ZOLL Medical Israel UCOR30 uCOR Monitoring System, uCor 3.0 User Manual EMC TEST REPORT

ZOLL Medical Israel Ltd. uCOR Monitoring System, uCor 3.0 EMC TEST REPORT

Manual

Download: ZOLL Medical Israel UCOR30 uCOR Monitoring System, uCor 3.0 User Manual EMC TEST REPORT
Mirror Download [FCC.gov]ZOLL Medical Israel UCOR30 uCOR Monitoring System, uCor 3.0 User Manual EMC TEST REPORT
Document ID3326679
Application IDBHCvz93GQ5mhgxj0FBpCRw==
Document DescriptionManual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize410.21kB (5127684 bits)
Date Submitted2017-03-22 00:00:00
Date Available2017-09-15 00:00:00
Creation Date2017-03-09 21:01:09
Producing SoftwareMicrosoft® Word 2010
Document Lastmod2017-03-09 21:01:09
Document TitleEMC TEST REPORT
Document CreatorMicrosoft® Word 2010
Document Author: EMC

EMC TEST REPORT
(FULL COMPLIANCE)
Report Number: 102716227BOX-002
Project Number: G102716227
Report Issue Date: 03/09/2017
Model(s) Tested:
Model(s) Partially Tested:
Model(s) Not Tested but declared equivalent by the
client:
Standards:
µCor 3.0
None
None
FCC Part 15 Subpart C: 01/2017
FCC Part 15 Subpart B: 01/2017
Tested by:
Intertek Testing Services NA, Inc.
70 Codman Hill Road
Boxborough, MA 01719
USA
Client:
Zoll Medical Israel Ltd.
14 Atir Yeda Street
Kfar-Saba 4464313
Israel
Client:
Zoll Medical Corporation
121 Gamma Drive Pittsburgh
PA 15238
USA
Report prepared by Naga Suryadevara
Report reviewed by Vathana Ven
Naga Suryadevara/EMC Engineer
Vathana Ven/Staff Engineer, EMC
This report is for the exclusive use of Intertek's Client and is provided pursuant to the agreement between
Intertek and its Client. Intertek's responsibility and liability are limited to the terms and conditions of the
agreement. Intertek assumes no liability to any party, other than to the Client in accordance with the agreement,
for any loss, expense or damage occasioned by the use of this report. Only the Client is authorized to permit
copying or distribution of this report and then only in its entirety. Any use of the Intertek name or one of its
marks for the sale or advertisement of the tested material, product or service must first be approved in writing by
Intertek. The observations and test results in this report are relevant only to the sample tested. This report by
itself does not imply that the material, product, or service is or has ever been under an Intertek certification
program.
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 1 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Table of Contents
Introduction and Conclusion .............................................................................................................. 3
Test Summary ...................................................................................................................................... 3
Client Information ................................................................................................................................ 4
Description of Equipment Under Test and Variant Models ............................................................. 4
System Setup and Method .................................................................................................................. 6
Output Power and Human RF Exposure ........................................................................................... 7
Occupied and 20dB Bandwidth ........................................................................................................ 15
Channel Separation ........................................................................................................................... 29
Number of Hopping Channels .......................................................................................................... 32
10
Average Channel Occupancy Time .............................................................................................. 34
11
Out of Band Conducted Emissions .............................................................................................. 38
12
Radiated Spurious Emissions....................................................................................................... 72
13
Radiated Emissions (Digital Device and Receiver) ..................................................................... 99
14
AC Mains Conducted Emissions ................................................................................................ 109
15
Revision History ........................................................................................................................... 117
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 2 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Introduction and Conclusion
The tests indicated in section 2.0 were performed on the product constructed as described in section 4.0.
The remaining test sections are the verbatim text from the actual data sheets used during the
investigation. These test sections include the test name, the specified test Method, a list of the actual Test
Equipment Used, documentation Photos, Results and raw Data. No additions, deviations, or exclusions
have been made from the standard(s) unless specifically noted.
Based on the results of our investigation, we have concluded the product tested complies with the
requirements of the standard(s) indicated. The results obtained in this test report pertain only to the
item(s) tested. Intertek does not make any claims of compliance for samples or variants which were not
tested.
Test Summary
Section
Test full name
Result
Client Information
Description of Equipment Under Test and Variant Models
System Setup and Method
Output Power and Human RF Exposure
(CFR47 FCC Part 15 Subpart C (15.247): 01/2017)
Pass
Occupied (99%) and 20 dB Bandwidth
(CFR47 FCC Part 15 Subpart C (15.247): 01/2017)
Pass
Channel Separation
(CFR47 FCC Part 15 Subpart C (15.247): 01/2017)
Pass
Number of Hopping Channels
(CFR47 FCC Part 15 Subpart C (15.247): 01/2017)
Pass
10
Average Channel Occupancy Time
(CFR47 FCC Part 15 Subpart C (15.247): 01/2017)
Pass
11
Out of Band Conducted Emissions
(CFR47 FCC Part 15 Subpart C (15.247): 01/2017)
Pass
12
Radiated Spurious Emissions
(CFR47 FCC Part 15 Subpart C (15.247): 01/2017)
Pass
13
Radiated Emissions (Digital parts and Receiver)
(FCC Part 15 Subpart B: 01/2017)
Pass
14
AC Mains Conducted Emissions
(FCC Part 15 Subpart B: 01/2017)
Pass
15
Revision History
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 3 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Client Information
This EUT was tested at the request of:
Client:
Zoll Medical Israel Ltd.
14 Atir Yeda Street
Kfar-Saba 4464313
Israel
Contact:
Telephone:
Fax:
Email:
Moshik Mosesko
+972 9 9603900
None
moshik@zoll.com
Description of Equipment Under Test and Variant Models
Manufacturer:
Zoll Medical Israel Ltd.
14 Atir Yeda Street
Kfar-Saba 4464313
Israel.
Description
µCor 3.0 sensor with
cable for Bluetooth
communication
µCor 3.0 open sensor
with cable for Bluetooth
communication and
antenna cable
µCor 3.0 sensor
µCor 3.0 charger (with
XT1527 (LTE) cellular
phone)
µCor 3.0 charger(with
XT1505 cellular phone)
AC/DC adapter
Receive Date:
Received Condition:
Type:
Equipment Under Test
Manufacturer
Model Number
Zoll
AS4200-01
Serial Number
D3-61653-0002
Zoll
AS4200-01
N/A
Zoll
Zoll
AS4200-01
CH0202-01
D3-61652-0095
C3-61520-0124
Zoll
CH0202-01
C3-61520-0127
Fuhua Electronic
UE24WCP1050300SPA
N/A
01/26/2017
Good
Production
Description of Equipment Under Test (provided by client)
The µCor 3.0 System is intended to record, store, and transmit the following physiological data to medical
professionals: i) Thoracic Impedance, ii) ECG; iii) Heart Rate; iv) Respiration Rate; v) Activity; and vi)
Posture. The µCor 3.0 System is indicated for patients who are 21 years of age or older: i) with fluidmanagement problems; ii) taking diuretic medication; iii) living with heart failure; iv) living with end-stage
renal disease; v) recovering from a coronary artery disease-related event; and/or vi) suffering from
recurrent dehydration.
Rated Voltage
100-240VAC (I/P to
Power Supply)
Internal Battery Li-Pol
Equipment Under Test Power Configuration
Rated Current
Rated Frequency
0.8 Amps
50/60 Hz
370 mA
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
N/A
Number of Phases
Single
N/A
Page 4 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
3.7V
Operating modes of the EUT:
No. Descriptions of EUT Exercising
Transmit mode – operating on internal battery
Receive mode – operating on internal battery
Transmit mode – Charging
Receive mode – Charging
Software used by the EUT:
Sensor:
No. Descriptions of EUT Exercising
Microcontroller version : 1.0.3
FPGA version: 3.0
BT version: 5.8
Fuel Gauge Version : 10082014
Gateway (cellular phone):
No. Descriptions of EUT Exercising
Access point: 1.3
Server:
No. Descriptions of EUT Exercising
ServerApp: 2.0.0.5
Radio/Receiver Characteristics
2412 – 2480 MHz
π/4 DQPSK(low baud rate) and 8DPSK(high baud rate)
Data Rate type(low baud rate) – Basic (DH3)
Data Rate type(high baud rate) – EDR (3-DH5)
3.863 mW
Maximum Output Power
Channel low - 2402 MHz
Test Channels
Channel middle - 2441 MHz
Channel high - 2480 MHz
1.2307 MHz (OBW)
Occupied Bandwidth
1.35 MHz (20dB)
Frequency Hopper: Number of Hopping
79
Channels
0.290 sec
Frequency Hopper: Channel Dwell Time
N/A
MIMO Information (# of Transmit and
Receive antenna ports)
Bluetooth single chip radio based on BlueCore 4.
Equipment Type
ETSI LBT/Adaptivity
N/A
ETSI Adaptivity Type
N/A
ETSI Temperature Category (I, II, III)
N/A
ETSI Receiver Category (1, 2, 3)
N/A
Integrated Antenna chip - 2450AT18B100E (Johanson)
Antenna Type and Gain
Peak Gain 0.5 dBi.
Frequency Band(s)
Modulation Type(s)
Data rates
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 5 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Variant Models:
The following variant models were not tested as part of this evaluation, but have been identified by the
manufacturer as being electrically identical models, depopulated models, or with reasonable similarity to
the model(s) tested. Intertek does not make any claims of compliance for samples or variants which were
not tested.
None
System Setup and Method
ID
Description
Power Cable
Description
Laptop
Cables
Length
(m)
Ferrites
Termination
None
None
AC Mains
Support Equipment
Manufacturer
Model Number
Dell
Latitude E7450
Phantom + Patch
Zoll
Gateway (Cellular
phone) into the charger
Motorola
5.1
Shielding
Patch PT0203-01
Phantom JIG0021
XT1505
XT1527
Serial Number
D48ZG72
N/A
GWID:1041
GWID:1047
Method:
Configuration as required by FCC Part 15 Subpart C: 01/2017, FCC Part 15 Subpart B: 01/2017, ANSI C
63.10: 2013 and ANSI C 63.4: 2014.
5.2
EUT Block Diagram:
EUT
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 6 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Output Power and Human RF Exposure
6.1
Method
Tests are performed in accordance with FCC Part 15 Subpart C (15.247).
TEST SITE: EMC Lab
The EMC Lab has one Semi-anechoic Chamber and one Shielded Chamber. AC Mains Power is available at 120,
230, and 277 Single Phase; 208, 400, and 480 3-Phase. Large reference ground-planes are installed in the general
lab area to facilitate EMC work not requiring a shielded environment.
6.2
Test Equipment Used:
Asset
Description
DAV004’
CBLHF2012
-2M-2’
ROS005’
Manufacturer
Weather Station
Davis Instruments
7400
Serial
PE80529A61
2m 9kHz-40GHz Coaxial Cable - SET2
ETSI Test System
Huber & Suhner
Rhode & Schwartz
SF102
TS8997
252675002
N/A
Software Utilized:
Name
Model
Manufacturer
Cal Date
Cal Due
05/02/2016
05/02/2017
02/09/2016
09/15/2016
02/09/2017
09/15/2017
Version
None
6.3
Results:
The sample tested was found to Comply. For systems operating in the 2400-2483.5 MHz band employing
at least 75 hopping channels, the maximum peak output power is 1 watt (30 dBm), for all other systems
0.125 W (21 dBm).
Device has 79 hopping channels
Frequency
2402
2402
2402
2402
2441
2441
2441
2441
2480
2480
2480
2480
Mode of
operation
Charging
Battery
Charging
Battery
Charging
Battery
Data Rate
Low
High
Low
High
Low
High
Low
High
Low
High
Low
High
Output Power
in dBm
5.87
3.83
5.78
3.86
5.38
3.29
5.34
3.32
4.26
2.64
4.71
2.64
Output Power in
mW
3.863
2.415
3.784
2.432
3.451
2.133
3.419
2.147
2.666
1.836
2.958
1.8365
Limit
1W
1W
1W
1W
1W
1W
1W
1W
1W
1W
1W
1W
Note: Devices antenna gain is less than 6 dBi.
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 7 of 117
Intertek
Report Number: 102716227BOX-002
6.4
Issued: 03/09/2017
Plots/Data:
Output Power @ 2402 MHz, battery mode and low data rate = 5.78 dBm
Output Power @ 2402 MHz, battery mode and high data rate = 3.86 dBm
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 8 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Output Power @ 2402 MHz, Charging mode and low data rate = 5.87 dBm
Output Power @ 2402 MHz, Charging mode and high data rate = 3.83 dBm
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 9 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Output Power @ 2441 MHz, battery mode and low data rate = 5.34 dBm
Output Power @ 2441 MHz, battery mode and high data rate = 3.32 dBm
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 10 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Output Power @ 2441 MHz, charging mode and low data rate = 5.38 dBm
Output Power @ 2441 MHz, charging mode and high data rate = 3.29 dBm
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 11 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Output Power @ 2480 MHz, battery mode and low data rate = 4.71 dBm
Output Power @ 2480 MHz, battery mode and high data rate = 2.64 dBm
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 12 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Output Power @ 2480 MHz, charging mode and low data rate = 4.26 dBm
Output Power @ 2480 MHz, charging mode and high data rate = 2.39 dBm
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 13 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Human RF Exposure/SAR Exemption
Maximum measured output power is 3.863 mW @ 2402 MHz
FCC SAR Exemption per KDB 447498
= (3.863/5)*(sqrt(2.402))
= 1.19 < 3.0 (below the limit, SAR Exempt per FCC)
Test Personnel:
Supervising/Reviewing
Engineer:
(Where Applicable)
Product Standard:
Input Voltage:
Pretest Verification w/
Ambient Signals or
BB Source:
Naga Suryadevara
Test Date:
01/26/2017
N/A
FCC Part 15 Subpart C
(15.247)
120VAC 60Hz
Internal Battery
Yes – Signal generator
Limit Applied:
See section 6.3
Ambient Temperature:
19 ºC
Relative Humidity:
30 %
Atmospheric Pressure:
981 mbars
Deviations, Additions, or Exclusions: None
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 14 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Occupied and 20dB Bandwidth
7.1
Method
Tests are performed in accordance with FCC Part 15 Subpart C (15.247).
