Honeywell 50025034 50025034 User Manual D
Honeywell International Inc. 50025034 D
Contents
User Manual D
OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 34-XY-25-16 Revision 3 6/24/08 Notices and Trademarks Copyright 2007 by Honeywell International Inc. Revision 3 June 24, 2008 While this information is presented in good faith and believed to be accurate, Honeywell disclaims the implied warranties of merchantability and fitness for a particular purpose and makes no express warranties except as may be stated in its written agreement with and for its customers. In no event is Honeywell liable to anyone for any indirect, special or consequential damages. The information and specifications in this document are subject to change without notice. Honeywell, PlantScape, Experion PKS, and TotalPlant are registered trademarks of Honeywell International Inc. Other brand or product names are trademarks of their respective owners. Honeywell International Process Solutions 2500 West Union Hills Phoenix, AZ 85027 1-800 343-0228 ii OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 About This Document This document describes preparation, operation and maintenance of the XYR 6000 Wireless Temperature/DI Transmitters. Mounting, installation and wiring are covered in other documents. Honeywell does not recommend using devices for critical control where there is a single point of failure or where single points of failure result in unsafe conditions. OneWireless is targeted at open loop control, supervisory control, and controls that do not have environmental or safety consequences. As with any process control solution, the end-user must weigh the risks and benefits to determine if the products used are the right match for the application based on security, safety, and performance. Additionally, it is up to the end-user to ensure that the control strategy sheds to a safe operating condition if any crucial segment of the control solution fails. Revision Information Document Name XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Document ID Revision Number Publication Date 34-XY-25-16 6/7/07 8/7/07 6/24/08 References The following list identifies all documents that may be sources of reference for material discussed in this publication. Document Title XYR 6000 Transmitters Quick Start Guide Getting Started with Honeywell OneWireless Solutions OneWireless Wireless Builder User’s Guide OneWireless Builder Parameter Reference Support and contact info United States and Canada Contact: Honeywell Solution Support Center Phone: 1-800 822-7673. In Arizona: 602- 313-5558 Calls are answered by dispatcher between 6:00 am and 4:00 pm Mountain Standard Time. Emergency calls outside normal working hours are received by an answering service and returned within one hour. Facsimile: (602) 313-3293 Mail: Honeywell TAC, MS P13 2500 West Union Hills Drive Phoenix, AZ, 85027 Revision 4OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 6/8/08 iii Europe Contact: Phone: Facsimile: Mail: Honeywell TAC-EMEA +32-2-728-2732 +32-2-728-2696 TAC-BE02 Hermes Plaza Hermeslaan, 1H B-1831 Diegem, Belgium Contact: Phone: Honeywell Global TAC – Pacific 1300-300-4822 (toll free within Australia) +61-8-9362-9559 (outside Australia) +61-8-9362-9564 Honeywell Limited Australia 5 Kitchener Way Burswood 6100, Western Australia GTAC@honeywell.com Pacific Facsimile: Mail: Email: India Contact: Phone: Facsimile: Mail: Email: Honeywell Global TAC – India +91-20- 6603-9400 +91-20- 6603-9800 Honeywell Automation India Ltd. 56 and 57, Hadapsar Industrial Estate Hadapsar, Pune –411 013, India Global-TAC-India@honeywell.com Korea Contact: Phone: Facsimile: Mail: Email: Honeywell Global TAC – Korea +82-2-799-6317 +82-11-9227-6324 +82-2-792-9015 Honeywell Co., Ltd 17F, Kikje Center B/D, 191, Hangangro-2Ga Yongsan-gu, Seoul, 140-702, Korea Global-TAC-Korea@honeywell.com People’s Republic of China Contact: Honeywell Global TAC – China Phone: +86- 21-5257-4568 Mail: Honeywell (China) Co., Ltd 33/F, Tower A, City Center, 100 Zunyi Rd. Shanghai 200051, People’s Republic of China Email: Global-TAC-China@honeywell.com iv OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Revision 4 6/24/08 Singapore Contact: Phone: Facsimile: Mail: Email: Honeywell Global TAC – South East Asia +65-6580-3500 +65-6580-3501 +65-6445-3033 Honeywell Private Limited Honeywell Building 17, Changi Business Park Central 1 Singapore 486073 GTAC-SEA@honeywell.com Taiwan Contact: Phone: Facsimile: Mail: Email: Honeywell Global TAC – Taiwan +886- 7- 536-2567 +886-7-536-2039 Honeywell Taiwan Ltd. 17F-1, No. 260, Jhongshan 2nd Road. Cianjhen District Kaohsiung, Taiwan, ROC Global-TAC-Taiwan@honeywell.com Japan Contact: Phone: Facsimile: Mail: Email: Honeywell Global TAC – Japan +81-3-6730-7160 +81-3-6730-7228 Honeywell Japan Inc. New Pier Takeshiba, South Tower Building, 20th Floor, 1-16-1 Kaigan, Minato-ku, Tokyo 105-0022, Japan Global-TAC-JapanJA25@honeywell.com World Wide Web Honeywell Solution Support Online: http://www.honeywell.com/ps Elsewhere Call your nearest Honeywell office. Training Classes Honeywell Automation College: http://www.automationcollege.com Revision 4OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 6/24/08 Symbol Definitions The following table lists those symbols used in this document to denote certain conditions. Symbol Definition ATTENTION: Identifies information that requires special consideration. TIP: Identifies advice or hints for the user, often in terms of performing a task. CAUTION Indicates a situation which, if not avoided, may result in equipment or work (data) on the system being damaged or lost, or may result in the inability to properly operate the process. CAUTION: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices. CAUTION symbol on the equipment refers the user to the product manual for additional information. The symbol appears next to required information in the manual. WARNING: Indicates a potentially hazardous situation, which, if not avoided, could result in serious injury or death. WARNING symbol on the equipment refers the user to the product manual for additional information. The symbol appears next to required information in the manual. WARNING, Risk of electrical shock: Potential shock hazard where HAZARDOUS LIVE voltages greater than 30 Vrms, 42.4 Vpeak, or 60 VDC