BTS BTSWEMG2 Wireless EMG sensor for EMG system User Manual ERFNC 00782 16 FREEEMG 300 ENG v 4 0 3

BTS SpA Wireless EMG sensor for EMG system ERFNC 00782 16 FREEEMG 300 ENG v 4 0 3

ERFNC-00782-16 FREEEMG 300 User Manual ENG v.4.0.3

BTS FREEEMGversion 4.0.3user manualenglish
Document Number : ERFNC-00782-16Published: June 2013 Copyright © 2010 - 2013 BTS SpA. All Rights Reserved.
1BTS BioengineeringBTS FREEEMGcontents              1icons, symbols and acronyms          5radio regulation              9disposal (WEEE)            10intended use                   11regulatory label              12  acquisition unit regulatory label        12  EMG wireless probes regulatory label      12  AUX wireless probes regulatory label      13  BTS charger regulatory label        14  BTS docking station regulatory label      14warnings                    15copyright                    18introduction                19  general  description           19   case contents            20 system components            24   physical description           24   acquisition unit            25  PDA (Personal Digital Assistant)        27  wireless EMG probes          29  wireless FSW/EGN probes (optional)      33  on-o analysis (optional)         34  electrogoniometers (optional)        35  charger              36contents
2BTS BioengineeringBTS FREEEMGcontents  docking station (optional)        38installation              43  user PC minimum conguration        43  installation of the Wi-Fi interface       43  description of the software on the user PC     45  switching on BTS FREEEMG 300      45   rst connection            47  connection check           48  correct data visualization check        50acquisition protocol            52 tting the patient            57guide to the use of onboard software        60  initial screen             60   “Active” button            60   selecting the work mode          64   description of the main bars and menu      65  protocol             68  probes setup            69  digital oscilloscope (gane)         72  digital oscilloscope (cross check)        75  footswitches            77  “Local” button: holter mode          78  “Remote” button: lab mode        91  “Download” button          95  “Cong” button           97  “Exit” button            106BTS FREEEMG 300 updating           107  updating the onboard software        107  creating a backup image          108
3BTS BioengineeringBTS FREEEMGcontentsmaintenance and cleaning          110appendix A – technical specications         113  wireless probes            113  receiving unit            113  BTS internal coding          114appendix B – environmental specications      116appendix C – power supply and switch o       117appendix D – battery             118appendix E – declaration of conformity         120 appendix F – troubleshooting guide          121  1708 error - Incorrect IP address or port set up     121  1709 error - Impossibility to connect to the unit    122  backup restoring           122  receiving unit hardware reset        124
5BTS BioengineeringBTS FREEEMGicons, symbols and acronymsSymbol in the instructions for the function. e icon represents the information which requires special attention.             Symbol in the instructions for the function. is icon makes reference to a more detailed discussion of the subject in hand.Symbol on the equipment:e data appearing next to the manufacturer’s symbol refer to the place of manufacture of the equipment itself.  Symbol on the equipment:Attention, read the information in the users’ manual carefully before using the equipment.
6BTS BioengineeringBTS FREEEMGicons, symbols and acronymsSymbol on the equipment and in the users’ instructions.Symbol for the separate disposal of electrical and electronic equipment, in accordance with Directive 2002/96/CE (WEEE). e equipment belongs to Group 8 (medical equipment). In force in the nations of the European Union, Norway and Switzerland.  Symbol on the equipment:e gure in the square indicates the insulation class and the part types used. In accordance with Standard ISO 60601-1, the equipment has an internal power supply and the parts used are type BF. REFSymbol on the equipment:Symbol located next to the model number (ref. to catalogue).SNSymbol on the equipment:Symbol located next to the series number on the equipment.
7BTS BioengineeringBTS FREEEMGicons, symbols and acronyms   Symbol on the equipment:CE mark with the code of the Notied Body. e CE mark certies that the product conforms to the standards applicable in the member states of the European Union (see Declaration of Conformity). Symbol on the equipment:CE mark with the code of the Notied Body. e CE mark certies that the product conforms to the Directive  99/05/EEC - R&TTE and obtained the Expert Opinion by IMQ. Symbol on the equipment:e double square indicates that the product is a medical device of II Class (In accordance with the law EN 60601-1). Symbol on the equipment:e “FCC” symbol refers to the Federal Communication Commission of the USA. e device complies with the relevant regulations put forth by the FCC as long as it is operated according to the instructions contained in this manual and to all national and local regulations.
8BTS BioengineeringBTS FREEEMGicons, symbols and acronymsSymbol on the equipment:e symbol indicates that U.S. Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner.Acronyms used in this manual:AP Access Point PDA Personal Digital AssistantAU Acquisition Unit RU Receiving UnitEMG Electromyography SW WorkstationFSW Footswitch
9BTS BioengineeringBTS FREEEMGRadio equipment identication:- EMG probes:FCC ID: YQH-BTSWEMG2 IC: 9188A-BTSWEMG2- FSWEGN probes:FCC ID: YQH-BTSWAUX IC: 9188A-BTSWAUX- Acquisition Unit contains:FCC ID: TFB-MATRIXLPIC: 5969A-MATRIXLPis device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) is device may not cause harmful interference,  and  (2)  this  device  must  accept  any  interference  received, including interference that may cause undesired operation.Modications not expressly approved by BTS SpA could void the user’s authority to operate the equipment under FCC rules.radio regulation
10BTS BioengineeringBTS FREEEMGIn disposing of the equipment observe the legal prescriptions. In accordance with Directive 2002/96/CE (WEEE) all equipment supplied after 13/08/2005 may not be disposed of in general domestic waste. is equipment belongs to Category 8 (medical equipment) and is classied in the Business-to-Business sector. e  symbol  of  the  crossed  out  rubbish  bin indicates  that  the  equipment  must  not  be disposed of in normal domestic waste. e regulations for disposal may dier between individual countries in the EU. In cases of doubt, refer to the respective sales outlet. is is a battery-powered equipment. See Appendix D for information about the batteries used. Operate and dispose of this equipment according to the instructions set in the “warnings” section.disposal (WEEE)
11BTS BioengineeringBTS FREEEMGis equipment is an instrument for the EMG surface analysis, classied as medical equipment in accordance with European Directive 93/42/CE (and its amendments).e BTS FREEEMG 300 must always be used only for this purpose, by qualied persons, in an environment suitable for the execution of EMG analyses and respecting the prevailing regulations in the countries in which it is being utilized.U.S. Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner.intended use
12BTS BioengineeringBTS FREEEMGAcquisition Unit Regulatory labelEMG Wireless Probes Regulatory labelLabel on the probes:regulatory labelProbe ID
13BTS BioengineeringBTS FREEEMGregulatory labelLabel not on actual probes due to size constraints:AUX Wireless Probes Regulatory labelLabel on the probes:Label not on actual probes due to size constraints:Probe ID
14BTS BioengineeringBTS FREEEMGregulatory labelBTS Charger Regulatory labelBTS Docking Station Regulatory label    3A  3A         9V       3A         9V
15BTS BioengineeringBTS FREEEMGWe  recommend  to  carry  out  any  kind  of  operation  keeping strictly to the security regulations contained in this manual. e safety of the instrument cannot be guaranteed if these conditions are not respected.BTS FreeEMG 300 is a medical device (EU Directive 93/42/CE and its amendments, including Directive 2007/47/CE) which use must be at all times be supervised by qualied and authorized personnel, according to the laws in force in the nation it is in use. e EMG probes are classied as ETSI EN 300 440 “Receiver category 3” according to Directive R&TTE 99/5/EEC.e results of the acquisitions must be assessed by people legally authorised by national law, who possess the suitable necessary knowledge of anatomy and muscular function.e instrument must be used in a medical environment, since it has a high level of sensitivity (measured voltage levels of between 1 microvolt and 6 millivolt).e uses of the device for other purposes and with methodologies dierent  from  of  those  indicated  in  this  manual  are  not  to  be considered congruent with the precise use of the device.During the preparation of the patient, take particular care that the system’s components  do  not  impede  in  any  way  the  normal  movements  of  the subject. Apply the probes only on undamaged skin.warnings
16BTS BioengineeringBTS FREEEMGwarnings•  Only  use  CE  branded  probes  and  hypoallergenic  double-sided  tape, compatible with the usage on undamaged skin for brief periods of time.•  Periodically verify the integrity of the system and of its components.•  To not wet or dip in water the parts that make up the system.No modication of this equipment is allowed.•  Only  BTS  S.p.A.  authorized  technicians  may  maintain  and  operate servicing  to  the  instrument.  BTS  S.p.A.  cannot  be  held  responsible for system safety should the instrument be opened, repairs carried out, third parties software be installed, or system components be replaced by persons other than those authorized by BTS S.p.A.•  Users cannot  change any software conguration (including  O.S. and CD writer software). •  In case the device accidentally falls, tear of the probes or other accidents always address authorized technical support.•  Use  only  the  provided  power  supply  unit  FW7363M/09  (FRIWO) or the one provided by BTS S.p.A. for supplying docking station and charger unit. If a dierent power supply unit is used, the compliance to IEC 60601-1 is not ensured.•  Only original cables must be used, otherwise BTS S.p.A. cannot assure the safety of the instrument. Should it be necessary to replace any part of the system, only original BTS S.p.A. parts may be used. In addition to the users’ instructions, the prescriptions regarding accident  prevention  and  technical  regulations  regarding occupational safety must also be complied with. e appertaining national  regulations  and  standards  of  the  country  of  use,  with regards prevention of accidents and environment, are an extension of the users’ instructions.
17BTS BioengineeringBTS FREEEMGwarnings•  Make  sure  that  the  cables  have  been  connected  correctly.  When disconnecting cables, use the connectors and not the cables themselves to unplug the connectors.•  Mains plug of external power supply unit is considered as disconnecting device.•  Avoid connecting the probes to the charger with inverted polarity with respect to that shown on the cover of the recharger – is could cause irreparable damage to them.•  For a safe use and adequate maintenance of rechargeable batteries strictly follow the instructions given in this manual. If rechargeable batteries are used in such a way that is not the one specied by BTS S.p.A. the shelf life, functionality and the integrity of the batteries is not ensured.•  ESD application  to EGN  probes, causes a loss of  link to  the device. Temporary loss of function or performance which is recoverable with operator action.BTS  FREEEMG  300  is  a  device  that  is  able  to  function COUNTINUOUSLY,  this  is  of  course  limited  by  the  battery duration  and  by  the  memory  available  for  the  acquisition  data storing.e device uses lithium ion battery. For the battery replacement and  disposal  please  contact  the  technical  support.  At  any  rate, ensure that device component  (i.e.)  probes, receiving unit,  …) integrity is never compromised.e information contained  in this manual  is  subject to change without notice and does not constitute product specications or any obligation on the part of BTS S.p.A.