TEST SITE: EMC Lab
The EMC Lab has one Semi-anechoic Chamber and one Shielded Chamber. AC Mains Power is available at 120,
230, and 277 Single Phase; 208, 400, and 480 3-Phase. Large reference ground-planes are installed in the general
lab area to facilitate EMC work not requiring a shielded environment.
7.2
Test Equipment Used:
Asset
Description
DAV004’
CBLHF2012
-2M-2’
ROS005’
Manufacturer
Weather Station
Davis Instruments
7400
Serial
PE80529A61
2m 9kHz-40GHz Coaxial Cable - SET2
ETSI Test System
Huber & Suhner
Rhode & Schwartz
SF102
TS8997
252675002
N/A
Software Utilized:
Name
Model
Manufacturer
Cal Date
Cal Due
05/02/2016
05/02/2017
02/09/2016
09/15/2016
02/09/2017
09/15/2017
Version
None
7.3
Results:
The sample tested was found to Comply.
Frequency
2402
2402
2402
2402
2441
2441
2441
2441
2480
2480
2480
2480
Mode of
operation
Charging
Battery
Charging
Battery
Charging
Battery
Data Rate
Low
High
Low
High
Low
High
Low
High
Low
High
Low
High
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Occupied
Bandwidth
919.08 kHz
1.2147 MHz
929.07 kHz
1.2147 MHz
919.08 kHz
1.214 MHz
919.08 kHz
1.218 MHz
927.07 kHz
1.2307 MHz
931.068 kHz
1.2307 MHz
20dB Bandwidth
987.00 kHz
1.34 MHz
989.00 kHz
1.34 MHz
983.00 kHz
1.34 MHz
991.00 kHz
1.33 MHz
983.00 kHz
1.33 MHz
999.0 kHz
1.35 MHz
Page 15 of 117
Intertek
Report Number: 102716227BOX-002
7.4
Issued: 03/09/2017
Plots/Data:
Occupied Bandwidth @ 2402 MHz, battery mode and low data rate = 929.07 kHz
20dB Bandwidth @ 2402 MHz, battery mode and low data rate = 989.00 kHz
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 16 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Occupied Bandwidth @ 2402 MHz, battery mode and high data rate = 1.2147 MHz
20dB Bandwidth @ 2402 MHz, battery mode and high data rate = 1.34 MHz
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 17 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Occupied Bandwidth @ 2402 MHz, charging mode and low data rate = 919.08 kHz
20dB Bandwidth @ 2402 MHz, charging mode and low data rate = 987.00 kHz
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 18 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Occupied Bandwidth @ 2402 MHz, charging mode and high data rate = 1.2147 MHz
20dB Bandwidth @ 2402 MHz, charging mode and high data rate = 1.34 MHz
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 19 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Occupied Bandwidth @ 2441 MHz, battery mode and low data rate = 919.08 KHz
20dB @ 2441 MHz, battery mode and low data rate = 991.00 KHz
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 20 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Occupied Bandwidth @ 2441 MHz, battery mode and high data rate = 1.218 MHz
20dB Bandwidth @ 2441 MHz, battery mode and high data rate = 1.33 MHz
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 21 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Occupied Bandwidth @ 2441 MHz, charging mode and low data rate = 919.08 KHz
20dB Bandwidth @ 2441 MHz, charging mode and low data rate = 983.00 KHz
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 22 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Occupied Bandwidth @ 2441 MHz, charging mode and high data rate = 1.214 MHz
20dB Bandwidth @ 2441 MHz, charging mode and high data rate = 1.34 MHz
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 23 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Occupied Bandwidth @ 2480 MHz, battery mode and low data rate = 931.068 kHz
20dB Bandwidth @ 2480 MHz, battery mode and low data rate = 999.000 kHz
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 24 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Occupied Bandwidth @ 2480 MHz, battery mode and high data rate = 1.23076 MHz
20dB Bandwidth @ 2480 MHz, battery mode and high data rate = 1.35 MHz
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 25 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Occupied Bandwidth @ 2480 MHz, charging mode and low data rate = 927.07 kHz
20dB Bandwidth @ 2480 MHz, charging mode and low data rate = 983.00 kHz
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 26 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Occupied Bandwidth @ 2480 MHz, charging mode and high data rate = 1.2307 MHz
20dB Bandwidth @ 2480 MHz, charging mode and high data rate = 1.33 MHz
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 27 of 117
Intertek
Report Number: 102716227BOX-002
Test Personnel:
Supervising/Reviewing
Engineer:
(Where Applicable)
Product Standard:
Input Voltage:
Pretest Verification w/
Ambient Signals or
BB Source:
Naga Suryadevara
Issued: 03/09/2017
Test Date:
01/26/2017
N/A
FCC Part 15 Subpart C
(15.247)
120VAC 60Hz
Internal Battery
Yes – Signal generator
Limit Applied:
Ambient Temperature:
19 ºC
Relative Humidity:
30 %
Atmospheric Pressure:
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
See section 7.3
981 mbars
Page 28 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Channel Separation
8.1
Method
Tests are performed in accordance with FCC Part 15 Subpart C (15.247).
TEST SITE: EMC Lab
The EMC Lab has one Semi-anechoic Chamber and one Shielded Chamber. AC Mains Power is available at 120,
230, and 277 Single Phase; 208, 400, and 480 3-Phase. Large reference ground-planes are installed in the general
lab area to facilitate EMC work not requiring a shielded environment.
8.2
Test Equipment Used:
Asset
Description
DAV004’
CBLHF2012
-2M-2’
ROS005’
Weather Station
Davis Instruments
7400
Serial
PE80529A61
2m 9kHz-40GHz Coaxial Cable - SET2
ETSI Test System
Huber & Suhner
Rhode & Schwartz
SF102
TS8997
252675002
N/A
Software Utilized:
Name
Manufacturer
Manufacturer
Model
Cal Date
Cal Due
05/02/2016
05/02/2017
02/09/2016
09/15/2016
02/09/2017
09/15/2017
Version
None
8.3
Results:
The sample tested was found to Comply. Systems shall have hopping channel carrier frequencies that are
separated by 25 kHz or two-thirds of the 20-dB bandwidth of the hopping channel, whichever is greater.
The measured channel frequency separation is 931 kHz @ low data rate and 979 kHz @ high data rate
which is greater than 900 kHz (2/3rd of 20dB BW(maximum))
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 29 of 117
Intertek
Report Number: 102716227BOX-002
8.4
Issued: 03/09/2017
Plots/Data:
Channel Separation, Low data rate = 931.10 kHz
Channel Separation, High data rate = 979.00 kHz
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 30 of 117
Intertek
Report Number: 102716227BOX-002
Test Personnel:
Supervising/Reviewing
Engineer:
(Where Applicable)
Product Standard:
Input Voltage:
Pretest Verification w/
Ambient Signals or
BB Source:
Naga Suryadevara
Issued: 03/09/2017
Test Date:
01/26/2017
N/A
FCC Part 15 Subpart C
(15.247)
120VAC 60Hz
Internal Battery
Yes – Signal generator
Limit Applied:
Ambient Temperature:
19 ºC
Relative Humidity:
30 %
Atmospheric Pressure:
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
See section 8.3
981 mbars
Page 31 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Number of Hopping Channels
9.1
Method
Tests are performed in accordance with FCC Part 15 Subpart C (15.247).
TEST SITE: EMC Lab
The EMC Lab has one Semi-anechoic Chamber and one Shielded Chamber. AC Mains Power is available at 120,
230, and 277 Single Phase; 208, 400, and 480 3-Phase. Large reference ground-planes are installed in the general
lab area to facilitate EMC work not requiring a shielded environment.
9.2
Test Equipment Used:
Asset
Description
DAV004’
CBLHF2012
-2M-2’
ROS005’
Weather Station
Davis Instruments
7400
Serial
PE80529A61
2m 9kHz-40GHz Coaxial Cable - SET2
ETSI Test System
Huber & Suhner
Rhode & Schwartz
SF102
TS8997
252675002
N/A
Software Utilized:
Name
Manufacturer
Manufacturer
Model
Cal Date
Cal Due
05/02/2016
05/02/2017
02/09/2016
09/15/2016
02/09/2017
09/15/2017
Version
None
9.3
Results:
The sample tested was found to Comply. Systems operating in the 2400-2483.5 MHz band shall use at
least 15 hopping channels.
The System has 79 hopping channels.
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 32 of 117
Intertek
Report Number: 102716227BOX-002
9.4
Issued: 03/09/2017
Plots/Data:
Number of hopping channels = 79
Test Personnel:
Supervising/Reviewing
Engineer:
(Where Applicable)
Product Standard:
Input Voltage:
Pretest Verification w/
Ambient Signals or
BB Source:
Naga Suryadevara
Test Date:
01/26/2017
N/A
FCC Part 15 Subpart C
(15.247)
120VAC 60Hz
Internal Battery
Yes – Signal generator
Limit Applied:
Ambient Temperature:
19 ºC
Relative Humidity:
30 %
Atmospheric Pressure:
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
See section 9.3
981 mbars
Page 33 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
10 Average Channel Occupancy Time
10.1 Method
Tests are performed in accordance with FCC Part 15 Subpart C (15.247).
TEST SITE: EMC Lab
The EMC Lab has one Semi-anechoic Chamber and one Shielded Chamber. AC Mains Power is available at 120,
230, and 277 Single Phase; 208, 400, and 480 3-Phase. Large reference ground-planes are installed in the general
lab area to facilitate EMC work not requiring a shielded environment.
10.2 Test Equipment Used:
Asset
DAV004’
CBLHF2012
-2M-2’
ROS005’
Description
Weather Station
Davis Instruments
7400
Serial
PE80529A61
2m 9kHz-40GHz Coaxial Cable - SET2
ETSI Test System
Huber & Suhner
Rhode & Schwartz
SF102
TS8997
252675002
N/A
Software Utilized:
Name
Manufacturer
Model
Manufacturer
Cal Date
Cal Due
05/02/2016
05/02/2017
02/09/2016
09/15/2016
02/09/2017
09/15/2017
Version
None
10.3 Results:
The sample tested was found to Comply. For systems operating in the 2400-2483.5 MHz band, the
average time of occupancy on any channel shall not be greater than 0.4 second within a period of 0.4
second multiplied by the number of hopping channels employed
Since the radio employs 79 channels, Occupancy time was calculated during the period of 0.4 * 79 = 31.6
sec.
Occupancy Time @ low data rate = 0.000510 * 31 *10 = 0.1581 sec
Occupancy Time @ high data rate = 0.00290 * 10 *10 = 0.290 sec
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 34 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
10.4 Plots/Data:
Dwell time, Low data rate = 510.00 us
Number of bursts in 3.16 second sweep, Low data rate = 31
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 35 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Dwell time, High data rate = 2.90 ms
Number of bursts in a 3.16 s sweep, High data rate = 10
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 36 of 117
Intertek
Report Number: 102716227BOX-002
Test Personnel:
Supervising/Reviewing
Engineer:
(Where Applicable)
Product Standard:
Input Voltage:
Pretest Verification w/
Ambient Signals or
BB Source:
Naga Suryadevara
Issued: 03/09/2017
Test Date:
01/26/2017
N/A
FCC Part 15 Subpart C
(15.247)
120VAC 60Hz
Internal Battery
Yes – Signal generator
Limit Applied:
Ambient Temperature:
19 ºC
Relative Humidity:
30 %
Atmospheric Pressure:
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
See section 10.3
981 mbars
Page 37 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
11 Out of Band Conducted Emissions
11.1 Method
Tests are performed in accordance with FCC Part 15 Subpart C (15.247) and RSS 247.
TEST SITE: EMC Lab
The EMC Lab has one Semi-anechoic Chamber and one Shielded Chamber. AC Mains Power is available at 120,
230, and 277 Single Phase; 208, 400, and 480 3-Phase. Large reference ground-planes are installed in the general
lab area to facilitate EMC work not requiring a shielded environment.
11.2 Test Equipment Used:
Asset
DAV004’
CBLHF2012
-2M-2’
ROS005’
Description
Weather Station
Davis Instruments
7400
Serial
PE80529A61
2m 9kHz-40GHz Coaxial Cable - SET2
ETSI Test System
Huber & Suhner
Rhode & Schwartz
SF102
TS8997
252675002
N/A
Software Utilized:
Name
Manufacturer
Manufacturer
Model
Cal Date
Cal Due
05/02/2016
05/02/2017
02/09/2016
09/15/2016
02/09/2017
09/15/2017
Version
None
11.3 Results:
The sample tested was found to Comply. In any 100 kHz bandwidth outside the EUT pass-band, the RF
power shall be at least 20 dB below that of the maximum in-band 100 kHz emission.
All emissions measured were 20 dB below fundamental as indicated in the plots in sections 11.5.
