United Imaging Healthcare MARS1717XU Wireless digital flat panel detector User Manual Title

Shanghai United Imaging Healthcare Co., Ltd. Wireless digital flat panel detector Title

User Manual

     Wireless Digital Flat Panel Detector Mars1717XU-VSI  User Manual                       Before operating, please read this user manual and pay attention to all safety precautions.   Please ensure that this user's manual is properly maintained so that it can be accessed at any time ( reserve ).   Please use it correctly on the basis of full understanding of the content.  E
To Customers iRay Technology Co. Ltd.                                                                                                                                                                                       1   Congratulations  on  your  purchase  of  the  Mars1717XU-VSI  Wireless  Digital  Flat  Panel detector  (hereinafter  referred  to  as  Mars1717XU)  which  is  manufactured  by  iRay Technology  Co.Ltd.  (Hereinafter referred to as iRay). Please take time to read through this user guide in order to utilize the product effectively. We hope you enjoy the experience with iRay Mars1717XU. If you have any questions or suggestions, please feel free to contact us.    Notes on usage and management of the equipment 1.  Read all of the instructions in the user guide before your operation. Give particular attention to all safety precautions. 2.  Only a physician or a legally certified operator should use this product. 3.  The equipment should be maintained in a safe and operable condition by maintenance personnel. 4.  Use  only  computers  and  image  display  monitors  complying  with  IEC 60601-1 or IEC 60950-1. For details, consult our sales representative or local iRay dealer. 5.  Use only the dedicated cables. Do not use any cables other than those supplied with this product. 6.  The equipment should not be used in inflammable gas or corrosive gas environment. 7.  Give  particular  attention  to  avoiding  liquid  or  conductive  materials invading into inside of the equipment in order to prevent the equipment from short circuit. 8.  House of the equipment forbids opening by unauthorized personnel. 9.  Request  your  sales  representative  or  local  iRay  dealer  to  install  this product。   Service Office Tel: +86 21 50720560 Fax: +86 21 50720561 E-mail: service@iraychina.com  Location: 2F, Building 9, No.590, Ruiqing Rd, Pudong, Shanghai,
Mars1717XU Wireless Digital Flat Panel Detector User Manual 2                                                                                                                                                                                       iRay Technology Co. Ltd.        Caring for your environment This  symbol  indicates  that  this  product  is  not  to  be  disposed  of  with  your residential or commercial waste.  Recycling iRay Equipment Please  do  not  dispose  of  this  product  with  your  residential  or  commercial waste. Improper handling of this type of waste could have a negative impact on health and on the environment. Some countries or regions, such as the European  Union,  have  set  up  systems  to  collect  and  recycle  electrical  or electronic  waste  items.  Contact  your local  authorities  for information  about practices established in your region. If collection systems are not available, call iRay Customer Service for assistance. Disclaimer 1.  iRay shall not be liable to the purchaser of this product or third parties for any damage, losse, or injury incurred by purchaser or third parties as a  result  of  fire,  earthquake,  any  accident,  misuse  or  abuse  of  this product. 2.  iRay  shall  not  be  liable  to  any  damage,  loss,  or  injury  arising  from unauthorized  modifications,  repairs,  or  alterations  to  this  product  or failure  to  strictly  comply  with  iRay’s  operating  and  maintenance instructions. 3.  iRay shall not be liable for any damage or loss arising from the use of any  options  or  consumable  products  other  than  those  dedicated  as Original iRay Products by iRay Technology. 4.  It is the responsibilities of the user/attending physicians for maintaining the  privacy  of  image  data  and  providing  medical  care  services.  iRay shall  not be  responsible  for the  legality  of image processing ,  reading and storage  nor it  shall  be  responsible for  loss of  image  data for  any reason. 5.  Information  regarding  specification,  compositions,  and  appearance  of this product is subject to change without prior notice.。 Copyright All rights reserved No part of this publication may be reproduced in any form or by any means without the  written permission  of iRay. The  information contained  herein is designed only for use with iRay Mars1717XU. Trademarks The iRay name and iRay logo are registered trademarks of iRay Technology  Co.Ltd.
To Customers iRay Technology Co. Ltd.                                                                                                                                                                                          3 Symbols and Conventions The following symbols and conventions are used throughout the user guide.  This symbol is used to identify conditions under which improper use of the product may cause death or serious personal injury.  This notice is used to identify conditions under which improper use of the product may cause minor personal injury.  This notice is used to identify conditions under which improper use of the product may cause property damage.  This is used to indicate a prohibited operation.  This is used to indicate an action that must be performed.  This is used to indicate important operations and restrictions.  This is used to indicate operations for reference and complementary information.
Mars1717XU Wireless Digital Flat Panel Detector User Manual 4                                                                                                                                                                                          iRay Technology Co. Ltd..  Labels and markings on the equipment The contents of the labels and markings on iRay Mars1717XU product are indicated below: Symbol Indication   This symbol is used to identify the manufactuer's series number which is after, below or adjacent to the symbol. The series number of iRay products is usually made of thirteen digits as shown below:             This symbol is used to indicate the name and address of the manufacturer,.the date of manufacture can be combined in this symbol.  This symbol is used to indicate consultation of the user guide for general information.  Safety Signs: please refer to the user guide for safety instructions.  This symbol is used to indicate detector is allowed to withstand 100 kg on the surface.  Caution: please refer to the instructions in the user manual.  This symbol is used to indicate the operational temperature limits.  This symbol is used to indicate non-ionizing radiation. Numerical Order Year Date Month Version Product Code A1A2A3A4      C1C2  M DD  Y   XXX
To Customers iRay Technology Co. Ltd.                                                                                                                                                                                       5  This symbol is used to indicate when end-user plans to discard this product, this product should be sent to the appropriate facilities for recovery and recycling. IPX4 This symbol is used to indicate the degrees of protection provided by enclosures.  This symbol is used to indicate B type applied part.  Handled with care.  Caution: refer to appending files for important information about safety, such as warnings and matters needing attention which is not suitable to be shown on the instrument for various reasons.  Package symbol, fragile, handle with care.  Package symbol, keep away from sunlight.  Package symbol, keep dry.  Package symbol on the the detector packing-case: storage temperature limits    Package symbol, this symbol is used to indicate the humidity limits.  Package symbol, keep the equipment up right.
Mars1717XU Wireless Digital Flat Panel Detector User Manual 6                                                                                                                                                                                          iRay Technology Co. Ltd..   Package symbol, do not roll the transportation package.  Package symbol, this symbol is used to indicate stacking limit number.