may be accessible. ESD HAZARD: Danger of an electro-static discharge to which equipment may be sensitive. Observe precautions for handling electrostatic sensitive devices. Protective Earth (PE) terminal: Provided for connection of the protective earth (green or green/yellow) supply system conductor. Functional earth terminal: Used for non-safety purposes such as noise immunity improvement. NOTE: This connection shall be bonded to Protective Earth at the source of supply in accordance with national local electrical code requirements. Earth Ground: Functional earth connection. NOTE: This connection shall be bonded to Protective Earth at the source of supply in accordance with national and local electrical code requirements. Chassis Ground: Identifies a connection to the chassis or frame of the equipment shall be bonded to Protective Earth at the source of supply in accordance with national and local electrical code requirements. continued vi OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Revision 4 6/24/08 Symbol Description ® The Factory Mutual Approval mark means the equipment has been rigorously tested and certified to be reliable. The Canadian Standards mark means the equipment has been tested and meets applicable standards for safety and/or performance. The Ex mark means the equipment complies with the requirements of the European standards that are harmonised with the 94/9/EC Directive (ATEX Directive, named after the French "ATmosphere EXplosible"). For radio equipment used in the European Union in accordance with the R&TTE Directive the CE Mark and the notified body (NB) identification number is used when the NB is involved in the conformity assessment procedure. The alert sign must be used when a restriction on use (output power limit by a country at certain frequencies) applies to the equipment and must follow the CE marking. The C-Tick mark is a certification trade mark registered to ACMA (Australian Communications and Media Authority) in Australia under the Trade Marks Act 1995 and to RSM in New Zealand under section 47 of the NZ Trade Marks Act. The mark is only to be used in accordance with conditions laid down by ACMA and RSM. This mark is equal to the CE Mark used in the European Union. N314 directly under the logo is Honeywell’s unique supplier identification number. Revision 4OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 6/24/08 vii Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual ix Contents 1. INTRODUCTION .................................................................................................... 1 1.1 Purpose ..........................................................................................................................................1 1.2 Scope..............................................................................................................................................1 1.3 OneWireless network overview ...................................................................................................1 1.4 About the transmitter....................................................................................................................1 Inputs......................................................................................................................................................................1 2. SPECIFICATIONS ................................................................................................. 3 2.1 European Union Usage.................................................................................................................3 2.2 Certifications and approvals........................................................................................................4 Transmitter .............................................................................................................................................................4 Authentication Device...........................................................................................................................................10 3. PREPARATION ................................................................................................... 11 3.1 Installation ...................................................................................................................................11 3.2 Configuration...............................................................................................................................11 3.3 Connecting to network ...............................................................................................................11 3.4 Calibrating the transmitter .........................................................................................................11 Methods of calibration ..........................................................................................................................................11 User calibration ....................................................................................................................................................12 Wiring for calibration.............................................................................................................................................14 Restore calibration................................................................................................................................................17 Clear calibration....................................................................................................................................................18 4. FUNCTION BLOCKS........................................................................................... 19 4.1 Introduction .................................................................................................................................19 4.2 Block description ........................................................................................................................19 Block types ...........................................................................................................................................................19 Block diagram.......................................................................................................................................................19 4.3 5. Parameter details ........................................................................................................................20 OPERATION ........................................................................................................ 