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 38 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
11.4 Plots/Data:
Low Channel (2402 MHz) – Battery mode, In band emissions, Low data rate
Low Channel (2402 MHz) – Battery mode, Out of band emissions, Low data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 39 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel (2402 MHz) – Battery mode, Out of band emissions, Low data rate
Low Channel (2402 MHz) – Battery mode, Out of band emissions, Low data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 40 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel (2402 MHz) – Battery mode, Band edge emissions, Low data rate
Low Channel (2402 MHz) – Battery mode, In band emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 41 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel (2402 MHz) – Battery mode, Out of band emissions, High data rate
Low Channel (2402 MHz) – Battery mode, Out of band emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 42 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel (2402 MHz) – Battery mode, Out of band emissions, High data rate
Low Channel (2402 MHz) – Battery mode, Band edge emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 43 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel (2402 MHz) – Charging mode, In band emissions, Low data rate
Low Channel (2402 MHz) – Charging mode, Out of band emissions, Low data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 44 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel (2402 MHz) – Charging mode, Out of band emissions, Low data rate
Low Channel (2402 MHz) – Charging mode, Out of band emissions, Low data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 45 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel (2402 MHz) – Charging mode, Band edge emissions, Low data rate
Low Channel (2402 MHz) – Charging mode, In Band emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 46 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel (2402 MHz) – Charging mode, Out of Band emissions, High data rate
Low Channel (2402 MHz) – Charging mode, Out of Band emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 47 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel (2402 MHz) – Charging mode, Out of Band emissions, High data rate
Low Channel (2402 MHz) – Charging mode, Band edge emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 48 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Mid Channel (2441 MHz) – Battery mode, In Band emissions, Low data rate
Mid Channel (2441 MHz) – Battery mode, Out of Band emissions, Low data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 49 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Mid Channel (2441 MHz) – Battery mode, Out of Band emissions, Low data rate
Mid Channel (2441 MHz) – Battery mode, Out of Band emissions, Low data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 50 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Mid Channel (2441 MHz) – Battery mode, In Band emissions, High data rate
Mid Channel (2441 MHz) – Battery mode, Out of Band emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 51 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Mid Channel (2441 MHz) – Battery mode, Out of Band emissions, High data rate
Mid Channel (2441 MHz) – Battery mode, Out of Band emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 52 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Mid Channel (2441 MHz) – Charging mode, In Band emissions, Low data rate
Mid Channel (2441 MHz) – Charging mode, Out of Band emissions, Low data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 53 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Mid Channel (2441 MHz) – Charging mode, Out of Band emissions, Low data rate
Mid Channel (2441 MHz) – Charging mode, Out of Band emissions, Low data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 54 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Mid Channel (2441 MHz) – Charging mode, In Band emissions, High data rate
Mid Channel (2441 MHz) – Charging mode, Out of Band emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 55 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Mid Channel (2441 MHz) – Charging mode, Out of Band emissions, High data rate
Mid Channel (2441 MHz) – Charging mode, Out of Band emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 56 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel (2480 MHz) – Battery mode, In Band emissions, Low data rate
High Channel (2480 MHz) – Battery mode, Out of Band emissions, Low data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 57 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel (2480 MHz) – Battery mode, Out of Band emissions, Low data rate
High Channel (2480 MHz) – Battery mode, Out of Band emissions, Low data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 58 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel (2480 MHz) – Battery mode, Band edge emissions, Low data rate
High Channel (2480 MHz) – Battery mode, In Band emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 59 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel (2480 MHz) – Battery mode, Out of emissions, High data rate
High Channel (2480 MHz) – Battery mode, Out of emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 60 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel (2480 MHz) – Battery mode, Out of emissions, High data rate
High Channel (2480 MHz) – Battery mode, Band edge emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 61 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel (2480 MHz) – Charging mode, In Band emissions, Low data rate
High Channel (2480 MHz) – Charging mode, Out of Band emissions, Low data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 62 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel (2480 MHz) – Charging mode, Out of Band emissions, Low data rate
High Channel (2480 MHz) – Charging mode, Out of Band emissions, Low data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 63 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel (2480 MHz) – Charging mode, Band edge emissions, Low data rate
High Channel (2480 MHz) – Charging mode, In Band emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 64 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel (2480 MHz) – Charging mode, Out of Band emissions, High data rate
High Channel (2480 MHz) – Charging mode, Out of Band emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 65 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel (2480 MHz) – Charging mode, Out of Band emissions, High data rate
High Channel (2480 MHz) – Charging mode, Band edge emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 66 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Frequency hopping – Battery mode, Lower Band edge emissions, Low data rate
Frequency hopping – Battery mode, Lower Band edge emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 67 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Frequency hopping – Charging mode, Lower Band edge emissions, Low data rate
Frequency hopping – Charging mode, Lower Band edge emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 68 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Frequency hopping – Battery mode, Higher Band edge emissions, Low data rate
Frequency hopping – Battery mode, Higher Band edge emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 69 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Frequency hopping – Charging mode, Higher Band edge emissions, Low data rate
Frequency hopping – Charging mode, Higher Band edge emissions, High data rate
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 70 of 117
Intertek
Report Number: 102716227BOX-002
Test Personnel:
Supervising/Reviewing
Engineer:
(Where Applicable)
Product Standard:
Input Voltage:
Pretest Verification w/
Ambient Signals or
BB Source:
Naga Suryadevara
Issued: 03/09/2017
Test Date:
01/26/2017
N/A
FCC Part 15 Subpart C
(15.247)
120VAC 60Hz
Internal Battery
Yes – Signal generator
Limit Applied:
Ambient Temperature:
19 ºC
Relative Humidity:
30 %
Atmospheric Pressure:
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
See section 11.3
981 mbars
Page 71 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
12 Radiated Spurious Emissions
12.1 Method
Tests are performed in accordance with FCC Part 15 Subpart C (15.247).
TEST SITE: 10m ALSE
The 10m ALSE is 13m (Length) x 21m (Depth) x 10m (Height) with the effective size in terms of space from the
tips of the absorber is 12m (Length) x 20m (Depth) x 8.5m (Height). This chamber achieves broadband
performance using a unique arrangement of hybrid and ferrite tile absorber. This chamber has a built in 3m
diameter turntable (Embedded type). The metal structure of the table makes electrical connection around the
entire circumference of the turntable to the ground plane with a metal brush type connection. The turntable is
located on one end of the chamber and the antennas are mounted 3 and 10 meters away at the other end of
the chamber on the adjustable an Antenna Mast. The antenna mast is a non-conductive bore sighted type with
remote control of antenna height and polarization. The Antenna Mast and the turntable can be remotely
controlled through the controller located in the adjacent Control room. A Styrofoam table 80 cm high is used for
table-top equipment.
Measurement Uncertainty
Measurement
Radiated Emissions, 10m
Radiated Emissions, 3m
Radiated Emissions, 3m
Radiated Emissions, 3m
Radiated Emissions, 3m
Radiated Emissions, 3m
Frequency
Range
30-1000 MHz
30-1000 MHz
1-6 GHz
6-15 GHz
15-18 GHz
18-40 GHz
As shown in the table above our radiated emissions
Expanded
Uncertainty
(k=2)
4.6 dB
5.3 dB
4.5 dB
5.2 dB
5.0 dB
5.0 dB
Ucispr
6.3 dB
6.3 dB
5.2 dB
5.5 dB
5.5 dB
5.5 dB
U lab is less than the corresponding U CISPR reference
value in CISPR 16-4-2 Table 1, hence the compliance of the product is only based on the measured
value, and no measurement uncertainty correction is required, based on CISPR 22 and CISPR 11 (for
2006 and later revisions) Clause 11.
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 72 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Sample Calculation
The field strength is calculated by adding the Antenna Factor and Cable Factor, and subtracting the
Amplifier Gain (if any) from the measured reading. The basic equation with a sample calculation is as
follows:
FS = RA + AF + CF - AG
Where
FS = Field Strength in dBV/m
RA = Receiver Amplitude (including preamplifier) in dBV
CF = Cable Attenuation Factor in dB
AF = Antenna Factor in dB
AG = Amplifier Gain in dB
In the following table(s), the reading shown on the data table reflects the preamplifier gain. An example
for the calculations in the following table is as follows.
Assume a receiver reading of 52.0 dBV is obtained. The antenna factor of 7.4 dB and cable factor of
1.6 dB is added. The amplifier gain of 29 dB is subtracted, giving a field strength of 32 dBV/m. This
value in dBV/m was converted to its corresponding level in V/m.
RA = 52.0 dBV
AF = 7.4 dB/m
CF = 1.6 dB
AG = 29.0 dB
FS = 32 dBV/m
To convert from dBV to V or mV the following was used:
where UF = Net Reading in V
NF = Net Reading in dBV
(NF / 20)
UF = 10
Example:
FS = RA + AF + CF – AG = 52.0 + 7.4 + 1.6 – 29.0 = 32.0
(32 dBV / 20)
UF = 10
= 39.8 V/m
Alternately, when C5 Software is used, the “Level” includes all losses and gains and is compared directly
in the “Margin” column to the “Limit”. “AF” is the Antenna Factor; “PA+CL” are Preamp and Cable Loss.
These are already accounted for in the “Level” column.
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 73 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
12.2 Test Equipment Used:
Asset
Description
Manufacturer
Model
Cal Date
Cal Due
05/02/2016
03/09/2016
05/02/2017
03/09/2017
multiple
PRE10
00143259
07/30/2016
12/16/2016
02/10/2016
07/30/2017
12/16/2017
02/10/2017
multiple
3008A00232
2090
07/30/2016
05/27/2016
09/14/2016
07/30/2017
05/27/2017
09/14/2017
06-1
PRE10
01/25/2016
12/16/2016
01/25/2017
12/16/2017
DAV004’
145145’
Weather Station
Broadband Hybrid Antenna 30 MHz - 3 GHz
Davis Instruments
Sunol Sciences Corp.
145-410’
PRE10’
ETS001’
Cables 145-420 145-421 145-422 145-406
30-1000MHz pre-amp
1-18GHz DRG Horn Antenna
Huber + Suhner
ITS
ETS-Lindgren
145-416’
145014’
EMC04’
Cables 145-420 145-423 145-424 145-408
Preamplifier (1 GHz to 26.5 GHz)
ANTENNA, RIDGED GUIDE, 18-40 GHZ
Huber + Suhner
Hewlett Packard
EMCO
3GHz High Pass Filter
30-1000MHz pre-amp
Reactel, Inc
ITS
2m 9kHz-40GHz Coaxial Cable - SET2
Huber & Suhner
SF102
252675002
02/09/2016
02/09/2017
5m 9kHz-40GHz Coaxial Cable - SET2
Huber & Suhner
SF102
252676002
02/19/2016
02/19/2017
REA004’’
PRE10’
CBLHF2012
-2M-2’
CBLHF2012
-5M-2’
Software Utilized:
Name
Compliance5
7400
JB3
10m Track A
Cables
PRE10
3117
3m Track B
cables
8449B
3116
7HSX3G/18G-S11
PRE10
Serial
PE80529A61
A122313
Manufacturer
Version
Teseq
5.26.46.46
12.3 Results:
The sample tested was found to Comply.