Catalogue    iRay Technology Co. Ltd.                                                                                                                                                              7        Contents  CONTENTS ....................................................................................................................... 7 1. SAFETY INFORMATION ............................................................................................ 9 1.1 Safety precautions ........................................................................... 10 1.2  Notes for Using ................................................................................ 14 2. GENERAL DESCRIPTION ........................................................................................... 16 2.1  Scope ............................................................................................... 17 2.2  Characteristic ................................................................................... 17 2.3  Intended use .................................................................................... 18 2.4  Standard Product Components ....................................................... 18 2.5  Components Description.................................................................. 19 2.6  Product Specification ....................................................................... 20 3. INSTALLATION ............................................................................................................ 24 3.1  Detector Installation ......................................................................... 25 3.2  Software Installation ......................................................................... 27 3.3  Detector Configuration ..................................................................... 27 4. OPERATION ................................................................................................................. 38 4.1 Main Operation ................................................................................. 39 4.2 Block diagram ................................................................................... 39 4.3 Work Flow ......................................................................................... 40 4.4 Description of workflow ..................................................................... 40 5. REGULATORY INFORMATION .................................................................................. 42 5.1 Registrant Information ....................................................................... 43 5.2 Manufacturer Information .................................................................. 43 5.3 Medical equipment safety standards ................................................ 43 5.4 References harmonized standards under Directive 93/42/EEC ....... 44 5.5 The compliance for each EMISSIONS and IMMUNITY standard or test specified by IEC60601-1-2 standard ................................................ 45 5.6 Lithium-ion Capacitor Reference Standards ..................................... 47 5.7 Product Label .................................................................................... 48 5.8 Radio Frequency Compliance Information ....................................... 48 5.9 INFORMATION TO USER. ............................................................... 49 6. SOFTWARE OPERATION INFORMATION ................................................................ 51 6.1  Default Settings ................................................................................ 52
Mars1717XU Wireless Digital Flat Panel Detector User Manual  8                                                                                                                                                                                          iRay Technology Co. Ltd..  7. TROUBLE SHOOTING ................................................................................................. 55 8. SERVICE INFORMATION ............................................................................................ 56 8.1  Product lifetime ................................................................................ 57 8.2  Regular inspection and Maintenance ............................................... 57 8.3  Repair ............................................................................................... 58 8.4  Replacement parts support .............................................................. 58 APPENDIX ........................................................................................................................ 59 Appendix A Information of Manufactures ................................................ 60
1. Safety Information     iRay Technology Co. Ltd.                                                                                                                                                            9                                                                                                                                                                                                   1.  Safety Information 1.1 Safety precautions ........................................................................... 10 1.2  Notes for Using ................................................................................ 14
Mars1717XU Wireless Digital Flat Panel Detector User Manual  iRay Technology Co. Ltd.                                                                                                                                                                                  10 1.1  Safety precautions Follow these safeguards and properly use the equipment to prevent injury and damage to any equipment/data. WARNING Installation and environment of use  Do not use or store the equipment near flammable chemicals such as alcohol, thinner, benzene, etc. If chemicals are spilled or evaporate, it may result in fire or electric shock through contact with electric parts inside the equipment. Also, some disinfectants are flammable. Be sure to take care when using them.  Do not connect the equipment with anything other than specified. Doing so may result in fire or electric shock.  All the patients with active implantable medical devices should be kept away from the equipment. Power supply   Do not operate the equipment using any type of power supply other than the one indicated on the rating label. Otherwise, it may result in fire or electric shock.  Do not handle the equipment with wet hands. You may experience electric shock that could result in death or serious injury.  Do not place heavy object such as medical equipment on cables and cords. Do not pull, bend, bundle, or step on them to prevent their sheath from being damaged, and do not alter them neither. Doing so may damage the cords which could result in fire or electric shock.  Do not supply power to more than one piece of equipment using the same AC outlet. Doing so may result in fire or electric shock.  Do not turn ON the system power when condensation has formed on the equipment. Doing so may result in fire or electric shock.  Do not connect a multiple portable socket-outlet or extension cord to the system. Doing so may result in fire or electric shock.  To avoid the risk of electric shock, this equipment must only be connected to power supply with protective earth. Not doing so may result in fire or electric shock.   Securely plug the power cord into the AC outlet. If contact failure occurs, or if metal objects come into contact with the exposed metal prongs of the plug, fire or electric shock may result.  Be sure to turn OFF the power to each piece of equipment before connecting or disconnecting the cords. Otherwise, you may get an electric shock that could result in death or serious injury.  Be sure to hold the plug or connector to disconnect the cord. If you pull the cord, the core wire may be damaged, resulting in fire or electric shock.
1. Safety Information iRay Technology Co. Ltd.                                                                                                                                                                              11 WARNING Handling   Never disassemble or modify the equipment. No modification of this equipment is allowed. Parts of the Mars1717XU that are not serviced or maintained while in use with the patient. Doing so may result in fire or electric shock. Also, since the equipment incorporates parts that may cause electric shock as well as other hazardous parts, touching them may cause death or serious injury.  Do not place anything on top of the equipment. The object may fall and cause an injury. Also, if metal objects such as needles or clips fall into the equipment, or if liquid is spilled, it may result in fire or electric shock.  Do not hit or drop the equipment. The equipment may be damaged if it receives a strong jolt, which may result in fire or electric shock if the equipment is used without being repaired.  Do not put the equipment and pointed objects together. The equipment may be damaged. If so, the equipment should be used in bucky.   Have the patient take a fixed posture and do not let the patient touch parts unnecessarily. If the patient touches connectors or switches, it may result in electric shock or malfunction of the equipment. When a problem occurs   Should any of the following occurs, immediately unplug the power cord of Control Box, and contact your sales representative or local iRay dealer: When there is smoke, an odd smell or abnormal sound. When liquid has been spilled into the equipment or a metal object has entered through an opening. When the equipment has been dropped and damaged. Maintenance and inspection    Please turn OFF the power of the equipment and unplug the power cord of adaptor before cleaning.  NEVER use alcohol, ether and other flammable cleaning agent for safety. NEVER use methanol, benzene, acid and base because they will erode the equipment.  DON'T dip the equipment into the liquid.  Please make sure that the equipment's surface & plugs are dry before turning ON. Otherwise, it may result in fire or electric shock.   Clean the plug of the power cord periodically by unplugging it from the AC outlet and removing dust or dirt from the plug, its periphery and AC outlet with a dry cloth. If the cord is kept plugged in for a long time in a dusty, humid or sooty place, dust around the plug will attract moisture; this could cause insulation failure that may result in a fire.  For safety reasons, be sure to turn OFF the power to each piece of equipment when performing inspections indicated in this manual. Otherwise, electric shocks may occur.
Mars1717XU Wireless Digital Flat Panel Detector User Manual 12                                                                                                                                                                                         iRay Technology Co. Ltd. CAUTION Installation and environment of use   Do not install the equipment in any of the locations listed below. Doing so may result in failure, malfunction, equipment falling, fire or injury. Close to facilities where water is used Where it will be exposed to direct sunlight Close to the air outlet of an air-conditioner or ventilation equipment Close to heat source such as a heater Where the power supply is unstable In a dusty environment  In a saline or sulfurous environment Where temperature or humidity is high Where there is freezing or condensation In areas prone to vibration  On an incline or in an unstable area  Take care that cables do not become tangled during use. Also, be careful not to get your feet caught by cable. Otherwise, it may cause a malfunction of the equipment or injury of the user due to tripping over the cable.        Power supply   Always connect the three-core power cord plug to a grounded AC power outlet.  To make it easy to disconnect the plug at any time, avoid putting any obstacles near the outlet. Otherwise, it may not be possible to disconnect the plug in an emergency.  Be sure to ground the equipment to an indoor grounded connector. Also, be sure to connect all the grounds for the system to a common ground.  Do not use any power source other than the one provided with this equipment. Otherwise, fire or electric shock may be caused due to leakage.  Handling  Do not spill liquid or chemicals onto the equipment. In case the patient is injured, it is not allowed to contact with blood or other body fluids. Doing so may result in fire or electric shock. In such a situation, protect the equipment with a disposable cover as necessary.  Turn OFF the power and pull out the plug to each piece of equipment for safety when not used.
