope and offset NO2 RANGE 22 (HIGH) N2CNC2 Concentration - Computed with data from NOx Range 2 and NO Range 2 N2ZSC2 Concentration during calibration, prior to computing new slope and offset NOx Range 1 (LOW) NOx RANGE 2 (HIGH) NO Range 1 (LOW) NO RANGE 2 (HIGH) 3 O2 Range3 O2CONC Concentration O2OFST3 Slope 3 Offset 3 Concentration during calibration, prior to computing new slope and offset O2SLPE O2ZSCN 1 Parameters are not listed in the order they appear on the iDAS list (see Table A-6 or Appendix A.5 for the proper order of the full list of parameters) 2 Since NO2 values are computed rather than measured directly, no separate slope or offset exist. 3 Only available on instruments with O2 sensor options installed. 119 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual To assign an iDAS parameter to a specific analog output channel, press, From the AIO CONFIGURATION SUBMENU (See Section 6.13.4.1) DIAG ANALOG I/O CONFIGURATION PREV NEXT DIAG AIO ENTR EXIT AOUTS CALIBRATED: NO SET> CAL EXIT Continue pressing SET> until you reach the output to be configured DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL EDIT DIAG AIO EXIT DATA_OUT_2 OUTPUT: 5V SET> EDIT EXIT Continue pressing SET> until ... DIAG AIO DATA_OUT_2 DATA: NOCNC1 EDIT DIAG AIO EXIT DATA_OUT_2 DATA: NOCNC1 PREV NEXT ENTR EXIT Use these keys to move up and down the list if available iDAS parameters (See Table A-6 of Appendix A.5) EXAMPLE DIAG AIO DIAG AIO DATA_OUT_2 DATA: STABIL INS DEL [1] ENTR EXIT DATA_OUT_2 DATA: STABIL EDIT EXIT 120 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.13.5.6. Setting the Reporting Range Scale for an Analog Output Once the iDAS parameter has been set, the top end of the scale must be selected. For concentration values this should be equal to the expected maximum value for the application. The analog channel will scale its output accordingly. EXAMPLE: IDAS parameter being output: NXCNC1 Maximum value expected: 800 ppm Output range; 10 V Output: 0 ppm......... 0.000 V 100 ppm...... 1.250 V 200 ppm...... 2.500 V 400 ppm...... 5.000 V 750 ppm...... 9.375 V NOTE Regardless of how the reporting range for an analog output channel is set, the instrument will continue to measure NO, NO2 and NOx accurately for the entire physical range of the instrument (See Section 6.13.3.2 for information on Physical range versus reporting range) Each output channel can be programmed for a separate gas with independent reporting range. EXAMPLE: A1 NXCNC1 (NOx Range 1) 0-1000 ppm NOX. A1 NXCNC2 (NOx Range 2) 0-1250 ppm NOX. A3 NOCNC1 (NOx Range 1) 0-500 ppm NO. A4 N2CNC1 (NO2 Range 1) 0-750 ppm NO2. NOTE: While Range 1 for each gas type is often referred to as the LOW range and Range 2 as the HIGH range, this is simply a naming convention. The upper limit for each range can be set to any value. EXAMPLE: A1 NXCNC1 (NOx Range 1) 0-1500 ppm NOX A2 NXCNC2 (NOx Range 2) 0-1000 ppm NOX. 121 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual To set the reporting range for an analog output, press: From the AIO CONFIGURATION SUBMENU (See Section 6.13.4.1) DIAG ANALOG I/O CONFIGURATION PREV NEXT DIAG AIO SET> ENTR EXIT DIAG AIO AOUTS CALIBRATED: NO CAL DATA_OUT_2 OUTPUT: 5V SET> EDIT EXIT Continue pressing SET> until ... Continue pressing SET> until you reach the output to be configured DIAG AIO DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL EDIT EXIT DATA_OUT_2 SCALE: 100.00 PPM EDIT EXIT EXIT DIAG AIO INS DEL [1] ENTR EXIT EXAMPLE DIAG AIO DIAG AIO INS DEL [1] ENTR EXIT DATA_OUT_2 SCALE: 1250.00 PPM EDIT EXIT RANGE SELECTION KEYPAD FUNCTIONS KEY FUNCTION Moves the cursor one character to the right. INS Inserts a character before the cursor location. DEL Deletes a character at the cursor location. [?] Press this key to cycle through the range of numerals and characters available for insertion: 0-9; as well as “+” & “-“. ENTR Accepts the new setting and returns to the previous menu. EXIT Ignores the new setting and returns to the previous menu. Some keys only appear as needed. 122 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.13.5.7. Setting Data Update Rate for an Analog Output The data update rate for the M200EH/EM analog outputs can be adjusted to match the requirements of the specific iDAS parameter chosen for each channel. For instance, if the parameter NXCNC1 (NOx concentration; Range 1) is chosen for channel A1 on an instrument set for dual gas measurement mode, it would be meaningless to have an update rate of less than 30 seconds, since the NOx-No measurement cycle takes that long to complete. On the other hand, if the channel was set to output the PMTDET voltage or the temperature of the moly converter it might be useful to have output updated more frequently. To change the update rate for an individual analog output channel, press: From the AIO CONFIGURATION SUBMENU (See Section 6.13.4.1) DIAG ANALOG I/O CONFIGURATION PREV NEXT DIAG AIO ENTR EXIT AOUTS CALIBRATED: NO SET> CAL EXIT Continue pressing SET> until you reach the output to be configured DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL EDIT DIAG AIO EXIT DATA_OUT_2 OUTPUT: 5V SET> EDIT EXIT Continue pressing SET> until ... DIAG AIO DATA_OUT_2 UPDATE: 5 SEC EDIT Toggle these keys to set the data update rate for this channel. DIAG AIO 0 EXIT DATA_OUT_2 UPDATE: 5 SEC 0 5 ENTR EXIT EXAMPLE DIAG AIO 0 DIAG AIO DATA_OUT_2 UPDATE: 30 SEC 3 0 ENTR EXIT DATA_OUT_2 UPDATE: 30 SEC EDIT EXIT 123 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.13.5.8. Turning an Analog Output On or Off Each output can be temporarily turned off. When off no data is sent to the output. Electronically, it is still active, there is simply no data so the signal level at the rear of the instrument will fall to zero. To turn an individual analog output channel ON/OFF, press: From the AIO CONFIGURATION SUBMENU (See Section 6.13.4.1) DIAG ANALOG I/O CONFIGURATION PREV NEXT DIAG AIO SET> ENTR EXIT AOUTS CALIBRATED: NO CAL EXIT Continue pressing SET> until you reach the output to be configured DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL EDIT DIAG AIO EXIT DATA_OUT_2 OUTPUT: 5V SET> EDIT EXIT Continue pressing SET> until ... DIAG AIO DATA_OUT_2 OUTPUT: ON EDIT Toggle this key to turn the channel ON/OFF DIAG AIO ON DIAG AIO OFF EXIT DATA_OUT_2 OUTPUT: ON ENTR EXIT DATA_OUT_2 OUTPUT: OFF ENTR EXIT 124 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.13.6. ANALOG OUTPUT CALIBRATION Analog calibration needs to be carried out on first startup of the analyzer (performed in the factory as part of the configuration process) or whenever re-calibration is required. The analog outputs can be calibrated automatically or adjusted manually (see Section 6.13.5). During automatic calibration the analyzer tells the output circuitry to generate a zero mV signal and high-scale point signal (usually about 90% of chosen analog signal scale) then measures actual signal of the output. Any error at zero or highscale is corrected with a slope and offset. Automatic calibration can be performed a group via the AOUTS CALIBRATION command, or individually by using the CAL button located inside each output channels submenu. By default, the analyzer is configured so that calibration of all four of the outputs can be initiated with the AOUTS CALIBRATION command. To enable or disable the Auto-Cal feature for one output channel, press. From the AIO CONFIGURATION SUBMENU (See Section 6.13.4.1) DIAG ANALOG I/O CONFIGURATION PREV NEXT DIAG AIO SET> ENTR EXIT AOUTS CALIBRATED: NO CAL EXIT Continue pressing SET> until you reach the output to be configured DIAG AIO DATA_OUT_3: 5V, NXCNC1, NOCAL EDIT DIAG AIO EXIT DATA_OUT_3 RANGE: 5V SET> EDIT EXIT Continue pressing SET> until ... DIAG AIO DATA_OUT_3 AUTO CAL.:ON EDIT Toggle this key to turn AUTO CAL ON or OFF DIAG AIO EXIT DATA_OUT_3 AUTO CAL.:ON ON ENTR EXIT (OFF = manual calibration mode). DIAG AIO ENTR accepts the new setting. EXIT ignores the new setting DATA_OUT_3 AUTO CAL.:OFF OFF ENTR EXIT NOTE: Channels with current loop output options cannot be calibrated automatically. Outputs Configured for 0.1V full scale should always be calibrated manually. 125 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.13.6.1. Automatic Analog Output Calibration To calibrate the outputs as a group with the AOUTS CALIBRATION command, press: From the AIO CONFIGURATION SUBMENU (See Section 6.13.4.1) DIAG ANALOG I/O CONFIGURATION PREV NEXT ENTR DIAG AIO AOUTS CALIBRATED: NO SET> CAL DIAG AIO Analyzer automatically calibrates all channels for which AUTO-CAL is turned ON EXIT AUTO CALIBRATING DATA_OUT_1 DIAG AIO AUTO CALIBRATING DATA_OUT_2 DIAG AIO NOT AUTO CAL. DATA_OUT_3 DIAG AIO If any of the channels have not been calibrated ot if at least one channel has AUTO-CAL turned OFF, this message will read NO. EXIT DIAG AIO AUTO CALIBRATING DATA_OUT_4 This message appears when AUTO-CAL is Turned OFF for a channel AOUTS CALIBRATED: YES SET> CAL EXIT NOTE: Manual calibration should be used for the 0.1V range or in cases where the outputs must be closely matched to the characteristics of the recording device. To initiate an automatic calibration for an individual output channel, press: From the AIO CONFIGURATION SUBMENU (See Section 6.13.4.1) DIAG ANALOG I/O CONFIGURATION PREV NEXT DIAG AIO SET> ENTR EXIT AOUTS CALIBRATED: NO CAL EXIT DIAG AIO DATA_OUT_2 CALIBRATED:NO CAL EXIT Continue pressing SET> until you reach the output to be configured DIAG AIO DIAG AIO AUTO CALIBRATING DATA_OUT_2 DATA_OUT_2 5V, NXCNC1, NOCAL EDIT EXIT DIAG AIO DIAG AIO DATA_OUT_2 RANGE: 5V SET> EDIT DATA_OUT_2 CALIBRATED: YES CAL EXIT EXIT Continue pressing SET> until ... 126 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.13.6.2. Manual Calibration of Analog Output configured for Voltage Ranges For highest accuracy, the voltages of the analog outputs can be manually calibrated. NOTE: The menu for manually adjusting the analog output signal level will only appear if the AUTO-CAL feature is turned off for the channel being adjusted (see Section 6.13.5.1) Calibration is performed with a voltmeter connected across the output terminals (See Figure 6-14) and by changing the actual output signal level using the front panel keys in 100, 10 or 1 count increments. See Table 3-1 for pin assignments of Analog Out connector on the rear panel V +DC Gnd V OUT + V IN + V OUT - V IN - Recording Device ANALYZER Figure 6-6-14: Setup for Calibrating Analog Outputs Table 6-24: Voltage Tolerances for Analog Output Calibration FULL SCALE ZERO TOLERANCE SPAN VOLTAGE SPAN TOLERANCE MINIMUM ADJUSTMENT (1 count) 0.1 VDC ±0.0005V 90 mV ±0.001V 0.02 mV 1 VDC ±0.001V 900 mV ±0.001V 0.24 mV 5 VDC ±0.002V 4500 mV ±0.003V 1.22 mV 10 VDC ±0.004V 4500 mV ±0.006V 2.44 mV 127 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual To manually adjust the signal levels of an analog output channel, press: From the AIO CONFIGURATION SUBMENU (See Section 6.13.4.1) DIAG ANALOG I/O CONFIGURATION PREV NEXT ENTR EXIT DIAG AIO DIAG AIO SET> AOUTS CALIBRATED: NO CAL Continue pressing SET> until ... DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL EDIT Continue adjustments until the voltage measured at the output of the analyzer and/or the input of the recording device matches the value in the upper right hand corner of the display (within the tolerances listed in Table 6-24. DATA_OUT_2 CALIBRATED:NO CAL EXIT EXIT DIAG AIO These keys increase / decrease the analog output signal level (not the value on the display) by 100, 10 or 1 counts. EXIT EXIT Continue pressing SET> until you reach the output to be configured DIAG AIO DATA_OUT_2 RANGE: 5V SET> EDIT DATA_OUT_2 VOLT-Z: 0 mV U100 UP10 UP DIAG AIO DATA_OUT_2 VOLT-S: 4500 mV U100 UP10 UP DIAG AIO DOWN DN10 D100 ENTR EXIT These menu’s only appear if AUTO-CAL is turned OFF DOWN DN10 D100 ENTR EXIT DATA_OUT_2 CALIBRATED: YES CAL EXIT 6.13.6.3. Manual Calibration of Analog Outputs configured for Current Loop Ranges The current loop output option (see Section 5.4) uses a small converter assembly to change the DC voltage output by the standard voltage output to a current signal ranging between 0-20 mA. Since the exact current increment per voltage count varies from converter to converter and from instrument to instrument, analog outputs with this option installed cannot be calibrated automatically and must be adjusted manually. Adjusting the signal zero and full scale values of the current loop output is done in a similar manner as manually adjusting analog outputs configured for voltage output except that: In this case calibration is performed with a current meter connected in series with the output circuitry (See Figure 6-6-15). Adjustments to the output are made using the front panel keys, also in 100, 10 or 1 count increments, but the change in the voltage driving the converter assembly is displayed on the front panel. As before, adjustment of the output is performed until the current reading of the meter reaches the desired point (e.g. 2 mA, 4 mA, 20 mA, etc.) 128 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual See Table 3-2 for pin assignments of the Analog Out connector on the rear panel. Operating Instructions mA Current Meter IN OUT V OUT + I IN + V OUT - I IN - Recording Device Analyzer Figure 6-6-15: Setup for Calibrating Current Outputs CAUTION Do not exceed 60 V between current loop outputs and instrument ground. If a current meter is not available, an alternative method for calibrating the current loop outputs is to connect a 250 1% resistor across the current loop output. Using a voltmeter, connected across the resistor, follow the procedure above but adjust the output to the following values: V +DC Gnd V OUT + Volt Meter V IN + 250 O V OUT - V IN - ANALYZER Recording Device Figure 6-6-16: Alternative Setup for Calibrating Current Outputs 129 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual Table 6-25: Current Loop Output Calibration with Resistor FULL SCALE VOLTAGE FOR 2-20 MA (measured across 250Ω resistor) VOLTAGE FOR 4-20 MA (measured across 250Ω resistor) 0% 0.5 V 1.0 V 100% 5.0 V 5.0 V To adjust the zero and span values of the current outputs, press: From the AIO CONFIGURATION SUBMENU (See Section 6.13.4.1) DIAG ANALOG I/O CONFIGURATION PREV NEXT DIAG AIO SET> ENTR EXIT AOUTS CALIBRATED: NO CAL EXIT DIAG AIO U100 UP10 Continue pressing SET> until you reach the output to be configured DIAG AIO DATA_OUT_2 CURR-Z: 13 mV UP DOWN DN10 D100 ENTR EXIT DATA_OUT_2: CURR, NXCNC1, NOCAL EDIT EXIT DIAG AIO U100 UP10 DIAG AIO DOWN DN10 D100 ENTR EXIT EXAMPLE U100 UP10 DIAG AIO DATA_OUT_2 CURR-Z: 0 mV UP DATA_OUT_2 CURR-S: 5000 mV UP Increase or decrease the current output by 100, 10 or 1 counts. The resulting change in output voltage is displayed in the upper line. Continue adjustments until the correct current is measured with the current meter. DOWN DN10 D100 ENTR EXIT DATA_OUT_2 RANGE: CURR SET> EDIT EXIT EXAMPLE DIAG AIO U100 UP10 DATA_OUT_2 CURR-S: 4866 mV UP DOWN DN10 D100 ENTR EXIT Continue pressing SET> until ... DIAG AIO DIAG AIO DATA_OUT_2 CALIBRATED:NO CAL DATA_OUT_2 CALIBRATED: YES CAL EXIT EXIT 130 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.13.6.4. AIN Calibration This is the sub-menu calibrates the analyzer’s A-to-D conversion circuitry. This calibration should only be necessary after major repair such as a replacement of CPU, motherboard or power supplies. To perform a AIN CALIBRATION, press: From the AIO CONFIGURATION SUBMENU (See Section 6.13.4.1) DIAG ANALOG I/O CONFIGURATION PREV NEXT DIAG AIO ENTR EXIT AOUTS CALIBRATED: NO SET> CAL EXIT Continue pressing SET> until …. DIAG AIO NOX=XXX.X CAL SETUP DIAG DISP SETUP X.X CFG PRIMARY SETUP MENU DAS RNGE PASS CLK MORE SETUP X.X COMM SETUP X.X 8 ALRM EXIT 8 DIAG EDIT ENTR EXIT Moves back and forth along existing list of display values ENTR EXIT DIAG DISP SIGNAL I/O NEXT DEL INSERT adds a new entry on the display list before the currently selected value. ENTER PASSWORD:818 1 4 SEC INS EXIT SECONDARY SETUP MENU VARS DIAG 1) NOX, PREV NEXT ENTR EXIT DISPLAY DATA: NOX PREV NEXT ENTR EXIT Toggle PREV and NEXT keys until desired display value appears (See Table Continue pressing NEXT until ... DIAG DISP DIAG DISPLAY SEQUENCE CONFIG. PREV NEXT ENTR PREV NEXT 6-23 ). DISPLAY DATA: N2H ENTR EXIT EXIT DIAG DISP 0 4 DISPLAY DURATION: 4 SEC ENTR Accepts the new setting. EXIT discards the new setting. ENTR EXIT Toggle these keys to set desired display duration in seconds 133 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual To delete an entry in the display value list, press: SAMPLE A1:NXCNC1=100PPM < TST TST > SETUP X.X NOX=XXX.X CAL SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X COMM SECONDARY SETUP MENU VARS DIAG SETUP X.X 8 EXIT ALRM EXIT ENTER PASSWORD:818 1 8 DIAG ENTR EXIT SIGNAL I/O NEXT ENTR EXIT Continue pressing NEXT until ... DIAG DISPLAY SEQUENCE CONFIG. PREV NEXT DIAG DISP PREV NEXT Moves back and forth along existing list of display values DIAG DISP YES ENTR 1) NOX, EXIT 4 SEC INS DEL EDIT ENTR EXIT DELETE? NO DIAG DISP DELETED 134 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.13.7.2. Optic Test The optic test function tests the response of the PMT sensor by turning on an LED located in the cooling block of the PMT (Fig. 10-15). The analyzer uses the light emitted from the LED to test its photo-electronic subsystem, including the PMT and the current to voltage converter on the pre-amplifier board. To make sure that the analyzer measures only the light coming from the LED, the analyzer should be supplied with zero air. The optic test should produce a PMT signal of about 2000±1000 mV. To activate the electrical test press the following key sequence. SAMPLE RANGE = 500.0 PPB NOX=X.X < TST TST > CAL SETUP X.X SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X SECONDARY SETUP MENU COMM VARS DIAG ALRM SETUP X.X 8 1 EXIT EXIT ENTER DIAG PASS: 818 8 ENTR EXIT DIAG SIGNAL I / O PREV NEXT JUMP ENTR EXIT Press NEXT until… DIAG OPTIC TEST PREV NEXT DIAG OPTIC ENTR EXIT A1:NXCNC1=100PPM NOX=XXX.X EXIT Press TST until… While the optic test is activated, PMT should be 2000 mV ± 1000 mV DIAG ELEC PMT = 2751 MV NOX=X.X EXIT NOTE This is a coarse test for functionality and not an accurate calibration tool. The resulting PMT signal can vary significantly over time and also changes with low-level calibration. 135 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.13.7.3. Electrical Test The electrical test function creates a current, which substitutes the PMT signal, and feeds it into the preamplifier board. This signal is generated by circuitry on the pre-amplifier board itself and tests the filtering and amplification functions of that assembly along with the A/D converter on the motherboard. It does not test the PMT itself. The electrical test should produce a PMT signal of about 2000 ±1000 mV. To activate the electrical test press the following keys. SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP X.X SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X EXIT SECONDARY SETUP MENU COMM VARS DIAG SETUP X.X 8 1 EXIT ENTER DIAG PASS: 818 8 ENTR EXIT DIAG SIGNAL I / O PREV NEXT JUMP ENTR EXIT Press NEXT until… DIAG ELECTRICAL TEST PREV NEXT DIAG ELEC ENTR EXIT A1:NXCNC1=100PPM NOX=XXX.X EXIT Press TST until… While the electrical test is activated, PMT should equal: DIAG ELEC PMT = 1732 MV NOX=X.X 2000 mV ± 1000 mV EXIT 136 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.13.7.4. Ozone Generator Override This feature allows the user to manually turn the ozone generator off and on. This can be done before disconnecting the generator, to prevent ozone from leaking out, or after a system restart if the user does not want to wait for 30 minutes during warm-up time. Note that this is one of the two settings in the DIAG menu that is retained after you exit the menu. To access this feature press the following keys. Also note that the ozone generator does not turn on if the ozone flow conditions are out of specification (e.g., if there is no flow through the system or the pump is broken). SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP X.X SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X EXIT SECONDARY SETUP MENU COMM VARS DIAG SETUP X.X 8 1 EXIT ENTER DIAG PASS: 818 8 ENTR EXIT DIAG SIGNAL I / O PREV NEXT JUMP ENTR EXIT Press NEXT until… DIAG OZONE GEN OVERRIDE PREV NEXT DIAG OZONE OFF ENTR EXIT OZONE GEN OVERRIDE ENTR EXIT Toggle this key to turn the O3 generator ON/OFF. 137 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.13.7.5. Flow Calibration The flow calibration allows the user to adjust the values of the sample flow rates as they are displayed on the front panel and reported through COM ports to match the actual flow rate measured at the sample inlet. This does not change the hardware measurement of the flow sensors, only the software-calculated values. To carry out this adjustment, connect an external, sufficiently accurate flow meter to the sample inlet (see Chapter 11 for more details). Once the flow meter is attached and is measuring actual gas flow, press: SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP X.X SETUP PRIMARY SETUP MENU CFG ACAL DAS RNGE PASS CLK SETUP X.X MORE EXIT SECONDARY SETUP MENU COMM VARS DIAG SETUP X.X 8 1 EXIT ENTER DIAG PASS: 818 8 ENTR EXIT DIAG SIGNAL I / O NEXT ENTR EXIT Repeat Pressing NEXT until . . . DIAG DIAG ENTR EXIT SAMPLE OZONE 0 Exit returns to the previous menu FLOW SENSOR TO CAL: SAMPLE DIAG FCAL Adjust these values until the displayed flow rate equals the flow rate being measured by the independent flow meter. Adjust these values until the displayed flow rate equals the flow rate being measured by the independent flow meter. FLOW CALIBRATION PREV NEXT Choose between sample and ozone flow sensors. Exit at any time to return to main the SETUP menu 4 ENTR EXIT ACTUAL FLOW: 480 CC / M 8 0 ENTR EXIT ENTR accepts the new value and returns to the previous menu EXIT ignores the new value and returns to the previous menu 138 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.14. SETUP – ALRM: USING THE OPTIONAL GAS CONCENTRATION ALARMS (OPT 67) The optional alarm relay outputs (Option 67) are installed includes two concentration alarms Each alarm has a user settable limit, and is associated with an opto-isolated TTL relay accessible via the status output connector on the instrument’s back panel (see Section 6.15.1.1). If the concentration measured by the instrument rises above that limit, the alarm‘s status output relay is closed NO2. The default settings for ALM1 and ALM2 are: Table 6-28: Concentration Alarm Default Settings ALARM STATUS ALM1 Disabled ALM2 1 LIMIT SET POINT Disabled OUTPUT RELAY DESIGNATION 1 100 ppm 133.9 mg/m3 AL2 300 ppm 3 AL3 401.6 mg/m Set points listed are for PPM. Should the reporting range units of measure be changed (see Section 6.13.4.5) the analyzer will automatically scale the set points to match the new range unit setting. NOTE To prevent the concentration alarms from activating during span calibration operations make sure to press CAL or CALS button prior to introducing span gas into the analyzer. To enable either of the concentration alarms and set the Limit points, press: SAMPLE A1:NXCNC1=100PPM < TST TST > CAL NOX=XXX.X SETUP SETUP X.X SETUP X.X ALARM MENU PRIMARY SETUP MENU ALM1 CFG DAS RNGE PASS CLK MORE ALM2 SETUP X. SETUP X.X EXIT EXIT ALARM 1 LIMIT: OFF SECONDARY SETUP MENU OFF COMM VARS DIAG ALRM ENTR EXIT EXIT ALARM 1 LIMIT: ON SETUP X. ON Toggle these keys to cycle through the available character set: 0-9 ENTR EXIT ALARM 1 LIMIT: 200,00 PPM SETUP X. 0 1 0 0 .0 0 ENTR EXIT ENTR key accepts the new settings EXIT key ignores the new settings 139 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.15. REMOTE OPERATION OF THE ANALYZER 6.15.1. REMOTE OPERATION USING THE EXTERNAL DIGITAL I/O 6.15.1.1. Status Outputs The status outputs report analyzer conditions via optically isolated NPN transistors, which sink up to 50 mA of DC current. These outputs can be used interface with devices that accept logic-level digital inputs, such as programmable logic controllers (PLC’s). Each Status bit is an open collector output that can withstand up to 40 VDC. All of the emitters of these transistors are tied together and available at D. NOTE Most PLC’s have internal provisions for limiting the current that the input will draw from an external device. When connecting to a unit that does not have this feature, an external dropping resistor must be used to limit the current through the transistor output to less than 50 mA. At 50 mA, the transistor will drop approximately 1.2V from its collector to emitter. The status outputs are accessed through a 12 pin connector on the analyzer’s rear panel labeled STATUS (see Figure 6-17). The function of each pin is defined in Table 6–29 STATUS + GROUND D EMITTERS 8 COMMON 7 LOW SPAN 6 DIAG MODE 5 SPAN CAL 4 ZERO CAL 3 HIGH RANGE 2 CONC VALID SYSTEM OK 1 Figure 6-6-17: Status Output Connector 140 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Table 6-29: Operating Instructions Status Output Pin Assignments CONNECTO R PIN STATUS 1 SYSTEM OK ON if no faults are present. 2 CONC VALID ON if concentration measurement is valid, OFF when invalid. 3 HIGH RANGE ON if unit is in high range of any AUTO range mode. 4 ZERO CAL ON whenever the instrument is in ZERO calibration mode. 5 SPAN CAL ON whenever the instrument is in SPAN calibration mode. CONDITION (ON=CONDUCTING) 6 DIAG MODE ON whenever the instrument is in DIAGNOSTIC mode. 7 LOW RANGE ON if unit is in low range of any AUTO range mode. 8 Unused. The emitters of the transistors on pins 1-8 are bussed together. For most applications, this pin should be connected to the circuit ground of the receiving device. D EMITTER BUS + DC POWER + 5 VDC, 30 mA maximum (combined rating with Control Inputs). DIGITAL GROUND The ground from the analyzer’s internal, 5/±15 VDC power supply. 6.15.1.2. Control Inputs Control inputs allow the user to remotely initiate ZERO and SPAN calibration modes are provided through a 10pin connector labeled CONTROL IN on the analyzer’s rear panel. These are opto-isolated, digital inputs that are activated when a 5 VDC signal from the “U” pin is connected to the respective input pin. Table 6-30: Control Input Pin Assignments INPUT STATUS CONDITION WHEN ENABLED A EXTERNAL ZERO CAL Zero calibration mode is activated. The mode field of the display will read ZERO CAL R. B EXTERNAL SPAN CAL Span calibration mode is activated. The mode field of the display will read SPAN CAL R. C EXTERNAL LOW SPAN CAL Low span (mid-point) calibration mode is activated. The mode field of the display will read LO CAL R. D, E & F Unused DIGITAL GROUND Provided to ground an external device (e.g., recorder). U DC power for Input pull ups Input for +5 VDC required to activate inputs A - F. This voltage can be taken from an external source or from the “+” pin. + Internal +5V Supply Internal source of +5V which can be used to activate inputs when connected to pin U. 141 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual There are two methods to activate control inputs. The internal +5V available from the “+” pin is the most convenient method (see Figure 6-18). However, to ensure that these inputs are truly isolated, a separate, external 5 VDC power supply should be used (see Figure 6-19). CONTROL IN ZERO C D E F U + SPAN B LOW SPAN A Figure 6-6-18: Control Inputs with local 5 V power supply CONTROL IN C D E F U + SPAN B LOW SPAN ZERO A - 5 VDC Power Supply + Figure 6-6-19: Control Inputs with external 5 V power supply 6.15.2. REMOTE OPERATION USING THE EXTERNAL SERIAL I/O 6.15.2.1. Terminal Operating Modes The Model 200EH/EM can be remotely configured, calibrated or queried for stored data through the serial ports. As terminals and computers use different communication schemes, the analyzer supports two communicate modes specifically designed to interface with these two types of devices. Computer mode is used when the analyzer is connected to a computer with a dedicated interface program such as APICOM. More information regarding APICOM can be found in later in this section or on the Teledyne Instruments website at http://www.teledyneapi.com/software/apicom/. Interactive mode is used with a terminal emulation programs such as HyperTerminal or a “dumb” computer terminal. The commands that are used to operate the analyzer in this mode are listed in Table 6-31 and in Appendix A-6. 142 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.15.2.2. Help Commands in Terminal Mode Table 6-31: Terminal Mode Software Commands COMMAND Control-T Control-C CR (carriage return) BS (backspace) ESC (escape) ? [ID] CR Control-C Control-P FUNCTION Switches the analyzer to terminal mode (echo, edit). If mode flags 1 & 2 are OFF, the interface can be used in interactive mode with a terminal emulation program. Switches the analyzer to computer mode (no echo, no edit). A carriage return is required after each command line is typed into the terminal/computer. The command will not be sent to the analyzer to be executed until this is done. On personal computers, this is achieved by pressing the ENTER key. Erases one character to the left of the cursor location. Erases the entire command line. This command prints a complete list of available commands along with the definitions of their functionality to the display device of the terminal or computer being used. The ID number of the analyzer is only necessary if multiple analyzers are on the same communications line, such as the multi-drop setup. Pauses the listing of commands. Restarts the listing of commands. 6.15.2.3. Command Syntax Commands are not case-sensitive and all arguments within one command (i.e. ID numbers, keywords, data values, etc.) must be separated with a space character. All Commands follow the syntax: X [ID] COMMAND Where X is the command type (one letter) that defines the type of command. Allowed designators are listed in Table 6-31 and Appendix A-6. [ID] is the analyzer identification number (see Section 6.11.1.). Example: the Command “? 200” followed by a carriage return would print the list of available commands for the revision of software currently installed in the instrument assigned ID Number 200. COMMAND is the command designator: This string is the name of the command being issued (LIST, ABORT, NAME, EXIT, etc.). Some commands may have additional arguments that define how the command is to be executed. Press ? or refer to Appendix A-6 for a list of available command designators. is a carriage return. All commands must be terminated by a carriage return (usually achieved by pressing the ENTER key on a computer). Table 6-32: Command Types COMMAND C D L T V W COMMAND TYPE Calibration Diagnostic Logon Test measurement Variable Warning 143 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.15.2.4. Data Types Data types consist of integers, hexadecimal integers, floating-point numbers, Boolean expressions and text strings. Integer data are used to indicate integral quantities such as a number of records, a filter length, etc. They consist of an optional plus or minus sign, followed by one or more digits. For example, +1, -12, 123 are all valid integers. Hexadecimal integer data are used for the same purposes as integers. They consist of the two characters “0x,” followed by one or more hexadecimal digits (0-9, A-F, a-f), which is the ‘C’ programming language convention. No plus or minus sign is permitted. For example, 0x1, 0x12, 0x1234abcd are all valid hexadecimal integers. Floating-point numbers are used to specify continuously variable values such as temperature set points, time intervals, warning limits, voltages, etc. They consist of an optional plus or minus sign, followed by zero or more digits, an optional decimal point, and zero or more digits. (At least one digit must appear before or after the decimal point.) Scientific notation is not permitted. For example, +1.0, 1234.5678, 0.1, 1 are all valid floating-point numbers. Boolean expressions are used to specify the value of variables or I/O signals that may assume only two values. They are denoted by the keywords ON and OFF. Text strings are used to represent data that cannot be easily represented by other data types, such as data channel names, which may contain letters and numbers. They consist of a quotation mark, followed by one or more printable characters, including spaces, letters, numbers, and symbols, and a final quotation mark. For example, “a”, “1”, “123abc”, and “()[]<>” are all valid text strings. It is not possible to include a quotation mark character within a text string. Some commands allow you to access variables, messages, and other items, such as iDAS data channels, by name. When using these commands, you must type the entire name of the item; you cannot abbreviate any names. 144 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.15.2.5. Status Reporting Reporting of status messages as an audit trail is one of the three principal uses for the RS-232 interface (the other two being the command line interface for controlling the instrument and the download of data in electronic format). You can effectively disable the reporting feature by setting the interface to quiet mode (see Section 6.11.8., Table 6-14). Status reports include iDAS data (when reporting is enabled), warning messages, calibration and diagnostic status messages. Refer to Appendix A-3 for a list of the possible messages, and this section for information on controlling the instrument through the RS-232 interface. GENERAL MESSAGE FORMAT All messages from the instrument (including those in response to a command line request) are in the format: X DDD:HH:MM [Id] MESSAGE Where X is a command type designator, a single character indicating the message type, as shown in the Table 6-31. DDD:HH:MM is the time stamp, the date and time when the message was issued. It consists of the Day-of-year (DDD) as a number from 1 to 366, the hour of the day (HH) as a number from 00 to 23, and the minute (MM) as a number from 00 to 59. [ID] is the analyzer ID, a number with 1 to 4 digits. MESSAGE is the message content that may contain warning messages, test measurements, iDAS reports, variable values, etc. is a carriage return / line feed pair, which terminates the message. The uniform nature of the output messages makes it easy for a host computer to parse them into an easy structure. Keep in mind that the front panel display does not give any information on the time a message was issued, hence it is useful to log such messages for trouble-shooting and reference purposes. Terminal emulation programs such as HyperTerminal can capture these messages to text files for later review. 6.15.2.6. Remote Access by Modem The M200EH/EM can be connected to a modem for remote access. This requires a cable between the analyzer’s COM port and the modem, typically a DB-9F to DB-25M cable (available from Teledyne Instruments with part number WR0000024). Once the cable has been connected, check to make sure the DTE-DCE is in the correct position. Also make sure the M200EH/EM COM port is set for a baud rate that is compatible with the modem, which needs to operate with an 8-bit word length with one stop bit. The first step is to turn on the MODEM ENABLE communication mode (Mode 64, Section 6.11.8). Once this is completed, the appropriate setup command line for your modem can be entered into the analyzer. The default setting for this feature is AT Y0 &D0 &H0 &I0 S0=2 &B0 &N6 &M0 E0 Q1 &W0 This string can be altered to match your modem’s initialization and can be up to 100 characters long. 145 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual NOTE: If Hessen Protocol Mode is active for a COMM port, operation via a modem is not available on that port. To change this setting press: SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP SETUP X.X SET> SETUP X.X COM1 MODE:0 EDIT CFG DAS RNGE PASS CLK MORE EXIT SETUP X.X EXIT returns to the previous menu SETUP X.X EXIT PRIMARY SETUP MENU COM1 BAUD RATE:19200 EDIT COMM VARS DIAG ALRM EXIT SETUP X.X Select which COM Port is tested SETUP X.X ID COM1 EXIT SECONDARY SETUP MENU COMMUNICATIONS MENU COM2 COM1 MODEM INIT:AT Y &D &H EDIT EXIT EXIT SETUP X.X The keys move the [ ] cursor left and right along the text string COM1 MODEM INIT:[A]T Y &D &H INS The INS key inserts a character before the cursor location. DEL [A] ENTR The DEL key deletes a character at the cursor location. EXIT ENTR accepts the new string and returns to the previous menu. EXIT ignores the new string and returns to the previous menu. Press the [?] key repeatedly to cycle through the available character set: 0-9 A-Z space ’ ~ ! # $ % ^ & * ( ) - _ = +[ ] { } < >\ | ; : , . / ? 146 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions To Initialize the modem press: SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP SETUP X.X SETUP X.X SET> PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE COM1 MODE:0 EDIT EXIT SETUP X.X EXIT returns to the previous menu SETUP X.X SECONDARY SETUP MENU COMM VARS DIAG ALRM COM1 BAUD RATE:19200 EDIT SETUP X.X ID COM1 EXIT EXIT COM1 MODEM INIT:AT Y &D &H SETUP X.X Select which COM Port is tested EXIT COMMUNICATIONS MENU COM2 EDIT EXIT EXIT SETUP X.X COM1 INITIALIZE MODEM INIT SETUP X.X EXIT returns to the Communications Menu. EXIT INITIALIZING MODEM INIT EXIT 6.15.2.7. COM Port Password Security In order to provide security for remote access of the M200EH/EM, a LOGON feature can be enabled to require a password before the instrument will accept commands. This is done by turning on the SECURITY MODE (see Section 6.11.8). Once the SECURITY MODE is enabled, the following items apply. A password is required before the port will respond or pass on commands. If the port is inactive for one hour, it will automatically logoff, which can also be achieved with the LOGOFF command. Three unsuccessful attempts to log on with an incorrect password will cause subsequent logins to be disabled for 1 hour, even if the correct password is used. If not logged on, the only active command is the '?' request for the help screen. The following messages will be returned at logon: o LOGON SUCCESSFUL - Correct password given o LOGON FAILED - Password not given or incorrect o LOGOFF SUCCESSFUL - Connection terminated successfully To log on to the M200EH/EM analyzer with SECURITY MODE feature enabled, type: LOGON 940331 940331 is the default password. To change the default password, use the variable RS232_PASS issued as follows: V RS232_PASS=NNNNNN Where N is any numeral between 0 and 9. 147 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.15.2.8. APICOM Remote Control Program APICOM is an easy-to-use, yet powerful interface program that allows to access and control any of Teledyne Instruments’ main line of ambient and stack-gas instruments from a remote connection through direct cable, modem or Ethernet. Running APICOM, a user can: Establish a link from a remote location to the M200EH/EM through direct cable connection via RS-232 modem or Ethernet. View the instrument’s front panel and remotely access all functions that could be accessed when standing in front of the instrument. Remotely edit system parameters and set points. Download, view, graph and save data for predictive diagnostics or data analysis. Retrieve, view, edit, save and upload iDAS configurations. Check on system parameters for trouble-shooting and quality control. APICOM is very helpful for initial setup, data analysis, maintenance and trouble-shooting. Figure 6-20 shows examples of APICOM’s main interface, which emulates the look and functionality of the instruments actual front panel Figure 6-6-20: APICOM Remote Control Program Interface APICOM is included free of cost with the analyzer and the latest versions can also be downloaded for free at http://www.teledyne-api.com/software/apicom/. 6.15.3. ADDITIONAL COMMUNICATIONS DOCUMENTATION Table 6-33: Serial Interface Documents Interface / Tool Document Title Part Number Available Online* APICOM APICOM User Manual 039450000 YES Multi-drop RS-232 Multi-drop Documentation 021790000 YES DAS Manual Detailed description of the iDAS. 028370000 YES * These documents can be downloaded at http://www.teledyne-api.com/manuals/ 148 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.15.4. USING THE M200EH/EM WITH A HESSEN PROTOCOL NETWORK 6.15.4.1. General Overview of Hessen Protocol The Hessen protocol is a multidrop protocol, in which several remote instruments are connected via a common communications channel to a host computer. The remote instruments are regarded as slaves of the host computer. The remote instruments are unaware that they are connected to a multidrop bus and never initiate messages. They only respond to commands from the host computer and only when they receive a command containing their own unique ID number. The Hessen protocol is designed to accomplish two things: to obtain the status of remote instruments, including the concentrations of all the gases measured; and to place remote instruments into zero or span calibration or measure mode. API’s implementation supports both of these principal features. The Hessen protocol is not well defined, therefore while API’s application is completely compatible with the protocol itself, it may be different from implementations by other companies. The following subsections describe the basics for setting up your instrument to operate over a Hessen Protocol network. For more detailed information as well as a list of host computer commands and examples of command and response message syntax, download the Manual Addendum for Hessen Protocol from the Teledyne Instruments’ web site: http://www.teledyne-api.com/manuals/index.asp . 6.15.4.2. Hessen COMM Port Configuration Hessen protocol requires the communication parameters of the M200EH/EM’s COMM ports to be set differently than the standard configuration as shown in the table below. Table 6-34: RS-232 Communication Parameters for Hessen Protocol Parameter Standard Hessen Data Bits 8 7 Stop Bits 1 2 Parity None Even Duplex Full Half To change the rest of the COMM port parameters and modes (see Section 6.11.8). To change the baud rate of the M200EH/EM’s COMM ports (see Section 6.11.9.) NOTES Make sure that the communication parameters of the host computer are properly set. The instrument software has a 200 ms. latency before it responds to commands issued by the host computer. Activating Hessen Protocol. Operation via modem is not available over any COMM port on which HESSEN protocol is active. 149 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual The first step in configuring the M200EH/EM to operate over a Hessen protocol network is to activate the Hessen mode for COMM ports and configure the communication parameters for the port(s) appropriately. Press: SAMPLE Repeat the entire process to set up the COM2 port A1:NXCNC1=100PPM < TST TST > CAL SETUP X.X NOX=XXX.X SETUP X.X SETUP NEXT OFF PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE Continue pressing next until … PREV NEXT SETUP X.X ID The sum of the mode IDs of the selected modes is displayed here COM1 ALRM COM2 SETUP X.X ENTR EXIT COM1 HESSEN PROTOCOL : ON PREV NEXT ON EXIT COM1 MODE:0 EDIT OFF EXIT COMMUNICATIONS MENU SETUP X.X SET> COM1 HESSEN PROTOCOL : OFF SECONDARY SETUP MENU COMM VARS DIAG Select which COMM port to configure ENTR EXIT EXIT SETUP X.X SETUP X.X COM1 QUIET MODE: OFF EXIT ENTR EXIT SETUP X.X COM1 E,7,1 MODE: OFF PREV NEXT OFF SETUP X.X COM1 E,7,1 MODE: ON Toggle OFF/ON keys to change activate/deactivate selected mode. ENTR EXIT PREV NEXT ON ENTR key accepts the new settings ENTR EXIT EXIT key ignores the new settings 6.15.4.3. Selecting a Hessen Protocol Type Currently there are two version of Hessen Protocol in use. The original implementation, referred to as TYPE 1, and a more recently released version, TYPE 2 that has more flexibility when operating with instruments that can measure more than one type of gas. For more specific information about the difference between TYPE 1and TYPE 2 download the Manual Addendum for Hessen Protocol from the Teledyne Instruments’ web site: http://www.teledyne-api.com/manuals/index.asp . To select a Hessen Protocol Type press: SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP SETUP X. SETUP X.X PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SET> HESSEN VARIATION: TYPE 1 EDIT EXIT ENTR key accepts the new settings SETUP X.X SETUP X.X SECONDARY SETUP MENU COMM VARS DIAG SETUP X.X ID HESN EXIT ALRM COMMUNICATIONS MENU COM1 COM2 EXIT HESSEN VARIATION: TYPE 1 TYE1 TYPE 2 EXIT key ignores the new settings ENTR EXIT EXIT Press to change protocol type. SETUP X.X PREV NEXT HESSEN VARIATION: TYPE 2 OFF ENTR EXIT NOTE 150 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions While Hessen Protocol Mode can be activated independently for COM1 and COM2, The TYPE selection affects both Ports. 6.15.4.4. Setting The Hessen Protocol Response Mode The Teledyne Instruments’ implementation of Hessen Protocol allows the user to choose one of several different modes of response for the analyzer. Table 6-28: M200EH/EM Hessen Protocol Response Modes MODE ID MODE DESCRIPTION CMD This is the Default Setting. Reponses from the instrument are encoded as the traditional command format. Style and format of responses depend on exact coding of the initiating command. BCC Responses from the instrument are always delimited with (at the beginning of the response, (at the end of the response followed by a 2 digit Block Check Code (checksum), regardless of the command encoding. TEXT Responses from the instrument are always delimited with at the beginning and the end of the string, regardless of the command encoding. To Select a Hessen response mode, press: SAMPLE RANGE = 500.000 PPB SO2 =XXX.X < TST TST > CAL SAMPLE 8 SETUP X.X SETUP ENTER SETUP PASS : 818 1 8 ENTR EXIT ID PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X EXIT ALRM HESN SETUP X.X SET> SECONDARY SETUP MENU COMM VARS DIAG SETUP X.X COMMUNICATIONS MENU COM1 COM2 EXIT HESSEN VARIATION: TYPE 1 EDIT EXIT ENTR key accepts the new settings EXIT Press to change response mode. SETUP X.X HESSEN RESPONSE MODE :CMD EDIT SETUP X.X HESSEN RESPONSE MODE :CMD BCC TEXT EDIT EXIT key ignores the new settings EXIT ENTR EXIT 151 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual 6.15.4.5. Hessen Protocol Gas ID Since the M200EH/EM measures both NO, NO2 NO and O2 (if the optional sensor is installed), all of these gases are listed in the Hessen protocol gas list. In its default state the Hessen protocol firmware assigns each of these gases a Hessen ID number and actively reports all of them even if the instrument is only measuring one (see MEASURE_MODE, Section 6.12) . To change or edit these settings press: SAMPLE A1:NXCNC1=100PPM NOX=XXX.X KEY < TST TST > CAL SETUP X.X SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X ID HESN SETUP X.X SET> Moves to next gas entry in list NEXT> Moves the cursor previous gas entry in list INS Inserts a new gas entry into the list. DEL Deletes the >>>>>. ENTR Accepts the new setting and returns to the previous menu. EXIT Ignores the new setting and returns to the previous menu. SECONDARY SETUP MENU COMM VARS DIAG SETUP X.X EXI FUNCTION EDIT SETUP X.X HESSEN GAS LIST EDIT EXIT EXIT HESSEN VARIATION: TYPE 1 EDIT SETUP X.X EXIT EXIT SETUP X.X NOX, 211, REPORTED INS DEL EDIT PRNT EXIT Use the PREV & NEXT keys to cycle existing entries in Hessen gas list SETUP X.X GAS TYPE NOX ENTR EXIT Use the PREV & NEXT keys to cycle through available gases SETUP X.X 0 0 ENTR key accepts the new settings GAS ID: 211 0 ENTR EXIT EXIT key ignores the new settings Toggle these keys to change the gas ID number for the chosen gas. SETUP X.X REPORTED : ON ON ENTR EXIT Toggle this key to switch reporting Between ON and OFF Table 6-35: M200EH/EM Hessen GAS ID List GAS DEFAULT HESSEN GAS ID NOx 211 NO 212 NO2 213 O2 214 152 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Operating Instructions 6.15.4.6. Setting Hessen Protocol Status Flags Teledyne Instruments’ implementation of Hessen protocols includes a set of status bits that are included in responses to inform the host computer of the M200EH/EM’s condition. The default settings for these bit/flags are: Table 6-36: Default Hessen Status Bit Assignments STATUS FLAG NAME DEFAULT BIT ASSIGNMENT WARNING FLAGS SAMPLE FLOW WARNING 0001 OZONE FLOW WARNING 0002 RCELL PRESS WARN 0004 BOX TEMP WARNING 0008 RCELL TEMP WARNING 0010 PMT TEMP WARNING 0040 CONVERTER TEMP WARNING 0080 WARMUP MODE 1000 INVALID CONC 8000 OPERATIONAL FLAGS In Manual Calibration Mode 0200 In O2 Calibration Mode 0400 In Zero Calibration Mode 0400 In Low Span Calibration Mode 0800 In Span Calibration Mode 0800 UNITS OF MEASURE FLAGS UGM 0000 MGM 2000 PPB 4000 PPM 6000 0020, 0100 SPARE/UNUSED BITS UNASSIGNED FLAGS Box Temp Warning Front Panel Warning System Reset Analog Cal Warning Rear Board Not Detected Cannot Dyn Zero Relay Board Warning Cannot Dyn Span Manifold Temp Warn O2 Cell Temp Warn Ozone Gen Off AutoZero Warning Conc Alarm 1 Conc Alarm 2 HVPS Warning In MP Calibration Mode NOTES: It is possible to assign more than one flag to the same Hessen status bit. This allows the grouping of similar flags, such as all temperature warnings, under the same status bit. Be careful not to assign conflicting flags to the same bit as each status bit will be triggered if any of the assigned flags is active. 153 04521C (DCN5731) Operating Instructions Teledyne API - Model 200EH/EM Operation Manual To assign or reset the status flag bit assignments, press: SAMPLE A1:NXCNC1=100PPM NOX=XXX.X < TST TST > CAL SETUP X.X SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X SECONDARY SETUP MENU COMM VARS DIAG SETUP X.X ID EXIT ALRM EXIT COMMUNICATIONS MENU HESN COM1 COM2 EXIT Repeat pressing SET> until … SETUP X. HESSEN STATUS FLAGS EDIT SETUP X. PMT DET WARNING: 0002 PREV NEXT EXIT EDIT PRNT EXIT Repeat pressing NEXT or PREV until the desired message flag is displayed. See Table 6-29. For example … SETUP X. PREV NEXT The keys move the [ ] cursor left and right along the bit string. SETUP X. SYSTEM RESET: 0000 EDIT PRNT EXIT SYSTEM RESET: [0]000 [0] ENTR key accepts the new settings ENTR EXIT EXIT key ignores the new settings Press the [?] key repeatedly to cycle through the available character set: 0-9 Note: Values of A-F can also be set but are meaningless. 6.15.4.7. Instrument ID Code Each instrument on a Hessen Protocol network must have a unique ID code. The M200EH/EM is programmed with a default ID code of 200. To change this code see Section 6.11.1 User Notes: 154 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Calibration Procedures 7. CALIBRATION PROCEDURES This chapter describes calibration procedures for the M200EH/EM. All of the methods described here can be initiated and controlled through the front panel or the COM ports. NOTE CALIBRATION vs. CALIBRATION CHECK Pressing the ENTR key during the following procedures re-calculates the stored values for OFFSET and SLOPE and alters the instrument’s calibration. If you wish to perform a calibration check, DO NOT press the ENTR button. 