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 74 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
12.4 Plots/Data:
Low Channel, Charging mode, Tx (Low data rate, 30 MHz – 1 GHz)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 10m
Zoll medical_01/26/2017
Tx Low channel, charging mode
19 C
30% 981 mbars
Naga Suryadevara
26 Jan 2017 13 : 01
Additional Information
Prescan Emission Graph
100
26 Jan 2017 13 : 01 Test: Radiated - FCC15 Class B @ 10m Project: Zoll medical_01/26/2017 Tested by: Naga Suryadevara Test Notes: Tx Low channel (HDR), charging mode
90
80
70
60
50
40
Level (dBuV/m)
30 FCC Part 15 Class B QP Radiated Emissions @10m
20
10
Trace3: Measured Average
Trace2: Measured Quasi Peak
0 Trace1: Measured Peak
30M
Frequency (Hz)
Measured
Measured
Measured
Maximum
100M
1G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace2: Measured Quasi Peak
Frequency(Hz)
319.978357156 M
463.994589287 M
335.972946186 M
303.958917427 M
431.946092589 M
815.994588912 M
Level(dBuV/m)
16.70
24.22
24.35
29.04
30.31
31.49
AF
20.100
23.660
20.219
19.758
22.739
28.000
PA+CL
-38.178
-37.590
-38.136
-38.210
-37.777
-35.661
Limit(dBuV/m)
36.020
36.020
36.020
36.020
36.020
36.020
Margin(dBuV/m)
-19.32
-11.80
-11.67
-6.98
-5.71
-4.53
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Hor ( -- ), Ver ( | )
-------
Azimuth (deg)(Deg)
310
96
97
96
94
286
Mast Height(m)
1.15
2.37
1.24
2.36
2.39
1.15
RBW(Hz)
120 k
120 k
120 k
120 k
120 k
120 k
Page 75 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel, Battery mode, Tx (Low data rate, 30 MHz – 1 GHz, X-axis)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 10m
Zoll medical 01/29/2017
Tx mode X axis Battery
19 C
19% 999 mbars
Naga Suryadevara
29 Jan 2017 19 : 11
Additional Information
Prescan Emission Graph
100
29 Jan 2017 19 : 11 Test: Radiated - FCC15 Class B @ 10m Project: Zoll medical 01/29/2017 Tested by: Naga Suryadevara Test Notes: Tx mode X axis Battery
90
80
70
60
50
40
Level (dBuV/m)
30 FCC Part 15 Class B QP Radiated Emissions @10m
20
10
Trace3: Measured Average
Trace2: Measured Quasi Peak
0 Trace1: Measured Peak
30M
Frequency (Hz)
Measured
Measured
Measured
Maximum
100M
1G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace2: Measured Quasi Peak
Frequency(Hz)
335.870340976 M
638.554108525 M
431.934870144 M
815.867134002 M
30.0 M
848.001403242 M
Level(dBuV/m)
10.71
17.08
17.32
19.63
17.24
26.34
AF
20.217
26.100
22.739
28.000
27.500
28.000
PA+CL
-38.137
-36.623
-37.778
-35.662
-40.080
-35.554
Limit(dBuV/m)
36.020
36.020
36.020
36.020
30.000
36.020
Margin(dBuV/m)
-25.31
-18.94
-18.70
-16.39
-12.76
-9.68
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Hor ( -- ), Ver ( | )
-|
Azimuth (deg)(Deg)
185
94
265
73
342
62
Mast Height(m)
1.08
1.27
1.05
1.50
2.40
1.77
RBW(Hz)
120 k
120 k
120 k
120 k
120 k
120 k
Page 76 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel, Battery mode, Tx (Low data rate, 30 MHz – 1 GHz, Y-axis)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 10m
Zoll medical 01/29/2017
Tx mode Y axis Battery
19 C
19% 999 mbars
Naga Suryadevara
29 Jan 2017 18 : 16
Additional Information
Prescan Emission Graph
100
29 Jan 2017 18 : 16 Test: Radiated - FCC15 Class B @ 10m Project: Zoll medical 01/29/2017 Tested by: Naga Suryadevara Test Notes: Tx mode Y axis Battery
90
80
70
60
50
40
Level (dBuV/m)
30 FCC Part 15 Class B QP Radiated Emissions @10m
20
10
Trace3: Measured Average
Trace2: Measured Quasi Peak
0 Trace1: Measured Peak
30M
Frequency (Hz)
Measured
Measured
Measured
Maximum
100M
1G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace2: Measured Quasi Peak
Frequency(Hz)
463.968136381 M
527.954309076 M
559.953907978 M
847.950100637 M
495.96452887 M
815.984969673 M
Level(dBuV/m)
21.33
23.63
24.17
26.42
27.15
27.33
AF
23.659
24.218
25.097
28.000
23.919
28.000
PA+CL
-37.591
-37.241
-37.096
-35.554
-37.396
-35.661
Limit(dBuV/m)
36.020
36.020
36.020
36.020
36.020
36.020
Margin(dBuV/m)
-14.69
-12.39
-11.85
-9.60
-8.87
-8.69
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Hor ( -- ), Ver ( | )
---|
-|
Azimuth (deg)(Deg)
40
63
53
75
63
72
Mast Height(m)
1.57
1.32
1.35
1.65
1.41
3.51
RBW(Hz)
120 k
120 k
120 k
120 k
120 k
120 k
Page 77 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel, Battery mode, Tx (Low data rate, 30 MHz – 1 GHz, Z-axis)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 10m
Zoll medical_01/27/2017
Z-axis Bluetooth
19C
27% 987mbars
Naga Suryadevara
27 Jan 2017 15 : 24
Additional Information
Prescan Emission Graph
100
27 Jan 2017 15 : 24 Test: Radiated - FCC15 Class B @ 10m Project: Zoll medical_01/27/2017 Tested by: Naga Suryadevara Test Notes: Y-axis Bluetooth
90
80
70
60
50
40
Level (dBuV/m)
30 FCC Part 15 Class B QP Radiated Emissions @10m
20
10
Trace3: Measured Average
Trace2: Measured Quasi Peak
0 Trace1: Measured Peak
30M
Frequency (Hz)
100M
1G
Measured Peak Value
Measured Quasi Peak Value
Measured Average Value
Maximum Value of Mast and Turntable
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace2: Measured Quasi Peak
Frequency(Hz)
703.911422617 M
463.940080269 M
624.030059689 M
559.974749661 M
495.980560934 M
713.079358611 M
Level(dBuV/m)
17.85
18.83
19.65
22.89
23.13
29.88
AF
26.700
23.658
25.542
25.098
23.920
26.823
PA+CL
-36.148
-37.591
-36.729
-37.096
-37.396
-36.101
Limit(dBuV/m)
36.020
36.020
36.020
36.020
36.020
36.020
Margin(dBuV/m)
-18.17
-17.19
-16.37
-13.13
-12.89
-6.14
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Hor ( -- ), Ver ( | )
----|
Azimuth (deg)(Deg)
40
75
62
73
73
Mast Height(m)
3.97
2.39
1.97
1.24
1.77
1.16
RBW(Hz)
120 k
120 k
120 k
120 k
120 k
120 k
Page 78 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel, Charging mode, Tx (Low data rate, above 1 GHz)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
Low Channel Charger
20C
21% 989 mbars
Naga Suryadevara
28 Jan 2017 17 : 17
Additional Information
Prescan Emission Graph
100
28 Jan 2017 17 : 17 Test: Radiated - FCC15 Class B @ 3m Project: Zoll medical_01/28/2017 Tested by: Naga Suryadevara Test Notes: Low Channel Charger
90
80
70
60
50
40
Level (dBuV/m)
30
20
10
Trace3: Measured Average
0 Trace1: Measured Peak
3G
Frequency (Hz)
Measured
Measured
Measured
Maximum
4G
5G
6G
7G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
8G
9G
10G
18G
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
7.203687375 G
4.803961255 G
9.599385438 G
Level(dBuV/m)
54.48
55.85
56.67
AF
35.644
34.192
36.681
PA+CL
-23.895
-25.631
-22.087
Limit(dBuV/m)
74.000
74.000
74.000
Margin(dBuV/m)
-19.52
-18.15
-17.33
Hor ( -- ), Ver ( | )
-|
Azimuth (deg)(Deg)
310
81
188
Mast Height(m)
4.00
3.00
1.23
RBW(Hz)
1M
1M
1M
AF
34.192
35.644
36.681
PA+CL
-25.631
-23.895
-22.087
Limit(dBuV/m)
54.000
54.000
54.000
Margin(dBuV/m)
-16.36
-13.33
-10.82
Hor ( -- ), Ver ( | )
-|
Azimuth (deg)(Deg)
81
310
188
Mast Height(m)
3.00
4.00
1.23
RBW(Hz)
1M
1M
1M
Trace3: Measured Average
Frequency(Hz)
4.803961255 G
7.203687375 G
9.599385438 G
Level(dBuV/m)
37.64
40.67
43.18
Note: Manual Scans were performed from 1-3 GHz and 18 – 40 GHz, no emissions were detected above the noise floor
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 79 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Mid Channel, Charging mode, Tx (Low data rate, above 1 GHz)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
Mid Channel Charger
20C
21% 989 mbars
Naga Suryadevara
28 Jan 2017 17 : 40
Additional Information
Prescan Emission Graph
100
28 Jan 2017 17 : 40 Test: Radiated - FCC15 Class B @ 3m Project: Zoll medical_01/28/2017 Tested by: Naga Suryadevara Test Notes: Mid Channel Charger
90
80
70
60
50
40
Level (dBuV/m)
30
20
10
Trace3: Measured Average
0 Trace1: Measured Peak
3G
Frequency (Hz)
Measured
Measured
Measured
Maximum
4G
5G
6G
7G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
8G
9G
10G
18G
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
7.323473614 G
4.843987975 G
9.759338677 G
Level(dBuV/m)
54.96
55.34
57.44
AF
35.622
34.192
36.829
PA+CL
-23.659
-25.630
-21.849
Limit(dBuV/m)
74.000
74.000
74.000
Margin(dBuV/m)
-19.04
-18.66
-16.56
Hor ( -- ), Ver ( | )
--|
Azimuth (deg)(Deg)
331
107
48
Mast Height(m)
3.87
3.88
2.96
RBW(Hz)
1M
1M
1M
AF
34.192
35.622
36.829
PA+CL
-25.630
-23.659
-21.849
Limit(dBuV/m)
54.000
54.000
54.000
Margin(dBuV/m)
-16.68
-12.95
-10.96
Hor ( -- ), Ver ( | )
--|
Azimuth (deg)(Deg)
107
331
48
Mast Height(m)
3.88
3.87
2.96
RBW(Hz)
1M
1M
1M
Trace3: Measured Average
Frequency(Hz)
4.843987975 G
7.323473614 G
9.759338677 G
Level(dBuV/m)
37.32
41.05
43.04
Note: Manual Scans were performed from 1-3 GHz and 18 – 40 GHz, no emissions were detected above the noise floor
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 80 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel, Charging mode, Tx (Low data rate, above 1 GHz)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
High Channel Charger
20C
21% 989 mbars
Naga Suryadevara
28 Jan 2017 18 : 04
Additional Information
Prescan Emission Graph
100
28 Jan 2017 18 : 04 Test: Radiated - FCC15 Class B @ 3m Project: Zoll medical_01/28/2017 Tested by: Naga Suryadevara Test Notes: High Channel Charger
90
80
70
60
50
40
Level (dBuV/m)
30
20
10
Trace3: Measured Average
0 Trace1: Measured Peak
3G
Frequency (Hz)
Measured
Measured
Measured
Maximum
4G
5G
6G
7G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
8G
9G
10G
18G
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
7.449833 G
9.921329325 G
4.9598998 G
Level(dBuV/m)
55.17
56.83
62.40
AF
35.632
37.064
34.213
PA+CL
-23.588
-21.749
-25.336
Limit(dBuV/m)
74.000
74.000
74.000
Margin(dBuV/m)
-18.83
-17.17
-11.60
Hor ( -- ), Ver ( | )
--|
Azimuth (deg)(Deg)
134
228
187
Mast Height(m)
4.00
1.11
3.55
RBW(Hz)
1M
1M
1M
AF
34.213
35.632
37.064
PA+CL
-25.336
-23.588
-21.749
Limit(dBuV/m)
54.000
54.000
54.000
Margin(dBuV/m)
-15.13
-12.87
-10.63
Hor ( -- ), Ver ( | )
---
Azimuth (deg)(Deg)
187
134
228
Mast Height(m)
3.55
4.00
1.11
RBW(Hz)
1M
1M
1M
Trace3: Measured Average
Frequency(Hz)
4.9598998 G
7.449833 G
9.921329325 G
Level(dBuV/m)
38.87
41.13
43.37
Note: Manual Scans were performed from 1-3 GHz and 18 – 40 GHz, no emissions were detected above the noise floor
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 81 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel, Battery mode, Tx (Low data rate, above 1 GHz, X axis)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
Low Channel X-axis
20C
21% 989 mbars
Naga Suryadevara
29 Jan 2017 11 : 52
Additional Information
Prescan Emission Graph
Measured
Measured
Measured
Maximum
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
4.801616566 G
7.20492986 G
9.605290581 G
Level(dBuV/m)
49.44
54.49
56.79
AF
34.191
35.644
36.686
PA+CL
-25.635
-23.889
-22.078
Limit(dBuV/m)
74.000
74.000
74.000
Margin(dBuV/m)
-24.56
-19.51
-17.21
Hor ( -- ), Ver ( | )
-|
Azimuth (deg)(Deg)
302
102
154
Mast Height(m)
1.56
2.34
3.05
RBW(Hz)
1M
1M
1M
AF
34.191
35.644
36.686
PA+CL
-25.635
-23.889
-22.078
Limit(dBuV/m)
54.000
54.000
54.000
Margin(dBuV/m)
-18.34
-13.33
-10.81
Hor ( -- ), Ver ( | )
-|
Azimuth (deg)(Deg)
302
102
154
Mast Height(m)
1.56
2.34
3.05
RBW(Hz)
1M
1M
1M
Trace3: Measured Average
Frequency(Hz)
4.801616566 G
7.20492986 G
9.605290581 G
Level(dBuV/m)
35.66
40.67
43.19