1. Safety Information iRay Technology Co. Ltd.                                                                                                                                                                              13 CAUTION  Handling  Handle the equipment carefully.  Do not submerge the equipment in water.  The internal image sensor may be damaged if something hits against it or it is dropped.              Do not place excessive weight on the equipment. Otherwise, the internal image sensor may be damaged and image collection may be malfunction.                           Be sure to use the equipment on a flat surface so it will not bend. Otherwise, the internal image sensor may be damaged. Be sure to securely hold the detector while using it in upright positions. Otherwise, the detector may flip over, resulting in injury to the user or patient as well as damage of the inner device.        Keep the same load (same pressure) on the detector when acquiring the image. Or the image will be incorrect. CAUTION <Load Limit> Uniform load:135 kg over the whole area of the detector surface. Local load:100 kg on an area 4 cm diameter.
Mars1717XU Wireless Digital Flat Panel Detector User Manual 14                                                                                                                                                                                         iRay Technology Co. Ltd.   Do not close to fire, do not use in high temperature  Do not invert positive and negative pole   Do not contact with metal in case of short circuit  Do not insert sharp objects into the Lithium-ion capacitor  Do not beat the Lithium-ion capacitor  Do not squeeze the Lithium-ion capacitor  Do not use the Lithium-ion capacitor out of rules  Do not change internal structure of the  Lithium-ion capacitor  Do not submerge the  Lithium-ion capacitor in water, please keep dry in storage and do not contact with water in use  Please charge the  Lithium-ion capacitor with charger following IEC60601-1 Standards provide by us  Do not mix the  Lithium-ion capacitor with ones not provided by iRay  Do not charge the  Lithium-ion capacitor already damaged    Adverse events may be caused by the product Due to operation or corresponding device malfunction, no valuable clinical image is acquired after exposure accomplished. Due to disturbance from a device which does not complied with IEC60601-1-2 standard , functions of the product are abnormal, which results in no valuable clinical image is acquired after exposure accomplished   The product is intended to be used with registered X-ray instruments. Please refer to this manual for installment and software operation part of the product.   For the rest part of operation, please refer to the operation manual of the applied X-ray instrument.   Means for monitoring, evaluation, and control  The product status indication is referred in “Product installation, connection” part of this manual. If the indication is not correct, please do not use.   The measuring function must be re-calibrated after exchanging or re-installation of FPD.  The measuring function should be calibrated every half year if the product structure is not disassembled.  1.2  Notes for Using When using the equipment, take the following precautions. Otherwise, problems may occur and the equipment may not function correctly.  Before exposure   Be sure to check the equipment daily and confirm that it works properly.
1. Safety Information iRay Technology Co. Ltd.                                                                                                                                                                              15   Sudden heating of the room in cold areas will cause condensation to form on the equipment. In this case, wait until the condensation evaporates before performing an exposure. If the equipment is used while condensation is formed on it, problems may occur in the quality of captured images. When an air-conditioner is used, be sure to raise/lower the temperature gradually so that a difference of temperature in the room and equipment does not occur, to prevent condensation.   The detector should warm up for 30 minutes before exposure or updating the gain map or defect map.  During exposure   Do not move the power or Ethernet Cables during exposure, or it may cause image noise or artifacts, even incorrect images.   Do not use the devices near the equipment generating a strong magnetic field. Otherwise, it may cause image noise, artifacts or even incorrect images.  After exposure   Recommend to shut down the detector’s power for storage when no use.  Disinfection and Cleaning   After every examination, wipe the patient contact surfaces of the detector using disinfectants such as ethanol, to prevent the risk of infection. For details on how to sterilize, consult a specialist.   Do not spray the detector directly with disinfectants or detergents.   Wipe it with a cloth slightly damped with a neutral detergent. Do not use solvents such as alcohol, thinner, benzene, acid and base. Doing so may damage the surface of the equipment.  It's recommended to use a waterproof non-woven cover as the isolated layer between detector and the blooding patient.
2. General Description  16                                                                                                                                                                                        iRay Technology Co. Ltd.                                                                                         2. General Description 2.1  Scope ............................................................................................... 17 2.2  Characteristic ................................................................................... 17 2.3  Intended use..................................................................................... 18 2.4  Standard Product Components ........................................................ 18 2.5  Components Description .................................................................. 19 2.5.1 Detector .................................................................................. 19 2.5.2 Power Supply ......................................................................... 20 2.6  Product Specification ....................................................................... 20 2.6.1 Detector .................................................................................. 20 2.6.1.1  Basic ................................................................................... 20 2.6.1.2  MTF .................................................................................... 21 2.6.1.3  DQE .................................................................................... 21 2.6.2  Power supply ......................................................................... 21 2.6.3  AP Router (Optional) ............................................................. 22 2.6.4  Wireless Communication ...................................................... 22 2.6.5  Recommended Appliance Condition ..................................... 23 2.6.6  Use Environment ................................................................... 23
Mars1717XU Wireless Digital Flat Panel Detector User Manual  iRay Technology Co. Ltd.                                                                                                                                                                                   17                                                                                                                                                                                  Mars1717XU is a cassette-size wireless X-ray flat panel detector based on amorphous silicon thin-film transistor technologies. It is developed to provide the highest quality of radiographic image, which contains an active matrix of 3072×3072 with 139um pixel pitch. Detectors’ scintillator is CsI(Caesium Iodide). The greatest improvement is Mars1717XU supports wireless communication between detectors and PC, and employs Lithium-ion capacitor, which makes Mars1717XU a real portable panel.   2.1  Scope This manual contains information about the iRay Mars1717XU. All operators must read and understand this manual before using equipment. All information in this manual, including the illustrations, is based on equipment prototype. If configuration of your equipment does not have any of these items, information about these items in the manual does not apply to your equipment.              2.2  Characteristic Mars □  □   □ VSI- CsI scintillator screen Product Application:XU, cassette size detector Product dimension:1717,17inch×17inch Product  serials:wireless  digital  flat  panel  detector serials   Wireless static Flat Panel Detector used for general radiography.   17 × 17 inch   Built-in non-detachable Lithium-ion capacitor
Mars1717XU Wireless Digital Flat Panel Detector User Manual 18                                                                                                                                                                                         iRay Technology Co. Ltd.  2.3  Intended use The Mars1717XU Wireless Digital Flat Panel detector (hereinafter referred to as Mars1717XU) is compatible with digital medical X-ray radiography system for a medical institute to acquire digital X-ray imaging of patient body. The Flat Panel Detector (FPD) of Mars1717XU is one of the flat panel detector series based on amorphous silicon technology, which are used in DR system to provide clinical X-ray images. The host computer of Mars1717XU-VSI flat panel detector employs Caesium Iodide (CsI) scintillation screen, and provides WIFI data port and DC power supply. This panel is not intended for mammography or dental applications,and prohibited for pregnant women. According to the Mars1717XU intended use and the result of risk management, identifying and describing the essential performance as the following: a)  To perform communication and instruction transmission between the flat plane detector and computer system through wireless router b) To perform imaging acquisition by the flat panel detector This equipment provides digital X-ray imaging for diagnosis of disease, injury, or any applicable health problem. The image is obtained as the result of X-ray passing through the human body and detected by the equipment. iRay will provide equipment and software support for integration of system. The length of\ DC Power Cable cannot exceed 3.5 m. Or the impedance of protective earth connections may exceed the safety threshold.  2.4  Standard Product Components Mars1717XU comes with power supply by both 12V DC and bulit-in Lithium-ion capacitor.  Once powered on, it would build a connection with PC through Wireless connection. Item Description Flat Panel Detector  Mars1717XU
2. General Description iRay Technology Co. Ltd.                                                                                                                                                                              19 CD-Rom   Defect correction map SDK Manual  2.5  Components Description 2.5.1 DETECTOR   Item Name Description A DC Input Interface DC input B Detector Indicator Power/Internet Connection/Status C Detector indicator of detector D Button Power button / Reset  Switch hole E Front surface film In line with biological compatibility F Middle frame Supporting and fixation G Back surface film In line with biological compatibility H Detector Label Product information.