7.1. CALIBRATION PREPARATIONS 7.1.1. REQUIRED EQUIPMENT, SUPPLIES, AND EXPENDABLES Calibration of the Model 200EH/EM analyzer requires a certain amount of equipment and supplies. These include, but are not limited to, the following: Zero-air source (defined in Section 7.1.2). Span gas source (defined in Section 7.1.3). Gas lines - all gas line materials should be stainless steel or Teflon-type (PTFE or FEP). High concentration NO gas transported over long distances may require stainless steel to avoid oxidation of NO with O2 diffusing into the tubing. A recording device such as a strip-chart recorder and/or data logger (optional). For electronic documentation, the internal data acquisition system can be used. 7.1.2. ZERO AIR Zero air is similar in chemical composition to the Earth’s atmosphere but scrubbed of all components that might affect the analyzer’s readings. For NOX measuring devices, zero air should be devoid of NOX and large amounts of CO2, NH3 and water vapor. Water vapor and moderate amounts of NH3 can be removed using a sample gas conditioner (Section 5.10). Devices such as the API Model 701 zero air generator that condition ambient air by drying and removal of pollutants are available. We recommend this type of device for generating zero air. Please contact our sales department for more information on this. 155 04521C (DCN5731) Calibration Procedures Teledyne API - Model 200EH/EM Operation Manual 7.1.3. SPAN CALIBRATION GAS STANDARDS & TRACEABILITY NOTE We strongly recommend that span calibration is carried out with NO span gas, although it is possible to use NO2 or a gas phase titration (GPT) system. Quick span checks may be done with either NO, NO2 or a mixture of NO and NO2 as is used in GPT. Span gas is specifically mixed to match the chemical composition of the gas being measured at about 80% of the desired full measurement range. For example, if the measurement range is 120 ppm, the span gas should have an NO concentration of about 96 ppm. Span gases should be certified to a specific accuracy to ensure accurate calibration of the analyzer. Typical gas accuracy for NOX gases is 1 or 2%. NO standards should be mixed in nitrogen (to prevent oxidation of NO to NO2 over time), whereas NO2 standards should be mixed in air (to keep it oxidized). For oxygen measurements, we recommend s reference gas of 21% O2 in N2. the user can either utilize the NOX standards (if mixed in air). For quick checks. ambient air can be used at an assumed concentration of 20.8%. Generally, O2 concentration in dry, ambient air varies by less than 1%. 7.1.3.1. Traceability All equipment used to produce calibration gases should be verified against standards of the National Institute for Standards and Technology (NIST). To ensure NIST traceability, we recommend to acquire cylinders of working gas that are certified to be traceable to NIST standard reference materials (SRM). These are available from a variety of commercial sources. Table 7-1: NIST-SRM's Available for Traceability of NOx Calibration Gases NIST-SRM4 TYPE NOMINAL CONCENTRATION 2627a 2628a 2629a Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 5 ppm 10 ppm 20 ppm 1683b 1684b 1685b 1686b 1687b Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 50 ppm 100 ppm 250 ppm 5000 ppm 1000 ppm 2630 2631a 2635 2636a Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 Nitric Oxide (NO) in N2 1500 ppm 3000 ppm 800 ppm 2000 ppm 2656 2660a Oxides of Nitrogen (NOx) in Air Oxides of Nitrogen (NOx) in Air 2500 ppm 100 ppm 2659a Oxygen in Nitrogen (O2) 21 mol % 156 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Calibration Procedures 7.1.4. DATA RECORDING DEVICES A strip chart recorder, data acquisition system or digital data acquisition system should be used to record data from the serial or analog outputs of the M200EH/EM. If analog readings are used, the response of the recording system should be checked against a NIST traceable voltage source or meter. Data recording devices should be capable of bi-polar operation so that negative readings can be recorded. For electronic data recording, the M200EH/EM provides an internal data acquisition system (iDAS), which is described in detail in Section 6.7. APICOM, a remote control program, is also provided as a convenient and powerful tool for data handling, download, storage, quick check and plotting. 7.1.5. NO2 CONVERSION EFFICIENCY To ensure accurate operation of the M200EH/EM, it is important to check the NO2 conversion efficiency (CE) periodically and to update this value as necessary. The default setting for the NO2 converter efficiency is 1.0000. For the analyzer to function correctly, the converter efficiency must be between 0.9600 and 1.0200 (96-102% conversion efficiency) as per US-EPA requirements. If the converter’s efficiency is outside these limits, the NO2 converter should be replaced. NOTE The currently programmed CE is recorded along with the calibration data in the iDAS for documentation and performance analysis 7.1.5.1. Determining / Updating the NO2 Converter Efficiency The following procedure will cause the Model 200EH/EM to automatically calculate the current NO2 conversion efficiency. STEP ONE: Connect a source of calibrated NO2 span gas as shown below. Source of MODEL 700 Gas Dilution Calibrator SAMPLE GAS VENT here if input is pressurized Removed during calibration NOx Gas (High Concentration) SAMPLE MODEL 701 Zero Gas Generator VENT EXHAUST MODEL 200EH/EM PUMP Figure 7-1: Gas Supply Setup for Determination of NO2 Conversion Efficiency 157 04521C (DCN5731) Calibration Procedures Teledyne API - Model 200EH/EM Operation Manual STEP TWO: Set the expected NO2 span gas concentration: SAMPLE < TST TST > A1:NXCNC1=100PPM CAL SAMPLE NOX M-P CAL NOX=XXX.X M-P CAL GAS TO CAL:NOX SAMPLE NOX ENTR EXIT NO2 ENTR EXIT EXIT CONCENTRATION MENU NO M-P CAL RANGE TO CAL:LOW LOW HIGH NOX=X.XXX ZERO SPAN CONC SETUP O2 A1:NXCNC1 =100PPM EXIT CONVERTER EFFICIENCY MENU CAL M-P CAL 0 CONV SET EXIT NO2 CE CONC:80.0 Conc 0 8 0 .0 EXIT ignores the new setting and returns to the previous display. ENTR EXIT ENTR accepts the new setting and returns to the CONVERTER EFFICIENCY MENU. If using NO span gas in addition to NO X repeat last step. The NO X & NO span concentration values automatically default to 80.0 Conc. If this is not the the concentration of the span gas being used, toggle these buttons to set the correct concentration of the NO X and NO calibration gases. STEP THREE Activate NO2 measurement stability function. SAMPLE RANGE = 50.000 PPM < TST TST > SETUP X.X CO =X.XXX CAL SETUP COMM 0) DAS_HOLD_OFF=15.0 Minutes EDIT PRNT EXIT PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X SETUP X.X JUMP EXIT Continue pressing NEXT until ... SECONDARY SETUP MENU VARS DIAG ALRM EXIT SETUP X.X 2) STABIL_GAS=NOX JUMP SETUP X.X 8 1 EDIT PRNT EXIT ENTER PASSWORD:818 8 ENTR EXIT SETUP X.X NO NO2 SETUP X.X Press EXIT 3 times to return to SAMPLE menu NO NO2 STABIL_GAS:NOX NOX O2 ENTR EXIT STABIL_GAS:NO2 NOX O2 ENTR EXIT 158 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Calibration Procedures STEP FOUR: Perform the converter efficiency calculation procedure: SAMPLE A1:NXCNC1=100PPM < TST TST > NOX=XXX.X CAL SETUP Toggle TST> button until ... SAMPLE NOX STB= XXX.X PPM < TST TST > NOX NOX=XXX.X CAL SAMPLE SETUP GAS TO CAL:NOX O2 ENTR EXIT SAMPLE RANGE TO CAL:LOW LOW HIGH ENTR EXIT M-P CAL NOX STB= XXX.X PPM OX=X.XXX ZERO SPAN CONC M-P CAL NOX EXIT CONCENTRATION MENU NO M-P CAL NO2 Set the Display to show the NOX STB test function. This function calculates the stability of the NO/NO x measurement CONV CONVERTER EFFICIENCY MENU CAL SET M-P CAL 1 EXIT EXIT CE FACTOR:1.000 Gain .0 0 0 0 ENTR EXIT Allow NO 2 to enter the sample port at the rear of the analyzer. M-P CAL NO2 When ENTR is pressed, the ratio of observed NO 2 concentration to expected NO 2 concentration is calculated and stored. CONVERTER EFFICIENCY MENU CAL SET SAMPLE NOX STB= XXX.X PPM < TST TST > M-P CAL NO2 ENTR NOX=XXX.X Wait until NO2 STB falls below 0.5 ppm and the ENTR button appears. This may take several minutes. SETUP CONVERTER EFFICIENCY MENU CAL M-P CAL 1 EXIT SET EXIT CE FACTOR:1.012 Gain .0 0 1 2 ENTR EXIT Press EXIT 3 times top return to the SAMPLE display 159 04521C (DCN5731) Calibration Procedures Teledyne API - Model 200EH/EM Operation Manual 7.2. MANUAL CALIBRATION The following section describes the basic method for manually calibrating the Model 200EH/EM NOX analyzer. If both available iDAS parameters for a specific gas type are being reported via the instruments analog outputs e.g. NXCNC1 and NXCNC2, separate calibrations should be carried out for each parameter. Use the LOW button when calibrating for NXCNC1 Use the HIGH button when calibrating for NXCNC2. See Section 6.13.3 & 6.13.4 for more information on analog output reporting ranges STEP ONE: Connect the sources of zero air and span gas as shown below. VENT here if input is pressurized Source of SAMPLE Gas VENT at HIGH Span Concentration Calibrated NO MODEL 700 Gas Dilution Calibrator MODEL 701 Zero Gas Generator Sample PUMP Exhaust Span Point External Zero Air Scrubber Zero Air Filter Pneumatic Connections–With Zero/Span Valve Option (50) On/Off Valves Source of SAMPLE Gas VENT at LOW Span Concentration VENT here if input is pressurized PUMP VENT Calibrated NO at HIGH Span Concentration Calibrated NO Figure 7-2: MODEL 200EH/EM Sample Exhaust High Span Point Low Span Point External Zero Air Scrubber Figure 7-3: Filter Zero Air MODEL 200EH/EM Pneumatic Connections–With 2-Span point Option (52) –Using Bottled Span Gas 160 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Calibration Procedures STEP TWO: Set Expected NO and NOX Span Gas Concentrations Set the expected NO and NOx span gas concentration. These should be 80% of range of concentration values likely to be encountered in this application. The default factory setting is 100 ppm. If one of the configurable analog outputs is to be set to transmit concentration values, use 80% of the reporting range set for that output (see Section 6.13.4.5) SAMPLE A1:NXCNC1=100PPM < TST TST > CAL SAMPLE NOX NOX=XXX.X SETUP GAS TO CAL:NOX O2 ENTR EXIT SAMPLE RANGE TO CAL:LOW LOW HIGH M-P CAL ENTR EXIT A1:NXCNC1 =100PPM NOX=X.XXX ZERO SPAN CONC M-P CAL NOX CONCENTRATION MENU NO CONV M-P CAL 0 The NOX & NO span concentration values automatically default to 80.0 Conc. If this is not the the concentration of the span gas being used, toggle these buttons to set the correct concentration of the NO X and NO calibration gases. EXIT EXIT NOX SPAN CONC:80.0 Conc 0 8 0 .0 ENTR EXIT EXIT ignores the new setting and returns to the previous display. ENTR accepts the new setting and returns to the CONCENTRATION MENU. If using NO span gas in addition to NOX repeat last step. NOTE The expected concentrations for both NOX and NO are usually set to the same value unless the conversion efficiency is not equal to 1.000 or not entered properly in the conversion efficiency setting. When setting expected concentration values, consider impurities in your span gas source (NO often contains 1-3% NO2 and vice versa). 161 04521C (DCN5731) Calibration Procedures Teledyne API - Model 200EH/EM Operation Manual STEP THREE: Perform Zero/Span Calibration: SAMPLE Analyzer continues to cycle through NO x, NO, and NO 2 measurements throughout this procedure. A1:NXCNC1=100PPM < TST TST > NOX=XXX.X CAL SETUP Toggle TST> button until ... SAMPLE NOX STB= XXX.X PPM < TST TST > Set the Display to show the NOX STB test function. This function calculates the stability of the NO/NO x measurement NOX=XXX.X CAL SETUP Allow zero gas to enter the sample port at the rear of the analyzer. Wait until NOX STB falls below 0.5 ppm. This may take several minutes. SAMPLE NOX STB= XXX.X PPM < TST TST > SAMPLE NOX NOX=XXX.X CAL SETUP GAS TO CAL:NOX O2 ENTR EXIT SAMPLE RANGE TO CAL:LOW LOW HIGH M-P CAL ENTR EXIT NOX STB= XXX.X PPM M-P CAL ZERO CONC NOX STB= XXX.X PPM ENTR NOX=XXX.X EXIT NOX=X.XXX CONC EXIT Allow span gas to enter the sample port at the rear of the analyzer. Press ENTR to changes the OFFSET & SLOPE values for both the NO and NO x measurements. Press EXIT to leave the calibration unchanged and return to the previous menu. Wait until NOX STB falls below 0.5 ppm. This may take several minutes. SAMPLE NOX STB= XXX.X PPM < TST TST > SAMPLE NOX You may see both keys. If either the ZERO or SPAN buttons fail to appear see Section 11 for troubleshooting tips. SETUP GAS TO CAL:NOX O2 ENTR EXIT SAMPLE The SPAN key now appears during the transition from zero to span. CAL RANGE TO CAL:LOW LOW HIGH M-P CAL ENTR EXIT NOX STB= XXX.X PPM ZERO SPAN CONC M-P CAL NOX STB= XXX.X PPM ENTR M-P CAL NOX=XXX.X CONC NOX STB= XXX.X PPM ENTR CONC NOX=X.XXX EXIT NOX=X.XXX EXIT NOX=X.XXX EXIT Press ENTR to changes the OFFSET & SLOPE values for both the NO and NO x measurements. Press EXIT to leave the calibration unchanged and return to the previous menu. EXIT at this point returns to the SAMPLE menu. 162 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Calibration Procedures 7.3. CALIBRATION CHECKS Informal calibration checks, which only evaluate but do not alter the analyzer’s response curve, are recommended as a regular maintenance item and in order to monitor the analyzer’s performance. To carry out a calibration check rather than a full calibration, follow these steps. STEP ONE: Connect the sources of zero air and span gas as shown in Figure 7.2 or 7.3. STEP TWO: Perform the zero/span calibration check procedure: SAMPLE < TST TST > Analyzer display continues to cycle through all of the available gas measurements throughout this procedure. A1:NXCNC1=100PPM NOX=XXX.X CAL SETUP Toggle TST> button until ... SAMPLE < TST TST > NOX STB= XXX.X PPM Set the Display to show the NOX STB test function. This function calculates the stability of the NO/NO x measurement NOX=XXX.X CAL SETUP Allow zero gas to enter the sample port at the rear of the analyzer. Wait until NOX STB falls below 0.5 ppm. This may take several minutes. Record NO X , NO, NO 2 or O 2 zero point readings Wait until NOX STB falls below 0.5 ppm. Allow span gas to enter the sample port at the rear of the analyzer. This may take several minutes. The ZERO and/or SPAN keys will appear at various points of this process. It is not necessary to press them. Record NO X, NO, NO 2 or O 2 span point readings\ 163 04521C (DCN5731) Calibration Procedures Teledyne API - Model 200EH/EM Operation Manual 7.4. MANUAL CALIBRATION WITH ZERO/SPAN VALVES Zero and Span calibrations using the Zero/Span Valve option are similar to that described in Section 7.2, except that: Zero air and span gas is supplied to the analyzer through the zero gas and span gas inlets rather than through the sample inlet. The zero and cal operations are initiated directly and independently with dedicated keys (CALZ & CALS) If both available iDAS parameters for a specific gas type are being reported via the instruments analog outputs e.g. NXCNC1 and NXCNC2, separate calibrations should be carried out for each parameter. Use the LOW button when calibrating for NXCNC1 Use the HIGH button when calibrating for NXCNC2. See Section 6.13.3 & 6.13.4 for more information on analog output reporting ranges STEP ONE: Connect the sources of zero air and span gas to the respective ports on the rear panel (Figure 3-1) as shown below. VENT here if input VENT at HIGH Span Concentration Calibrated NO MODEL 700 Gas Dilution Calibrator MODEL 701 Zero Gas Generator is pressurized Source of SAMPLE Gas PUMP Sample Exhaust Span Point External Zero Air Scrubber Figure7-4: Filter Zero Air MODEL 200EH/EM Pneumatic Connections–With Zero/Span Valve Option (50) 164 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Calibration Procedures STEP TWO: Set Expected NO and NOX Span Gas Concentrations. Set the expected NO and NOx span gas concentration. These should be 80% of range of concentration values likely to be encountered in this application. The default factory setting is 100 ppm. If one of the configurable analog outputs is to be set to transmit concentration values, use 80% of the reporting range set for that output (see Section 6.13.4.5) SAMPLE A1:NXCNC1=100PPM < TST TST > CAL CALZ CALS SAMPLE NOX NOX=XXX.X SETUP GAS TO CAL:NOX O2 SAMPLE ENTR EXIT RANGE TO CAL:LOW LOW HIGH ENTR EXIT SPAN CAL M A1:NXCNC1 =100PPM NOX=X.XXX ZERO SPAN CONC EXIT SPAN CAL M CONCENTRATION MENU NOX NO CONV EXIT SPAN CAL M NOX SPAN CONC:80.0 Conc 0 The NOX & NO span concentration values automatically default to 80.0 Conc. If this is not the the concentration of the span gas being used, toggle these buttons to set the correct concentration of the NO X and NO calibration gases. 0 8 0 .0 ENTR EXIT EXIT ignores the new setting and returns to the previous display. ENTR accepts the new setting and returns to the CONCENTRATION MENU. If using NO span gas in addition to NOX repeat last step. NOTE The expected concentrations for both NOX and NO are usually set to the same value unless the conversion efficiency is not equal to 1.000 or not entered properly in the conversion efficiency setting. When setting expected concentration values, consider impurities in your span gas source (NO often contains 1-3% NO2 and vice versa). 165 04521C (DCN5731) Calibration Procedures Teledyne API - Model 200EH/EM Operation Manual STEP THREE: Perform Zero/Span Calibration: SAMPLE Analyzer continues to cycle through NO x, NO, and NO 2 measurements throughout this procedure. A1:NXCNC1=100PPM < TST TST > NOX=XXX.X CAL CALZ CALS SETUP Toggle TST> button until ... SAMPLE NOX STB= XXX.X PPM < TST TST > Set the Display to show the NOX STB test function. This function calculates the stability of the NO/NO x measurement NOX=XXX.X CAL CALZ CALS SETUP Allow zero gas to enter the sample port at the rear of the analyzer. Wait until NOX STB falls below 0.5 ppm. This may take several minutes. SAMPLE NOX STB= XXX.X PPM < TST TST > SAMPLE NOX SETUP GAS TO CAL:NOX O2 ENTR EXIT SAMPLE Analyzers enters ZERO cal mode. NOX=XXX.X CAL CALZ CALS RANGE TO CAL:LOW LOW HIGH ENTR EXIT ZERO CAL M NOX STB= XXX.X PPM NOX=XXX.X ZERO ZERO CAL M CONC EXIT NOX STB= XXX.X PPM NOX=XXX.X ENTR CONC EXIT Allow span gas to enter the sample port at the rear of the analyzer. Press ENTR to changes the OFFSET & SLOPE values for both the NO and NO x measurements. Press EXIT to leave the calibration unchanged and return to the previous menu. Wait until NOX STB falls below 0.5 ppm. This may take several minutes. SAMPLE NOX STB= XXX.X PPM < TST TST > SAMPLE NOX Analyzers enters SPAN cal mode and the SPAN key appears. You may see both keysduring the transition from ZERO to SPAN modes. If either the ZERO or SPAN buttons fail to appear see Section 11 for troubleshooting tips. CAL CALZ CALS NOX=XXX.X SETUP GAS TO CAL:NOX O2 SAMPLE ENTR EXIT RANGE TO CAL:LOW LOW HIGH ENTR EXIT SPAN CAL M NOX STB= XXX.X PPM ZERO SPAN CONC SPAN CAL M NOX STB= XXX.X PPM ENTR CONC SPAN CAL M NOX STB= XXX.X PPM ENTR CONC NOX=X.XXX EXIT NOX=X.XXX EXIT NOX=X.XXX EXIT Press ENTR to changes the OFFSET & SLOPE values for both the NO and NO x measurements. Press EXIT to leave the calibration unchanged and return to the previous menu. EXIT at this point returns to the SAMPLE menu. 166 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Calibration Procedures 7.5. CALIBRATION CHECKS WITH ZERO/SPAN VALVES Zero and span checks using the zero/span valve option are similar to that described in Section 7.4, except that zero air and span gas are supplied to the analyzer through the zero gas and span gas inlets from two different sources. Informal calibration checks, which only evaluate but do not alter the analyzer’s response curve, are recommended as a regular maintenance item and in order to monitor the analyzer’s performance. To carry out a calibration check rather than a full calibration, follow these steps. To perform a manual calibration check with zero/span valve or IZS option installed: STEP ONE: Connect the sources of Zero Air and Span Gas as shown in section 7-4. STEP TWO: Perform the zero/span check. Set the Display to show the NOX STB test function. This function calculates the stability of the NO/NO x measurement SAMPLE A1:NXCNC1=100PPM < TST TST > NOX=XXX.X CAL CALZ CALS SETUP Toggle TST> button until ... SAMPLE A1:NXCNC1=100PPM < TST TST > ZERO CAL M NOX STB= XXX.X PPM NOX=XXX.X ZERO CAL CALZ CALS SETUP SAMPLE < TST TST > Allow zero gas to enter the sample port at the rear of the analyzer. Wait until NOX STB falls below 0.5 ppm. CONC EXIT A1:NXCNC1=100PPM NOX=XXX.X NOX=XXX.X CAL CALZ CALS SETUP Allow span gas to enter the sample port at the rear of the analyzer. Wait until NOX STB falls below 0.5 ppm. This may take several minutes. SAMPLE A1:NXCNC1=100PPM < TST TST > This may take several minutes. NOX=XXX.X CAL CALZ CALS SETUP SAMPLE A1:NXCNC1=100PPM < TST TST > The ZERO and/or SPAN keys will appear at various points of this process. It is not necessary to press them. SAMPLE NOX Analyzers enters ZERO cal mode. NOX=XXX.X CAL CALZ CALS SETUP GAS TO CAL:NOX O2 ENTR EXIT SAMPLE NOX SAMPLE Return to SAMPLE Display GAS TO CAL:NOX O2 ENTR EXIT RANGE TO CAL:LOW LOW HIGH ENTR EXIT SAMPLE RANGE TO CAL:LOW LOW HIGH ZERO CAL M NOX STB= XXX.X PPM NOX=XXX.X ZERO CONC EXIT ENTR EXIT SPAN CAL M NOX STB= XXX.X PPM NOX=X.XXX ZERO SPAN CONC Record NO X, NO, NO 2 or O 2 zero point readings Analyzers enters SPAN cal mode. EXIT Record NO X, NO, NO 2 or O 2 span point readings\ SPAN CAL M NOX STB= XXX.X PPM NOX=XXX.X ZERO CONC EXIT Return to SAMPLE Display 167 04521C (DCN5731) Calibration Procedures Teledyne API - Model 200EH/EM Operation Manual 7.6. CALIBRATION WITH REMOTE CONTACT CLOSURES Contact closures for controlling calibration and calibration checks are located on the rear panel CONTROL IN connector. Instructions for setup and use of these contacts can be found in Section 6.15.1.2. When the appropriate contacts are closed for at least 5 seconds, the instrument switches into zero, low span or high span mode and internal zero/span valves (if installed) will be automatically switched to the appropriate configuration. The remote calibration contact closures may be activated in any order. It is recommended that contact closures remain closed for at least 10 minutes to establish a reliable reading; the instrument will stay in the selected mode for as long as the contacts remain closed. If contact closures are used in conjunction with the analyzer’s AutoCal (Section 7.7) feature and the AutoCal attribute CALIBRATE is enabled, the M200EH/EM will not re-calibrate the analyzer until the contact is opened. At this point, the new calibration values will be recorded before the instrument returns to SAMPLE mode. If the AutoCal attribute CALIBRATE is disabled, the instrument will return to SAMPLE mode, leaving the instrument’s internal calibration variables unchanged. 168 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Calibration Procedures 7.7. AUTOMATIC CALIBRATION (AUTOCAL) The AutoCal feature allows unattended, periodic operation of the zero/span valve options by using the analyzer’s internal time of day clock. The AutoCal feature is only available on the front panel menu (ACAL) if either the zero/span valve or the IZS option is installed. AutoCal operates by executing user-defined sequences to initiate the various calibration modes of the analyzer and to open and close valves appropriately. It is possible to program and run up to three separate sequences (SEQ1, SEQ2 and SEQ3). Each sequence can operate in one of three modes or be disabled (Table 7-2). Table 7-2: AutoCal Modes MODE ACTION DISABLED ZERO ZERO-LO 1 ZERO-LO-HI1 ZERO-HI LO1 LO-HI1 HI Disables the sequence Causes the sequence to perform a zero calibration or check Causes the sequence to perform a zero calibration or check followed by a mid-span concentration calibration or check Causes the sequence to perform a zero calibration or check followed by a mid-span concentration calibration or check and finally a high-span point calibration or check. Causes the sequence to perform a zero calibration or check followed by a high-span point calibration or check. Causes the sequence to perform a mid-span concentration calibration or check Causes the sequence to perform a mid-span concentration calibration or check followed by a high-span point calibration or check Causes the sequence to perform a high-span point calibration or check. O2 –ZERO2 Causes the sequence to do a zero-point calibration for the O2 sensor. Causes the sequence to perform a zero calibration of the or check O2 sensor followed O2 ZERO-SP by a mid-span concentration calibration or check of the O2 sensor. O2 SPAN2 Causes the sequence to perform a zero calibration or check of the O2 sensor. 1 Only applicable if analyzer is equipped with the second span point valve option (52) 2 Only applicable if instrument is equipped wit the O2 sensor option (65(. 2 Each mode has seven parameters that control operational details of the sequence(Table 7-3). Table 7-3: AutoCal Attribute Setup Parameters PARAMETER TIMER ENABLED STARTING DATE STARTING TIME DELTA DAYS DELTA TIME DURATION CALIBRATE RANGE TO CAL ACTION Turns on the sequence timer Sequence will operate on Starting Date Sequence will operate at Starting Time Number of days between each sequence trigger. If set to 7, for example, the AutoCal feature will be enabled once every week on the same day. Incremental delay on each delta day that the sequence starts. If set to 0, the sequence will start at the same time each day. Delta Time is added to Delta Days for the total time between cycles. This parameter prevents the analyzer from being calibrated at the same daytime of each calibration day and prevents a lack of data for one particular daytime on the days of calibration. Duration of the each sequence step in minutes. This parameter needs to be set such that there is enough time for the concentration signal to stabilize. The STABIL parameter shows if the analyzer response is stable at the end of the calibration. This parameter is logged with calibration values in the iDAS. Enable to do a true, dynamic zero or span calibration; disable to do a calibration check only. LOW calibrates the low range, HIGH calibrates the high range. Applies only to auto and remote range modes; this property is not available in single and independent range modes. 169 04521C (DCN5731) Calibration Procedures Teledyne API - Model 200EH/EM Operation Manual The following example sets sequence #2 to carry out a zero-span calibration every other day starting at 14:00 on 01 January, 2003, lasting 30 minutes (15 for zero and 15 for span). This sequence will start 30 minutes later each day. Table 7-4: Example Auto-Cal Sequence MODE AND ATTRIBUTE VALUE SEQUENCE 2 COMMENT Define sequence #2 MODE ZERO-HI TIMER ENABLE ON STARTING DATE 01-JAN-03 STARTING TIME 14:00 DELTA DAYS 2 DELTA TIME 00:30 Repeat sequence 30 minutes later each time (every 2 days and 30 minutes) DURATION 15.0 Each sequence step will last 15 minutes (total of 30 minutes when using zero-span mode) CALIBRATE ON The instrument will recalculate the slope and offset values for the NO and NOX channel at the end of the AutoCal sequence. Select zero and span mode Enable the timer Start on or after 01 January 2003 First sequence starts at 14:00 (24-hour clock format) Repeat this sequence every 2 days Please note the following suggestions for programming the AutoCal feature. The programmed Starting Time must be 5 minutes later than the real time clock (Section 6.10). Avoid setting two or more sequences at the same time of the day. Any new sequence which is initiated from a timer, the COM ports, or the contact closures will override any sequence in progress. Note that two sequences with different daily increments may eventually overlap. If at any time an illegal entry is selected, (for example: Delta Days > 366) the ENTR key will disappear from the display. With CALIBRATE turned on, the state of the internal setup variables DYN_SPAN and DYN_ZERO is set to ON and the instrument will reset the slope and offset values for the NO and NOX response each time the AutoCal program runs. This continuous re-adjustment of calibration parameters can often mask subtle fault conditions in the analyzer. It is recommended that, if CALIBRATE is enabled, the analyzer’s test functions, slope and offset values be checked frequently to assure high quality and accurate data from the instrument. 170 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Calibration Procedures To program the sample sequence shown above, follow this flow chart: SAMPLE RANGE = 500.0 PPB NOX=X.X SETUP < TST TST > CAL CALZ CZLS PRIMARY SETUP MENU SETUP X.X SEQ 1) DISABLED EXIT SEQ 2) DISABLED SETUP X.X EXIT MODE: DISABLED SETUP X.X 0 ENTR EXIT 0 MODE: ZERO – HI SETUP X.X SEQ 2) ZERO – HI, 1:00:00 SETUP X.X EXIT TIMER ENABLE: ON SETUP X.X EXIT STARTING DATE: 01 – JAN – 02 EDIT Toggle keys to set day, month & year: DDMON-Y Y SETUP X.X 0 4 SETUP X.X STARTING DATE: 01–JAN–02 SEP 0 3 ENTR SETUP C.4 STARTING DATE: 04 – SEP – 03 Toggle keys to set time: HH:MM. This is a 24 hr clock. PM hours are 13-24. Example: 2:15 PM = 14:15 SETUP C.4 STARTING TIME:00:00 SETUP C.4 1 4 EXIT 5 EXIT DELTA TIME: 00:00 :3 0 ENTR EXIT DURATION:15.0 MINUTES EXIT DURATION 15.0MINUTES .0 ENTR EXIT DURATION:30.0 MINUTES Toggle keys to set duration for each iteration of the sequence: Set in Decimal minutes from 0.1 – 60.0 EXIT CALIBRATE: OFF EXIT CALIBRATE: OFF ENTR EXIT Toggle key between Off and ON CALIBRATE: ON EXIT EDIT SEQ 2) ZERO – SPAN, 2:00:30 EXIT PREV NEXT MODE SET Mode ENTR EXIT Toggle keys to set delay time for each iteration of the sequence: HH:MM (0 – 24:00) DELTA TIEM:00:30 Sequence # STARTING TIME:00:00 :1 DELTA TIME00:00 ON SETUP C.4 EDIT EXIT EDIT SETUP C.4 EXIT EXIT Toggle keys to set number of days between procedures (1-367) DELTA DAYS:2 EDIT SETUP C.4 EXIT EDIT 0 SETUP C.4 EXIT STARTING DATE: 04 – SEP – 03 EDIT 3 SETUP C.4 EXIT ENTR EDIT SETUP C.4 SET> EDIT DELTA DAYS: 1 2 EDIT SETUP C.4 PREV NEXT MODE SET Default value is ON 0 SETUP C.4 ENTR EXIT EXIT EDIT SETUP C.4 Toggle NEXT button until ... PREV NEXT DELTA DAYS: 1 EDIT SETUP C.4 NEXT SETUP X.X 0 SETUP C.4 PREV NEXT MODE EXIT EDIT SETUP C.4 NEXT MODE STARTING TIME:14:15 EDIT SETUP C.4 CFG ACAL DAS RNGE PASS CLK MORE EXIT SETUP X.X SETUP C.4 EXIT returns to the SETUP Menu Delta Time Delta Days EXIT 171 04521C (DCN5731) Calibration Procedures Teledyne API - Model 200EH/EM Operation Manual 7.8. CALIBRATION QUALITY ANALYSIS After completing one of the calibration procedures described above, it is important to evaluate the analyzer’s calibration SLOPE and OFFSET parameters. These values describe the linear response curve of the analyzer, separately for NO and NOX. The values for these terms, both individually and relative to each other, indicate the quality of the calibration. To perform this quality evaluation, you will need to record the values of the following test functions (Section 6.2.1 or Appendix A-3), all of which are automatically stored in the iDAS channel CALDAT for data analysis, documentation and archival. NO OFFS NO SLOPE NOX OFFS NOX SLOPE Make sure that these parameters are within the limits listed in Table 7-5 and frequently compare them to those values on the Final Test and Checkout Sheet that came attached to your manual, which should not be significantly different. If they are, refer to the troubleshooting Chapter 11. Table 7-5: Calibration Data Quality Evaluation FUNCTION MINIMUM VALUE OPTIMUM VALUE MAXIMUM VALUE NOX SLOPE -0.700 1.000 1.300 NO SLOPE -0.700 1.000 1.300 NOX OFFS -20.0 mV 0.0 mV 150.0 mV NO OFFS -20.0 mV 0.0 mV 150.0 mV The default iDAS configuration records all calibration values in channel CALDAT as well as all calibration check (zero and span) values in its internal memory. Up to 200 data points are stored for up 4 years of data (on weekly calibration checks) and a lifetime history of monthly calibrations. Review these data to see if the zero and span responses change over time. These channels also store the STABIL value (standard deviation of NOX concentration) to evaluate if the analyzer response has properly leveled off during the calibration procedure. Finally, the CALDAT channel also stores the converter efficiency for review and documentation. USER NOTES: 172 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual EPA Protocol Calibration 8. EPA PROTOCOL CALIBRATION At the writing of this manual there is no EPA requirements for the monitoring of NOX or published calibration protocols. User Notes 173 04521C (DCN5731) 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Instrument Maintenance 9. INSTRUMENT MAINTENANCE Predictive diagnostic functions, including data acquisition records, failure warnings and test functions built into the analyzer, allow the user to determine when repairs are necessary without performing unnecessary, preventative maintenance procedures. There is, however, a minimal number of simple procedures that, when performed regularly, will ensure that the analyzer continues to operate accurately and reliably over its lifetime. Repair and troubleshooting procedures are covered in Chapter 11 of this manual. NOTE A span and zero calibration check must be performed following some of the maintenance procedures listed below. Refer to Chapter 0. CAUTION Risk of electrical shock. Disconnect power before performing any operations that require entry into the interior of the analyzer. NOTE The operations outlined in this chapter must be performed by qualified maintenance personnel only. 9.1. MAINTENANCE SCHEDULE Table 9-1 shows the recommended maintenance schedule for the M200EH/EM. Please note that in certain environments with high levels of dust, humidity or pollutant levels some maintenance procedures may need to be performed more often than shown. 175 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Instrument Maintenance Table 9-1: M200EH/EM Preventive Maintenance Schedule ITEM ACTION FREQUENCY CAL CHECK MANUAL SECTION Particulate Filter Change filter Weekly No 9.3.1 Verify Test Functions Review and evaluate Weekly No 9.2; Appendix C Zero/Span Check Evaluate offset and slope Weekly -- 7.3, 7.5, 7.7 Zero/Span Calibration Zero and span calibration Every 3 months -- 7.2, 7.4, 7.6, 7.7, 7,8 NO2 Converter Replace converter & check efficiency Every 3 years or if conversion efficiency < 96% Yes if CE factor is used -- 1 External Zero Air Scrubber (Optional) Exchange chemical Every 3 months No 9.3.5 External Dryer (Optional) Replace chemical When indicator color changes No Reaction Cell Window Clean optics, Change O-rings Annually or as necessary Yes 9.3.7 1 Air Inlet Filter Of Perma Pure Dryer Change particle filter Annually No 9.3.2 Pneumatic SubSystem Check for leaks in gas flow paths Annually or after repairs involving pneumatics Yes on leaks, else no 0, 0 1 All Critical Flow Orifice O-Rings & Sintered Filters Replace Annually Yes 9.3.8 Rebuild head Annually Yes 9.3.4 Inline Exhaust Scrubber Replace Annually No Pmt Sensor Hardware Calibration Low-level hardware calibration On PMT/ preamp changes & if 0.7< SLOPE >1.3 Yes 1 1 1 1, 2 1 2 Pump DATE PERFORMED 11.6.5 These Items are required to maintain full warranty, all other items are strongly recommended. A pump rebuild kit is available from Teledyne Instruments Customer Service including all instructions and required parts (see Appendix B for part numbers). 176 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Instrument Maintenance 9.2. PREDICTIVE DIAGNOSTICS The analyzer’s test functions can be used to predict failures by looking at trends in their values. Initially it may be useful to compare the state of these test functions to the values measured on your instrument at the factory and recorded on the M200EH/EM Final Test and Validation Data Form (Teledyne Instruments part number 04490, attached to the manual). Table 9-2 can be used as a basis for taking action as these values change with time. The internal data acquisition system (iDAS) is a convenient way to record and track these changes. APICOM control software can be used to download and review these data even from remote locations (Section 6.15.2.8 describes APICOM). Table 9-2: Predictive Uses for Test Functions FUNCTION EXPECTED RCEL pressure Constant to within ± 0.5 SAMPLE pressure Constant within atmospheric changes Ozone Flow Constant to within ± 15 ACTUAL INTERPRETATION & ACTION Fluctuating Developing leak in pneumatic system. Check for leaks Slowly increasing Pump performance is degrading. Replace pump head when pressure is above 10 in-Hg-A Fluctuating Developing leak in pneumatic system. Check for leaks Slowly increasing Flow path is clogging up. Replace orifice filters Slowly decreasing Developing leak in pneumatic system to vacuum (developing valve failure). Check for leaks Slowly decreasing Flow path is clogging up. Replace orifice filters Developing AZERO valve failure. Replace valve AZERO Constant within ±20 of check-out value Significantly increasing PMT cooler failure. Check cooler, circuit, and power supplies Developing light leak. Leak check. O3 air filter cartridge is exhausted. Change chemical Slowly decreasing signal for same concentration NO2 CONC Constant for constant concentrations NO2 CONC (IZS) Constant response from day to day Decreasing over time NO2 CONC (IZS) Constant response from day to day Heavily fluctuating from day to day NO CONC Constant for constant concentration Decreasing over time Converter efficiency may be degrading. Replace converter. Change in instrument response. Low level (hardware) calibrate the sensor Degradation of IZS permeation tube. Change permeation tube Ambient changes in moisture are affecting the performance. Add a dryer to the zero air inlet. Drift of instrument response; clean RCEL window, change O3 air filter chemical. 9.3. MAINTENANCE PROCEDURES The following procedures need to be performed regularly as part of the standard maintenance of the Model 200EH/EM. 177 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Instrument Maintenance 9.3.1. CHANGING THE SAMPLE PARTICULATE FILTER The particulate filter should be inspected often for signs of plugging or excess dirt. It should be replaced according to the service interval in Table 9-1 even without obvious signs of dirt. Filters with 1 and 5 µm pore size can clog up while retaining a clean look. We recommend to handle the filter and the wetted surfaces of the filter housing with gloves and tweezers. We recommend not to touch any part of the housing, filter element, PTFE retaining ring, glass cover and the O-ring with bare hands as this may cause the pores to clog quicker and surfaces to become dirty due to possible oils from your hands. Figure 9-1: Sample Particulate Filter Assembly To change the filter according to the service interval in Table 9-1, follow this procedure: 1. Turn OFF the pump to prevent drawing debris into the sample line. 2. Remove the CE Mark locking screw in the center of the front panel and open the hinged front panel and unscrew the knurled retaining ring of the filter assembly. 3. Carefully remove the retaining ring, glass window, PTFE O-ring and filter element. We recommend to clean the glass and O-rings at least once monthly, weekly in very polluted areas. 4. Install a new filter element, carefully centering it in the bottom of the holder. 5. Re-install the PTFE O-ring with the notches facing up (important!), the glass cover, then screw on the hold-down ring and hand-tighten the assembly. Inspect the (visible) seal between the edge of the glass window and the O-ring to assure proper gas tightness. 6. To fulfill CE Mark safety requirements, the front panel locking screw must be installed at all times during operation of the analyzer. 178 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Instrument Maintenance 7. Re-start the analyzer. 9.3.2. CHANGING THE O3 DRYER PARTICULATE FILTER The air for the O3 generator passes through a Perma Pure© dryer, which is equipped with a small particulate filter at its inlet. This filter prevents dust from entering the Perma Pure© dryer and degrading the dryer’s performance over time. To change the filter according to the service interval in Table 9-1: 1. Check and write down the average RCEL pressure and the OZONE flow values. 2. Turn off the analyzer, unplug the power cord and remove the cover. 3. Unscrew the nut around the port of the filter using 5/8” and 9/16” wrenches and by holding the actual fitting body steady with a 7/16” wrench. CAUTION Risk of significant leak. Make sure to use proper wrenches and to not turn the fitting against the Perma Pure© dryer. This may loosen the inner tubing and cause large leaks. 4. Take off the old filter element and replace it with a suitable equivalent (TAPI part# FL-3). Figure 9-2: Particle Filter on O3 Supply Air Dryer 5. Holding the fitting steady with a 5/8” wrench, tighten the nut with your hands. If necessary use a second wrench but do not over-tighten the nut. 6. Replace the cover, plug in the power cord and restart the analyzer. 7. Check the O3 flow rate, it should be around 80 cm³/min ± 15. Check the RCEL pressure, it should be the same value as before. 179 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Instrument Maintenance 9.3.3. MAINTAINING THE EXTERNAL SAMPLE PUMP 9.3.3.1. Rebuilding the Pump The sample pump head periodically wears out and must be replaced when the RCEL pressure exceeds 10 inHg-A (at sea level, adjust this value accordingly for elevated locations). A pump rebuild kit is available from the factory. Appendix B of this manual lists the part numbers of the pump rebuild kit. Instructions and diagrams are included in the kit. A flow and leak check after rebuilding the sample pump is recommended. A span check and re-calibration after this procedure is necessary as the response of the analyzer changes with the RCEL pressure. 9.3.3.2. Changing the Inline Exhaust Scrubber CAUTION! Do NOT attempt to change the contents of the inline exhaust scrubber cartridge; change the entire cartridge. 1. Through the SETUP>MORE>DIAG menu turn OFF the OZONE GEN OVERRIDE. Wait 10 minutes to allow pump to pull room air through scrubber before proceeding to step 2. 2. Disconnect exhaust line from analyzer. 3. Turn off (unplug) analyzer sample pump. 4. Disconnect tubing from (NOx or charcoal) scrubber cartridge. 5. Remove scrubber from system. 6. Dispose of according to local laws. 7. Install new scrubber into system. 8. Reconnect tubing to scrubber and analyzer. 9. Turn on pump. 10. Through the SETUP menu (per Step 1 above) turn ON the OZONE GEN OVERRIDE. NOTE: The inline exhaust scrubber is strictly intended for Nitric Acid and NO2 only. 9.3.4. CHANGING THE PUMP AND IZS DUST FILTERS The exhaust air from the analyzer passes a small particle filter (DFU filter, part # FL3) before entering the pump. When this particle filter becomes visibly dirty or the pressure drop between SAMP and RCEL pressure increases significantly, it needs replacement in order to prevent a large pressure drop with degraded analyzer performance. 180 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Instrument Maintenance 1. Power down the analyzer and pump. 2. For internally mounted filters, skip the next two steps. 3. For externally mounted filters on the pump housing, remove the analyzer exhaust tube from the dust filter. Remove the particle filter from the pump. To do so, push the white plastic ring into the fitting and pull the filter out of the fitting. If necessary, use needle-nose pliers to pry the filter out of the fittings. 4. Push a new filter into the pump fitting and make sure that the arrow on the filter points towards the pump. Push the exhaust tubing onto the filter. Skip the next two steps. 5. For internally mounted filters at the inside rear panel, remove the chassis and locate the filter between the vacuum manifold and the exhaust port fitting. 6. Disconnect the clear tubing from the filter body and change the filter with the arrow pointing against the gas flow. To remove the hose clamps, slide the two clamp ends in opposite directions with a needlenose pliers until the clamp comes apart. Reconnect the tubing by using the same or new clamps and pushing tightening them until a good seal is achieved. 7. Restart the pump and clear any error warnings from the front panel display. 8. After about 5 minutes, check the RCEL pressure reading and ensure that it is similar to its value before changing the filter but less than 10 in-Hg-A. 181 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Instrument Maintenance 9.3.5. CHANGING THE EXTERNAL ZERO AIR SCRUBBER The external zero air scrubber contains two chemicals, pink Purafil© (Part # CH 9) and black, charcoal (Part # CH 1). The Purafil© converts NO in the ambient air to NO2 and the following charcoal absorbs any NO2. The chemicals need to be replaced periodically according to Table 9-1 or as needed. This procedure can be carried out while the instrument is running. Make sure that the analyzer is not in ZERO calibration mode. CAUTION! The following procedures apply only to the External Zero Air Scrubber and NOT to the inline exhaust scrubber cartridge (Section 9.3.3.2) that is part of the pump pack assembly. 1. Locate the scrubber on the outside rear panel (for location, see Scrubber Cartridge in Figure 3-2). Figure 9-3 shows the exploded scrubber assembly. 2. Remove the old scrubber by disconnecting the 1/4” plastic tubing from the particle filter using 9/16” and 1/2" wrenches. 3. Remove the particle filter from the cartridge using 9/16” wrenches. 4. Unscrew the top of the scrubber canister and discard the Purafil© and charcoal contents. Make sure to abide to local laws about discarding these chemicals. The rebuild kit (listed in Appendix B) comes with a Material and Safety Data Sheet, which contains more information on these chemicals. 5. Refill the scrubber with charcoal at the bottom and with the Purafil© chemical at the top, and use three white retainer pads (Figure 9-3) to separate the chemicals. 6. Replace the screw-top cap and tighten the cap - hand-tight only. 7. If necessary, replace the DFU filter with a new unit and discard the old. The bottom retainer pad should catch most of the dust, the filter should not be visibly dirty (on the inside) 8. Replace the scrubber assembly into its clips on the rear panel. 9. Reconnect the plastic tubing to the fitting of the particle filter. 10. Adjust the scrubber cartridge such that it does not protrude above or below the analyzer in case the instrument is mounted in a rack. If necessary, squeeze the clips for a tighter grip on the cartridge. Figure 9-3: 182 Zero Air Scrubber Assembly Teledyne API - Model 200EH/EM Operation Manual Instrument Maintenance 9.3.6. CHANGING THE NO2 CONVERTER The NO2 converter is located in the center of the instrument, see Figure 3-1 for location, and Figure 9-4 for the assembly. The converter is designed for replacement of the cartridge only, the heater with built-in thermocouple can be reused. 1. Turn off the analyzer power, remove the cover and allow the converter to cool. 2. Remove the top lid of the converter as well as the top layers of the insulation until the converter cartridge can be seen. CAUTION THE CONVERTER OPERATES AT 315º C. SEVERE BURNS CAN RESULT IF THE ASSEMBLY IS NOT ALLOWED TO COOL. DO NOT HANDLE THE ASSEMBLY UNTIL IT IS AT ROOM TEMPERATURE. THIS MAY TAKE SEVERAL HOURS. 3. Remove the tube fittings from the converter. 4. Disconnect the power and the thermocouple of the converter. Unscrew the grounding clamp of the power leads with a Phillips-head screw driver. 5. Remove the converter assembly (cartridge and band heater) from the can. Make a note of the orientation of the tubes relative to the heater cartridge. 6. Unscrew the band heater and loosen it, take out the old converter cartridge. Figure 9-4: NO2 Converter Assembly 7. Wrap the band heater around the new replacement cartridge and tighten the screws using a hightemperature anti-seize agent such as copper paste. Make sure to use proper alignment of the heater with respect to the converter tubes. 8. Replace the converter assembly, route the cables through the holes in the can and reconnect them properly. Reconnect the grounding clamp around the heater leads for safe operation. 9. Re-attach the tube fittings to the converter and replace the insulation and cover. 10. Replace the instrument cover and power up the analyzer. 11. Allow the converter to burn-in for 24 hours, then re-calibrate the instrument. 183 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Instrument Maintenance 9.3.7. CLEANING THE REACTION CELL The reaction cell should be cleaned whenever troubleshooting suggests. A dirty reaction cell will cause excessive noise, drifting zero or span values, low response or a combination of all. To clean the reaction cell, it is necessary to remove it from the sensor housing. refer to Section 11.6.6. for an overview of the entire sensor assembly. Use the following guide to clean the reaction cell: 1. Turn off the instrument power and vacuum pump. Refer to the Figure 9-5 for the following procedure. 2. Disconnect the black 1/4" exhaust tube and the 1/8” sample and ozone air tubes from the reaction cell. Disconnect the heater/thermistor cable. 3. Remove four screws holding the reaction cell to the PMT housing and lift the cell and manifold out as shown in the inset of Figure 9-5. 000940500 – Ozone Critical Flow Orifice Figure 9-5: Reaction Cell Assembly 4. The reaction cell will separate into two halves, the stainless steel manifold assembly and the black plastic reaction cell with window, stainless steel cylinder and O-rings. 5. The reaction cell (both plastic part and stainless steel cylinder) and optical glass filter should be cleaned with methanol and a clean tissue and dried thereafter. 6. Usually it is not necessary to clean the ozone flow orifice since it is protected by a sintered filter. If tests show that cleaning is necessary, refer to Section 9.3.8 on how to clean the critical flow orifice. 184 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Instrument Maintenance 7. Do not remove the sample and ozone nozzles. They are Teflon threaded and require a special tool for reassembly. If necessary, the manifold with nozzles attached can be cleaned in an ultrasonic bath. 8. Reassemble in proper order and re-attach the reaction cell to the sensor housing. Reconnect pneumatics and heater connections, then re-attach the pneumatic sensor assembly and the cleaning procedure is complete. 9. After cleaning the reaction cell, it is also recommended to exchange the ozone supply air filter chemical 10. After cleaning, the analyzer span response may drop 10 - 15% in the first 10 days as the reaction cell window conditions. This is normal and does not require another cleaning. 9.3.8. CHANGING CRITICAL FLOW ORIFICES There are several critical flow orifices installed in the M200EH/EM, Figure 9-5 shows one of the two most important orifice assemblies, located on the reaction cell. Refer to Section 10.3.3 for a detailed description on functionality and locations. Despite the fact that these flow restrictors are protected by sintered stainless steel filters, they can, on occasion, clog up, particularly if the instrument is operated without sample filter or in an environment with very fine, sub-micron particle-size dust. The M200EH/EM introduces an orifice holder that makes changing the orifice very easy. In fact, it is recommended to keep spare orifice holder assemblies at hand to minimize downtime and swap orifices in a matter of a few minutes. Appendix B lists several complete spare part kits for this purpose. To replace a critical flow orifice, do the following: 1. Turn off power to the instrument and vacuum pump. Remove the analyzer cover and locate the reaction cell (Figure 9-5, Figure 11- and Figure 3-4). 