21 5.1 Overview ......................................................................................................................................21 Transmitter display modes ...................................................................................................................................21 Authentication Device...........................................................................................................................................21 5.2 Transmitter connection status...................................................................................................22 5.3 Transmitter PV display ...............................................................................................................23 5.4 Transmitter quick view of parameters ......................................................................................27 Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual xi Contents 5.5 Transmitter menu ........................................................................................................................28 Menu tree............................................................................................................................................................. 28 5.6 Authentication device menus ....................................................................................................29 Overview .............................................................................................................................................................. 29 Main menu ........................................................................................................................................................... 29 Security and Node Deployment ........................................................................................................................... 30 Device Local Configuration .................................................................................................................................. 32 Read Node Information ........................................................................................................................................ 33 Advanced Options................................................................................................................................................ 35 6. MAINTENANCE/REPAIR .................................................................................... 36 6.1 Parts .............................................................................................................................................36 6.2 Replacing batteries .....................................................................................................................37 When to replace................................................................................................................................................... 37 Tools required ...................................................................................................................................................... 37 Procedure ............................................................................................................................................................ 37 6.3 Replacing sensor module ..........................................................................................................39 When to replace................................................................................................................................................... 39 Tools required ...................................................................................................................................................... 39 Procedure ............................................................................................................................................................ 39 6.4 Replacing antenna ......................................................................................................................41 Tools required ...................................................................................................................................................... 41 Procedure ............................................................................................................................................................ 41 6.5 Replacing terminal board ...........................................................................................................44 When to replace................................................................................................................................................... 44 Tools required ...................................................................................................................................................... 44 Procedure ............................................................................................................................................................ 44 xii OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 Contents Tables Tables Table 1 User calibration: T/C and RTD .......................................................................................................12 Table 2 User calibration: ohms (STTW400 only).........................................................................................13 Table 3 Calibration error messages ............................................................................................................16 Table 4 Transmitter connection status ........................................................................................................22 Table 5 Transmitter PV Display...................................................................................................................23 Table 6 PV engineering units ......................................................................................................................23 Table 7 PV status ........................................................................................................................................24 Table 8 Device status ..................................................................................................................................24 Table 9 Menu tree........................................................................................................................................28 Table 10 Buttons for Device Local Configuration ........................................................................................32 Table 11 Read Node Information ................................................................................................................33 Table 12 Advanced Options ........................................................................................................................35 Table 13 Battery replacement procedure ....................................................................................................38 Table 14 Sensor module replacement.........................................................................................................40 Table 15 Antenna replacement procedure ..................................................................................................42 Table 16 Terminal board replacement procedure .......................................................................................44 Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual xiii Contents Figures Figures Figure 1 XYR 6000 Functional Diagram .......................................................................................................2 Figure 2 Block Diagram ...............................................................................................................................19 Figure 3 Main menu.....................................................................................................................................29 Figure 4 Security and Node Deployment.....................................................................................................30 Figure 5 Device Local Configuration screen................................................................................................32 Figure 6 Read Node Information .................................................................................................................33 Figure 7 Advanced Options .........................................................................................................................35 Figure 8 Battery replacement ......................................................................................................................38 Figure 9 Sensor module removal and replacement ....................................................................................40 Figure 10 Antenna replacement ..................................................................................................................43 Figure 11 Terminal board replacement .......................................................................................................45 xiv OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 1. Introduction 1.1. Purpose 1. Introduction 1.1 Purpose This manual describes the Honeywell OneWireless XYR 6000 Temperature/DI Transmitter function, operation and maintenance. 1.2 Scope The manual includes: 1.3 • Details of topics that relate uniquely to the Honeywell XYR 6000 Temperature/DI Transmitter, • This manual does not cover installation, mounting, or wiring. See XYR 6000 Transmitter Quick Start Guide (document 34-XY-25-21). OneWireless network overview OneWireless is an all digital, serial, two-way communication mesh network that interconnects industrial field sensors to a central system. OneWireless has defined standards to which field devices and operator stations communicate with one another. The communications protocol is built as an "open system" to allow all field devices and equipment that are built to OneWireless standard to be integrated into a system, regardless of the device manufacturer. This interoperability of devices using OneWireless technology is to become an industry standard for automation systems. 1.4 About the transmitter The XYR 6000 Temperature/DI Transmitter is furnished with OneWireless interface to operate in a compatible distributed OneWireless system. The transmitter will interoperate with any OneWirelessregistered device. The transmitter includes OneWireless electronics for operating in a 2.4GHz network. It features function block architecture. Inputs The transmitter supports three input channels in various configurations depending on model. • Up to 3 T/C or millivolt channels (STTW400) • Up to 2 RTD channels (STTW400) • Channel 3 must be T/C or mV (STTW400) • Channels 2 and 3 are discrete inputs (DI) and channel 1 is T/C or millivolt (STTW401) continued Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 1. Introduction 1.4. About the transmitter • Input types − Thermocouple (B ,E, J, K, N, R, S, T) (all models) − RTD (Pt100, Pt200, Pt500) (STTW400) − mV (0 to 10, 0 to 50, 0 to 100) (all models) − Ohm (0 to 100, 0 to 200, 0 to 500, 0 to 1000) (STTW400) The transmitter measures the analog signal from temperature sensors, discrete inputs, millivolt or ohm values and transmits a digital output signal proportional to the measured value for direct digital communications with systems. The discrete input channels support voltage-free floating contacts. Maximum ON contact resistance is 300 ohms. Minimum OFF contact resistance is 5000 ohms. The Process Variable (PV) is available for monitoring and alarm purposes. The cold junction temperature is also available as a fourth channel PV. Available PV update rates are 1, 5, 10, 30 seconds and are set on Wireless Builder. Slower update rates extend battery life. Figure 1 shows a block diagram of the XYR 6000 Temperature transmitter’s operating functions. Radio Board Measurement Board CH 1 EEPROM Input TC RTD mV Ohm CH 2 CH 3 Multiplexer A/D Microprocessor Microprocessor EEPROM Antenna Battery Electronics Housing Configuration Data Figure 1 XYR 6000 Functional Diagram OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 2. Specifications 2.1. European Union Usage 2. Specifications 2.1 European Union Usage This product may be used in any of the following European Union nations. ISO 3166 ISO 