18BTS BioengineeringBTS FREEEMGe software of the system described in this manual is supplied with the “licence  to  use”  contract.  e  software  may  be  used  or  copied  only  as stipulated under the terms of this contract. No part of this manual may be copied or transmitted in any form or means, electronic or mechanical, including photocopying, without prior written permission from BTS S.p.A. Unless otherwise specied, any reference to companies, names, data and addresses used  in  the  reproduction of  the  screens and  the  examples  are purely incidental, and has the sole purpose of illustrating the use of the BTS product. All trademarks are registered by the respective owners.is  publication  contains  reserved  information  which  is  the property of BTS S.p.A.e recipient acknowledges that the illustrations and information supplied in this manual shall not be made available to third parties without explicit written agreement by BTS S.p.A.copyright
19BTS BioengineeringBTS FREEEMGGeneral descriptionBTS  FREEEMG  300  is  an  indispensable  instrument  for  laboratories which  are  concerned  with  the  study  of  muscular  activity  in  the  elds of  rehabilitation,  sports  medicine,  ergonomics,  clinical  research,  and assessment of functional ability and muscular fatigue.BTS  FREEEMG  300  makes  the  selection  of  muscles,  the  setting  of duration and frequency of acquisition, amplication gain and the correct positioning of electrodes, simple and rapid.  BTS  FREEEMG  300  can  be  applied  in  research,  sports,  occupational medicine, gnatology, neurology, and orthopaedics.With  the  software  tools  available,  BTS  FREEEMG  300  becomes  an advanced  diagnostic  tool  to  evaluate  neurological  and  orthopaedic disorders,  pharmacological  therapy,  progression  of  motor  decits,  use of  ortheses,  rehabilitative  follow-up,  and  to  optimize  sports  training programmes,...BTS FREEEMG 300 is applied with EMG-Analyzer the most complete software  solution  for  analyzing  electromyographic  signals.  It  includes predened templates for evaluations in the clinical, sports, and research eld: Jump, plyometrics, walking, fatigue analysis, isokinetic, etc...BTS EMG-Analyzer also has an editor for creating elaboration protocols:  introduction
20BTS BioengineeringBTS FREEEMGintroduction thanks to an innovative object interface, that translates the biomechanical analysis language into graphical form, the user can develop quickly and eectively customized analysis protocols.BTS  FREEEMG  300  seamlessly  integrates  with  BTS  motion  analysis systems,  through the SMART (and ELITE) dedicated software.Case contents     Standard components:• Receiving Unit• Acquisition Unit• HP iPAQ hx4700 Pocket PC• Belts for attachment to patient  • Stylus (located in the Pocket PC)  • Probes Kit  • up to 16 wireless probes (identicative labels available in 4  dierent colors)    Acquisition UnitPocket PC
21BTS BioengineeringBTS FREEEMGintroduction• Probes Chargers:• up to 2 “Charger” units: each charger handles up to 8 probes and 2 FSW/EGN. e second unit is included in the standard  kit  only  for  FREEEMG  with  at  least  12  probes,  or available as an option.• e optional “Docking Station” version is also available• Accessories             • USB-PDA cable                    •AC adapter                                                               • Access Point• Set of disposable electrodes
22BTS BioengineeringBTS FREEEMGintroduction • Manuals• BTS FREEEMG 300 user        manual complete of CD             Optional components:• Footswitch Kit  • 2  FSW/EGN Wireless Probes      • 2 Connectors of 4 Switches           • 10  single Switches• Fine-Wire Adaptor• Fine-Wire Kit  • xing set for Fine-Wire adaptor  • set of disposal Fine-Wire electrodes with applicator       • Electrogoniometer Kit
23BTS BioengineeringBTS FREEEMGintroduction  • 1  FSW/EGN Wireless Probe      • 1 Electrogoniometer   • 1 Connector                 • Supplementary Receiving Unit• Secure Digital      You will receive the instructions for use for other possible op-tional components not mentioned in this manual.
24BTS BioengineeringBTS FREEEMGBTS FREEEMG 300 electromyograph consists of a receiving unit which utilizes a PocketPC platform and wireless probes designed and developed by BTS SpA.Physical descriptionactivesync connector battery door lock (in lock position to switch on BTS FREEEMG 300)12power button  (on/o)Compact Flash (CF) slot type IISecure Digital (SD) slotstylus pen1234123456wireless probessystem componentsLED - WIFI (solid blue = WiFi Active)LED - charging and notication indicators (blinking amber = charging; solid amber = full charge)VGA color displayantennaservice slotgrip1225 243121346
25BTS BioengineeringBTS FREEEMGsystem componentsAcquisition unite acquisition unit consists of:  • a digital card• a receiver card• a plastic casing • a guide for the PDA insertion.e digital card is connected to the palmtop, via the Compact Flash port, when this is made to slide in the proper guide.When  the  acquisition  unit  is  connected  to  the  palmtop,  the  card communicates, “describing” itself to the driver in operation and thereby allowing the system to self-congure and to carry out a self-diagnosis. In addition to serving as a Compact Flash interface the card allows control
26BTS BioengineeringBTS FREEEMGsystem componentsof  the  receiver  card  and  handles  acquisition  of  data  coming  from  the probes.e  receiver  card  is  comprised  of  4  elements  that  act  as  coordinators of  the  wireless  connection  network  with  the  probes,  thus  managing communications with the probes, both for activation and data acquisition.e acquisition unit is able to handle:•  up to 16 probes, used for acquisition of the activity of the same number of electromyographic signals•  up to 2 probes of 4  auxiliary  channels for switch connection, used for the basographic acquisition •  as an alternative to the EMG channels it is possible to acquire up  to  8  electrogoniometers  (each  EGN  occupies  2  analogue channels).  In particular, the following probe combinations are possible:EGN  EMG FSW0 up to 16 probes 2 probes of 4 switches1-2 up to 12 probes 2 probes of 4 switches3-4 up to 8 probes 2 probes of 4 switches5-6 up to 4 probes 1 probe of 4 switches7-8 0 0
27BTS BioengineeringBTS FREEEMGsystem componentsPDA (Personal Digital Assistant)HP  iPAQ  hx4700  Pocket  PC  is  based  on Microsoft® Windows®. It  uses  the  Intel®  PXA270  processor  and the  new  transective  display  VGATFT,  4” with  64k  colors,  to  provide  exceptional graphics  power  and  performance  to  display large  dimension  images  at  highest  quality resolution. HP  hx4700  is  tted  with  excellent  expansion  capacity  (Compact  Flash Type II and SD slot) to provide more storage space and greater function range. 64 MB of SDRAM and 128 MB of ROM memory allow a greater number of programmes and les to be stored, while the 1800 mAh lithium ion battery ensures longer uninterrupted use of the Pocket PC.
28BTS BioengineeringBTS FREEEMGsystem componentsFor further details please refer to the HP iPAQ hx4700 Pocket PC Manual included.
29BTS BioengineeringBTS FREEEMGsystem componentsWireless EMG ProbesBTS FREEEMG 300 utilizes miniaturized probes with active electrodes weighing less than 13 grams.  e special design ensures maximum space-saving and comfort for the patient who is free to move around without obstacles.e probes can be hooked on directly to the pre-gelled electrodes without requiring additional xing with plasters or double-sided tape. is together with the total absence of cables enables a much faster patient preparation, drastically reducing the time of each session.Each receiving unit can control up to 16 probes, for a total therefore of 16 analysable muscles in a single exam (standard conguration with one receiving unit).Each probe consists of a mother electrode and a satellite electrode, each tted with a clip. e two parts, connected via a exible cable, may be positioned as needed by the  user at adjustable distance (electrodes with variable geometry).In  the  mother  electrode  there  are  the  A/D  converter,  the  antenna  and the  battery.    In  the  satellite  electrode  there  is  the  Signal  Conditioning (preamplier and lter) and is reported the probe ID.All probes are also equipped with a solid state memory buer, to prevent data loss for problems due to the Wi-Fi network or due to exceeding the useful operating range.12345mother electrodeLEDsatellite electrodemother electrode clip satellite electrode clipexible cableprobe ID6751225 3467
30BTS BioengineeringBTS FREEEMGsystem componentsEach probe is tted with an LED indicating its state. e probes can be in one of a number of dierent states: •  Charge: steady blue LED. During the recharging phase the steady blue LED is on.is phase occurs when the probes are connected to the charger turned on, and the charge level is less than 90%.When the charge level reaches 90% the led turns OFF.Since, by connecting a probe to the charger on, it enters in “Deep Sleep” mode, even while charging, the probe will be completely passive and does not respond to any commands.•  Inactive: white LED which cyclically lights with a steady light for a few seconds every 3 minutes.  In the “Inactive” mode the probes cannot be used directly by the patient unit but are in constant standby to be activated by the activation software. When the probe is inactive, it carries out a scan of the radio frequencies: during the scan the white LED is lit continuously.•  Active-Scanning: white  LED  which  cyclically  lights  for  a  few seconds.  In this  mode  the probe is  searching for  the  receiving unit  on the channel assigned during the activation phase. At intervals of about 1 minutes it carries out a scan of the frequencies of few seconds. During the scan the white LED ashes quickly. •  Active-Connected: white LED which ashes slowly.  When the probe and the receiving unit establish a connection, the white LED begins to pulse slowly: the probe is waiting for
31BTS BioengineeringBTS FREEEMGsystem componentscommands.   If the connection is interrupted, the probe returns to  “Active-Scanning”  mode  and  attempts  to  re-establish  the connection. •  Active-Capturing:  white  LED  which  lights  and  goes  out  at regular intervals. During acquisition the white LED ashes at regular intervals of approximately one second. At the conclusion of the acquisition, the  probe  returns  to  the  “Active-Connected”  condition.    If during the  acquisition, connection  to the  receiver unit is  lost, the probe continues to acquire, storing the data locally for one minute and at the same time scans the assigned channel trying to reconnect to the receiving unit. If after one minute the scan is  unsuccessful,  the  probe  returns  to  the  “Active-Scanning” condition interrupting the storage of data.•  Completely discharged or in “Deep Sleep” mode: LED is o.  If the probe is completely discharged the LED does not display any ashing cycle and is o.  e same happens when it is in “Deep  Sleep”  mode  (except  during  the  recharging  phase  in witch the led is steady blue). e probes in “Deep Sleep” mode do not perform any scan cycle, but are turned o. Is therefore guaranteed energy savings.  To put the probe in “Deep Sleep” mode it is necessary to connect them  to the Charger  switched  on,  or  to put them in  contact with a magnet for half a second.
32BTS BioengineeringBTS FREEEMGsystem componentsBefore  the  next  use  is  necessary  to  reactivate  the probes, putting them in contact with a magnet. After this operation they will pass to “Inactive” mode.e probes are charged  by a dedicated charger to  which the probes are connected via their respective clips.When the probes are in charge and the charger is turned on, they are in “Deep Sleep” mode and all the LEDs are o. it is necessary to reactivate the probes, prior to use them, using a magnet.e EMG probes can be used also to acquire the EMG signal by means of Fine-Wire electrodes. With this aim, there are  available,  in  option,  adaptors  allowing for the connection with Fine-Wire electrodes. e EMG probes can be connected to them using the clips. Each adaptor can be xed to the subject by means of proper elastics.For these applications the acquisition frequency must be set on 4  KHz  (the  guidelines  recommend for Fine-Wire  electrodes an acquisition frequency major or equal to 2 KHz).
33BTS BioengineeringBTS FREEEMGsystem componentsWireless FSW/EGN Probes (optional)For  collecting  the  on-o  analysis  signals  coming  from  the  Footswitch or  for  collecting  data  from  the  Electrogoniometers  (optional  system components) BTS FREEEMG  300 uses  wireless probes which must be connected to the FSW or EGN probes using a special connector.e  probe  will  work  dierently  if  used  with  one  or  the  other probe and will receive from the same receiving unit information on its work modality during activation.e probe consists of a single parallelepiped-shaped block. e upper face has an ID tag characterized by a color (Green, Red, Yellow, Blue) and a letter (A, B, C, D, E or F) and a status LED which operates similar to that described for the EMG probes (see previous paragraph).e  two  lateral  faces  have  two  connectors;  the  one  on  the  ID  tag  side serves to charge the probe, and a cable will be connected to this to enable connection to the charger (Charger or Docking Station).e one on the other side is for the probe connections (FSW or EGN).Finally, the ID identier of the probe is on the bottom left corner of the back side.