2. Unscrew the 1/8” sample and ozone air tubes from the reaction cell 3. For orifices on the reaction cell (Figure 9-5): Unscrew the orifice holder with a 9/16” wrench. This part holds all components of the critical flow assembly as shown in Figure 9-6. Appendix B contains a list of spare part numbers. 4. For orifices in the vacuum manifold: the assembly is similar to the one shown in Figure 9-6, but without the orifice holder, part number 04090, and bottom O-ring OR34 and with an NPT fitting in place of the FT 10 fitting. After taking off the connecting tube, unscrew the NPT fitting. 185 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Instrument Maintenance Figure 9-6: Critical Flow Orifice Assembly 5. Take out the components of the assembly: a spring, a sintered filter, two O-rings and the orifice. For the vacuum manifold only, you may need to use a scribe or pressure from the vacuum port to get the parts out of the manifold. 6. Discard the two O-rings and the sintered filter and the critical flow orifice. 7. Re-assemble the flow control assembly with new the parts (see Appendix B for part number or replacement kit) as shown in Figure 9-6 and re-connect them to the reaction cell manifold or the vacuum manifold. 8. Reconnect all tubing, power up the analyzer and pump and - after a warm-up period of 30 minutes, carry out a leak test as described in Section 0. 9.3.9. CHECKING FOR LIGHT LEAKS When re-assembled or operated improperly, the M200EH/EM can develop small leaks around the PMT, which let stray light from the analyzer surrounding into the PMT housing. To find such light leaks, follow the below procedures. CAUTION: this procedure can only be carried out with the analyzer running and its cover removed. This procedure should only be carried out by qualified personnel. 1. Scroll the TEST functions to PMT. 2. Supply zero gas to the analyzer. 3. With the instrument still running, carefully remove the analyzer cover. Take extra care not to touch any of the inside wiring with the metal cover or your body. Do not drop screws or tools into a running analyzer! 4. Shine a powerful flashlight or portable incandescent light at the inlet and outlet fitting and at all of the joints of the reaction cell as well as around the PMT housing. The PMT value should not respond to the light, the PMT signal should remain steady within its usually noise. 186 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Instrument Maintenance 5. If there is a PMT response to the external light, symmetrically tighten the reaction cell mounting screws or replace the 1/4” vacuum tubing with new, black PTFE tubing (this tubing will fade with time and become transparent). Often, light leaks are also caused by O-rings being left out of the assembly. 6. Carefully replace the analyzer cover. 7. If tubing was changed, carry out a leak check (Section 0). USER NOTES: 187 04521C (DCN5731) 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10. THEORY OF OPERATION The M200EH/EM Nitrogen Oxides Analyzer is a microprocessor controlled instrument that determines the concentration of nitric oxide (NO), total nitrogen oxides (NOX, the sum of NO and NO2) and nitrogen dioxide (NO2) in a sample gas drawn through the instrument. It requires that sample and calibration gases are supplied at ambient atmospheric pressure in order to establish a constant gas flow through the reaction cell where the sample gas is exposed to ozone (O3), initiating a chemical reaction that gives off light (chemiluminescence). The instrument measures the amount of chemiluminescence to determine the amount of NO in the sample gas. A catalytic-reactive converter converts any NO2 in the sample gas to NO, which is then – including the NO in the sample gas – is then reported as NOX. NO2 is calculated as the difference between NOX and NO. Calibration of the instrument is performed in software and usually does not require physical adjustments to the instrument. During calibration, the microprocessor measures the sensor output signal when gases with known amounts of NO or NO2 are supplied and stores these results in memory. The microprocessor uses these calibration values along with the signal from the sample gas and data of the current temperature and pressure of the gas to calculate a final NOX concentration. The concentration values and the original information from which it was calculated are stored in the unit’s internal data acquisition system (iDAS Section 6.7.2) and are reported to the user through a vacuum fluorescence display or several output ports. 10.1. MEASUREMENT PRINCIPLE 10.1.1. CHEMILUMINESCENCE The principle of the M200EH/EM’s measurement method is the detection of chemiluminescence, which occurs when nitrogen oxide (NO) reacts with ozone (O3). This reaction is a two-step process. In the first step, one molecule of NO and one molecule of O3 collide and chemically react to produce one molecule of oxygen (O2) and one molecule of nitrogen dioxide (NO2). Some of the NO2 retains a certain amount of excess energy from the collision and, hence, remains in an excited state, which means that one of the electrons of the NO2 molecule resides in a higher energy state than is normal (denoted by an asterisk in Equation 10-1). NO + O3 → NO2* + O2 (Equation 10-1) Thermodynamics requires that systems seek the lowest stable energy state, hence, the NO2 molecule quickly returns to its ground state in a subsequent step, releasing the excess energy in form of a quantum of light (h) with wavelengths between 600 and 3000 nm, with a peak at about 1200 nm (Equation 10-2, Figure 10-10-1). NO2* → NO2 + hν (Equation 10-2) All things being constant, the relationship between the amount of NO present in the reaction cell and the amount of light emitted from the reaction is very linear. More NO produces more light, which can be measured with a light-sensitive sensor in the near-infrared spectrum (Figure 10-10-1). In order to maximize the yield of reaction (1), the M200EH/EM supplies the reaction cell with a large, constant excess of ozone (about 3000-5000 ppm) from the internal ozone generator. 189 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation Model 200E Instrument Response Intensity 140 a.u. 120 a.u. NO + O3 Emission Spectrum 100 a.u. 80 a.u. 60 a.u. PMT Response 40 a.u. Optical Hi-Pass Filter Performance 20 a.u. 0 a.u. 0.5µm 0.7µm 0.9µm 1.1µm 1.3µm 1.5µm 1.7µm 1.9µm Wavelength M200EH/EM Sensitivity Window Figure 10-10-1: M200EH/EM Sensitivity Spectrum However, only about 20% of the NO2 that is formed through reaction 10-1 is in the excited state. In addition, the excited NO2 can collide with another collision partner M in the reaction cell (mostly other molecules but also cell walls) and transfer its excess energy to its collision partner without emitting any light at all (Equation 10-3). In fact, by far the largest portion of the NO2* returns to the ground state this way, leaving only a few percent yield of usable chemiluminescence. NO2* + M → NO2 + M (Equation 10-3) In order to enhance the light yield of the reaction, the reaction cell is maintained at reduced pressure. The probability of a collision between the NO2* molecule and a collision partner M increases proportionally with the reaction cell pressure. This non-radiating collision with the NO2* molecules is usually referred to as quenching, an unwanted process further described in Section 10.2.4.2. 10.1.2. NOX AND NO2 DETERMINATION The only gas that is truly measured in the M200EH/EM is NO. Any NO2 contained in the gas is not detected in the above process since NO2 does not react with O3 to undergo chemiluminescence. In order to measure the concentration of NO or NOX (which is defined here as the sum of NO and NO2 in the sample gas), the M200EH/EM periodically switches the sample gas stream through a converter cartridge filled with molybdenum (Mo, “moly”) chips heated to a temperature of 315° C. The heated molybdenum reacts with NO2 in the sample gas and produces a variety of molybdenum oxides and NO according to Equation 10-4. xNO2 + yMo → xNO + M y Oz (at 315° C ) (Equation 10-4) Once the NO2 in the sample gas has been converted to NO, it is routed to the reaction cell where it undergoes the chemiluminescence reaction described in Equations 10-1 and 10-2. 190 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation Figure 10-10-2: NO2 Conversion Principle By converting the NO2 in the sample gas into NO, the analyzer can measure the total NOX (NO+NO2) content of the sample gas. By switching the NO2 converter in and out of the sample gas stream every 6 - 10 seconds, the M200EH/EM analyzer is able to quasi-continuously measure both the NO and the total NOX content. The NO2 concentration, finally, is not measured but calculated by simply subtracting the known NO content of the sample gas from the known NOX content. 10.2. CHEMILUMINESCENCE DETECTION 10.2.1. THE PHOTO MULTIPLIER TUBE The M200EH/EM uses a photo-multiplier tube (PMT) to detect the amount of light created by the NO and O3 reaction in the reaction cell. A PMT is typically a vacuum tube containing a variety of specially designed electrodes. Photons enter the PMT and strike a negatively charged photo cathode causing it to emit electrons. These electrons are accelerated by an applied high voltage and multiply through a sequence of such acceleration steps (dynodes) until a useable current signal is generated. This current increases or decreases with the amount of detected light (Section 10.4.3 for more details), is converted to a voltage and amplified by the preamplifier board and then reported to the motherboard’s analog inputs. 191 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation Figure 10-10-3: Reaction Cell with PMT Tube 10.2.2. OPTICAL FILTER Another critical component in the method by which your M200EH/EM detects chemiluminescence is the optical filter that lies between the reaction cell and the PMT (Figure: 10-3). This filter is a high pass filter that is only transparent to wavelengths of light above 645 nm. In conjunction with the response characteristics of the PMT, this filter creates a very narrow window of wavelengths of light to which the M200EH/EM will respond (Figure 1010-1). The narrow band of sensitivity allows the M200EH/EM to ignore extraneous light and radiation that might interfere with the M200EH/EM’s measurement. For instance, some oxides of sulfur can also undergo chemiluminescence when in contact with O3 but emit light at shorter wavelengths (usually around 260 nm to 480 nm). 10.2.3. AUTO ZERO Inherent in the operation of any PMT is a certain amount of noise. This is due to a variety of factors such as black body infrared radiation given off by the metal components of the reaction cell, unit to unit variations in the PMT units and even the constant universal background radiation that surrounds us at all times. In order to reduce this amount of noise and offset, the PMT is kept at a constant 7° C (45° F) by a thermo-electric cooler (TEC). While this intrinsic noise and offset is significantly reduced by cooling the PMT, it is not eradicated. To determine how much noise remains, the M200EH/EM diverts the sample gas flow directly to the vacuum manifold without passing the reaction cell once every minute for about 5 seconds (Figure 10-10-4). During this time, only O3 is present in the reaction cell, effectively turning off the chemiluminescence reaction. Once the chamber is completely dark, the M200EH/EM records the output of the PMT and keeps a running average of these AZERO values. This average offset value is subtracted from the raw PMT readings while the instrument is measuring NO and NOX to arrive at a auto-zero corrected reading. 192 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation Figure 10-10-4: Reaction Cell During the AutoZero Cycle. 10.2.4. MEASUREMENT INTERFERENCES It should be noted that the chemiluminescence method is subject to interferences from a number of sources. The M200EH/EM has been successfully tested for its ability to reject interference from most of these sources. Table 10-1 lists the most important gases, which may interfere with the detection of NO in the M200EH/EM. 10.2.4.1. Direct Interference Some gases can directly alter the amount of light detected by the PMT due to chemiluminescence in the reaction cell. This can either be a gas that undergoes chemiluminescence by reacting with O3 in the reaction cell or a gas that reacts with other compounds and produces excess NO upstream of the reaction cell. 10.2.4.2. Third Body Quenching As shown in Equation 10-3, other molecules in the reaction cell can collide with the excited NO2*, preventing the chemiluminescence of Equation 10-2, a process known as quenching. CO2 and H2O are the most common quenching interferences, but N2 and O2 also contribute to this interference type. Quenching is an unwanted phenomenon and the extent to which it occurs depends on the properties of the collision partner. larger, more polarized molecules such as H2O and CO2 quench NO chemiluminescence more effectively than smaller, less polar and electronically “harder” molecules such as N2 and O2. The influence of water vapor on the M200EH/EM measurement can be eliminated with an optional, internal sample gas dryer. The concentrations of N2 and O2 are virtually constant in ambient air measurements, hence provide a constant amount of quenching and the interference of varying CO2 amounts is negligible at low concentrations. 193 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation The M200EH and 200EM analyzers are typically used in high CO2 concentration environments. The pneumatic setup of these two analyzer models minimizes the interference from CO2 such that the analyzers conform to the standards set forth by the US-EPA in Method 20 - NOx from Stationary Gas Turbines, available at http://www.epa.gov/ttn/emc/promgate.html Table 10-1: GAS CO2 SOX List of Interferents INTERFERENCE TYPE REJECTION METHOD Dilution: Viscosity of CO2 molecules causes them to collect in aperture of Critical Flow Orifice altering flow rate of NO. If high concentrations of CO2 are suspected, special calibration methods must be performed to account for the affects of the CO2. 3rd Body Quenching: CO2 molecules collide with NO2* molecules absorbing excess energy kinetically and preventing emission of photons. Contact Teledyne Instruments Customer Service department for details. Some SOX variants can also initiate a chemiluminescence reaction upon exposure to O3 producing excess light. Wavelengths of light produced by chemiluminescence of SOX are screened out by the Optical Filter. Chemically reacts with NH3, O2 and H2O in O3 generator to create (NH3)2SO4 (ammonium sulfate) and NH3NO2 (ammonium nitrate) which form opaque white deposits on optical filter window. Also forms highly corrosive HNO3 (Nitric Acid) Most of the ammonium sulfate and ammonium nitrate produced is removed from the sample gas by an air purifier located between the O3 Generator and the reaction cell. 3rd Body quenching: SOX molecules collide with NO2* molecules absorbing excess energy kinetically and preventing emission of photons. If high concentrations of SOX are suspected, special calibration methods must be performed to account for the affects of the SO2. Contact Teledyne Instruments Customer Service department for details. H20 NH3 3rd Body quenching: H2O molecules collide with NO2* molecules absorbing excess energy kinetically and preventing emission of photons. Analyzer’s operating in high humidity areas must have some method of drying applied to the sample gas supply (Section 5.10 for more details). Chemically reacts with NH3 and SOX in O3 generator to create (NH3)2SO4 (ammonium sulfate) and NH3NO2 (ammonium nitrate) which form opaque white deposits on optical filter Window. Also forms highly corrosive HNO3 (nitric acid) Removed from the O3 gas stream by the Perma Pure® Dryer (Section 10.3.7 for more details). Direct Interference: NH3 is converted to H2O and NO by the NO2 converter. Excess NO reacts with O3 in reaction cell creating excess chemiluminescence. If a high concentration of NH3 is suspected, steps must be taken to remove the NH3 from the sample gas prior to its entry into the NO2 converter. Chemically reacts with H2O, O2 and SOX in O3 generator to create (NH3)2SO4 (ammonium sulfate) and NH3NO2 (ammonium nitrate) which form opaque white deposits on optical filter window. Also forms highly corrosive HNO3 (nitric acid). The Perma Pure® dryer built into the M200EH/EM is sufficient for removing typical ambient concentration levels of NH3. In cases with excessively high CO2 concentrations (larger than 0.5%), the effect can be calibrated out by using calibration gases with a CO2 content equal to the measured air. Only very high and highly variable CO2 concentrations will then be cause of measurable interference. For those applications, we recommend to use other analyzer models. Please consult sales or our website. 10.2.4.3. Light Leaks The M200EH/EM sensitivity curve includes a small portion of the visible light spectrum (Figure 10-1), hence, it is important to make sure than the reaction cell is completely sealed with respect to light. To ensure this, all pneumatic tubing leading into the reaction cell is either opaque (vacuum exit tubing) in order to prevent light from entering the cell or light penetration is prevented by stainless steel filters and orifices (gas entries). 194 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.3. PNEUMATIC OPERATION CAUTION It is important that the sample airflow system is leak-tight and not pressurized over ambient pressure. Regular leak checks should be performed on the analyzer as described in the maintenance schedule, Table 9-1. Procedures for correctly performing leak checks can be found in Section 11.5. 10.3.1. PUMP AND EXHAUST MANIFOLD NOTE Relative Pressure versus absolute pressure. In this manual vacuum readings are given in inches of mercury absolute pressure (in-Hg-A), i.e. indicate an absolute pressure referenced against zero (a perfect vacuum). The gas flow for the M200EH/EM is created by an external pump (Figure 10-10-5) that is pneumatically connected through a 6.4 mm / 0.25” tube to the analyzer’s exhaust port located on the rear panel (Figure 3-1). This pump creates a vacuum of approximately 5 in-Hg-A at one standard liter/minute, which is provided to various pneumatic components by a vacuum manifold located just in front of the rear panel (Figure 3-1). Gas flow is created by keeping the analyzer’s sample gas inlet near ambient pressure, usually by means of a small vent installed in the sample line at the inlet, in effect pulling the gas through the instrument’s pneumatic systems. There are several advantages to this external pump / pull-through configuration. By using an external pump, it is possible to remove a significant source of acoustic noise and vibration from the immediate vicinity of the sensor. The PMT can act as a “microphone”, amplifying noise and vibration within the chassis. This is one of the main reasons, why the M200EH/EM has an external pump. Pumping heats and compresses the sample air, complicating the measurement process if the pump is upstream. Most importantly, however, certain physical parts of the pump itself are made of materials that might chemically react with the sample gas. Placing the pump downstream of the reaction cell avoids these problems. To M200EH/EM Exhaust Port Figure 10-10-5: External Pump Pack 195 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation Finally, the M200EH/EM requires a steady, high under-pressure, which cannot be achieved reliably over extended periods of time with small vacuum pumps. The external pump used for the M200EH/EM has a very long lifetime and duty cycle and provides a very good vacuum for its entire lifetime. However, the pump is too large to fit into the chassis of the analyzer. 10.3.2. SAMPLE GAS FLOW The sample gas is the most critical flow path in the analyzer, as the medium has to be routed through a variety of valves and tubes for the measurement of zero offset and concentrations of both NO and NOX (and possibly the drying of the gas if the optional sample dryer is installed). At any point before and in the reaction cell, the integrity of the sample gas cannot be compromised. Sample gas flow in the M200EH/EM analyzer is not a directly measured value, but is rather calculated from the sample pressure using the flow principle across a critical orifice. In general, the differential pressure ratio between sample pressure and reaction cell pressure needs to exceed 2:1 to allow critical flow. The actual flow rate is then only dependent on the size of the orifice and the upstream pressure. Refer to Section 10.3.3 for a detailed description of the instrument’s method of gas flow rate control. 10.3.2.1. NO/NOx and AutoZero cycles For the routing of the sample gas flow, the analyzer uses a variety of valves. The NO/NOX valve directs the sample gas either directly to the reaction cell or through the unit’s NO2 converter, alternating every ~4 s. The AutoZero valve directs the sample gas stream to completely bypass the reaction cell for dark noise measurement once every minute, which is then subtracted as a measurement offset from the raw concentration signal. The valve cycle phases are summarized in the following table. Table 10-2: M200EH/EM Valve Cycle Phases PHASE NO/ NOX VALVE STATUS NO Measure Open to AutoZero valve NOX Measure Open to NO2 converter AUTOZERO VALVE STATUS Open to reaction cell Open to reaction cell TIME INDEX ACTIVITY 0-2s Wait period (NO dwell time). Ensures reaction cell has been flushed of previous gas. 2-4s Analyzer measures chemiluminescence in reaction cell. 4–6s Wait period (NOX dwell time). Ensures reaction cell has been flushed of previous gas. 6–8s Analyzer measures NO + O3 chemiluminescence in reaction cell. 0–4s Wait period (AZERO dwell time). Ensures reaction cell has been flushed of sample gas and chemiluminescence reaction is stopped. FIGURE Figure 10-10-2 Figure 10-10-2 Cycle repeats every ~8 seconds AutoZero Open to AutoZero valve Open to vacuum manifold 4-6s Figure 10-10-4 Analyzer measures background noise without sample gas Cycle repeats every minute 196 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.3.3. FLOW RATE CONTROL - CRITICAL FLOW ORIFICES The Model M200EH/EM analyzers use special flow control assemblies located at various locations around the instrument to maintain constant flow rates for both the O3 supply air and the sample gas. These assemblies consists of: A critical flow orifice. Two o-rings: Located just before and after the critical flow orifice, the o-rings seal the gap between the walls of assembly housing and the critical flow orifice. A spring: Applies mechanical force needed to form the seal between the o-rings, the critical flow orifice and the assembly housing. See Figures 10-6 through 10-9 For the location of these flow control assemblies: VACUUM PRESSURE SENSOR SAMPLE PRESSURE SENSOR GAS FLOW CONTROL ASSEMBLIES Figure 10-10-6: Location of Gas Flow Control Assemblies for M200EH 197 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation VACUUM PRESSURE SENSOR SAMPLE PRESSURE SENSOR GAS FLOW CONTROL ASSEMBLIES Figure 10-10-7: Location of Gas Flow Control Assemblies for M200EM GAS FLOW CONTROL ASSEMBLIES VACUUM PRESSURE SENSOR SAMPLE PRESSURE SENSOR Figure 10-10-8: Location of Gas Flow Control Assemblies for M200EH with O2 sensor Option 65 198 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation VACUUM PRESSURE SENSOR SAMPLE PRESSURE SENSOR GAS FLOW CONTROL ASSEMBLIES Figure 10-10-9: Location of Gas Flow Control Assemblies for M200EH with Second Span Point Option 52 NOTE: Location of flow control assemblies in the M200EH/EM with zero/span option 50 installed are the same as shown in Figures 10-6 and 10-7. 10.3.3.1. Critical Flow Orifice The most important component of the flow control assemblies is the critical flow orifice. Critical flow orifices are a remarkably simple way to regulate stable gas flow rates. They operate without moving parts by taking advantage of the laws of fluid dynamics. By restricting the flow of gas though the orifice, a pressure differential is created. This pressure differential combined with the action of the analyzer’s pump draws the gas through the orifice. As the pressure on the downstream side of the orifice (the pump side) continues to drop, the speed that the gas flows though the orifice continues to rise. Once the ratio of upstream pressure to downstream pressure is greater than 2:1, the velocity of the gas through the orifice reaches the speed of sound. As long as that ratio stays at least 2:1 the gas flow rate is unaffected by any fluctuations, surges, or changes in downstream pressure because such variations only travel at the speed of sound themselves and are therefore cancelled out by the sonic shockwave at the downstream exit of the critical flow orifice. 199 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation Figure 10-10-10: Flow Control Assembly & Critical Flow Orifice The actual flow rate of gas through the orifice (volume of gas per unit of time), depends on the size and shape of the aperture in the orifice. The larger the hole, the more gas molecules, moving at the speed of sound, pass through the orifice. With nominal pressures of 28 and 4 in-Hg-A for the sample and reaction cell pressures, respectively the necessary ratio of sample to reaction cell pressure of 2:1 is largely exceeded and accommodates a wide range of possible variability in atmospheric pressure and pump degradation extending the useful life of the pump. Once the pump does degrades to the point where the vacuum pressure exceeds 14 in-Hg-A so that the ratio between sample and vacuum pressures is less than 2:1 a critical flow rate can no longer be maintained. At this point, the instrument will display “XXXX" indicating an invalid sample flow rate. The following table lists the gas flow rates of the critical flow orifices in the standard M200EH/EM Table 10-3: M200EH/EM Critical Flow Orifice Diameters and Gas Flow Rates LOCATION PURPOSE ORIFICE DIAMETER NOMINAL FLOWRATE (cm³/min) M200EH M200EM M200EH M200EM Bypass Manifold out to No/NOx valve and NO2 converter Controls rate of flow of sample gas into the NO2 converter and reaction cell. 0.003” 0.007” 40 250 Vacuum Manifold: Bypass Manifold Port Controls rate of sample gas flow that bypasses the analyzer when bypassing the reaction cell during the auto-zero cycle. 0.007” N/A 250 N/A 290 250 80 80 370 330 TOTAL INLET GAS FLOW – Standard Configuration Controls rate of flow of zero purge gas through the O2 sensor (when installed and enabled) when inactive. Vacuum manifold: O2 sensor port 0.004" 0.004" TOTAL INLET GAS FLOW – With O2 Sensor Option O3 supply inlet of reaction cell. Dry air return of Perma Pure® dryer Controls rate of flow of ozone gas into the reaction cell. 0.007” 0.007” 250 250 Controls flow rate of dry air return / purge air of the dryer. 0.004" 0.004" 80 80 200 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation In addition to controlling the gas flows, the two critical flow orifices at the inlets of the reaction cell also maintain an under-pressure inside the reaction cell, effectively reducing the number of molecules in the chamber and therefore increasing the chemiluminescence yield as the likelihood of third body quenching is reduced (Section 10.2.4.1). The M200EH/EM sensitivity reaches a peak at about 2 in-Hg-A, below which the sensitivity drops due to a low number of molecules and decreased yield in the chemiluminescence reaction. EFFECT OF TEMPERATURE ON CRITICAL FLOW Changes in temperature will cause the critical flow orifice materials to expand or contract. Even though these changes are extremely small, they can alter the diameter of the critical flow orifice enough to cause noticeable changes in the flow rate though the orifice. To alleviate this problem the two most important of the flow assemblies (those controlling the sample gas an O3 gas flow)in the M200EH/EM are maintained at a constant temperature. 10.3.4. SAMPLE PARTICULATE FILTER To remove particles in the sample gas, the analyzer is equipped with a PTFE membrane filter of 47 mm diameter (also referred to as the sample filter) with a 1 µm pore size. The filter is accessible through the front panel, which folds down (after removal of the CE Mark safety screw), and should be changed according to the maintenance schedule in Table 9-1. 5 10.3.5. OZONE GAS AIR FLOW The excess ozone needed for reaction with NO in the reaction cell is generated inside the analyzer because of the instability and toxicity of ozone. Besides the ozone generator itself, this requires a dry air supply and filtering of the gas before it is introduced into the reaction cell. Due to its toxicity and aggressive chemical behavior, O3 must also be removed from the gas stream before it can be vented through the exhaust outlet. In contrast to the sample flow, the ozone flow is measured with a mass flow sensor, which is mounted on the pneumatic sensor board (Figure 11-), just behind the PMT sensor assembly. This mass flow sensor has a full scale range of 0-1000 cm³/min and can be calibrated through software to its span point (Section 6.13.7.5). As the flow value displayed on the front panel is an actual measurement (and not a calculated value), the flow variability may be higher than that of the sample flow, which is based on a calculation from (more stable) differential pressures. On the other hand, the drift, i.e. long-term change, in the ozone flow rate may be higher and usually indicates a flow problem. As with all other test parameters, we recommend to monitor the ozone flow over time for predictive diagnostics and maintenance evaluation. CAUTION Ozone (O3) is a toxic gas. Obtain a Material and Safety Data Sheet (MSDS) for this gas. Read and rigorously follow the safety guidelines described there. Always make sure that the plumbing of the O3 generation and supply system is maintained and leak-free. 201 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.3.6. O3 GENERATOR The M200EH/EM uses a corona discharge (CD) tube for creating its O3. Corona discharge generation is capable of producing high concentrations of ozone efficiently and with low excess heat. Although there are many cell designs, the fundamental principle remains the same (Figure 10-10-11). Figure 10-10-11: Ozone Generator Principle The M200EH/EM utilizes a dual-dielectric design. This method utilizes a glass tube with hollow walls. The outermost and innermost surfaces are coated with electrically conductive material. The air flows through the glass tube, between the two conductive coatings, in effect creating a capacitor with the air and glass acting as the dielectric. The layers of glass also separate the conductive surfaces from the air stream to prevent reaction with the O3. As the capacitor charges and discharges, electrons are created and accelerated across the air gap and collide with the O2 molecules in the air stream splitting them into elemental oxygen. Some of these oxygen atoms recombine with O2 to O3. The quantity of ozone produced is dependent on factors such as the voltage and frequency of the alternating current applied to the CD cells. When enough high-energy electrons are produced to ionize the O2 molecules, a light emitting, gaseous plasma is formed, which is commonly referred to as a corona, hence the name corona discharge generator. 202 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.3.7. PERMA PURE® DRYER The air supplied to the O3 generation system needs to be as dry as possible. Normal room air contains a certain amount of water vapor, which greatly diminishes the yield of ozone produced by the ozone generator. Also, water can react with other chemicals inside the O3 Generator to produce chemicals that damage the optical filter located in the reaction cell (Table 10-1) such as ammonium sulfate or highly corrosive nitric acid. To accomplish this task the M200EH/EM uses a Perma Pure® single tube permeation dryer. The dryer consists of a single tube of Nafion® , a co-polymer similar to Teflon® that absorbs water very well but not other chemicals. The Nafion® tube is mounted within an outer, flexible plastic tube. As gas flows through the inner Nafion® tube, water vapor is absorbed into the membrane walls. The absorbed water is transported through the membrane wall and evaporates into the dry, purge gas flowing through the outer tube, countercurrent to the gas in the inner tube (Figure 10-10-12). Figure 10-10-12: Semi-Permeable Membrane Drying Process This process is called per-evaporation and is driven by the humidity gradient between the inner and outer tubes as well as the flow rates and pressure difference between inner and outer tubing. Unlike micro-porous membrane permeation, which transfers water through a relatively slow diffusion process, per-evaporation is a simple kinetic reaction. Therefore, the drying process occurs quickly, typically within milliseconds. The first step in this process is a chemical reaction between the molecules of the Nafion® material and water, other chemical components of the gases to be dried are usually unaffected. The chemical reaction is based on hydrogen bonds between the water molecule and the Nafion material. Other small polar gases that are capable of hydrogen bonds can be absorbed this way, too, such as ammonia (NH3) and some low molecular amines. The gases of interest, NO and NO2, do not get absorbed and pass the dryer unaltered. To provide a dry purge gas for the outer side of the Nafion tube, the M200EH/EM returns some of the dried air from the inner tube to the outer tube (Figure 10-10-13). When the analyzer is first started, the humidity gradient between the inner and outer tubes is not very large and the dryer’s efficiency is low at first but improves as this cycle reduces the moisture in the sample gas and settles at a minimum humidity. 203 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation Figure 10-10-13: M200EH/EM Perma Pure® Dryer Just like on startup, if the instrument is turned on after having been off for more than 30 minutes, it takes a certain amount of time for the humidity gradient to become large enough for the Perma Pure® Dryer to adequately dry the air. In this case, called a cold start, the O3 Generator is not turned on for 30 minutes. When rebooting the instrument within less than 30 minutes of power-down, the generator is turned on immediately. The Perma Pure® Dryer used in the M200EH/EM is capable of adequately drying ambient air to a dew point of ≤ -5˚C (~4000 ppm residual H2O) at a flow rate of 1 standard liter per minute (slpm) or down to ≤ -15˚C (~1600 ppm residual H2O) at 0.5 slpm. The Perma Pure® Dryer is also capable of removing ammonia from the sample gas up to concentrations of approximately 1 ppm. 204 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.3.8. OZONE SUPPLY AIR FILTER The M200EH/EM uses ambient air as the supply gas for the O3 generator and may produce a variety of byproducts. Small amounts of water, ammonia and various sulfur oxides can combine to create ammonium sulfate, ammonium nitrate, nitric acid and other compounds. Whereas sulfates and nitrates can create powdery residues inside the reaction cell causing sensitivity drift, nitric acid is a very aggressive compound, which can deteriorate the analyzer’s components. In order to remove these chemical byproducts from the O3 gas stream, the output of the O3 generator flows through a special filter between the generator and the reaction cell. Any NOX that may be produced in the generator (from reaction of O2 or O3 and N2 in the air) and may cause an artifact in the measurement, is calibrated out through the Auto-zero functionality, which checks the background signal of the O3 stream only once per minute. 10.3.9. OZONE SCRUBBER Even though ozone is unstable and typically reacts to form O2, the break-down is not quite fast enough to ensure that it is completely removed from the exhaust gas stream of the M200EH/EM by the time the gas exits the analyzer. Due to the high toxicity and reactivity of O3, a special catalytic ozone scrubber is used to remove all of the O3 exiting the reaction cell. Besides its efficient destruction of O3, this catalyst does not produce any toxic or hazardous gases as it only converts ozone to oxygen. The O3 scrubber is located inside the NO2 converter housing next to the NO2 converter in order to utilize residual heat given of by the converter heater. Even though the catalyst is 100% efficient at scrubbing ozone at room temperature, heating it significantly reduces the necessary residence time (the amount of time the gas must be in contact with the catalyst) for 100% efficiency and full efficiency can be maintained at higher gas flow rates. As this is a true catalytic converter, there are no maintenance requirements as would be required for charcoalbased scrubbers. A certain amount of fine, black dust may exit the catalyst, particularly if the analyzer is subjected to sudden pressure drops (for example, when disconnecting the running pump without letting the analyzer properly and slowly equilibrate to ambient pressure). To avoid the dust from entering the reaction cell or the pump, the scrubber is equipped with sintered stainless steel filters of 20 µm pore size on either end and on some models, an additional dust filter may be attached to the exhaust port. 205 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.3.10. PNEUMATIC SENSORS NOTE The M200EH/EM displays all pressures in inches of mercury absolute (in-Hg-A), i.e. absolute pressure referenced against zero (a perfect vacuum). The M200EH/EM uses three pneumatic sensors to verify gas streams. These sensors are located on a printed circuit assembly, called the pneumatic pressure/flow sensor board, located just behind the sensor assembly. 10.3.10.1. Vacuum Manifold The vacuum manifold is the central exit port for all analyzer pneumatics. All gas streams of the analyzer exit through this assembly and connect to the instrument’s pump. Figure 10-10-14 shows the standard configuration. Configurations will vary depending on the optional equipment that is installed. An IZS option, for example, will add another FT8 connector and orifice assembly to the manifold, an optional sample dryer may add a Tee-fitting so that two ¼” tubes can be connected to the same port. At this time, the vacuum manifold does not yet support the orifice holder shown in Figure 9-6. To exchange the critical orifice installed in the vacuum manifold, the user needs to either blow the orifice out with reversed pressure or remove the entire manifold for this task. However, orifices installed in the vacuum manifold should not have to be cleaned under normal circumstances. Figure 10-10-14: Vacuum Manifold 10.3.10.2. Sample Pressure Sensor An absolute pressure transducer connected to the input of the NO/NOX valve is used to measure the pressure of the sample gas before it enters the analyzer’s reaction cell. This is the “upstream” pressure mentioned above, which is used to compute sample flow rate. In conjunction with the vacuum pressure sensor, it is also used to validate the critical flow condition (2:1 pressure ratio) through the sample gas critical flow orifice (Section 10.3.3). If the temperature/pressure compensation (TPC) feature is turned on (Section 10.7.3), the output of this sensor is also used to supply pressure data for that calculation. The actual pressure value is viewable through the analyzer’s front panel display as the test function SAMP. The flow rate of the sample gas is displayed as SAMP FLW. 206 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.3.10.3. Vacuum Pressure Sensor An absolute pressure transducer connected to the exhaust manifold is used to measure the pressure downstream from and inside the instrument’s reaction cell. The output of the sensor is used by the CPU to calculate the pressure differential between the gas upstream of the reaction cell and the gas downstream from it and is also used as the main diagnostic for proper pump operation. If the ratio between the upstream pressure and the downstream pressure falls below 2:1, a warning message (SAMPLE FLOW WARN) is displayed on the analyzer’s front panel (Section6.2.2) and the sample flow rate will display XXXX instead of an actual value. If this pressure exceeds 10 in-Hg-A, an RCEL PRESSURE WARNING Is issued, even though the analyzer will continue to calculate a sample flow up to ~14 in Hg. Also, if the temperature/pressure compensation (TPC) feature is turned on (Section 10.7.3), the output of this sensor is used to supply pressure data for that calculation. This measurement is viewable through the analyzer’s front panel as the test function RCEL. 10.3.10.4. O3 Supply Air Flow Sensor A mass flow meter connected between the Perma Pure® dryer and the O3 generator measures the flow rate of O3 supply air through the analyzer. This information is used to validate the O3 gas flow rate. If the flow rate exceeds ±15% of the nominal flow rate (80 cm³/min), a warning message OZONE FLOW WARNING is displayed on the analyzer’s front panel (Section 6.2.2) and the O3 generator is turned off. As second warning, OZONE GEN OFF, is displayed. This flow measurement is viewable through instrument’s front panel display as the test function OZONE FL. 10.3.11. DILUTION MANIFOLD Certain applications require to measure NOX in sample gases that do not contain any oxygen. However, the molybdenum NO2 converter requires a minimum amount of oxygen to operate properly and to ensure constant conversion efficiency. For these special applications, the M200E analyzer may be equipped with a dilution manifold (Figure 10-10-15) to provide the instrument with an internal sample stream that contains about 2.5% O2. This manifold is mounted between converter housing and vacuum manifold on a small mounting bracket. If the dilution manifold is to be mounted in the M200EH/EM analyzer, it will fit on the back of the shown bracket as the front of the bracket is occupied by the bypass manifold. The manifold is equipped with two orifice holders that control the flow of the O2-free sample gas and the bleeds in a small amount of zero air before the combined sample stream goes to the NO/NOX valve for measurement. The zero air is produced by an external zero air scrubber cartridge, mounted on the rear panel (Figure 3-2 and )LJXUH). The dilution manifold is not temperature controlled, although the residual heat of the NO2 converter housing provides some temperature stability. Tight temperature stability is not critical to the dilution application. 207 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation Figure 10-10-15: Dilution Manifold Please inquire with Teledyne-API sales if the M200E can be modified to fit your application. 208 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.4. ELECTRONIC OPERATION Figure 10-10-16 shows a block diagram of the major electronic components of the M200EH/EM. Back Panel Connectors Analog Outputs A1 COM1 Optional 4-20 mA A2 COM2 Control Inputs: 1–6 A3 Optional Ethernet Interface Status Outputs: 1–8 A4 Analog Outputs (D/A) External Digital I/O) RS–232 ONLY RS–232 or RS–485 A/D Converter( V/F) Power-Up Circuit Box Temp MOTHER BOARD PC 104 CPU Card Disk On Chip CPU STATUS LED Flash Chip PC 104 Bus PMT TEMPERATURE PMT Temperature Sensor (Externally Powered) I2C Bus Pneumatic Sensor Board PMT OUTPUT (PMT DET) O2 OPTION TEMPERATURE OPTIC TEST CONTROL ELECTRIC TEST CONTROL REACTION CELL TEMPERATURE IZS OPTION PERMEATION TUBE TEMPERATURE PUMP Analog Sensor Inputs Internal Digital I/O HIGH VOLTAGE POWER SUPPLY LEVEL Thermistor Interface Sample Pressure Sensor Vacuum Pressure Sensor O3 Flow Sensor PMT I2C Status LED Keybd & Display RELAY BOARD TEMPERATURE SIGNAL NO/NOx Valve Reaction Cell Heater Autozero Valve MOLYBDENUM CONVERTER Molybdenum Converter Heater PREAMP PCA PMT PMT TEC TEC Drive PCA Figure 10-10-16: IZS Option Permeation Tube Heater O2 Sensor Option Sample Cal Valve Option Option IZS Valve Option MOLYBDENUM CONVERTER TEMPERATURE M200EH/EM Electronic Block Diagram The core of the analyzer is a microcomputer (CPU) that controls various internal processes, interprets data, calculates data, and reports results using specialized firmware developed by Teledyne Instruments. It communicates with the user, receives data from and issues commands to a variety of peripheral devices through the motherboard, the main printed circuit assembly on the rear panel (Figure 3-1). 209 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.4.1. CPU The CPU is a low power (5 VDC, 0.8A max), high performance, 386-based microcomputer running a version of the DOS operating system. Its operation and assembly conform to the PC-104 specification, version 2.3 for embedded PC and PC/AT applications. It has 2 MB of DRAM memory on board and operates at 40 MHz clock rate over an internal, 32-bit data and address bus. Chip to chip data handling is performed by two 4-channel, direct memory access (DMA) devices over data busses of either 8-bit or 16-bit bandwidth. The CPU supports both RS-232 and RS-485 serial protocols. Figure 10-10-17 shows the CPU board. The CPU communicates with the user and the outside world in a variety of ways: Through the analyzer’s keyboard and vacuum fluorescence display over a clocked, digital, serial I/O bus using the I2C protocol (read I-square-C bus) RS-232 and/or RS-485 serial ports (one of which can be connected to an Ethernet converter) Various analog voltage and current outputs Several digital I/O channels Figure 10-10-17: M200EH/EM CPU Board Annotated Finally, the CPU issues commands (also over the I2C bus) to a series of relays and switches located on a separate printed circuit assembly, the relay board (located in the right rear of the chassis on its own mounting bracket) to control the function of heaters and valves. The CPU includes two types of non-volatile data storage, one disk-on-chip and one or two flash chips. 210 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.4.1.1. Disk On Chip Technically, the disk-on-chip is an EEPROM, but appears to the CPU as, behaves as, and performs the same functions in the system as an 8 mb disk drive, internally labeled as DOS drive C:\. It is used to store the computer’s operating system files, the Teledyne Instruments firmware and peripheral files, and the operational data generated by the analyzer’s internal data acquisition system (iDAS - Sections 10.7.5 and 6.7). 10.4.1.2. Flash Chip The flash chip is another, smaller EEPROM with about 64 kb of space, internally labeled as DOS drive B:\. The M200EH/EM CPU board can accommodate up to two EEPROM flash chips. The M200EH/EM standard configuration is one chip with 64 kb of storage capacity, which is used to store the analyzer configuration as created during final checkout at the factory. Separating these data onto a less frequently accessed chip significantly decreases the chance of data corruption through drive failure. In the unlikely event that the flash chip should fail, the analyzer will continue to operate with just the DOC. However, all configuration information will be lost, requiring the unit to be recalibrated. 10.4.2. SENSOR MODULE, REACTION CELL Electronically, the M200EH/EM sensor assembly (see Figure 9-6) consists of several subassemblies with different tasks: to detect the intensity of the light from the chemiluminescence reaction between NO and O3 in the reaction cell, to produce a current signal proportional to the intensity of the chemiluminescence, to control the temperature of the PMT to ensure the accuracy and stability of the measurements and to drive the high voltage power supply that is needed for the PMT. The individual functions are described individually below, Section 11.6.6 shows the sensor assembly and its components. 10.4.2.1. Reaction Cell Heating Circuit The stability of the chemiluminescence reaction between NO and O3 can be affected by changes in the temperature and pressure of the O3 and sample gases in the reaction cell. In order to reduce temperature effects, the reaction cell is maintained at a constant 50 C, just above the high end of the instrument’s operation temperature range. Two AC heaters, one embedded into the bottom of the reaction cell, the other embedded directly above the chamber’s exhaust fitting, provide the heat source. These heaters operate off of the instrument’s main AC power and are controlled by the CPU through a power relay on the relay board (Section 10.4.7). A thermistor, also embedded in the bottom of the reaction cell, reports the cell’s temperature to the CPU through the thermistor interface circuitry of the motherboard (Section 0). 211 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.4.3. PHOTO MULTIPLIER TUBE (PMT) The M200EH/EM uses a photo multiplier tube (PMT) to detect the amount of chemiluminescence created in the sample chamber. PMT Housing End Plate This is the entry to the PMT Exchange PMT Output Connector PMT Preamp PCA PMT Power Supply & Aux. Signal Connector High voltage Power Supply (HVPS) PMT O-Test LED PMT Cold Block Connector to PMT Pre Amp PCA 12V Power Connector Insulation Gasket PMT Temperature Sensor Light from Reaction Chamber shines through hole in side of Cold Block Thermo-Electric Cooler (TEC) PMT Heat Exchange Fins TEC Driver PCA Cooling Fan Housing Figure 10-10-18: PMT Housing Assembly A typical PMT is a vacuum tube containing a variety of specially designed electrodes. Photons from the reaction are filtered by an optical high-pass filter, enter the PMT and strike a negatively charged photo cathode causing it to emit electrons. A high voltage potential across these focusing electrodes directs the electrons toward an array of high voltage dynodes. The dynodes in this electron multiplier array are designed so that each stage multiplies the number of emitted electrons by emitting multiple, new electrons. The greatly increased number of electrons emitted from one end of electron multiplier are collected by a positively charged anode at the other end, which creates a useable current signal. This current signal is amplified by the preamplifier board and then reported to the motherboard. Figure 10-10-19: Basic PMT Design 212 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation A significant performance characteristic of the PMT is the voltage potential across the electron multiplier. The higher the voltage, the greater is the number of electrons emitted from each dynode of the electron multiplier, making the PMT more sensitive and responsive to small variations in light intensity but also more noisy (dark noise). The gain voltage of the PMT used in the M200EH/EM is usually set between 450 V and 800 V. This parameter is viewable through the front panel as test function HVPS (see Section 6.2.1). For information on when and how to set this voltage, see Section 11.6.3.8. The PMT is housed inside the PMT module assembly (see Figure 10-18). This assembly also includes the high voltage power supply required to drive the PMT, an LED used by the instrument’s optical test function, a thermistor that measures the temperature of the PMT and various components of the PMT cooling system including the thermo-electric cooler (TEC). 