Mars1717XU Wireless Digital Flat Panel Detector User Manual 20                                                                                                                                                                                         iRay Technology Co. Ltd.  2.5.2 POWER SUPPLY Mars1717XU supports both DC and bulit-in Lithiu-ion capacitor power supply. The DC power supply is not in the product package, and the external power adapter is only used for customer service maintenance and production.  2.6  Product Specification  2.6.1 DETECTOR 2.6.1.1  BASIC Item Specification Model Mars1717XU-VSI  Image Sensor a-Si (Amorphous Silicon) TFT Pixel Size 139 μm Pixel Matrix 3072 x 3072 Effective Area (H x V) 427 x 427 mm Greyscales 16bit Spatial Resolution 3.6 Lp/mm  Preview Acquisition Time@Client (5GHz) 2.5 sec (since exposure ends) Processed Acquisition Time@Client (5GHz) 5 sec. (since exposure ends) Cycle Time Min. 8.5s Power Consumption Max. 30W (including charging) Stand-by Power Consumption Max.5W Charging Time Max. 50minutes Discharging Time Cycle time=30s, 30min, 60images Cycle time=90s, 45min, 30images Exposure Preparation Time 1s Sleeping Awakening Time 2s Sleeping Endurance Time 1h Dimension (L × W × H) 460 x 460 x 15.4 mm Weight 3.9kg
2. General Description iRay Technology Co. Ltd.                                                                                                                                                                              21 WIFI IEEE802.11a/b/g/n/ac (2.4G/5G) X-ray Energy 40kV to 150kV  2.6.1.2  MTF Spatial Frequency (lp/mm) MTF(CsI) 1.0 0.65 2.0 0.43 3.0 0.17  2.6.1.3  DQE Spatial Frequency (lp/mm)@RQA5 DQE (CsI)@2.5uGy 0 0.65 1.0 0.45 2.0 0.30 3.0 0.20   2.6.2  POWER SUPPLY Mars1717XU supports both DC and built-in Lithium-ion capacitor power supply.. Item Specifications DC Power 12V(DC),2.5A Lithium-ion Capacitor 7.6V(DC),0.93A
Mars1717XU Wireless Digital Flat Panel Detector User Manual 22                                                                                                                                                                                         iRay Technology Co. Ltd. 2.6.3  AP ROUTER (OPTIONAL) Mars1717XU  does  not  include  a  wireless  AP  Router.  Users  can  choose according to their requirement, but the following specification must be matched. Item Specifications Wireless Standard IEEE 802.11 a/b/g/n/ac Frequency Range 2.400  ~  2.4835  GHz  and  5.15~5.25  GHz 5.725~5.85 GHz Wireless Data Rate 802.11b : Max. 11Mbps 802.11a/g : Max. 54Mbps 802.11n : Max. 300Mbps (MIMO 2x2)   2.6.4  WIRELESS COMMUNICATION Item Description Wireless Standard IEEE802.11a/b/g/n/ac Frequency Range 2.4G: 2.400 ~ 2.4835 GHz @CH1-CH13 HT20 Section @CH3-CH11 HT40 Section 5G: 5.15~5.25 GHz and 5.725~5.85 GHz @CH36-CH165 HT20 Section @CH38-CH163 HT40 Section Modulation 802.11b: CCK, DQPSK, DBPSK 802.11a/g: 64 QAM, 16 QAM, QPSK, BPSK 802.11n: 64 QAM, 16 QAM, QPSK, BPSK Antenna 2 Dual Band internal antenna
2. General Description iRay Technology Co. Ltd.                                                                                                                                                                              23 2.6.5  RECOMMENDED APPLIANCE CONDITION Item Description Operating System Windows 8  64bit CPU Intel Core i7 3.6G Memory 8G DDR3 Hard Disk 640 G LAN Card Intel Pro EXP9301CT PRO   2.6.6  USE ENVIRONMENT  Operating Storage Temperature 5~35℃ -30~55℃ Temperature change <1k/min <1k/min Humidity 30%~80% RH 10%~90% RH Atmospheric Pressure 700~1060hPa 700~1060hPa Pressure Change <10kp/min (1kp=1.0197E-5Pa) <10kp/min (1kp=1.0197E-5Pa) The Mars1717XU serial detectors shall operate at an altitude specified NO more than 3000m.
 3. Installation and Operation  iRay Technology Co. Ltd.                                                                                                                                                                                               24                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          3. Installation  3.1  Detector Installation ......................................................................... 25 3.1.1 Attach dc power...................................................................... 25 3.1.1 booting up ............................................................................... 25 3.2  Software Installation ......................................................................... 27 3.3  Detector Configuration ..................................................................... 27 3.3.1 AP Mode ................................................................................. 27 3.3.1.1 External Wireless LAN Setting ............................................ 27 3.3.1.2 AP Mode Detector Configuration ........................................ 29 3.3.2 Client Mode ............................................................................ 32 3.3.2.1 Configuration of External wireless router ............................ 32 3.3.2.2 Client mode configuration of detector ................................. 36
 Mars1717XU Wireless Digital Flat Panel Detector User Manual       25                                                                                                                                                                                              iRay Technology Co. Ltd.                                                                                                                                                                                                                                                  3.1  Detector Installation 3.1.1 ATTACH DC POWER Insert the DC power cable into the DC input port of the detector     3.1.1 BOOTING UP On the control panel, user can press power button to power on/off the detector. If detector is powered off, users can press the button for 3 seconds to power on the detector when a Lithium-ion capacity is inserted and the Lithium-ion capacitor is no less than 20%, or direct current power is connected. If  detector  is  powered  on,  users can press the button for 5 seconds to  shut  down the  detector.  On  the other hand, it can also be used as  internal  control  IC  reset  button when button is active for 8s.    After booting up the detector, user can check the status LED indicator of detector.