Note: Manual Scans were performed from 1-3 GHz and 18 – 40 GHz, no emissions were detected above the noise floor
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 82 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel, Battery mode, Tx (Low data rate, above 1 GHz, Y axis)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
Low channel Y axis
20C
21% 989 mbars
Naga Suryadevara
28 Jan 2017 13 : 38
Additional Information
Prescan Emission Graph
100
28 Jan 2017 13 : 38 Test: Radiated - FCC15 Class B @ 3m Project: Zoll medical_01/28/2017 Tested by: Naga Suryadevara Test Notes: Low channel Y axis
90
80
70
60
50
40
Level (dBuV/m)
30
20
10
Trace3: Measured Average
0 Trace1: Measured Peak
3G
Frequency (Hz)
Measured
Measured
Measured
Maximum
4G
5G
6G
7G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
8G
9G
10G
18G
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
7.2198998 G
9.606666667 G
4.803914495 G
Level(dBuV/m)
54.41
57.61
59.47
AF
35.641
36.687
34.192
PA+CL
-23.826
-22.076
-25.631
Limit(dBuV/m)
74.000
74.000
74.000
Margin(dBuV/m)
-19.59
-16.39
-14.53
Hor ( -- ), Ver ( | )
---
Azimuth (deg)(Deg)
118
263
Mast Height(m)
3.70
1.94
1.95
RBW(Hz)
1M
1M
1M
AF
34.192
35.641
36.687
PA+CL
-25.631
-23.826
-22.076
Limit(dBuV/m)
54.000
54.000
54.000
Margin(dBuV/m)
-13.62
-13.10
-10.80
Hor ( -- ), Ver ( | )
-|
--
Azimuth (deg)(Deg)
263
118
Mast Height(m)
1.95
3.70
1.94
RBW(Hz)
1M
1M
1M
Trace3: Measured Average
Frequency(Hz)
4.803914495 G
7.2198998 G
9.606666667 G
Level(dBuV/m)
40.38
40.90
43.20
Note: Manual Scans were performed from 1-3 GHz and 18 – 40 GHz, no emissions were detected above the noise floor
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 83 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel, Battery mode, Tx (Low data rate, above 1 GHz, Z axis) – Scan1
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
Low channel Z axis
20C
21% 989 mbars
Naga Suryadevara
28 Jan 2017 13 : 03
Additional Information
Prescan Emission Graph
100
28 Jan 2017 13 : 27 Test: Radiated - FCC15 Class B @ 3m Project: Zoll medical_01/28/2017 Tested by: Naga Suryadevara Test
90
80
70
60
50
40
Level (dBuV/m)
30
20
10
Trace3: Measured Average
0 Trace1: Measured Peak
3G
Frequency (Hz)
Measured
Measured
Measured
Maximum
4G
5G
6G
7G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
8G
9G
10G
18G
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
7.21490982 G
4.803941215 G
Level(dBuV/m)
54.39
56.80
AF
35.642
34.192
PA+CL
-23.847
-25.631
Limit(dBuV/m)
74.000
74.000
Margin(dBuV/m)
-19.61
-17.20
Hor ( -- ), Ver ( | )
-|
Azimuth (deg)(Deg)
232
181
Mast Height(m)
1.59
2.05
RBW(Hz)
1M
1M
AF
34.192
35.642
PA+CL
-25.631
-23.847
Limit(dBuV/m)
54.000
54.000
Margin(dBuV/m)
-15.02
-13.12
Hor ( -- ), Ver ( | )
--
Azimuth (deg)(Deg)
181
232
Mast Height(m)
2.05
1.59
RBW(Hz)
1M
1M
Trace3: Measured Average
Frequency(Hz)
4.803941215 G
7.21490982 G
Level(dBuV/m)
38.98
40.88
Note: Manual Scans were performed from 1-3 GHz and 18 – 40 GHz, no emissions were detected above the noise floor
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 84 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel, Battery mode, Tx (Low data rate, above 1 GHz, Z axis) – Scan2
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
Low channel Z axis
20C
21% 989 mbars
Naga Suryadevara
28 Jan 2017 13 : 27
Additional Information
Prescan Emission Graph
100
28 Jan 2017 13 : 27 Test: Radiated - FCC15 Class B @ 3m Project: Zoll medical_01/28/2017 Tested by: Naga Suryadevara Test Notes: Low channel Z axis
90
80
FCC Part 15 Class B Peak Radiated Emissions @3m
70
60
FCC Part 15 Class B Average Radiated Emissions @3m
50
40
Level (dBuV/m)
30
20
10
Trace3: Measured Average
0 Trace1: Measured Peak
1G
Frequency (Hz)
Measured
Measured
Measured
Maximum
10G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
18G
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
9.602932532 G
Level(dBuV/m)
56.78
AF
36.684
PA+CL
-22.081
Limit(dBuV/m)
74.000
Margin(dBuV/m)
-17.22
Hor ( -- ), Ver ( | )
--
Azimuth (deg)(Deg)
360
Mast Height(m)
1.92
RBW(Hz)
1M
AF
36.684
PA+CL
-22.081
Limit(dBuV/m)
54.000
Margin(dBuV/m)
-10.81
Hor ( -- ), Ver ( | )
--
Azimuth (deg)(Deg)
360
Mast Height(m)
1.92
RBW(Hz)
1M
Trace3: Measured Average
Frequency(Hz)
9.602932532 G
Level(dBuV/m)
43.19
Note: Manual Scans were performed from 1-3 GHz and 18 – 40 GHz, no emissions were detected above the noise floor
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 85 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Mid Channel, Battery mode, Tx (Low data rate, above 1 GHz, X axis)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
Middle Channel X-axis
20C
21% 989 mbars
Naga Suryadevara
29 Jan 2017 11 : 01
Additional Information
Prescan Emission Graph
Measured
Measured
Measured
Maximum
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
7.323146293 G
9.774395457 G
4.882030728 G
Level(dBuV/m)
55.46
56.52
57.54
AF
35.622
36.843
34.238
PA+CL
-23.659
-21.840
-25.483
Limit(dBuV/m)
74.000
74.000
74.000
Margin(dBuV/m)
-18.54
-17.48
-16.46
Hor ( -- ), Ver ( | )
Azimuth (deg)(Deg)
294
319
129
Mast Height(m)
2.75
4.00
1.80
RBW(Hz)
1M
1M
1M
AF
34.238
35.622
36.843
PA+CL
-25.483
-23.659
-21.840
Limit(dBuV/m)
54.000
54.000
54.000
Margin(dBuV/m)
-15.39
-12.95
-10.94
Hor ( -- ), Ver ( | )
Azimuth (deg)(Deg)
129
294
319
Mast Height(m)
1.80
2.75
4.00
RBW(Hz)
1M
1M
1M
Trace3: Measured Average
Frequency(Hz)
4.882030728 G
7.323146293 G
9.774395457 G
Level(dBuV/m)
38.61
41.05
43.06
Note: Manual Scans were performed from 1-3 GHz and 18 – 40 GHz, no emissions were detected above the noise floor
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 86 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Mid Channel, Battery mode, Tx (Low data rate, above 1 GHz, Y axis)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
Mid channel Y axis
20C
21% 989 mbars
Naga Suryadevara
28 Jan 2017 14 : 05
Additional Information
Prescan Emission Graph
100
28 Jan 2017 14 : 05 Test: Radiated - FCC15 Class B @ 3m Project: Zoll medical_01/28/2017 Tested by: Naga Suryadevara Test Notes: Mid channel Y axis
90
80
70
60
50
40
Level (dBuV/m)
30
20
10
Trace3: Measured Average
0 Trace1: Measured Peak
3G
Frequency (Hz)
Measured
Measured
Measured
Maximum
4G
5G
6G
7G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
8G
9G
10G
18G
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
4.886813627 G
7.400487642 G
7.35746159 G
9.757675351 G
Level(dBuV/m)
50.05
54.86
55.23
57.84
AF
34.241
35.611
35.617
36.827
PA+CL
-25.474
-23.616
-23.640
-21.850
Limit(dBuV/m)
74.000
74.000
74.000
74.000
Margin(dBuV/m)
-23.95
-19.14
-18.77
-16.16
Hor ( -- ), Ver ( | )
Azimuth (deg)(Deg)
330
319
96
40
Mast Height(m)
3.87
4.01
3.16
3.75
RBW(Hz)
1M
1M
1M
1M
AF
34.241
35.617
35.611
36.827
PA+CL
-25.474
-23.640
-23.616
-21.850
Limit(dBuV/m)
54.000
54.000
54.000
54.000
Margin(dBuV/m)
-17.54
-12.94
-12.76
-10.96
Hor ( -- ), Ver ( | )
Azimuth (deg)(Deg)
330
96
319
40
Mast Height(m)
3.87
3.16
4.01
3.75
RBW(Hz)
1M
1M
1M
1M
Trace3: Measured Average
Frequency(Hz)
4.886813627 G
7.35746159 G
7.400487642 G
9.757675351 G
Level(dBuV/m)
36.46
41.06
41.24
43.04
Note: Manual Scans were performed from 1-3 GHz and 18 – 40 GHz, no emissions were detected above the noise floor
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 87 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Mid Channel, Battery mode, Tx (Low data rate, above 1 GHz, Z axis)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
Mid channel Z axis
20C
21% 989 mbars
Naga Suryadevara
28 Jan 2017 14 : 35
Additional Information
Prescan Emission Graph
100
28 Jan 2017 14 : 35 Test: Radiated - FCC15 Class B @ 3m Project: Zoll medical_01/28/2017 Tested by: Naga Suryadevara Test Notes: Mid channel Z axis
90
80
70
60
50
40
Level (dBuV/m)
30
20
10
Trace3: Measured Average
0 Trace1: Measured Peak
3G
Frequency (Hz)
4G
5G
6G
7G
Measured Peak Value
Measured Quasi Peak Value
Measured Average Value
Maximum Value of Mast and Turntable
8G
9G
10G
18G
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
7.32998664 G
9.771663326 G
4.881877087 G
Level(dBuV/m)
54.70
56.93
58.77
AF
35.621
36.840
34.238
PA+CL
-23.655
-21.842
-25.483
Limit(dBuV/m)
74.000
74.000
74.000
Margin(dBuV/m)
-19.30
-17.07
-15.23
Hor ( -- ), Ver ( | )
-|
Azimuth (deg)(Deg)
230
166
152
Mast Height(m)
2.18
2.69
3.17
RBW(Hz)
1M
1M
1M
AF
34.238
35.621
36.840
PA+CL
-25.483
-23.655
-21.842
Limit(dBuV/m)
54.000
54.000
54.000
Margin(dBuV/m)
-14.82
-12.95
-10.94
Hor ( -- ), Ver ( | )
--
Azimuth (deg)(Deg)
152
230
166
Mast Height(m)
3.17
2.18
2.69
RBW(Hz)
1M
1M
1M
Trace3: Measured Average
Frequency(Hz)
4.881877087 G
7.32998664 G
9.771663326 G
Level(dBuV/m)
39.18
41.05
43.06
Note: Manual Scans were performed from 1-3 GHz and 18 – 40 GHz, no emissions were detected above the noise floor
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 88 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel, Battery mode, Tx (Low data rate, above 1 GHz, X axis)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
High Channel X-axis Battery
20C
21% 989 mbars
Naga Suryadevara
29 Jan 2017 10 : 09
Additional Information
Prescan Emission Graph
Measured
Measured
Measured
Maximum
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
7.452124249 G
9.91260521 G
4.95997328 G
Level(dBuV/m)
55.04
57.22
60.06
AF
35.633
37.047
34.213
PA+CL
-23.587
-21.754
-25.336
Limit(dBuV/m)
74.000
74.000
74.000
Margin(dBuV/m)
-18.96
-16.78
-13.94
Hor ( -- ), Ver ( | )
Azimuth (deg)(Deg)
360
331
338
Mast Height(m)
1.67
1.38
2.91
RBW(Hz)
1M
1M
1M
AF
34.213
35.633
37.047
PA+CL
-25.336
-23.587
-21.754
Limit(dBuV/m)
54.000
54.000
54.000
Margin(dBuV/m)
-15.57
-12.71
-10.47
Hor ( -- ), Ver ( | )
Azimuth (deg)(Deg)
338
360
331
Mast Height(m)
2.91
1.67
1.38
RBW(Hz)
1M
1M
1M
Trace3: Measured Average
Frequency(Hz)
4.95997328 G
7.452124249 G
9.91260521 G
Level(dBuV/m)
38.43
41.29
43.53
Note: Manual Scans were performed from 1-3 GHz and 18 – 40 GHz, no emissions were detected above the noise floor
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 89 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel, Battery mode, Tx (Low data rate, above 1 GHz, Y axis)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
High Channel Y-axis
20C
21% 989 mbars
Naga Suryadevara
28 Jan 2017 19 : 03
Additional Information
Prescan Emission Graph
100
28 Jan 2017 19 : 03 Test: Radiated - FCC15 Class B @ 3m Project: Zoll medical_01/28/2017 Tested by: Naga Suryadevara Test Notes: High Channel Y-axis
90
80
70
60
50
40
Level (dBuV/m)
30
20
10
Trace3: Measured Average
0 Trace1: Measured Peak
3G
Frequency (Hz)
Measured
Measured
Measured
Maximum
4G
5G
6G
7G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
8G
9G
10G
18G
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
7.447775551 G
9.919271877 G
4.95997328 G
Level(dBuV/m)
55.17
57.09
63.87
AF
35.632
37.060
34.213
PA+CL
-23.589
-21.750
-25.336
Limit(dBuV/m)
74.000
74.000
74.000
Margin(dBuV/m)
-18.83
-16.91
-10.13
Hor ( -- ), Ver ( | )
-|
Azimuth (deg)(Deg)
319
69
Mast Height(m)
1.32
2.27
2.88
RBW(Hz)
1M
1M
1M
AF
34.213
35.632
37.060
PA+CL
-25.336
-23.589
-21.750
Limit(dBuV/m)
54.000
54.000
54.000
Margin(dBuV/m)
-14.30
-12.88
-10.63
Hor ( -- ), Ver ( | )
-|
Azimuth (deg)(Deg)
69
319
Mast Height(m)
2.88
1.32
2.27
RBW(Hz)
1M
1M
1M
Trace3: Measured Average