10.4.4. PMT COOLING SYSTEM. The performance of the analyzer’s PMT is significantly affected by temperature. Variations in PMT temperature are directly reflected in the signal output of the PMT. Also the signal to noise ratio of the PMT output is radically influenced by temperature as well. The warmer The PMT is, the noisier its signal becomes until the noise renders the concentration signal useless. To alleviate this problem a special cooling system exists that maintains the PMT temperature at a stable, low level TEC PCA sets appropriate drive voltage for cooler Preamp PCA sends buffered and amplified thermistor signal to TEC PCA TEC Control PCA PMT Preamp PCA Heat Sink ThermoElectric Cooler Thermistor outputs temp of cold block to preamp PCA PMT Cold Block Heat form PMT is absorbed by the cold block and transferred to the heat sink via the TEC then bled off into the cool air stream. Cooling Fan Figure 10-10-20: PMT Cooling System 10.4.4.1. TEC Control Board The TEC control printed circuit assembly is located ion the sensor housing assembly, under the slanted shroud, next to the cooling fins and directly above the cooling fan. Using the amplified PMT temperature signal from the PMT preamplifier board (see Section 10.4.5), it sets the drive voltage for the thermoelectric cooler. The warmer the PMT gets, the more current is passed through the TEC causing it to pump more heat to the heat sink. A red LED located on the top edge of this circuit board indicates that the control circuit is receiving power. Four test points are also located at the top of this assembly. For the definitions and acceptable signal levels of these test points see Chapter 11. 213 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.4.5. PMT PREAMPLIFIER The PMT preamplifier board amplifies the PMT signal into a useable analog voltage (PMT) that can be processed by the motherboard into a digital signal to be used by the CPU to calculate the NO, NO2 and NOx concentrations of the gas in the sample chamber. The output signal of the PMT is controlled by two different adjustments. First, the voltage across the electron multiplier array of the PMT is adjusted with a set of two hexadecimal switches. Adjusting this voltage directly affects the HVPS voltage and, hence, the signal from the PMT. Secondly, the gain of the amplified signal can further be adjusted through a potentiometer. These adjustments should only be performed when encountering problems with the software calibration that cannot be rectified otherwise. See Section 11.6.3.8 for this hardware calibration. O Test Control From CPU PMT Fine Gain Set PMT Coarse Gain Set To (Rotary Switch) (Rotary O Test LED Motherboard PMT Preamp PCA O-Test Generator PMT HVPS Drive Voltage D-A Converter PMT Output E Test Control From CPU MUX Amp to Voltage Converter/ Amplifier E-Test Generator PMT Temp Analog Signal TEC Control PCA PMT Signal Offset to Motherboard PMT Temp Sensor Low Pass Noise Filter PMT Temperature Feedback Circuit PMT Output Signal (PMT) to Motherboard Figure 10-10-21: PMT Preamp Block Diagram The PMT temperature control loop maintains the PMT temperature around 7° C and can be viewed as test function PMT TEMP on the front panel (see Section 6.2.1). The electrical test (ETEST) circuit generates a constant, electronic signal intended to simulate the output of the PMT (after conversion from current to voltage). By bypassing the detector’s actual signal, it is possible to test most of the signal handling and conditioning circuitry on the PMT preamplifier board. See section 6.9.6 for instructions on performing this test. The optical test (OTEST) feature causes an LED inside the PMT cold block to create a light signal that can be measured with the PMT. If zero air is supplied to the analyzer, the entire measurement capability of the sensor module can be tested including the PMT and the current to voltage conversion circuit on the PMT preamplifier board. See section 6.9.5 for instructions on performing this test. 214 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.4.6. PNEUMATIC SENSOR BOARD The flow and pressure sensors of the M200EH/EM are located on a printed circuit assembly just behind the PMT sensor. Refer to Section 11.5.15 for information on how to test this assembly. The signals of this board are supplied to the motherboard for further signal processing. All sensors are linearized in the firmware and can be span calibrated from the front panel. 10.4.7. RELAY BOARD The relay board is the central switching and power distribution unit of the analyzer. It contains power relays, valve drivers and status LEDs for all heated zones and valves, as well as thermocouple amplifiers, power distribution connectors and the two switching power supplies of the analyzer. The relay board communicates with the motherboard over the I2C bus and can be used for detailed trouble-shooting of power problems and valve or heater functionality. See Figure 11-4 for an annotated view of the relay board. 10.4.7.1. Relay PCA Location and Layout Generally the relay PCA is located in the right-rear quadrant of the analyzer and is mounted vertically on the back side of the same bracket as the instrument’s DC power supplies, however the exact location of the relay PCA may differ from model to model (see Figure 3-1) 10.4.7.2. Heater Control The heater control loop is illustrated in Figure 10-22. Two T/C inputs can be configured for either type-T or typeK thermocouples. Additionally: Both T/C’s can be configured as either grounded or ungrounded thermocouples. Standard configuration of the both type of thermocouples is 10 mV/°C. In order to accommodate the M200EH’s Mini High-Con converter option, a type-K; 5mV/°C output configuration has been added. Thermistor(s) – Low Temperature Sensing: (e.g. Sample Chamber and Reaction Cell temperatures) MOTHER BOARD A/D Converter (V/F) RELAY PCA Preamplifiers and Signal Conditioning THERMOCOUPLE CONFIGURATION JUMPER (JP5) Themocouple(s) (High Temperature Sensing; e.g. Moly and HiCon Converter temperatures) CPU Cold Junction Compensation DC Control Logic Solid State AC Relays DC HEATERS Figure 10-10-22: AC HEATERS Heater Control Loop Block Diagram. 215 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.4.7.3. Thermocouple Inputs and Configuration Jumper (JP5) Although the relay PCA supports two thermocouple inputs, the current M200EH/EM series analyzers only utilize one. By default, this single thermocouple input is plugged into the TC1 input (J15). TC2 (J16) is currently not used. See Figure 11-4 for location of J15 and J16 Table 10-4: Thermocouple Configuration Jumper (JP5) Pin-Outs TC INPUT JUMPER PAIR DESCRIPTION 1 – 11 FUNCTION Gain Selector Selects preamp gain factor for J or K TC - IN = J TC gain factor Output Scale Selector Selects preamp gain factor for J or K TC - IN = 5 mV / °C - OUT = K TC gain factor 2 – 12 - OUT = 10 mV / °C TC1 3 – 13 Type J Compensation When present, sets Cold Junction Compensation for J type Thermocouple 4 – 14 Type K Compensation When present, sets Cold Junction Compensation for K type Thermocouple Selects between Isolated and grounded TC - IN = Isolate TC Termination Selector 5 – 15 - OUT = Grounded TC Gain Selector Same as Pins 1 – 11 above. 7 – 17 Output Scale Selector Same as Pins 2 – 12 above. 8 – 18 Type J Compensation Same as Pins 3 – 13 above. 9 – 19 Type K Compensation Same as Pins 4 – 14 above. 10 – 20 Termination Selector Same as Pins 5 – 15 above. Figure 10-10-23: Termination Selector 10 – 20 Type J Compensation 9 – 19 Output Scale Selector 7 – 17 Input Gain Selector 6 – 16 Termination Selector 5 – 15 TC2 Type J Compensation 4 – 14 Type J Compensation 3 – 13 Output Scale Selector 2 – 12 Input Gain Selector 1 – 11 TC1 Type J Compensation 8 – 18 TC2 6 – 16 Thermocouple Configuration Jumper (JP5) Pin-Outs 216 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation Table 10-5: Typical Thermocouple Settings For M200E Series Analyzers TC TYPE TERMINATION TYPE OUTPUT SCALE TYPE JUMPER BETWEEN PINS USED ON JUMPER COLOR INPUT TC1 (J15) K GROUNDED 5mV / °C 2 – 12 4 – 14 M200EH/EM with Mini HiCon Converter BROWN K ISOLATED 5mV / °C 2 – 12 4 – 14 5 – 15 M200EH/EM with Mini HiCon Converter GREY K ISOLATED 10mV / °C 4 – 14 5 – 15 M200EH/EM models with Moly Converter PURPLE J ISOLATED 10mV / °C 1 – 11 3 – 13 5 – 15 M200EH/EM models with Moly Converter RED J GROUNDED 10mV / °C 1 – 11 3 – 13 M200EH/EM models with Moly Converter GREEN 10.4.7.4. Valve Control The relay board also hosts two valve driver chips, each of which can drive up four valves. The main valve assembly in the M200EH/EM is the NO/NOX - Auto-zero solenoid valve component mounted right in front of the NO2 converter housing. These two valves are actuated with 12 V supplied from the relay board and driven by the CPU through the I2Z bus. A second set of valves may be installed if the zero/span valve or the IZS option is enabled in the analyzer. Specialty manifold valves may be present in the analyzer. 217 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.4.8. STATUS LEDS & WATCH DOG CIRCUITRY Thirteen LEDs are located on the analyzer’s relay board to indicate the status of the analyzer’s heating zones and valves as well as a general operating watchdog indicator. Table 11-2 shows the states of these LEDs and their respective functionality. D7 (Green) – Zero / Span Valv D8 (Green) – Sample / Cal D4 (Yellow) – Manifold Heater D3 (Yellow) – NO2 Converter Heater D9 (Green ) – Auto / Z D2 (Yellow) – Reaction Cell Heater D10 (Green) – NOx D5(Yellow) D6 (Yellow) – O2 Sensor Heater D1 (RED) Watchdog Indicator Figure 10-10-24: Status LED Locations – Relay PCA 10.4.8.1. Watchdog Indicator (D1) The most important of the status LED’s on the relay board is the red I1C Bus watch-dog LED. It is controlled directly analyzer’s CPU over the I2C bus. Special circuitry on the relay PCA watches the status of D1. Should this LED ever stay ON or OFF for 30 seconds, indicating that the CPU or I2C bus has stopped functioning, this Watchdog Circuit automatically shuts all valves and turn off all heaters and lamps. 218 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.4.9. MOTHERBOARD This is the largest electronic assembly in the analyzer and is mounted to the rear panel as the base for the CPU board and all I/O connectors. This printed circuit assembly provides a multitude of functions including A/D conversion, digital input/output, PC-104 to I2C translation, temperature sensor signal processing and is a pass through for the RS-232 and RS-485 signals. 10.4.9.1. A to D Conversion Analog signals, such as the voltages received from the analyzer’s various sensors, are converted into digital signals that the CPU can understand and manipulate by the analog to digital converter (A/D).Under the control of the CPU, this functional block selects a particular signal input and then coverts the selected voltage into a digital word. The A/D consists of a voltage-to-frequency (V-F) converter, a programmable logic device (PLD), three multiplexers, several amplifiers and some other associated devices. The V-F converter produces a frequency proportional to its input voltage. The PLD counts the output of the V-F during a specified time period, and sends the result of that count, in the form of a binary number, to the CPU. The A/D can be configured for several different input modes and ranges but in the is used in uni-polar mode with a +5V full scale. The converter includes a 1% over and under-range. This allows signals from -0.05V to +5.05V to be fully converted. For calibration purposes, two reference voltages are supplied to the A/D converter: Reference ground and +4.096 VDC. During calibration, the device measures these two voltages, outputs their digital equivalent to the CPU. The CPU uses these values to compute the converter’s offset and slope and uses these factors for subsequent conversions. See Section 6.13.5.4 for instructions on performing this calibration. 10.4.9.2. Sensor Inputs The key analog sensor signals are coupled to the A/D converter through the master multiplexer from two connectors on the motherboard. Terminating resistors (100 kΩ ) on each of the inputs prevent cross-talk between the sensor signals. PMT DETECTOR OUTPUT: This signal, output by the PMT preamp PCA, is used in the computation of the NO, NO2 and NOx concentrations displayed at the top right hand corner of the front panel display and output through the instruments analog outputs and COMM ports. PMT HIGH VOLTAGE POWER SUPPLY LEVEL: This input is based on the drive voltage output by the PMT pram board to the PMT’s high voltage power supply (HVPS). It is digitized and sent to the CPU where it is used to calculate the voltage setting of the HVPS and stored in the instruments memory as the test function HVPS. HVPS is viewable as a test function (see Section 6.2.1) through the analyzer’s front panel. PMT TEMPERATURE: This signal is the output of the thermistor attached to the PMT cold block amplified by the PMT temperature feedback circuit on the PMT preamp board. It is digitized and sent to the CPU where it is used to calculate the current temperature of the PMT. This measurement is stored in the analyzer. Memory as the test function PMT TEMP and is viewable as a test function (see Section 6.2.1) through the analyzer’s front panel. 219 04521C (DCN5731) Theory of Operation Teledyne API - Model 200EH/EM Operation Manual NO2 CONVERTER TEMPERATURE: This parameter is measured with a Type-K thermocouple attached to the NO2 converter heater and its analog signal is amplified by the circuitry on the relay board. It is sent to the CPU and then digitized and is used to calculate the current temperature of the NO2 converter. It is also stored in the iDAS and reported as test function MOLY TEMP. SAMPLE GAS PRESSURE: This is measured upstream of the reaction cell, stored in the iDAS and reported as SAMPLE. The vacuum gas pressure is measured downstream of the reaction cell and is stored in the iDAS and reported as RCEL. For more information on these sensor’s functions see Section 10.3.10. O3 GAS FLOW This sensor measures the gas flow upstream of the ozone generator, stored in the iDAS and reported as test function OZONE FL. For more information on this sensor’s function see Section 10.3.10. 10.4.9.3. Thermistor Interface This circuit provides excitation, termination and signal selection for several negative-coefficient, thermistor temperature sensors located inside the analyzer. They are: REACTION CELL TEMPERATURE SENSOR A thermistor embedded in the reaction cell manifold. This temperature is used by the CPU to control the reaction cell heating circuit and as a parameter in the temperature/pressure compensation algorithm. This measurement is stored in the analyzer’s iDAS and reported as test function RCEL TEMP. BOX TEMPERATURE SENSOR: A thermistor is attached to the motherboard. It measures the analyzer’s inside temperature. This information is stored by the CPU and can be viewed by the user for troubleshooting purposes through the front panel display. It is also used as part of the NO, NOX and NO2 calculations when the instrument’s Temperature/Pressure Compensation feature is enabled. This measurement is stored in the analyzer. Memory as the test function BOX TEMP and is viewable as a test function (Section 6.2.1) through the analyzer’s front panel. The thermistor inside the PMT cold block as well as the thermistor located on the preamplifier board are both converted to analog signals on the preamplifier board before being sent to the motherboard’s A/D converter. O2 SENSOR TEMPERATURE: For instruments with the oxygen sensor option installed, the thermistor measuring the temperature of the heating block mounted to the sensor is reported as test function O2 TEMP on the front panel. This temperature is maintained at 50° C. 10.4.10. ANALOG OUTPUTS The analyzer comes equipped with four Analog Outputs: A1, A2, A3 and a fourth that is a spare. A1 and A2 Outputs: The first two, A1 and A2 are normally set up to operate in parallel so that the same data can be sent to two different recording devices. While the names imply that one should be used for sending data to a chart recorder and the other for interfacing with a datalogger, either can be used for both applications. Output Loop-back: All of the functioning analog outputs are connected back to the A/D converter through a Loop-back circuit. This permits the voltage outputs to be calibrated by the CPU without need for any additional tools or fixtures (see Section 6.13.5.4) 220 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.4.11. EXTERNAL DIGITAL I/O The external digital I/O performs two functions. The STATUS outputs carry logic-level (5V) signals through an optically isolated 8-pin connector on the rear panel of the analyzer. These outputs convey on/off information about certain analyzer conditions such as CONC VALID. They can be used to interface with certain types of programmable devices (Section 6.15.1.1). The CONTROL inputs can be initiated by applying 5V DC power from an external source such as a PLC or data logger (Section 6.15.1.2). Zero and span calibrations can be initiated by contact closures on the rear panel. 10.4.12. I2C DATA BUS I2C is a two-wire, clocked, digital serial I/O bus that is used widely in commercial and consumer electronic systems. A transceiver on the motherboard converts data and control signals from the PC-104 bus to I2C. The data are then fed to the keyboard/display interface and finally onto the relay board. Interface circuits on the keyboard/display interface and relay board convert the I2C data to parallel inputs and outputs. An additional interrupt line from the keyboard to the motherboard allows the CPU to recognize and service key strokes on the keyboard. 10.4.13. POWER-UP CIRCUIT This circuit monitors the +5V power supply during analyzer start-up and sets the analog outputs, external digital I/O ports, and I2C circuitry to specific values until the CPU boots and the instrument software can establish control. 221 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.5. POWER DISTRIBUTION &CIRCUIT BREAKER The analyzer operates in two main AC power ranges: 100-120 VAC and 220-240 VAC (both ± 10%) between 47 and 63 Hz. A 5 ampere circuit breaker is built into the ON/OFF switch. In case of a wiring fault or incorrect supply power, the circuit breaker will automatically turn off the analyzer. CAUTION Should the power circuit breaker trip correct the condition causing this situation before turning the analyzer back on. SENSOR SUITES ANALOG SENSORS (e.g. UV sensors, Temp Sensors, Flow Sensors, PMT HVPS, etc.) KEY Sensor Control & I/O Logic Pre-Amplifiers & Amplifiers AC POWER LOGIC DEVICES DC POWER (e.g. CPU, I2 C bus, Keyboard, Display, MotherBoard, etc.) PS 1 +5 VDC PUMP AC HEATERS AC HEATERS for O2 SENSOR UV Lamp P/S ±15 VDC Configuration Jumpers ON / OFF SWITCH Configuration Jumpers Configuration Jumpers PS 2 (+12 VDC) RELAY PCA Solenoid Drivers AC POWER IN MODEL SPECIFIC VALVES (e.g. NO X – NO Valves, Auto-zero valves, etc.) Figure 10-10-25: OPTIONAL VALVES (e.g. Sample/Cal, Zero/Spans, etc.) TEC and Cooling Fan(s) Power Distribution Block Diagram Under normal operation, the M200EH/EM draws about 1.5 A at 115 V and 2.0 A during start-up. 222 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.6. COMMUNICATIONS INTERFACE The analyzer has several ways to communicate the outside world, see Figure 10-26. Users can input data and receive information directly through the front panel keypad and display. Direct, two-way communication with the CPU is also available by way of the analyzer’s RS232 & RS485 I/O ports (see Section 6.11 and 6.15). Alternatively, an Ethernet communication option can be substituted for one of the COMM ports. The analyzer can also send status information and data via the eight digital status output lines (see Section 6.15.1.1) and the three analog outputs (see Section 6.7) located on the rear panel as well as receive commands by way of the six digital control inputs also located on the rear pane (see Section 6.15.1.2). Figure 10-10-26: Interface Block Diagram 223 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.6.1. FRONT PANEL INTERFACE MODE FIELD MESSAGE FIELD CONCENTRATION FIELD FASTENER FASTENER KEY DEFINITIONS SAMPLE A1:NXCNC1=100PPM < TST TST > CAL NOX=XXX.X SETUP SAMPLE CAL FAULT STATUS LED’s KEYBOARD POWER ON / OFF SWITCH ? ? ?? ?? ??? ? ?? ?? ?? ?? ?? ?? ?? ?? ?? ?? ??? ? ?? CHEMILUMINESENCE NOx ANALYZER – M200EH HINGE Figure 10-10-27: M200EH/EM Front Panel Layout The most commonly used method for communicating with the M200EH/EM UV Chemiluminescence NOx Analyzer is via the instrument’s front panel which includes a set of three status LEDs, a vacuum florescent display and a keyboard with 8 context sensitive keys. 10.6.1.1. Analyzer Status LED’s Three LEDS are used to inform the user of the instruments basic operating status Table 10-6: Front Panel Status LED’s NAME SAMPLE COLOR Green STATE Unit is not operating in sample mode, iDAS is disabled. On Sample Mode active; Front Panel Display being updated, iDAS data being stored. Blinking CAL Yellow Red Unit is operating in sample mode, front panel display being updated, iDAS hold-off mode is ON, iDAS disabled Off Auto Cal disabled On Auto Cal enabled Blinking FAULT DEFINITION Off Off Blinking Unit is in calibration mode No warnings exist Warnings exist 10.6.1.2. Keyboard A row of eight keys just below the vacuum florescent display (see Figure 10-27) is the main method by which the user interacts with the analyzer. As the software is operated, labels appear on the bottom row of the display directly above each active key, defining the function of that key as it is relevant for the operation being performed. Pressing a key causes the associated instruction to be performed by the analyzer. 224 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation Note that the keys do not auto-repeat. In circumstances where the same key must be activated for two consecutive operations, it must be released and re-pressed. 10.6.1.3. Display The main display of the analyzer is a vacuum florescent display with two lines of 40 text characters each. Information is organized in the following manner (see Figure 10-27): MODE FIELD: Displays the name of the analyzer’s current operating mode. MESSAGE FIELD: Displays a variety of informational messages such as warning messages, operation data and response messages during interactive tasks. CONCENTRATION FIELD: Displays the actual concentration of the sample gas currently being measured by the analyzer KEYPAD DEFINITION FIELD: Displays the definitions for the row of keys just below the display. These definitions dynamic, context sensitive and software driven. I2C to/from CPU I2C Interface Serial Data Display Controller Display Power Watchdog Clock Display Data Decoder Display Write Keypad Decoder 2 I C to Relay Board Parallel Data Key Press Detect Keyboard Interrupt Status Bit 10.6.1.4. Keyboard/Display Interface Electronics From 5 VDC Power Supply Sample LED (Green) Cal LED (Yellow) KEYBOARD Maint. Switch 2nd Lang. Switch 2 x 40 CHAR. VACUUM FLUORESCENT DISPLAY Fault LED (Red) Beeper Figure 10-10-28: Optional Maintenance LED FRONT PANEL Keyboard and Display Interface Block Diagram The keyboard/display interface electronics of the M200EH/EM Analyzer watches the status of the eight front panel keys, alerts the CPU when keys are depressed, translates data from parallel to serial and back and manages communications between the keyboard, the CPU and the front panel display. Except for the Keyboard interrupt status bit, all communication between the CPU and the keyboard/display is handle by way of the instrument’s I2C buss. The CPU controls the clock signal and determines when the various devices on the bus are allowed to talk or required to listen. Data packets are labeled with addresses that identify for which device the information is intended. 225 04521C (DCN5731) Theory of Operation Teledyne API - Model 200EH/EM Operation Manual KEYPAD DECODER Each key on the front panel communicates with a decoder IC via a separate analog line. When a key is depressed the decoder chip notices the change of state of the associated signal; latches and holds the state of all eight lines (in effect creating an 8-bit data word); alerts the key-depress-detect circuit (a flip-flop IC); translates the 8-bit word into serial data and; sends this to the I2C interface chip. KEY-DEPRESS-DETECT CIRCUIT This circuit flips the state of one of the inputs to the I2C interface chip causing it to send an interrupt signal to the CPU I2C INTERFACE CHIP This IC performs several functions: Using a dedicated digital status bit, it sends an interrupt signal alerting the CPU that new data from the keyboard is ready to send. Upon acknowledgement by the CPU that it has received the new keyboard data, the I2C interface chip resets the key-depress-detect flip-flop. In response to commands from the CPU, it turns the front panel status LEDs on and off and activates the beeper. Informs the CPU when the optional maintenance and second language switches have been opened or closed (see Chapter 5 for information on these options). DISPLAY DATA DECODER This decoder translates the serial data sent by the CPU (in TTY format) into a bitmapped image which is sent over a parallel data bus to the display. DISPLAY CONTROLLER This circuit manages the interactions between the display data decoder and the display itself. It generates a clock pulse that keeps the two devices synchronized. It can also, in response to commands from the CPU turn off and/or reset the display. Additionally, for analyzers with the optional maintenance switch is installed (See Chapter 5), the display controller turns on an LED located on the back of the keyboard interface PCA whenever the instrument is placed in maintenance mode. DISPLAY POWER WATCHDOG The Model 200EH/EM’s display can begin to show garbled information or lock-up if the DC voltage supplied to it falls too low, even momentarily. To alleviate this, a brown-out watchdog circuit monitors the level of the power supply and in the event that the voltage level falls below a certain level resets the display by turning it off, then back on. I2C LINK TO THE RELAY PCA While the CPU’s I2C communication with the relay board is also routed through the keyboard/display interface, information passed to and from the relay board via this channel is not recognized by, acted upon or affected by the circuitry of the keyboard. 226 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.7. SOFTWARE OPERATION The M200EH/EM NOX analyzer’s core module is a high performance, 386-based microcomputer running a version of DOS. On top of the DOS shell, special software developed by Teledyne Instruments interprets user commands from various interfaces, performs procedures and tasks, stores data in the CPU’s memory devices and calculates the concentrations of NOX in the sample gas. Figure 10-10-29 shows a block diagram of this software functionality. Figure 10-10-29: Schematic of Basic Software Operation 227 04521C (DCN5731) Theory of Operation Teledyne API - Model 200EH/EM Operation Manual 10.7.1. ADAPTIVE FILTER The M200EH/EM NOX analyzer software processes sample gas concentration data through a built-in adaptive filter. Unlike other analyzers that average the output signal over a fixed time period, the M200EH/EM averages over a defined number of samples, with samples being about 8 seconds apart (reflecting the switching time of 4 s each for NO and NOX). This technique is known as boxcar filtering. During operation, the software may automatically switch between two different filters lengths based on the conditions at hand. During constant or nearly constant concentrations, the software, by default, computes an average of the last 42 samples, or approximately 5.6 minutes. This provides smooth and stable readings and averages out a considerable amount of random noise for an overall less noisy concentration reading. If the filter detects rapid changes in concentration the filter reduces the averaging to only 6 samples or about 48 seconds to allow the analyzer to respond more quickly. Two conditions must be simultaneously met to switch to the short filter. First, the instantaneous concentration must differ from the average in the long filter by at least 50 ppb. Second, the instantaneous concentration must differ from the average in the long filter by at least 10% of the average in the long filter. If necessary, these boxcar filter lengths can be changed between 1 (no averaging) and 1000 samples but with corresponding tradeoffs in rise time and signal-to-noise ratio. Signal noise increases accordingly when in adaptive filter mode, but remains within the official M200EH/EM specifications as long as the filter size remains at or above 3 samples. In order to avoid frequent switching between the two filter sizes, the analyzer has a delay of 120 s before switching out of adaptive filter mode, even if the two threshold conditions are no longer met. Note that the filter settings in NOX only or NO only 10.7.2. CALIBRATION - SLOPE AND OFFSET Aside from the hardware calibration of the preamplifier board (Section 11.6.5) upon factory checkout, calibration of the analyzer is usually performed in software. During instrument calibration (Chapters 7) the user enters expected values for span gas concentration through the front panel keypad and supplies the instrument with sample gas of know NO and NOX concentrations. The readings are then compared to the expected values and the software computes values for the new instrument slope and offset for both NO and NOX response. These values are stored in memory for use in calculating the NO, NOX and NO2 concentration of the sample gas. By default, the iDAS stores 200 software calibration settings for documentation, review and data analysis. Instrument slope and offset values recorded during the last calibration can be viewed on the front panel. NO SLOPE, NOX SLOPE, NO OFFS and NOX OFFS are four of the test parameters accessible through the buttons. 228 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Theory of Operation 10.7.3. TEMPERATURE/PRESSURE COMPENSATION (TPC) The software features a compensation of some temperature and pressure changes critical in the measurement of NO and NOX concentration. When the TPC feature is enabled (default setting), the analyzer divides the value of the PMT output signal (PMTDET) by a value called TP_FACTOR. TP_FACTOR is calculated according to the following equation. TP _ FACTOR= A RCELLTEMP(K) 7 (in Hg) SAMP(in Hg BOXTEMP(K) ×B ×C ×D 323(K) RCEL(in Hg) 29.92(in Hg) 298(K) (Equation 10-5) Where A, B, C, D are gain functions. The four parameters used to compute TP_FACTOR are: RCELL TEMP: The temperature of the reaction cell, measured in K. RCEL: The pressure of the gas in the vacuum manifold, measured in in-Hg-A. SAMP: The pressure of the sample gas before it reaches the reaction cell, measured in in-Hg-A. This measurement is ~1 in-Hg-A lower than atmospheric pressure. BOX TEMP: The temperature inside the analyzer’s case measured in K. This is typically about 5 K higher than room temperature. The current value of all four of these measurements are viewable as TEST FUNCTIONS through the instrument’s front panel display. Note that, as RCEL TEMP, BOX TEMP and SAMP pressure increase, the value of TP_FACTOR increases and, hence, the PMTDET value decreases. Conversely, increases in the reaction cell pressure (RCEL) decrease TP_FACTOR and, hence increase the PMTDET value. These adjustments are meant to counter-act changes in the concentrations caused by these parameters. Each of the terms in the above equation is attenuated by a gain function with a numerical value based on a preset gain parameter (shown below in CAPITALIZED ITALICS) normalized to the current value of the parameter being attenuated. The gain functions A, B, C and D are defined as: A = 1+ [( rcell _ temp(K ) 1) × RCTEMP _ TPC _ GAIN ] 323(K ) (Equation 10-6) 5(" Hg ) 1) × RCPRESS _ TPC _ GAIN ] B = 1+ [( rcell _ pressure(" Hg ) (Equation 10-7) rcell _ temp(K ) 1) × SPRESS _ TPC _ GAIN ] C = 1+ [( 323(K ) (Equation 10-8) D = 1+ [( box _ temp(K ) 1) × BXTEMP _ TPC _ GAIN ] 298(K ) (Equation 10-9) The preset gain parameters are set at the factory and may vary from analyzer to analyzer. Section 6.12 describes the method for enabling/disabling the TPC feature. 229 04521C (DCN5731) Theory of Operation Teledyne API - Model 200EH/EM Operation Manual 10.7.4. NO2 CONVERTER EFFICIENCY COMPENSATION Over time, the molybdenum in the NO2 converter oxidizes and looses its original capacity of converting NO2 into NO, eventually resulting in a decreased converter efficiency (CE). Even though we recommend to replace the converter if CE drops below 96%, the analyzer’s firmware allows adjusting minor deviations of the CE from 1.000 and enables reporting the true concentrations of NO2 and NOX. Converter efficiency is stored in the instrument’s memory as a decimal fraction that is multiplied with the NO2 and NOX measurements to calculate the final concentrations for each. Periodically, this efficiency factor must be measured and - if it has changed from previous measurements - entered into the analyzer’s memory (Section 7.1.5). 10.7.5. INTERNAL DATA ACQUISITION SYSTEM (IDAS) The iDAS is designed to implement predictive diagnostics that stores trending data for users to anticipate when an instrument will require service. Large amounts of data can be stored in non-volatile memory and retrieved in plain text format for further processing with common data analysis programs. The iDAS has a consistent user interface among all Teledyne Instruments A- and E-series instruments. New data parameters and triggering events can be added to the instrument as needed. Section 6.7 describes the iDAS and its default configuration in detail, Chapter 8 shows the parameters that can be used for predictive diagnostics. Depending on the sampling frequency and the number of data parameters, the iDAS can store several months of data, which are retained even when the instrument is powered off. However, if new firmware or a new iDAS configuration are uploaded to the analyzer, we recommend retrieving data before doing so to avoid data loss. The iDAS permits users to access the data through the instrument’s front panel or the remote interface. The latter can automatically report stored data for further processing. APICOM, a user-friendly remote control program is the most convenient way to view, retrieve and store iDAS data (Section 6.15.2.8) USER NOTES: 230 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11. TROUBLESHOOTING & REPAIR This section contains a variety of methods for identifying and solving performance problems with the analyzer. NOTE The operations outlined in this chapter must be performed by qualified maintenance personnel only. CAUTION Risk of electrical shock. Some operations need to be carried out with the analyzer open and running. Exercise caution to avoid electrical shocks and electrostatic or mechanical damage to the analyzer. Do not drop tools into the analyzer or leave those after your procedures. Do not shorten or touch electric connections with metallic tools while operating inside the analyzer. Use common sense when operating inside a running analyzer. 11.1. GENERAL TROUBLESHOOTING The analyzer has been designed so that problems can be rapidly detected, evaluated and repaired. During operation, the analyzer continuously performs diagnostic tests and provides the ability to evaluate its key operating parameters without disturbing monitoring operations. A systematic approach to troubleshooting will generally consist of the following five steps: Note any warning messages and take corrective action as necessary. Examine the values of all TEST functions and compare them to factory values. Note any major deviations from the factory values and take corrective action. Use the internal electronic status LED’s to determine whether the electronic communication channels are operating properly. Verify that the DC power supplies are operating properly by checking the voltage test points on the relay board. Note that the analyzer’s DC power wiring is color-coded and these colors match the color of the corresponding test points on the relay board. Suspect a leak first! Customer service data indicate that the majority of all problems are eventually traced to leaks in the pneumatic system of the analyzer (including the external pump), the source of zero air or span gases or the sample gas delivery system. Check for gas flow problems such as clogged or blocked internal/external gas lines, damaged seals, punctured gas lines, a damaged pump diaphragm, etc. Follow the procedures defined in Section 3.2.4. to confirm that the analyzer’s vital functions are working (power supplies, CPU, relay board, keyboard, PMT cooler, etc.). See Figure 3-1, Figure 3-2, and Figure 3-3 for general layout of components and sub-assemblies in the analyzer. See the wiring interconnect diagram (document 04504) and interconnect list (document 04496) in Appendix D. 231 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.1.1. WARNING MESSAGES The most common and/or serious instrument failures will result in a warning message displayed on the front panel. Table A-2 in Appendix A.3 contains a list of warning messages, along with their meaning and recommended corrective action. It should be noted that if more than two or three warning messages occur at the same time, it is often an indication that some fundamental analyzer sub-system (power supply, relay board, motherboard) has failed rather than an indication of the specific failures referenced by the warnings. In this case, a combined-error analysis needs to be performed. The analyzer will alert the user that a warning is active by displaying the keypad labels MSG and CLR on the front panel and a text message in the top center line of the display as shown in this example: SAMPLE AZERO WARNING < TST TST > CAL NOX =123.4 MSG CLR SETUP The analyzer will also issue a message to the serial port and cause the red FAULT LED on the front panel to blink. To view or clear a warning messages press: SAMPLE keys replaced with TEST key. Pressing TEST deactivates warning messages until new warning(s) are activated. TEST SAMPLE SYSTEM RESET CAL If warning messages re-appear, the cause needs to be found. Do not repeatedly clear warnings without corrective action. MSG SYSTEM RESET < TST TST > CAL Figure 11-1: MSG A1:NXCNC1=100PPM < TST TST > CAL SAMPLE NOX = XXX.X CLR SETUP NOX=XXX.X CLR SETUP NOX = XXX.X MSG CLR SETUP MSG indicates that warning messages are active. All Warning messages are hidden, but MSG button appears Press CLR to clear the current warning message. If more than one warning is active, the next message will take its place. Once the last warning has been cleared, the analyzer returns to SAMPLE Mode. Viewing and Clearing Warning Messages 11.1.2. FAULT DIAGNOSIS WITH TEST FUNCTIONS Besides being useful as predictive diagnostic tools, the TEST functions, viewable from the front panel, can be used to isolate and identify many operational problems when combined with a thorough understanding of the analyzer’s theory of operation (Chapter 10). We recommend to use the APICOM remote control program to download, graph and archive TEST data for analysis and long-term monitoring of diagnostic data ( Section 6/15.2.8). The acceptable ranges for these test functions are listed in Appendix A-3. The actual values for these test functions on checkout at the factory were also listed in the Final Test and Validation Data Sheet, which was shipped with the instrument. Values outside the acceptable ranges indicate a failure of one or more of the analyzer’s subsystems. Functions with values that are within the acceptable range but have significantly changed from the measurements recorded on the factory data sheet may also indicate a failure or a 232 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair maintenance item. A problem report worksheet has been provided in Appendix C (Teledyne Instruments part number 04503) to assist in recording the value of these test functions. The following table contains some of the more common causes for these values to be out of range. Table 11-1: Test Functions - Possible Causes for Out-Of-Range Values TEST FUNCTION NOX STB INDICATED FAILURE(S) Unstable concentrations; leaks SAMPLE FL Leaks; clogged critical flow orifice OZONE FL Leaks; clogged critical flow orifice PMT NORM PMT AZERO HVPS RCELL TEMP Calibration off; HVPS problem; no flow (leaks) AutoZero too high Leaks; malfunctioning NO/NOx or AutoZero valve; O3 air filter cartridge exhausted HVPS broken; calibration off; preamp board circuit problems Malfunctioning heater; relay board communication (I2C bus); relay burnt out BOX TEMP Environment out of temperature operating range; broken thermistor PMT TEMP TEC cooling circuit broken; relay board communication (I2C bus); 12 V power supply IZS TEMP (OPTION) MOLY TEMP Malfunctioning heater; relay board communication (I2C bus); relay burnt out Malfunctioning heater; disconnected or broken thermocouple; relay board communication (I2Z bus); relay burnt out; incorrect AC voltage configuration RCEL (PRESSURE) Leak; malfunctioning valve; malfunctioning pump; clogged flow orifices SAMP (PRESSURE) Leak; malfunctioning valve; malfunctioning pump; clogged flow orifices; sample inlet overpressure; NOX SLOPE NOX OFF NO SLOPE NO OFFS TIME OF DAY HVPS out of range; low-level (hardware) calibration needs adjustment; span gas concentration incorrect; leaks Incorrect span gas concentration; low-level calibration off HVPS out of range; low-level calibration off; span gas concentration incorrect; leaks Incorrect span gas concentration; low-level calibration off Internal clock drifting; move across time zones; daylight savings time? 11.1.3. USING THE DIAGNOSTIC SIGNAL I/O FUNCTION The signal I/O parameters found under the diagnostics (DIAG) menu combined with a thorough understanding of the instrument’s theory of operation (Chapter 10) are useful for troubleshooting in three ways: The technician can view the raw, unprocessed signal level of the analyzer’s critical inputs and outputs. All of the components and functions that are normally under instrument control can be manually changed. Analog and digital output signals can be manually controlled. This allows to systematically observe the effect of these functions on the operation of the analyzer. Figure 11-2 shows an example of how to use the signal I/O menu to view the raw voltage of an input signal or to control the state of an output voltage or control signal. The specific parameter will vary depending on the situation. 233 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair SAMPLE A1:NXCNC1=100PPM < TST TST > SETUP X.X NOX=XXX.X CAL SETUP PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X COMM SECONDARY SETUP MENU VARS DIAG SETUP X.X 8 1 DIAG EXIT ALRM EXIT ENTER PASSWORD:818 8 ENTR EXIT SIGNAL I/O NEXT DIAG I/O ENTR 0) EXT_ZERO_CAL =OFF NEXT JUMP DIAG I/O 0 EXIT ENTR EXIT JUMP TO:0 0 ENTR EXIT Enter 07 to Jump to Signal 7: (CAL_LED) DIAG I/O 0 DIAG AIO JUMP TO:7 7 ENTR EXIT 7) CAL LED=OFF PREV NEXT JUMP OFF PRNT EXIT Toggle this Key to turn the CAL LED ON/OFF Figure 11-2: Switching Signal I/O Functions 234 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.1.4. STATUS LED’S Several color-coded, light-emitting diodes (LED) are located inside the instrument to determine if the analyzer’s CPU, I2C communications bus and the relay board are functioning properly. 11.1.4.1. Motherboard Status Indicator (Watchdog) A red LED labeled DS5 in the upper portion of the motherboard (Figure 11-3), just to the right of the CPU board, flashes when the CPU is running the main program. After power-up, DS5 should flash on and off about once per second. If characters are visible on the front panel display but DS5 does not flash then the program files have become corrupted. Contact customer service because it may be possible to recover operation of the analyzer. If 30 - 60 seconds after a restart neither DS5 is flashing nor any characters are visible on the front panel display, the firmware may be corrupted or the CPU may be defective. If DS5 is permanently off or permanently on, the CPU board is likely locked up and the analyzer should not respond (either with locked-up or dark front panel). Figure 11-3: Motherboard Watchdog Status Indicator 11.1.4.2. CPU Status Indicator The CPU board has two red LEDs, the lower of which is the watchdog timer (the device that pulses the motherboard watchdog). This LED is labeled LED2 and blinks about twice per second (twice as fast as the motherboard LED) when operating normally. LED1 above LED2 should always be on. However, both CPU LEDs only indicate if the CPU is powered up properly and generally working. The lower LED can continue to blink even if the CPU or firmware are locked up. 11.1.4.3. Relay Board and Status LEDs The most important status LED on the relay board is the red I2C Bus watch-dog LED, labeled D1, which indicates the health of the I2C communications bus. This LED is the left-most in LED row 1 in the center of the relay board when looking at the electronic components. If D1 is blinking, then the other LEDs can be used in 235 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair conjunction with the DIAG menu I/O functions to test hardware functionality by manually switching devices on and off and watching the corresponding LED go on or off. Figure 11-4 illustrates the relay board layout including the two rows of LEDs, Table 11-2 lists the individual LED functions and the menu tree below shows how to access the manual control of the I/O functions. Note that only some or the LEDs may be functional in your analyzer model; the relay board layout is conceptualized for spare, future functionality and is also common to many of the E-series analyzers. Thermocouple Signal Output Status LED’s (D2 through D16) Watchdog Status LED (D1) (JP5) Thermocouple Configuration Jumpers DC Power Supply Test Points I2C Connector (J15) TC1 Input Power Connection for DC Heaters (J16) TC2 Input Shutter Control Connector (JP7) Pump AC Configuration Jumper (M100E Series Only) Valve Control Drivers Pump Power Output Valve Option Control Connector AC Power IN AC Heater Power Output Solid State AC Power Relays (Not Present on P/N 45230100) (JP6) (JP2) AC Configuration Jumpers for Optional IZS Valve Heaters & 02Sensors Figure 11-4: DC Power Distribution Connectors Main AC Heater Configuration Jumpers AC Power Output for Optional IZS Valve Heaters & 02 sensors Relay Board PCA 236 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair Table 11-2: Relay Board Status LEDs COLOR FUNCTION FAULT STATUS INDICATED FAILURE(S) D1 Red Watchdog Circuit; I2C bus operation. Continuously ON or OFF Failed or halted CPU; faulty motherboard, keyboard, relay board; wiring between motherboard, keyboard or relay board; +5 V power supply D2 Yellow Relay 0 - reaction cell heater Continuously ON or OFF Heater broken, thermistor broken D3 Yellow Relay 1 - NO2 converter heater Continuously ON or OFF Heater broken, thermocouple broken D4 Yellow Relay 2 - manifold heater Continuously ON or OFF Heater broken, thermistor broken D7 1 Green Valve 0 - zero/span valve status Continuously ON or OFF Valve broken or stuck, valve driver chip broken D8 1 Green Valve 1 - sample/cal valve status Continuously ON or OFF Valve broken or stuck, valve driver chip broken D9 Green Valve 2 - auto-zero valve status Continuously ON or OFF Valve broken or stuck, valve driver chip broken D10 Green Valve 3 - NO/NOx valve status Continuously ON or OFF Valve broken or stuck, valve driver chip broken D5 Yellow Relay 3 - IZS heater Continuously ON or OFF Heater broken, thermistor broken D6 Yellow Relay 4 – (O2 sensor heater 200EH/EM) N/A N/A D11- 16 Green Spare N/A N/A LED LED ROW 1 LED ROW 2 1 Only active for instruments with Z/S valve options installed To enter the signal I/O test mode to manually control I/O functions such as valves and heaters, press the following keys while observing the relay board LEDs when toggling: SAMPLE A1:NXCNC1=100PPM < TST TST > SETUP X.X NOX=XXX.X CAL PRIMARY SETUP MENU COMM EXIT 0 1 JUMP TO:0 0 ENTR EXIT Enter 07 to Jump to Signal 7: (CAL_LED) ALRM EXIT DIAG I/O 0 8 JUMP TO:25 7 ENTR EXIT ENTER PASSWORD:818 8 ENTR EXIT DIAG AIO 25) RELAY_WATCHDOG=ON PREV NEXT JUMP DIAG SIGNAL I/O NEXT ENTR EXIT SECONDARY SETUP MENU VARS DIAG SETUP X.X 0) EXT_ZERO_CAL =OFF NEXT JUMP DIAG I/O CFG DAS RNGE PASS CLK MORE SETUP X.X DIAG I/O SETUP ENTR EXIT Toggle this Key to turn the CAL LED ON/OFF ON PRNT EXIT See Menu Tree A-6 in Appendix A.1 for a list of I/O Signals 237 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.2. GAS FLOW PROBLEMS The M200EH/EM has two main flow paths, the sample flow and the flow of the ozone supply air. With IZS or zero/span valve option installed, there is a third (zero air) and a fourth (span gas) flow path, but either one of those is only controlled by critical flow orifices and not displayed on the front panel or stored to the iDAS. The full flow diagrams of the standard configuration and with options installed (Appendix D, document 04574) help in trouble-shooting flow problems. In general, flow problems can be divided into three categories: Flow is too high Flow is greater than zero, but is too low, and/or unstable Flow is zero (no flow) When troubleshooting flow problems, it is essential to confirm the actual flow rate without relying on the analyzer’s flow display. The use of an independent, external flow meter to perform a flow check as described in Section 6.13.7.5 is essential. The flow diagrams found in a variety locations within this manual depicting the M200EH and M200EM in their standard configuration and with options installed can help in trouble-shooting flow problems. For your convenience they are colleted here in Sections 11.2.1 (M200EH) and 11.2.2 (M200EM) 11.2.1. M200EH INTERNAL GAS FLOW DIAGRAMS Figure 11-5: M200EH – Basic Internal Gas Flow 238 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair Figure 11-6: M200EH – Internal Gas Flow With OPT 50 Figure 11-7: M200EH – Internal Gas Flow With OPT 52 239 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair Figure 11-8: Figure 11-9: M200EH – Internal Gas Flow With OPT 65 M200EH – Internal Gas Flow With OPT 50 + OPT 65 240 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.2.2. M200EM INTERNAL GAS FLOW DIAGRAMS Figure 11-10: M200EM – Basic Internal Gas Flow Figure 11-11: M200EM – Internal Gas Flow With OPT 50 241 04521C (DCN5731) Troubleshooting & Repair Teledyne API - Model 200EH/EM Operation Manual Figure 11-12: M200EM – Internal Gas Flow With OPT 52 Figure 11-13: M200EM – Internal Gas Flow With OPT 65 242 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual SAMPLE/ CAL VALVE BYPASS MANIFOLD Orifice Dia. 0.003" SPAN GAS INLET FLOW PRESSURE SENSOR PCA NO/NOX VALVE VACUUM PRESSURE SENSOR NO2 Converter ZERO GAS INLET ZERO/SPAN VALVE O2 Sensor EXHAUST GAS OUTLET SAMPLE PRESSURE SENSOR AUTOZERO VALVE EXHAUST MANIFOLD O3 Purifier NOX Exhaust Scrubber O3 FLOW SENSOR SAMPLE GAS INLET Troubleshooting & Repair Orifice Dia. 0.007" Orifice Dia. 0.004" O3 GENERATOR O3 Scrubber REACTION CELL Orifice Dia. 0.004" PUMP PMT Filter PERMAPURE DRYER INSTRUMENT CHASSIS Figure 11-14: M200EM – Internal Gas Flow With OPT 50 + OPT 65 11.2.3. ZERO OR LOW FLOW PROBLEMS 11.2.3.1. Sample Flow is Zero or Low The M200EH/EM does not actually measure the sample flow but rather calculates it from a differential pressure between sample and vacuum manifold. On flow failure, the unit will display a SAMPLE FLOW WARNING on the front panel display and the respective test function reports XXXX instead of a value “0”. This message applies to both a flow rate of zero as well as a flow that is outside the standard range (200-600 cm³/min; 300-700 cm³/min with O2 option installed). If the analyzer displays XXXX for the sample flow, confirm that the external sample pump is operating and configured for the proper AC voltage. Whereas the M200EH/EM can be internally configured for two different power regimes (100-120 V and 220-240 V, either 50 or 60 Hz), the external pump is physically different for each of three power regimes (100 V / 50 Hz, 115 V / 60 Hz and 230 V / 50 Hz). If the pump is not running, use an AC Voltmeter to make sure that the pump is supplied with the proper AC power. If AC power is supplied properly, but the pump is not running, replace the pump. NOTE Sample and vacuum pressures mentioned in this chapter refer to operation of the analyzer at sea level. Pressure values need to be adjusted for elevated locations, as the ambient pressure decreases by about 1 in-Hg per 300 m / 1000 ft. If the pump is operating but the unit reports a XXXX gas flow, do the following three steps: 243 04521C (DCN5731) Troubleshooting & Repair Teledyne API - Model 200EH/EM Operation Manual Check for actual sample flow. To check the actual sample flow, disconnect the sample tube from the sample inlet on the rear panel of the instrument. Make sure that the unit is in basic SAMPLE mode. Place a finger over the inlet and see if it gets sucked in by the vacuum or, more properly, use a flow meter to measure the actual flow. If there is proper flow (see Table 10-3 for flow rates), contact customer service. If there is no flow or low flow, continue with the next step. Check pressures. Check that the sample pressure is at or around 28 in-Hg-A at sea level (adjust as necessary when in elevated location, the pressure should be about 1” below ambient atmospheric pressure) and that the RCEL pressure is below 10 in-Hg-A. The M200EH/EM will calculate a sample flow up to about 14 in-Hg-A RCEL pressure but a good pump should always provide less than 10 in. If both pressures are the same and around atmospheric pressure, the pump does not operate properly or is not connected properly. The instrument does not get any vacuum. If both pressures are about the same and low (probably under 10 in-Hg-A, or ~20” on sample and 15” on vacuum), there is a cross-leak between sample flow path and vacuum, most likely through the Perma Pure dryer flow paths. See troubleshooting the Perma Pure dryer later in this chapter. If the sample and vacuum pressures are around their nominal values (28 and <10 in-Hg-A, respectively) and the flow still displays XXXX, carry out a leak check as described in Section 0 If gas flows through the instrument during the above tests but goes to zero or is low when it is connected to zero air or span gas, the flow problem is not internal to the analyzer but likely caused by the gas source such as calibrators/generators, empty gas tanks, clogged valves, regulators and gas lines. If an IZS or Zero/Span valve option is installed in the instrument, press CALZ and CALS. If the sample flow increases, suspect a bad Sample/Cal valve. If none of these suggestions help, carry out a detailed leak check of the analyzer as described in Section 0. 