34BTS BioengineeringBTS FREEEMGsystem componentsOn-o analysis (optional)e  footswitches  are  useful  in  dening  the  contact  points  during  the contact phases of deambulation. e footswitches consist  of  a  resistive membrane,  (FSR technology), of diameter 18 mm and thickness less than 0.5 mm, expressly designed for applications in the analysis of movement. e compact size of the instrument permits a maximum of exibility in positioning on the patient’s foot.For  applications  other  than  gait  analysis,  there  are  available  on  request smaller  diameter  (8  mm)  switch  probes  (appliable,  for  example,  to  the nger), and square (useful for tapping tests), 44 mm x 44 mm.BTS FREEEMG 300 permits up to 8 basographic zones to be measured, through 2 connectors from 4 single switches (usually right and left side) that are connected to the two FSW/EGN wireless probes. e  footswitch  channels  are  supplementary  to  the  16 electromyographic channels. Refer to § “Acquisition unit” for the list of probe combinations possible in case of acquisition also with EGN.
35BTS BioengineeringBTS FREEEMGsystem componentsElectrogoniometers (optional)e electrogoniometer is an easy to use device that allows the measurement of joint angle progress over time. ere are  primarily  two  types of electrogoniometers: the  potentiometer and  the  strain  gauge.  BTS  FREEEMG  300  uses  the  strain  gauge electrogoniometer of Biometrics LTD.  ere are single-axle models for the neck (axial rotation) and the forearm (prone-supination) and biaxial models for other main joints: wrist, elbow, knee, ankle, hip and back.e strain gauge electrogoniometers are made up of two sensors, connected to each other, that are xed to the bone segments involved in the joint to value.e measure  of the angle  is provided by the  relative angle  between the axes of the two sensors and, unlike the potentiometric electrogoniometers, it doesn’t depend on the linear slidings in which the two extremities can incur.Each  electrogoniometer  is  connected  to  a  FSW/EGN  wireless  probe using the appropriate connector. Each receiving unit can handle up to 8 electrogoniometers.
36BTS BioengineeringBTS FREEEMGsystem componentse electrogoniometer channels are not supplementary to the 16 electromyographic  channels,  but  each  electrogoniometer  used engages 2 EMG channels.  Refer to § “Acquisition unit” for the detailed list of probe combinations possible.Charger e Charger, included with the product, charges the FREEEMG probes.e Charger can simultaneously charge 8 EMG probes and 2 FSW/EGN probes. Two units are provided for systems with more than 12 probes.More  units  can  be  connected  in  series  for  simultaneous  power  supply through the same AC adapter, using the cable included.e EMG probes are connected to the charger using the same clips that normally collect the EMG signal, while the FSW/EGN probes connect to the charger using the special connector as shown in the gure below:
37BTS BioengineeringBTS FREEEMGsystem componentsFor correct connection simply follow the outline of the probes shown on the base of the Charger:e  Charger  comes  with  an  output  short  circuit  protection system also in case of reversed recharging poles. At any rate, poles connected incorrectly will not recharge.To recharge connect all the probes that you would like to charge to the Charger (follow the instructions  described above) and connect the AC/DC adaptor to the mains and turn on the switch located on the rear panel.When the Charger is properly connected to the mains and has been turned on, the status LED “Power” will show a steady GREEN light.e charging status of the EMG probes and of the FSW/EGN probes is indicated by the respective status LED (see § “Wireless EMG Probes” and “Wireless FSW/EGN Probes”).
38BTS BioengineeringBTS FREEEMGsystem componentsNote that inserting the EMG probes into the Charger when it is on, these come in “Deep Sleep” mode. It is necessary to reactivate these  probes,  prior  to  use  them,  using  a  magnet.  After  this operation they will pass to “Inactive” mode. While, connecting the FSW/EGN probes to the Charger (connected to the mains and  turned  on) if they are  charged  less  than  90%  they will be automatically reset and will pass to the  “Inactive” mode. To proceed with another acquisitions it is necessary to repeat the activation procedure.Docking Station (optional)As in alternative to the Charger an optional Docking Station is available. is unit has additional features that are described below.Each Docking Station can simultaneously charge 8 EMG probes and 2 FSW/EGN probes.Unlike the standard Charger, each Docking Station will have to be powered separately; therefore two units cannot be connected in series for contemporary power supply.
39BTS BioengineeringBTS FREEEMGsystem componentse EMG probes are connected to the charger using the clips for collecting the electromyographic signal, while the FSW/EGN probes connect to the charger using the special connector. For correct positioning simply follow the outline shown on the base of the Docking Station as shown in the following gure: e Docking Station comes with an output short circuit protection system also in case of reversed recharging poles. At any rate, poles connected incorrectly will not recharge.e  Docking  Station  also  allows  for  recharging  the  receiving  unit  by connecting it with the relevant cable (USB/PDA cable) in the USB port located on the rear panel.It  should  be  noted  that  you  can  use  the  Docking  Station  to recharge the PDA only when by pressing the power button the palmtop turns on. Otherwise you will need to recharge the PDA with the supplied HP  by  connecting  it  to  the  network.  In  order  to  go  back  and recharge  using  the  Docking  Station,  simply  charge  with  the power supply unit for about one minute, and then check that it is
40BTS BioengineeringBTS FREEEMGsystem componentssuciently charged by unplugging it from the power supply unit and turning it o and then back on.ere are two ways to recharge the probes:•  using the medical power supply unit from the network •  using the internal batterye rst method involves the connection of the Docking Station by means of the AC/DC adaptor to the mains.  In this case not only the connected probes will be charged but also the internal battery of the unit itself.e  second  method  lets  you  recharge  the  probes  and/or  the  receiving unit when you are not at the laboratory and  a mains power network is unavailable.  In  particular  with  the  battery  fully  charged  you  can  fully recharge all the probes and receiving unit or fully charge the probes only twice.To recharge a probe simply connect it as described in the method above.Connection to the Docking  Station resets the probe if charged less than 90%, to use it again for acquisitions you must repeat the activation procedure.e Docking System is equipped with an “Activate” button and 3 status LEDs: “Charge”, “Battery” and “Power”.
41BTS BioengineeringBTS FREEEMGsystem componentse  “Activate”  button  allows  immediate  activation  of  the  FSW/EGN probes connected to the Docking Station and selected to activate in the receiving unit (see § “Active” Button); you  do not need to wait for the software cycle of the “inactive” state probe to be complete. By pressing the button the FSW/EGN probe will immediately scan the radio frequencies and enter immediately in communication with the receiving unit.e “Charge” LED gives information on the probes charging and consists of a BLUE light which signals the following:•  BLUE LED “breathing”: indicates that the probes are charging. Two situations can be veried in this case:•  Status LED of the probe on: e probe is charged less than 90%•  Status LED of the probe o: e probe is more than 90% charged but still not complete.It is important to remember that the LED of the probe stays on until it is 90% charged. is function allows you to have an indication on the full charge of the probe, therefore increasing the autonomy of the probe by about 10%.•  BLUE LED o: indicates that all probes are fully recharged. •  BLUE LED x on: indicates that the “activate” button has been pressed. In this case the probes are not powered by the Docking Station but by the probe’s battery.e  “Battery”  LED  is  an  indicator  of  the  state  of  the  auxiliary  battery and consists of a two-tone RED- GREEN LED, whose signals have the
42BTS BioengineeringBTS FREEEMGsystem componentsfollowing signicance:•  LED o (both the RED light and the GREEN light are o): in this case the battery is in sleep mode or has been removed from the Docking Station•  GREEN light of LED on: the auxiliary battery is charging.•  Steady  RED  light  of  LED  on:  the  auxiliary  battery  is  fully charged.•  Blinking RED light of LED on: indicates battery malfunction or a battery charging problem.e “Power” LED is on (GREEN light) when the switch located on the rear panel of the Docking Station is ON and it is connected to the mains or, if this is not the case, the auxiliary battery is suciently charged.e Docking Station has a lid, which opens automatically by means of a spring-  brake mechanism,  which works  by pressing close  to the  word “Push”, located on the left shoulder of the unit.e probes should be put back in the Docking Station when not in use for acquisition sessions. Moreover, if left for a long period, we recommend turning of the unit by moving the switch located on the rear panel of the Docking Station to OFF. Close the lid for better protection of the probes.
43BTS BioengineeringBTS FREEEMGUser PC minimum conguration Operating system    Windows XP        Microsoft.NET framework 1.1Processor      P IV 1600 MHzRAM        256 MBVideo card      32 MB, Open GLDisk space      100 MB for the application,         not including storage for acquired dataUSB        2.0Installation of the Wi-Fi Interfacee real time data transmission between the BTS FREEEMG 300 and the Workstation occurs thanks to the Wi-Fi network. is type of connection is permitted using an Access Point supplied with the system.An AP is an Ethernet device which works with devices present in a small area. e AP, starting point of the service, operates from bridges between the wireless devices and the systems physically connected to the network with an RJ45 connector. e AP transmits the signals to the surrounding area via radio waves. All the wireless devices within the area can receive signals and communicate with  the  AP,  and  through  them,  with  all  the  other  devices,  creating  a network without cables. installation
44BTS BioengineeringBTS FREEEMGinstallationConnecting a Workstation (PC laptop or desktop), not provided with a Wi-Fi interface, to Access Point with a normal network cable is created through the wireless connection with BTS FREEEMG 300.e  access  point  included  in  the  system  is  already  congured  with  the network  name  (SSID)  “BTSFREEEMG300”  with  transmission  on channel 6. e base conguration prepared in BTS (factory settings) uses the following IP addresses:  Access Point     192.168.1.1   BTS FREEEMG 300   192.168.1.2  Workstation    assigned automatically         by the Access PointYou should always use the AP included even if the Workstation is equipped with a Wi-Fi interface.Receiving UnitWorkstation Access Point Wireless Probes
45BTS BioengineeringBTS FREEEMGinstallationDescription of the software on the user PCBTS  EMG-Analyzer  is  the  complete  and  highly  exible  solution  for making advanced elaborations of electromyographic signals and angular measurements of body segments. Includes  predened  templates  for  evaluations  in  clinics,  sports,  and research and an editor to develop customized elaboration protocols.However BTS FREEEMG 300 is manageable by all BTS applications of  SMART family dedicated to the motion analysis with which is possible to  acquire  and  process  the  electromyography  signals  in  a  integrate  and synchronous way with the other devices connected.e SMART-Analyzer software allows to perform advanced elaborations of the EMG signals, integrating the electromyography information with the cinematic and kinetic data. It  is  possible  to  construct  customized  analysis  protocols  complete  of advanced multimedial report.Switching on BTS FREEEMG 300e  patients  unit  consists  of  two  main  components:  the  PDA  and  the Acquisition Unit and is usually given with the two parts already connected. Anyway, to connect the Acquisition Unit to the palmtop, run the PDA along  the  AU  guides,  being  particularly  careful  that  the  CF  enters  the For more details concerning the use of this software, please refer to its specic manual.
46BTS BioengineeringBTS FREEEMGinstallationappropriate slot without being forced or receives any sudden impact, and avoid any excessive pressure on the palmtop screen.To avoid damaging the Compact Flash we suggest only to remove the PDA from the AU only if necessary and in any case, even if expressly requested, it is suggested not to move it completely from its place, but to let it slip just enough to disconnect it from the digital card (more or less 1 cm) event that will be signalled by an audio beep. e recharge of the battery and the transfer of data with the USB can also be done through the specic cable, without disconnecting the AU or the PDA. It is also advisable to perform a connecting/separating operation of these parts only when the device is turned o.  Check the connection between PDA and AU. Switch on the BTS FREEEMG 300: check that the slot switch, located on the side of the BTS FREEEMG 300 shown in the diagram, is in the position closest to the closed lock symbol. If not, move it to this position.