Frequency(Hz)
4.95997328 G
7.447775551 G
9.919271877 G
Level(dBuV/m)
39.70
41.12
43.37
Note: Manual Scans were performed from 1-3 GHz and 18 – 40 GHz, no emissions were detected above the noise floor
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 90 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel, Battery mode, Tx (Low data rate, above 1 GHz, Z axis)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
High channel Z axis
20C
21% 989 mbars
Naga Suryadevara
28 Jan 2017 15 : 05
Additional Information
Prescan Emission Graph
100
28 Jan 2017 15 : 05 Test: Radiated - FCC15 Class B @ 3m Project: Zoll medical_01/28/2017 Tested by: Naga Suryadevara Test Notes: High channel Z axis
90
80
70
60
50
40
Level (dBuV/m)
30
20
10
Trace3: Measured Average
0 Trace1: Measured Peak
3G
Frequency (Hz)
Measured
Measured
Measured
Maximum
4G
5G
6G
7G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
8G
9G
10G
18G
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
7.446472946 G
9.765190381 G
4.95995324 G
Level(dBuV/m)
55.17
56.51
63.74
AF
35.631
36.834
34.213
PA+CL
-23.590
-21.846
-25.336
Limit(dBuV/m)
74.000
74.000
74.000
Margin(dBuV/m)
-18.83
-17.49
-10.26
Hor ( -- ), Ver ( | )
-|
Azimuth (deg)(Deg)
229
19
173
Mast Height(m)
1.69
2.05
1.93
RBW(Hz)
1M
1M
1M
AF
34.213
35.631
36.834
PA+CL
-25.336
-23.590
-21.846
Limit(dBuV/m)
54.000
54.000
54.000
Margin(dBuV/m)
-13.91
-12.88
-10.95
Hor ( -- ), Ver ( | )
-|
Azimuth (deg)(Deg)
173
229
19
Mast Height(m)
1.93
1.69
2.05
RBW(Hz)
1M
1M
1M
Trace3: Measured Average
Frequency(Hz)
4.95995324 G
7.446472946 G
9.765190381 G
Level(dBuV/m)
40.09
41.12
43.05
Note: Manual Scans were performed from 1-3 GHz and 18 – 40 GHz, no emissions were detected above the noise floor
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 91 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel, Charging mode, Lower Band edge emissions (High data rate)
RBW
100 kHz
7.30 dB æ V
VBW
300 kHz
2.40000000 GHz
SWT
300 ms
Marker 2 [T2]
Ref Lvl
72 dB æ V
RF Att
0 dB
Unit
dB æ V
72
2 [T2]
7.30 dB æ V
2.40000000 GHz
1 [T2]
60
51.68 dB æ V
2.40181764 GHz
50
IN1
40
2MAX
2MA
30
P20
20
10
-10
-20
-28
Start 2.398 GHz
Date:
29.JAN.2017
500 kHz/
Stop 2.403 GHz
13:30:38
Emissions at lower band edge were 20 dB below the peak
Low Channel, Battery mode, Lower Band edge emissions (High data rate X axis)
RBW
100 kHz
24.05 dB æ V
VBW
300 kHz
2.40000000 GHz
SWT
300 ms
Marker 2 [T2]
Ref Lvl
80 dB æ V
RF Att
0 dB
Unit
dB æ V
80
2 [T2]
24.05 dB æ V
2.40000000 GHz
70
1 [T2]
51.46 dB æ V
2.40183768 GHz
60
50
IN1
2MAX
2MA
40
30
20
10
-10
-20
Start 2.393 GHz
Date:
29.JAN.2017
1 MHz/
Stop 2.403 GHz
12:55:04
Emissions at lower band edge were 20 dB below the peak
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 92 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Low Channel, Battery mode, Lower Band edge emissions (High data rate Y axis)
RBW
100 kHz
6.36 dB æ V
VBW
300 kHz
2.40000000 GHz
SWT
Marker 2 [T2]
Ref Lvl
72 dB æ V
5 ms
RF Att
0 dB
Unit
dB æ V
72
2 [T2]
6.36 dB æ V
2.40000000 GHz
1 [T2]
60
55.77 dB æ V
2.40183768 GHz
50
IN1
40
2MAX
2MA
30
P20
20
10
-10
-20
-28
Start 2.398 GHz
Date:
29.JAN.2017
500 kHz/
Stop 2.403 GHz
14:09:05
Emissions at lower band edge were 20 dB below the peak
Low Channel, Battery mode, Lower Band edge emissions (High data rate Z axis)
RBW
100 kHz
6.39 dB æ V
VBW
300 kHz
2.40000000 GHz
SWT
Marker 2 [T2]
Ref Lvl
72 dB æ V
5 ms
RF Att
0 dB
Unit
dB æ V
72
2 [T2]
6.39 dB æ V
2.40000000 GHz
1 [T2]
60
49.58 dB æ V
2.40216834 GHz
50
IN1
40
2MAX
2MA
30
P20
20
10
-10
-20
-28
Start 2.398 GHz
Date:
29.JAN.2017
500 kHz/
Stop 2.403 GHz
14:03:31
Emissions at lower band edge were 20 dB below the peak
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 93 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel, Charging mode, Upper Band edge emissions (High data rate)
RBW
1 MHz
18.84 dB æ V
VBW
3 MHz
2.48350000 GHz
SWT
Marker 2 [T2]
Ref Lvl
72 dB æ V
RF Att
300 ms
Unit
0 dB
dB æ V
72
2 [T2]
18.84 dB æ V
2.48350000 GHz
1 [T2]
60
54.35 dB æ V
2.47980561 GHz
50
IN1
40
2MAX
2MA
30
P20
20
10
-10
-20
-28
Start 2.479 GHz
Date:
Detector
Type
PK
AVG
Ant.
Pol.
(V/H)
29.JAN.2017
Frequency Reading
MHz
dB(uV)
2483.500
18.84
2483.500
4.56
600 kHz/
Stop 2.485 GHz
13:21:22
Antenna
Factor
dB(1/m)
32.22
32.22
Cable
Loss
dB
5.18
5.18
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Pre-amp
Factor
dB
0.00
0.00
Distance
Factor
Net
Limit
dB
dB(uV/m) dB(uV/m)
0.00
56.25
74.00
0.00
41.97
54.00
Margin
dB
-17.75
-12.03
Bandwidth
1/3 MHz
1/3 MHz
Page 94 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel, Battery mode, Upper Band edge emissions (High data rate, X axis)
RBW
1 MHz
18.12 dB æ V
VBW
3 MHz
2.48350000 GHz
SWT
5 ms
Marker 2 [T2]
Ref Lvl
72 dB æ V
RF Att
Unit
0 dB
dB æ V
72
2 [T2]
18.12 dB æ V
2.48350000 GHz
1 [T2]
60
54.33 dB æ V
2.47986573 GHz
50
IN1
40
2MAX
2MA
30
P20
20
10
-10
-20
-28
Start 2.479 GHz
Date:
Detector
Type
PK
AVG
Ant.
Pol.
(V/H)
29.JAN.2017
Frequency Reading
MHz
dB(uV)
2483.500
18.12
2483.500
4.34
600 kHz/
Stop 2.485 GHz
13:44:12
Antenna
Factor
dB(1/m)
32.22
32.22
Cable
Loss
dB
5.18
5.18
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Pre-amp
Factor
dB
0.00
0.00
Distance
Factor
Net
Limit
dB
dB(uV/m) dB(uV/m)
0.00
55.53
74.00
0.00
41.75
54.00
Margin
dB
-18.47
-12.25
Bandwidth
1/3 MHz
1/3 MHz
Page 95 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel, Battery mode, Upper Band edge emissions (High data rate, Y axis)
RBW
1 MHz
17.53 dB æ V
VBW
3 MHz
2.48350000 GHz
SWT
5 ms
Marker 2 [T2]
Ref Lvl
72 dB æ V
RF Att
Unit
0 dB
dB æ V
72
2 [T2]
17.53 dB æ V
2.48350000 GHz
1 [T2]
60
52.74 dB æ V
2.47984168 GHz
50
IN1
40
2MAX
2MA
30
P20
20
10
-10
-20
-28
Start 2.479 GHz
Date:
Detector
Type
PK
AVG
Ant.
Pol.
(V/H)
29.JAN.2017
Frequency Reading
MHz
dB(uV)
2483.500
17.53
2483.500
4.22
600 kHz/
Stop 2.485 GHz
13:50:55
Antenna
Factor
dB(1/m)
32.22
32.22
Cable
Loss
dB
5.18
5.18
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Pre-amp
Factor
dB
0.00
0.00
Distance
Factor
Net
Limit
dB
dB(uV/m) dB(uV/m)
0.00
54.94
74.00
0.00
41.63
54.00
Margin
dB
-19.06
-12.37
Bandwidth
1/3 MHz
1/3 MHz
Page 96 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
High Channel, Battery mode, Upper Band edge emissions (High data rate, Z axis)
RBW
1 MHz
18.89 dB æ V
VBW
3 MHz
2.48350000 GHz
SWT
5 ms
Marker 2 [T2]
Ref Lvl
72 dB æ V
RF Att
Unit
0 dB
dB æ V
72
2 [T2]
18.89 dB æ V
2.48350000 GHz
60
1 [T2]
56.78 dB æ V
2.48015431 GHz
50
IN1
40
2MAX
2MA
30
P20
20
10
-10
-20
-28
Start 2.479 GHz
Date:
Detector
Type
PK
AVG
Ant.
Pol.
(V/H)
29.JAN.2017
Frequency Reading
MHz
dB(uV)
2483.500
18.89
2483.500
5.68
600 kHz/
Stop 2.485 GHz
13:58:48
Antenna
Factor
dB(1/m)
32.22
32.22
Cable
Loss
dB
5.18
5.18
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Pre-amp
Factor
dB
0.00
0.00
Distance
Factor
Net
Limit
dB
dB(uV/m) dB(uV/m)
0.00
56.30
74.00
0.00
43.09
54.00
Margin
dB
-17.70
-10.91
Bandwidth
1/3 MHz
1/3 MHz
Page 97 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Test Date:
Test Personnel:
Supervising/Reviewing
Engineer:
(Where Applicable)
Product Standard:
Input Voltage:
Pretest Verification w/
Ambient Signals or
BB Source:
Naga Suryadevara
01/27/2017
01/28/2017
01/29/2017
N/A
FCC Part 15 Subpart C
(15.247)
120VAC 60Hz
Internal Battery
BB Source
Limit Applied:
As specified in the plots (Section
12.4)
Ambient Temperature:
19, 20 19 ºC
Relative Humidity:
30, 21 19 %
Atmospheric Pressure:
981, 989, 999 mbars
Deviations, Additions, or Exclusions: None
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 98 of 117
Intertek
Report Number: 102716227BOX-002
13
Issued: 03/09/2017
Radiated Emissions (Digital Device and Receiver)
13.1 Method
Tests are performed in accordance with FCC Part 15 Subpart B.
TEST SITE: 10m ALSE
The 10m ALSE is 13m (Length) x 21m (Depth) x 10m (Height) with the effective size in terms of space from the
tips of the absorber is 12m (Length) x 20m (Depth) x 8.5m (Height). This chamber achieves broadband
performance using a unique arrangement of hybrid and ferrite tile absorber. This chamber has a built in 3m
diameter turntable (Embedded type). The metal structure of the table makes electrical connection around the
entire circumference of the turntable to the ground plane with a metal brush type connection. The turntable is
located on one end of the chamber and the antennas are mounted 3 and 10 meters away at the other end of
the chamber on the adjustable an Antenna Mast. The antenna mast is a non-conductive bore sighted type with
remote control of antenna height and polarization. The Antenna Mast and the turntable can be remotely
controlled through the controller located in the adjacent Control room. A Styrofoam table 80 cm high is used for
table-top equipment.
Measurement Uncertainty
Measurement
Radiated Emissions, 10m
Radiated Emissions, 3m
Radiated Emissions, 3m
Radiated Emissions, 3m
Radiated Emissions, 3m
Radiated Emissions, 3m
Frequency
Range
30-1000 MHz
30-1000 MHz
1-6 GHz
6-15 GHz
15-18 GHz
18-40 GHz
As shown in the table above our radiated emissions
Expanded
Uncertainty
(k=2)
4.6 dB
5.3 dB
4.5 dB
5.2 dB
5.0 dB
5.0 dB
Ucispr
6.3 dB
6.3 dB
5.2 dB
5.5 dB
5.5 dB
5.5 dB
U lab is less than the corresponding U CISPR reference
value in CISPR 16-4-2 Table 1, hence the compliance of the product is only based on the measured
value, and no measurement uncertainty correction is required, based on CISPR 22 and CISPR 11 (for
2006 and later revisions) Clause 11.
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 99 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Sample Calculation
The field strength is calculated by adding the Antenna Factor and Cable Factor, and subtracting the
Amplifier Gain (if any) from the measured reading. The basic equation with a sample calculation is as
follows:
FS = RA + AF + CF - AG
Where
FS = Field Strength in dBV/m
RA = Receiver Amplitude (including preamplifier) in dBV
CF = Cable Attenuation Factor in dB
AF = Antenna Factor in dB
AG = Amplifier Gain in dB
In the following table(s), the reading shown on the data table reflects the preamplifier gain. An example
for the calculations in the following table is as follows.
Assume a receiver reading of 52.0 dBV is obtained. The antenna factor of 7.4 dB and cable factor of
1.6 dB is added. The amplifier gain of 29 dB is subtracted, giving a field strength of 32 dBV/m. This
value in dBV/m was converted to its corresponding level in V/m.
RA = 52.0 dBV
AF = 7.4 dB/m
CF = 1.6 dB
AG = 29.0 dB
FS = 32 dBV/m
To convert from dBV to V or mV the following was used:
where UF = Net Reading in V
NF = Net Reading in dBV
(NF / 20)
UF = 10
Example:
FS = RA + AF + CF – AG = 52.0 + 7.4 + 1.6 – 29.0 = 32.0
(32 dBV / 20)
UF = 10
= 39.8 V/m
Alternately, when C5 Software is used, the “Level” includes all losses and gains and is compared directly
in the “Margin” column to the “Limit”. “AF” is the Antenna Factor; “PA+CL” are Preamp and Cable Loss.
These are already accounted for in the “Level” column.
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 100 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
13.2 Test Equipment Used:
Asset
Description
Manufacturer
DAV004’
145145’
Weather Station
Broadband Hybrid Antenna 30 MHz - 3 GHz
Davis Instruments
Sunol Sciences Corp.