11.2.3.2. Ozone Flow is Zero or Low If there is zero or a low (<200 cm³/min) ozone flow, the unit displays an OZONE FLOW WARNING message on the front panel and a value between 0.0 and 200 cm³/min for the actual ozone flow as measured by the internal mass flow meter. In this case, carry out the following steps: Check the actual flow rate through the ozone dryer by using an external flow meter to the inlet port of the dryer. This inlet port is inside the analyzer at the end of the plastic particle filter (Section 9.3.2 for illustration). If there is nominal flow (see Table 10-3 for flow rates), consult customer service as there is a problem with the firmware or electronics. If the actual flow is low or zero, check if the pump operates properly. The RCEL pressure should be below 10 in-Hg-A at sea level. If it is above 10”, rebuild the pump (Section 9.3.3). Check the spare parts list in Appendix B on how to order pump rebuild kits. Check if the particle filter is clogged. Briefly remove the particle filter to see if this improves the flow. Be very cautious about handling the Perma Pure dryer fittings - refer to Section 9.3.2 on proper handling instructions. If the filter is clogged, replace it with a new unit. If taking off this filter does not solve the problem, continue to the next step. Do not leave the Perma Pure dryer without filter for more than a few seconds, as you may draw in dust, which will reduce the performance of the dryer. A leak between the flow meter and the reaction cell (where the flow-determining critical orifice is located) may cause a low flow (the system draws in ambient air through a leak after the flow meter). Check for leaks as described in Section 0. Repair the leaking fitting, line or valve and re-check. 244 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair The most likely cause for zero or low ozone flow is a clogged critical flow orifice or sintered filter within the orifice assembly. The orifice that sets the ozone flow is located on the reaction cell. Check the actual ozone flow by disconnecting the tube from the reaction cell and measuring the flow going into the cell. If this flow is correct (see Table 10-3 for flow rates), the orifice works properly. If this flow is low, replace or clean the orifice. The orifice holder assembly allows a quick and easy replacement of the orifice, refer to Section 9.3.8 on how to do this. Appendix B lists a spare part kit with a complete orifice assembly that allows a quick replacement with minimum instrument down-time. The clogged orifice can then be cleaned while the instrument is running with the replacement. 11.2.4. HIGH FLOW Flows that are significantly higher than the allowed operating range (typically ±10-11% of the nominal flow) should not occur in the M200EH/EM unless a pressurized sample, zero or span gas is supplied to the inlet ports. Ensure to vent excess pressure and flow just before the analyzer inlet ports. When supplying sample, zero or span gas at ambient pressure, a high flow would indicate that one or more of the critical flow orifices are physically broken (very unlikely case), allowing more than nominal flow, or were replaced with an orifice of wrong specifications. If the flows are within 15% higher than normal, we recommend to re-calibrate the flow electronically using the procedure in Section 6.13.7.5, followed by a regular review of these flows over time to see if the new setting is retained properly. 11.2.5. SAMPLE FLOW IS ZERO OR LOW BUT ANALYZER REPORTS CORRECT FLOW Note that the M200EH/EM analyzer can report a correct flow rate even if there is no or a low actual sample flow through the reaction cell. The sample flow on the M200EH/EM is only calculated from the sample pressure and critical flow condition is verified from the difference between sample pressure and vacuum pressure. If the critical flow orifice is partially or completely clogged, both the sample and vacuum pressures are still within their nominal ranges (the pump keeps pumping, the sample port is open to the atmosphere), but there is no flow possible through the reaction cell. Although measuring the actual flow is the best method, in most cases, this fault can also be diagnosed by evaluating the two pressure values. Since there is no longer any flow, the sample pressure should be equal to ambient pressure, which is about 1 in-Hg-A higher than the sample pressure under normal operation. The reaction cell pressure, on the other hand, is significantly lower than under normal operation, because the pump no longer has to remove 500 cm³/min of sample gas and evacuates the reaction cell much better. Those two indicators, taken together with a zero or low actual flow, indicate a clogged sample orifice. The M200EH/EM features a new orifice holder, which makes switching sample and ozone flow orifices very easy, refer to Section 9.3.8 on how to change the sample orifices and Appendix B for part numbers of these assemblies. Again, monitoring the pressures and flows regularly will reveal such problems, because the pressures would slowly or suddenly change from their nominal, mean values. Teledyne Instruments recommends to review all test data once per week and to do an exhaustive data analysis for test and concentration values once per month, paying particular attention to sudden or gradual changes in all parameters that are supposed to remain constant, such as the flow rates. 245 04521C (DCN5731) Troubleshooting & Repair Teledyne API - Model 200EH/EM Operation Manual 11.3. CALIBRATION PROBLEMS 11.3.1. NEGATIVE CONCENTRATIONS Negative concentration values can be caused for several reasons: A slight, negative signal is normal when the analyzer is operating under zero gas and the signal is drifting around the zero calibration point. This is caused by the analyzer’s zero noise and may cause reported concentrations to be negative for a few seconds at a time down to -0.2 ppm, but should randomly alternate with similarly high, positive values. The M200EH/EM has a built-in Auto-zero function, which should take care of most of these deviations from zero, but may yield a small, residual, negative value. If larger, negative values persist continuously, check if the Auto-zero function was accidentally turned off using the remote variables in Appendix A-2. In this case, the sensitivity of the analyzer may be drifting negative. A corruption of the Auto-zero filter may also cause negative concentrations. If a short, high noise value was detected during the AutoZero cycle, that higher reading will alter the Auto-zero filter value. As the value of the Auto-zero filter is subtracted from the current PMT response, it will produce a negative concentration reading. High AutoZero readings can be caused by a leaking or stuck AutoZero valve (replace the valve), by an electronic fault in the preamplifier causing it to have a voltage on the PMT output pin during the AutoZero cycle (replace the preamplifier), by a reaction cell contamination causing high background (>40 mV) PMT readings (clean the reaction cell), by a broken PMT temperature control circuit, allowing high zero offset (repair the faulty PMT cooler). After fixing the cause of a high Auto-zero filter reading, the M200EH/EM will take 15 minutes for the filter to clear itself, or by an exhausted chemical in the ozone scrubber cartridge (Section Mis-calibration is the most likely explanation for negative concentration values. If the zero air contained some NO or NO2 gas (contaminated zero air or a worn-out zero air scrubber) and the analyzer was calibrated to that concentration as “zero”, the analyzer may report negative values when measuring air that contains little or no NOx. The same problem occurs, if the analyzer was zero-calibrated using zero gas that is contaminated with ambient air or span gas (cross-port leaks or leaks in supply tubing or user not waiting long enough to flush pneumatic systems). If the response offset test functions for NO (NO OFFS) or NOX (NOX OFFS) are greater than 150 mV, a reaction cell contamination is indicated. Clean the reaction cell according to Section 9.3.7. 11.3.2. NO RESPONSE If the instrument shows no response (display value is near zero) even though sample gas is supplied properly and the instrument seems to perform correctly. Check if the ozone generator is turned on. Usually, the analyzer issues a warning whenever the ozone generator is turned off. Go to SETUP-MORE-DIAG-ENTR, then scroll to the OZONE GEN OVERRIDE and see if it shows ON. If it shows OFF, turn it ON and EXIT the DIAG menu. If this is done and the ozone flow is correct, the analyzer should be properly supplied with ozone unless the generator itself is broken. A more detailed description of the ozone generator subsystem checks are in Section 11.5.17. Confirm the lack of response by supplying NO or NO2 span gas of about 80% of the range value to the analyzer. Check the sample flow and ozone flow rates for proper values. 246 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair Check for disconnected cables to the sensor module. Carry out an electrical test with the ELECTRICAL TEST procedure in the diagnostics menu, see Section 6.13.7.3. If this test produces a concentration reading, the analyzer’s electronic signal path is correct. Carry out an optical test using the OPTIC TEST procedure in the diagnostics menu, see Section 6.13.6.2. If this test results in a concentration signal, then the PMT sensor and the electronic signal path are operating properly. If the M200EH/EM passes both ETEST and OTEST, the instrument is capable of detecting light and processing the signal to produce a reading. Therefore, the problem must be in the pneumatics or the ozone generator. If NO2 signal is zero while NO signal is correct, check the NO/NOX valve and the NO2 converter for proper operation. 11.3.3. UNSTABLE ZERO AND SPAN Leaks in the M200EH/EM or in the external gas supply and vacuum systems are the most common source of unstable and non-repeatable concentration readings. Check for leaks in the pneumatic systems as described in Section 0. Consider pneumatic components in the gas delivery system outside the M200EH/EM such as a change in zero air source (ambient air leaking into zero air line or a worn-out zero air scrubber) or a change in the span gas concentration due to zero air or ambient air leaking into the span gas line. Once the instrument passes a leak check, do a flow check (this chapter) to make sure that the instrument is supplied with adequate sample and ozone air. Confirm the sample pressure, sample temperature, and sample flow readings are correct and steady. Verify that the sample filter element is clean and does not need to be replaced. 11.3.4. INABILITY TO SPAN - NO SPAN KEY In general, the M200EH/EM will not display certain keyboard choices whenever the actual value of a parameter is outside of the expected range for that parameter. If the calibration menu does not show a SPAN key when carrying out a span calibration, the actual concentration must be outside of the range of the expected span gas concentration, which can have several reasons. Verify that the expected concentration is set properly to the actual span gas concentration in the CONC sub-menu. Confirm that the NOx span gas source is accurate. This can be done by comparing the source with another calibrated analyzer, or by having the NOx source verified by an independent traceable photometer. Check for leaks in the pneumatic systems as described in Section 0. Leaks can dilute the span gas and, hence, the concentration that the analyzer measures may fall short of the expected concentration defined in the CONC sub-menu. If the low-level, hardware calibration has drifted (changed PMT response) or was accidentally altered by the user, a low-level calibration may be necessary to get the analyzer back into its proper range of expected values. One possible indicator of this scenario is a slope or offset value that is outside of its allowed range (0.7-1.3 for slope, -20 to 150 for offsets). See Section 11.6.5 on how to carry out a lowlevel hardware calibration. 247 04521C (DCN5731) Troubleshooting & Repair Teledyne API - Model 200EH/EM Operation Manual 11.3.5. INABILITY TO ZERO - NO ZERO KEY In general, the M200EH/EM will not display certain keyboard choices whenever the actual value of a parameter is outside of the expected range for that parameter. If the calibration menu does not show a ZERO key when carrying out a zero calibration, the actual gas concentration must be significantly different from the actual zero point (as per last calibration), which can have several reasons. Confirm that there is a good source of zero air. If the IZS option is installed, compare the zero reading from the IZS zero air source to a zero air source using NOX-free air. Check any zero air scrubber for performance. It may need to be replaced (Section 9.3.5). Check to make sure that there is no ambient air leaking into zero air line. Check for leaks in the pneumatic systems as described in Section 11.5. 11.3.6. NON-LINEAR RESPONSE The M200EH/EM was factory calibrated to a high level of NO and should be linear to within 1% of full scale. Common causes for non-linearity are: Leaks in the pneumatic system. Leaks can add a constant of ambient air, zero air or span gas to the current sample gas stream, which may be changing in concentrations as the linearity test is performed. Check for leaks as described in Section 11.5. The calibration device is in error. Check flow rates and concentrations, particularly when using low concentrations. If a mass flow calibrator is used and the flow is less than 10% of the full scale flow on either flow controller, you may need to purchase lower concentration standards. The standard gases may be mislabeled as to type or concentration. Labeled concentrations may be outside the certified tolerance. The sample delivery system may be contaminated. Check for dirt in the sample lines or reaction cell. Calibration gas source may be contaminated (NO2 in NO gas is common). Dilution air contains sample or span gas. Ozone concentration too low because of wet air in the generator. Generator system needs to be cleaned and dried with dry supply air. Check the Perma Pure dryer for leaks. This mostly affects linearity at the low end. Sample inlet may be contaminated with NOX exhaust from this or other analyzers. Verify proper venting of the pump exhaust. Span gas overflow is not properly vented and creates a back-pressure on the sample inlet port. Also, if the span gas is not vented at all and does not supply enough sample gas, the analyzer may be evacuating the sample line. Make sure to create and properly vent excess span gas. Diffusion of oxygen into Teflon-type tubing over long distances. PTFE or related materials can act as permeation devices. In fact, the permeable membrane of NO2 permeation tubes is made of PTFE. When using very long supply lines (> 1 m) between high concentrations span gases and the dilution system, oxygen from ambient air can diffuse into the line and react with NO to form NO2. This reaction is dependent on NO concentration and accelerates with increasing NO concentration, hence, affects linearity only at high NO levels. Using stainless steel for long span gas supply lines avoids this problem. 5 5 248 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.3.7. DISCREPANCY BETWEEN ANALOG OUTPUT AND DISPLAY If the concentration reported through the analog outputs does not agree with the value reported on the front panel, you may need to re-calibrate the analog outputs. This becomes more likely when using a low concentration or low analog output range. Analog outputs running at 0.1 V full scale should always be calibrated manually. See Section 6.13.4.4 for a detailed description of this procedure. 11.3.8. DISCREPANCY BETWEEN NO AND NOX SLOPES If the slopes for NO and NOX are significantly different after software calibration (more than 1%), consider the following two problems NO2 impurities in the NO calibration gas. NO gases often exhibit NO2 on the order of 1-2% of the NO value. This will cause differences in the calibration slopes. If the NO2 impurity in NO is known, it can easily be accounted for by setting the expected values for NO and NO2 accordingly to different values, e.g., 0.448 ppm NO and 0.45 ppm NOX. This problem is worse if NO gas is stored in a cylinder with balance air instead of balance gas nitrogen or large amounts of nitrous oxide (N2O). The oxygen in the air slowly reacts with NO to yield NO2, increasing over time. The expected concentrations for NO and NOX in the calibration menu are set to different values. If a gas with 100% pure NO is used, this would cause a bias. See Section 7.2 on how to set expected concentration values. The converter efficiency parameter has been set to a value not equal to 1.000 even though the conversion efficiency is 1.0. The actual conversion efficiency needs to match the parameter set in the CAL menu. See Section 7.1.5 for more information on this feature. An instrument calibration with the IZS option (and expected concentrations set to the same amount) will always yield identical slopes for NO and NOX, as the instrument measures only NOX and assumes NO to be the same (with NO2 being zero). 11.4. OTHER PERFORMANCE PROBLEMS Dynamic problems (i.e. problems which only manifest themselves when the analyzer is monitoring sample gas) can be the most difficult and time consuming to isolate and resolve. The following section provides an itemized list of the most common dynamic problems with recommended troubleshooting checks and corrective actions. 11.4.1. EXCESSIVE NOISE Excessive noise levels under normal operation usually indicate leaks in the sample supply or the analyzer itself. Make sure that the sample or span gas supply is leak-free and carry out a detailed leak check as described earlier in this chapter. Another possibility of excessive signal noise may be the preamplifier board, the high voltage power supply and/or the PMT detector itself. Contact the factory on trouble-shooting these components. 11.4.2. SLOW RESPONSE If the analyzer starts responding too slow to any changes in sample, zero or span gas, check for the following: Dirty or plugged sample filter or sample lines. Sample inlet line is too long. Leaking NO/NOX valve. Carry out a leak check. 249 04521C (DCN5731) Troubleshooting & Repair Teledyne API - Model 200EH/EM Operation Manual Dirty or plugged critical flow orifices. Check flows, pressures and, if necessary, change orifices (Section 9.3.8). Wrong materials in contact with sample - use glass, stainless steel or Teflon materials only. Porous materials, in particular, will cause memory effects and slow changes in response. Dirty reaction cell. Clean the reaction cell. Insufficient time allowed for purging of lines upstream of the analyzer. Wait until stability is low. Insufficient time allowed for NO or NO2 calibration gas source to become stable. Wait until stability is low. NO2 converter temperature is too low. Check for proper temperature. 11.4.3. AUTO-ZERO WARNINGS Auto-zero warnings occur if the signal measured during an auto-zero cycle is lower than –20 mV or higher than 200 mV. The Auto-Zero warning displays the value of the auto-zero reading when the warning occurs. If this value is higher than 150 mV, check that the auto-zero valve is operating properly. To do so, use the SIGNAL I/O functions in the DIAG menu to toggle the valve on and off. Listen if the valve is switching, see if the respective LED on the relay board is indicating functionality. Scroll the TST functions until PMT is displayed and observe the PMT value change between the two valve states. If the valve is operating properly, you should be able to hear it switch (once a minute under normal operation or when manually activated from the SIGNAL I/O menu), the PMT value should drop from its nominal reading for span gas level measurements to less than 150 mV and the LED on the relay board should light up when the valve is activated. If the PMT value drops significantly but not to less than 150 mV, the valve is probably leaking across its ports. In this case, replace the valve. If the PMT value does not change at all, the valve is probably not switching at all. Check the power supply to the valve (12 V to the valve should turn on and off when measured with a voltmeter). Note that it takes only a small leak across the ports of the valve to show excessive auto-zero values when supplying high concentrations of span gas. Another reason for high (although not necessarily out-of-range) values for AutoZero could be the ozone air filter cartridge, if its contents has been exhausted and needs to be replaced. This filter cartridge (Figure 3-1) filters chemicals that can cause chemiluminescence and, if saturated, these chemicals can break through to the reaction cell, causing an erroneously high AutoZero value (background noise). A dirty reaction cell can cause high AutoZero values. Clean the reaction cell according to Section 9.3.7. Finally, a high HVPS voltage value may cause excess background noise and a high AZERO value. The HVPS value changes from analyzer to analyzer and could show nominal values between 450 and 800 V. Check the low-level hardware calibration of the preamplifier board and, if necessary, recalibrate exactly as described in Section 11.6.5 in order to minimize the HVPS. 250 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.5. SUBSYSTEM CHECKOUT The preceding sections of this manual discussed a variety of methods for identifying possible sources of failures or performance problems within the analyzer. In most cases this included a list of possible causes and, in some cases, quick solutions or at least a pointer to the appropriate sections describing them. This section describes how to determine if a certain component or subsystem is actually the cause of the problem being investigated. 11.5.1. SIMPLE VACUUM LEAK AND PUMP CHECK Leaks are the most common cause of analyzer malfunction; This section presents a simple leak check, whereas Section 0 details a more thorough procedure. The method described here is easy, fast and detects, but does not locate, most leaks. It also verifies the sample pump condition. Turn the analyzer ON, and allow at least 30 minutes for flows to stabilize. Cap the sample inlet port (cap must be wrench-tight). After several minutes, when the pressures have stabilized, note the SAMP (sample pressure) and the RCEL (vacuum pressure) readings. If both readings are equal to within 10% and less than 10 in-Hg-A, the instrument is free of large leaks. It is still possible that the instrument has minor leaks. If both readings are < 10 in-Hg-A, the pump is in good condition. A new pump will create a pressure reading of about 4 in-Hg-A (at sea level). 11.5.2. DETAILED PRESSURE LEAK CHECK If a leak cannot be located by the above procedure, obtain a leak checker similar to Teledyne Instruments part number 01960, which contains a small pump, shut-off valve, and pressure gauge to create both over-pressure and vacuum. Alternatively, a tank of pressurized gas, with the two stage regulator adjusted to ≤ 15 psi, a shutoff valve and pressure gauge may be used. CAUTION Once tube fittings have been wetted with soap solution under a pressurized system, do not apply or re-apply vacuum as this will cause soap solution to be sucked into the instrument, contaminating inside surfaces. Do not exceed 15 psi when pressurizing the system. Turn OFF power to the instrument and remove the instrument cover. Install a leak checker or a tank of gas (compressed, oil-free air or nitrogen) as described above on the sample inlet at the rear panel. Disconnect the pump tubing on the outside rear panel and cap the pump port. If IZS or zero/span valves are installed, disconnect the tubing from the zero and span gas ports and plug them (Figure 3-2). Cap the DFU particle filter on the Perma Pure dryer (Figure 9-2). 251 04521C (DCN5731) Troubleshooting & Repair Teledyne API - Model 200EH/EM Operation Manual Pressurize the instrument with the leak checker or tank gas, allowing enough time to fully pressurize the instrument through the critical flow orifice. Check each tube connection (fittings, hose clamps) with soap bubble solution, looking for fine bubbles. Once the fittings have been wetted with soap solution, do not re-apply vacuum as it will draw soap solution into the instrument and contaminate it. Do not exceed 15 psi pressure. If the instrument has the zero and span valve option, the normally closed ports on each valve should also be separately checked. Connect the leak checker to the normally closed ports and check with soap bubble solution. If the analyzer is equipped with an IZS Option Connect the leak checker to the Dry Air inlet and check with soap bubble solution. Once the leak has been located and repaired, the leak-down rate of the indicated pressure should be less than 1 in-Hg-A (0.4 psi) in 5 minutes after the pressure is turned off. Clean surfaces from soap solution, re-connect the sample and pump lines and replace the instrument cover. Restart the analyzer. 11.5.3. PERFORMING A SAMPLE FLOW CHECK CAUTION Use a separate, calibrated flow meter capable of measuring flows between 0 and 1000 cm³/min to measure the gas flow rate though the analyzer. Do not use the built in flow measurement viewable from the front panel of the instrument. This value is only calculated, not measured. Sample flow checks are useful for monitoring the actual flow of the instrument, as the front panel display shows only a calculated value. A decreasing, actual sample flow may point to slowly clogging pneumatic paths, most likely critical flow orifices or sintered filters. To perform a sample flow check: Disconnect the sample inlet tubing from the rear panel SAMPLE port shown in Figure 3-2. Attach the outlet port of a flow meter to the sample inlet port on the rear panel. Ensure that the inlet to the flow meter is at atmospheric pressure. The sample flow measured with the external flow meter should be within 10% of the nominal values shown in Table 10-3. Low flows indicate blockage somewhere in the pneumatic pathway. ] 252 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.5.4. AC POWER CONFIGURATION The E-Series digital electronic systems will operate with any of the specified power regimes. As long as instrument is connected to 100-120 VAC or 220-240 VAC at either 50 or 60 Hz it will turn on and after about 30 seconds show a front panel display. Internally, the status LEDs located on the Relay PCA, Motherboard and CPU should turn on as soon as the power is supplied. On the other hand, some of the analyzer’s non-digital components, such as the pump and the various AC powered heaters must be properly configured for the type of power being supplied to the instrument. Figure 1116shows the location of the various sets of AC Configuration jumpers. JP6 IZS Permeation Tube Heater and O2 Sensor Connection. (optional) JP7 Pump Configuration JP2 Main AC Heater Configuration Figure 11-15: Location of AC power Configuration Jumpers There are several changes between the Relay PCA 04523 and previous version regarding AC power configuration and distribution. Previously, in analyzer models with internal pumps, the AC power for the pump came directly from the instrument back panel. The 04523 version handles all AC and DC power distribution including power to the pump. Prior to this change, configuring the pump for compatibility with various line voltages and frequencies was done with a set of hard-wired, in-line connections. The Relay PCA 04523, now includes a set of jumpers that perform this function. This change increase reliability and simplifies troubleshooting and repair operations. 253 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair The Relay PCA 04523, includes a set of jumpers that connect AC power to heaters included in several optional items, such as the zero/span valve options and the O2 sensor option available on the M200EH/EM analyzers. In earlier versions of the relay PCA this was also handled by in-line connections. 11.5.4.1. AC configuration – Internal Pump (JP7) AC power configuration for internal pumps is set using Jumper set JP7 (see Figure 11-4 for the location of JP7). Table 11-3: AC Power Configuration for Internal Pumps (JP7) LINE POWER LINE FREQUENCY 60 HZ WHITE 110VAC 115 VAC 1 50 HZ 220VAC 240 VAC 1 60 HZ 50 HZ1 FUNCTION JUMPER BETWEEN PINS Connects pump pin 3 to 110 / 115 VAC power line 2 to 7 Connects pump pin 3 to 110 / 115 VAC power line 3 to 8 Connects pump pins 2 & 4 to Neutral 4 to 9 Connects pump pin 3 to 110 / 115 VAC power line 2 to 7 Connects pump pin 3 to 110 / 115 VAC power line 3 to 8 Connects pump pins 2 & 4 to Neutral 4 to 9 Connects pump pins 3 and 4 together 1 to 6 Connects pump pin 1 to 220 / 240VAC power line 3 to 8 Connects pump pins 3 and 4 together 1 to 6 Connects pump pin 1 to 220 / 240VAC power line 3 to 8 JUMPER COLOR BLACK BROWN BLUE A jumper between pins 5 and 10 may be present on the jumper plug assembly, but is only functional on the M300E and has no function on the M200EH/EM analyzers. 110 VAC /115 VAC 220 VAC /240 VAC 1 6 1 6 2 7 2 7 8 3 3 8 4 9 4 9 5 10 5 10 Present on 50 Hz version of jumper set, and functional for M300E but not Models M100E, M200E & M400E Figure 11-16: Pump AC Power Jumpers (JP7) 254 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.5.4.2. AC Configuration – Standard Heaters (JP2) Power configuration for the AC the standard heaters is set using Jumper set JP2 (see Figure 11-4 for the location of JP2). Table 11-4: Power Configuration for Standard AC Heaters (JP2) LINE VOLTAGE JUMPER BETWEEN PINS FUNCTION 1 to 8 Common 2 to 7 Neutral to Load 3 to 10 Common 4 to 9 Neutral to Load 3 to 10 Common 4 to 9 Neutral to Load 5 to 12 Common 6 to 11 Neutral to Load Reaction Cell / Sample Chamber Heaters2 1 to 7 Load Hi Concentration Converter 3 to 9 Load Moly Converter 3 to 9 Load Bypass Manifold 5 to 11 Load JUMPER COLOR HEATER(S) Reaction Cell / Sample Chamber Heaters 110 VAC / 115 VAC 50Hz & 60 Hz Mini Hi-Con Converter WHITE Moly Converter Bypass Manifold 220 VAC / 240 VAC 50Hz & 60 Hz Reaction Cell or Sample Chamber Heaters Mini Hi-Con or Moly Converter Heaters 200EM/EH By Pass Manifold Heater BLUE 1 7 1 7 2 8 2 8 Reaction Cell or Sample Chamber Heaters 3 9 3 9 4 10 4 10 5 11 5 11 6 12 6 12 110 VAC /115 VAC Mini Hi-Con or Moly Converter Heaters 200EM/EH By Pass Manifold Heater 220 VAC / 240 VAC Figure 11-17: Typical Set Up of AC Heater Jumper Set (JP2) 255 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.5.4.3. AC Configuration –Heaters for Option Packages (JP6) Both the IZS valve option or an O2 sensor options include AC heaters that maintain an optimum operating temperature for key components of those options. Jumper set JP6 is used to connect the heaters associated with those options to AC power. Since these heaters work with either 110/155 VAC or 220/240 VAC, there is only one jumper configuration. Table 11-5: Power Configuration for Optional AC Heaters (JP6) JUMPER COLOR FUNCTION M100E’s, M200E’s & M400E 1 to 8 Common 2 to 7 Neutral to Load M100E’s & M200E’s 3 to 10 Common 4 to 9 Neutral to Load MODEL’S USED ON1 IZS1 Permeation Tube Heater RED O2 Sensor Heater 1 JUMPER BETWEEN PINS HEATER(S) This Option Not Available on the M200EH/EM 10 IZS Permeation Tube 12 Heater 11 6 5 4 9 3 8 7 2 1 O2 Sensor Heater Figure 11-18: Typical Set Up of AC Heater Jumper Set (JP2) 256 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.5.5. DC POWER SUPPLY TEST POINTS Table 11-6: DC Power Test Point and Wiring Color Code NAME TEST POINT# COLOR DEFINITION DGND 1 Black Digital ground +5V 2 Red AGND 3 Green +15V 4 Blue -15V 5 Yellow +12R 6 Purple +12V 7 Orange Analog ground 12 V return (ground) line Table 11-7: DC Power Supply Acceptable Levels CHECK RELAY BOARD TEST POINTS POWER SUPPLY VOLTAG E FROM TO Test Point Test Point MIN V MAX V NAME # NAME # DGND 1 +5 2 +4.80 +5.25 PS1 +5 PS1 +15 AGND 3 +15 4 +13.5 +16.0 PS1 -15 AGND 3 -15V 5 -14.0 -16.0 PS1 AGND AGND 3 DGND 1 -0.05 +0.05 PS1 Chassis DGND 1 Chassis N/A -0.05 +0.05 PS2 +12 +12V Ret 6 +12V 7 +11.8 +12.5 PS2 DGND +12V Ret 6 DGND 1 -0.05 +0.05 The test points are located at the top, right-hand corner of the PCA (see Figure 11-4) 11.5.6. I2C BUS Operation of the I2C bus can be verified by observing the behavior of the LED labeled D1 on the relay board in conjunction with the performance of the front panel display. Assuming that the DC power supplies are operating properly and the wiring from the motherboard to the keyboard as well as from the keyboard to the relay board is intact, the I2C bus is operating properly if: D1 on the relay board is flashing or D1 is not flashing but pressing a key on the front panel results in a change to the display. If the display is locked up or if the analyzer is not booting up at all, the I2C bus may be the cause. Contact customer service if you suspect a problem with the I2C bus. 257 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.5.7. KEYBOARD / DISPLAY INTERFACE The front panel keyboard, the display and the keyboard/display circuit board can be verified by observing the operation of the display when power is applied to the instrument and when a key is pressed on the front panel. Assuming that there are no wiring problems and that the DC power supplies are operating properly: The vacuum fluorescence display is working properly if, on power-up, a “-“ character is visible on the upper left hand corner of the display. If there is no “-“ character on the display at power-up but the D1 LED on the relay board is flashing, the keyboard/display circuit may be bad. If the analyzer starts operation with a normal display but pressing a key on the front panel does not change the display, then there are three possible problems: One or more of the keys is bad, The interrupt signal between the keyboard and the motherboard is broken or The keyboard circuit is bad. You can verify this failure by logging on to the instrument using APICOM or a terminal program. If the analyzer responds to remote commands and the display changes accordingly, the display wiring or the I2C bus may be faulty. 11.5.8. GENREAL RELAY BOARD DIAGNOSTIC The relay board circuit can most easily be checked by observing the condition of its status LEDs as described in Section 11.1.4.3, and the associated output when toggled on and off through the SIGNAL I/O function in the DIAG menu, see Section 6.13.1. If the front panel display responds to key presses and D1 on the relay board is not flashing, then either the wiring between the keyboard and the relay board is bad, or the relay board itself is bad. If D1 on the Relay board is flashing and the status indicator for the output in question (heater, valve, etc.) does not toggle properly using the Signal I/O function, then the associated device (valve or heater) or its control device (valve driver, heater relay) is malfunctioning. Several of the control devices are in sockets and can easily be replaced. The table below lists the control device associated with a particular function: Table 11-8: Relay Board Control Devices Function Control Device Socketed All valves U5 Yes All heaters K1-K5 Yes 258 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.5.9. MOTHERBOARD 11.5.9.1. A/D functions A basic check of the analog to digital (A/D) converter operation on the motherboard is to use the Signal I/O function under the DIAG menu. Check the following two A/D reference voltages and input signals that can be easily measured with a voltmeter. Using the Signal I/O function (Section 6.13.1 Appendix D), view the value of REF_4096_MV and REF_GND. If both are within 3 mV of their nominal values (4096 and 0) and are stable to within ±0.5 mV, the basic A/D converter is functioning properly. If these values fluctuate largely or are off by more than 3 mV, one or more of the analog circuits may be overloaded or the motherboard may be faulty. Choose one parameter in the Signal I/O function such as SAMPLE_PRESSURE (see previous section on how to measure it). Compare its actual voltage with the voltage displayed through the SIGNAL I/O function. If the wiring is intact but there is a difference of more than ±10 mV between the measured and displayed voltage, the motherboard may be faulty. 11.5.9.2. Analog Output Voltages To verify that the analog outputs are working properly, connect a voltmeter to the output in question and perform an analog output step test as described in Section 6.13.3. For each of the steps, taking into account any offset that may have been programmed into the channel (Section 6.13.4.4), the output should be within 1% of the nominal value listed in the table below except for the 0% step, which should be within 2-3 mV. If one or more of the steps is outside of this range, a failure of one or both D/A converters and their associated circuitry on the motherboard is likely. Table 11-9: Analog Output Test Function - Nominal Values FULL SCALE OUTPUT VOLTAGE 100mV STEP % 1V 5V 10V NOMINAL OUTPUT VOLTAGE 1 0 0 mV 0 0 0 2 20 20 mV 0.2 1 2 3 40 40 mV 0.4 2 4 4 60 60 mV 0.6 3 6 5 80 80 mV 0.8 4 8 6 100 100 mV 1.0 5 10 259 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.5.9.3. Status Outputs The procedure below can be used to test the Status outputs. V +DC Gnd Figure 11-19: Typical Set Up of Status Output Test 1. Connect a cable between the “D“ pin and the “” pin on the status output connector. 2. Connect a 1000 Ω resistor between the “+” pin and the pin for the status output that is being tested. 3. Connect a voltmeter between the “D“ pin and the pin of the output being tested (Table 11-10). 4. Under the DIAG / SIGNAL I/O menu (Section 6.13.1), scroll through the inputs and outputs until you get to the output in question. Alternately turn the output on and off. The Voltmeter will read approximately 5 VDC when the output is OFF. The Voltmeter will read approximately 0 VDC when the output is ON. Table 11-10: Status Outputs Pin Assignments PIN # STATUS 1 SYSTEM OK 2 CONC VALID 3 HIGH RANGE 4 ZERO CAL 5 SPAN CAL 6 DIAG MODE 7 LOW 8 SPARE 260 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.5.9.4. Control Inputs The control input bits can be tested by the following procedure: Connect a jumper from the +5 V pin on the STATUS connector to the +5 V on the CONTROL IN connector. Connect a second jumper from the ‘-‘ pin on the STATUS connector to the A pin on the CONTROL IN connector. The instrument should switch from SAMPLE mode to ZERO CAL R mode. Connect a second jumper from the ‘-‘ pin on the STATUS connector to the B pin on the CONTROL IN connector. The instrument should switch from SAMPLE mode to SPAN CAL R mode. In each case, the M200EH/EM should return to SAMPLE mode when the jumper is removed. 11.5.10. CPU There are two major types of CPU board failures, a complete failure and a failure associated with the Disk-OnChip (DOC). If either of these failures occur, contact the factory. For complete failures, assuming that the power supplies are operating properly and the wiring is intact, the CPU is faulty if on power-on: The vacuum fluorescence display does not show a dash in the upper left hand corner There is no activity from the primary RS-232 port (COM1) on the rear panel even if “? ” is pressed. In some rare circumstances, this failure may be caused by a bad IC on the motherboard, specifically U57, the large, 44 pin device on the lower right hand side of the board. If this is true, removing U57 from its socket will allow the instrument to start up but the measurements will be incorrect. If the analyzer stops during initialization (the vacuum fluorescence display shows some text), it is likely that the DOC, the firmware or the configuration and data files have been corrupted or that the wrong firmware was uploaded or does not have the correct filename. 261 04521C (DCN5731) Troubleshooting & Repair Teledyne API - Model 200EH/EM Operation Manual 11.5.11. RS-232 COMMUNICATION 11.5.11.1. General RS-232 Troubleshooting Teledyne Instruments analyzers use the RS-232 protocol as the standard, serial communications protocol. RS232 is a versatile standard, which has been used for many years but, at times, is difficult to configure. Teledyne Instruments conforms to the standard pin assignments in the implementation of RS-232. Problems with RS-232 connections usually center around 4 general areas: Incorrect cabling and connectors. This is the most common problem. See 6.11.3 for connector and pinout information. The communications (baud) rate and protocol parameters are incorrectly configured. See Section 6.11.9 on how to set the baud rate. The COM port communications mode is set incorrectly (Section 6.11.8). If a modem is used, additional configuration and wiring rules must be observed. See Section 6.15.2.6. Incorrect setting of the DTE - DCE switch. Typically, the red LED is on as soon as you power up the analyzer. If not, contact the factory, as this indicates a problem with the motherboard. As the analyzer is connected to the computer with a cable, the green LED should also illuminate. If not, set the DCE/DTE switch to the other position. See also Section 6.11.5. Note that some laptops do not enable their RS-232 port when in power-saving mode. In this case, connect the laptop and start either APICOM or a Hyperterminal window and start communicating with the analyzer. This will enable the serial port on the laptop and the green LED should illuminate. You may have to switch back and forth while communicating to get the right setting. 11.5.11.2. Modem or Terminal Operation These are the general steps for troubleshooting problems with a modem connected to a Teledyne Instruments analyzer. Check cables for proper connection to the modem, terminal or computer. Check the correct position of the DTE/DCE as described in Section 6.11.5. Check the correct setup command (Section 6.15.2.6). Verify that the Ready to Send (RTS) signal is at logic high. The M200EH/EM sets pin 7 (RTS) to greater than 3 volts to enable modem transmission. Make sure the baud rate, word length, and stop bit settings between modem and analyzer match, see Section 6.15.2.6 and 6.11.8. Use the RS-232 test function to send “w” characters to the modem, terminal or computer; See Section 6.11.10. Get your terminal, modem or computer to transmit data to the analyzer (holding down the space bar is one way). The green LED on the rear panel should flicker as the instrument is receiving data. Make sure that the communications software is functioning properly. Further help with serial communications is available in a separate manual “RS-232 Manual”, Teledyne Instruments part number 013500000, available online at http://www.Teledyne-api.com/manuals/. 262 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.5.12. PMT SENSOR The photo multiplier tube detects the light emitted by the reaction of NO with ozone. It has a gain of about 1: 500000 to 1:1000000. It is not possible to test the detector outside of the instrument in the field. The best way to determine if the PMT is working properly is by using the optical test (OTEST), which is described in Section 6.13.6.2. The basic method to diagnose a PMT fault is to eliminate the other components using ETEST, OTEST and specific tests for other sub-assemblies. 11.5.13. PMT PREAMPLIFIER BOARD To check the correct operation of the preamplifier board, we suggest to carry out the optical and electrical tests described in Sections 6.13.6.2 and 6.13.7.3. If the ETEST fails, the preamplifier board may be faulty. Refer to Section 11.6.5 on hardware calibration through the preamplifier board. 11.5.14. HIGH VOLTAGE POWER SUPPLY The HVPS is located in the interior of the sensor module and is plugged into the PMT tube (Section 10.4.2). It requires 2 voltage inputs. The first is +15 V, which powers the supply. The second is the programming voltage which is generated on the preamplifier board. Adjustment of the HVPS is covered in the factory calibration procedure in Section 11.6.5. This power supply has 10 independent power supply steps, one to each pin of the PMT. The following test procedure below allows you to test each step. Turn off the instrument. Remove the cover and disconnect the 2 connectors at the front of the NOX sensor module. Remove the end cap from the sensor (4 screws). Remove the HVPS/PMT assembly from the cold block inside the sensor (2 plastic screws). Re-connect the 7 pin connector to the sensor end cap, and power-up the instrument. Scroll the front panel display to the HVPS test parameter. Divide the displayed HVPS voltage by 10 and test the pairs of connector points as shown in Table 11-11. Check the overall voltage (should be equal to the HVPS value displayed on the front panel, for example 700 V) and the voltages between each pair of pins of the supply (should be 1/10th of the overall voltage, in this example 70 V): Table 11-11: Example of HVPS Power Supply Outputs If HVPS reading = 700 VDC PIN PAIR NOMINAL READING 12 70 VDC 23 70 VDC 34 70 VDC 45 70 VDC 56 70 VDC 67 70 VDC 78 70 VDC 6 7 5 8 4 3 9 2 10 11 1 KEY Turn off the instrument power, and reconnect the PMT, then reassemble the sensor. If any faults are found in the test, you must obtain a new HVPS as there are no user serviceable parts inside the supply. 263 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.5.15. PNEUMATIC SENSOR ASSEMBLY The pressure/flow sensor circuit board, located behind the sensor assembly, can be checked with a voltmeter using the following procedure, which assumes that the wiring is intact and that the motherboard and the power supplies are operating properly. Refer to Figure 11- for trouble-shooting. Measure the voltage across TP1 and TP2, it should be 10.0 0.25 V. If not, the board is faulty. Measure the voltage across the leads of capacitor C2. It should be 5.0 ± 0.25 V, if not, the board may be faulty. 11.5.15.1. Reaction Cell Pressure Measure the voltage across test points TP1 and TP5. With the sample pump disconnected or turned off, the voltage should be 4500 250 mV. With the pump running, it should be 800-1700 mV depending on the performance of the vacuum pump. The lower the reaction cell pressure, the lower the resulting voltage is. If this voltage is significantly different, the pressure transducer S1 or the board may be faulty. If this voltage is between 2 and 5 V, the pump may not be performing well, check that the reaction cell pressure is less than 10 in-Hg-A (at sea level). Ensure that the tubing is connected to the upper port, which is closer to the sensor’s contacts; the lower port does not measure pressure. 