47BTS BioengineeringBTS FREEEMGinstallationSwitch  on  by  actuating  the  button  on  the  opposite  side  of  the  BTS FREEEMG 300, as shown here:At this point wait until the application BTS FREEEMG 300 is completely charged. On the BTS FREEEMG 300 display the following screen will be visualized:First Connectione  BTS FREEEMG  300  is  supplied  with  the  onboard  PDA  software preinstalled and congured. In any case for the rst connection we suggest to perform some verications.
48BTS BioengineeringBTS FREEEMGinstallationConnection checke user PC communicates with the BTS FREEEMG 300 by means of Wi-Fi protocol through the Access Point (see § “Installation of the Wi-Fi Interface”). Verify  that  the  local  net  resources  are  laid  out,  like  assumed    in  the conguration of default according to BTS, the PC user IP address must assigned automatically by the Access Point.Verify the software conguration installed on the PC.  To  do  so,  launch  the  program  that  you  want  to  congure,  i.e.  EMG-Analyzer, and on  the  menu  bar at the  voice “Laboratory”  select “Select Emg Device” and verify that the check is on the voice “FREEEMG”.en, again from the menu bar at the voice “Laboratory”, select “Set Emg Device” and verify that the check is on “EMG device”:
49BTS BioengineeringBTS FREEEMGinstallationVerify that the IP address and the port are set with the following connection parameters:    IP Address:  192.168.1.2    Port:     8000Moreover, if there is a second BTS FREEEMG 300 receiving unit available (mind that in this case it is possible to manage up to 8 channels with each unit), select the option “Activate Second Unit (only for BTS FREEEMG)” and check that the IP address and the port for the second device are set up with the following connection parameters:    IP Address:  192.168.1.3    Port:     9000If the values do not correspond change them and click on  “Update” before closing the window. Switch-on the BTS FREEEMG 300 (see § “switching on BTS FREEEMG 300”),  wait  until  the  application  of  the  acquisition  management  is completely loaded and select the remote application clicking on “Remote”. If the Wi-Fi connection between the Receiving Unit and the Workstation is assigned in the correct way, after clicking on the “Remote” button, the connection led of the Wi-Fi net, that you can nd on the front of the BTS FREEEMG 300 and that presents the Wi-Fi logo (     ), will result on (solid blue).
50BTS BioengineeringBTS FREEEMGinstallationDuring the start-up of the Remote mode dierent wireless nets can be found present in the working area. Always use the “BTSFREEEMG 300” net ignoring the other nets.From the Workstation launch the BTS EMG-Analyzer application.Wait until, in the application being used, the respective LED indicating the  state  of  the  Wi-Fi  connection  between  the  receiving  unit  and  the workstation changes from red to ashing green. Correct data visualization checkDuring the initial connection, check the correct operation of the probes in use, and the correct display of data on the BTS FREEEMG 300.To do this, activate all the EMG probes and, if the system is also provided with footswitches, two FSW/EGN probes, enabled as FSW, to be tested all previously charged, as indicated in § ““Active” Button”. When all the probes are shown to be active on the display, select the Remote mode by clicking  on  the  “Remote”  button.  Select  the  default  protocol  and  then “Setup Probes” from the menu and wait for all the activated probes to be acknowledged (the respective box must be green). To verify the correct functioning of the EMG probes, position a pair of electrodes (see § “Fitting the Patient”) and connect Probe 1. Next, select “Gains” from the menu and check the progress of the EMG1 signal. Repeat the procedure connecting the other probes one at a time.  After  checking  the  electromyograph  channels,  if  the  system  is  also
51BTS BioengineeringBTS FREEEMGinstallationprovided  with  footswitches,  attach  the  connectors  with  the  4  switches already inserted to the 2 corresponding FSW/EGN probes. After selecting from  the  menu  the  item  “Footswitch”  (as  shown  later  in  §  “Guide  to using the software on the BTS FREEEMG 300”), check each contact is functioning  correctly  by  pressing  rmly  each  switch  and  observing  the corresponding red LED light on the display of the BTS FREEEMG 300 (see § “Footswitch”). If  your  system  has  one  or  more  electrogoniometers,  repeat  the  probe activation  procedures  deactivating  all  the  EMG  probes  and  enabling as  EGN  a  number  of  FSW/EGN  probes  equal  to  the  number  of electrogoniometers included (maximum 8), refer to  § “Active” Button”. When  all  the  probes  required  are  active  go  back  to  Remote  mode  by clicking  on  the  “Remote”  button.  From  the  Workstation  prepare  an acquisition protocol that  entails  the use of  a  number of EGN  equal  to those provided with your system and transfer it on the receiving unit (see §  “Create a new protocol”). Selection the newly transferred protocol, select “Probes setup” on the menu and assign each channel to an EGN probe, checking that they are all recognized (the respective box must be green)   (see § “Probes Setup”).Finally,  connect  each  electrogoniometer  to  a  connector  and  then  to  a FSW/EGN probe. Select “Gains” from the menu and check the related signal to the EGN channel that is testing.  Repeat the procedure for each electrogoniometer provided.
52BTS BioengineeringBTS FREEEMGNow we will describe a basic procedure about how to create a new protocol using BTS EMG-Analyzer software. Refer to the EMG-Analyzer manual for further details about the “Protocol Builder” function.e  procedure  may  be  dierent  if  performed  with  other applications.  In  this  case,  refer  to  the  specic  manual  of  the software you are using.Please note that with BTS FREEEMG 300 it is possible to acquire:- up to 16  analog channels: electromyographic signals - EMG (each EMG  employs  1  analog  channel),  up  to  16  electrogoniometer signals - EGN (each EGN employs 2 analog channels) - 8 on-o analysis areas  - FSW.Refer to § “Acquisition unit” for the detailed list of possible probe combinations.Launch the software EMG-Analyzer and wait for the led of communication with the receiving unit to be green.Put BTS FREEEMG 300 in Lab Modality by clicking from the main page on the “Remote” button; if two receiving units are being used, make sure that both of them have the Remove Mode activated.acquisition protocol
53BTS BioengineeringBTS FREEEMGacquisition protocolTo create a new protocol, from the menu voice “Laboratory” of  EMG-Analyzer, select “Create Emg Protocol” .A window will open, which allows to create or modify protocols.To create a new protocol follow this procedure:- Click on “New”. e elds to be lled for the protocol creation will be enabled:
54BTS BioengineeringBTS FREEEMGacquisition protocol-select the anatomical map containing the muscle you want to add to the protocol and click on the little yellow square that identied it; immediately the rst free row available in the protocol table will be lled with the info related to the selected muscle.- repeat this operation for all the muscles to be acquired.-If  you  want  to  acquire  Footswitches,  select  the  “District” anatomical  map  and  proceed  insert  the  8  contact  point  in  the protocol by selecting the corresponding square in the anatomical map:-Finally, if you want to acquire electrogoniometers check the rst free line in the protocol table, click on the “Type” eld and select EGN among the voices available in the drop box menu:
55BTS BioengineeringBTS FREEEMGacquisition protocolIn the same way, click on the “Description” eld and select among the drop box menu voices the angle you want to acquire with the EGN, then on the “Side” eld to select the body side to which the data refers.- indicate a name for the protocol specifying it on the “Name” box-it is possible to insert also a short description of the protocol using the appropriate box  “Description”When every signal of the protocol is inserted,  proceed saving the protocol, following these instructions:- make sure that the BTS FREEEMG 300 is in the Lab Modality - click on “Save and Transfer”: the protocol will be both saved on the PC and transferred on to the BTS FREEEMG 300. If two receiving units are being used, it is necessary to wait for the protocol to  be transferred  to both units.  At the end  of the transfer, the protocol will be visible in both, but in each unit will be displayed only the respective signals from the channels they are
56BTS BioengineeringBTS FREEEMGacquisition protocolconnected to. Note that If you are working with two receiving units the rst 8 channels and the FSW will always be assigned to the rst unit, while the other, starting from the 9 to the 16th, will be loaded on the second unit. For example, if a protocol with 10 EMG channels and the footswitches has been created, the protocol on the rst unit will register 8 EMG channels and the FSWs, while the protocol on the second unit will register only 2 EMG channels.
57BTS BioengineeringBTS FREEEMGacquisition protocolHere some guidelines to follow for a correct application of the probes.Carefully prepare the skin of the subject in the area of the muscle in which the pre-gelled electrodes will be positioned.Note  that  the  condition  of  the  skin  where  the  electrodes  are applied inuences the quality of the acquired signal. erefore  you  should  pay  careful  attention  to  this  phase  in preparing the subject.Attach the probes,  previously activated  (see § ““Active” Button), to the electrodes using the appropriate clips.Find  the  position  in  which  to  place  the  electrode  on  each  muscle  that you want to analyze and apply the probe in the direction of muscle bers through the pre-gelled already applied to the probe, trying to keep the two electrodes as close as possible (tangent to each other).anks to the special hardware architecture of the probes and the much reduced weight, normally no extra xing is required.If the analysis protocol being used requires it, proceed to the positioning of the footswitches to the respective contact areas. When FSWs are used, these are xed to the skin of the subject used tape tting the patient
58BTS BioengineeringBTS FREEEMGacquisition protocolor hyper-allergenic bi-adhesive tape. Never  apply  the  tape  on  the  black  membrane  of  the  sensor, removing the tape could cause the detachment of the membrane making the switch unusable. After positioning the sensors, connect them by means of the appropriate connector to the corresponding preactivated probe. Finally,  attach  the  probe  and  the  switch  box  with  the  hypo-allergenic adhesive tape close to the same switches (i.e. close to the ankle when the sensor have been placed on the foot sole) at all times taking care to leave the patient the greatest degree of freedom of movement. Finally,  If  the  analysis  protocol  being  used  requires  it,  proceed  to  the positioning of the Electrogoniometers. Each EGN should be positioned  straddling the joint with the  two sti ends positioned on the sagittal plane, parallel to the axes of the two body segments that make up the joint under investigation, securing them to the patient’s skin using the plaster or hypoallergenic double-sided tape.After positioning the sensors, connect them by means of the appropriate connector to the corresponding preactivated probe.
59BTS BioengineeringBTS FREEEMGacquisition protocolFinally,  attach  the  probe  and  the  switch  box  with  the  hypo-allergenic adhesive tape close to the sensor on the side where the connection cable comes out.roughout the dressing procedure of the patient or at the end of it you can use the receiving unit to check, using the oscilloscope of the “Gains” function of the RU, the correct placement of the electrodes and control that the acquisition gain of the EMG channels is properly set.It’s possible also to compare two channels at a time using the “Cross check” function (see § “Guide to the use of the onboard software”).After making all the checks, it is possible to proceed with the acquisition session.