 Mars1717XU Wireless Digital Flat Panel Detector User Manual 26                                                                                                                                                                                         iRay Technology Co. Ltd. The power indicator C shows as following: Power Indicator Lighting Status Operating Battery Capacity DC Input OFF  Power OFF/Sleep / / Orange On  Power ON/Wake up ≤30% NO Green  On  Power ON/Wake up ≧30% NO Orange Slow Blinking  Power ON ≤30% YES Green Slow Blinking  Power ON ≧30% YES OFF after Orange ON with 1sec  Power OFF ≤20% NO OFF after Green ON with 1sec  Power OFF ≧20% NO   Link indicator D shows as following: LINK Indicator Lighting Status Description OFF  Shut down/Wired&Wireless connection not ready Blue ON  Client mode, Wireless connection is ready Blue Blinking  AP mode, Wireless AP is ready Green ON  Wired Connection is ready Green Blinking  Or RST key(AP/Client switch) active   Status indicator E shows as following: Status indicator Lighting Status Description OFF  Shut down/Idle Green ON  Data Transmission/RST Key(Recovery) active Orange ON  ERROR
 3. Installation and Operation   iRay Technology Co. Ltd.                                                                                                                                                                       27 KEY button as following table: PWB key-B; RST key-A Key  Push Time Description PWB key ≤1.5s None action ≧1.5s - <3s Power ON or Wake up ≧3s -  <7.5s Power OFF ≧7.5s Power Reset Key  Push Time Description RST key <3s None action ≧3s - <7.5s AP/Client mode switch ≧7.5s  Recovery Factory Parameters(include AP address)   3.2  Software Installation In  the  case  of  IDetector  doesn’t  work,  please  install  following  VC  redistribute package .NET Framework. ( IDetector should not be used for terminal hospital)  1.  Install Microsoft  .NET Framework 4.5  before first  use, which needs to be downloaded from the Microsoft website. 2.  Need  to  install  VC  redistribution  package  vcredist_x86_2013  (or vcredist_x64_vs2013).   3.3  Detector Configuration 3.3.1 AP MODE AP Mode is the default set mode of the product during purchasing. The default SSID  is  XU-SN,  password  is  12345678,  and  IP  address  is  192.168.8.8.  The detector will directly connect to wireless LAN when power on. A user should accomplish wireless client configuration of AP mode as following steps:  3.3.1.1 EXTERNAL WIRELESS LAN SETTING
 Mars1717XU Wireless Digital Flat Panel Detector User Manual 28                                                                                                                                                                                         iRay Technology Co. Ltd. Open the wireless signal list  Select  SSID  belong  to  the detector  and  input  the password  Open wireless network status
 3. Installation and Operation   iRay Technology Co. Ltd.                                                                                                                                                                       29 open IPV4 setting   IP setting Network mask setting   IP address: 192.168.8.188 Subnet mask: 255.255.255.0 Open  SDK,  select  a corresponding  product  and start  connection  by  clicking “connect”    3.3.1.2 AP MODE DETECTOR CONFIGURATION Connect  the  detector  and Indetector  in  default  AP mode
 Mars1717XU Wireless Digital Flat Panel Detector User Manual 30                                                                                                                                                                                         iRay Technology Co. Ltd. Click  “Detector”  button  of IDetector software  Click “wifi” button  Select “AP mode”  1.  Click  “Read  Config” button to read the default setting; 2.  Change  SSID  and password  setting  to make  sure  SSID  not  to conflict  with  other existing devices’ SSID;
 3. Installation and Operation   iRay Technology Co. Ltd.                                                                                                                                                                       31 Change  the  channel  and frequency settings.  Click  “Chanel”  to  select  a clean wireless frequency and channel.  Click “write Config”
 Mars1717XU Wireless Digital Flat Panel Detector User Manual 32                                                                                                                                                                                         iRay Technology Co. Ltd. Unplug  wire  cable  until  the status of the panel change to Ready from Busy.  Since the default SSID and password have been changed, a revised AP SSID will be shown when the detector is power on next time.  3.3.2 CLIENT MODE The user needs to complete the wireless client mode configuration according to the  following  operation.  When  different  detectors  are  used,  it  is  strictly forbidden to use the same SSID name for connection.  3.3.2.1 CONFIGURATION OF EXTERNAL WIRELESS ROUTER  Connect  one  end  of  Gigabit Ethernet  Cable  to  PC, Connect another end to LAN port  of  External  wireless router  / Open  network  management interface
 3. Installation and Operation   iRay Technology Co. Ltd.                                                                                                                                                                       33 Open  Local  network configuration  Open IPV4 setting   IPSetting Network mask setting  Select “Obtain an IP address automatically”
 Mars1717XU Wireless Digital Flat Panel Detector User Manual 34                                                                                                                                                                                         iRay Technology Co. Ltd. Open  browser  and  type 192.168.1.1 Log  into  external  wireless router  to  set  configuration through the interface.  Wireless setup  Configure  2.4GHz  wireless network SSID: NETGEAR_BIG_24 (For example, set up a wifi named Mars2.4G) Security: WPA2-PSK Password: 12345678 Channel: [Please check the current Wi-Fi environment, and choose a relatively clean channel] Configure  5GHz  wireless network SSID: NETGEAR_BIG_50 (For example, set up a wifi named Mars5G) Security: WPA2-PSK Password: 12345678
 3. Installation and Operation   iRay Technology Co. Ltd.                                                                                                                                                                       35 Channel: [Please check the current Wi-Fi environment, and choose a relatively clean channel] LAN Setup  Configure LAN IP address IP address: 192.168.100.1 Subnet Mask: 255.255.255.0 External Wireless AP Reboot Apply above settings and reboot your wireless router. Recover  Local  Network  IPv4 setting  of  PC  wired  Ethernet interface  IP setting Network mask setting IP address: 192.168.100.188 Subnet mask: 255.255.255.0
 Mars1717XU Wireless Digital Flat Panel Detector User Manual 36                                                                                                                                                                                         iRay Technology Co. Ltd. 3.3.2.2 CLIENT MODE CONFIGURATION OF DETECTOR Connect to  Idetector through wireless  network  card  under AP mode of the detector  Click  “Detector”  button  of IDetector;  Click  “wifi”  button  and  “Read Config” button; Select “Client mode”; Change  the  default  wireless router:      SSID: IRAY-China      Key: 1234-qwer to current wireless router (for example: set up a wifi named Mars5G)      SSID: NETGEAR_BIG_50      Password: 12345678  Click “write Config ”
 3. Installation and Operation   iRay Technology Co. Ltd.                                                                                                                                                                       37 Reconnect IDetector and the flat panel detector  Enter  acquisition  interface  to set up connection
 4. Operation   iRay Technology Co. Ltd.                                                                                                                                                             38                                                                                                                                                                                                4. Operation  4.1 Main Operation .................................................................................. 39 4.2 Block diagram.................................................................................... 39 4.3 Work Flow ......................................................................................... 40 4.4 Description of workflow ..................................................................... 40
Mars1717XU Wireless Digital Flat Panel Detector User Manual  39                                                                                                                                                                                                 iRay Technology Co. Ltd.                                                                                                                                                                                              Mars1717XU provides SDK for users to integrate detector into their DR system. Additionally, it also provides applications for demonstration, iDetector. User can use iDetector to control detector without DR system.   4.1 Main Operation To Acquire X-ray image is the main operation of Mars1717XU. Most importantly, detector should build synchronization with X-ray generator.Mars1717XU is born with PREP_ACQ2 Mode to acquire X-ray image.   4.2 Block diagram Software mode is the basic way to acquire X-ray image. Please see figure below for general feature PC Wireless APRouterHand SwitchPanel BPanel A Workstation is a host PC device installed with IDetector or SDK. In RREP_ACQ2 Mode, the workstation communicates with GCU through SDK (double panel mode) or the flat panel detector to control X-ray imaging time of X-ray generator.