11.5.15.2. Sample Pressure Measure the voltage across test points TP1 and TP4. With the sample pump disconnected or turned off, this voltage should be 4500 250 mV. With the pump running, it should be about 0.2 V less as the sample pressure drops by about 1 in-Hg-A below ambient pressure. If this voltage is significantly different, the pressure transducer S2 or the board may be faulty. A leak in the sample system to vacuum may also cause this voltage to be between about 0.6 and 4.5. Make sure that the front panel reading of the sample pressure is at about 1 inHg-A less than ambient pressure. Ensure that the tubing is connected to the upper port, which is closer to the sensor’s contacts; the lower port does not measure pressure. Figure 11-20: Pressure / Flow Sensor Assembly 264 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.5.15.3. Ozone Flow Measure the voltage across TP1 and TP3. With proper ozone flow (250 cm3/min), this should be approximately 3.0 ± 0.3 V (this voltage will vary with altitude). With flow stopped (pump turned off), the voltage should be approximately 0 V. If the voltage is incorrect, the flow sensor or the board may be faulty. A cross-leak to vacuum inside the Perma Pure dryer may also cause this flow to increase significantly, and the voltage will increase accordingly. Also, make sure that the gas flows from P1 to P2 as labeled on the flow sensor (“high” pressure P1 to “low” pressure P2 or “Port” 1 to “Port” 2). 11.5.16. NO2 CONVERTER The NO2 converter assembly can fail in two ways, an electrical failure of the band heater and/or the thermocouple control circuit and a performance failure of the converter itself. 1) NO2 converter heater failures can be divided into two possible problems: Temperature is reported properly but heater does not heat to full temperature. In this case, the heater is either disconnected or broken or the power relay is broken. Disconnect the heater cable coming from the relay board and measure the resistance between any two of the three heater leads with a multi-meter. The resistance between A and B should be about 1000 Ω and that between A and C should be the same as between B and C, about 500 Ω each. If any of these resistances is near zero or without continuity, the heater is broken. Temperature reports zero or overload (near 500° C). This indicates a disconnected or failing thermocouple or a failure of the thermocouple circuit. First, check that the thermocouple is connected properly and the wire does not show signs of a broken or kinked pathway. If it appears to be properly connected, disconnect the yellow thermocouple plug (marked K) from the relay board and measure the voltage (not resistance) between the two leads with a multi-meter capable of measuring in the low mV range. The voltage should be about 12 mV (ignore the sign) at 315° C and about 0 mV at room temperature. Measure the continuity with an Ohm-meter. It should read close to zero Ω. If the thermocouple does not have continuity, it is broken. If it reads zero voltage at elevated temperatures, it is broken. To test the thermocouple at room temperature, heat up the converter can (e.g., with a heat gun) and see if the voltage across the thermocouple leads changes. If the thermocouple is working properly, the electronic circuit is broken. In both cases, consult the factory. 2) If the converter appears to have performance problems (conversion efficiency is outside of allowed range of 96-102%), check the following: Conversion efficiency setting in the CAL menu. If this value is different from 1.000, this correction needs to be considered. Section 7.1.5 describes this parameter in detail. Accuracy of NO2 source (GPT or gas tank standard). NO2 gas standards are typically certified to only ±2% and often change in concentrations over time. You should get the standard re-certified every year. If you use GPT, check the accuracy of the ozone source. Age of the converter. The NO2 converter has a limited operating life and may need to be replaced every ~3 years or when necessary (e.g., earlier if used with continuously high NO2 concentrations). We estimate a lifetime of about 10000 ppm-hours (a cumulative product of the NO2 concentration times the exposure time to that concentration). However, this lifetime heavily depends on many factors such as absolute concentration (temporary or permanent poisoning of the converter is possible), sample flow rate and pressure inside the converter, converter temperature, duty cycle etc. This lifetime is only an estimated reference and not a guaranteed lifetime. 265 04521C (DCN5731) Troubleshooting & Repair Teledyne API - Model 200EH/EM Operation Manual In some cases with excessive sample moisture, the oxidized molybdenum metal chips inside the converter cartridge may bake together over time and restrict air flow through the converter, in which case it needs to be replaced. To avoid this problem, we recommend the use of a sample gas conditioner (Section 5.10). Section 9.3.6 describes how to replace the NO2 converter cartridge. With no NO2 in the sample gas and a properly calibrated analyzer, the NO reading is negative, while the NO2 reading remains around zero. The converter destroys NO and needs to be replaced. With no NO2 in the sample gas and a properly calibrated analyzer, the NOX reading is significantly higher than the actual (gas standard) NO concentration. The converter produces NO2 and needs to be replaced. 11.5.17. O3 GENERATOR The ozone generator can fail in two ways, electronically (printed circuit board) and functionally (internal generator components). Assuming that air is supplied properly to the generator, the generator should automatically turn on 30 minutes after the instrument is powered up or if the instrument is still warm. See Section 10.3.6 for ozone generator functionality. Accurate performance of the generator can only be determined with an ozone analyzer connected to the outlet of the generator. However, if the generator appears to be working properly but the sensitivity or calibration of the instrument is reduced, suspect a leak in the ozone generator supply air. A leak in the dryer or between the dryer and the generator can cause moist, ambient air to leak into the air stream, which significantly reduces the ozone output. The generator will produce only about half of the nominal O3 concentration when run with moist, ambient air instead of dried air. In addition, moist supply air will produce large amounts of nitric acid in the generator, which can cause analyzer components downstream of the generator to deteriorate and/or causes significant deposit of nitrate deposits on the reaction cell window, reducing sensitivity and causing performance drift. Carry out a leak check as described earlier in this chapter. 11.5.18. BOX TEMPERATURE The box temperature sensor (thermistor) is mounted on the motherboard below the bottom edge of the CPU board when looking at it from the front. It cannot be disconnected to check its resistance. Box temperature will vary with, but will usually read about 5° C higher than, ambient (room) temperature because of the internal heating zones from the NO2 converter, reaction cell and other devices. To check the box temperature functionality, we recommend to check the BOX_TEMP signal voltage using the SIGNAL I/O function under the DIAG Menu (Section 6.13.1). At about 30° C, the signal should be around 1500 mV. We recommend to use a certified or calibrated external thermometer / temperature sensor to verify the accuracy of the box temperature by placing it inside the chassis, next to the thermistor labeled XT1 (above connector J108) on the motherboard. 11.5.19. PMT TEMPERATURE PMT temperature should be low and constant. It is more important that this temperature is maintained constant than it is to maintain it low. The PMT cooler uses a Peltier, thermo-electric cooler element supplied with 12 V DC power from the switching power supply PS2. The temperature is controlled by a proportional temperature controller located on the preamplifier board. Voltages applied to the cooler element vary from 0.1 to 12 VDC. The temperature set point (hard-wired into the preamplifier board) will vary by ±1C due to component tolerances. The actual temperature will be maintained to within 0.1° C around that set point. On power-up of the analyzer, the front panel enables the user to watch that temperature drop from about ambient temperature down to its set point of 6-8° C. If the temperature fails to adjust after 30 minutes, there is a problem in the cooler circuit. If the control circuit on the preamplifier board is faulty, a temperature of –1° C is reported. 266 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.6. REPAIR PROCEDURES This section contains some procedures that may need to be performed when a major component of the analyzer requires repair or replacement. Note that maintenance procedures (e.g., replacement of regularly changed expendables) are discussed in Chapter 8 (Maintenance) are not listed here. Also note that Teledyne-API customer service may have a more detailed service note for some of the below procedures. Contact customer service. 11.6.1. DISK-ON-CHIP REPLACEMENT Replacing the Disk-on-Chip (DOC) will cause all of the instrument configuration parameters to be lost unless the replacement chip carries the exact same firmware version. iDAS data will always be lost and, if possible, should be downloaded prior to changing the DOC. If the analyzer is equipped with at least one EEPROM flash chip (standard configuration), the configuration settings are stored on the EEPROM. It is recommended to document all analyzer parameters that may have been changed, such as calibration, range, auto-cal, analog output, serial port and other settings before replacing the CPU chip. Refer to Figure 10-10-17 for locating the DOC and other CPU components. 1. Ground yourself to prevent electrostatic damage to electronic components. 2. Turn off power to the instrument. 3. Fold down the rear panel by loosening the mounting screws. You may have to lift up the analyzer cover to prevent some connectors on the CPU board to brush against the cover. 4. Locate the Disk-on-Chip on the CPU board. The chip should carry a label with analyzer model number (M200EH/EM), firmware revision (example: M200EH/EM_C7.EXE), date and initials of the programmer. 5. Remove the IC with a dedicated IC removal tool or by gently prying it up from the socket. Do not bend the connector pins. 6. Reinstall the new Disk-on-Chip, making sure the notch at the end of the chip matches the notch in the socket. It may be necessary to straighten the pins somewhat to fit them into the socket. Gently but firmly press the chip all the way in. Do not bend the pins. 7. Close the rear panel, replace the cover and turn on power to the machine. Generally, all of the setup information will need to be re-entered, including analog input and output calibration unless the firmware revision has not changed and the analyzer is equipped and properly configured with an EEPROM chip. Note especially that the A/D converter must be re-calibrated, and all information collected in step 1 above must be re-entered before the instrument will function correctly. The analyzer typically issues an ANALOG CALIBRATION WARNING if the analog circuitry was not calibrated within 10 minutes after restart. 267 04521C (DCN5731) Troubleshooting & Repair Teledyne API - Model 200EH/EM Operation Manual 11.6.2. FLASH CHIP REPLACEMENT OR UPGRADE The M200EH/EM CPU board can accommodate up to two EEPROM flash chips. The standard configuration is one chip with 64 kb of storage capacity, which is used to store the analyzer configuration as created during final checkout at the factory. Replacing this chip will erase that configuration, which will be recreated with a new copy when restarting the analyzer. However, if the firmware and/or the DOC is changed at the same time, all analyzer configuration settings and iDAS data will be lost. Adding a second EEPROM chip to the existing chip will double memory but this procedure will require a BIOS configuration and is not a standard sales option. Also make sure that you receive a fully formatted EEPROM chip for replacement. Contact the factory for details. 1. Ground yourself to prevent electrostatic damage to electronic components. 2. Turn off power to the instrument, fold down the rear panel by loosening the mounting screws. If necessary, lift the cover to prevent the rear panel connectors from brushing against it. 3. Locate the EEPROM chip in the left-most socket of the CPU board. The chip is almost square with one corner cut off, the socket is shaped accordingly and the chip is recessed into the socket. 4. Remove the old chip by using a special tool or gently pry the chip out using a very fine screwdriver. Make sure not to bend or destroy any of the contacts of the socket. When upgrading the CPU with a second chip, no removal is necessary as the second socket should be empty. 5. Reinstall the new or additional EEPROM chip, making sure the cut-off edge matches that of the socket. Press the chip symmetrically and straight all the way in. 6. Close the rear panel and cover and turn on power to the machine. If a front panel message Flash Format INVALID appears on start-up, the EEPROM was not properly formatted. Contact the factory for a proper replacement. 11.6.3. O3 GENERATOR REPLACEMENT The ozone generator is a black, brick-shaped device with printed circuit board attached to its rear and two tubes extending out the right side in the front of the analyzer. To replace the ozone generator: 1. Turn off the analyzer power, remove the power cord and the analyzer cover. 2. Disconnect the 1/8” black tube from the ozone scrubber cartridge and the ¼” clear tube from the plastic extension tube at the brass fitting nearest to the ozone generator. 3. Unplug the electrical connection on the rear side of the brick. 4. Unscrew the two mounting screws that attach the ozone generator to the chassis and take out the entire assembly. 5. If you received a complete replacement generator with circuit board and mounting bracket attached, simply reverse the above steps to replace the current generator. 6. Make sure to carry out a leak check and a recalibration after the analyzer warmed up for about 30 minutes. 268 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.6.4. SAMPLE AND OZONE DRYER REPLACEMENT The M200EH/EM standard configuration is equipped with a dryer for the ozone supply air. An optional dryer is available for the sample stream and a combined dryer for both gas streams can also be purchased. To change one or all of these options: 1. Turn off power to the analyzer and pump, remove the power cord and the analyzer cover. 2. Locate the dryers in the center of the instrument, between sensor and NO2 converter. They are mounted to a bracket, which can be taken out when unscrewing the two mounting screws (if necessary). 3. Disconnect all tubing that extends out of the dryer assembly, These are usually the purge tube connecting to the vacuum manifold, the tube from the exit to the ozone flow meter (ozone dryer) or to the NO/NOx valve (sample dryer) or two tubes to the ozone flow meter and the NO/NOX valve (combo-dryer). Take extra care not to twist any of the white plastic fittings on the dryer, which connect the inner drying tube to the outer purge tube. See Section 9.3.2 and Figure 9-2. 4. Note the orientation of the dryer on the bracket. 5. Cut the tie wraps that hold the dryer to the mounting bracket and take out the old dryer. If necessary, unscrew the two mounting screws on the bracket and take out the entire assembly. 6. Attach the replacement dryer to the mounting bracket in the same orientation as the old dryer. 7. Fix the dryer to the bracket using new tie wraps. 8. Cut off excess length of the wraps. 9. Put the assembly back into the chassis and tighten the mounting screws. 10. Re-attach the tubes to vacuum manifold, flow meter and/or NO/NOx valve using at least two wrenches. Ttake extra care not to twist the dryer’s white plastic fittings, as this will result in large leaks that are difficult to trouble-shoot and fix. 11. Carry out a detailed leak check (Section 0), 12. Close the analyzer. 13. Power up pump and analyzer and re-calibrate the instrument after it stabilizes 269 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.6.5. PMT SENSOR HARDWARE CALIBRATION The sensor module hardware calibration is used in the factory to adjust the slope and offset of the PMT output and to optimize the signal output and HVPS. If the instrument’s slope and offset values are outside of the acceptable range and all other more obvious causes for this problem have been eliminated, the hardware calibration can be used to adjust the sensor as has been done in the factory. This procedure is also recommended after replacing the PMT or the preamplifier board. 1. Perform a full zero calibration using zero air (Section 7.2, 7.4, or 7.6). 2. Locate the preamplifier board (Figure 3-1). 3. Locate the following components on the preamplifier board (Figure 11-): HVPS coarse adjustment switch (Range 0-9, then A-F). HVPS fine adjustment switch (Range 0-9, then A-F). Gain adjustment potentiometer (Full scale is 10 turns). 4. Turn the gain adjustment potentiometer 12 turns clockwise to its maximum setting. 5. Feed NO to the analyzer: For the M200EH use 450 ppm NO. For the M200Em use 18 ppm NO. 6. Wait until the STABIL value is below 0.5 ppm 7. Scroll to the NORM PMT value on the analyzer’s front panel. 8. With the NO gas concentrations mentioned instep 5 above, the NORM PMT value should be 3600 mV. 9. Set the HVPS coarse adjustment to its minimum setting (0). Set the HVPS fine adjustment switch to its maximum setting (F). 10. Set the HVPS coarse adjustment switch to the lowest setting that will give you just above 3600 mV NORM PMT signal. The coarse adjustment typically increments the NORM PMT signal in 100-300 mV steps. Figure 11-22: Pre-Amplifier Board Layout 270 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11. Adjust the HVPS fine adjustment such that the NORM PMT value is 3600-3700 mV. The fine adjustment typically increments the NORM PMT value by about 30 mV. It may be necessary to go back and forth between coarse and fine adjustments if the proper value is at the threshold of the min/max coarse setting. NOTE Do not overload the PMT by accidentally setting both adjustment switches to their maximum setting. Start at the lowest setting and increment slowly. Wait 10 seconds between adjustments. 12. If the NORM PMT value set above is now between 3560-3640 mV, skip this step. Otherwise, adjust the NORM PMT value with the gain potentiometer down to 3600±10 mV. his is the final very-fine adjustment. 13. Note that during adjustments, the NORM PMT value may be fluctuating, as the analyzer continues to switch between NO and NOX streams as well as between measure and AutoZero modes. You may have to mentally average the values of NO and NOX response for this adjustment. 14. Perform a software span calibration (Section 7.2, 7.4, or 7.6) to normalize the sensor response to its new PMT sensitivity. 15. Review the slope and offset values, the slopes should be 1.000±0.300 and the offset values should be 0.0±20 mV (-20 to +150 mV is allowed). 11.6.6. REPLACING THE PMT, HVPS OR TEC The photo multiplier tube (PMT) should last for the lifetime of the analyzer. However, in some cases, the high voltage power supply (HVPS) or the thermo-electric cooler (TEC) may fail. In case of PMT, HVPS or TEC failure, the sensor assembly needs to be opened in order to change one of these components. Refer to Figure 11- for the structure of the 200EH/EM sensor assembly and follow the steps below for replacement of one of its components. We recommend to ensure that the PMT, HVPS or TEC modules are, indeed, faulty to prevent unnecessary opening of the sensor. NOTE Whereas it is possible for a skilled technician to change the PMT or HVPS through the front panel with the sensor assembly mounted to the analyzer, we recommend to remove the entire assembly and carry this procedure out on a clean, anti-static table with the user wearing an anti-static wrist strap to prevent static discharge damage to the assembly or its circuits. 1. Power down the analyzer, disconnect the power cord. 2. Remove the cover and disconnect all pneumatic and electrical connections from the sensor assembly. 271 04521C (DCN5731) Troubleshooting & Repair Teledyne API - Model 200EH/EM Operation Manual 3. If the TEC is to be replaced, remove the reaction cell assembly at this point by unscrewing two holding screws. This is necessary only if the PMT cold block is to be removed. This step is not necessary if the HVPS or the PMT only are exchanged. Figure 11-22: M200EH/EM Sensor Assembly 4. Remove the two connectors on the PMT housing end plate facing towards the front panel. 5. Remove the end plate itself (4 screws with plastic washers). 6. Remove the dryer packages inside the PMT housing. 7. Along with the plate, slide out the OPTIC TEST LED and the thermistor that measures the PMT temperature. Both may be coated with a white, thermal conducting paste. Do not contaminate the inside of the housing with this grease, as it may contaminate the PMT glass tube on re-assembly. 8. Unscrew the PMT assembly, which is held to the cold block by two plastic screws. 9. Discard the plastic screws and replace with new screws at the end of this procedure (the threads get stripped easily and it is recommended to use new screws). 10. Carefully take out the assembly consisting of the HVPS, the gasket and the PMT. 11. Change the PMT or the HVPS or both, clean the PMT glass tube with a clean, anti-static wipe and do not touch it after cleaning. 272 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 12. If the cold block or TEC is to be changed: a) Disconnect the TEC driver board from the preamplifier board, remove the cooler fan duct (4 screws on its side) including the driver board. b) Disconnect the driver board from the TEC and set the sub-assembly aside. 13. Remove the end plate with the cooling fins (4 screws) and slide out the PMT cold block assembly, which contains the TEC. 14. Unscrew the TEC from the cooling fins and the cold block and replace it with a new unit. 15. Re-assemble this TEC subassembly in reverse order. Make sure to use thermal grease between TEC and cooling fins as well as between TEC and cold block and that the side opening in the cold block will face the reaction cell when assembled. Evenly tighten the long mounting screws for good thermal conductivity. CAUTION The thermo-electric cooler needs to be mounted flat to the heat sink. If there is any significant gap, the TEC might burn out. Make sure to apply heat sink paste before mounting it and tighten the screws evenly and cross-wise. 16. Re-insert the TEC subassembly in reverse order. Make sure that the O-ring is placed properly and the assembly is tightened evenly. 17. Re-insert the PMT/HVPS subassembly in reverse order and don’t forget the gasket between HVPS and PMT. Use new plastic screws to mount the PMT assembly on the PMT cold block. 18. Insert the LED and thermistor into the cold block, insert new drying packages and carefully replace the end plate by making sure that the O-ring is properly in place. Improperly placed O-rings will cause leaks, which – in turn – cause moisture to condense on the inside of the cooler and likely cause a short in the HVPS. 19. Reconnect the cables and the reaction cell (evenly tighten these screws). 20. Replace the sensor assembly into the chassis and fasten with four screws and washers. 21. Reconnect all electrical and pneumatic connections. 22. leak check the system. 23. Power up the analyzer. 24. Verify the basic operation of the analyzer using the ETEST and OTEST features or zero and span gases, then carry out a hardware calibration of the analyzer (Section 11.6.5) followed by a software calibration. 273 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.7. REMOVING / REPLACING THE RELAY PCA FROM THE INSTRUMENT This is the most commonly used version of the Relay PCA. It includes a bank of solid state AC relays. This version is installed in analyzers where components such as AC powered heaters must be turned ON & OFF. A retainer plate is installed over the relay to keep them securely seated in their sockets. Retainer Mounting Screws AC Relay Retainer Plate Figure 11-23: Relay PCA with AC Relay Retainer In Place The Relay retainer plate installed on the relay PCA covers the lower right mounting screw of the relay PCA. Therefore, when removing the relay PCA, the retainer plate must be removed first. Mounting Screws AC Relay Retain Occludes Mounting Screw on P/N 045230200 Figure 11-24: Relay PCA Mounting Screw Locations 274 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual Troubleshooting & Repair 11.8. TECHNICAL ASSISTANCE If this manual and its trouble-shooting / repair sections do not solve your problems, technical assistance may be obtained from Teledyne-API, Customer Service, 9480 Carroll Park Drive, San Diego, CA 92121. Phone: +1 858 657 9800 or 1-800 324 5190. Fax: +1 858 657 9816. Email: api-customerservice@teledyne.com. Before you contact customer service, fill out the problem report form in Appendix C, which is also available online for electronic submission at http://www.teledyne-api.com/forms/. USER NOTES: 275 04521C (DCN5731) 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual A Primer on Electro-Static Discharge 12. A PRIMER ON ELECTRO-STATIC DISCHARGE Teledyne Instruments considers the prevention of damage caused by the discharge of static electricity to be extremely important part of making sure that your analyzer continues to provide reliable service for a long time. This section describes how static electricity occurs, why it is so dangerous to electronic components and assemblies as well as how to prevent that damage from occurring. 12.1. HOW STATIC CHARGES ARE CREATED Modern electronic devices such as the types used in the various electronic assemblies of your analyzer, are very small, require very little power and operate very quickly. Unfortunately, the same characteristics that allow them to do these things also make them very susceptible to damage from the discharge of static electricity. Controlling electrostatic discharge begins with understanding how electro-static charges occur in the first place. Static electricity is the result of something called triboelectric charging which happens whenever the atoms of the surface layers of two materials rub against each other. As the atoms of the two surfaces move together and separate, some electrons from one surface are retained by the other. Materials Makes Contact + Materials Separate + + + PROTONS = 3 ELECTRONS = 3 PROTONS = 3 ELECTRONS = 3 NET CHARGE = 0 NET CHARGE = 0 PROTONS = 3 ELECTRONS = 2 PROTONS = 3 ELECTRONS = 4 NET CHARGE = -1 NET CHARGE = +1 Figure 12-1: Triboelectric Charging If one of the surfaces is a poor conductor or even a good conductor that is not grounded, the resulting positive or negative charge cannot bleed off and becomes trapped in place, or static. The most common example of triboelectric charging happens when someone wearing leather or rubber soled shoes walks across a nylon carpet or linoleum tiled floor. With each step, electrons change places and the resulting electro-static charge builds up, quickly reaching significant levels. Pushing an epoxy printed circuit board across a workbench, using a plastic handled screwdriver or even the constant jostling of StyrofoamTM pellets during shipment can also build hefty static charges Table 12-1: Static Generation Voltages for Typical Activities MEANS OF GENERATION 65-90% RH 10-25% RH 1,500V 35,000V Walking across vinyl tile 250V 12,000V Worker at bench 100V 6,000V Poly bag picked up from bench 1,200V 20,000V Moving around in a chair padded with urethane foam 1,500V 18,000V Walking across nylon carpet 277 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual A Primer on Electro-Static Discharge 12.2. HOW ELECTRO-STATIC CHARGES CAUSE DAMAGE Damage to components occurs when these static charges come into contact with an electronic device. Current flows as the charge moves along the conductive circuitry of the device and the typically very high voltage levels of the charge overheat the delicate traces of the integrated circuits, melting them or even vaporizing parts of them. When examined by microscope the damage caused by electro-static discharge looks a lot like tiny bomb craters littered across the landscape of the component’s circuitry. A quick comparison of the values in Table 12-1 with the those shown in the Table 12-2, listing device susceptibility levels, shows why Semiconductor Reliability News estimates that approximately 60% of device failures are the result of damage due to electro-static discharge. Table 12-2: Sensitivity of Electronic Devices to Damage by ESD DEVICE DAMAGE SUSCEPTIBILITY VOLTAGE RANGE DAMAGE BEGINS OCCURRING AT CATASTROPHIC DAMAGE AT MOSFET 10 100 VMOS 30 1800 NMOS 60 100 GaAsFET 60 2000 EPROM 100 100 JFET 140 7000 SAW 150 500 Op-AMP 190 2500 CMOS 200 3000 Schottky Diodes 300 2500 Film Resistors 300 3000 This Film Resistors 300 7000 ECL 500 500 SCR 500 1000 Schottky TTL 500 2500 Potentially damaging electro-static discharges can occur: Any time a charged surface (including the human body) discharges to a device. Even simple contact of a finger to the leads of a sensitive device or assembly can allow enough discharge to cause damage. A similar discharge can occur from a charged conductive object, such as a metallic tool or fixture. When static charges accumulated on a sensitive device discharges from the device to another surface such as packaging materials, work surfaces, machine surfaces or other device. In some cases, charged device discharges can be the most destructive. A typical example of this is the simple act of installing an electronic assembly into the connector or wiring harness of the equipment in which it is to function. If the assembly is carrying a static charge, as it is connected to ground a discharge will occur. Whenever a sensitive device is moved into the field of an existing electro-static field, a charge may be induced on the device in effect discharging the field onto the device. If the device is then momentarily grounded while within the electrostatic field or removed from the region of the electrostatic field and grounded somewhere else, a second discharge will occur as the charge is transferred from the device to ground. 278 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual A Primer on Electro-Static Discharge 12.3. COMMON MYTHS ABOUT ESD DAMAGE I didn’t feel a shock so there was no electro-static discharge: The human nervous system isn’t able to feel a static discharge of less than 3500 volts. Most devices are damaged by discharge levels much lower than that. I didn’t touch it so there was no electro-static discharge: Electro-static charges are fields whose lines of force can extend several inches or sometimes even feet away from the surface bearing the charge. It still works so there was no damage: Sometimes the damaged caused by electro-static discharge can completely sever a circuit trace causing the device to fail immediately. More likely, the trace will be only partially occluded by the damage causing degraded performance of the device or worse, weakening the trace. This weakened circuit may seem to function fine for a short time, but even the very low voltage and current levels of the device’s normal operating levels will eat away at the defect over time causing the device to fail well before its designed lifetime is reached. These latent failures are often the most costly since the failure of the equipment in which the damaged device is installed causes down time, lost data, lost productivity, as well as possible failure and damage to other pieces of equipment or property. Static Charges can’t build up on a conductive surface: There are two errors in this statement. Conductive devices can build static charges if they are not grounded. The charge will be equalized across the entire device, but without access to earth ground, they are still trapped and can still build to high enough levels to cause damage when they are discharged. A charge can be induced onto the conductive surface and/or discharge triggered in the presence of a charged field such as a large static charge clinging to the surface of a nylon jacket of someone walking up to a workbench. As long as my analyzer is properly installed, it is safe from damage caused by static discharges: It is true that when properly installed the chassis ground of your analyzer is tied to earth ground and its electronic components are prevented from building static electric charges themselves. This does not prevent discharges from static fields built up on other things, like you and your clothing, from discharging through the instrument and damaging it. 12.4. BASIC PRINCIPLES OF STATIC CONTROL It is impossible to stop the creation of instantaneous static electric charges. It is not, however difficult to prevent those charges from building to dangerous levels or prevent damage due to electro-static discharge from occurring. 279 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual A Primer on Electro-Static Discharge 12.4.1. General Rules Only handle or work on all electronic assemblies at a properly set up ESD station. Setting up an ESD safe workstation need not be complicated. A protective mat properly tied to ground and a wrist strap are all that is needed to create a basic anti-ESD workstation (see figure 12-2). P r o t e c t iv e M a t W r is t S t r a p G r o u n d P o in t Figure 12-2: Basic anti-ESD Work Station For technicians that work in the field, special lightweight and portable anti-ESD kits are available from most suppliers of ESD protection gear. These include everything needed to create a temporary anti-ESD work area anywhere. Always wear an Anti-ESD wrist strap when working on the electronic assemblies of your analyzer. An anti-ESD wrist strap keeps the person wearing it at or near the same potential as other grounded objects in the work area and allows static charges to dissipate before they can build to dangerous levels. Anti-ESD wrist straps terminated with alligator clips are available for use in work areas where there is no available grounded plug. Also, anti-ESD wrist straps include a current limiting resistor (usually around one meg-ohm) that protects you should you accidentally short yourself to the instrument’s power supply. Simply touching a grounded piece of metal is insufficient. While this may temporarily bleed off static charges present at the time, once you stop touching the grounded metal new static charges will immediately begin to re-build. In some conditions, a charge large enough to damage a component can rebuild in just a few seconds. Always store sensitive components and assemblies in anti-ESD storage bags or bins: Even when you are not working on them, store all devices and assemblies in a closed anti-Static bag or bin. This will prevent induced charges from building up on the device or assembly and nearby static fields from discharging through it. Use metallic anti-ESD bags for storing and shipping ESD sensitive components and assemblies rather than pink-poly bags. The famous, “pink-poly” bags are made of a plastic that is impregnated with a liquid (similar to liquid laundry detergent) which very slowly sweats onto the surface of the plastic creating a slightly conductive layer over the surface of the bag. While this layer may equalizes any charges that occur across the whole bag, it does not prevent the build up of static charges. If laying on a conductive, grounded surface, these bags will allow charges to bleed away but the very charges that build up on the surface of the bag itself can be transferred through the bag by induction onto the circuits of your ESD sensitive device. Also, the liquid impregnating the plastic is eventually used up after which the bag is as useless for preventing damage from ESD as any ordinary plastic bag. Anti-Static bags made of plastic impregnated with metal (usually silvery in color) provide all of the charge equalizing abilities of the pink-poly bags but also, when properly sealed, create a Faraday cage that completely isolates the contents from discharges and the inductive transfer of static charges. 280 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual A Primer on Electro-Static Discharge Storage bins made of plastic impregnated with carbon (usually black in color) are also excellent at dissipating static charges and isolating their contents from field effects and discharges. Never use ordinary plastic adhesive tape near an ESD sensitive device or to close an anti-ESD bag. The act of pulling a piece of standard plastic adhesive tape, such as Scotch® tape, from its roll will generate a static charge of several thousand or even tens of thousands of volts on the tape itself and an associated field effect that can discharge through or be induced upon items up to a foot away. 12.4.2. Basic anti-ESD Procedures for Analyzer Repair and Maintenance 12.4.2.1. Working at the Instrument Rack When working on the analyzer while it is in the instrument rack and plugged into a properly grounded power supply. 1. Attach your anti-ESD wrist strap to ground before doing anything else. Use a wrist strap terminated with an alligator clip and attach it to a bare metal portion of the instrument chassis. This will safely connect you to the same ground level to which the instrument and all of its components are connected. 2. Pause for a second or two to allow any static charges to bleed away. 3. Open the casing of the analyzer and begin work. Up to this point, the closed metal casing of your analyzer has isolated the components and assemblies inside from any conducted or induced static charges. 4. If you must remove a component from the instrument, do not lay it down on a non-ESD preventative surface where static charges may lie in wait. 5. Only disconnect your wrist strap after you have finished work and closed the case of the analyzer. 12.4.2.2. Working at an Anti-ESD Work Bench. When working on an instrument of an electronic assembly while it is resting on an anti-ESD work bench: 1. Plug your anti-ESD wrist strap into the grounded receptacle of the work station before touching any items on the work station and while standing at least a foot or so away. This will allow any charges you are carrying to bleed away through the ground connection of the workstation and prevent discharges due to field effects and induction from occurring. 2. Pause for a second or two to allow any static charges to bleed away. 3. Only open any anti-ESD storage bins or bags containing sensitive devices or assemblies after you have plugged your wrist strap into the workstation. Lay the bag or bin on the workbench surface. Before opening the container, wait several seconds for any static charges on the outside surface of the container to be bled away by the workstation’s grounded protective mat. 4. Do not pick up tools that may be carrying static charges while also touching or holding an ESD Sensitive Device. Only lay tools or ESD-sensitive devices and assemblies on the conductive surface of your workstation. Never lay them down on any non-ESD preventative surface. 5. Place any static sensitive devices or assemblies in anti-static storage bags or bins and close the bag or bin before unplugging your wrist strap. 281 04521C (DCN5731) A Primer on Electro-Static Discharge Teledyne API - Model 200EH/EM Operation Manual 6. Disconnecting your wrist strap is always the last action taken before leaving the workbench. 12.4.2.3. Transferring Components from Rack to Bench and Back When transferring a sensitive device from an installed Teledyne Instruments analyzer to an Anti-ESD workbench or back: 1. Follow the instructions listed above for working at the instrument rack and workstation. 2. Never carry the component or assembly without placing it in an anti-ESD bag or bin. 3. Before using the bag or container allow any surface charges on it to dissipate: If you are at the instrument rack, hold the bag in one hand while your wrist strap is connected to a ground point. If you are at an anti-ESD workbench, lay the container down on the conductive work surface. In either case wait several seconds. 4. Place the item in the container. 5. Seal the container. If using a bag, fold the end over and fastening it with anti-ESD tape. Folding the open end over isolates the component(s) inside from the effects of static fields. Leaving the bag open or simply stapling it shut without folding it closed prevents the bag from forming a complete protective envelope around the device. 6. Once you have arrived at your destination, allow any surface charges that may have built up on the bag or bin during travel to dissipate: Connect your wrist strap to ground. If you are at the instrument rack, hold the bag in one hand while your wrist strap is connected to a ground point. If you are at a anti-ESD work bench, lay the container down on the conductive work surface In either case wait several seconds 7. Open the container. 12.4.2.4. Opening Shipments from Teledyne Instruments Customer Service. Packing materials such as bubble pack and Styrofoam pellets are extremely efficient generators of static electric charges. To prevent damage from ESD, Teledyne Instruments ships all electronic components and assemblies in properly sealed anti-ESD containers. Static charges will build up on the outer surface of the anti-ESD container during shipping as the packing materials vibrate and rub against each other. To prevent these static charges from damaging the components or assemblies being shipped make sure that you always unpack shipments from Teledyne Instruments Customer Service by: 1. Opening the outer shipping box away from the anti-ESD work area. 2. Carry the still sealed ant-ESD bag, tube or bin to the anti-ESD work area. 3. Follow steps 6 and 7 of Section 12.4.2.3 above when opening the anti-ESD container at the work station. 282 04521C (DCN5731) Teledyne API - Model 200EH/EM Operation Manual A Primer on Electro-Static Discharge 4. Reserve the anti-ESD container or bag to use when packing electronic components or assemblies to be returned to Teledyne Instruments. 12.4.2.5. Packing Components for Return to Teledyne Instruments Customer Service. Always pack electronic components and assemblies to be sent to Teledyne Instruments Customer Service in antiESD bins, tubes or bags. WARNING DO NOT use pink-poly bags. NEVER allow any standard plastic packaging materials to touch the electronic component/assembly directly This includes, but is not limited to, plastic bubble-pack, Styrofoam peanuts, open cell foam, closed cell foam, and adhesive tape DO NOT use standard adhesive tape as a sealer. Use ONLY anti-ESD tape 1. Never carry the component or assembly without placing it in an anti-ESD bag or bin. 2. Before using the bag or container allow any surface charges on it to dissipate: If you are at the instrument rack, hold the bag in one hand while your wrist strap is connected to a ground point. If you are at an anti-ESD workbench, lay the container down on the conductive work surface. In either case wait several seconds. 3. Place the item in the container. 4. Seal the container. If using a bag, fold the end over and fastening it with anti-ESD tape. Folding the open end over isolates the component(s) inside from the effects of static fields. Leaving the bag open or simply stapling it shut without folding it closed prevents the bag from forming a complete protective envelope around the device. NOTE If you do not already have an adequate supply of anti-ESD bags or containers available, Teledyne Instruments’ Customer Service department will supply them. Follow the instructions listed above for working at the instrument rack and workstation. User Notes: 283 04521C (DCN5731) A Primer on Electro-Static Discharge Teledyne API - Model 200EH/EM Operation Manual USER NOTES: 284 04521C (DCN5731) Addendum to the M200EM/EH Operators Manual (P/N 04521) (Ref: 06116A) ADDENDUM TO THE M200EM/EH OPERATORS MANUAL (P/N 04521) 1. PREFACE This addendum is an update to the Model M200EM/EH Operators Manual. It documents the following improvements: STAINLESS STEEL OZONE DESTRUCT P/N 051210000: The Ozone Destruct now resides down stream of the vacuum manifold. Previously, it was located downstream of the reaction cell. BYPASS MANIFOLD ASSEMBLY P/N 044430100: As a cost reduction, the bypass manifold is now incorporated into the three port reaction cell. 2. CHANGES AND UPDATES 2.1. OZONE DESTRUCT The following photograph identifies the Ozone Destruct assembly and pneumatic connections. FIGURE 1.0 - OZONE DESTRUCT ASSY P/N 05121 Addendum-1 04521C (DCN5731) (Ref: 06116A) Addendum to the M200EM/EH Operators Manual (P/N 04521) 2.2. THREE PORT REACTION CELL By incorporating the bypass flow orifice into the reaction cell, the bypass manifold assembly, which previously resided between the vacuum manifold and Molycon container, is no longer required. Ozone Flow Orifice: 7 Mil for M200EM Or EH Sample Flow: No Orifice Bypass Flow Orifice: 7 Mil for M200EM 3 Mil for M200EH FIGURE 2.0 – THREE PORT REACTION CELL ASSY P/N 06028 Addendum-2 04521C (DCN5731) Model 200EH/EM (Ref: 05147F) APPENDIX A - Version Specific Software Documentation APPENDIX A - Version Specific Software Documentation APPENDIX A-1: Model 200EH/EM Software Menu Trees APPENDIX A-2: Model 200EH/EM Setup Variables Available Via Serial I/O APPENDIX A-3: Model 200EH/EM Warnings and Test Measurements Via Serial I/O APPENDIX A-4: Model 200EH/EM Signal I/O Definitions APPENDIX A-5: Model 200EH/EM iDAS Functions APPENDIX A-6: Model 200EH/EM Terminal Command Designators A-1 04521C (DCN5731) APPENDIX A-1: M200EH/EM Software Menu Trees, Revision F.0 Model 200EH/EM (Ref: 05147F) APPENDIX A-1: M200EH/EM Software Menu Trees, Revision F.0 SAMPLE TEST 1 A1: A2: A3: A4: User User User User CALS 4 MSG 1 HIGH LOW HIGH LOW HIGH SPAN CONC 2 Selectable Range Selectable Range 2 Selectable Range 2 Selectable Range 2 NOX STB SAMP FLOW 0ZONE FLOW PMT NORM PMT AZERO HVPS RCELL TEMP BOX TEMP PMT TEMP MF TEMP O2 CELL TEMP 3 MOLY TEMP RCEL SAMP NOX SLOPE NOX OFFSET NO SLOPE NO OFFSET O2 SLOPE 3 O2 OFFSET 3 TIME ZERO CLR 1 SETUP O2 3 NOX LOW CALZ 4 SPAN CONC NOX NO NO2 ZERO Press to cycle through the active warning messages. Press to clear an active warning messages. CONV CAL CFG PRIMARY SETUP MENU SET ACAL 4 DAS RANGE PASS CLK MORE SECONDARY SETUP MENU 1 Only appears when warning messages are active. User selectable analog outputs A1 – A4 (see Section X.X.X) 3 Only appears if analyzer is equipped with O2 sensor option. 4 Only appears if analyzer is equipped with Zero/Span or IZS valve options. 2 Figure A-1: COMM VARS DIAG ALAR Basic Sample Display Menu A-2 04521C (DCN5731) APPENDIX A-1: M200EH/EM Software Menu Trees, Revision F.0 Model 200EH/EM (Ref: 05147F) SAMPLE ACAL 1 CFG PREV DAS NEXT RNGE PASS Go to iDAS Menu Tree MODE MORE ON OFF SEQ 1) SEQ 2) SEQ 3) MODEL TYPE AND NUMBER PART NUMBER SERIAL NUMBER SOFTWARE REVISION LIBRARY REVISION iCHIP SOFTWARE PREV REVISION HESSEN PROTOCOL REVISION 2 CPU TYPE & OS REVISION DATE FACTORY CONFIGURATION SAVED 1 CLK ACAL menu and its submenus only appear if analyzer is equipped with Zero/Span or IZS valve options. 2 Only appears if Dilution option is active 3 Only appears if Hessen protocol is active. 4 O 2 Modes only appear if analyzer is equipped with O2 sensor option. 5 DOES NOT appear if one of the three O2 modes is selected TIME NEXT UNIT DISABLED ZERO ZERO-LO ZERO-LO-HI ZERO-HI LO LO-HI HI O2 ZERO 4 O2 ZERO-SP 4 O2 SPAN 4 Figure A-2: PPM DIL 3 MGM DATE Go to SECONDARY SETUP Menu Tree SET ON TIMER ENABLE DURATION CALIBRATE 5 RANGE TO CAL OFF STARTING DATE STARTING TIME DELTA DAYS DELTA TIME LOW 5 HIGH 5 Primary Setup Menu (Except iDAS) A-3 04521C (DCN5731) APPENDIX A-1: M200EH/EM Software Menu Trees, Revision F.0 Model 200EH/EM (Ref: 05147F) SAMPLE SETUP CFG ACAL 1 RNGE PASS CLK MORE DAS VIEW PREV EDIT NEXT ENTER PASSWORD: 818 CONC CALDAT CALCHE HIRES DIAG PREV CONC CALDAT CALCHE HIRES DIAG VIEW PV10 PREV NEXT NX10 Selects the data point to be viewed Cycles through parameters assigned to this iDAS channel PREV INS DEL YES NEXT NX10 NAME EVENT PARAMETERS REPORT PERIOD NUMBER OF RECORDS RS-232 REPORT CHANNEL ENABLE NO CAL MODE YES 2 Cycles through list of available trigger events 3 NEXT Create/edit the name of the channel Sets the time lapse between each report ON PREV NEXT INS DEL EDIT 2 PRNT OFF YES 2 Cycles through list of currently active parameters for this channel YES NO Sets the maximum number of records recorded by this channel EDIT PRNT SAMPLE MODE PRECISION 1 2 Cycles through list of available & currently active parameters for this channel PREV NEXT Figure A-3: INST NO AVG MIN MAX 3 ACAL menu only appear if analyzer is equipped with Zero/Span or IZS valve options. Editing an existing iDAS channel will erase any data stored on the channel options. Changing the event for an existing iDAS channel DOES NOT erase the data stored on the channel. Primary Setup Menu iDAS Submenu A-4 04521C (DCN5731) APPENDIX A-1: M200EH/EM Software Menu Trees, Revision F.0 Model 200EH/EM (Ref: 05147F) SAMPLE CFG DAS RNGE PASS ACAL SETUP MORE CLK COMM VARS INET 1 HESN 2 ENTER PASSWORD: 818 Go to COMM / Hessen Menu Tree ID EDIT ENTER PASSWORD: 818 1 COM1 COM2 EDIT PREV DHCP OFF EDIT EDIT INSTRUMENT IP 3 GATEWAY IP 3 SUBNET MASK 3 TCP PORT 4 HOSTNAME 5 300 1200 2400 4800 9600 19200 38400 57600 115200 QUIET COMPUTER SECURITY HESSEN PROTOCOL E, 7, 1 RS-485 MULTIDROP PROTOCOL ENABLE MODEM ERROR CHECKING XON/XOFF HANDSHAKE HARDWARE HANDSHAKE HARDWARE FIFO COMMAND PROMPT ON OFF Figure A-4: NEXT 0) 1) 2) 3) 4) 5) 6) BAUD RATE TEST PORT TEST ON DIAG JUMP EDIT PRNT DAS_HOLD_OFF TPC_ENABLE RCELL_SET DYN_ZERO DYN_SPAN CONC_PRECISION CLOCK_ADJ ENTER PASSWORD: 818 Go to DIAG Menu Tree 1 Only appears if optional Ethernet PCA is installed. NOTE: When Ethernet PCA is present COM2 submenu disappears. 2 Only appears if HESSEN PROTOCOL mode is ON (See COM1 & COM2 – MODE submenu above). 3 INSTRUMENT IP, GATEWAY IP & SUBNET MASK are only editable when DHCP is OFF. 4 Although TCP PORT is editable regardless of the DHCP state, do not change the setting for this property. 5 HOST NAME is only editable when DHCP is ON. Secondary Setup Menu COMM and VARS Submenus A-5 04521C (DCN5731) APPENDIX A-1: M200EH/EM Software Menu Trees, Revision F.0 Model 200EH/EM (Ref: 05147F) SAMPLE CFG ACAL DAS RNGE PASS SETUP MORE CLK COMM HESN 2 INET 1 ID COM1 COM2 ENTER PASSWORD: 818 ENTER PASSWORD: 818 ENTER PASSWORD: 818 RESPONSE MODE BCC TEXT PREV NOX, 211, REPORTED EDIT Go to COMM / VARS Menu Tree GAS LIST NO2, 213 REPORTED STATUS FLAGS CMD NEXT INS DEL YES NO, 212, REPORTED NO EDIT PRNT GAS TYPE GAS ID REPORTED O2, 214, REPORTED ON OFF 1 Only appears if Ethernet Option is installed. 2 Only appears if HESSEN PROTOCOL mode is ON. Figure A-5: Go to DIAG Menu Tree Set/create unique gas ID number NOX NO NO2 O2 Secondary Setup Menu Hessen Protocol Submenu A-6 04521C (DCN5731) APPENDIX A-1: M200EH/EM Software Menu Trees, Revision F.0 Model 200EH/EM (Ref: 05147F) SETUP SAMPLE CFG ACAL DAS RNGE PASS CLK MORE DIAG COMM VARS ENTER PASSWORD: 818 PREV DISPLAY SEQUENCE CONFIGURATION ANALOG CONFIGURATION ANALOG OUTPUT SIGNAL I/O Press ENTR to start test PREV NEXT 0) 1) 2) 3) 4) 5) EXT ZERO CAL EXT SPAN CAL EXT LOW SPAN REMOTE RANGE HI MAINT MODE LANG2 SELECT 6) 7) 8) 9) 10) 11) 12) 13) 14) 15) 16) 17) 18) 19) 20) 21) 22) 23) 24) 25) 26) 27) 28) 29) 30) 31) 32) 33) 34) 35) SAMPLE LED CAL LED FAULT LED AUDIBLE BEEPER ELEC TEST OPTIC TEST PREAMP RANGE HIGH O3GEN STATUS ST SYSTEM OK ST CONC VALID ST HIGH RANGE ST ZERO CAL ST SPAN CAL ST DIAG MODE ST LOW SPAN CAL ST O2 CAL ST SYSTEM OK2 ST CONC ALARM 1 ST CONC ALARM 2 RELAY WATCHDOG RCELL HEATER CONV HEATER MANIFOLD HEATER O2 CELL HEATER ZERO VALVE CAL VALVE AUTO ZERO VALVE NOX VALVE LOW SPAN VALVE HIGH SPAN VALVE NEXT FLOW ELECTRICAL OZONE GEN OVERRIDE CALIBRATION TEST OPTIC TEST Press ENTR to start test ON Press ENTR to start test EDIT PREV AOUTS CALIBRATED DATA DATA DATA DATA ON OUT OUT OUT OUT NEXT INS PREV AIN CALIBRATED OFF DEL YES Cycles through list of already programmed display sequences 11 21 31 41 SAMP OFF EDIT PRNT NO NOX NXL NXH NO NOL NOH NO2 N2L N2H O2 NEXT DISPLAY DATA CAL RANGE OVER RANGE AUTO 2 CALIBRATED OUTPUT RANGE OFFSET 2 CAL ON OFF ON ON OFF OFF Sets the degree of offset 36 INTERNAL ANALOG to VOLTAGE SIGNALS 61 (see Appendix A) CAL 2 Auto Cal 0.1V 1V 5V 10V 1 Correspond to analog Output A1 – A4 on back of analyzer 2 Only appears if one of the voltage ranges is selected. 