145-410’
PRE10’
ETS001’
Cables 145-420 145-421 145-422 145-406
30-1000MHz pre-amp
1-18GHz DRG Horn Antenna
Huber + Suhner
ITS
ETS-Lindgren
145-416’
145014’
PRE10’
Cables 145-420 145-423 145-424 145-408
Preamplifier (1 GHz to 26.5 GHz)
30-1000MHz pre-amp
Huber + Suhner
Hewlett Packard
ITS
Software Utilized:
Name
Compliance5
Model
7400
JB3
10m Track A
Cables
PRE10
3117
3m Track B
cables
8449B
PRE10
Serial
PE80529A61
A122313
Cal Date
Cal Due
05/02/2016
03/09/2016
05/02/2017
03/09/2017
multiple
PRE10
00143259
07/30/2016
12/16/2016
02/10/2016
07/30/2017
12/16/2017
02/10/2017
multiple
3008A00232
PRE10
07/30/2016
05/27/2016
12/16/2016
07/30/2017
05/27/2017
12/16/2017
Manufacturer
Version
Teseq
5.26.46.46
13.3 Results:
The sample tested was found to Comply.
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 101 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
13.4 Plots/Data:
Battery mode, Rx, 30 – 1000 MHz
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 10m
Zoll medical_01/26/2017
Rx Low channel, battery with phantom
19 C
30% 981 mbars
Naga Suryadevara
26 Jan 2017 15 : 36
Additional Information
Prescan Emission Graph
100
26 Jan 2017 15 : 36 Test: Radiated - FCC15 Class B @ 10m Project: Zoll medical_01/26/2017 Tested by: Naga Suryadevara Test Notes: Rx Low channel, battery with phantom
90
80
70
60
50
40
Level (dBuV/m)
30 FCC Part 15 Class B QP Radiated Emissions @10m
20
10
Trace3: Measured Average
Trace2: Measured Quasi Peak
0 Trace1: Measured Peak
30M
Frequency (Hz)
100M
1G
Measured Peak Value
Measured Quasi Peak Value
Measured Average Value
Maximum Value of Mast and Turntable
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace2: Measured Quasi Peak
Frequency(Hz)
336.049098491 M
368.030260519 M
495.908416645 M
399.953707116 M
431.929258922 M
463.996994096 M
Level(dBuV/m)
15.21
16.62
16.62
17.99
18.28
21.77
AF
20.221
21.100
23.918
21.698
22.739
23.660
PA+CL
-38.136
-38.045
-37.396
-37.940
-37.778
-37.590
Limit(dBuV/m)
36.020
36.020
36.020
36.020
36.020
36.020
Margin(dBuV/m)
-20.81
-19.40
-19.40
-18.03
-17.74
-14.25
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Hor ( -- ), Ver ( | )
-------
Azimuth (deg)(Deg)
20
184
293
164
141
Mast Height(m)
2.09
3.04
1.45
2.50
2.28
1.56
RBW(Hz)
120 k
120 k
120 k
120 k
120 k
120 k
Page 102 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Charging mode, Rx, 30 – 1000 MHz (XT 1505)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 10m
Zoll medical_01/26/2017
Rx Low channel, charging mode
19 C
30% 981 mbars
Naga Suryadevara
26 Jan 2017 13 : 56
Additional Information
Prescan Emission Graph
100
26 Jan 2017 13 : 56 Test: Radiated - FCC15 Class B @ 10m Project: Zoll medical_01/26/2017 Tested by: Naga Suryadevara Test Notes: Rx Low channel, charging mode
90
80
70
60
50
40
Level (dBuV/m)
30 FCC Part 15 Class B QP Radiated Emissions @10m
20
10
Trace3: Measured Average
Trace2: Measured Quasi Peak
0 Trace1: Measured Peak
30M
Frequency (Hz)
Measured
Measured
Measured
Maximum
100M
1G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace2: Measured Quasi Peak
Frequency(Hz)
335.933667629 M
880.022645431 M
303.958115824 M
431.966934273 M
847.946894224 M
815.972144022 M
Level(dBuV/m)
20.46
28.95
29.04
30.24
30.58
31.24
AF
20.219
28.300
19.758
22.739
28.000
28.000
PA+CL
-38.137
-35.446
-38.210
-37.777
-35.554
-35.661
Limit(dBuV/m)
36.020
36.020
36.020
36.020
36.020
36.020
Margin(dBuV/m)
-15.56
-7.07
-6.98
-5.78
-5.44
-4.78
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Hor ( -- ), Ver ( | )
-------
Azimuth (deg)(Deg)
97
130
261
85
121
297
Mast Height(m)
2.51
1.26
3.83
1.87
1.15
1.05
RBW(Hz)
120 k
120 k
120 k
120 k
120 k
120 k
Page 103 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Charging mode, Rx, 30 – 1000 MHz (XT 1527)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 10m
Zoll medical_01/30/2017
Bluetooth Rx mode_Cellphone with LTE
20 C
16% 996 mbars
Naga Suryadevara
30 Jan 2017 12 : 12
Additional Information
Prescan Emission Graph
100
30 Jan 2017 12 : 12 Test: Radiated - FCC15 Class B @ 10m Project: Zoll medical_01/30/2017 Tested by: Naga Suryadevara Test Notes: Bluetooth Rx mode_Cellphone with LTE
90
80
70
60
50
40
Level (dBuV/m)
30 FCC Part 15 Class B QP Radiated Emissions @10m
20
10
Trace3: Measured Average
Trace2: Measured Quasi Peak
0 Trace1: Measured Peak
30M
Frequency (Hz)
Measured
Measured
Measured
Maximum
100M
1G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace2: Measured Quasi Peak
Frequency(Hz)
288.553907667 M
617.134268321 M
699.310220764 M
766.514829782 M
460.825450673 M
32.015030228 M
Level(dBuV/m)
14.77
15.95
17.76
18.58
19.78
16.47
AF
19.400
25.128
26.700
27.400
23.533
25.888
PA+CL
-38.255
-36.778
-36.175
-35.861
-37.611
-40.044
Limit(dBuV/m)
36.020
36.020
36.020
36.020
36.020
30.000
Margin(dBuV/m)
-21.25
-20.07
-18.26
-17.44
-16.24
-13.53
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Hor ( -- ), Ver ( | )
-|
-|
-|
Azimuth (deg)(Deg)
36
277
176
96
286
Mast Height(m)
1.04
1.24
3.04
2.56
2.16
3.01
RBW(Hz)
120 k
120 k
120 k
120 k
120 k
120 k
Page 104 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Battery mode, Rx, 1 – 13 GHz
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
Low Channel Rx Battery
20C
21% 989 mbars
Naga Suryadevara
29 Jan 2017 12 : 17
Additional Information
Prescan Emission Graph
Measured
Measured
Measured
Maximum
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
3.43263193 G
2.562845692 G
4.627862391 G
10.761322645 G
6.647040748 G
12.527869071 G
Level(dBuV/m)
46.90
47.90
50.31
58.64
59.14
64.82
AF
32.802
32.360
34.427
37.706
35.533
39.115
PA+CL
-27.303
-28.116
-25.895
-20.189
-24.361
-17.549
Limit(dBuV/m)
74.000
74.000
74.000
74.000
74.000
74.000
Margin(dBuV/m)
-27.10
-26.10
-23.69
-15.36
-14.86
-9.18
Hor ( -- ), Ver ( | )
--|
Azimuth (deg)(Deg)
142
320
315
306
321
Mast Height(m)
3.28
4.00
2.65
3.06
3.02
2.17
RBW(Hz)
1M
1M
1M
1M
1M
1M
AF
32.802
32.360
34.427
37.706
35.533
39.115
PA+CL
-27.303
-28.116
-25.895
-20.189
-24.361
-17.549
Limit(dBuV/m)
54.000
54.000
54.000
54.000
54.000
54.000
Margin(dBuV/m)
-20.81
-20.05
-17.97
-8.79
-8.39
-3.35
Hor ( -- ), Ver ( | )
--|
Azimuth (deg)(Deg)
142
320
315
306
321
Mast Height(m)
3.28
4.00
2.65
3.06
3.02
2.17
RBW(Hz)
1M
1M
1M
1M
1M
1M
Trace3: Measured Average
Frequency(Hz)
3.43263193 G
2.562845692 G
4.627862391 G
10.761322645 G
6.647040748 G
12.527869071 G
Level(dBuV/m)
33.19
33.95
36.03
45.21
45.61
50.65
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 105 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Charging mode, Rx, 1 – 13 GHz (XT 1505)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
Low Channel X-axis charging cellphone w/o LTE
20C
21% 989 mbars
Naga Suryadevara
29 Jan 2017 16 : 19
Additional Information
Prescan Emission Graph
100
29 Jan 2017 16 : 19 Test: Radiated - FCC15 Class B @ 3m Project: Zoll medical_01/28/2017 Tested by: Naga Suryadevara Test Notes: Low Channel X-axis charging cellphone w/o LTE
90
80
FCC Part 15 Class B Peak Radiated Emissions @3m
70
60
FCC Part 15 Class B Average Radiated Emissions @3m
50
40
Level (dBuV/m)
30
20
10
Trace3: Measured Average
0 Trace1: Measured Peak
1G
Frequency (Hz)
Measured
Measured
Measured
Maximum
10G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
13G
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
1.887882431 G
2.439425518 G
6.625210421 G
12.348550435 G
Level(dBuV/m)
45.60
49.78
59.21
63.27
AF
30.801
32.257
35.548
39.103
PA+CL
-28.451
-28.153
-24.447
-18.019
Limit(dBuV/m)
74.000
74.000
74.000
74.000
Margin(dBuV/m)
-28.40
-24.22
-14.79
-10.73
Hor ( -- ), Ver ( | )
--|
--
Azimuth (deg)(Deg)
96
128
253
163
Mast Height(m)
1.31
1.07
1.07
1.67
RBW(Hz)
1M
1M
1M
1M
AF
30.801
32.257
35.548
39.103
PA+CL
-28.451
-28.153
-24.447
-18.019
Limit(dBuV/m)
54.000
54.000
54.000
54.000
Margin(dBuV/m)
-21.80
-13.59
-8.73
-4.50
Hor ( -- ), Ver ( | )
--|
--
Azimuth (deg)(Deg)
96
128
253
163
Mast Height(m)
1.31
1.07
1.07
1.67
RBW(Hz)
1M
1M
1M
1M
Trace3: Measured Average
Frequency(Hz)
1.887882431 G
2.439425518 G
6.625210421 G
12.348550435 G
Level(dBuV/m)
32.20
40.41
45.27
49.50
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 106 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Charging mode, Rx, 1 – 13 GHz (XT 1527)
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
Radiated - FCC15 Class B @ 3m
Zoll medical_01/28/2017
Low Channel X-axis charging cellphone vt LTE
20C
21% 989 mbars
Naga Suryadevara
29 Jan 2017 15 : 19
Additional Information
Prescan Emission Graph
100
29 Jan 2017 15 : 19 Test: Radiated - FCC15 Class B @ 3m Project: Zoll medical_01/28/2017 Tested by: Naga Suryadevara Test Notes: Low Channel X-axis charging cellphone vt LTE
90
80
FCC Part 15 Class B Peak Radiated Emissions @3m
70
60
FCC Part 15 Class B Average Radiated Emissions @3m
50
40
Level (dBuV/m)
30
20
10
Trace3: Measured Average
0 Trace1: Measured Peak
1G
Frequency (Hz)
Measured
Measured
Measured
Maximum
10G
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
13G
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace1: Measured Peak
Frequency(Hz)
12.102204409 G
2.612738811 G
2.433547095 G
6.75744823 G
12.101162325 G
12.539812959 G
Level(dBuV/m)
58.03
47.07
47.23
60.88
61.71
64.07
AF
38.865
32.437
32.262
35.524
38.864
39.126
PA+CL
-18.672
-28.097
-28.155
-23.979
-18.675
-17.523
Limit(dBuV/m)
74.000
74.000
74.000
74.000
74.000
74.000
Margin(dBuV/m)
-15.97
-26.93
-26.77
-13.12
-12.29
-9.93
Hor ( -- ), Ver ( | )
----|
--
Azimuth (deg)(Deg)
139
52
49
29
225
Mast Height(m)
0.00
1.31
1.31
1.42
1.57
3.71
RBW(Hz)
1M
1M
1M
1M
1M
1M
AF
32.262
32.437
35.524
38.864
39.126
PA+CL
-28.155
-28.097
-23.979
-18.675
-17.523
Limit(dBuV/m)
54.000
54.000
54.000
54.000
54.000
Margin(dBuV/m)
-21.14
-20.91
-6.96
-5.75
-3.48
Hor ( -- ), Ver ( | )
---|
--
Azimuth (deg)(Deg)
52
139
49
29
225
Mast Height(m)
1.31
1.31
1.42
1.57
3.71
RBW(Hz)
1M
1M
1M
1M
1M
Trace3: Measured Average
Frequency(Hz)
2.433547095 G
2.612738811 G
6.75744823 G
12.101162325 G
12.539812959 G
Level(dBuV/m)
32.86
33.09
47.04
48.25
50.52
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 107 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Test Date:
Test Personnel:
Supervising/Reviewing
Engineer:
(Where Applicable)
Product Standard:
Input Voltage:
Pretest Verification w/
Ambient Signals or
BB Source:
Naga Suryadevara
01/26/2017
01/29/2017
01/30/2017
N/A
FCC Part 15 Subpart B
120VAC 60Hz
Internal Battery
BB Source
Limit Applied:
Ambient Temperature:
19, 20, 20 ºC
Relative Humidity:
30, 21, 16 %
Atmospheric Pressure:
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Class A
981, 989, 996 mbars
Page 108 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
14 AC Mains Conducted Emissions
14.1 Method
Tests are performed in accordance with FCC Part 15 Subpart C and FCC Part 15 Subpart B.
TEST SITE: EMC Lab
The EMC Lab has one Semi-anechoic Chamber and one Shielded Chamber. AC Mains Power is available at 120,
230, and 277 Single Phase; 208, 400, and 480 3-Phase. Large reference ground-planes are installed in the general
lab area to facilitate EMC work not requiring a shielded environment.