Mars1717XU Wireless Digital Flat Panel Detector User Manual 40                                                                                                                                                                                         iRay Technology Co. Ltd. 4.3 Work Flow  4.4 Description of workflow 1. Workstation PC receives “Awaken hand switch I prep ” request, sends command “Awaken” to the detector; the detector starts internal clear after receiving “Awaken” command sent from the workstation;  2.  Meanwhile the workstation informs GCU (I/O Box) to control high voltage generator to start anode rotation;   Push down “hand switch I” button GCU  informs  high  voltage  exposure  preparation,  meanwhile  informs  the Panel to prepare for exposure through AP The  panel  starts  an  action  of  “exposure  preparation”;  after  the  action finishes, the panel informs GCU to make exposure enable through AP. Enable exposure, open the window, start “hand switch II” to expose  End the exposure window, upload the acquired image      The system go to sleep  SDK uploads pre-viewed image   SDK uploads full image
 4. Operation    iRay Technology Co. Ltd.                                                                                                                                                                                              41 3. Workstation PC receives “Hand switch II exposure” request, confirms whether GCU anode rotation has ended and whether the detector has completed awaken preparation; 4. If the preparation is completed, the workstation PC asks GCU to send “Hand switch II exposure” request to high voltage generator; 5. Send “Exposure” request to the flat panel detector if preparation of the high voltage generator is confirmed to be completed; 6. The detector sends “Exposure” information to GCU, and then corresponding information will be shown in IDetector information bar, and the user is noticed that the detector is ready to receive X-ray; 7. GCU receives “Exposure” to start generating X-ray, and the X-ray generator starts to release X-ray; 8.  After the X-ray generator finishes X-ray shooting, GCU send “End exposure” request to the detector; 9. The detector receives “End exposure” request from GCU and then perform “Acquire” action; 10.  The detector completes image acquisition and begins to send data to the workstation; 11.  Workstation receives all image data from the detector (including preview image and full image).   Typical Cycle Analysis  Hand Switch I IIIIRotating Anode Filament
5. Regulatory Information   iRay Technology Co. Ltd.                                                                                                                                                                                                 42           5. Regulatory Information 5.1 Registrant Information ....................................................................... 43 5.2 Manufacturer Information .................................................................. 43 5.3 Medical equipment safety standards ................................................ 43 5.4 References harmonized standards under Directive 93/42/EEC ....... 44 5.5 Guidance and manufacture’s declaration for EMC ........................... 45 5.6 Lithium-ion Capacitor Reference Standards ..................................... 47 5.7 Product Label .................................................................................... 48 5.7.1 Detector Host Label ............................................................... 48 5.7.2 Host Housing Label ................................................................ 48 5.8 Radio Frequency Compliance Information ........................................ 48 5.8.1 SRRC Compliance ................................................................. 48
Mars1717XU Wireless Digital Flat Panel Detector User Manual      43                                                                                                                                                                                                iRay Technology Co. Ltd.                                                                                                                                                                                  5.1 Registrant Information Registrant: iRay Technology Co., Ltd . Address: Room 202, Building 9, No.590, Ruiqing Rd, East area of Zhangjiang Hi-tech industrial district, Pudong, Shanghai, PRC, 201201 Contact: +86-21-50720560 After-sales Service Office: iRay Technology Co., Ltd .  Address of After-sales Service Office: Building 45, No.1000, Jinhai Rd, Pudong New Area, Shanghai, PRC, 201206 Contact of After-sales Service Office: +86-21-50720560  5.2 Manufacturer Information Manufacturer: iRay Technology Co., Ltd .  Address: Room 202, Building 9, No.590, Ruiqing Rd, East area of Zhangjiang Hi-tech industrial district, Pudong, Shanghai, PRC, 201201 Production Address: Room 202, Building 9, No.590, Ruiqing Rd, East area of Zhangjiang Hi-tech industrial district, Pudong, Shanghai, PRC, 201201 Contact: +86-21-50720560 Production Licence Number: Shanghai Food and Drug Machinery Safety Production Permission No. 20121946  5.3 Medical equipment safety standards   Medical equipment classification Type  of  protection  against  electrical shock Class  I  Equipment  (using  Mars1717XU medical approved adaptor supply) Internally  powered  Equipment  (using Mars1717XU  Lithium-ion  capacitor  power supply) Degree  of  protection  against  electrical shock B-type Applied Parts Degree  of  protection  against  ingress  of water IPX4(Mars1717XU) Mode of operation Continuous operation There are two modes for power supply and signal transmission of the product: internal
Mars1717XU Wireless Digital Flat Panel Detector User Manual 44                                                                                                                                                                                                  iRay Technology Co. Ltd. Lithium-ion capacitor power supply and adaptor supply, wireless transmission. Any one of power supply modes may combine with wireless transmission.  Not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.  5.4 References harmonized standards under Directive 93/42/EEC Applicable safety standards of Mars1717XU series digital wireless X-ray imaging device cover the product host and other accessories. MDD (93/42/EEC) Medical Device Directive EN ISO 13485:2012/EN ISO 13485:2012/AC:2012 Medical  devices  --- Quality management systems  --- Requirements for regulatory purposes EN ISO14971: 2012 Medical  device  –  Application  of  risk  management  to medical devices IEC 60601-1:2005+ Amendment 1:2012/EN 60601-1:2006+ Amendment 1:2013 Medical  electrical  equipment  --  Part  1:  General requirements  for  basic  safety  and  essential performance IEC  60601-1-2:2014/EN60601-1-2:2015 Medical  electrical  equipment  –  Part  1-2:  General requirements  for basic  safety  and  essential  performance  –  Collateral standard: Electromagnetic disturbances – Requirements and tests IEC 60601-2-54:2015/EN 60601-2-54:2015 Medical  electrical  equipment  --  Part  2-54:  Particular requirements  for  the  basic  safety  and  essential performance of X-ray equipment for radiography and radioscopy IEC 62133:2012 Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made  from them,  for use in portable applications EN 62220-1:2004 Medical electrical equipment - Characteristics of
5. Regulatory Information    iRay Technology Co. Ltd.                                                                                                                                                                          45 digital X-ray imaging devices - Part 1: Determination of the detective quantum efficiency EN 62304:2006/AC:2008 Medical  device  software  -  Software  life-cycle processes EN 62366:2008 Medical devices - Application of usability engineering to medical devices  5.5 The compliance for each EMISSIONS and IMMUNITY standard or test specified by IEC60601-1-2 standard  EMI Compliance Table Emission Phenomenon Compliance Electromagnetic environment RF emissions CISPR 11 Group 1, Class B Professional healthcare facility environment Harmonic distortion IEC 61000-3-2 Class A Professional healthcare facility environment Voltage fluctuations and flicker IEC 61000-3-3 Compliance Professional healthcare facility environment  EMS Compliance Table Enclosure Port Phenomenon Basic EMC standard Immunity test levels Professional healthcare facility environment Electrostatic Discharge IEC 61000-4-2 ±8 kV contact ±2kV, ±4kV, ±8kV, ±15kV air Radiated RF EM field IEC 61000-4-3 3V/m 80MHz-2.7GHz 80% AM at 1kHz Proximity fields from RF wireless communications equipment IEC 61000-4-3 Refer to table “Proximity fields from RF wireless communications equipment” Rated power frequency magnetic IEC 61000-4-8 30A/m
Mars1717XU Wireless Digital Flat Panel Detector User Manual 46                                                                                                                                                                                                  iRay Technology Co. Ltd. fields 50Hz or 60Hz  Proximity fields from RF wireless communications equipment  Test frequency (MHz) Band (MHz) Immunity test levels Professional healthcare facility environment 385 380-390 Pulse modulation 18Hz, 27V/m 450 430-470 FM, ±5kHz deviation, 1kHz sine, 28V/m 710 704-787 Pulse modulation 217Hz, 9V/m 745 780 810 800-960 Pulse modulation 18Hz, 28V/m 870 930 1720 1700-1990 Pulse modulation 217Hz, 28V/m 1845 1970 2450 2400-2570 Pulse modulation 217Hz, 28V/m 5240 5100-5800 Pulse modulation 217Hz, 9V/m 5500 5785  Input d.c. power Port Phenomenon Basic EMC standard Immunity test levels Professional healthcare facility environment Electrical fast transients/burst IEC 61000-4-4 ±2 kV 100kHz repetition frequency Conducted disturbances induced by RF fields IEC 61000-4-6 3V, 0.15MHz-80MHz 6V in ISM bands between 0.15MHz and 80MHz 80%AM at 1kHz   Cables information below is provided for EMC reference.