60BTS BioengineeringBTS FREEEMGInitial screenAt  the  start-up  of  the  BTS  FREEEMG  300  you  see  the  initial  screen comprised  of  six  buttons:  “Local”,  “Remote”,  “Download”,  “Cong”, “Activate” and “Exit”. In the lower right hand corner of the main screen there is the version num-ber of the software installed on board.“Active” buttone “Active”  button allows only  those probes to be activated which are to be used during the acquisition and to assign each of these to a specic channel.Click on this button to call up the following display: guide to the use of onboard software
61BTS BioengineeringBTS FREEEMGguide to the use of onboard software e  probes  are  divided  by  type:  EMG,  up  to  16  probes  for  collecting electromyographic signals, FSW, up to 8 probes for collecting basographic signals  (up  to  2  can  be  selected  at  the  same  time),  and  EGN,  up  to  8 probes for collecting signals from the electrogoniometers. A number (or  letter)  and the color  corresponding to  the  ID tag of  the probe is shown next to each probe. By clicking  on Options and  then “Show probe ID”  you can pass  from viewing by label to that for viewing by “ProbeID” identication code, the code shown on the clip of each probe and vice versa:
62BTS BioengineeringBTS FREEEMGguide to the use of onboard software To activate, select the probe that you would like to use for the acquisition by enabling the check mark in the appropriate box:It is important to remember that only certain probe congurations can be activated (for the detailed list of possible congurations, see § “acquisition unit”). is  procedure  will  be  guided  by  the  software  disabling,  upon reaching  a  conguration  limit,  all  the  checkboxes  that  can  no longer  be  selected,  furthermore,  in  the  notication  area  at  the bottom, any time the maximum conguration for a type of probe (EMG, EGN, or FSW) is reached, a warning message will appear as shown in the following gure:
63BTS BioengineeringBTS FREEEMGguide to the use of onboard software Before proceeding to activation, connect the probes to the Charger (or to the Docking Station) included, use two units if required.  is operation resets the  probes, deactivating any  probes already active, except  for  those  probes  that  are  more  than  90%  charged  (in  case  of Charger to reset these probes you must wait for the battery level to below the threshold of 90%). It is advisable to connect all the probes and not just those used during acquisition to ensure that you reset any previously activated probes, which will not be used.An  already  active  probe  can  be  reused  and  does  not  require reactivation.  In  this  case  it  must  be  selected  in  the  activation program, but do not connect it to the Charger or Docking Station so it doesn’t reset.It is in any case necessary that the probes are suciently charged for the activation to succeed.Click Start to initiate the probe activation procedure.Now we would like to distinguish between operations to perform based on if you use the Charger or Docking Station:- Charger procedures:e probes will be activated only when removed from the Charger.Remove the probes, therefore, from the charger, and wait for the BTS FREEEMG 300 to activate all the selected probes.
64BTS BioengineeringBTS FREEEMGguide to the use of onboard software To facilitate the activation and to verify that all the required probes are active, we suggest detaching the probes from the charger one at a time.- Docking Station procedures:Click on the Docking Station “Activate” button. Immediately  all  the  probes  selected  in  the  activation  probe  will  be automatically activated and all the other probes will be reset, regardless of the charge level.When a probe has been recognized and activated, it will be highlight by a green rectangle, as shown in the following gure:When  all  the  selected  EMG,  EGN  and  FSW  probes  are  indicated  as activated, click “Stop” to terminate the activation procedure. Finally, click “Exit” to return to the initial screen.Selecting the work modee rst two buttons (Red and Yellow) allows the user to select the desired
65BTS BioengineeringBTS FREEEMGguide to the use of onboard software working mode: button “Local” Holter mode, button “Remote” Lab mode.Clicking on the button for the desired mode, you can enter the software of relative management. In particular by clicking on the “Local” and “Remote” buttons you will see the following screens:Description of the main bars and menu All of the BTS FREEEMG 300 pages have a button bar on the right and a status bar at the top.Button Bar Local Acquisition (Holter Mode)Remote Acquisition (Lab Mode)
66BTS BioengineeringBTS FREEEMGguide to the use of onboard software e side arrows (Toggles) can be used to change the active eld.e up and down buttons make it possible to:•  change the value in the active eld • navigate around the main menue menu button opens the main menuStatus Bar e status bar displays:• active page • the condition of the 4 transmitters indicated with A, B, C, D. One transmitter can serve more than one probe. When a probe is  transmitting  the  LED  of  its  corresponding  transmitter  turns steady green• e status of the FSWs (steady green LED if at least 1 FSW probe is active and recognized by the system)• battery levelMain MenuClicking on the “menu”  button of the buttons  bar  you enter the  main menu.ere are two menus: one for the Holter mode and one for the Lab mode. e two are dierent for the items: “Database” and “PlayBack” which exist only in the Holter mode.
67BTS BioengineeringBTS FREEEMGguide to the use of onboard software e main menu allows you to:- explore the functions of the BTS FREEEMG 300. To  enter a page select  the  corresponding  icon by clicking  with the stylus on the desired function or by using the Up and Down Buttons.- lock up the palm by clicking with the stylus on the icon “Lock”.Accomplishing this function will result in the following screen:    To  unlock  the  palm,  clik  on  the  “Unlock”  button,  this  will Lab Mode Menu Holter Mode Menu
68BTS BioengineeringBTS FREEEMGguide to the use of onboard software immediately move to another position.   To complete this procedure it is necessary to press it again before it goes back to the original position.- close the application by clicking on the “Exit” button.e following describes the menu functions starting from the ones that you can nd both in the Holter Mode and in the Lab Mode that have the same characteristics.  Protocolis  item  allows the  user  to  select  a  protocol  from the ones  previously created by the user  (see § “Acquisition protocol”). In any case if a protocol hasn’t yet been created it is possible to use the default protocol.Selecting a protocol is achieved by  scrolling through the list on the left panel by clicking on the Up and Down buttons of the buttons bar. e protocol you select will be highlighted in blue.
69BTS BioengineeringBTS FREEEMGguide to the use of onboard software Probes Setup When a new protocol is dened, the user needs to associate with the items of the protocol (muscles, joints and foot areas) the channel that physically acquires the data.By  clicking  “Probes  setup”  from  the  menu  you  will  see  the  following screens:In particular we can identify two zones:  • the probe-protocol association table• e “Probes” areae table, for each signal of the acquisition protocol, previously selected in the “protocol” section shows:• the  identier  of  the  assigned  probe  (comprised  of  the  color  and number/letter of the label), if no probe has been assigned the eld is set on “--” • the type of probe: Surface, Footswitch, Electrogoniometer,...• the state of the probe:• OK:  Signal/probe  properly  assigned,  the  probe  has  been
70BTS BioengineeringBTS FREEEMGguide to the use of onboard software selected in the activator and is connected to the RU.• NOT CONNECTED: Signal/probe properly assigned, the probe has been selected in the activator, but does not appear to be connected to the RU.• “--” : no assigned probe.• NOT  ACTIVATED:  the  probe  is  assigned  to the protocol signal, but has not been selected in the activator.• ERROR: the signal/probe has not been properly assigned as the type of probe expected by the protocol does not match the actual probeIn the “Probes” area (rectangle at the bottom) where the selected probes in the activator are listed and the state is described. In particular, the boxes associated to channels can have dierent colors:• Green: the  selected  probe  in  the  activator  is  connected  to  the  RU  and  is assigned  to  a  protocol  signal    (the  “Status”  eld  of  the  protocol  is “OK”)• Grey: the selected probe in the activator is connected to the RU but is not assigned to any protocol signal• Orange: the selected probe  in the activator  is not connected  to the RU but is assigned to a protocol signal (the “Status” eld of the protocol is “NOT CONNECTED”)• White: the selected probe in the activator is not connected to the RU and has not been assigned to any protocol signal
71BTS BioengineeringBTS FREEEMGguide to the use of onboard software • Red: the selected probe in the activator is connected to the RU and assigned to a protocol signal, but the type of probe expected by the protocol does not match the actual probe (the “Status” eld of the protocol is “ERROR”).To make the association probe-signal:1.  use Up and Down buttons to select a signal2.  use toggle to select the “Probe” eld3.  choose  among  the  available  ones  (all  the  probes  present  in the  probe  area  will  be  proposed),  the  probe  to  pair  to  the highlighted signal4.  use toggle to select the description (the whole row)5.  move on the next description with the arrow Down and restart from point 2 to assign the channel to the next item.Active elds (Toggle) Value range (Select)Description Signals ListProbe 1, 2, ..., 15, 16, for EMG channelsA.1, A.2, A.3, A.4, B.1, B.2, B.3, B.4, C.1, C.2, C.3, C.4, D.1, D.2, D.3, D.4, E.1, E.2, E.3, E.4, F.1, F.2, F.3, F.4, for EGN and FSWall available in 4 colors: Green, Red, Blue, Yellow
72BTS BioengineeringBTS FREEEMGguide to the use of onboard software Each Electrogoniometer generates two signals, it is important that you assign two channels of the same probe to the signals generated by the same EGN, e.g. C.1, C.2To  properly  assign  the  basographic  signals  to  the  FSW  probe channels please remember to match the 4 signals related to one probe to the protocol signals related to the physically close contact area, e.g. A.1, A.2, A,3, A.4  for the Right foot and  B.1, B.2, B.3, B.4 for the Left foot.Before  starting  acquisition,  the  user  needs  to  conrm  that  the “Status” of each protocol voice is OK.Digital Oscilloscope (gain)With BTS FREEEMG 300 the user can set up to 3 dierent ranges of values to see the muscular contractions better. To setting this value see § “ Cong” Button”.  e selectable values are:  • S1: 6mV;• S2: 3mV;• S3: 1,5mV.By  accessing  the  gains  page  it  is  possible  to  verify  that  electrodes  have been well positioned (no noise or motion artefacts) and that the signals are properly displayed and if they are not you can change the range of value
73BTS BioengineeringBTS FREEEMGguide to the use of onboard software set in the conguration section.To do this, simply view via the oscilloscope the real-time muscular activity of each muscle you would  like to analyze for a few seconds, asking the patient to perform several contractions.If the intensity of the muscle is very low, so that it is dicult to distinguish between when the muscle is in contraction from when it is at rest, the range needs to be reduced (the system will increase the gain correspondingly). If, on the other hand, the signal in the current amplitude tends to saturate (meaning  during the contraction there are peaks in the orange zone or beyond), it is necessary to increase the range (the system will reduce the signal gain accordingly). Keep in mind that the base range value will be the same for all signals acquired. erefore  select  the  value  that  lets  you  best  see  the  channels simultaneously without saturation. If  the  selected  range  is  insucient  to  contain  the  entire  signal dynamic,  the  acquisition  data  will  tend  to  saturate  during the  contraction,  which  will  appear  as  unnatural  plateaus  in measurement values at the limits of the selected range.On the other hand, if the signal is very weak, maintaining a high range is not a mistake. It is recommended, however, that the range be reduced whenever possible, in order to make better use of the digitalization of the acquisition card.
74BTS BioengineeringBTS FREEEMGguide to the use of onboard software e gain used in the acquisition of the EMG signals is managed by the application on the basis of the range selected, the choice of range should be done with great care since it aects not only the display but also the acquisition signal (it is not a just a simple zoom!).By clicking the  voice “Gain” from the menu you will see  the following screens:To scroll the dierent signals, click on the Toggle to activate the “Signal” active eld and select the protocol item desired by actuating the Up and Down buttons.To  make  the  signal  visualization  easier  it  is  possible  to  use  the  “Time” active eld (by selecting it with the Toggle) this allows to adjust the speed at which the signals run on the screen  (this is only for display purposes, and will have no eect on the capture). It  is  possible  to  set  the  time  value  interval  shown  on  the  axes  of  the oscilloscope  abscissas,  selecting  from  a  range  of  values  from  1  to  10 seconds,   that can be selected with the Up and Down buttons.
75BTS BioengineeringBTS FREEEMGguide to the use of onboard software Active elds (Toggle) Value range (Select)Signal Protocol signalsTime 1 sec - 10 secDigital Oscilloscope (cross check)e Cross Check  is very useful during the patient setup phase, allows the operator to verify, prior to the actual acquisition of the signals, that the electrodes have been positioned correctly. e  possibility  of  displaying  two  traces  simultaneously  prevents  the problem of cross-talk: this phenomenon occurs as a detection of electrical activity on the trace of a muscle  during a movement which should not involve the muscle in question. Cross-talk is occurring if, for example, during the movement of extension of the wrist, one observes the activity of the ulnar carpus exor.