3 Manual adjustment menu only appears if either the Auto Cal feature is OFF or the range is set for CURRent. Manual Cal3 DATA SCALE UPDATE Sets the scale width of the reporting range. Cycles through the list of iDAS data types. OZONE ENTR DISPLAY DURATION Sets time lapse between data updates on selected output CURR U100 Figure A-6: UP10 UP DOWN DN10 D100 DIAG Menu A-7 04521C (DCN5731) APPENDIX A-2: Setup Variables For Serial I/O, Revision F.0 Model 200EH/EM (Ref: 05147F) APPENDIX A-2: Setup Variables For Serial I/O, Revision F.0 Table A-1: M200EH/EM Setup Variables, Revision F.0 Setup Variable Numeric Units Default Value Value Range DAS_HOLD_OFF Minutes 15 0.5–20 MEASURE_MODE — NO-NOX, NO, NOX, NOX-NO, NON-OX Gas measure mode. Enclose value in double quotes (") when setting from the RS-232 interface. TPC_ENABLE — ON OFF, ON ON enables temperature/ pressure compensation; OFF disables it. DYN_ZERO — OFF ON, OFF ON enables remote dynamic zero calibration; OFF disables it. DYN_SPAN — OFF ON, OFF ON enables remote dynamic span calibration; OFF disables it. CONC_PRECISION — 3 CLOCK_ADJ Sec./Day 0 AUTO, 0, 1, 2, 3, 4 -60–60 ENGL, SECD, Description Duration of DAS hold off period. Number of digits to display to the right of the decimal point for concentrations on the display. Enclose value in double quotes (") when setting from the RS-232 interface. Time-of-day clock speed adjustment. Selects the language to use for the user interface. Enclose value in double quotes (") when setting from the RS-232 interface. LANGUAGE_SELECT — ENGL MAINT_TIMEOUT Hours 2 0.1–100 BXTEMP_TPC_GAIN — 0 0–10 Box temperature compensation attenuation factor. RCTEMP_TPC_GAIN — 0 0–10 Reaction cell temperature compensation attenuation factor. RCPRESS_TPC_GAIN — 1 0–10 Reaction cell pressure compensation attenuation factor. SPRESS_TPC_GAIN — 1 0–10 Sample pressure compensation attenuation factor. CE_FACTOR1 — 1 0.8–1.2 Moly converter efficiency factor for range 1. CE_FACTOR2 — 1 0.8–1.2 Moly converter efficiency factor for range 2. 1 SEC 33 MS, 66 MS, 133 MS, 266 MS, 533 MS, 1 SEC, 2 SEC CONV_TIME SG_CONV_TIME — — 33 MS EXTN Same as above. NEG_NO2_SUPPRESS — ON ON, OFF FILT_SIZE Samples 1 2 5 , 10 1–500 Time until automatically switching out of software-controlled maintenance mode. Conversion time for PMT detector channel. Enclose value in double quotes (") when setting from the RS232 interface. Conversion time for PMT detector channel in single-gas measure modes. Enclose value in double quotes (") when setting from the RS232 interface. ON suppresses negative NO2 in during switching mode; OFF does not suppress negative NO2 readings Moving average filter size. A-8 04521C (DCN5731) Model 200EH/EM (Ref: 05147F) APPENDIX A-2: Setup Variables For Serial I/O, Revision F.0 Setup Variable Numeric Units Default Value Value Range SG_FILT_SIZE Samples 60 1–500 FILT_ADAPT — ON ON, OFF FILT_OMIT_DELTA PPM FILT_OMIT_PCT % FILT_SHORT_DELT PPM 5 , 0.5 FILT_SHORT_PCT % 5 ,7 1 2 1–100 Percent change in concentration to shorten filter. FILT_ASIZE Samples 2 ,3 1 2 1–500 Moving average filter size in adaptive mode. SG_FILT_ASIZE Samples 10 1–500 Moving average filter size in adaptive mode, in single-gas measure modes. 60 1,80 2 0–200 Delay before leaving adaptive filter mode. Seconds 60 0–200 Delay before leaving adaptive filter mode in single-gas measure modes. — ON ON, OFF O2_FILT_SIZE 4 Samples 60 1–500 O2 moving average filter size in normal mode. O2_FILT_ASIZE 4 Samples 10 1–500 O2 moving average filter size in adaptive mode. O2_FILT_DELTA 4 % 2 0.1–100 Absolute change in O2 concentration to shorten filter. % 2 0.1–100 Relative change in O2 concentration to shorten filter. Seconds 20 0–300 Delay before leaving O2 adaptive filter mode. 0.1–30 Dwell time after switching valve to NOX position. 1 10 , 0.8 2 10 Moving average filter size in singlegas measure modes. 1 5–100 , 0.1–100 2 1–100 1 Seconds FILT_DELAY SG_FILT_DELAY O2_FILT_ADAPT O2_FILT_PCT 4 4 O2_FILT_DELAY 4 1 1 5–100 , 2 0.1–100 2 2 Description ON enables adaptive filter; OFF disables it. Absolute change in concentration to omit readings. Percent change in concentration to omit readings. Absolute change in concentration to shorten filter. ON enables O2 adaptive filter; OFF disables it. NOX_DWELL Seconds SG_NOX_DWELL Seconds 1 0.1–30 Dwell time after switching valve to NOX position in single-gas measure modes. NOX_SAMPLE Samples 2 1–30 Number of samples to take in NOX mode. SG_NOX_SAMPLE Samples 2 1–30 Number of samples to take in NOX mode in single-gas measure modes. NO_DWELL Seconds SG_NO_DWELL Seconds NO_SAMPLE 4.2 , 3.5 0.1–30 Dwell time after switching valve to NO position. 1 0.1–30 Dwell time after switching valve to NO position in single-gas measure modes. Samples 2 1–30 Number of samples to take in NO mode. SG_NO_SAMPLE Samples 2 1–30 Number of samples to take in NO mode in single-gas measure modes. USER_UNITS — PPM PPM, MGM Concentration units for user interface. Enclose value in double quotes (") when setting from the RS232 interface. 1 4.2 , 3.0 2 A-9 04521C (DCN5731) APPENDIX A-2: Setup Variables For Serial I/O, Revision F.0 Model 200EH/EM (Ref: 05147F) Setup Variable Numeric Units Default Value Value Range Description DIL_FACTOR — 1 1–1000 Dilution factor. Used only if is dilution enabled with FACTORY_OPT variable. AZERO_ENABLE — ON ON, OFF ON enables auto-zero; OFF disables it. AZERO_FREQ Minutes 2 0–60 AZERO_DWELL Seconds 4 0.1–60 Dwell time after opening auto-zero valve. AZERO_POST_DWELL Seconds 4 0–60 Dwell time after closing auto-zero valve. AZERO_SAMPLE Samples 2 1–10 Number of auto-zero samples to average. SG_AZERO_SAMP Samples 2 1–10 Number of auto-zero samples to average in single-gas measure modes. AZERO_FSIZE Samples 8 1–50 Moving average filter size for autozero samples. AZERO_LIMIT mV 200 0–1000 Maximum auto-zero offset allowed. NOX_SPAN1 Conc. 1 2 80 , 16 1–5000 Target NOX concentration during span calibration of range 1. NO_SPAN1 Conc. 1 2 80 , 16 1–5000 Target NO concentration during span calibration of range 1. NO2_SPAN1 Conc. 1 2 80 , 16 1–5000 Target NO2 concentration during converter efficiency calibration of range 1. NOX_SLOPE1 PPM/mV 1 0.25–4 NOX slope for range 1. NOX_OFFSET1 mV 0 -10000–10000 NOX offset for range 1. NO_SLOPE1 PPM/mV 1 0.25–4 NO slope for range 1. NO_OFFSET1 mV 0 -10000–10000 NO offset for range 1. NOX_SPAN2 Conc. 1 2 80 , 16 1–5000 Target NOX concentration during span calibration of range 2. NO_SPAN2 Conc. 1 2 80 , 16 1–5000 Target NO concentration during span calibration of range 2. NO2_SPAN2 Conc. 1 2 80 , 16 1–5000 Target NO2 concentration during converter efficiency calibration of range 2. NOX_SLOPE2 PPM/mV 1 0.25–4 NOX slope for range 2. NOX_OFFSET2 mV 0 -10000–10000 NOX offset for range 2. NO_SLOPE2 PPM/mV 1 0.25–4 NO slope for range 2. mV 0 -10000–10000 NO offset for range 2. % 20.95 0–100 Target O2 concentration during span calibration. NO_OFFSET2 O2_TARG_SPAN_CONC O2_SLOPE 4 4 Auto-zero frequency. — 1 0.5–2 O2 slope. O2_OFFSET 4 % 0 -10–10 O2 offset. O2_RANGE 4 % 100 0.1–500 O2 concentration range. STD_O2_CELL_TEMP 4 ºK 323 1–500 PHYS_RANGE1 PPM 1 2 500 , 20 PHYS_RANGE2 CONC_RANGE1 PPM Conc. 1 5000 , 200 1 2 100 , 20 2 Standard O2 cell temperature for temperature compensation. 5–5000 Low pre-amp range. 5–5000 High pre-amp range. 1 5–5000 , 2 1–500 D/A concentration range 1 or range for NOX. A-10 04521C (DCN5731) Model 200EH/EM (Ref: 05147F) Setup Variable APPENDIX A-2: Setup Variables For Serial I/O, Revision F.0 Numeric Units Default Value Value Range 50 RCELL_SET ºC Warnings: 30–70 Description Reaction cell temperature set point and warning limits. 45–55 50 MANIFOLD_SET ºC Warnings: 30–70 Manifold temperature set point and warning limits. 45–55 50 O2_CELL_SET 4 ºC Warnings: 30–70 O2 sensor cell temperature set point and warning limits. 45–55 CONV_TYPE — MOLY NONE, MOLY, CONV, O3KL 315 CONV_SET ºC Warnings: 0–800 Converter type. “CONV” is minihicon. Enclose value in double quotes (") when setting from the RS232 interface. Converter temperature set point and warning limits. 305–325 CONV_TEMP_TRIG Cycles 10 0–100 30 BOX_SET ºC Warnings: 0–70 Number of converter temperature errors required to trigger warning. Nominal box temperature set point and warning limits. 7–48 7 PMT_SET ºC Warnings: 0–40 PMT temperature warning limits. Set point is not used. 5–12 1 1+4 Sample flow warning limits. Set point is not used. 290 , 360 , 250 2, 320 2+4 SFLOW_SET cc/m Warnings: 350–600, 100–1000 200–600 1,2, 300–700 1+4, 2+4 SAMP_FLOW_SLOPE — 1 0.001–100 250 OFLOW_SET cc/m OZONE_FLOW_SLOPE — 1 0.001–100 RCELL_SAMP_RATIO — 0.53 0.1–2 Warnings: 100–1000 Slope term to correct sample flow rate. Ozone flow warning limits. Set point is not used. 200–600 298 STD_BOX_TEMP ºK Valid limits: 1–500 Slope term to correct ozone flow rate. Maximum reaction cell pressure / sample pressure ratio for valid sample flow calculation. Standard box temperature and valid limits for temperature compensation. 278–338 323 STD_RCELL_TEMP ºK Valid limits: 1–500 Standard reaction cell temperature and valid limits for temperature compensation. 278–338 A-11 04521C (DCN5731) APPENDIX A-2: Setup Variables For Serial I/O, Revision F.0 Setup Variable Numeric Units Model 200EH/EM (Ref: 05147F) Default Value Value Range 5 STD_RCELL_PRESS "Hg Valid limits: 0.1–50 Description Standard reaction cell pressure and valid limits for pressure compensation. 0.5–12 29.92 STD_SAMP_PRESS "Hg Valid limits: 0.1–50 Standard sample pressure and valid limits for pressure compensation. 0.5–32 RS-232 COM1 mode flags. Add values to combine flags. 1 = quiet mode 2 = computer mode 4 = enable security 16 = enable Hessen protocol 3 32 = enable multidrop RS232_MODE — 0 0–65535 64 = enable modem 128 = ignore RS-232 line errors 256 = disable XON / XOFF support 512 = reserved 1024 = enable RS-485 mode 2048 = even parity, 7 data bits, 1 stop bit 4096 = enable command prompt BAUD_RATE — 19200 300, 1200, 2400, 4800, 9600, 19200, 38400, RS-232 COM1 baud rate. Enclose value in double quotes (") when setting from the RS-232 interface. 57600, 115200 Any character in the allowed character set. Up to 100 characters long. RS-232 COM1 modem initialization string. Sent verbatim plus carriage return to modem on power up or manually. Enclose value in double quotes (") when setting from the RS232 interface. MODEM_INIT — “AT Y0 &D0 &H0 &I0 S0=2 &B0 &N6 &M0 E0 Q1 &W0” RS232_MODE2 BitFlag 0, 0–65535 — 19200 300, 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200 RS-232 COM2 baud rate. Enclose value in double quotes (") when setting from the RS-232 interface. MODEM_INIT2 — “AT Y0 &D0 &H0 &I0 S0=2 &B0 &N6 &M0 E0 Q1 &W0” Any character in the allowed character set. Up to 100 characters long. RS-232 COM2 modem initialization string. Sent verbatim plus carriage return to modem on power up or manually. Enclose value in double quotes (") when setting from the RS232 interface. RS232_PASS Password 940331 0–999999 MACHINE_ID ID 200 0–9999 “Cmd> ” Any character in the allowed character set. Up to 100 characters long. BAUD_RATE2 COMMAND_PROMPT — RS-232 COM2 mode flags. (Same settings as RS232_MODE) RS-232 log on password. Unique ID number for instrument. RS-232 interface command prompt. Displayed only if enabled with RS232_MODE variable. Enclose value in double quotes (") when setting from the RS-232 interface. A-12 04521C (DCN5731) Model 200EH/EM (Ref: 05147F) APPENDIX A-2: Setup Variables For Serial I/O, Revision F.0 Setup Variable Numeric Units Default Value Value Range Description REMOTE_CAL_MODE — LOW LOW, HIGH Range to calibrate during remote calibration. Enclose value in double quotes (") when setting from the RS232 interface. PASS_ENABLE — OFF ON, OFF STABIL_GAS — NOX NO, NO2, NOX STABIL_FREQ Seconds 10 1–300 STABIL_SAMPLES Samples 25 2–40 1 550 , 600 HVPS_SET Volts 2 Warnings: 400–700 1, ON enables passwords; OFF disables them. Selects gas for stability measurement. Enclose value in double quotes (") when setting from the RS-232 interface. Stability measurement sampling frequency. Number of samples in concentration stability reading. High voltage power supply warning limits. Set point is not used. 0–2000 450–750 2 6 RCELL_PRESS_SET In-Hg Warnings: 0–100 Reaction cell pressure warning limits. Set point is not used. 0.5–15 Reaction cell temperature control cycle period. RCELL_CYCLE Seconds 10 0.5–30 RCELL_PROP 1/ºC 1 0–10 Reaction cell PID temperature control proportional coefficient. RCELL_INTEG — 0.1 0–10 Reaction cell PID temperature control integral coefficient. RCELL_DERIV — 0 (disabled) 0–10 Reaction cell PID temperature control derivative coefficient. MANIFOLD_CYCLE Seconds 5 0.5–30 MANIFOLD_PROP 1/ºC 0.2 0–10 Manifold PID temperature control proportional coefficient. MANIFOLD_INTEG — 0.1 0–10 Manifold PID temperature control integral coefficient. MANIFOLD_DERIV — 0.5 0–10 Manifold PID temperature control derivative coefficient. O2_CELL_CYCLE 4 Seconds 10 0.5–30 O2 cell temperature control cycle period. Manifold temperature control cycle period. O2_CELL_PROP 4 — 1 0–10 O2 cell PID temperature control proportional coefficient. O2_CELL_INTEG 4 — 0.1 0–10 O2 cell PID temperature control integral coefficient. O2_CELL_DERIV 4 — 0 (disabled) 0–10 O2 cell PID temperature control derivative coefficient. — 8 0.1–100 Any character in the allowed character set. Up to 100 characters long. Unique serial number for instrument. Enclose value in double quotes (") when setting from the RS-232 interface. HIGH,MED, Front panel display intensity. Enclose value in double quotes (") when setting from the RS-232 interface. SLOPE_CONST SERIAL_NUMBER — “00000000 ” DISP_INTENSITY — HIGH LOW, DIM Slope constant factor to keep visible slope near 1. A-13 04521C (DCN5731) APPENDIX A-2: Setup Variables For Serial I/O, Revision F.0 Model 200EH/EM (Ref: 05147F) Setup Variable Numeric Units Default Value Value Range I2C_RESET_ENABLE — ON OFF, ON ALARM_TRIGGER Cycles 10 1–100 Description 2 I C bus automatic reset enable. Number of valve cycles to trigger concentration alarm. Time-of-day clock format flags. Enclose value in double quotes (") when setting from the RS-232 interface. “%a” = Abbreviated weekday name. “%b” = Abbreviated month name. “%d” = Day of month as decimal number (01 – 31). “%H” = Hour in 24-hour format (00 – 23). “%I” = Hour in 12-hour format (01 – 12). CLOCK_FORMAT — “TIME=%H:%M: %S” Any character in the allowed character set. Up to 100 characters long. “%j” = Day of year as decimal number (001 – 366). “%m” = Month as decimal number (01 – 12). “%M” = Minute as decimal number (00 – 59). “%p” = A.M./P.M. indicator for 12hour clock. “%S” = Second as decimal number (00 – 59). “%w” = Weekday as decimal number (0 – 6; Sunday is 0). “%y” = Year without century, as decimal number (00 – 99). “%Y” = Year with century, as decimal number. “%%” = Percent sign. Factory option flags. Add values to combine flags. 1 = enable dilution factor 2 = display units in concentration field 4 = zero/span valves installed 8 = low span valve installed 16 = IZS and zero/span valves installed 32 = enable software-controlled maintenance mode FACTORY_OPT — 0, 512 0–65535 64 = display temperature in converter warning message 128 = enable switch-controlled maintenance mode 256 = enable simultaneous display of all gas concentrations 512 = enable manifold temperature control 1024 = enable O2 sensor cell temperature control (temporarily removed) 2048 = enable Internet option A-14 04521C (DCN5731) Model 200EH/EM (Ref: 05147F) Setup Variable 1 M200EH. 2 M200EM. APPENDIX A-2: Setup Variables For Serial I/O, Revision F.0 Numeric Units Default Value 3 Must power-cycle instrument for these options to fully take effect. 4 O2 option. Value Range Description A-15 04521C (DCN5731) APPENDIX A-3: Warnings and Test Functions, Revision F.0 Model 200EH/EM (Ref: 05147F) APPENDIX A-3: Warnings and Test Functions, Revision F.0 Table A-2: M200EH/EM Warning Messages, Revision F.0 Name Message Text WSYSRES SYSTEM RESET Description Instrument was power-cycled or the CPU was reset. WDATAINIT DATA INITIALIZED WCONFIGINIT CONFIG INITIALIZED WSAMPFLOW SAMPLE FLOW WARN Sample flow outside of warning limits specified by SFLOW_SET variable. WOZONEFLOW OZONE FLOW WARNING Ozone flow outside of warning limits specified by OFLOW_SET variable. WOZONEGEN OZONE GEN OFF WRCELLPRESS RCELL PRESS WARN Reaction cell pressure outside of warning limits specified by RCELL_PRESS_SET variable. WBOXTEMP BOX TEMP WARNING Chassis temperature outside of warning limits specified by BOX_SET variable. WRCELLTEMP RCELL TEMP WARNING Reaction cell temperature outside of warning limits specified by RCELL_SET variable. WMANIFOLDTEMP 4 MANIFOLD TEMP WARN Bypass or dilution manifold temperature outside of warning limits specified by MANIFOLD_SET variable. WO2CELLTEMP 5 O2 CELL TEMP WARN O2 sensor cell temperature outside of warning limits specified by O2_CELL_SET variable. WIZSTEMP IZS TEMP WARNING IZS temperature outside of warning limits specified by IZS_SET variable. WCONVTEMP CONV TEMP WARNING WPMTTEMP PMT TEMP WARNING PMT temperature outside of warning limits specified by PMT_SET variable. WAUTOZERO AZERO WRN XXX.X MV Auto-zero reading above limit specified by AZERO_LIMIT variable. Value shown in message indicates auto-zero reading at time warning was displayed. WHVPS HVPS WARNING WDYNZERO CANNOT DYN ZERO Contact closure zero calibration failed while DYN_ZERO was set to ON. WDYNSPAN CANNOT DYN SPAN Contact closure span calibration failed while DYN_SPAN was set to ON. 1 Data storage was erased. Configuration storage was reset to factory configuration or erased. Ozone generator is off. This is the only warning message that automatically clears itself. It clears itself when the ozone generator is turned on. Converter temperature outside of warning limits specified by CONV_SET variable. High voltage power supply output outside of warning limits specified by HVPS_SET variable. WREARBOARD REAR BOARD NOT DET WRELAYBOARD RELAY BOARD WARN Firmware is unable to communicate with the relay board. WFRONTPANEL FRONT PANEL WARN Firmware is unable to communicate with the front panel. WANALOGCAL ANALOG CAL WARNING Rear board was not detected during power up. The A/D or at least one D/A channel has not been calibrated. O2 option. A-16 04521C (DCN5731) Model 200EH/EM (Ref: 05147F) APPENDIX A-3: Warnings and Test Functions, Revision F.0 Table A-3: TEST Function M200EH/EM Test Functions, Revision F.0 Message Text Description RNG_DATA_OUT_1 “A1:= ” D/A #1 range. RNG_DATA_OUT_2 “A2:= ” D/A #2 range. RNG_DATA_OUT_3 “A3:= ” D/A #3 range. RNG_DATA_OUT_4 “A4:= ” D/A #4 range. NONOXCONC 1 NO=396.5 NOX=396.5 3 Simultaneously displays NO and NOX concentrations. RANGE RANGE=500.0 PPB 3 RANGE1 RANGE1=500.0 PPB 3 D/A #1 range in independent range mode. RANGE2 RANGE2=500.0 PPB 3 D/A #2 range in independent range mode. RANGE3 RANGE3=500.0 PPB 3 O2RANGE 2 O2 RANGE=200.00 % STABILITY NOX STB=0.0 PPB 2 D/A range in single or auto-range modes. 3 D/A #3 range in independent range mode. D/A #4 range for O2 concentration. Concentration stability (standard deviation based on setting of STABIL_FREQ and STABIL_SAMPLES). Select gas with STABIL_GAS variable. RSP=8.81(1.30) SEC Instrument response. Length of each signal processing loop. Time in parenthesis is standard deviation. SAMPFLOW SAMP FLW=460 CC/M Sample flow rate. OZONEFLOW OZONE FL=87 CC/M Ozone flow rate. PMT PMT=800.0 MV Raw PMT reading. NORMPMT NORM PMT=793.0 MV PMT reading normalized for temperature, pressure, auto-zero offset, but not range. AUTOZERO AZERO=1.3 MV Auto-zero offset. HVPS HVPS=650 V High voltage power supply output. RCELLTEMP RCELL TEMP=50.8 C Reaction cell temperature. BOXTEMP BOX TEMP=28.2 C Internal chassis temperature. PMTTEMP PMT TEMP=7.0 C PMT temperature. MANIFOLDTEMP MF TEMP=50.8 C Bypass or dilution manifold temperature. O2 CELL TEMP=50.8 C O2 sensor cell temperature. RESPONSE O2CELLTEMP 2 IZSTEMP IZS TEMP=50.8 C IZS temperature. CONVTEMP MOLY TEMP=315.0 C Converter temperature. Converter type is MOLY, CONV, or O3KL. RCELLPRESS RCEL=7.0 IN-HG-A Reaction cell pressure. SAMPPRESS SAMP=29.9 IN-HG-A Sample pressure. NOXSLOPE NOX SLOPE=1.000 NOX slope for current range, computed during zero/span calibration. NOXOFFSET NOX OFFS=0.0 MV NOX offset for current range, computed during zero/span calibration. NOSLOPE NO SLOPE=1.000 NO slope for current range, computed during zero/span calibration. NOOFFSET NO OFFS=0.0 MV NO offset for current range, computed during zero/span calibration. NO2 NO2=0.0 PPB NOX 3 NOX=396.5 PPB NO2 concentration for current range. 3 NOX concentration for current range. A-17 04521C (DCN5731) APPENDIX A-3: Warnings and Test Functions, Revision F.0 TEST Function Message Text Model 200EH/EM (Ref: 05147F) Description NO=396.5 PPB 3 NO concentration for current range. O2 SLOPE=1.000 O2 slope computed during zero/span calibration. O2OFFSET 2 O2 OFFSET=0.00 % O2 offset computed during zero/span calibration. O2 2 O2=0.00 % O2 concentration. TESTCHAN TEST=3627.1 MV Value output to TEST_OUTPUT analog output, selected with TEST_CHAN_ID variable. CLOCKTIME TIME=10:38:27 Current instrument time of day clock. NO O2SLOPE 2 1 Factory option. 2 O2 option. A-18 04521C (DCN5731) Model 200EH/EM (Ref: 05147F) APPENDIX A-4: M200EH/EM Signal I/O Definitions, Revision F.0 APPENDIX A-4: M200EH/EM Signal I/O Definitions, Revision F.0 Table A-4: M200EH/EM Signal I/O Definitions, Revision F.0 BIT OR CHANNEL NUMBER SIGNAL NAME DESCRIPTION Internal inputs, U7, J108, pins 9–16 = bits 0–7, default I/O address 322 hex 0–7 Spare Internal outputs, U8, J108, pins 1–8 = bits 0–7, default I/O address 322 hex ELEC_TEST 0 1 = electrical test on OPTIC_TEST 1 1 = optic test on PREAMP_RANGE_HI 2 1 = select high preamp range O3GEN_STATUS 3 0 = ozone generator on 0 = off 0 = off 0 = select low range 1 = off 4–5 I2C_RESET 6 Spare 2 1 = reset I C peripherals 0 = normal I2C_DRV_RST 7 0 = hardware reset 8584 chip 1 = normal Control inputs, U11, J1004, pins 1–6 = bits 0–5, default I/O address 321 hex EXT_ZERO_CAL 0 0 = go into zero calibration 1 = exit zero calibration EXT_SPAN_CAL 1 0 = go into span calibration 1 = exit span calibration EXT_LOW_SPAN 2 0 = go into low span calibration 1 = exit low span calibration REMOTE_RANGE_HI 3 0 = remote select high range 1 = default range 4–5 Spare 6–7 Always 1 Control inputs, U14, J1006, pins 1–6 = bits 0–5, default I/O address 325 hex 0–5 Spare 6–7 Always 1 Control outputs, U17, J1008, pins 1–8 = bits 0–7, default I/O address 321 hex 0–7 Spare Control outputs, U21, J1008, pins 9–12 = bits 0–3, default I/O address 325 hex 0–3 Spare Alarm outputs, U21, J1009, pins 1–12 = bits 4–7, default I/O address 325 hex ST_SYSTEM_OK2 2 4 1 = system OK 0 = any alarm condition or in diagnostics mode ST_CONC_ALARM_1 5 1 = conc. limit 1 exceeded 0 = conc. OK ST_CONC_ALARM_2 6 1 = conc. limit 2 exceeded 0 = conc. OK A-19 04521C (DCN5731) APPENDIX A-4: M200EH/EM Signal I/O Definitions, Revision F.0 Model 200EH/EM (Ref: 05147F) BIT OR CHANNEL NUMBER SIGNAL NAME 7 DESCRIPTION Spare A status outputs, U24, J1017, pins 1–8 = bits 0–7, default I/O address 323 hex ST_SYSTEM_OK 0 0 = system OK 1 = any alarm condition ST_CONC_VALID 1 0 = conc. valid 1 = hold off or other conditions ST_HIGH_RANGE 2 0 = high auto-range in use ST_ZERO_CAL 3 0 = in zero calibration 1 = low auto-range 1 = not in zero ST_SPAN_CAL 4 0 = in span calibration 1 = not in span ST_DIAG_MODE 5 ST_LOW_SPAN_CAL 6 0 = in diagnostic mode 1 = not in diagnostic mode 0 = in low span calibration 1 = not in low span ST_O2_CAL 1 7 0 = in O2 calibration mode 1 = in NOX calibration mode B status outputs, U27, J1018, pins 1–8 = bits 0–7, default I/O address 324 hex 0–7 Spare 2 Front panel I C keyboard, default I2C address 4E hex MAINT_MODE 5 (input) 0 = maintenance mode 1 = normal mode LANG2_SELECT 6 (input) SAMPLE_LED 8 (output) 0 = select second language 1 = select first language (English) 0 = sample LED on 1 = off CAL_LED 9 (output) 0 = cal. LED on 1 = off FAULT_LED 10 (output) 0 = fault LED on AUDIBLE_BEEPER 14 (output) 0 = beeper on (for diagnostic testing only) 1 = off 1 = off Relay board digital output (PCF8575), default I2C address 44 hex RELAY_WATCHDOG 0 RCELL_HEATER 1 Alternate between 0 and 1 at least every 5 seconds to keep relay board active 0 = reaction cell heater on 1 = off CONV_HEATER 2 0 = converter heater on 1 = off MANIFOLD_HEATER 3 0 = bypass or dilution manifold heater on 1 = off IZS_HEATER 4 0 = IZS heater on 1 = off A-20 04521C (DCN5731) Model 200EH/EM (Ref: 05147F) APPENDIX A-4: M200EH/EM Signal I/O Definitions, Revision F.0 SIGNAL NAME BIT OR CHANNEL NUMBER O2_CELL_HEATER 1 5 DESCRIPTION 0 = O2 sensor cell heater on 1 = off ZERO_VALVE 6 0 = let zero gas in 1 = let sample gas in SPAN_VALVE 6 0 = let span gas in 1 = let zero gas in CAL_VALVE 7 0 = let cal. gas in 1 = let sample gas in AUTO_ZERO_VALVE 8 0 = let zero air in 1 = let sample gas in NOX_VALVE 9 0 = let NOX gas into reaction cell 1 = let NO gas into reaction cell LOW_SPAN_VALVE 10 0 = let low span gas in 1 = let sample gas in HIGH_SPAN_VALVE 11 0 = let high span gas in 1 = let sample gas in 12–15 Spare Rear board primary MUX analog inputs, MUX default I/O address 32A hex PMT_SIGNAL 0 PMT detector HVPS_VOLTAGE 1 HV power supply output PMT_TEMP 2 PMT temperature 3 Spare 4 Temperature MUX 5 Spare O2_SENSOR 1 6 O2 concentration sensor SAMPLE_PRESSURE 7 Sample pressure RCELL_PRESSURE 8 Reaction cell pressure REF_4096_MV 9 4.096V reference from MAX6241 OZONE_FLOW 10 Ozone flow rate TEST_INPUT_11 11 Diagnostic test input CONV_TEMP 12 Converter temperature TEST_INPUT_13 13 Diagnostic test input 14 DAC loopback MUX REF_GND 15 Ground reference Rear board temperature MUX analog inputs, MUX default I/O address 326 hex BOX_TEMP 0 Internal box temperature RCELL_TEMP 1 Reaction cell temperature IZS_TEMP 2 IZS temperature 3 Spare 4 O2 sensor cell temperature TEMP_INPUT_5 5 Diagnostic temperature input TEMP_INPUT_6 6 Diagnostic temperature input MANIFOLD_TEMP 7 Bypass or dilution manifold temperature O2_CELL_TEMP 1 A-21 04521C (DCN5731) APPENDIX A-4: M200EH/EM Signal I/O Definitions, Revision F.0 SIGNAL NAME Model 200EH/EM (Ref: 05147F) BIT OR CHANNEL NUMBER DESCRIPTION Rear board DAC MUX analog inputs, MUX default I/O address 327 hex DAC_CHAN_1 0 DAC channel 0 loopback DAC_CHAN_2 1 DAC channel 1 loopback DAC_CHAN_3 2 DAC channel 2 loopback DAC_CHAN_4 3 DAC channel 3 loopback Rear board analog outputs, default I/O address 327 hex CONC_OUT_1, 0 DATA_OUT_1 CONC_OUT_2, Data output #1 1 DATA_OUT_2 CONC_OUT_3, TEST_OUTPUT, DATA_OUT_4 1 O2 option. 2 Optional Concentration output #2 (NO) , Data output #2 2 DATA_OUT_3 CONC_OUT_4 1, Concentration output #1 (NOX), Concentration output #3 (NO2) , Data output #3 3 Test measurement output, Concentration output #4 (O2) , Data output #4 A-22 04521C (DCN5731) Model 200EH/EM (Ref: 05147F) APPENDIX A-5: M200EH/EM iDAS Functions, Revision F.0 APPENDIX A-5: M200EH/EM iDAS Functions, Revision F.0 Table A-5: M200EH/EM DAS Trigger Events, Revision F.0 NAME DESCRIPTION ATIMER Automatic timer expired EXITZR Exit zero calibration mode EXITLS Exit low span calibration mode EXITHS Exit high span calibration mode EXITMP Exit multi-point calibration mode EXITO2 1 Exit O2 calibration mode SLPCHG Slope and offset recalculated O2SLPC 1 O2 slope and offset recalculated EXITDG Concentration exceeds limit 1 warning CONC2W Concentration exceeds limit 2 warning AZEROW Auto-zero warning OFLOWW Ozone flow warning RPRESW Reaction cell pressure warning RTEMPW Reaction cell temperature warning MFTMPW Bypass or dilution manifold temperature warning O2TMPW 1 Exit diagnostic mode CONC1W 1 O2 sensor cell temperature warning IZTMPW IZS temperature warning CTEMPW Converter temperature warning PTEMPW PMT temperature warning SFLOWW Sample flow warning BTEMPW Box temperature warning HVPSW HV power supply warning O2 option. A-23 04521C (DCN5731) APPENDIX A-5: M200EH/EM iDAS Functions, Revision F.0 Table A-6: Model 200EH/EM (Ref: 05147F) M200EH/EM iDAS Data Types, Revision F.0 NAME DESCRIPTION UNITS PMTDET PMT detector reading mV NXSLP1 NOX slope for range #1 — NXSLP2 NOX slope for range #2 — NOSLP1 NO slope for range #1 — NOSLP2 NO slope for range #2 — NXOFS1 NOX offset for range #1 mV NXOFS2 NOX offset for range #2 mV NOOFS1 NO offset for range #1 mV NOOFS2 NO offset for range #2 mV O2SLPE 1 O2 slope — O2OFST 1 O2 offset Weight % NXZSC1 Concentration for NOX reporting range #1 during zero/span calibration, just before computing new slope and offset PPB NXZSC2 Concentration for NOX reporting range #2 during zero/span calibration, just before computing new slope and offset PPB NOZSC1 Concentration for NO reporting range #1 during zero/span calibration, just before computing new slope and offset PPB NOZSC2 Concentration for NO reporting range #2 during zero/span calibration, just before computing new slope and offset PPB N2ZSC1 Concentration for NO2 reporting range #1 during zero/span calibration, just before computing new slope and offset PPB N2ZSC2 Concentration for NO2 reporting range #2 during zero/span calibration, just before computing new slope and offset PPB O2 concentration during zero/span calibration of the O2 sensor, just before computing new slope and offset Weight % NXCNC1 Concentration for NOX reporting range #1 PPB NXCNC2 Concentration for NOX reporting range #2 PPB NOCNC1 Concentration for NO reporting range #1 PPB NOCNC2 Concentration for NO reporting range #2 PPB N2CNC1 Concentration for NO2 reporting range #1 PPB N2CNC2 Concentration for NO2 reporting range #2 PPB O2 concentration Weight % Concentration stability PPB O2ZSCN 1 1 O2CONC STABIL AZERO Auto zero offset (range de-normalized) mV O3FLOW Ozone flow rate cc/m RCPRES Reaction cell pressure "Hg RCTEMP Reaction cell temperature °C MFTEMP Bypass or dilution manifold temperature °C O2TEMP 1 O2 sensor cell temperature °C IZTEMP IZS block temperature °C CNVEF1 Converter efficiency factor for range #1 — CNVEF2 Converter efficiency factor for range #2 — CNVTMP Converter temperature °C PMTTMP PMT temperature °C A-24 04521C (DCN5731) Model 200EH/EM (Ref: 05147F) NAME DESCRIPTION UNITS SMPFLW Sample flow rate cc/m SMPPRS Sample pressure "Hg BOXTMP HVPS REFGND 1 APPENDIX A-5: M200EH/EM iDAS Functions, Revision F.0 Internal box temperature °C High voltage power supply output Volts Ground reference (REF_GND) mV RF4096 4096 mV reference (REF_4096_MV) mV TEST11 Diagnostic test input (TEST_INPUT_11) mV TEST13 Diagnostic test input (TEST_INPUT_13) mV TEMP5 Diagnostic temperature input (TEMP_INPUT_5) °C TEMP6 Diagnostic temperature input (TEMP_INPUT_6) °C O2 option. A-25 04521C (DCN5731) APPENDIX A-6: Terminal Command Designators, Revision F.0 Model 200EH/EM (Ref: 05147F) APPENDIX A-6: Terminal Command Designators, Revision F.0 Table A-7: Terminal Command Designators, Revision F.0 COMMAND ADDITIONAL COMMAND SYNTAX ? [ID] LOGON [ID] password LOGOFF [ID] T [ID] W [ID] C [ID] D [ID] V [ID] DESCRIPTION Display help screen and this list of commands Establish connection to instrument Terminate connection to instrument SET ALL|name|hexmask Display test(s) LIST [ALL|name|hexmask] [NAMES|HEX] Print test(s) to screen name Print single test CLEAR ALL|name|hexmask Disable test(s) SET ALL|name|hexmask Display warning(s) LIST [ALL|name|hexmask] [NAMES|HEX] Print warning(s) name Clear single warning CLEAR ALL|name|hexmask Clear warning(s) ZERO|LOWSPAN|SPAN [1|2] Enter calibration mode ASEQ number Execute automatic sequence COMPUTE ZERO|SPAN Compute new slope/offset EXIT Exit calibration mode ABORT Abort calibration sequence LIST Print all I/O signals name[=value] Examine or set I/O signal LIST NAMES Print names of all diagnostic tests ENTER name Execute diagnostic test EXIT Exit diagnostic test RESET [DATA] [CONFIG] [exitcode] Reset instrument PRINT ["name"] [SCRIPT] Print iDAS configuration RECORDS ["name"] Print number of iDAS records REPORT ["name"] [RECORDS=number] [FROM= ][TO= ][VERBOSE|COMPACT|HEX] (Print DAS records)(date format: MM/DD/YYYY(or YY) [HH:MM:SS] Print iDAS records CANCEL Halt printing iDAS records LIST Print setup variables name[=value [warn_low [warn_high]]] Modify variable name="value" Modify enumerated variable CONFIG Print instrument configuration MAINT ON|OFF Enter/exit maintenance mode MODE Print current instrument mode DASBEGIN [] DASEND Upload iDAS configuration CHANNELBEGIN propertylist CHANNELEND Upload single iDAS channel CHANNELDELETE ["name"] Delete iDAS channels A-26 04521C (DCN5731) Model 200EH/EM (Ref: 05147F) APPENDIX A-6: Terminal Command Designators, Revision F.0 The command syntax follows the command type, separated by a space character. Strings in [brackets] are optional designators. The following key assignments also apply. TERMINAL KEY ASSIGNMENTS ESC CR (ENTER) Ctrl-C Abort line Execute command Switch to computer mode COMPUTER MODE KEY ASSIGNMENTS LF (line feed) Ctrl-T Execute command Switch to terminal mode USER NOTES: A-27 04521C (DCN5731) APPENDIX A-6: Terminal Command Designators, Revision F.0 Model 200EH/EM (Ref: 05147F) A-28 04521C (DCN5731) Model 200EH/EM Instruction Manual APPENDIX B: Spare Parts and Expendables APPENDIX B: Spare Parts and Expendables NOTE Use of replacement parts other than those supplied by Teledyne-API may result in non-compliance with European standard EN 61010-1. The following lists contain spare parts recommended for the proper care and maintenance of your M200EH/EM: 05480 – Spare Parts List, M200EH 04416 – Recommended Spare Parts Stocking Level, M200EH 05483 – Spare Parts List, M200EM 04415 – Recommended Spare Parts Stocking Level, M200EM 04715 – Expendables Kit M200E/EH/EM 04521C (DCN5731) B-1 APPENDIX B: Spare Parts and Expendables B-2 Model 200EH/EM Instruction Manual 04521C (DCN5731) M200EH Spare Parts List (Ref: 05480S) Part Number 000940100 000940300 000940400 000940500 001761800 002270100 002730000 003290000 005960000 005970000 008830000 009690200 009690300 009810300 009810600 009810700 010680100 010820000 011630000 011930100 013140000 014080100 016290000 016301400 016680600 018080000 018720100 02190020A 022630200 037860000 040010000 040030800 040400000 040410200 040420200 040900000 041800500 041920000 042580000 042680100 042900100 043170000 043220100 043420000 043940000 044340000 044430100 044440000 04521C (DCN5731) Description ORIFICE, 3 MIL, BYPASS MANIFOLD, SAMPLE FLOW CD, ORIFICE, .020 VIOLET ORIFICE, 4 MIL, OZONE DRYER FLOW, O2 OPTION ORIFICE, 7 MIL, OZONE FLOW/BYPASS FLOW ASSY, FLOW CTL, 90CC, OZONE DRYER AKIT, GASKETS, WINDOW, (12) CD, FILTER, 665NM (KB) THERMISTOR, BASIC (VENDOR ASSY)(KB) AKIT, EXPEND, 6LBS ACT CHARCOAL AKIT, EXPENDABLE, 6LB PURAFIL COLD BLOCK (KB) AKIT, TFE FLTR (FL19) ELEM, 47MM, (100) AKIT, TFE FLTR ELEM (FL19), 47MM, 1UM (3 ASSY, PUMP PK, 115V/60HZ w/FL34/NO/SO ASSY, PUMP PACK, 100V/60HZ w/FL34 ASSY, PUMP, 220-240V/50-60HZ (wo) BAND HTR W/TC, 50W @115V, CE/VDE * ASSY, THERMOCOUPLE, HICON, M501 HVPS INSULATOR GASKET (KB) CD, PMT (R928), NOX, M200AH, M200EM/EH * ASSY, COOLER FAN (NOX/SOX) ASSY, HVPS, SOX/NOX WINDOW, SAMPLE FILTER, 47MM (KB) ASSY, SAMP FILT, 47MM, ANG BKT, 1UM, TEE PCA, O3 GEN DRIVER, NOX, E SERIES AKIT, DESSICANT BAGGIES, (12) ASSY, MOLY CONVERTER W/03 DESTRUCTOR ASSY, TC, TYPE K, LONG, WELDED MOLY PCA, TEMP CONTROL BOARD, W/PS, M501 ORING, TFE RETAINER, SAMPLE FILTER ASSY, FAN REAR PANEL, E SERIES PCA, PRESS SENSORS (2X), FLOW, E (NOX) ASSY, HEATERS/THERMAL SWITCH, RX CELL ASSY, VACUUM MANIFOLD, M200EH ASSY, O3 GEN BRK, M200E, HIGH-O/P ORIFICE HOLDER, M200E REACTION CELL (KB) PCA, PMT PREAMP, VR, M200E/EM/EH ASSY, THERMISTOR, REACTION CELL PCA, KEYBOARD, E-SERIES, W/V-DETECT ASSY, VALVE (SS), M200E PROGRAMMED FLASH, E SERIES MANIFOLD, RCELL, M200E, (KB) * THERMOCOUPLE INSULATING SLEEVE, M501NH * ASSY, HEATER/THERM, O2 SEN, "E" SERIES PCA, INTERFACE, ETHERNET, E-SERIES ASSY, HTR, BYPASS MANIFOLD, M200EH ASSY, BYPASS MANIFOLD, M200EH (KB) ASSY, HI-CON CONVERTER W/03 DESTRUCTOR B-3 M200EH Spare Parts List (Ref: 05480S) Part Number 044530000 044540000 044610100 045210000 045230200 045500200 045500400 045500500 046030000 047050100 047210000 048830000 049310100 049760300 050610700 050610900 050611100 051210000 051990000 052930200 054250000 055740000 055740100 055740200 058021100 059940000 061400000 062390000 062420200 062870000 063540100 064540000 064540100 064540200 065190100 065200100 CN0000458 CN0000520 CP0000014 DS0000025 FL0000001 FL0000003 FL0000034 FM0000004 FT0000010 HW0000005 HW0000020 HW0000030 HW0000036 HW0000041 B-4 Description OPTION, O2 SENSOR ASSY, M200EX (KB) ASSY, THERMISTOR, BYPASS MANIFOLD ASSY, VALVES, MOLY/HICON, M200EM/H MANUAL, OPERATORS, M200EH/EM PCA, RELAY CARD, M100E/200E ASSY, ORIFICE HOLDER, 7 MIL, OZONE FLOW ASSY, ORIFICE HOLDER, 3 MIL ASSY, ORIFICE HOLDER, NOX ORIFICE AKIT, CH-43, 3 REFILLS ASSY, ORIFICE HOLDER, BYPASS MANIFOLD ASSY, MINI-HICON GUTS, GROUNDED, M200EH AKIT, EXP KIT, EXHAUST CLNSR, SILCA GEL PCA, TEC CONTROL, E SERIES ASSY, TC PROG PLUG, MOLY,TYP K, TC1 CONFIGURATION PLUGS, 115V, M200E CONFIGURATION PLUGS, 220-240V, M200E CONFIGURATION PLUGS, 100V, M200E ASSY, OZONE DESTRUCTOR ASSY, SCRUBBER, INLINE, PUMP PACK ASSY, BAND HEATER TYPE K, M200EX OPTION, CO2 SENSOR (20%) ASSY, PUMP, NOx PUMP PACK, 115V/60HZ ASSY, PUMP, NOx PUMP PACK, 220V/60HZ ASSY, PUMP, NOx PUMP PACK, 220V/50HZ PCA, E-SERIES MOTHERBD, GEN 5-ICOP OPTION, SAMPLE GAS CONDITIONER, M200A/E ASSY, DUAL HTR, MINI-HICON, 120/240VAC ASSY, MOLY GUTS w/WOOL, M101E/M200EX PCA, SER INTRFACE, ICOP CPU, E- (OPTION) CPU, PC-104, VSX-6150E, ICOP *(KB) DOM, w/SOFTWARE, M200EH * ASSY, PUMP NOX INTERNAL, 115V/60HZ ASSY, PUMP NOX INTERNAL, 230V/60HZ ASSY, PUMP NOX INTERNAL, 230V/50HZ ASSY, NOX CELL TOP-FLO, M200EH >S/N612 ASSY SENSOR, TOP-FLOW, M200EH PLUG, 12, MC 1.5/12-ST-3.81 (KB) PLUG, 10, MC 1.5/10-ST-3.81 (KB) CONTROLLER, TEMP, W/PG-08 (CN262) DISPLAY, E SERIES (KB) FILTER, FLOW CONTROL FILTER, DFU (KB) FILTER, DISPOSABLE, PENTEK (IC-101L)(KB) FLOWMETER (KB) FITTING, FLOW CONTROL FOOT, CHASSI/PUMP PACK SPRING, FLOW CONTROL ISOLATOR, SENSOR ASSY TFE TAPE, 1/4" (48 FT/ROLL) STNOFF,#6-32X3/4" 04521C (DCN5731) M200EH Spare Parts List (Ref: 05480S) Part Number HW0000099 HW0000101 HW0000453 KIT000095 KIT000219 KIT000231 KIT000253 KIT000254 OP0000030 OP0000033 OR0000001 OR0000002 OR0000025 OR0000027 OR0000034 OR0000039 OR0000044 OR0000083 OR0000086 OR0000094 OR0000101 PU0000005 PU0000011 PU0000052 PU0000054 PU0000083 RL0000009 RL0000015 SW0000006 SW0000040 SW0000051 SW0000058 SW0000059 WR0000008 04521C (DCN5731) Description STANDOFF, #6-32X.5, HEX SS M/F ISOLATOR, PUMP PACK SUPPORT, CIRCUIT BD, 3/16" ICOP AKIT, REPLACEMENT COOLER, A/E SERIES KIT, 4-20MA CURRENT OUTPUT (E SERIES) KIT, RETROFIT, M200E/EM/EH Z/S VALVE ASSY & TEST, SPARE PS37, E SERIES ASSY & TEST, SPARE PS38, E SERIES OXYGEN TRANSDUCER, PARAMAGNETIC CO2 MODULE, 0-20% ORING, FLOW CONTROL ORING, REACTION CELL SLEEVE ORING, 2-133V ORING, COLD BLOCK/PMT HOUSING & HEATSINK ORING, (USED W/FT10) ORING, FLOW CONTROL ORING, REACTION CELL MANIFOLD ORING, PMT SIGNAL & OPTIC LED ORING, 2-006, CV-75 COMPOUND(KB) ORING, SAMPLE FILTER ORING, CO2 OPTION PUMP, THOMAS 607, 115V/60HZ (KB) REBUILD KIT, THOMAS 607(KB) PUMP, THOMAS 688, 220/240V 50HZ/60HZ PUMP, THOMAS 688, 100V, 50/60HZ KIT, REBUILD, PU80, PU81, PU82 SSRT RELAY RELAY, DPDT, (KB) SWITCH, THERMAL, 60 C PWR SWITCH/CIR BRK, VDE CE (KB) SWITCH, POWER CIRC BREAK VDE/CE, w/RG(KB SWITCH, THERMAL/450 DEG F PRESSURE SENSOR, 0-15 PSIA, ALL SEN POWER CORD, 10A B-5 M200EH Recommended Spare Parts Stocking Levels (Ref: 04416N) Part Number 011310000 011930000 014080100 040010000 040030800 040400000 040420200 041800500 042580000 Description ASSY, DRYER, NOX CD, PMT (R928), NOX, M200A, M200E(KB) ASSY, HVPS, SOX/NOX ASSY, FAN REAR PANEL, E SERIES PCA, PRESS SENSORS (2X), FLOW, E (NOX) ASSY, HEATERS/THERMAL SWITCH, RX CELL ASSY, O3 GEN BRK, M200E, HIGH-O/P PCA, PMT PREAMP, VR, M200E/EH, (KB) PCA, KEYBOARD, E-SERIES, W/V-DETECT 042680100 ASSY, VALVE (SS), M200E 044440000 044610000 045230200 045500200 045500400 058021100 059940000 062870000 DS0000025 FM0000004 HE0000017 KIT000095 KIT000129 ASSY, HICON w/O3 DEST, M200EH/EM ASSY, VALVES, MOLY/HICON, M200E PCA, RELAY CARD, M100E/200E ASSY, ORIFICE HOLDER, 7 MIL ASSY, ORIFICE HOLDER, 3 MIL PCA, E-SERIES MOTHERBOARD, GEN 5-I OPTION, SAMPLE GAS CONDITIONER, M200A/E CPU, PC-104, VSX-6150E, ICOP *(KB) DISPLAY, E SERIES (KB) FLOWMETER (KB) HTR, 12W/120V (50W/240V), CE AP (KB) AKIT, REPLACEMENT COOLER, A/E SERIES REPLACEMENT, MOLY CONV WELDED CARTRIDGE OP0000030 OXYGEN TRANSDUCER, PARAMAGNETIC OR0000034 OR0000044 OR0000045 PS0000037 PS0000038 PU0000005 RL0000015 ORING, 2-011V FT10 ORING, 2-125V ORING, 2-226V PS, 40W SWITCHING, +5V, +/-15V(KB) * PS, 60W SWITCHING, 12V(KB) * PUMP, THOMAS 607, 115V/60HZ (KB) *1 RELAY, DPDT, (KB) 1 2-5 6-10 1 1 1 1 2 1 2 1 2 1 1 1 1 1 1 2 2 1 1 1 11-20 3 1 1 4 2 2 21-30 UNITS 1 1 4 1 1 4 3 3 1 1 1 2 1 1 1 2 2 1 1 1 1 2 3 3 1 4 2 2 2 4 4 2 2 1 1 3 3 3 1 10 10 10 2 2 1 3 3 1 1 1 2 2 1 1 2 2 2 1 1 5 5 5 1 1 1 1 1 1 2 2 With IZS, ZS Option * With O2 Option *1 * Use KIT000208 To upgrade from 039550200 to 045230200 Relay Board: *1 PU0000005 Use PU0000006 for 220V / 50Hz applications B-6 04521C (DCN5731) M200EM Spare Parts List (Ref: 05483S) Part Number 000940300 000940400 000940500 000941200 001761800 002270100 002730000 009690200 009690300 009810300 009810600 009810700 011630000 011930100 013140000 014080100 016290000 016301400 018080000 018720100 037860000 040010000 040030800 040400000 040410300 040420200 040900000 041800500 041920000 042580000 042680100 042900100 043170000 043420000 043940000 044340000 044430200 044530000 044540000 044610100 045210000 045230200 045500200 047050500 048830000 049310100 04521C (DCN5731) Description CD, ORIFICE, .020 VIOLET ORIFICE, 4 MIL, OZONE DRYER FLOW, O2 OPTION ORIFICE, 7 MIL, OZONE FLOW/SAMPLE FLOW CD, ORIFICE, .008, RED/NONE ASSY, FLOW CTL, 90CC, 1/4" TEE-TMT, B AKIT, GASKETS, WINDOW, (12) CD, FILTER, 665NM (KB) AKIT, TFE FLTR (FL19) ELEM, 47MM, (100) AKIT, TFE FLTR ELEM (FL19), 47MM, 1UM (3 ASSY, PUMP PK, 115V/60HZ w/FL34/NO/SO ASSY, PUMP PACK, 100V/60HZ w/FL34 ASSY, PUMP, 220-240V/50-60HZ (wo) HVPS INSULATOR GASKET (KB) CD, PMT (R928), NOX, M200AH, M200EM/EH * ASSY, COOLER FAN (NOX/SOX) ASSY, HVPS, SOX/NOX WINDOW, SAMPLE FILTER, 47MM (KB) ASSY, SAMP FILT, 47MM, ANG BKT, 1UM, TEE AKIT, DESSICANT BAGGIES, (12) ASSY, MOLY CONVERTER W/03 DESTRUCTOR ORING, TFE RETAINER, SAMPLE FILTER ASSY, FAN REAR PANEL, E SERIES PCA, FLOW/PRESSURE ASSY, HEATERS/THERMAL SWITCH, RX CELL ASSY, VACUUM MANIFOLD, M200EM ASSY, O3 GEN BRK, M200E, HIGH-O/P ORIFICE HOLDER, M200E REACTION CELL (KB) PCA, PMT PREAMP, VR, M200E/EM/EH ASSY, THERMISTOR, REACTION CELL PCA, KEYBOARD, E-SERIES, W/V-DETECT ASSY, VALVE (SS), M200E PROGRAMMED FLASH, E SERIES MANIFOLD, RCELL, M200E, (KB) * ASSY, HEATER/THERM, O2 SEN, "E" SERIES PCA, INTERFACE, ETHERNET, E-SERIES ASSY, HTR, BYPASS MANIFOLD, M200EH ASSY, BYPASS MANIFOLD, M200EM (KB) OPTION, O2 SENSOR ASSY, M200EX (KB) ASSY, THERMISTOR, BYPASS MANIFOLD ASSY, VALVES, MOLY/HICON, M200EM/H MANUAL, OPERATORS, M200EH/EM PCA, RELAY CARD, M100E/200E ASSY, ORIFICE HOLDER, 7 MIL, OZONE FLOW ASSY, ORIFICE HOLDER, BYPASS MANIFOLD AKIT, EXP KIT, EXHAUST CLNSR, SILCA GEL PCA, TEC CONTROL, E SERIES B-7 M200EM Spare Parts List (Ref: 05483S) Part Number 049760300 050610700 050610900 050611100 051210000 051990000 052930200 054250000 055740000 055740100 055740200 057660000 058021100 059940000 061400000 062390000 062420200 062870000 063530100 064540000 064540100 064540200 065190000 CN0000458 CN0000520 DS0000025 FL0000001 FL0000003 FM0000004 FT0000010 HW0000005 HW0000020 HW0000030 HW0000036 HW0000099 HW0000101 HW0000453 KIT000095 KIT000219 KIT000231 KIT000253 KIT000254 OP0000030 OP0000033 OR0000001 OR0000002 OR0000025 OR0000027 OR0000034 B-8 Description ASSY, TC PROG PLUG, MOLY,TYP K, TC1 CONFIGURATION PLUGS, 115V, M200E CONFIGURATION PLUGS, 220-240V, M200E CONFIGURATION PLUGS, 100V, M200E ASSY, OZONE DESTRUCTOR ASSY, SCRUBBER, INLINE, PUMP PACK ASSY, BAND HEATER TYPE K, M200EX OPTION, CO2 SENSOR (20%) ASSY, PUMP, NOx PUMP PACK, 115V/60HZ ASSY, PUMP, NOx PUMP PACK, 220V/60HZ ASSY, PUMP, NOx PUMP PACK, 220V/50HZ ASSY, DFU FILTER, M703E PCA, E-SERIES MOTHERBD, GEN 5-ICOP OPTION, SAMPLE GAS CONDITIONER, M200A/E ASSY, DUAL HTR, MINI-HICON, 120/240VAC ASSY, MOLY GUTS w/WOOL, M101E/M200EX PCA, SER INTRFACE, ICOP CPU, E- (OPTION) CPU, PC-104, VSX-6150E, ICOP *(KB) DOM, w/SOFTWARE, M200EM * ASSY, PUMP NOX INTERNAL, 115V/60HZ ASSY, PUMP NOX INTERNAL, 230V/60HZ ASSY, PUMP NOX INTERNAL, 230V/50HZ ASSY, NOX CELL TOP-FLO, M200EM >S/N417 CONNECTOR, REAR PANEL, 12 PIN CONNECTOR, REAR PANEL, 10 PIN DISPLAY, E SERIES (KB) FILTER, FLOW CONTROL FILTER, DFU (KB) FLOWMETER (KB) FITTING, FLOW CONTROL FOOT, CHASSI/PUMP PACK SPRING, FLOW CONTROL ISOLATOR, SENSOR ASSY TFE TAPE, 1/4" (48 FT/ROLL) STANDOFF, #6-32X.5, HEX SS M/F ISOLATOR, PUMP PACK SUPPORT, CIRCUIT BD, 3/16" ICOP AKIT, REPLACEMENT COOLER, A/E SERIES KIT, 4-20MA CURRENT OUTPUT (E SERIES) KIT, RETROFIT, M200E/EM/EH Z/S VALVE ASSY & TEST, SPARE PS37, E SERIES ASSY & TEST, SPARE PS38, E SERIES OXYGEN TRANSDUCER, PARAMAGNETIC CO2 MODULE, 0-20% ORING, FLOW CONTROL ORING, REACTION CELL SLEEVE ORING, 2-133V ORING, COLD BLOCK/PMT HOUSING & HEATSINK ORING, (USED W/FT10) 04521C (DCN5731) M200EM Spare Parts List (Ref: 05483S) Part Number OR0000039 OR0000044 OR0000083 OR0000086 OR0000094 OR0000101 PU0000005 PU0000011 PU0000052 PU0000054 PU0000083 RL0000015 SW0000051 SW0000059 WR0000008 04521C (DCN5731) Description ORING, FLOW CONTROL ORING, REACTION CELL MANIFOLD ORING, PMT SIGNAL & OPTIC LED ORING, 2-006, CV-75 COMPOUND(KB) ORING, SAMPLE FILTER ORING, CO2 OPTION PUMP, THOMAS 607, 115V/60HZ (KB) REBUILD KIT, THOMAS 607(KB) PUMP, THOMAS 688, 220/240V 50HZ/60HZ PUMP, THOMAS 688, 100V, 50/60HZ KIT, REBUILD, PU80, PU81, PU82 RELAY, DPDT, (KB) SWITCH, POWER CIRC BREAK VDE/CE, w/RG(KB PRESSURE SENSOR, 0-15 PSIA, ALL SEN POWER CORD, 10A B-9 M200EM Recommended Spare Parts Stocking Levels (Ref: 04415N) Part Number 011310000 011930000 014080100 040010000 040030800 040400000 040420200 041800500 042580000 Description ASSY, DRYER, NOX CD, PMT (R928), NOX, M200A, M200E(KB) ASSY, HVPS, SOX/NOX ASSY, FAN REAR PANEL, E SERIES PCA, PRESS SENSORS (2X), FLOW, E (NOX) ASSY, HEATERS/THERMAL SWITCH, RX CELL ASSY, O3 GEN BRK, M200E, HIGH-O/P PCA, PMT PREAMP, VR, M200E/EM/EH PCA, KEYBOARD, E-SERIES, W/V-DETECT 042680100 ASSY, VALVE (SS), M200E 044440000 044610000 045230200 045500200 058021100 059940000 062870000 DS0000025 FM0000004 KIT000095 KIT000129 ASSY, HICON w/O3 DEST, M200EH/EM ASSY, VALVES, MOLY/HICON, M200E PCA, RELAY CARD, M100E/200E ASSY, ORIFICE HOLDER, 7 MIL PCA, E-SERIES MOTHERBD, GEN 5-ICOP OPTION, SAMPLE GAS CONDITIONER, M200A/E CPU, PC-104, VSX-6150E, ICOP *(KB) DISPLAY, E SERIES (KB) FLOWMETER (KB) AKIT, REPLACEMENT COOLER, A/E SERIES REPLACEMENT, MOLY CONV WELDED CARTRIDGE OP0000030 OXYGEN TRANSDUCER, PARAMAGNETIC OR0000034 OR0000044 OR0000045 PS0000037 PS0000038 PU0000005 RL0000015 ORING, 2-011V FT10 ORING, 2-125V ORING, 2-226V PS, 40W SWITCHING, +5V, +/-15V(KB) * PS, 60W SWITCHING, 12V(KB) * PUMP, THOMAS 607, 115V/60HZ (KB) RELAY, DPDT, (KB) B-10 1 2-5 6-10 1 1 1 1 2 1 2 1 2 1 1 1 1 2 1 1 1 1 2 1 1 2 2 2 1 1 1 1 11-20 3 1 1 4 2 2 21-30 UNITS 1 1 4 1 1 4 3 3 1 1 1 2 1 1 1 2 1 1 1 1 2 3 1 4 2 2 2 4 2 2 1 1 3 3 1 1 5 5 5 1 1 1 2 1 10 10 10 2 2 1 3 With IZS, ZS Option With O2 Option 04521C (DCN5731) Part Number 018080000 002270100 009690300 046030000 FL0000001 FL0000003 HW0000020 OR0000086 OR0000034 OR0000039 04521C (DCN5731) Description KIT, DESSICANT BAGGIES (12) KIT, WINDOW GASKET (12) KIT, TFE FILTER ELEMENTS, 47MM, 1UM (30) KIT, CH-43, 3 REFILLS FILTER, SS FILTER, DFU SPRING ORING, FLOW CONTROL ORING, FLOW CONTROL ORING, FLOW CONTROL Quantity M200E M200EM/EH "00" "01" 1 1 1 1 4 1 4 8 2 2 1 1 1 4 1 4 8 2 2 B-11 This page intentionally left blank. B-12 04521C (DCN5731) Warranty/Repair Questionnaire Model 200EH/EM Appendix C (Ref: 05149A) TELEDYNE INSTRUMENTS Advanced Pollution Instrumentation A Teledyne Technologies Company CUSTOMER:_____________________________________ PHONE: ________________________________ CONTACT NAME: ________________________________ FAX NO. _______________________________ SITE ADDRESS:_____________________________________________________________________________ MODEL TYPE: ______________ SERIAL NO.: _________________ FIRMWARE REVISION: ___________ 1. Are there any failure messages? ______________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ ________________________________________________________________ (Continue on back if necessary) PLEASE COMPLETE THE FOLLOWING TABLE: TEST FUNCTION RECORDED VALUE NOx STAB UNITS ACCEPTABLE VALUE PPB/PPM 1 PPB WITH ZERO AIR 3 SAMPLE FLOW CM OZONE FLOW CM3 80 ± 15 PMT SIGNAL WITH ZERO AIR MV -20 to 150 MV 0-5000MV 1 2 0-5,000 PPM , 200 PPM PMT SIGNAL AT SPAN GAS CONC PPB 500 ± 50 NORM PMT SIGNAL AT SPAN GAS CONC MV PPB 0-5000MV 1 2 0-5,000 PPM , 200 PPM AZERO MV -20 to 150 HVPS V 400 to 900 RCELL TEMP ºC 50 ± 1 BOX TEMP ºC AMBIENT ± 5ºC PMT TEMP ºC 7 ± 2ºC ºC 30ºC to 70ºC 3 O2 CELL TEMP 3 IZS TEMP ºC 50 ± 1ºC MOLY TEMP ºC 315 ± 5ºC RCEL IN-HG-A <10 SAMP IN-HG-A AMBIENT ± 1 NOx SLOPE 1.0 ± 0.3 NOx OFFSET mV 50 to 150 NO SLOPE 1.0 ± 0.3 NO OFFSET mV 50 to 150 3 O2 SLOPE O2 OFFSET 0.5 to 2.0 3 % -10 to + 10 PMT SIGNAL DURING ETEST MV PMT SIGNAL DURING OTEST MV 2000 ± 1000 MV 4096mv ±2mv and Must be Stable MV 0± 0.5 and Must be Stable REF_4096_MV 4 REF_GND4 1 2 M200EH M200EM 4 Located in Signal I/O list under DIAG menu 3 2000 ± 1000 If option is installed TELEDYNE INSTRUMENTS CUSTOMER SERVICE EMAIL: api-customerservice@teledyne.com PHONE: (858) 657-9800 TOLL FREE: (800) 324-5190 FAX: (858) 657-9816 C-1 04521C (DCN5731) Appendix C (Ref: 05149A) Warranty/Repair Questionnaire Model 200EH/EM TELEDYNE INSTRUMENTS Advanced Pollution Instrumentation A Teledyne Technologies Company 2. What is the rcell & sample pressures with the sample inlet on rear of machine capped? RCELL PRESS - __________________ IN-HG-A SAMPLE PRESSURE: _______________ IN-HG-A 3. What are the failure symptoms? ______________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ 4. What test have you done trying to solve the problem? _____________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ 5. If possible, please include a portion of a strip chart pertaining to the problem. Circle pertinent data. Thank you for providing this information. Your assistance enables Teledyne Instruments to respond faster to the problem that you are encountering. OTHER NOTES: ____________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ C-2 TELEDYNE INSTRUMENTS CUSTOMER SERVICE EMAIL: api-customerservice@teledyne.com PHONE: (858) 657-9800 TOLL FREE: (800) 324-5190 FAX: (858) 657-9816 04521C (DCN5731) Model 200EH/EM (Ref 05150D) APPENDIX D: Diagrams and Schematics APPENDIX D: Diagrams and Schematics Table D-1: List of Included Diagrams and Schematics DOCUMENT # DOCUMENT TITLE 04504 Document, M200EH/EM Electronic Interconnect Diagram 04496 Document, M200EH/EM Electronic Interconnect Listing 01669 PCA 016680300, Ozone generator board 01840 PCA Thermo-electric cooler board 03632 PCA 03631, 0-20mA Driver 03956 PCA 039550200, Relay Board 05703 PCA 05702, Motherboard, E-series, Gen 4 04259 PCA 04258, Keyboard & Display Driver 04354 PCA 04003, Pressure/Flow Transducer Interface 04395 PCA 4394, Ethernet board (optional equipment) 04181 PCA 041800200, PMT pre-amplifier board 04468 PCA, 04467, Analog Output E-Series USER NOTES: 04521C (DCN5731) D-1 APPENDIX D: Diagrams and Schematics D-1 Model 200EH/EM (05150D) 04521C (DCN5731) M200E INTERCONNECT LIST (Ref: 04496D) CONNECTION FROM Cable Part Signal Assembly PN J/P Pin # 00729 CBL, KEYBOARD/DISPLAY D7 Display DS0000025 CN1 1 D6 Display DS0000025 CN1 2 D5 Display DS0000025 CN1 3 D4 Display DS0000025 CN1 4 D3 Display DS0000025 CN1 5 D2 Display DS0000025 CN1 6 D1 Display DS0000025 CN1 7 D0 Display DS0000025 CN1 8 DISP WRITE Display DS0000025 CN1 9 DGND Display DS0000025 CN1 10 Spare Display DS0000025 CN1 11 DISP_BUSY Display DS0000025 CN1 12 DISP_RETURN Display DS0000025 CN1 13 DISP_RETURN Display DS0000025 CN1 14 DISP_PWR Display DS0000025 CN1 15 DISP_PWR Display DS0000025 CN1 16 0364901 CBL, AC Power, E-series AC Line Power Entry CN0000073 L AC Neutral Power Entry CN0000073 N Power Grnd Power Entry CN0000073 Power Grnd Power Entry CN0000073 AC Line Switched Power Switch SW0000051 L AC Neutral Switched Power Switch SW0000051 N Power Grnd Power Entry CN0000073 AC Line Switched Power Switch SW0000051 L AC Neutral Switched Power Switch SW0000051 N Power Grnd Power Entry CN0000073 AC Line Switched Power Switch SW0000051 L AC Neutral Switched Power Switch SW0000051 N Power Grnd Power Entry CN0000073 03829 CBL, DC power to motherboard, E-series DGND Relay Board 045230100 P7 1 +5V Relay Board 045230100 P7 2 AGND Relay Board 045230100 P7 3 +15V Relay Board 045230100 P7 4 AGND Relay Board 045230100 P7 5 -15V Relay Board 045230100 P7 6 +12V RET Relay Board 045230100 P7 7 +12V Relay Board 045230100 P7 8 Chassis Gnd Relay Board 045230100 P7 10 04021 CBL, Preamp, O2 sensor, O3 generator, fan, relay board, motherboard, M200E DGND Relay Board 045230100 P12 1 +5V Relay Board 045230100 P12 2 +15V Relay Board 045230100 P12 4 AGND Relay Board 045230100 P12 3 +12V Relay Board 045230100 P12 8 +12V RET Relay Board 045230100 P12 7 O3GEN enable signal Ozone generator 040420200 P1 6 ETEST Motherboard 057020100 P108 8 OTEST Motherboard 057020100 P108 16 PHYSICAL RANGE Motherboard 057020100 P108 7 PMT TEMP Preamplifier board 041800500 P6 5 HVPS Preamplifier board 041800500 P6 6 PMT SIGNAL+ Preamplifier board 041800500 P6 7 AGND Preamplifier board 041800500 P6 S AGND Motherboard 057020100 P109 9 O2 SIGNAL Motherboard 057020100 P109 7 O2 SIGNAL + Motherboard 057020100 P109 1 DGND O2 Sensor (optional) OP0000030 P1 5 +5V O2 Sensor (optional) OP0000030 P1 6 04521C (DCN5731) Assembly CONNECTION TO PN J/P J3 J3 J3 J3 J3 J3 J3 J3 J3 J3 J3 J3 J3 J3 J3 J3 Pin Keyboard/Interface Keyboard/Interface Keyboard/Interface Keyboard/Interface Keyboard/Interface Keyboard/Interface Keyboard/Interface Keyboard/Interface Keyboard/Interface Keyboard/Interface Keyboard/Interface Keyboard/Interface Keyboard/Interface Keyboard/Interface Keyboard/Interface Keyboard/Interface 042580000 042580000 042580000 042580000 042580000 042580000 042580000 042580000 042580000 042580000 042580000 042580000 042580000 042580000 042580000 042580000 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 Power Switch Power Switch Shield Chassis PS2 (+12) PS2 (+12) PS2 (+12) PS1 (+5, ±15) PS1 (+5, ±15) PS1 (+5, ±15) Relay Board Relay Board Relay Board SW0000051 SW0000051 SW0000051 042190000 PS0000038 PS0000038 PS0000038 PS0000037 PS0000037 PS0000037 045230100 045230100 045230100 SK2 SK2 SK2 SK2 SK2 SK2 J1 J1 J1 1 3 2 1 3 2 1 3 2 Motherboard Motherboard Motherboard Motherboard Motherboard Motherboard Motherboard Motherboard Motherboard 057020100 057020100 057020100 057020100 057020100 057020100 057020100 057020100 057020100 P15 P15 P15 P15 P15 P15 P15 P15 P15 1 2 3 4 5 6 7 8 9 Ethernet board Ethernet board Ozone generator Ozone generator PMT cooling fan PMT cooling fan Motherboard Preamplifier board Preamplifier board Preamplifier board Motherboard Motherboard Motherboard Motherboard O2 Sensor (optional) O2 Sensor (optional) O2 Sensor (optional) Relay Board Relay Board 043940000 043940000 040420200 040420200 013140000 013140000 057020100 041800500 041800500 041800500 057020100 057020100 057020100 057020100 OP0000030 OP0000030 OP0000030 045230100 045230100 P102 P102 P1 P1 P1 P1 P108 P6 P6 P6 P109 P109 P109 P109 P1 P1 P1 P5 P5 1 2 4 5 1 2 15 1 2 4 4 5 6 11 S 9 10 1 2 L N D-3 M200E INTERCONNECT LIST (Ref: 04496D) CONNECTION FROM CONNECTION TO Cable Part Signal Assembly PN J/P Pin Assembly PN # 04022 CBL, DC Power, fan, keyboard, TEC, sensor board, M200E TEC +12V TEC board 049310100 P1 1 Relay Board 045230100 TEC +12V RET TEC board 049310100 P1 2 Relay Board 045230100 DGND Relay Board 045230100 P10 1 Keyboard 042580000 +5V Relay Board 045230100 P10 2 Keyboard 042580000 DGND Keyboard 042580000 P1 2 Relay Board 045230100 +5V Keyboard 042580000 P1 3 Relay Board 045230100 +12V RET Relay Board 045230100 P11 7 Chassis fan 040010000 +12V Relay Board 045230100 P11 8 Chassis fan 040010000 P/Flow Sensor AGND Relay Board 045230100 P11 3 P/Flow Sensor board 040030800 P/Flow Sensor +15V Relay Board 045230100 P11 4 P/Flow Sensor board 040030800 Pressure signal 1 P/Flow Sensor board 040030800 P1 2 Motherboard 057020100 Pressure signal 2 P/Flow Sensor board 040030800 P1 4 Motherboard 057020100 Flow signal 1 P/Flow Sensor board 040030800 P1 5 Motherboard 057020100 Flow signal 2 P/Flow Sensor board 040030800 P1 1 Motherboard 057020100 Shield P/Flow Sensor board 040030800 P1 S Motherboard 057020100 Shield Motherboard 057020100 P110 9 Relay Board 045230100 Thermocouple signal 1 Motherboard 057020100 P110 2 Relay Board 045230100 TC 1 signal DGND Motherboard 057020100 P110 8 Relay Board 045230100 Thermocouple signal 2 Motherboard 057020100 P110 1 Relay Board 045230100 TC 2 signal DGND Motherboard 057020100 P110 7 Relay Board 045230100 04023 CBL, I2C, relay board to motherboard, E-series I2C Serial Clock Motherboard 057020100 P107 3 Relay Board 045230100 I2C Serial Data Motherboard 057020100 P107 5 Relay Board 045230100 I2C Reset Motherboard 057020100 P107 2 Relay Board 045230100 I2C Shield Motherboard 057020100 P107 6 Relay Board 045230100 CBL, Nox, zero/span, IZS valves, M200E 04024 Zero/Span valve +12V Relay Board 045230100 P4 1 Zero/Span valve 042680100 Zero/Span valve +12V RET Relay Board 045230100 P4 2 Zero/Span valve 042680100 Sample valve +12V Relay Board 045230100 P4 3 Sample valve 042680100 Sample valve +12V RET Relay Board 045230100 P4 4 Sample valve 042680100 AutoZero valve +12V Relay Board 045230100 P4 5 AutoZero valve 042680100 AutoZero valve +12V RET Relay Board 045230100 P4 6 AutoZero valve 042680100 NONOx valve +12V Relay Board 045230100 P4 7 NONOx valve 042680100 NONOx valve +12V RET Relay Board 045230100 P4 8 NONOx valve 042680100 0402603 CBL, IZS & O2 sensor heaters/thermistors; reaction cell & manifold thermistors, M200E Rcell thermistor A Reaction cell thermistor 041920000 P1 2 Motherboard 057020100 Rcell thermistor B Reaction cell thermistor 041920000 P1 1 Motherboard 057020100 IZS thermistor A Motherboard 057020100 P27 6 IZS thermistor/heater 003290000 IZS thermistor B Motherboard 057020100 P27 13 IZS thermistor/heater 003290000 IZS heater L IZS thermistor/heater 003290000 P1 4 Relay Board 045230100 IZS heater N IZS thermistor/heater 003290000 P1 1 Relay Board 045230100 Shield Relay Board 045230100 O2 sensor heater Relay Board 045230100 P18 6 O2 sensor therm./heater 043420000 O2 sensor heater Relay Board 045230100 P18 7 O2 sensor therm./heater 043420000 Shield Relay Board 045230100 P18 12 O2 sensor therm./heater 043420000 O2 sensor thermistor A O2 sensor therm./heater 043420000 P1 3 Motherboard 057020100 O2 sensor thermistor B O2 sensor therm./heater 043420000 P1 1 Motherboard 057020100 Byp/dil. man. thermistor A Motherboard 057020100 P27 1 Manifold thermistor 044530000 Byp/dil. man. thermistor B Motherboard 057020100 P27 8 Manifold thermistor 044530000 Configuration jumper intern. Relay Board 045230100 P18 3 Relay Board 045230100 Configuration jumper intern. Relay Board 045230100 P18 8 Relay Board 045230100 04027 CBL, NO2 converter, reaction cell & manifold heaters, M200E Bypass/dil. manifold heater L Manifold heater 1 044340000 P1 1 Relay Board 045230100 Bypass/dil. manifold heater N Manifold heater 1 044340000 P1 2 Relay Board 045230100 Bypass/dil. manifold heater L Relay Board 045230100 P2 11 Manifold heater 2 044340000 Bypass/dil. manifold heater N Relay Board 045230100 P2 15 Manifold heater 2 044340000 Moly heater A Relay Board 045230100 P2 7 Moly heater A 039700100 Moly heater C Relay Board 045230100 P2 6 Moly heater C 039700100 Moly heater B Relay Board 045230100 P2 10 Moly heater B 039700100 Configuration jumper intern. Relay Board 045230100 P2 13 Relay Board 045230100 Configuration jumper intern. Relay Board 045230100 P2 8 Relay Board 045230100 Reaction cell heater/switch Relay Board 045230100 P2 1 Reaction cell heater 1B 040400000 Reaction cell heater/switch Relay Board 045230100 P2 1 Reaction cell heater 2B 040400000 Reaction cell heater/switch Relay Board 045230100 P2 2 Reaction cell heater 1A 040400000 Reaction cell heater/switch Relay Board 045230100 P2 3 Reaction cell heat switch 040400000 Reaction cell heater/switch Relay Board 045230100 P2 4 Reaction cell heat switch 040400000 Reaction cell heater/switch Relay Board 045230100 P2 5 Reaction cell heater 2A 040400000 D-3 J/P Pin P10 P10 P1 P1 P11 P11 P1 P1 P1 P1 P110 P110 P110 P110 P110 P17 P17 P17 P17 P17 8 7 8 1 1 2 1 2 3 6 6 5 4 3 12 S 1 2 3 4 P3 P3 P3 P3 1 2 4 5 P1 P1 P1 P1 P1 P1 P1 P1 1 2 1 2 1 2 1 2 P27 P27 P1 P1 P18 P18 P18 P1 P1 P1 P27 P27 P1 P1 P18 P18 7 14 2 3 1 2 11 4 2 P2 P2 P1 P1 P1 P1 P1 P2 P2 P1 P1 P1 P1 P1 P1 11 12 1 2 1 2 3 14 9 4 6 3 1 2 5 4 11 1 2 4 9 04521C (DCN5731) M200E INTERCONNECT LIST (Ref: 04496D) CONNECTION FROM Cable Part Signal Assembly PN # 04105 CBL, Keyboard, display to motherboard, E-series Kbd Interrupt Keyboard 042580000 DGND Keyboard 042580000 SDA Keyboard 042580000 SCL Keyboard 042580000 Shld Keyboard 042580000 04176 CBL, DC power to relay board, E-series DGND Relay Board 045230100 +5V Relay Board 045230100 +15V Relay Board 045230100 AGND Relay Board 045230100 -15V Relay Board 045230100 +12V RET Relay Board 045230100 +12V Relay Board 045230100 04211 CBL, Serial data, motherboard to CPU, E-series RXD(0) CPU board CP0000026 RTS(0) CPU board CP0000026 TXD(0) CPU board CP0000026 CTS(0) CPU board CP0000026 GND(0) CPU board CP0000026 RXD(1) CPU board CP0000026 RTS(1) CPU board CP0000026 TXD(1) CPU board CP0000026 CTS(1) CPU board CP0000026 GND(1) CPU board CP0000026 NET+ CPU board CP0000026 NETCPU board CP0000026 GND CPU board CP0000026 Shield CPU board CP0000026 04339 CBL, CPU to Ethernet (optional), E-series Ethernet DCD CPU board CP0000026 Ethernet DSR CPU board CP0000026 Ethernet RXD CPU board CP0000026 Ethernet RTS CPU board CP0000026 Ethernet TXD CPU board CP0000026 Ethernet CTS CPU board CP0000026 Ethernet DTR CPU board CP0000026 Ethernet GND CPU board CP0000026 Ground CPU board CP0000026 04433 CBL, preamplifier to relay board, M200E Preamplifier DGND Relay Board 045230100 Preamplifier +5V Relay Board 045230100 Preamplifier AGND Relay Board 045230100 Preamplifier +15V Relay Board 045230100 Preamplifier -15V Relay Board 045230100 04437 CBL, preamplifier to TEC, M200E Preamp TEC drive VREF Preamplifier board 041800500 Preamp TEC drive CTRL Preamplifier board 041800500 Preamp TEC drive AGND Preamplifier board 041800500 04521C (DCN5731) J/P Pin Assembly J2 J2 J2 J2 J2 7 2 5 6 10 Motherboard Motherboard Motherboard Motherboard Motherboard P8 P8 P8 P8 P8 P8 P8 1 2 4 5 6 7 8 CN3 CN3 CN3 CN3 CN3 CN4 CN4 CN4 CN4 CN4 CN5 CN5 CN5 CN5 CONNECTION TO PN J/P Pin 057020100 057020100 057020100 057020100 057020100 J106 J106 J106 J106 J106 1 8 2 6 5 Power Supply Triple Power Supply Triple Power Supply Triple Power Supply Triple Power Supply Triple Power Supply Single Power Supply Single PS0000037 PS0000037 PS0000037 PS0000037 PS0000037 PS0000038 PS0000038 J1 J1 J1 J1 J1 J1 J1 3 1 6 4 5 3 1 3 4 5 6 9 3 4 5 6 9 2 4 6 Motherboard Motherboard Motherboard Motherboard Motherboard Motherboard Motherboard Motherboard Motherboard Motherboard Motherboard Motherboard Motherboard Motherboard 057020100 057020100 057020100 057020100 057020100 057020100 057020100 057020100 057020100 057020100 057020100 057020100 057020100 057020100 J12 J12 J12 J12 J12 J12 J12 J12 J12 J12 J12 J12 J12 J12 14 13 12 11 10 9 8 7 6 5 9 7 5 2 CN4 CN4 CN4 CN4 CN4 CN4 CN4 CN4 CN4 1 2 3 4 5 6 7 9 Ethernet board Ethernet board Ethernet board Ethernet board Ethernet board Ethernet board Ethernet board Ethernet board Ethernet board 043940000 043940000 043940000 043940000 043940000 043940000 043940000 043940000 043940000 P101 P101 P101 P101 P101 P101 P101 P101 P101 6 4 3 10 8 5 9 16 2 P9 P9 P9 P9 P9 1 2 3 4 6 Preamplifier board Preamplifier board Preamplifier board Preamplifier board Preamplifier board 041800500 041800500 041800500 041800500 041800500 P5 P5 P5 P5 P5 1 2 3 4 6 J1 J1 J1 1 2 3 TEC board TEC board TEC board 049310100 049310100 049310100 J3 J3 J3 1 2 3 D-5 D-6 04521C (DCN5731) 1 2 3 4 6 5 VERSION TABLE 016680000 - CE MARK VERSION STD PROD. VERSION UP TO 10/99 016680100 - NON CE MARK (OBSOLETE) +15V 016680200 - SUB PS 17 SWITCHER FOR LINEAR SUPPLY DELETE COMPONENTS T1, D1, D2, C9, C11, PTC1, PTC2, U2 ADD COMPONENTS PS1 +15V +15V D R1 R5 TP1 016680300 - LOW OUTPUT + FIXED FREQ REPLACE VR2 WITH A WIRE JUMPER REPLACE R4 WITH RS297 127KOHM 1.2K 4.7K 1% +15V TP6 R6 + R2 10K 1% Q1 IRFZ924 C2 .01 L1 C7 J2 1000uF/25V 1 2 3 4 .1 10 16 2 9 6 7 1 4 C3 .1 VR2 100K "FREQ" 6 5 4 3 2 1 VIN C_B C_A E_B E_A OSC -SEN GND SD VREF INV+ COMP RT CT INV+SEN C5 .1 J1 68uH TP2 U1 C 016680400 - HI OUTPUT + FIXED FREQ REPLACE VR2 WITH A WIRE JUMPER REPLACE R4 WITH RS13 11 KOHM 10 C1 15 13 12 14 11 3 5 8 Q2 IRFZ24 R7 + 016680600 - HI OUTPUT,E SERIES DELETE COMPONENTS T1,D1,D2,C9,PTC1,PTC2,U2 C8 1000uF/25V 10 R8 1.2K C SG3524B + D C6 100pF R10 C4 4.7uF/16V 3K TP3 Text R11 150K R4 10K 1% TP4 115V 15V B 1 1.1A 1N4007 IN Text R9 3 OUT .1 R13 10K 1% R12 7 6 2 3 115V D1 8 GND PTC2 T1 + C9 2200uF/35V 10K 1% 2 1 +15V TP5 LM7815 U2 C10 .1 C11 Text B 15V 5 4 PWR XFRMR PTC1 D2 1.1A 1N4007 .22 R14 VR1 1K 20T 4.7K 1% "PW" C12 .22 R15 4.7K 1% Error : LOGO.BMP file not found. 10/15/96 REV. D: Added PTC1,2 secondary overcurrent protection. 11/21/96 REV. E: Minor cosmetic fixes The information herein is the property of API and is submitted in strictest confidence for reference only. Unauthorized use by anyone for any other purposes is prohibited. This document or any information contained in it may not be duplicated without proper authorization. A 10/01/99 REV. F 1 04521C (DCN5731) 2 ADDED VERSION TABLE AT D6 3 4 5 APPROVALS DATE OZON_ GEN A DRAWN DRIVER CHECKED SIZE B APPROVED DRAWING NO. REVISION 01669 G LAST MOD. SHEET 30-Nov-2006 1 of 1 6 D-7 1 2 3 4 5 6 D 1 2 D 2 7 2 5 1 4 +15 4 6 +15 3 5 1 6 7 +15 +15 3 +15 1 2 2 C 5 7 +15 +15 10 4 6 3 1 1 2 3 4 12 8 11 11 9 8 67 8 12 8 11 8 32 7 10 5 9 2 6 1 4 5 9 3 8 3 2 1 10 1 C B B 12 4 67 14 13 +15 +15 5 + +15 + The information herein is the property of API and is submitted in strictest confidence for reference only. Unauthorized use by anyone for any other purposes is prohibited. This document or any information contained in it may not be duplicated without proper authorization. A 1 D-8 +15 2 3 4 5 APPROVALS DRAW N DATE T H E R M O E L E C T A COOLER_CONTROL CH ECKED SIZ E DRAW ING NO. APPROVED LAST MOD. REVISION B 01840 B 14-Jul-1999 SH EET 1 of 1 6 04521C (DCN5731) 1 2 3 4 6 5 D 1 0.1 C4 1000PF U4 U3 ISO_-15V +12V 9 C6 ISO_+15V D 15 12 11 VOUT 7 4 VIN(10) GATEDRV U2 2 R1 R2 4.75K 9.76K GND TP6 C5 220PF 3 5 6 3 OPA277 8 +VS2 VIN 15 TESTPOINT TP1 7 1 TESTPOINT TP2 +V SR SSENSE 4 +VS1 VREF SENSE VRADJ 2 D1 1N914 OFFADJ OFFADJ SPAN 4MA 16MA VREFIN VIN(5V) GND 16 1 ISO_+15V 13 14 Q1 MOSFETP 7 6 8 10 9 IOUT+ XTR110 J1 +12V -VS1 GND1 -VS2 GND2 C7 0.1 -12V C ISO_+15V HEADER 4X2 IOUT- VINVIN+ ISO124 10 8 2 4 6 8 -12V ISO_-15V +15V 1 3 5 7 2 C 16 IOUTIOUT+ +15V U1 C1 0.47 ISO+15 TP3 1 2 5 6 7 ISO_+15V ISO_GND TP5 B C2 0.47 ISO_GND ISO_-15V 0V +VOUT -VOUT SIN SOUT 14 8 B DCP010515 C3 0.47 VIN- TP4 ISO-15 VS 0V JP1 JUMPER2 Error : LOGO.BMP file not found. A 04521C (DCN5731) 1 2 Date Rev. Change Description Engineer 8/9/00 A INITIAL RELEASE (FROM 03039) KL 3 4 The information herein is the property of API and is submitted in strictest confidence for reference only. Unauthorized use by anyone for any other purposes is prohibited. This document or any information contained in it may not be duplicated without proper authorization. 5 APPROVALS DATE PCA 03631, Isolated 0-20ma, E Series A DRAWN CHECKED SIZE B APPROVED DRAWING NO. REVISION 03632 A LAST MOD. SHEET 19-Jul-2002 1 6 of D-9 1 1 2 J1 1 2 3 4 4 PIN D 3 4 6 5 General Trace Width Requirements 1. Vcc (+5V) and I2C VCC should be 15 mil 2. Digitial grounds should be at least 20 mils 3. +12V and +12V return should be 30 mils 4. All AC lines (AC Line, AC Neutral, RELAY0 - 4, All signals on JP2) should be 30 mils wide, with 120 mil isolation/creepage distance around them 5. Traces between J7 - J12 should be top and bottom and at least 140 mils. 6. Traces to the test points can be as small as 10 mils. AC_Line AC_Neutral RELAY0 VCC RN1 330 R1 R2 2.2K 2.2K RELAY1 RELAY0 K1 9 1 4 3 2 1 4 3 K3 JP2 Heater Config Jumper 2 COMMON0 LOAD0 TS0 RELAY0 RELAY2 1 2 3 4 5 6 7 8 9 10 11 12 2 K2 RELAY2 I2C_Vcc 10 8 7 6 5 4 3 I2C_Vcc 2 1 1 JP1 1 2 3 4 5 6 7 8 HEADER 4X2 D RELAY1 3 +- SLD-RLY +- 4 TS0 TS1 TS2 SLD-RLY COMMON1 LOAD1 TS1 RELAY1 A SLD-RLY +- YEL RL0 YEL RL1 D8 D9 YEL RL2 GRN VA0 GRN VA1 GRN VA2 D10 GRN VA3 1 IO10 IO11 IO12 IO13 IO14 IO15 2 SN74HC04 VCC U2B Q1 VCC 4 11 3 R5 10K JP4 1 2 3 C3 1 U2D R6 10K 1 11 CON10THROUGH 2 1 C6 2000/25 VCC J11 1 SPARE J10 1 2 3 4 5 6 7 8 9 10 14 TP1 TP2 TP3 TP4 TP5 TP6 TP7 DGND +5V AGND +15V -15V +12RT +12V 1 SYNC DEMOD J9 1 2 3 4 5 6 7 8 9 10 U2F REV B J12 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9 10 13 + AUTH CAC DATE 10/3/02 CE MARK LINE VOLTAGE TRACE SPACING FIX 12 A Title CON10THROUGH CON10THROUGH CON10THROUGH 3 Te T D-10 8 PIN 10 1 CON10THROUGH VLV_ENAB U2E 1 CON10THROUGH 8 + 1 MTHR BRD J8 1 2 3 4 5 6 7 8 9 10 B VALVE3 7 A KEYBRD J7 1 2 3 4 5 6 7 8 9 10 VALVE2 2 1 C5 10/16 2 DC PWR IN J5 DGND 1 VCC 2 AGND 3 +15V 4 AGND 5 -15V 6 +12RET 7 +12V 8 EGND 9 CHS_GND 10 CON10THROUGH VALVE1 2 + C4 10/16 1 R4 1M 2 1 MAX693 AK C2 0.001 D17 RLS4148 VALVE0 UDN2540B(16) 9 A JP3 1 2 HEADER 1X2 J4 1 2 3 4 5 6 7 8 WTCDG OVR 1 16 15 14 13 12 11 10 9 K VBATT RESET VOUT RESET' VCC WDO' GND CD IN' BATT_ONCD OUT' LOW LINE' WDI OSC IN PFO' OSC SEL PFI 6 IN 4 OUT4 IN 3 K ENABLE OUT 3 IN 2 OUT 2 IN 1 K OUT 1 VCC U2C I2C_Vcc IRF7205 1 2 3 6 7 8 GND GND GND GND U4 1 2 3 4 5 6 7 8 +12V U5 16 15 14 10 9 13 12 5 4 R3 20K VCC C U2A 5 B COMMON2 LOAD2 TS2 RELAY2 AC_Neutral IO3 IO4 PCF8575 12 D7 1 P00 P01 P02 P03 P04 SCL P05 SDA P06 P07 P10 P11 P12 P13 P14 P15 P16 P17 Vss 22 23 A0 A1 A2 INT 4 5 6 7 8 9 10 11 13 14 15 16 17 18 19 20 D4 KA 24 J3 1 2 3 4 5 CON5 21 2 3 1 D3 RED U1 Vdd C1 0.1 C D2 K D1 WDOG I2C_Vcc J216 PIN 1 2 RELAY0 3 4 5 6 7 RELAY1 8 9 10 11 12 RELAY2 13 14 15 16 Size B Date: File: APPLIES TO PCB 03954 4 5 M100E/M200E Relay PCB Number 03956 Revision A 3 3 30-Jun-2004 Sheet 1 of N:\PCBMGR\RELEASED\03954cc\PROTEL\03954a.ddb Drawn By: 6 Te T 04521C (DCN5731) 1 2 3 4 5 6 AC_Line J20 1 2 3 4 5 6 RELAY3 RN2 330 D RELAY4 10 9 8 7 6 5 4 3 2 1 RELAY3 1 K4 RELAY4 2 1 4 3 K5 Aux Relay Connector D MOLEX6 2 AC_Neutral I2C_Vcc 3 I2C_Vcc +- SLD-RLY RL3 RL4 VA4 D12 GRN D13 GRN D14 GRN D15 GRN D16 GRN VA5 VA6 VA7 TR0 TR1 C K C D11 GRN KA D6 YEL A SLD-RLY D5 YEL 4 +- IO3 IO4 VCC IO13 +12V 11 U3A 1 SN74HC04 16 15 14 10 9 VLV_ENAB IN 4 OUT4 IN 3 K ENABLE OUT 3 IN 2 OUT 2 IN 1 K OUT 1 GND GND GND GND U3D 9 J6 1 2 3 4 5 6 7 8 9 10 U6 2 VCC IO10 IO11 IO12 8 1 2 3 6 7 8 13 12 5 4 UDN2540B(16) U3B U3E IO14 3 Valve4 Valve5 Valve6 Valve7 CON10 4 11 10 B B U3C 14 VCC U3F 13 IO15 5 6 12 J13 1 2 MINIFIT-2 C13 0.1 7 +12V Q2 IRL3303 Use 50 mil traces +12V J14 1 2 MINIFIT-2 Q3 IRL3303 A A Title Use 40 mil traces Size B Date: File: +12RET 1 04521C (DCN5731) 2 3 Te T 4 Te T 5 100E/200E/400E RELAY PCB Number 03956 Revision A 3 3 30-Jun-2004 Sheet 2 of N:\PCBMGR\RELEASED\03954cc\PROTEL\03954a.ddb Drawn By: 6 D-11 1 2 3 4 R7 2.55K +15V 6 5 VDD_TC ZR1 C15 C7 D 0.1 0.1 +15V D LTC1050 U8 K 1 2 2 4 CCW CW JP5 1 2 JUMPER R13 332K 1K CCW K R17 R19 J17 1 2 3 4 MICROFIT-4 1 10K 5K C C9 0.1 ZR2 5.6V A AK VEE_TC W W C8 0.1 C R15 11K C17 CW R11 249K R9 TYPE k K TC Connector -15V CW 5 4 1 OPA2277 J18 - 2 + 1 ZR3 10V 3 6 TYPE J J TC Connector R21 20k U7A 3 KA C16 0.1 8 7 J15 2 + 1 - 8 A 5.6V R8 2.55K VDD_TC B 8 7 ZR4 LTC1050 U9 U7B 3 6 7 2 J16 - 2 + 1 20k R22 5 6 10V B K -15V KA A C10 0.1 J 8 K 7 R- 5 R14 676K 1K JP6 1 2 JUMPER R16 11K R20 10K R18 Vin Gnd C14 0.1 R10 U10 3 TOUT CW R12 249K 2 TYPE J J TC Connector 5 4 1 OPA2277 C20 1 uF 5K C11 LT1025 4 0.1 C12 0.1 A A VEE_TC Title TYPE K J19 - 2 + 1 K TC Connector Size B Date: File: 1 2 3 Te D-12 4 5 100E/200E/400E RELAY PAB Number 03956 Revision A 3 3 30-Jun-2004 Sheet 3 of N:\PCBMGR\RELEASED\03954cc\PROTEL\03954a.ddb Drawn By: 6 Te 04521C (DCN5731) 3 U2 74HC154 ENAB2 U4B 10 11 D0 12 13 PRE CLK D CLR 4 3 2 1 8 Q U51D AEN IOEN A1 A2 A3 A4 A5 A6 A7 A8 2 3 4 5 6 7 8 9 1 2 3 4 6 7 8 9 Y1 Y2 Y3 Y4 Y5 Y6 Y7 Y8 5 Q 74HC74 TP56 VCC 1.2 uF, 6.3V ceramic 2 4 74HC32 U50A 1 4 5 6 19 INT A0 6 5 17 16 18 U50B 2 1 shorted - sldr side JP4 IRQ10 JP5 8 10 74HC08 U6C 74HC08 12 A15 13 11 10 74HC32 74HC08 JP2 2 1 4 2 U3 LTC699CS8 7 8 9 11 12 13 14 15 D0 D1 D2 D3 D4 D5 D6 D7 47k, 5% R5 VCC VCC 74AHC1GU04 2 1 IOR IOW INLINE-6 J106 KBINT SDA 3 SCL VCC 2 SDA DB0 DB1 DB2 DB3 DB4 DB5 DB6 DB7 6 IDC-HEADER IOR IOW SCL DGND MICROFIT-8 2 10 VSS JP6 1 IDC-HEADER B WDI RESET C3 7 0.15 uF, ceramic I2C_RESET SHDN SHDN IOR IOW U5B 10 11 12 13 U51A 1 +12V 1 2 3 4 5 6 7 8 INT shorted - sldr side JP3 6 VCC 8 U39 Q 1 2 A14 Q 9 U50D VCC PRE CLK D CLR 5 20 VCC GND GND GND 64 63 62 61 60 59 58 57 56 55 54 53 52 51 50 49 48 47 46 45 44 43 42 41 40 39 38 37 36 35 34 33 5 U5A 74HC74 4 3 2 1 9 CS RD WR 1 2 3 4 5 6 3 4 8 GND GND J101B OSC PC104 +5V BALE TC DACK2 IRQ3 IRQ4 IRQ5 IRQ6 IRQ7 SYSCLK REFRESH DRQ1 DACK1 DRQ3 DACK3 IOR IOW SMEMR SMEMW (KEY) +12V ENDXFR -12V DRQ2 -5V IRQ9 +5V RESETDRV GND A13 I2C_DRV_RST U50C 6 CLK IACK INT A0 RESET C J107 DGND SDA VCC SCL I2C_RESET U10 PCF8584 SYSCLK U51B 1 R3 74HC08 4 R4 2.2K, 5% C39 2.2K, 5% R38 2.2K, 5% IOW 1 3 R25 R24 2 LED, RED, smt 1206 X3 NOT INSTALLED A12 1 1 JITO-2-DC5F-10OHM 4 10 U6D 3 12 IOEN 11 2 13 PRE CLK D CLR Q Q 9 8 SHDAC SHDAC 74HC74 74HC08 74HC32 R61 47k, 5% A KBINT IDC-HEADER Notes: Title 1) This schematic is for PCA #05702 2) This schematic is for PCB 05701 Size Schematic for E Series Motherboard PCA 05702 Orcad B Date: File: 1 04521C (DCN5731) D 2.2K, 5% 74HC08 VCC 1 2 3 HEADER3-DEFAULTED-1 DS5 VCC TC1 13 6 Q 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 MICROFIT-16 12 11 PRE CLK D CLR Pins 1&2 shorted on PCA JP7 JP1 2 10 5 9 Q 18 17 16 15 14 13 12 11 D0 D1 D2 D3 D4 D5 D6 D7 RN16 47Kx8 G1 G2 U4A D1 1 74HC74 I2C_RESET ADDR=0x360 (DEFAULT) ADDR = 0x320 (JP1 INSTALLED) EN 1 19 U8 J108 DI6 DI4 DI2 DI0 DO6 DO4 DO2 DO0 DI7 DI5 DI3 DI1 DO7 DO5 DO3 DO1 DO0 DO1 DO2 DO3 DO4 DO5 DO6 DO7 DI0 DI1 DI2 DI3 DI4 DI5 DI6 DI7 12 13 14 15 16 17 18 19 Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 D1 D2 D3 D4 D5 D6 D7 D8 74HC574 U7 74HC541 IOR 19 P=Q TP2 0X32F 20 VCC B0 B7 B1 B6 B2 B5 B3 B4 A0 A7 A1 A6 A2 A5 A3 A4 0X32C 9 8 7 6 5 4 3 2 D0 D1 D2 D3 D4 D5 D6 D7 74HC32 VCC U1 74HC688 3 18 5 16 7 14 9 12 2 17 4 15 6 13 8 11 IOW DIGIO2 DIGIO3 DIGIO4 TEMP DACV WRDAC VFPROG CHGAIN VFREAD 6 R59 47k, 5% G1 G2 DIGIO1 OC CLK 2 IDC-HEADER IRQ12 2 18 19 C38 0.15 uF, ceramic D[0..7] B 1 VCC GND PC104CD A TP44 1 11 3 C 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 6 1 DIGIO0 0X32D 0X32E C J102 VCC 1 D 32 31 A0 30 A1 29 A2 28 A3 27 A4 26 A5 25 A6 24 A7 23 A8 22 A9 21 A10 20 A11 19 A12 18 A13 17 A14 16 A15 15 14 13 12 AEN 11 10 9 D0 8 D1 7 D2 6 D3 5 D4 4 D5 3 D6 2 D7 1 GND A0 A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14 A15 A16 A17 A18 A19 AEN IOCHRDY D0 D1 D2 D3 D4 D5 D6 D7 IOCHECK 1 2 3 4 5 6 7 8 9 10 11 13 14 15 16 17 Y0 Y1 Y2 Y3 Y4 Y5 Y6 Y7 Y8 Y9 Y10 Y11 Y12 Y13 Y14 Y15 A B C D 5 U6A 1 23 22 21 20 J101A PC104 4 1 2 3 1 2 3 4 5 Number Revision A 05703 17-Jun-2008 Sheet 1of 8 N:\Pcbmgr\05701dn.E-motherboard.gen4\Source\05701a.DDB Drawn By: 6 D-13 1 2 3 4 5 6 1 2 3 4 5 6 7 8 9 RX1 TX1 RS-GND1 DS2 DS1 1 1 2 3 4 RX for Com1 R12 10k, 1% 4.9K, 5% 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 1 2 3 4 7 9 DB9M 1 2 3 4 DTE 10 TV ARRAY 11 8 7 6 5 R2 2.2K, 5%VCC R1 2.2K, 5% 8 7 6 5 VCC R14 R13 NOT INSTALLED 1 1 12 C TV2 SMDA15LCC SW1001 SW PUSHBUTTON-4PDT NOT INSTALLED DS3 DS4 R10 NOT INSTALLED 2 RX for Com2 1 TX for Com2 1 LED, RED, smt 1206 LED, GRN, smt 1206 1 2 NC RXD TXD NC GND NC RTS CTS NC 8 1 INLINE-12 J1013 DCE side of switch is side towards pin 1, RX0 RTS0 TX0 CTS0 RS-GND0 RX1 RTS1 TX1 CTS1 RS-GND1 1 2 3 4 5 6 7 8 9 10 11 12 13 14 R111 Com1 - RS232-A J12 D -15V 1 R11 4.9K, 5% LED, GRN, smt 1206 TV1 TV ARRAY SMDA15LCC 8 7 6 5 TX for Com1 1 2 LED, RED, smt 1206 2 8 7 6 5 1 2 2 C 1 2 3 4 RTS1 CTS1 D Com2 - RS232-B/RS485 J1010 DB9 FEMALE MT6 MT7 MT8 MT9 TP17 +12V +12VRET +15V -15V 1 1 1 1 VCC 1 MT1 MT2 MT3 MT4 MT5 TP18 B 1 1 TP16 1 TP15 1 TP14 1 TP13 MOUNTING HOLE 1 MOUNTING HOLE MOUNTING HOLE MOUNTING HOLE MOUNTING HOLE MOUNTING HOLE MOUNTING HOLE MOUNTING HOLE MOUNTING HOLE B J15 AUX DC POWER IN +12V +12RET DGND +15V -15V AGND +5V AGND EGND CHASGND 8 7 1 4 6 3 2 5 9 10 VCC U51C 9 8 10 uF, 35V, TANTALUM + C2 10 D1 C1 + 74HC08 10 uF, 35V, TANTALUM MOLEX-10 MBRS340CT D9 D1, D9 & R35 must be within 1" of J15 MBRS340CT R35 A A NOT INSTALLED Title Schematic for E Series Motherboard PCA 05702 Size Orcad B Date: File: 1 D-14 2 3 4 5 Number Revision A 05703 17-Jun-2008 Sheet 2of 8 N:\Pcbmgr\05701dn.E-motherboard.gen4\Source\05701a.DDB Drawn By: 6 04521C (DCN5731) 3 -15V 2 4 6 8 2 - OP-AMP, PRECISION DUAL 1 0.15 uF, ceramic U29B U30 74HC574 U20C 9 8 CLK 10 2 3 4 5 6 7 8 9 0.15 uF, ceramic + Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 D1 D2 D3 D4 D5 D6 D7 D8 19 18 17 16 15 14 13 12 CSDACA 2 4 6 8 - CSDACA CSRANGE1 CSDACB 2 4 6 8 4 3 2 1 5 6 7 8 SMDA15LCC L5 L6 L7 L15 SMDA15LCC 1 3 5 7 1 2 3 4 5 6 7 8 9 10 DAC3V DAC0 DAC1 DAC2 DAC3 1 2 3 4 5 6 7 8 9 10 0.15 uF, ceramic MICROFIT-10 +15V C D[0..7] 0.15 uF, ceramic C10 C9 C +15V 4 U35A OP-AMP, PRECISION QUAD 1 U32 D0 CLK 11 12 14 13 3 B 15 10 SHDAC W1 B1 AGND1 18.7K W2 B2 AGND2 5 4 2 1 10k, 1% 5 6 7 8 23 11 12 14 13 D0 CLK A1 - 3 OP-AMP, PRECISION QUAD 16 SHDAC A2 U36B 5 C17 10 W2 B2 AGND2 9 C11 0.15 uF, ceramic + 4 2 1 6 - -15V OP-AMP, PRECISION QUAD OP-AMP, PRECISION QUAD DGND 16 B +15V U36C VCC 0.15 uF, ceramic 10 + 9 - 8 9 R20 DAC3V DAC3 18.7K OP-AMP, PRECISION QUAD + 8 TP32 R18 10k, 1% 0.15 uF, ceramic U35C DAC1 +15V -15V 7 VCC +15V 18 20 17 DAC2V TP29 C16 C14 0.15 uF, ceramic 0.15 uF, ceramic 18.7K RS SHDN VCC VCC 9 15 10 22 24 21 CS SDI CLK SDO 7 6 W1 B1 AGND1 0.15 uF, ceramic -15V + C18 - 19 DAC1V A3 DAC1V R22 W3 B3 AGND3 18 20 17 -15V TP33 U36D +15V R21 10k, 1% 18.7K 1 -15V 7 +15V U35D POT, DIGITAL R23 10k, 1% + MBRS340CT 13 - 12 + 13 - 14 -15V OP-AMP, PRECISION QUAD 14 A 8 6 5 POT, DIGITAL D7 12 A4 W4 B4 AGND4 11 A4 8 6 5 4 7 W4 B4 AGND4 11 1 4 W3 B3 AGND3 VOA GND VCC VOB C12 -15V DGND A3 DOUT CS DIN CLK SOCKET U33 +15V U35B VCC 19 4 3 2 1 DAC, 12 BIT R19 RS SHDN VCC CSDACB D0 CLK TP28 CS SDI CLK SDO A2 U33 R17 22 24 21 11 A1 DAC 2 11 DAC, 12 BIT 23 DUAL DAC A2 U34 4 5 6 7 8 4 VOA GND VCC VOB 11 DOUT CS DIN CLK 1 U31 R16 DAC0V 11 DAC0V 4 - DUAL DAC A1 - 4 2 2 1 1 4 3 CSDACA 2 D0 1 CLK SOCKET U31 OP-AMP, PRECISION QUAD + 3 + 1 1 3 TP27 4 U36A TP26 D J22 CSRANGE2 74HC32 0 0G 1 1G 2 2G 3 3G TERMBLOCK-8 FE BEAD 1 3 5 7 IDC-8 OP-AMP, PRECISION DUAL CSDACB C19 C13 10000 pF 10000 pF C5 C4 10000 pF 10000 pF TV4 TV ARRAY J23 7 6 1 3 5 7 TV3 TV ARRAY IDC-8 4 D0 D1 D2 D3 D4 D5 D6 D7 5 OC CLK 2 4 6 8 1 3 5 7 J1020 1 2 3 4 5 6 7 8 11 1 11 74HC32 WRDAC -15V 8 U20B 5 L2 L3 L4 5 6 7 8 C53 IOW IOW 2 4 6 8 C8 TC2 6 1 3 5 7 J21 4 R63 10k, 1% 4 1 3 5 7 IDC-8 D DACV DACV 2 4 6 8 C20 10000 pF C15 10000 pF 4 3 2 1 + 4 3 2 1 1 VREF 3 C7 10000 pF 10000 pF C21 L1 FE BEAD +15V DAC RANGE & OFFSET PROGRAM 40K R15 6 ANALOG VOLTAGE & CURRENT OUTPUTS J19 0.15 uF, ceramic 8 U29A 5 5 6 7 8 TP21 4 ISOLATED 0-20MA OPTIONAL BOARDS C6 +15V 5 6 7 8 2 4 3 2 1 1 A D8 Title 11 Schematic for E Series Motherboard PCA 05702 D7 and D8 Must be located within 1" of U32 & U34 1 04521C (DCN5731) -15V OP-AMP, PRECISION QUAD MBRS340CT Size Orcad B Date: File: 2 3 4 5 Number Revision A 05703 17-Jun-2008 Sheet 3 of 8 N:\Pcbmgr\05701dn.E-motherboard.gen4\Source\05701a.DDB Drawn By: 6 D-15 1 2 3 4 5 5 10 6 +15V 5 10 - C46 0.15 uF, ceramic VCC U54 .022 uF, 50V VCC U55 DG444DY 3 14 11 6 1 16 9 8 2 15 10 7 12 4 5 13 D1 D2 D3 D4 VCC -VS GND +VS 100 R47 1 IOW 1 3 2 Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 19 18 17 16 15 14 13 12 D6 D2 VCC 10 uF, 35V, TANTALUM 10 5 C50 VCC C D4 D3 D7 C51 0.15 uF, ceramic D0 SEL60 D[0..7] IOW 5 D0 D1 D2 D3 D4 D5 D6 D7 74HC32 A 2 3 4 5 6 7 8 9 D1 D2 D3 D4 D5 D6 D7 D8 Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 19 18 17 16 15 14 13 12 7 8 9 10 11 12 13 14 15 16 17 TP54 R9 100 TP57 DB4 RDMBYTE DB3 GND U57 DB7 TIE TIE DB0 Xilinx CPLD TDI TMS TCK TC8 TIE TIE TIE TIE FREQ TIE TIE VCCIO GND TDO SEL60 B C52 0.15 uF, ceramic SEL60 IOR SA SB SC START VFREAD MSB MID LSB A Title Schematic for E Series Motherboard PCA 05702 Date: File: D-16 VCC D1 Orcad B 2 39 38 37 36 35 34 33 32 31 30 29 TP55 Size 1 X1 MB100H-4.8MHZ 5 18 19 20 21 22 23 24 25 26 27 28 TP53 1 OE CLK TP52 1 TP51 1 6 1 VFPROG 1 11 1 U60 74HC574 TP50 U59B 4 D5 RDMSB TIE DB1 VCCINT IOR GND SA SB SC READ START 74HC32 1 4 4 0.15 uF, ceramic R49 100 6 5 4 3 2 1 44 43 42 41 40 1 2 3 4 6 7 8 9 OE CLK D1 D2 D3 D4 D5 D6 D7 D8 C 1 TP48 PLACE 100 OHM RESISTOR AS CLOS AS POSSIBLE TO VCC X1 AND X2 R47 and R48 reduce the gain for analog inputs by 1%, so that we can read slightly above full scale, to prevent overflow of ADC reading + U59A CHGAIN 2 3 4 5 6 7 8 9 C54 X2 JITO-2-DCA5AE-4.8MHZ C49 -15V TC6 D0 D1 D2 D3 D4 D5 D6 D7 AD652KP 1 1.2 uF, 6.3V ceramic U58 74HC574 B VCC 8 +15V -15V RN17 100Kx8 1 11 10 uF, 35V, TANTALUM C48 1 2 3 4 VREF SHDN + R46 1.1K, 5% 0.15 uF, ceramic VOLTAGE REF TP49 C45 18 17 16 15 14 COMP+ COMPAGND GND FOUT R48 200 VREF DACMUX OP OUT OPOP+ 5VI 10VI 6 +15V S1 S2 S3 S4 IN1 IN2 IN3 IN4 4 5 6 7 8 1 TP1 VCC 3 2 1 20 19 -15V 1M, 1%, 1206 CHIP R45 C44 13 2 3 18 14 15 16 17 VREF NC NC ENB A3 A2 A1 A0 U56 C47 1.2 uF, 6.3V ceramic 8 NC NC 7 NC VIN 6 VOUT NR 5 TRIM GND TC7 NC +VS NC REF NC C43 0.15 uF, ceramic 12 GND R45 induces an offset in analog signal to give a 'live 0' for sensors with 0 or slightly negative output U53 27 -VSS AN MUX AGND 1 1 +VSS TP3 1 TEMPMUX 0.15 uF, ceramic2 OP-AMP, PRECISION 6 3 CH14 CH13 CH12 CH11 CH9 CH8 + C42 8VI OPT10V -VS COS CLK CH11 CH12 CH13 CH14 3 28 OUT 1 MICROFIT-12 10 uF, 35V, TANTALUM 9 10 11 12 13 CH7 CH8 D + RDLSB DB2 DB6 TIE TIE TIE DB5 VFCLK ICLK VCCINT TIE 9 8 7 6 4 3 2 1 CH6 IN 1 IN 2 IN 3 IN 4 IN 5 IN 6 IN 7 IN 8 IN 9 IN 10 IN 11 IN 12 IN 13 IN 14 IN 15 IN 16 C41 0.15 uF, ceramic 4 19 20 21 22 23 24 25 26 11 10 9 8 7 6 5 4 CH1 CH2 CH3 CH4 CH9 C C40 0.15 uF, ceramic U52 J110 100 C55 RN15 100Kx8 MICROFIT-12 1 2 3 4 5 6 7 8 9 10 11 12 R43 ANALOG INPUTS 9 8 7 6 4 3 2 1 D CH7 CH6 CH4 CH3 CH2 CH1 C 1 2 3 4 5 6 7 8 9 10 11 12 +15V -15V +15V 7 C J109 RN14 100Kx8 3 4 5 Number Revision 05703 A 4 17-Jun-2008 Sheet of 8 N:\Pcbmgr\05701dn.E-motherboard.gen4\Source\05701a.DDB Drawn By: 6 04521C (DCN5731) 1 2 3 4 5 6 +15V +5VANA U23 1 3 + C60 10 uF, 35V, TANTALUM LP2981IM5 D 2 D OUT IN ON/OFF NC GND BYPASS CAPS MUST BE WITHIN 1/2" OF THE REGULATOR INPUT/OUTPUT PINS 5 4 C29 1 uF D[0..7] +5VANA VCC +15V XT1 U48 MAX382CWN 9 14 15 4 3 2 17 16 18 1 TEMPMUX D0 D1 D2 C SHDN OUT +VSS GND VENB A0 A1 A2 RS WR J27 THERMISTER THERMISTER1 5 6 7 8 13 12 11 10 IN 1 IN 2 IN 3 IN 4 IN 5 IN 6 IN 7 IN 8 THERMISTER2 THERMISTER3 THERMISTER4 THERMISTER5 THERMISTER6 THERMISTER7 THERMISTER8 2 3 4 6 7 8 9 10 U59D TEMP 12 IOW THERMISTER6 THERMISTER5 11 1 C 74HC32 C MICROFIT-14 RN20 10Kx9, 2% 13 1 2 3 4 5 6 7 8 9 10 11 12 13 14 B B +15V-15V RN18 U49 DACMUX 10K R34 C36 0.15 uF, ceramic VCC C37 2 15 10 7 12 4 5 13 D1 D2 D3 D4 VCC -VS GND +VS S1 S2 S3 S4 IN1 IN2 IN3 IN4 3 14 11 6 1 16 9 8 1 2 3 4 8 7 6 5 1 2 3 4 1Kx4 8 7 6 5 DAC0V DAC0V DAC1V DAC2V DAC3V DAC1V DAC2V DAC3V DAC0 DAC1 DAC2 DAC3 0.15 uF, ceramic DG444DY 10Kx4 RN21 A A Title Schematic for E Series Motherboard PCA 05702 Size Orcad B Date: File: 1 04521C (DCN5731) 2 3 4 5 Number Revision A 05703 17-Jun-2008 Sheet 5of 8 N:\Pcbmgr\05701dn.E-motherboard.gen4\Source\05701a.DDB Drawn By: 6 D-17 1 2 3 4 5 6 CONTROL INPUTS 5 10 5 10 VCC C C RN3 510x8 TP7 RN2 15Kx8 D U11 1 D 74HC541 10000 pF C D[0..7] R27 R28 R29 100 100 100 C97 R26 100 D7 9 R31 R32 R33 100 100 100 R30 100 330 pF, 50V L23 L24 L26 C59 C62 L25 FE BEAD 16 2 3 15 14 4 5 13 12 6 7 11 10 8 9 330 pF, 50V C102 C98 C96 1 C100 330 pF, 50V U13 PS2702-4 C C103 8 D6 11 10 D5 6 7 D0 D1 D2 D3 D4 D5 D6 D7 D4 13 12 D0 C22 C56 4 5 10000 pF EXT_+5V_OUT TERMBLOCK-10 C34 L9 15 14 18 17 16 15 14 13 12 11 Y1 Y2 Y3 Y4 Y5 Y6 Y7 Y8 A1 A2 A3 A4 A5 A6 A7 A8 C101 C57 C23 L8 2 3 2 3 4 5 6 7 8 9 D3 L22 FE BEAD C35 EXTERNAL CONTROL IN A 16 DIGIO0 IOR C99 1 2 3 4 5 6 7 8 9 10 1 D2 L19 L20 L21 D1 J1004 1 19 G1 G2 9 8 7 6 4 3 2 1 9 8 7 6 4 3 2 1 U12 PS2702-4 330 pF, 50V Place these termination resistors at the end of each data line. Each data line should be laid out as a daisy-chain, the signal passing from one IC to the next. VCC C61 C58 10000 pF 10000 pF B 5 10 B 8 7 6 5 C RN4 15Kx8 U14 RN1 1 2 3 4 L28 L29 L30 L27 1 16 2 3 15 14 4 5 13 12 6 7 11 10 8 9 A1 A2 A3 A4 A5 A6 A7 A8 DIGIO4 D0 D1 D2 D3 D4 D5 D6 D7 L11 C66 10000 pF A EXT_+5V_OUT Title Schematic for E Series Motherboard PCA 05702 C65 C63 C64 C25 FE BEAD Size Orcad B Date: File: D-18 18 17 16 15 14 13 12 11 IOR 74HC541 10000 pF 1 Y1 Y2 Y3 Y4 Y5 Y6 Y7 Y8 1 19 D[0..7] L10 C24 A 1 2 3 4 5 6 7 8 9 10 TERMBLOCK-10 2 3 4 5 6 7 8 9 U15 PS2702-4 J1006 EXTERNAL CONTROL IN B G1 G2 9 8 7 6 4 3 2 1 510x4 2 3 4 5 Number Revision A 05703 17-Jun-2008 Sheet 6of 8 N:\Pcbmgr\05701dn.E-motherboard.gen4\Source\05701a.DDB Drawn By: 6 04521C (DCN5731) 1 2 3 4 5 6 5 10 VCC DIGITAL OUTPUTS C RN10 510x8 D D U22 9 8 7 6 4 3 2 1 1 C80 PS2702-4 16 C82 10000 pF TP19 SHDN SHDN 1 U6B 4 DIGIO2 IOW U24 74HC574 1 11 6 5 2 3 4 5 6 7 8 9 D0 D1 D2 D3 D4 D5 D6 D7 74HC32 OE CLK D1 D2 D3 D4 D5 D6 D7 D8 Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 19 18 17 16 15 14 13 12 2 3 15 14 4 5 13 12 6 7 11 10 8 9 U25 D[0..7] C 1 PS2702-4 16 2 3 15 14 4 5 13 12 6 7 11 10 8 C79 C81 10000 pF L43 L44 L45 L46 FE BEAD J1017 1 2 3 4 5 6 7 8 9 10 11 12 L48 L49 L50 L47 FE BEAD C84 C86 9 FE BEAD C83 C TERMBLOCK-12 10000 pF L12 A STATUS OUTPUTS C85 C26 10000 pF C27 RESETTABLE FUSE, 0.3A, 60V VCC 5 10 D6 F1 L13 VCC C FE BEAD RN12 510x8 9 8 7 6 4 3 2 1 U26 B SHDN U27 74HC574 U20D 12 DIGIO3 1 11 11 13 IOW IOW 74HC32 D0 D1 D2 D3 D4 D5 D6 D7 2 3 4 5 6 7 8 9 OE CLK D1 D2 D3 D4 D5 D6 D7 D8 Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 19 18 17 16 15 14 13 12 EXT_+5V_OUT DIODE, SCHOTTKY 1 PS2702-4 16 2 3 15 14 4 5 13 12 6 7 11 10 8 9 1 U28 PS2702-4 16 2 3 15 14 4 5 13 12 6 7 11 10 8 9 B C90 L52 L53 L54 C88 B STATUS OUTPUTS C89 C87 10000 pF J1018 L51 FE BEAD 1 2 3 4 5 6 7 8 9 10 L56 L57 L58 L55 FE BEAD C28 A 10000 pF L14 TERMBLOCK-10 C92 C91 1 2 3 4 5 6 7 8 RET GND C94 10000 pF C93 A 10000 pF Title Schematic for E Series Motherboard PCA 05702 Size Orcad B Date: File: 1 04521C (DCN5731) 2 3 4 5 Number Revision A 05703 17-Jun-2008 Sheet 7of 8 N:\Pcbmgr\05701dn.E-motherboard.gen4\Source\05701a.DDB Drawn By: 6 D-19 6 5 4 3 2 1 5 10 VCC DIGITAL C IOW 1 11 8 10 D0 D1 D2 D3 D4 D5 D6 D7 74HC32 2 3 4 5 6 7 8 9 OE CLK D1 D2 D3 D4 D5 D6 D7 D8 Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 19 18 17 16 15 14 13 12 15 14 4 5 13 12 6 7 11 10 8 9 U18 D[0..7] 1 PS2702-4 16 2 3 15 14 4 5 13 12 6 7 11 10 8 9 C70 2 3 D 10000 pF C69 9 8 7 6 4 3 2 1 U17 74HC574 U59C 9 DIGIO0 PS2702-4 16 C67 D 1 C68 RN7 510x8 U16 SHDN SHDN OUTPUTS 10000 pF L32 L33 L34 L31 FE BEAD J1008 L36 L37 L38 L35 FE BEAD CO_EXT_RET C 1 2 3 4 5 6 7 8 9 10 11 12 13 14 CONTROL OUTPUTS C TERMBLOCK-14 5 10 C74 C72 L59 FE BEAD VCC EXTERNAL CONNECTOR SOLDER SIDE C71 C95 C RN5 510x8 C73 10000 pF 10000 pF 10000 pF SHDN U21 74HC574 2 3 15 14 IOW 1 2 74HC32 B 1 11 3 D0 D1 D2 D3 D4 D5 D6 D7 2 3 4 5 6 7 8 9 OE CLK 4 5 13 12 D1 D2 D3 D4 D5 D6 D7 D8 Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 19 18 17 16 15 14 13 12 6 7 11 10 8 9 L40 L41 L42 L39 FE BEAD 10000 pF +12V D2 RELAY SPDT 4 1 3 K1 2 5 DIODE, SCHOTTKY B C77 DIGIO4 C75 U20A C78 PS2702-4 16 C76 U19 9 8 7 6 4 3 2 1 1 10000 pF J1009 Q1 R58 +12V 1 2 3 4 5 6 7 8 9 10 11 12 D3 RELAY SPDT 2.2K, 5% K2 SO2222 4 1 3 2 5 DIODE, SCHOTTKY RELAY SPDT Q2 R6 K3 +12V 4 1 3 2 5 D4 2.2K, 5% SO2222 DIODE, SCHOTTKY Q3 EXTERNAL REAR PANEL ALARM OUTPUTS TERMBLOCK-12 +12V RELAY SPDT D5 K4 R7 2.2K, 5% SO2222 DIODE, SCHOTTKY 2 5 4 1 3 Q4 A A R8 Title Schematic for E Series Motherboard PCA 05702 2.2K, 5% SO2222 +12VRET Size Orcad B Date: File: 1 D-20 2 3 4 5 Number Revision A 05703 17-Jun-2008 Sheet 8of 8 N:\Pcbmgr\05701dn.E-motherboard.gen4\Source\05701a.DDB Drawn By: 6 04521C (DCN5731) 1 M1 2 3 4 5 6 VCC M2 1 2 3 4 5 6 7 10uF DS3 S4 KBD_A0 KBD_A1 KBD_A2 21 2 3 1 SCL SDA 22 23 A0 A1 A2 INT P00 P01 P02 P03 P04 SCL P05 SDA P06 P07 P10 PCF8575 P11 P12 P13 P14 P15 P16 P17 M8 S3 VCC VCC S2 R2 1.0K U3A 1 4 3 2 1 C MF4 RN1 4.7K S1 C7 PRE CLK D CLR 5 6 Q Q + DS5 MAINT_SW LANG_SELCT DS6 GRN LED YEL LED RED LED LED 4 LED 5 LED 6 HORN SPR_I/O_0 RI-1000 ONLY Layout Instructions: A1 SONALERT Vss MF3 DS4 4 5 6 7 8 9 10 11 13 14 15 16 17 18 19 20 12 M10 220 2 3 4 5 6 MM74HC74A 300pF S9 VCC RI-1000 ONLY U4 VCC OPT. MAINT SWITCH S12 RN5 4.7K SPR_I/O_1 SDA TP3 BUSY SCL TP8 DISP_PWR DISP_RET 4.7K DISP_PWR_EN must be high for display to be powered. DISP_BUSY A VCC 6 7 8 9 10 J1 +5_DISP DISP_CN_A0 DISP_CN_A1 DISP_CN_A2 SCL SDA 1 2 3 14 15 AO A1 A2 SCL SDA SCL DISP_RET VCC JP3 1 2 3 4 SDA 5 6 7 8 9 10 DISP_RET (U1) DISP_RET SCL KYBRD_INT (U2) (U4) (U45 PCF8574 C11 C12 C10 220pF 220pF 220pF + C14 C4 100uF .1uF C17 VCC 4.7K SCL KYBRD_INT B Q1 C9 C8 220pF 220pF + C13 C2 C3 C5 C15 C16 .1uF .1uF .1uF .1uF P0 P1 P2 P3 P4 P5 P6 P7 INT 4 5 6 7 9 10 11 12 1 2 3 RN2 4.7K 1500uF VCC 1 2 3 4 SDA 5 DISP_PWR DISP_RET VCC +5_DISP +5_DISP DISP_WR DISP_BUSY 13 DISPLAY CONTROL U5 NOTES: 1. This schematic is based on the PWB PN, 03974 and applies to PCA PN, 03975 R4 1 3 5 7 9 11 13 15 +5_DISP DISPL CONTROL (DISP_CN_A0 -A1) 011 R3 TP9 2 4 6 8 10 12 14 16 4 KEYBOARD (KBD_A0 - A2) 111 KYBRD_INT INT 4 5 6 7 9 10 11 12 16 JP1 ADRS SLCTS TP5 SDA +5_DISP DISP_RET TP7 3 KYBRD INT TP4 VCC DISP_PWR JP2 I2C TERMINATION SCL 1 2 SDA DEFAULT ADDRESS SELECTS FOR I2C TO PARALLEL DECODERS: 2 4 6 8 10 12 14 16 18 +5_DISP TP2 TP6 PCF8574 Vss DISP_DA_A0 DISP_DA_A1 DISP_DA_A2 S13 SCL SDA DISPLAY DATA SPR_I/O_2 OPT. LANG. SWITCH 14 15 D G SI3443DV 1 2 MCP120T 1 U6 13 Vdd RST 3 MMBT3904 R20 Q2 1K 4.85V DTCT SPR_I/O_1 SPR_I/O_2 A 10uF .1uF Title J2 JP4 Schematic for PCA #04258 and PCB #04257, Keyboard/Display Interface for E series Size Number Revision 04259 B Date: File: 04521C (DCN5731) S +5_DISP 6 5 4 JP5 DISP_PWR_OVR DISP_WR DISP_BUSY DISP_PWR_EN MAINT_LED DISP_RET 1 C J3 TO/FRM DISPLAY P0 P1 P2 P3 P4 P5 P6 P7 8 SPR_I/O_0 VCC TP1 SCL SDA MM74HC74A KBD_A0 KBD_A1 KBD_A2 DISP_CN_A0 DISP_CN_A1 DISP_CN_A2 3M-2514-6002UB GND KYBRD_INT AO A1 A2 1 MAINT_LED_V+ MAINT_LED LANG_SELCT 9 8 Q Q 2 3 4 5 6 7 8 9 10 MAINT_SW PRE CLK D CLR 1 2 3 Vss B 1 2 3 4 5 6 7 8 9 10 11 12 13 14 T8201 1 3 5 7 9 11 13 15 17 MAINT SW MAINT SW RET MAINT LED V+ MAINT LED LANG SW LANG SW RET SPR I/O_0 SPR I/O RET SPR I/O_1 SPR I/O RET SPR I/O_2 SPR I/O RET 10 11 12 13 1 VCC J4 DISP_DA_A0 DISP_DA_A1 DISP_DA_A2 U3B 1. Minimum trace width 8 mil would like to have 10 mil traces if possible. 2. Please run traces on both and backside but where possible fill one side with GND. 3. Minimum width for +5_DISP, DISP_PWR, DISP_RET is 40 mil, except to test points. 4. Minimum width for VCC, GND, Vdd, Vss is 30 mil, except to test points 2 3 4 5 6 M9 D U2 S5 74C923 MAINT_LED_V+ RED KEYBOARD, LED & HORN 12 11 9 8 10 9 8 7 6 Vss M7 13 X1 X2 X3 X4 OE YEL VCC AVL Vss 14 1 2 3 4 5 2 + C6 .1uF RN3 GRN 16 C1 S6 DS2 Vdd M5 DS1 8 M6 19 18 17 16 15 D_A D_B D_C D_D D_E Vdd S7 Y1 Y2 Y3 Y4 Y5 OSC KBM 24 S8 10 D M4 Vdd M3 Vcc 20 VCC U1 2 3 4 5 a 21-Mar-2002 Sheet of N:\YHWork\M300B\keyboard\04257a\04259A.ddb Drawn By: 6 D-21 1 2 3 4 +15V D R2 1.1K S1 ASCX PRESSURE SENSOR 1 2 3 4 5 6 2 VR2 D 3 C2 1.0UF 1 LM4040CIZ TP4 TP5 S1/S4_OUT S2_OUT TP3 S3_OUT TP2 10V_REF TP1 GND 3 2 1 S2 ASCX PRESSURE SENSOR C 1 2 3 4 5 6 +15V J1 6 5 4 MINIFIT6 +15V C R1 499 S3 FLOW SENSOR FM_4 1 2 3 2 +15V 1 2 3 4 B 3 C1 1.0UF 1 CN_647 X 3 S4 VR1 LM4040CIZ C3 1.0 B CON4 The information herein is the property of API and is submitted in strictest confidence for reference only. Unauthorized use by anyone for any other purposes is prohibited. This document or any information contained in it may not be duplicated without proper authorization. A 1 D-22 2 3 APPROVALS DATE SCH, PCA 04003, PRESS/FLOW, 'E' SERIES DRAWN A CHECKED SIZE APPROVED LAST MOD. B DRAWING NO. REVISION 04354 D SHEET 3-Dec-2007 1 of 1 4 04521C (DCN5731) A +5V 1 6 4 5 8 A18 A19 IC102:A 74AC00D IC101 +5V 74ACT138 1 2 3 A15 A16 A17 IC102:D 74AC00D 1 3 11 2 2 13 12 6 4 5 6 IC103:C 74ACT32 9 8 43 44 52 68 +5V STATUS PL101:2 PL101:16 2 16 PL101:12 12 R101 +5V R102 4.99K 4.99K IC106 MAX237 10 S C120 1uF 16V C121 1uF 16V S TXD PL101:3 DTR PL101:4 RTS PL101:5 DCD PL101:6 RI PL101:7 C1V+ RS-232 3 4 5 6 7 2 3 1 24 20 PL101:8 PL101:9 PL101:10 8 9 10 4 23 16 RESET PL101:11 11 8 TTL TO1 TO2 TO3 TO4 TO5 TI1 TI2 TI3 TI4 TI5 RI1 RI2 RI3 RO1 RO2 RO3 R103 499 R104 499 DS103 TXD DS104 RXD 27 28 29 32 66 5 22 17 30 15 59 12 61 62 55 67 +5V IC107 TL7705 +5V 7 2 3 1 C124 1uF 16V S C125 1uF 16V C126 S 100nF S SENSE VCC RESIN RESET CT RESET REF URTINT -LMSEL S 26 63 58 60 20 R105 4.99K S -UCS -LCS S R106 4.99K +5V PL102-1 -WR -RD +5V 7 6 18 19 21 GND +5V 4 14 15 S (3) RXD (4) DSR (7) CTS C129 10uF 16V PL102-2 C2V- S DB-9 PIN NUMBERS IN PARENS. (2) (6) (8) (1) 9 13 S 12 11 +5V +5V VCC C2+ C1+ X2 CLKO S 40 65 +5V C119 100nF 14 21 C123 C122 1uF 16V 1uF 16V GND (5) GND VCC X1 GND S -BHE LANDRQ ALE 2 7 11 13 15 12 14 16 17 75 31 18 41 Y101 18.432MHz C105 22pF 1 15 2 3 1 42 13 +5V IC104 C0561AD-L +5V C104 22pF 14 PL101:1 PL101:15 61 62 29 28 10 10 +5V +5V 33 34 36 49 63 64 5 8 NC S IC103:B 74ACT32 4 IC102:C 74AC00D PL101:14 PL101:13 IC105 CS8900A-CQ C118 100nF NC1 NC2 NC3 NC4 NC5 TXD1 -DTR1 -RTS1 -DCD1 -RI1 RXD1 -DSR1 -CTS1 -RES HLDA HOLD GND GND A19 A18 A17 A16 A15 A14 A13 A12 A11 A10 A9 A8 A7 A6 A5 A4 A3 A2 A1 A0 D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0 LANINT 50 39 8 45 49 47 51 48 53 56 54 38 37 33 46 36 35 34 19 57 A19 A18 A17 A16 A15 60 59 58 54 53 52 51 50 48 47 46 45 44 43 42 41 40 39 38 37 25 10 11 23 13 16 17 22 9 24 2 3 4 5 6 7 18 19 20 21 24 25 26 27 74 73 72 71 68 67 66 65 64 35 30 31 32 9 7 IOCS1 6 MEMCS1 6 SBHE REFRESH AEN IOCHRDY C107 15pF 20.0 MHz X T A L 2 R108 10.0K 5 6 D V D D 1 D V D D 2 D V D D 3 D V D D 4 A V D D 2 A V D D 1 A V SLEEP D D TEST 3 LANLED ELCS CHIPSEL DMARQ2 DMARQ1 DMARQ0 DMACK2 DMACK1 DMACK0 CSOUT RESET LINKLED/ HC0 77 +5V R110 4K99 100 99 RES SA19 SA18 SA17 SA16 SA15 SA14 SA13 SA12 SA11 SA10 SA9 SA8 SA7 SA6 SA5 SA4 SA3 SA2 SA1 SA0 RXD- DS101 ACT, R113 499R DS102 LINK 04521C (DCN5731) B S 1 +5V C128 100nF C103 100nF 78 93 R114 4K99 C109 100nF T101 TG43-1406N 92 1 RXD+ TXD- TXD+ SD15 SD14 SD13 SD12 SD11 SD10 SD09 SD08 SD7 SD6 SD5 SD4 SD3 SD2 SD1 SD0 DODO+ CICI+ DIDI+ D V S S 1 D V S S 1 A D V S S 2 D V S S 3 D V S S 3 A D V S S 4 A V S S 0 A V S S 1 A V S S 2 A V S S 3 A V S S 4 EECS EESK EEDATAOUT EEDATAIN 91 88 87 R116 24R3 R117 24R3 1:1 J101 16 15 3 14 3 11 2 1: 2 7 10 8 9 C110 100nF 84 6 2 6 C108 68pF 83 2 RX- RX+ TX- 1 4 5 7 8 9 10 TX+ NC1 NC2 NC3 NC4 S1 S2 3 82 81 80 +5V S 3 4 5 6 1 8 8 9 9 9 6 4 6 +5V C111 100nF S 79 +5V S +5V C112 100nF S +5V C115 100nF S +5V C113 100nF +5V C116 100nF S C114 100nF +5V C117 100nF C127 100nF R109 10.0K TELEDYNE ADVANCED POLLUTION INSTRUMENTATION INC. Title 1 2 3 THIS SCHEMATIC APPLIES TO PWB 04393 REV. A. ALL RESISTANCES IN OHMS, 1% PARTS DENOTED "S" ON SECONDARY SIDE OF PCA C 4 ETHERNET INTERFACE SCHEMATIC Number Size B Date Filename A +5V C102 100nF MT1 NOTES: R107 10.0K +5V R112 499R R115 100R INTRQ3 INTRQ2 INTRQ1 INTRQ0 S R111 4K99 76 +5V C101 100nF S MT2 BSTATUS/ HC1 8 1 2 5 5 7 0 3 5 7 0 8 +5V 2 5 6 8 9 9 9 2 6 9 5 0 5 9 8 X T A L 1 +5V +5V IOR IOW MEMR MEMW +5V 4 D Y102 C106 15pF +5V 11 9 NC +5V IC103:D 74ACT32 13 IC102:B 74AC00D C IC103:A 74ACT32 1 12 3 16 15 14 13 12 11 10 9 7 VCC Y0 Y1 Y2 Y3 Y4 G1 Y5 Y6 G2 Y7 G3 GND A B C B Rev 04395 A Drawn by Sheet 1 Thu Jul 25 2002 SLAN.S03 of 1 D D-23 1 2 3 4 6 5 D D C C Interconnections 04181H-1-m100e200e.sch preamp cktry 04181H-2-m100e200e.SCH HVPS Cktry 04181H-3-m100e200e.SCH B B A A Title M100E/200E PMT Preamp PCA Size B Date: File: 1 D-24 2 3 4 5 Number Revision 04181 H Sheet 0 10-May-2007 of N:\PCBMGR\04179cc\Source\RevG\04179.ddb Drawn By: 3 6 04521C (DCN5731) 1 2 3 4 6 5 ON JP2: +15V PMT TEMPERATURE FEEDBACK FOR 100E/200E : SHORT PINS 2 &5 ONLY. FOR 200EU: SHORT PINS 3 & 6 and PINS 2 & 5. +12V_REF JP2 +15V R28 TH1 FSV +15V D1 6.2V ZENER 6.2V 1 2 OPTIC TEST 8 50K R8 150K D 3 1 2 3 4 5 6 TJP1A TJP2A U2A 2 R27 R18 SEE TABLE 1 499 PMT TEMP CONFIG JUMPER D 3 LF353 4 + C23 100 pF S R6 R15 SEE TABLE C1 +12V_REF TO TEC BOARD 100K C26 0.1 uF +12V_REF * J2 TP3 VREF 1 COOLER CONTROL 2 AGND 3 3 PIN INLINE 8 Q3 J176 D R35 1.0K N/I G U3B R41 300K R16 100K R2 51.1K 6 7 5 * TP24 TJP1A LF353 4 THERMISTOR+ +15V PREAMP1 LED+ TP23 * THERMISTOR+ U13 +15V b R23 1 4 2 +5V_SYS C6 COMP. 100E 200E 0200EU ------------------------------------------------R18 10K 10K 14K R15 55K 55K 47K R10 8.09K 8.09K 10K LED+ HVPS R7 10K R1 10K U3A 2 R9 1 PMT_TEMP 3 OPTIC_TEST 2.0K LF353 R10 4.99K 3 Q2 PN2222 R37 3.3K 4 INLINE-9-RA 74AHC1GU04 C D2 11DQ05 0.1 uF 8 -15V 2 C 9 8 7 6 5 4 3 2 1 Ec J3 RT1 R32 499 SEE TABLE TJP2A * TP18 * TP17 * TP25 * TP19 * TP22 TP21 * * TP20 Signal Connector J6 ETEST OPTIC_TEST HIGAIN PMT_TEMP B HVPS ELEC TEST OPTIC TEST PREAMP RNG BIT2 PREAMP RNG BIT1 PMT TEMP HVPS VOLTAGE PMT SIGNAL 1 2 3 4 5 6 7 8 VPMT B MICROFIT-8 J5 *TP11 L2 +15V 4.7 uH C21 + C49 0.68 uF 100uF * *TP16 TP15 * TP14 * TP13 1 2 3 4 5 6 7 8 9 10 Power Connector MINIFIT-10 L1 -15V 4.7 uH +5V_SYS C16 A Printed documents are uncontrolled + C46 0.68 uF 4.7uF, 16v 100E/200E PMT PREAMP PCA Schematic Size B Date: File: 1 04521C (DCN5731) 2 3 4 A Title 5 Number 04181 Revision H 10-May-2007 Sheet 1 of N:\PCBMGR\04179cc\Source\RevG\04179.ddb Drawn By: 3 6 D-25 1 2 3 4 6 5 D D VPMT 5 TP9 * 6 11 NC3 14 NC2 +15V 3 NC1 C31 0.68 uF 8 7 9 10 16 15 1 2 IN 4 COM4 IN 3 COM3 IN2 COM2 IN1 COM1 2 74AHC1GU04 U17 4 HIGAIN 13 12 4 -15V ETEST ETEST ETEST PREAMP2 HIGAIN DG444DY +15V U5 4 ETEST 2 HIGAIN -15V 74AHC1GU04 4 PREAMP1 NC4 V+ V(L) V- ETEST_SIGNAL GND U4 U9A 3 +5V_SYS C29 0.68 uF 1 2 -15V C 8 8 C LF353 U16B R11 100M 6 C4 0.001 uF 5 7 100 pF R48 1K R46 100 TP1 * 4 C2 +15V LF353, OPAMP R5 R29 50k, POT 1000M N/I, SHORTED R12 TP8 * +15V C28 10uF/25V +15V R50 N/I R44 + PREAMP2 SEE TABLE C48 R3 1 PMT Signal Connector 2 2 4.99K C5 0.68 uF U1 6 TP7 * SEE TABLE For 1.0 uF use C11. For 11 uF use C11A & C11B. PREAMP1 3 COAX R17 SEE TABLE 4 OPA124 C2710uF/25V R4 TP6 * C30 0.68 uF -15V ETEST_SIGNAL 3 R19 10K, POT A 1 VERSION TABLE: 0100 - M10XE 0200 - M20XE 3 R38 N/I 2 COMP. 0100 0200 ---------------------------------------------R17 20.0K 10.0 ohms R44 39.2K 25.5K R51 10K not installed C3 0.1 uF 0.012 C11 11.0 1.0 ELECT. TEST 1 D-26 R13 N/I, POT 2 100 SPAN ADJUST R43 4.99K 2 1.0uF C11 1 -2.5V C36 0.1 uF 5 LF353, OPAMP 250K C3 SEE TABLE U11 1 2 3 4 7 R36 + PMTGND BUFOUT FB OUT AGND V+ VDIV RATIO C OSC 8 7 6 5 LTC1062CN8 B U2B 6 + C11B 22uF/25V + C11A 22uF/25V 8 VREF 0.1 uF 8 J1 B PMTGND TP2 * 4 7 GUARD RING -15V C47 0.68 uF +12V_REF C9 3900 pF, FILM R51 SEE TABLE PMTGND NOTES: 1. UNLESS OTHERWISE SPECIFIED CAPACITANCE IS IN MICROFARADS. A Printed documents are uncontrolled PMTGND 2. RESISTORS ARE 1%, 1/4W. 3. RESISTANCE IS IN OHMS. 4. THIS CIRCUIT MUST BE USED AS A MATCHED PAIR WITH THE TEC CONTROL CIRCUIT Title M100E/200E PMT Preamp PCA Schematic Size 3 B Date: File: 4 5 Number Revision 04181 H Sheet 2 10-May-2007 of N:\PCBMGR\04179cc\Source\RevG\04179.ddb Drawn By: 3 6 04521C (DCN5731) 1 2 3 C45 4 6 5 HIGH VOLTAGE SUPPLY 100pF TP4 * VREF D R42 4.99K U16A 2 8 3 3 LF353, OPAMP 0.68 uF 2 Vrf(+) 4 R49 1.0K Vrf(-) C33 0.68 uF 16V 4 COMP 5 C24 0.1 uF TC 7 Vee -15V C GND C20 Iout 1 1 D7 K A C22 10uF/25V 4 IN 2 1 C51 0.1uF/ 50V CA0000192 U6 2 + R20 4.99K Vcc HVPS D 1 8 0.1 uF 3.92K C32 1.0uF/16V CA0000199 +5V_LOCAL C25 OUT GND GND 6 C7 0.68 uF +15V R47 +15V U22 LT1790AIS6-5 4.99K 9 10 11 12 13 14 15 16 D7 D6 D5 D4 D3 D2 D1 D0 9 8 7 6 4 3 2 1 RN1 C R33 5 10 100Kx8 +5V_LOCAL C DAC0802 8 6 -15V U9B 6 7 5 1 4 3 6 1 4 3 6 4 LF535 1 2 4 8 1 2 4 8 S2 S1 B B OUT 1 1 3 LM78L12ACZ(3) C34 10uF/25V + 2 + C15 10uF/25V OUT IN ON/OFF NC GND IN 5 2 2 +5V_LOCAL TP10 * U14 5 4 LP2981IM5 + 2 3 +15V GND U8 5 +12V_REF TP5 * C14 10uF/25V 2 C42 0.68 uF D6 11DQ05 C50 10uF/25V TP12 * 1 3 -2.5V A Printed documents are uncontrolled VR1 LM336Z-2.5 R24 2k M100E/200E PMT PREAMP PCA Schematic Size B -15V 1 04521C (DCN5731) 2 A Title 3 Date: File: 4 5 Number Revision 04181 H 10-May-2007 Sheet 3 of N:\PCBMGR\04179cc\Source\RevG\04179.ddb Drawn By: 3 6 D-27 1 2 3 4 A A B B JP1 R1 Not Used R2 22 1 2 3 4 5 6 7 8 C C Title D Size A Date: File: 1 D-28 2 3 SCH, E-Series Analog Output Isolator, PCA 04467 Number Revision 04468 6/28/2004 N:\PCBMGR\..\04468B.sch D B Sheet of Drawn By: 4 04521C (DCN5731)
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.6 Linearized : Yes Author : Karen Saucedo Category : Manuals Comments : Company : Teledyne-API Create Date : 2010:05:17 20:24:10-07:00 Keywords : M200E, Operator's, Manual Manager : Modify Date : 2010:05:20 12:41:24-07:00 Subject : Has XFA : No XMP Toolkit : Adobe XMP Core 4.2.1-c043 52.372728, 2009/01/18-15:08:04 Creator Tool : Acrobat PDFMaker 9.1 for Word Metadata Date : 2010:05:20 12:41:24-07:00 Producer : Acrobat Distiller 9.3.0 (Windows) Format : application/pdf Creator : Karen Saucedo Title : Teledyne API - Model 200EH/EM Operation Manual Description : Document ID : uuid:bb05150a-d1ec-4680-8a09-96fd03791c58 Instance ID : uuid:1a26cc3f-a550-45f7-9e8b-0cadbc02637e Page Layout : OneColumn Page Mode : UseNone Page Count : 370EXIF Metadata provided by EXIF.tools