5. Regulatory Information    iRay Technology Co. Ltd.                                                                                                                                                                          47 Cable Recommended cable length Shielded or Unshielded Number Cable classification AC Power Cable 3m Unshielded 1 Set AC Power DC Power Cable 3.5m Unshielded 1 pcs Signal   Important information regarding Electro Magnetic Compatibility (EMC) Mars1717XU require special precautions regarding EMC and needs to be installed only by iRay or authorized personnel and put into service according to EMC information provided in the user manual. Mars1717XU in use may be susceptible to electromagnetic interference from portable and mobile RF communications such as mobile (cellular) telephones. Electromagnetic interference may result in incorrect operation of the system and create a potentially unsafe situation. Mars1717XU conforms to this EN60601-1-2:2015 standard for both immunity and emissions. Nevertheless, special precautions need to be observed: The use of accessories, transmitters and cables other than those specified by this User Manual, with the exception of accessories and cables sold by iRay of Mars1717XU as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of Mars1717XU.  5.6 Lithium-ion Capacitor Reference Standards Standards Description IEC 62813, Ed. 1 Lithium  ion  capacitors  for  use  in  electric  and  electronic equipment - Test methods for electrical characteristics QC-T 741-2014 Ultra-capacitor for Electric Vehicles UN38.3 United  Nations  Recommendations  on  the  Transport  of dangerous goods Manual of tests and Criteria ST/SG/AC.10/11/Rev.5/Amend.1&Amend.2 GB31241-2014 Lithium  ion  cells  and  batteries  used  in  portable  electronic equipments - Safety requirements
Mars1717XU Wireless Digital Flat Panel Detector User Manual 48                                                                                                                                                                                                  iRay Technology Co. Ltd. 5.7 Product Label 5.7.1 DETECTOR HOST LABEL      5.7.2 HOST HOUSING LABEL    5.8 Radio Frequency Compliance Information 5.8.1 SRRC COMPLIANCE   This equipment has been tested and found to comply with CMIIT ID:2017AP3876 of SRRC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.   Operation is subject to the following tow conditions.
5. Regulatory Information    iRay Technology Co. Ltd.                                                                                                                                                                          49 This device may not cause harmful interference. This device must accept any interference received, including interference that may cause undesired operation.   This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measure. Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the distributor or an experienced X-ray technician for help. 5.9 INFORMATION TO USER.   The users manual or instruction manual for an intentional or unintentional radiator shall caution the user that changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. In cases where the manual is provided only in a form other than paper, such as on a computer disk or over the Internet, the information required by this section may be included in the manual in that alternative form, provided the user can reasonably be expected to have the capability to access information in that form.    (A) ALL OTHER DEVICES SHALL BEAR THE FOLLOWING STATEMENT IN A CONSPICUOUS LOCATION ON THE DEVICE:   THIS DEVICE COMPLIES WITH PART 15 OF THE FCC RULES. OPERATION IS SUBJECT TO THE FOLLOWING TWO CONDITIONS: (1) THIS DEVICE MAY NOT CAUSE HARMFUL INTERFERENCE, AND (2) THIS DEVICE MUST ACCEPT ANY INTERFERENCE RECEIVED, INCLUDING INTERFERENCE THAT MAY CAUSE UNDESIRED OPERATION.
Mars1717XU Wireless Digital Flat Panel Detector User Manual 50                                                                                                                                                                                                  iRay Technology Co. Ltd. (B) FOR A CLASS B DIGITAL DEVICE OR PERIPHERAL, THE INSTRUCTIONS FURNISHED THE USER SHALL INCLUDE THE FOLLOWING OR SIMILAR STATEMENT, PLACED IN A PROMINENT LOCATION IN THE TEXT OF THE MANUAL: NOTE: THIS EQUIPMENT HAS BEEN TESTED AND FOUND TO COMPLY WITH THE LIMITS FOR A CLASS B DIGITAL DEVICE, PURSUANT TO PART 15 OF THE FCC RULES. THESE LIMITS ARE DESIGNED TO PROVIDE REASONABLE PROTECTION AGAINST HARMFUL INTERFERENCE IN A RESIDENTIAL INSTALLATION. THIS EQUIPMENT GENERATES, USES AND CAN RADIATE RADIO FREQUENCY ENERGY AND, IF NOT INSTALLED AND USED IN ACCORDANCE WITH THE INSTRUCTIONS, MAY CAUSE HARMFUL INTERFERENCE TO RADIO COMMUNICATIONS. HOWEVER, THERE IS NO GUARANTEE THAT INTERFERENCE WILL NOT OCCUR IN A PARTICULAR INSTALLATION. IF THIS EQUIPMENT DOES CAUSE HARMFUL INTERFERENCE TO RADIO OR TELEVISION RECEPTION, WHICH CAN BE DETERMINED BY TURNING THE EQUIPMENT OFF AND ON, THE USER IS ENCOURAGED TO TRY TO CORRECT THE INTERFERENCE BY ONE OR MORE OF THE FOLLOWING MEASURES:   REORIENT OR RELOCATE THE RECEIVING ANTENNA.   INCREASE THE SEPARATION BETWEEN THE EQUIPMENT AND RECEIVER.   CONNECT THE EQUIPMENT INTO AN OUTLET ON A CIRCUIT DIFFERENT FROM THAT TO WHICH THE RECEIVER IS CONNECTED.   CONSULT THE DEALER OR AN EXPERIENCED RADIO/TV TECHNICIAN FOR HELP.   This device has been tested and meets the FCC RF exposure guidelines. The maximum SAR value reported is 0.28 w/kg.