76BTS BioengineeringBTS FREEEMGguide to the use of onboard software is is due to an incorrect positioning of the electrodes, as a consequence of which the channel dedicated to a certain muscle is in reality measuring the activity of another muscle. In  the  above  example,  the  activity  measured  on  the  trace  of  the  ulnar carpus exor is not in reality being produced by that muscle, but by one deeper down or by its antagonist.To appoint the signals to visualize select the “Signal 1” and then “Signal 2” eld with the Toggle and scroll the signal list included in the protocol using the Up and Down buttons. e current display can be enlarged or reduced via the zoom function: this does not aect in any way the data being read. e “Time” eld can be used to vary the speed at which the signal runs on the screen (again, this has no eect on the reading, but only on the display).  It is possible to x the time intervals represented on the axis of the abscissas of the oscilloscope, choosing from a range of values from 1 to 10 seconds. Active elds (Toggle) Value range (Select)Signal 1 Protocol signalsSignal 2 Protocol signalsZoom From 1/10 to x10Time From 1 sec  to 10 sec
77BTS BioengineeringBTS FREEEMGguide to the use of onboard software Footswitchese  Footswitches page shows  in  real time the activation  of  the  contact zone during the deambulation contact phase. When a zone is active, the foot is resting on it - the corresponding circle colors is red.If there isn’t correspondence between the contact points indicated on  the  display  and  the  movements  done  by  the  subject  (for example if with the right foot in stance the areas under the left foot of the display became red) we suggest you to review the FSW channels setup selecting the menu voice “Probes Setup”.Usually the Footswitches are used for studying the sole of the foot during deambulation.Four points have been identied for this purpose in order to be able to distinguish the foot tread while walking.ese 4 areas are:
78BTS BioengineeringBTS FREEEMGguide to the use of onboard software • Internal Contact, • External Contact, • Heel Contact, • Toe Contact. However,  while  preparing  the  protocol  you  can  indicate  the  general contact areas: • “Contact area 1”, • “Contact area 2”, • “Contact area 3”,• ...e Footswitch probes can also be used for other applications and placed on other parts of the body where you would like to record the moment of contact (for example on a nger during pointing exercises). In this case you should use a general “Contact Area 1”, “Contact Area 2” in the protocol, and in the FSW page refer to the rectangles located under the gure showing the foot to check that the probes are properly activated.“Local” button: Holter ModeIn this mode the RU is completely independent and it’s possible to acquire the data independently of the WS.Session preparation:selecting “Local” you can enter the patient’s database on the palm.
79BTS BioengineeringBTS FREEEMGguide to the use of onboard software To add a new patient click with the stylus on “New subject” button.A window will open which allows you to insert general information about the patient.To insert the patient data: • move up to the eld to insert• insert the value with the keypad, (each tap on the gender eld  will change from F to M and viceversa)• press OK to insert the patient into the databaseList of patients already included in the database  of the BTS FREEEMG 300List of sessions for the highlighted patientList of the tests of the highlighted session
80BTS BioengineeringBTS FREEEMGguide to the use of onboard software It is possible to modify or to delete the data of a patient already inserted. By clicking on the line of that patient, it will be underlined in light blue and the “Delete” and “Modify” buttons will appear.Clicking  on  “Modify”  a  window  will  open  that  allows  to  modify  the general information of this patient. Click on “Delete” and immediately a window with the request to conrm the deletion will appear:Clicking  on “Yes” the trial will be removed from the database, with all the related sessions and trials, while if you choose “No” the deletion will be aborted.
81BTS BioengineeringBTS FREEEMGguide to the use of onboard software To create a new test session click on “New Session”. A window will open that allows to the user to insert data linked to the session:• patient weight• patient height• pathology • possible notes of the sessionPress “OK” to add the session of the selected patient. e  default  protocol  is  linked  to  the  session;  to  change  the protocol click on the Menu button and enter the “Protocol” item (see § “Protocol”).Cautious that it is not possible, rather than, to change protocol of a session when this has already got some acquisitions. To continue working with the same patient but with a dierent protocol it is necessary to create a new acquisition session for that patient.
82BTS BioengineeringBTS FREEEMGguide to the use of onboard software To modify or to delete the data of a session created previously click on the correspondent line, it will be underlined in light blue and the “Delete” and “Modify” buttons will appear.Clicking on “Modify” a window will open that allows to modify the data related to this session:Click on “Delete” and immediately a window with the request to conrm the deletion of the session and the relate trials will appear:Clicking    on  “Yes”  the  session  and  the  trials  will  be removed from the database, with all the related sessions and trials, while if you choose “No” the deletion will be aborted.
83BTS BioengineeringBTS FREEEMGguide to the use of onboard software Latter  it  is  possible  to  delete  single  trials.  To  do  that  click  on  the correspondent line, it  will be underlined  in light blue and the “Delete” buttons will appear.Click on “Delete” and immediately a window with the request to conrm the deletion of the trial will appear:Clicking  on “Yes” the selected trial will be removed from the database, with  all  the  related  sessions  and  trials,  while  if  you  choose  “No”  the deletion will be aborted.Acquisition:to proceed with the acquisition in “Local” mode  it is necessary to have created a new session and to have selected an analysis protocol or to have selected a session already in the database.Proceed then with  the operations  described in the  previous paragraphs: “Probes Setup”, “Range”, “Digital Oscilloscope”, “Footswitches”. Once nished all the preliminary operations and the necessary checking it is possible to proceed with the acquisitions. To do that, select the item “Capture” from the Menu.
84BTS BioengineeringBTS FREEEMGguide to the use of onboard software A screen will appear, which gives the setup for that acquisition: For  every  test  it’s  possible  to  indicate  the  Start  and  Stop  mode,  the acquisition frequency and the type and notes of the test. To edit the notes click on “Edit Notes”, a window will open which allows you to edit the notes. Introduce the note using the keyboard and then click on “OK” to conrm or on “Cancel” to come back to the previous screen without saving the text just typed.
85BTS BioengineeringBTS FREEEMGguide to the use of onboard software To modify the other acquisition parameters select the relative eld  with the Toggle and then use the Up and Down buttons.Active elds (Toggle) Value range (Select)Acq. Frequency 1KHz, 2KHz, 3KHz, 4KHzFlag MMT, Resting, Walking, Undened, ExtraIf the acquisition frequency selected is 3 or 4KHz, the system can acquire 8 EMG channels maximum. If  the  selected  protocol  requires  a  higher  number  of  probes  a warning message will appear and the acquisition won’t be allowed.By clicking on the “Start” button,  the acquisition starts and the system automatically goes into Lock mode, showing the following:To unlock the palm click on the “Unlock” button, this will immediately
86BTS BioengineeringBTS FREEEMGguide to the use of onboard software move to another position. To complete this procedure it is necessary to press it again before it goes back to the original position.Before quitting this mode the screen with the acquiring setup reappears but the “Start” (or “Arm”) button is substituted with the “Unlock” and “Stop” buttons.e  “Unlock” button allows to come back to the previous screen.Clicking on “Stop”, instead, will end the acquisition.At this point the following screen will be visualized:
87BTS BioengineeringBTS FREEEMGguide to the use of onboard software To save data on BTS FREEEMG 300 select “Store” whereas to cancel the data just acquired click on “Delete”. In this case a conrmation for the cancellation will be asked of the user, select “Yes” if you want to delete the new acquired data. Click“No” button” and then “Store” to save the test.After  the  storing  or  the  cancellation  of  the  last  acquired  test  you  are redirected  to the screen containing the “Start” button.To  continue  with  another  acquisition  click  on  “Start”  and  repeat  the operations just described. If any other acquisition won’t be executed, select “Exit” from the menu.PlaybackIn the Holter mode it is possible to visualize a preview of the data acquired and saved in the database.To  do  that,  from  the  menu  select  the  voice  “Database”  and  select  the patient, the session and the trial that you want to visualize clicking with the stylus on the correspondent line, that will be underlined in light blue.Once the trial is selected, from the menu select the voice “Playback”.At  the  end  of  an  acquisition  session  it  is  possible  to  go  directly  to  the “Playback” page, without passing from the “database”: in this circumstance all the trials saved in the last session acquired will be loaded.In both these circumstances the following screen will appear:
88BTS BioengineeringBTS FREEEMGguide to the use of onboard software From  the  lower  panel  select  the  trial  to  visualize  by  tapping  on  the correspondent line.It is possible to visualize two signals at the same time. To do that select the eld “Signal 1” and than “Signal 2” using the Toggle and scroll the list of the available signals operating on the Up and Down buttons. After  that  to  improve  the  signal  visualization,  adjust  the  active  elds “Zoom” and “Time” selecting them with the Toggle and changing their value with the Up and Down buttons. In particular the “Zoom” eld allows increasing or reducing the current view area (it acts on the axis of ordinates), while the “Time” eld allows changing the temporal  value range  showed on the axis  of the abscissas, choosing among a range of value in between 1 and 10 seconds.is means that the selected trial will not be entirely visualized, but only bits of the length of 1, 2, 5 or 10 seconds according with the “Time” eld.
89BTS BioengineeringBTS FREEEMGguide to the use of onboard software Active elds (Toggle) Value range (Select)Signal 1 Protocol signalsSignal 2 Protocol signalsZoomTime From 1 sec  to 10 sec#  Trial Trials of the selected sessionHowever it is possible to scroll the whole trial changing time after time the frame from where the visualization starts. To do that click with the stylus on the grey band that represents the whole trial in the point correspondent to the rst frame to visualize.On the grey band at the two extremities, are indicated the beginning of the trial (that is 0:00) and the total length of it (like in the example in the gure 0:34, that means 0 minutes and 34 seconds). In the spot in which you tap, a grey rectangle will appear in which will be indicated the correspondent time of the selected rst frame (like  in the example in the gure 0:21). erefore,  if  the  “Time”  eld  is  set,  for  example,  on  “5sec”,  we  are visualizing 5 seconds of trial starting from the 21st second.It is possible, if you regard it relevant, delete some trials.
90BTS BioengineeringBTS FREEEMGguide to the use of onboard software Click on the trial, choosing it in the lower panel, that you want to delete. It will be underlined in light blue and the “Delete” button will appear. Click on “Delete” and immediately a window with the request to conrm the deletion will appear:Clicking  on “Yes” the trial will be removed from the database, while if you choose “No” the deletion will be aborted. Importing data to EMG-Analyzere data acquired in Holter mode is saved in a local database in the BTS FREEEMG 300. To execute the elaboration and the analysis of the data, you need to import the test using the software EMG-Analyzer. To do this, proceed as described further in § “Download button”.
91BTS BioengineeringBTS FREEEMGguide to the use of onboard software “Remote” button: Lab modeIn this mode the connection between RU and WS is always available.e specic characteristics are:• sending data boxes to the WS for the visualization in the remote oscilloscope, available in the applicative software.• data transfer at the end of acquisition. • the duration of acquisition and the Start/Stop are controlled by the PC. Selecting “Capture” from the main menu the following screen appears:is screen displays some information concerning the system: • Status:  Ready mode • Mode:  Stand Alone• Data: NoneTo make an acquisition in Lab mode, there must be no data saved on the BTS FREEEMG 300.
92BTS BioengineeringBTS FREEEMGguide to the use of onboard software is possibility can happen only with a loss of WiFi connection between BTS FREEEMG 300 and Workstation during the last acquisition, or if the data transfer has not gone well.In this case the screen displayed after having selected “Capture” from the main menu will contain instead of the “Arm” button, the “Delete” button and the “Data” eld will be “Ready”. If  you  want,  it  is  possible  to  recover  the  lost  data  saved  in  the  RU before proceeding  with the new acquisitions downloading them on  the Workstation using the EMG-Analyzer software. To know the download procedure consult the manual of EMG-Analyzer. Otherwise the data will have to be deleted and cannot be retrieved in the future. To delete the data, click on the “Delete” button; by doing this the system will ask you a conrmation for the deletion, select “Yes” if you really want to delete the data in a permanent way, otherwise click on “No” to come back to the previous screen.