Measurement Uncertainty
Measurement
AC Line Conducted
Emissions
Telco Port Emissions
Frequency
Range
Expanded
Uncertainty
(k=2)
Ucispr
150 kHz - 30 MHz
150 kHz - 30 MHz
2.8dB
3.2dB
3.4dB
5.0dB
As shown in the table above our conducted emissions
U lab is less than the corresponding U CISPR
reference value in CISPR 16-4-2 Table 1, hence the compliance of the product is only based on the
measured value, and no measurement uncertainty correction is required, based on CISPR 22 and CISPR
11 (for 2006 and later revisions) Clause 11.
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 109 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Sample Calculations
The following is how net line-conducted readings were determined:
NF = RF + LF + CF + AF
Where NF = Net Reading in dBV
RF = Reading from receiver in dBV
LF = LISN or ISN Correction Factor in dB
CF = Cable Correction Factor in dB
AF = Attenuator Loss Factor in dB
To convert from dBV to V or mV the following was used:
where UF = Net Reading in V
NF = Net Reading in dBV
(NF / 20)
UF = 10
Example:
NF = RF + LF + CF + AF = 28.5 + 0.2 + 0.4 + 20.0 = 49.1 dBV
(49.1 dBV / 20)
UF = 10
= 285.1 V/m
Alternately, when C5 Software is used, the “Level” includes all losses and gains and is compared directly
in the “Margin” column to the “Limit”. “TF” is the Transducer Factor; in this case LISN or ISN loss.
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 110 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
14.2 Test Equipment Used:
Asset
DAV002’
ROS002’
DS22’
CBLBNC2
012-4’
LISN31’
Weather Station
Description
Manufacturer
Davis Instruments
9kHz to 3GHz EMI Test Receiver
Attenuator, 20dB
Rohde & Schwartz
Mini Circuits
50 Ohm Coaxial Cable
LISN - CISPR16 Compliant 9kHz-30MHz
Pomona
Com-Power
Software Utilized:
Name
Compliance 5
Model
7400
ESCI
1166.5950K0
20dB, 50 ohm
RG58C/U
LI-215A
Serial
PE80519A93
Cal Date
06/01/2016
Cal Due
06/01/2017
100067
DS22
CBLBCN2012
-4
191957
07/29/2016
09/08/2016
07/29/2017
09/08/2017
03/21/2016
03/14/2016
03/21/2017
03/14/2017
Manufacturer
Version
Teseq
5.26.46.46
14.3 Results:
The sample tested was found to Comply.
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 111 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
14.4 Plots/Data:
120VAC 60Hz, Tx mode
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
LISN - FCC15 Class B
Zoll medical_01/27/2017
120VAC 60Hz_Tx mode
22 C
14% 987 mbars
Naga Suryadevara
27 Jan 2017 16 : 51
Additional Information
Prescan Emission Graph
100
27 Jan 2017 16 : 51 Test: LISN - FCC15 Class B Project: Zoll medical_01/27/2017 Tested by: Naga Suryadevara Test Notes: 120VAC 60Hz_Tx mode
90
80
70
FCC Part 15 Class B Conducted QP (Mains Ports)
60
FCC Part 15 Class B Conducted AV (Mains Ports)
50
40
30
Level (dBuV)
20
10
Trace3: Measured Average
Trace2: Measured Quasi Peak
0 Trace1: Measured Peak
150k
Frequency (Hz)
Measured
Measured
Measured
Maximum
1M
10M
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
30M
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace2: Measured Quasi Peak
Frequency(Hz)
393.95 k
161.05 k
200.15 k
191.65 k
174.65 k
166.15 k
374.4 k
Level(dBuV)
33.90
41.73
40.12
41.11
43.73
44.82
38.64
TF
0.027
0.071
0.040
0.047
0.060
0.067
0.028
PA+CL
20.580
20.594
20.591
20.592
20.593
20.593
20.581
Limit(dBuV)
57.980
65.410
63.604
63.965
64.736
65.151
58.403
Margin(dBuV)
-24.08
-23.68
-23.48
-22.85
-21.01
-20.33
-19.77
RBW(Hz)
9k
9k
9k
9k
9k
9k
9k
Comment
LINE
L1
TF
0.071
0.067
0.047
0.060
0.040
0.027
0.028
PA+CL
20.594
20.593
20.592
20.593
20.591
20.580
20.581
Limit(dBuV)
55.410
55.151
53.965
54.736
53.604
47.980
48.403
Margin(dBuV)
-28.91
-24.15
-23.94
-23.56
-23.49
-20.12
-17.57
RBW(Hz)
9k
9k
9k
9k
9k
9k
9k
Comment
LINE
L1
Trace3: Measured Average
Frequency(Hz)
161.05 k
166.15 k
191.65 k
174.65 k
200.15 k
393.95 k
374.4 k
Level(dBuV)
26.50
31.00
30.02
31.17
30.12
27.86
30.83
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 112 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
120VAC 60Hz, Rx mode
Test Information
Test Details
Test:
Project:
Test Notes:
Temperature:
Humidity:
Tested by:
Test Started:
User Entry
LISN - FCC15 Class B
Zoll medical_01/27/2017
120VAC 60Hz_Rx mode
22 C
14% 987 mbars
Naga Suryadevara
27 Jan 2017 17 : 08
Additional Information
Prescan Emission Graph
100
27 Jan 2017 17 : 08 Test: LISN - FCC15 Class B Project: Zoll medical_01/27/2017 Tested by: Naga Suryadevara Test Notes: 120VAC 60Hz_Rx mode
90
80
70
FCC Part 15 Class B Conducted QP (Mains Ports)
60
FCC Part 15 Class B Conducted AV (Mains Ports)
50
40
30
Level (dBuV)
20
10
Trace3: Measured Average
Trace2: Measured Quasi Peak
0 Trace1: Measured Peak
150k
Frequency (Hz)
Measured
Measured
Measured
Maximum
1M
10M
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
30M
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace2: Measured Quasi Peak
Frequency(Hz)
441.55 k
193.35 k
199.3 k
172.95 k
372.7 k
154.25 k
Level(dBuV)
31.46
39.43
39.69
43.14
39.12
47.86
TF
0.025
0.045
0.041
0.062
0.028
0.077
PA+CL
20.577
20.592
20.591
20.593
20.581
20.594
Limit(dBuV)
57.033
63.891
63.640
64.818
58.441
65.768
Margin(dBuV)
-25.57
-24.46
-23.95
-21.68
-19.32
-17.91
RBW(Hz)
9k
9k
9k
9k
9k
9k
Comment
LINE
L1
TF
0.045
0.062
0.041
0.077
0.025
0.028
PA+CL
20.592
20.593
20.591
20.594
20.577
20.581
Limit(dBuV)
53.891
54.818
53.640
55.768
47.033
48.441
Margin(dBuV)
-25.76
-23.82
-23.57
-21.29
-20.28
-16.86
RBW(Hz)
9k
9k
9k
9k
9k
9k
Comment
LINE
L1
Trace3: Measured Average
Frequency(Hz)
193.35 k
172.95 k
199.3 k
154.25 k
441.55 k
372.7 k
Level(dBuV)
28.13
31.00
30.07
34.48
26.76
31.58
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 113 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
120VAC 60Hz, Tx mode
Test Information
Test Details
Test:
Project:
Test Notes:
Tested by:
Test Started:
User Entry
LISN - FCC15 Class B
Zoll medical_01/30/2017
Tx mode Cellphone with LTE
Naga Suryadevara
30 Jan 2017 17 : 00
Additional Information
Prescan Emission Graph
100
30 Jan 2017 17 : 00 Test: LISN - FCC15 Class B Project: Zoll medical_01/30/2017 Tested by: Naga Suryadevara Test Notes: Tx mode Cellphone with LTE
90
80
70
FCC Part 15 Class B Conducted QP (Mains Ports)
60
FCC Part 15 Class B Conducted AV (Mains Ports)
50
40
30
Level (dBuV)
20
10
Trace3: Measured Average
Trace2: Measured Quasi Peak
0 Trace1: Measured Peak
150k
Frequency (Hz)
Measured
Measured
Measured
Maximum
1M
10M
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
30M
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace2: Measured Quasi Peak
Frequency(Hz)
429.65 k
3.169 M
250.3 k
178.9 k
154.25 k
367.6 k
Level(dBuV)
24.29
25.00
34.83
46.12
49.49
44.95
TF
0.026
0.020
0.035
0.057
0.077
0.028
PA+CL
20.577
20.580
20.588
20.593
20.594
20.581
Limit(dBuV)
57.259
56.000
61.747
64.537
65.768
58.555
Margin(dBuV)
-32.97
-31.00
-26.92
-18.42
-16.28
-13.61
RBW(Hz)
9k
9k
9k
9k
9k
9k
Comment
LINE
L1
L1
L1
L1
TF
0.026
0.020
0.035
0.077
0.057
0.028
PA+CL
20.577
20.580
20.588
20.594
20.593
20.581
Limit(dBuV)
47.259
46.000
51.747
55.768
54.537
48.555
Margin(dBuV)
-27.29
-26.57
-26.30
-21.06
-19.57
-8.38
RBW(Hz)
9k
9k
9k
9k
9k
9k
Comment
LINE
L1
L1
L1
L1
Trace3: Measured Average
Frequency(Hz)
429.65 k
3.169 M
250.3 k
154.25 k
178.9 k
367.6 k
Level(dBuV)
19.97
19.43
25.44
34.71
34.97
40.17
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 114 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
120VAC 60Hz, Rx mode
Test Information
Test Details
Test:
Project:
Test Notes:
Tested by:
Test Started:
User Entry
LISN - FCC15 Class B
Zoll medical_01/30/2017
Rx mode Cellphone with LTE
Naga Suryadevara
30 Jan 2017 17 : 22
Additional Information
Prescan Emission Graph
100
30 Jan 2017 17 : 22 Test: LISN - FCC15 Class B Project: Zoll medical_01/30/2017 Tested by: Naga Suryadevara Test Notes: Rx mode Cellphone with LTE
90
80
70
FCC Part 15 Class B Conducted QP (Mains Ports)
60
FCC Part 15 Class B Conducted AV (Mains Ports)
50
40
30
Level (dBuV)
20
10
Trace3: Measured Average
Trace2: Measured Quasi Peak
0 Trace1: Measured Peak
150k
Frequency (Hz)
Measured
Measured
Measured
Maximum
1M
10M
Peak Value
Quasi Peak Value
Average Value
Value of Mast and Turntable
30M
__ Swept Peak Data
__ Swept Quasi Peak Data
__ Swept Average Data
Emissions Test Data
Trace2: Measured Quasi Peak
Frequency(Hz)
224.8 k
198.45 k
170.4 k
176.35 k
155.1 k
371.85 k
Level(dBuV)
37.86
40.87
45.14
46.01
49.35
45.74
TF
0.038
0.041
0.064
0.059
0.076
0.028
PA+CL
20.590
20.591
20.593
20.593
20.594
20.581
Limit(dBuV)
62.640
63.675
64.941
64.656
65.722
58.460
Margin(dBuV)
-24.78
-22.80
-19.80
-18.65
-16.37
-12.72
RBW(Hz)
9k
9k
9k
9k
9k
9k
Comment
LINE
L1
L1
TF
0.038
0.064
0.041
0.076
0.059
0.028
PA+CL
20.590
20.593
20.591
20.594
20.593
20.581
Limit(dBuV)
52.640
54.941
53.675
55.722
54.656
48.460
Margin(dBuV)
-22.73
-21.87
-21.43
-20.63
-19.54
-8.11
RBW(Hz)
9k
9k
9k
9k
9k
9k
Comment
LINE
L1
L1
Trace3: Measured Average
Frequency(Hz)
224.8 k
170.4 k
198.45 k
155.1 k
176.35 k
371.85 k
Level(dBuV)
29.91
33.07
32.24
35.10
35.12
40.35
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 115 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
Test Date:
Test Personnel:
Supervising/Reviewing
Engineer:
(Where Applicable)
Product Standard:
Input Voltage:
Pretest Verification w/
Ambient Signals or
BB Source:
Naga Suryadevara
N/A
FCC Part 15 Subpart C
FCC Part 15 Subpart B
120VAC 60Hz
Signal generator
Limit Applied:
01/27/2017
01/30/2017
Class B
Ambient Temperature:
22, 20 ºC
Relative Humidity:
14, 17 %
Atmospheric Pressure:
987, 990 mbars
Deviations, Additions, or Exclusions: None
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Page 116 of 117
Intertek
Report Number: 102716227BOX-002
Issued: 03/09/2017
15 Revision History
Revision
Date
Level
02/12/2017
02/15/2017
03/09/2017
Report Number
Prepared
By
Reviewed
By
102716227BOX-002
VFV
102716227BOX-002
VFV
102716227BOX-002
VFV
Non-Specific Radio Report Shell Rev. August 2015
Company: Zoll medical Corporation Model: µCor 3.0
Notes
Original Issue
Fixed typographical
errors
Updated section 4,
5.1, 5.2 and pages
94-97.
Page 117 of 117

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : No
Page Count                      : 117
Language                        : en-US
Tagged PDF                      : Yes
Title                           : EMC TEST REPORT
Author                          : EMC
Creator                         : Microsoft® Word 2010
Create Date                     : 2017:03:09 21:01:09-05:00
Modify Date                     : 2017:03:09 21:01:09-05:00
Producer                        : Microsoft® Word 2010
EXIF Metadata provided by EXIF.tools
FCC ID Filing: 2ABHFUCOR30