6. Software Opeartion Information  iRay Technology Co. Ltd.                                                                                                                                                                                                51          6. Software Operation Information 6.1  Default Settings ................................................................................ 52
Mars1717XU Wireless Digital Flat Panel Detector User Manual  52iRay Technology Co. Ltd.                                                                                                                                                                                    Please refer to Chapter 3 of this manual “IDetector use method” and description of iRayDR user manual for operation of the software.  6.1  Default Settings Default setting parameters of part of detector software are as following:  [Battery]  Battery_Level_1_H=100    Battery_Level_1_L=60        ; normal operation, allow exposure Battery_Level_2_H=59      Battery_Level_2_L=40        ; normal operation, low power warning, allow exposure Battery_Level_3_H=39 Battery_Level_3_L=30        ; normal operation, low power warning, forbid exposure Battery_Level_4_H=29       ; forbid use, the system needs shut down (<20% auto shut down) Battery_Level_4_L=0 [Wireless]  Signal_Split_1=70                  ; normal operation, allow exposure Signal_Split_2=40                  ; normal operation, forbid exposure [Temperature]  Temp_NormalRange_H=50    ; normal operation, allow exposure Temp_NormalRange_L=10    ; normal operation, allow exposure Temp_FatalLimit_H=60         ; not recommend use, forbid exposure Temp_FatalLimit_L=5            ; not recommend use, forbid exposure [SDK] Cfg_FpsCheck_Enable=0 Cfg_FpsCheck_Tolerance=0 Cfg_FWUpdTimeOut=1800 Cfg_OfflineInspectTimeout=10 Cfg_AllowReconnectByOnlineNotice=1 Cfg_ResetTimeout=120 Cfg_PreviewImage_Enable=1 Cfg_PushImageAtExpTimeout_Enable=1
6. Software Opeartion Information iRay Technology (Shanghai) Ltd.       Doc N0.: 058-201-02       Version: A4       Date:2017.08.23                                                                               53 Cfg_RetransferCount=3 Cfg_ClearAcqParam_DelayTime=600 Cfg_ConnRecoverTimeout=60 Cfg_TemperatureHighThreshold=40 Cfg_AllowMismatchSN=1 Cfg_ImagePacketGapTimeout=1000 Cfg_FwAllowedDefectPoints=8000 Cfg_PostOffsetStart_DelayTime=300 Cfg_TotalAcqTimeout=30000 Cfg_PreExpImageAcqTimeout=1000 Cfg_CleanupProcessTime=2000  [FactoryParameters] Product No=45 Product No=45 Sub Product No=SubProductNo_CsI550 Serial No=VT36xxxxxxxxx Main Version=1.6.x.x Read Version=0.0.0.0 Master Build Time=17-03-02:1 Slave Build Time=00-00-00:0 MCU Build Time=00-00-00:0 CBX Build Time=00-00-00:0 Src IP=192.168.100.8 Src MAC=000FEAEF6FBE Dest Port=28000 Dest IP=192.168.100.188 Dest MAC=5CF9DD5FE30E Sync-Box IP=192.168.100.96 Auto Sleep Idle Time (minute)=0 Trigger Mode=TriggerMode_Prep
Mars1717XU Wireless Digital Flat Panel Detector User Manual 54                                                                                iRay Technology (Shanghai) Ltd.       Doc N0.: 058-201-02       Version: A4       Date:2017.08.23 Prep CapMode=PrepCapMode_Acq2 Self CapEnable=On Integrate Time (ms)=776 Self Clear Span Time (us)=100 Sequence Interval Time (ms)=8000 Tube Ready Time=500 Set Delay Time (ms)=600 Exp Window Time (ms)=10000 Sync Exp Time (ms)=65535 VT (v)=0.87 Dyna Offset Gap Time (ms)=100 Image Channel Protocol=Ethernet_TCP
7. Trouble Shooting  iRay Technology Co. Ltd.                                                                                                                                                                                    55       7. Trouble Shooting This section is detailed in the Mars1717XU service manual.
Mars1717XU Wireless Digital Flat Panel Detector User Manual  56                                                                                                                                                                                       iRay Technology Co. Ltd.            8. Service Information 8.1  Product lifetime ................................................................................ 57 8.2  Regular inspection and Maintenance ............................................... 57 8.2.1 Daily inspection ...................................................................... 57 8.2.2 Monthly and annual inspection............................................... 57 8.3  Repair ............................................................................................... 58 8.4  Replacement parts support .............................................................. 58
8. Service Information iRay Technology Co. Ltd.                                                                                                                                                                                                  57 8.1  Product lifetime The estimated product lifetime is up to 7 years under appropriate regular inspection and maintenance. (The Lithium-ion capacitor could last for 7 years)  8.2  Regular inspection and Maintenance In order to ensure the safety of patients, operating person and third parties, and to maintain the performance and reliability of the equipment, be sure to perform regular inspection at least once a year. If necessary, clean up the equipment, make adjustments, or replace consumables such as fuses, detector cable, etc. There may be cases where overhaul is recommended depending on conditions. Contact iRay service office or local iRay dealer for regular inspection or maintenance. Cause the Lithium-ion capacitor products have its own power consumption, Please make sure that Mars1717XU is fully charged before it is stored and shipment, alse it needs take a full charge each two months during storage.  8.2.1 DAILY INSPECTION Before and after using this product, following inspections should be taken: Item Operation Detector Ensure the detector has no screw loosening or cracks. Ensure there is no dust or dirt attached on connection pins of Lithium-ion capacitor Ensure there is no crack or short circuit in the Lithium-ion capacitor connection pin. Cables Ensure the cables have no damage and no torn shell Ensure  connection  between  power  cable  and  AC  I/O socket reliable  8.2.2 MONTHLY AND ANNUAL INSPECTION Item Frequency Operation Resolution Every month/ Every year Use resolution image or body model to inspect  Linearity Every month/ Every year Evaluate through checking the grey value of images  Correction Every month/ Every year When  X-ray  generator,  bulb, collimator  or  exposure environment changes  If there is any repair, maintenance and inspection involved with opening the housing of the instrument, please make sure to contact after-sale service engineering with corresponding qualification, or please contact Shanghai iRay after-sales service department or authorized product distributors.
Mars1717XU Wireless Digital Flat Panel Detector User Manual 58                                                                                                                                                                                                   iRay Technology Co. Ltd.    There is a Lithium-ion capacitor in the FPD, its lifetime is 7 years, when arrived in the lifetime of the capacitor it needs to be placed. And the placement needs contact Shanghai iRay after-sales service departments or authorized product distributors.  8.3  Repair The product faults may be decreased to the minimum if the user follow this manual to do maintenance on time. However, if a problem cannot be solved even taking the measures indicated in troubleshooting, contact your sales representative or local iRay dealer for repairs. Please refer to the name label and provide the following information: Product Name:  Series Number: Description of Problem: as clearly as possible  8.4  Replacement parts support Performance parts (parts required to maintain the function of the product) of this product will be stocked for 8 years after discontinuance of production, to allow for repair.
Appendix  iRay Technology Co. Ltd.                                                                                                                                                                                                  59                                                                                                                                                                                                                                                                            Appendix Appendix A Information of Manufactures ................................................ 60
Mars1717XU Wireless Digital Flat Panel Detector User Manual  60                                                                                                                                                                                                   iRay Technology Co. Ltd.                                                                                                                                                                                                                                                        Appendix A Information of Manufactures               COMPANY: iRay Technology  Co.LTD. ADDRESS: Rm202, Building 7, No. 590, Ruiqing Rd, Zhangjiang East, Pudong New Area, Shanghai, China ZIP CODE: 201201 TELEPHONE: +86-21-50720560 FAX: +86-21-50720561 HOMEPAGE: WWW.IRAYCHINA.COM SERVICE: SERVICE DEPARTMENT OF IRAY SERVICE TEL +86-21-50720560 MARKET TEL +86-21-50720560

Navigation menu