93BTS BioengineeringBTS FREEEMGguide to the use of onboard software Once deleted the le, the “Arm” button appears again and the “Data” eld will again become “None”; so it’s possible to proceed with the acquisition. Before arming  the  system it is  possible  to set the  acquisition  frequency using the Up and Down button of the button bar.Active elds (Toggle) Value range (Select)Acq. Frequency 1KHz, 2KHz, 3KHz, 4KHzIf  the  acquisition  frequency  selected  is  3  or  4KHz,  the  system can acquire 8 EMG channels maximum. If the selected protocol requires a higher number of probes a warning message will appear and the acquisition won’t be allowed.To   prepare the system for capture it is necessary to “arm” it by  simply tapping the “Arm” button that will  be replaced by the  “Lock”  and  the  “Cancel” buttons. Now it is possible to make the acquisition via remote control.
94BTS BioengineeringBTS FREEEMGguide to the use of onboard software At this level it is not possible to change the acquisition frequency.Tapping  on  the  “Cancel”  button  the  previous  screen  appears,  returns control to the RU. To proceed with the next acquisition, the system needs to be armed again.It is possible to lock the system by clicking   on “Lock” directly by this window, therefore at the beginning of the acquisition the BTS FREEEMG 300 locks automatically displaying:e acquisition starts-up through the remote control, directly by the Workstation.is display will remain on the BTS FREEEMG 300 until the acquiring session is over. It’s  possible to interrupt the acquisition  remotely and proceed  with the next ones.
95BTS BioengineeringBTS FREEEMGguide to the use of onboard software At  the  end  of  every  acquisition  an  automatic  data  download  will  be executed.  Only at the end of the download is it possible to launch another remote acquisition.Once the acquisition session is over, unlock the BTS FREEEMG 300. In order to do so, press  the “Unlock” button, this will immediately move to another position. To complete this procedure it is necessary to press it again before it goes back to the original position.Finally to quit from the “Armed” status click on the “Cancel” button.For the details of the remote management of the acquisition from the WS refer to the specic manual of the software meant to be used.Download ButtonIf the Holter mode is chosen (local acquisition), the acquisitions are saved in a local database to BTS FREEEMG 300.To  elaborate  the  acquired  data  with  the  dedicate  softwares installed on the Workstation, the user needs to download them with EMG-Analyzer  following these instructions:• Select from the main page the download mode clicking on the “Download” button • BTS  FREEEMG  300  keeps  waiting,  showing  the  following screen:
96BTS BioengineeringBTS FREEEMGguide to the use of onboard software • From  EMG-Analyzer,  select  the  “View  Local  Database”  item from the “Laboratory” menu. • A window will appear that will allow you to explore and modify the active database on the BTS FREEEMG 300, and download all the les that you are interested in.For further details concerning the data download refer to the user manuals of EMG-Analyzer software.e data import takes place with a wireless connection; if  the  import  operation  fails,  you  need  to  verify  that the RU and WS are connected  (see § “Appendix C – troubleshooting guide”).• When  the  download is nished  on  the  BTS FREEEMG 300 keep pressing the “Exit” button to go back to the main screen.
97BTS BioengineeringBTS FREEEMGguide to the use of onboard software “Cong” buttone  “Cong” button launches an application that allows the modication of some system settings. It uses ve tabs to reach the dierent information and features:• Info: Main screen that contains information of the installed versions.• General: General settings (i.e. language, audio settings...).• Radio: Associates the ID identier of each EMG and FSW/EGN probe to its label identied by color and number/letter;setting of the signal range;setting the radio channels utilizing the probes.• Database: Database management.• Maintenance: session permitted only to BTS Technician sta, for assistance. For this reason to access this area a Password is required. To execute any changes of the parts that can type on it is possible to use the touch-screen features of the display and where necessary, activate the keyboard,  clicking  on  the  low  right  corner  icon  represented  by  a  little keyboard.To exit this application and come back to the main screen, you must tap on the “Ok” button, placed in the top right corner. One tap on this button will conrm all the changes made.
98BTS BioengineeringBTS FREEEMGguide to the use of onboard software It won’t be possible to undo the changes made, apart setting again, manually the previous values.All  the  changes  made  are  immediately  activated,  therefore,  in  some circumstance,  it  will  be  necessary  to  restart  the  BTSFREEEMG300 application to see these changes applied (for example when you change the language).Let’s analyze now, more in depth, the four tabs that the user can modify.InfoClicking  on  the  “Cong”  button  on  the  main  screen,  it  automatically  selects the application screen in which the tab Info is selected:On this page it is possible to read all the information about the versions of each installed component, hardware or software.
99BTS BioengineeringBTS FREEEMGguide to the use of onboard software GeneralSelecting “General”, the following screen will be visualized:It is possible to set:• e language• “DHCP”,  for  automatic  host  conguration,  the  “IP  address” and “Netmask”, available only with the “DHCP” not checked, to  manually  assign  the  connection  parameters  and  the  “Port Number”. • “Quite  mode”:  Checking  this  voice,  the  program  will  play without any noise (instead of what normally happens when for example you Tap on a button or activate an acquisition…)• “Lock  on  Capture”:  Checking  this  voice  when  an  acquisition starts  the  BTS  FREEEMG  300  will  be  automatically  locked showing the “Unlock” button• “Video o on Capture”: if checked, the display  automatically will  switch-o  at  the  acquisition  beginning.  To  activate  the screen again a click on one of the PocketPC button is enough• e use of a Secure Digital memory card to store the temp le generated during the acquisition (check this voice only if there is a Secure Digital board in the related port).
100BTS BioengineeringBTS FREEEMGguide to the use of onboard software RadioWhen “Radio” is selected the following page is displayed:e “EMG probes” and “FSW/EGN probes” sections allow to associate the ID that uniquely identies each probe manufactured by BTS S.p.A. with the probe’s label. e ID is shown on the clip of every EMG probe and on the back of every FSW/EGN probe. Up to 16 EMG probes and up to 8 FSW/ EGN probes can be registered.  Probes that are not registered in this section will not be recognized by  the  Receiving  Unit  and  therefore  may  not  be  used  for acquisitions.Generally, the system is distributed to the customer with this association already made. However it could be useful if new probes are purchased or if probes are replaced due to a malfunction.
101BTS BioengineeringBTS FREEEMGguide to the use of onboard software However  when  you  receive  the  system  it  is  advisable  to  check that the association has been made correctly. If a probe has been incorrectly associated to a label or the ID number does not match that  on  the  clip  the  probe  may  not  work.  Instead,  changing two  IDs  compared  to  the  label  could  cause  an  exchange  of acquisition signals as well and therefore cause false results in the Electromyographic examination.  Follow  the  procedure  below,  to  register  a  probe  that  has  not  been introduced:• Enter the ID of the probe using the keypad that can be activated on the bottom right-hand side of the page.If  the  string  entered  is  not  admissible  or  has  already been entered the software will highlight it in red.  • Select  the  color  of  the  probe’s  label  (the  colors  can  be  Red, Green, Blue, and Yellow) using the drop down menu.
102BTS BioengineeringBTS FREEEMGguide to the use of onboard software • Finally, select the number or letter shown on the probe’s label (number selection from 1- 16, letter selection from A-F).If the  label  (identied by the  color/ number or  letter combination)  has  already been  used  the  software  will highlight the ID string in red.
103BTS BioengineeringBTS FREEEMGguide to the use of onboard software From the Radio page two other sections may be accessed.e  “Signal”  section  allows  to  select  the  Range  for  electromyographic signals. 3 dierent value are available:• 6,34mV;• 3,17mV;• 1,62mVFrom the “Radio Channels”  section  the  ZigBee channel with  which  to work can be assigned to the activator and the four transmitters.
104BTS BioengineeringBTS FREEEMGguide to the use of onboard software It  is  recommended  to  modify  this  section  only  if  absolutely necessary and eventually with referral to BTS technical support.DatabaseSelecting “Database” the following screen will be visualized:In this section it is possible to select the current database, to create new databases, to delete and rename existent databases.ere are two list-boxes, the rst one, “Media”, allows the selection of the box in which the database can be found or to insert a new one. In particular there are two possible medias: the ash memory, internal to the Pocket PC and the Secure Digital card. If the latter is not present, a message will be visualized indicating that the card is not inserted.
105BTS BioengineeringBTS FREEEMGguide to the use of onboard software Instead the error message:“ERROR: media not initialized” will be visualized if the card has been inserted but hasn’t still been initialized. In this circumstance a button will appear allowing the initialization.Once selected the “Media” it is possible operating on the second list-box to select the database to use. Mind that if the selected Media is empty, it is necessary to create a new one. To create a new database click on “New”. e new database will be created with the default name “TrialsDB”. To change the name select it and click on “Rename”.To delete a database select it and click on “Delete”. is operation removes completely the database and all its content (patients, sessions and trials). For  this  reason  the  system  will  ask  a  conrmation  before proceeding with the deletion.“Exit” buttone “Exit” button allows the closing of the application and the visualization
106BTS BioengineeringBTS FREEEMGguide to the use of onboard software of the operative system installed on the PDA. To do that it is enough to press and hold the “Exit” button for few seconds.
107BTS BioengineeringBTS FREEEMGUpdating the onboard softwareTo update the BTS FREEEMG 300, exit the application by selecting the entry  of  the  Exit  menu  and  holding  down  the  “Exit” button  for  some seconds.Before  carrying  out  the  upgrade  of  the  software  it  is  recommended  a backup be made of the software version currently installed.Slide the PDA out of the methacrylate guide enough to disconnect the Compact Flash. Connect the PDA via the dedicated cable to the USB port of the WS. On  the  WS,  from  Windows,  run  the  update  program,  launching  the “BTSFREEEMG300Setup.exe” le, and follow the video instructions till the update is complete.Reinsert the PDA in the AU.Restart  the  “BTSFREEEMG300”  application  on  the  BTS  FREEEMG 300:  from  the  Start  menu,  click  on  “Programs”,  search  for  the “BTSFREEEMG300” icon and single click it. e main screen will appear.BTS FREEEMG 300 updating
108BTS BioengineeringBTS FREEEMGBTS FREEEMG 300 updatingCreating a backup imageAt the end of each software upgrading, it is important to remember  to create a backup image.It  can  happen  that  the  Pocket  PC  battery    goes  completely dead.  When  this  happens  all  the  BTS  settings  as  also  the BTSFREEEMG300 software will be lost. To restore the software and bring the BTS FREEEMG 300 to the standard conguration it will be enough to proceed with the restore of the last backup done (see Appendix D – Troubleshooting guide).To do the backup proceed as follow:• exit  from  the  BTSFREEEMG300  application  holding  the button “Exit” for some seconds• select “Start” and then “Programs”• among the installed programs, choose “iPAQ Backup”: • with the stylus tap on “Restore Now” and wait the end of the backup procedure.  e  Pocket  PC  will  restart  and  then  the  BTSFREEEMG300 software will be automatically reloaded.
109BTS BioengineeringBTS FREEEMGBTS FREEEMG 300 updatinge backup procedure is now nished, therefore we suggest to rename the backup le created giving  it a more meaningful name,  in particular we recommend to indicate the version number of the software. To do that proceed in this way:• exit from the application• select “Start” and then “Programs”• among the installed programs, choose “File Explorer” click on the  arrow  that  appears  on  the  grey  bar,  under  the  title  “File Explorer” and choose “Device”• open the folder  “iPAQ File Store”• look for the backup le just created that will have a name like: Backup_2007-07-15,  where  the  date  is  the  one  during  the backup time• with the stylus, tap and hold for few seconds on the backup le icon till when the menu appears• choose  rename  and  type  the  new  name  (i.e.  Backup_v107 supposing  you  have  updated    the  software  with  the  version 1.2.107.0).
110BTS BioengineeringBTS FREEEMGPeriodically inspect the PDA and the slide of the receiving unit for any signs of detachment, cracks or other damage.Clean  the  external  parts  periodically  with  cleaning  liquid  approved  for computers. On the PDA monitor use a microbre cloth only, and do not use liquid detergents or solvents.Contact with the skin of  the patients is achieved via the pre-gelled Ag/AgCl electrodes, which, being single-use disposable, intrinsically guarantee a suciently high level of hygiene. Generally    you  use  hypoallergenic  band-aid  or  hypoallergenic  double-sided tape to apply the Footswitches in correspondence of the foot contact points.At the end of each work session, it is important to carefully remove every glue or gel residual from these parts and from the probes.After each use we suggest to clean the switches using a soft cloth moist with a disinfectant solution,  being careful removing any glue residual left.e BTS FREEEMG 300 cannot be sterilised. maintenance and cleaning
113BTS BioengineeringBTS FREEEMGappendix Atechnical specicationsWireless ProbesGeometry:       variableElectrodes:       standard with clip connectionSeparation:       min: 16mm - max: 66mmAutonomy:       8h of use        some days stand-by        some months deep sleepBattery:       rechargeable, lithium ionDimensions:       14x41,5x24,8mm mother electrode        diameter 16x12mm satellite electrodeWeight:       13g battery includedFrequency used:      ISM band        2.4GHz (standard IEEE802.15.4)Input impedance:     >10 GOhmCMMR:       >110 dB @ 50-60HzResolution:       16bitAcquisition frequency:    up to 4KHzSensitivity:       1μVAaccuracy of measurement*:  ± 2%Receiving UnitAutonomy:       up to 24h with rechargeable battery         up to 9h with single batteryBattery:       rechargeable, lithium ion
114BTS BioengineeringBTS FREEEMGappendixDimensions:       155x106x35mmWeight:       360 gramsFrequency used:      ISM band        2.4GHz (standard IEEE802.15.4         and IEEE802.11b)Display:       4” VGA color touch-screen* e system is calibrated at the factory. No further calibration is requiredBTS internal codingName of device, components, parts and/or accessories as per product labelIdentier for device (bar code, catalogue, model or part number)FREEEMG 300Pocket PC with Stylus Acquisition unitBelts and armbands for attachment  to patient• 16 EMG probesFREEEMG 300FootSwitch/Electrogoniometer Wireless ProbeFoot switches: Insole individual foot switches for the automatic identication of the gait events.Electrogoniometers: Sensors for the measurement of joint angles progress over the time.FRESPU12
115BTS BioengineeringBTS FREEEMGName of device, components, parts and/or accessories as per product labelIdentier for device (bar code, catalogue, model or part number)Fine-Wire adapter : Adapter for the Fine-Wire electrodes. FRESPU23O-Port expander : Digital to Analogue EMG data converter. FREALG16EMGenius: software application to acquire, elaborate and report electromyography and footswitch signals EMGE0901EMG-Analyzer: software application for  EMG signal analysis SMAN0901Docking station FRESPU05Charger FRESPU06
116BTS BioengineeringBTS FREEEMGappendix Benvironmental specicationsMin Max NoteOperating Temperature -20° +45°Operating Humidity 50% 80% Relative, non-condensingStorage and Transport Temperature 0° +40°Storage and Transport Humidity 50% 80% Relative, non-condensingAltitude 0m 2000mDegrees of protection provided by the dangerous enclosures of  water and dust (IEC 60529): IPX0.
117BTS BioengineeringBTS FREEEMGappendix Cpower supply and switch oe medical power supplier TR30RAM050 provided with the system has the following functionality:- Provide power to the Receiving Unit  and recharge its battery. -Guarantee the galvanic insulation of the Receiving Unit from the power.To switch o the system the following operation must be done:- Stand-By of the Receiving Unit:1) push the graphical button “MENU” and then select “Exit” to quit the application.2) set the Receiving Unit state on “stand-by” by clicking on the switch -on/o button of the PDA.- Stand-By of the Probes:place the probes on the “Charger” or “Docking Station”.
118BTS BioengineeringBTS FREEEMGBTS  FREEEMG  300  is  an  internally  powered  device:  several  of  its components use batteries as indicated below. e Probe battery replacement can be done only by BTS qualied personnel. e probes are sealed to avoid the access to the internal circuit components.  Batteries are equipped with battery protection circuit to: -over-voltage, threshold 4.3V -under-voltage, threshold 2.8V - short-circuitse  specic  characteristic  of  the  dierent  components  of  the  system  equipped with the battery are: - Receiving UnitQuantity:     1 per each unitTechnology:     lithium ion (Li-Ion)Removable:     yes, by removing the PDA from the acquisition       unit- Wireless ProbesQuantity:     1 per each EMG or FSW/EGN probe   Technology:     lithium polymer (Li-Poly)Removable:     NO, BTS technical service is required appendix Dbattery
119BTS BioengineeringBTS FREEEMGappendix- Docking Station (optional)Quantity:     1 for each unitTechnology:     lithium polymer (Li-Poly)Removable:     NO, BTS technical service is required
120BTS BioengineeringBTS FREEEMG  DECLARATION OF CONFORMITY BTS SpA Via della Croce Rossa 11,  35129 Padova (PD)  – Italy Tel. +39 049 981 5500  Fax +39 049 792 9260 declare under our sole responsibility that the product(s):  name / description:  Electromyographic system / device for recording myoelectrical activity. model:  FreeEMG300S/N:  SN  satisfies the essential  requirements of the Medical  Devices  Directive 93/42/EC (and its amendments inluding 2007/47/CE), and therefore carries the CE marking of the European Union. The conformity assessment procedure is according to the article 11 of the directive (Annex II.3 full quality assurance) and the article 12 it is not applicable. In accordance with Annex IX of the 93/42/EC directive it is classified as follow: CLASS “IIa”  (rule 10) In accordance with IEC 60601-1 is also classified as follow: Class: internally powered device  Applied part type: BF The product conforms to the following standards: EN ISO 14971  Medical Devices - Application of risk management to medical devices.IEC 60601-1  Medical Electrical Equipment - Part 1: General Requirements for basic safety and essential performance.IEC 60601-1-2  Medical Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: electromagnetic compatibility - Requirements and tests.IEC 60601-1-6  Medical Electrical Equipment - Part 1-6: General Requirements for basic safety and essential performance - collateral Standard: UsabilityEN 62304  Medical device software - Software life-cycle processesETSI EN 301 489-3  Electromagnetic compatibility and Radio spectrum Matters (ERM) – Electromagnetic Compatibility (EMC) – standard for radio equipment and services – Part 3: Specific conditions for Short-Range Devices (SRD) operating on frequencies between 9 KHz and 40 GHz.ETSI EN 301 440-2  Electromagnetic compatibility and Radio spectrum Matters (ERM) – Short Range Devices (SRD) – Radio equipment to be used in the 1 GHz to 40 GHz frequency range – Part 2: Harmonized EN covering essential requirements of Article 3(2) of the R&TTE Directive. This compliance is valid ONLY for the equipment identified when used in a manner consistent with the intent of the referenced documents and according to the product’s usage manual.  Notified Body : TÜV Product Service GmbH, Zertifizierstelle, Ridierstrasse 65, 80339 München – Germa-ny,  Identification N. 0123. EC certificate N. G1 12 10 65301 003 valid until January, 16 2018. Padova, DateBruno Ros CEO BTS S.p.A.appendix E declaration of conformity
121BTS BioengineeringBTS FREEEMG1708 Error – Incorrect IP address or port set upe software uses IP address parameters and communication ports to cre-ate network connections between the Workstation and the PDA. In the case of incorrect conguration, the correct parameter values need to be re-input. To do this, launch the software that you are going to use and then, by the menu voice “Laboratory” select “Set Emg Device”.Verify that the IP address and the port setting are correct. e values to specify are the following:          IP Address:   192.168.1.2          Port:           8000and if a second receiving unit is available:          IP Address:   192.168.1.3          Port:           9000If necessary, correct the values and click on “Upgrade” before closing the windows.appendix Ftroubleshooting guide
122BTS BioengineeringBTS FREEEMGappendix1709 Error – Impossibility to connect to the unitOne or more Wi-Fi network resources (Access Point, Workstation, PDA) are not available. In general, we recommend some simple checks as follows:- Access Point Check  they  are  correctly  connected  to  the  power  supply  and switched on. Try switching the AP o and then back on. Since some models are not tted with an On/O switch, it may be necessary to remove and reinsert the power cable.If on the rear of the model there is a selection button, verify it is on position AP- WorkstationIf  the  Workstation  is  tted  with  a  Wi-Fi  interface,  check  it  is active and connected to “BTSFREEEMG300”. As the connection is through the AP, check that the LAN (local area network) is active. -  BTS FREEEMG 300e Wi-Fi connection is active only when the BTS FREEEMG 300 is in Remote Mode or in Download Mode. From the main screen choose one of these two mode.If the connection is still not activated, also with all these settings, please contact the BTS Customer Care.Backup restoringe  backup  restoring  procedure  allows  the  reloading  of  the
123BTS BioengineeringBTS FREEEMGappendix“BTSFREEEMG300”  software  and  the  restoring  of  the  standard conguration  of the Pocket PC. is is  very  useful, for example, when the PDA battery goes completely dead. If this happens, when the PDA is switched on, the software “BTSFREEEMG300” won’t be loaded and, instead of it, a screen will appear in which the system will ask the user to execute some start-up operations.In  this  circumstance,  before  proceeding  with  the  restoring procedure,  it  is  necessary  to  complete  the  start-up  operations required  by  the  system  that  consists  in  the  execution  of  some simple  exercises  like  the stylus pointing, to  execute  a copy and paste, etc...(for further details refer to the PDA manual). When all these operation are nished, the windows main screen will  be  visualized  and  it  will  be  possible  to  proceed  with  the restoring of the last backup.To restore the backup follow these instructions:• select “Start” and then “Programs”• among the installed programs, choose “iPAQ Backup”: • click  on  “Restore Now”  to  restore  the last backup,    the  PDA will restart and then the “BTSFREEEMG300” software will be reloaded• at this point, exit from the application to set the date and the time, to do that select the date under the voice “Start”. For further  details about the  time  and  date setting refer  to  the PDA manual.
124BTS BioengineeringBTS FREEEMGappendixReceiving unit hardware resetIn case the PDA blocks during use and does not respond to any commands, you must perform an hardware reset operation.To do that follow these instructions:• place the receiver on a plane; • grip with your right hand the PDA stilus;• using index and middle nger of your left hand push the two buttons (marked by the icons keypad and arrow) positioned at the right of the PDA display and simultaneously, with your right hand, insert the stylus in the hole on the right side of the PDA, as shown in the gure below:Immediately the system will be rebooted;• when  the  PDA  will  switch  on  again  the  software “BTSFREEEMG300”  won’t  be  loaded  and,  instead  of  it,  a screen will appear in which the system will ask the user to execute some start-up operations; complete the operations required that consists in the execution of some simple exercises like the stylus
125BTS BioengineeringBTS FREEEMGappendixpointing, to execute a copy and paste, etc...(for further details refer to the PDA manual). • when all these operation are nished, wait for the windows main screen to be visualized;• nally proceed with the restoring of the last backup procedure described in the previous paragraph.
BTS Bioengineering Corp.147 Prince Street - Suite 1111201 Brooklyn NY USAinfo: +1 347 204 7027helpdesk: +1 646 575 0426BTS S.p.A.viale Forlanini 4020024 Garbagnate M.se MIItalytel +39 02 366 490 00fax +39 02 366 490 24www.btsbioengineering.cominfo@btsbioengineering.com