3166 Country Country 2 letter code 2 letter code Austria AT Latvia LV Belgium BE Liechtenstein LI Bulgaria BG Lithuania LT Cyprus CY Malta MT Czech Republic CZ Netherlands NL Denmark DK Norway NO Estonia EE Poland PL Finland FI Portugal PT France FR Romania RO Germany DE Slovakia SK Greece GR Slovenia SI Hungary HU Spain ES Iceland IS Sweden SE Ireland IE Switzerland CH Italy IT United Kingdom BG Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 2. Specifications 2.2. Certifications and approvals 2.2 Certifications and approvals Transmitter See the product label for applicable approvals and ratings. Approval / Item CSAcus Intrinsically Safe Ratings / Description CL I, Div 1, Groups A, B, C, & D; CL II, Div 1, Groups E, F & G; CL III, T4 CL I, Zone 0: Ex ia IIC, T4; CL I, Zone 0: AEx ia IIC, T4 CSAcus Explosionproof CL I, Div 1, Groups A, B, C, & D; CL II, Div 1, Groups E, F & G; CL III, T4 CL I, Zone 1: Ex d IIC, T4; CL I, Zone 1: AEx d IIC, T4 CSAcus Nonincendive CL I, Div 2, Groups A, B, C & D; CL II, Div 2, Groups F & G; CL III, Div 2, T4 CL I, Zone 2: Ex nA IIC, T4; CL I, Zone 2: AEx nA IIC, T4 FM Approvals CL I, Div 1, Groups A, B, C, & D; CL II, Div 1, Groups E, F & G; CL III, T4 Intrinsically Safe CL I, Zone 0: AEx ia IIC, T4 FM Approvals CL I, Div 1, Groups A, B, C, & D; CL II, Div 1, Groups E, F & G; CL III, T4 Explosionproof CL I, Zone 1: AEx d IIC, T4 FM Approvals CL I, Div 2, Groups A, B, C & D; CL II, Div 2, Groups F & G; CL III, Div 2, T4 Nonincendive CL I, Zone 2: AEx nA IIC, T4 HON – ATEX , Ex nA IIC, T4; Ta = 85°C, Zone 2 Non-Sparking Process Connections in Division 2 / Zone 2 Division 2 / Zone 2 apparatus may only be connected to processes classified as non-hazardous or Division 2 / Zone 2. Connection to hazardous (flammable or ignition capable) Division 1 / Zone 0, or 1 process is not permitted. Enclosure Type Type 4X, IP 66/67 CRN Canadian Registration Number Class II and III installations and for Type 4X/IP66 applications require that all cable and unused entires be sealed with an NRTL listed cable gland or seal fitting. Cable glands and seal fittings are not supplied by Honeywell. For detailed transmitter specifications see the following Specification and Model Selection Guide. • XYR 6000 Wireless Transmitter Temperature/DI (document 34-XY-03-29) OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 2. Specifications 2.3. Agency compliance information 2.3 Agency compliance information This section contains the Federal Communications Commission (FCC), Industry Canada (IC) and Radio Frequency compliance statements for the OneWireless Multinode device. ATTENTION XYR6000 units must be professionally installed in accordance with the requirements specified in the OneWireless XYR6000 Agency Compliance Professional Installation Guide. FCC compliance statements • This device complies with Part 15 of FCC Rules and Regulations. Operation is subject to the following two conditions: (1) This device may not cause harmful interference and (2) this device must accept any interference received, including interference that may cause undesired operation. • This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radiofrequency energy and, if not installed and used in accordance with these instructions, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. • Intentional or unintentional changes or modifications must not be made to the Multinode unless under the express consent of the party responsible for compliance. Any such modifications could void the user’s authority to operate the equipment and will void the manufacturer’s warranty. IC compliance statements • To reduce potential radio interference to other users, the antenna type and its gain should be so chosen that the equivalent isotropic radiated power (EIRP) is not more than that permitted for successful communication. • Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. • This Class A digital apparatus complies with Canadian ICES-003. • French: Cet appareil numérique de la classe A est conforme à la norme NMB-003 du Canada. Radio Frequency (RF) statement To comply with FCC’s and Industry Canada’s RF exposure requirements, the following antenna installation and device operating configurations must be satisfied. Rev. 3 6/24/08 • Remote Point-to-Multi-Point antenna(s) for this unit must be fixed and mounted on outdoor permanent structures with a separation distance between the antenna(s) of greater than 20cm and a separation distance of at least 20cm from all persons. • Remote Fixed Point–to-Point antenna(s) for this unit must be fixed and mounted on outdoor permanent structures with a separation distance between the antenna(s) of greater than 20cm and a separation distance of at least 100cm from all persons. OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 2. Specifications 2.3. Agency compliance information • Furthermore, when using integral antenna(s) the Multinode unit must not be co-located with any other antenna or transmitter device and have a separation distance of at least 20cm from all persons. European Union restriction France restricts outdoor use to 10mW (10dBm) EIRP in the frequency range of 2,454-2,483.5 MHz. Installations in France must limit EIRP to 10dBm, for operating modes utilizing frequencies in the range of 2,454 – 2,483.5MHz. OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 2. Specifications 2.4. Honeywell European (CE) Declaration of Conformity (DoC) 2.4 Honeywell European (CE) Declaration of Conformity (DoC) This section contains the European Declaration of Conformity (DoC) statement, for the OneWireless product line. R&TTE Directive 1999/5/EC LVD Directive 73/23/EEC EMC Directive 2004/108/EC ATEX Directive 94/9/EC Harmonized Standards Emissions Specification and Method: EN 300 328 V1.7.1 Emissions Spec and Method: EN 301 893 V1.4.1 Immunity Specification: EN 301 489-17 V1.2.1 Immunity Method: EN 301 489-1 V1.6.1 Product Standard: IEC61326-1 (1st Edition, 2002-02, Industrial Locations) EN 50014:1992, "Electrical Apparatus for Potentially Explosive Atmospheres – General Requirements" EN 50021:1999, "Electrical Apparatus for Potentially Explosive Atmospheres – Type of Protection "n" Manufacturer’s Name and Address Honeywell Process Solutions Compliance Statement The product herewith complies with the harmonized standards listed above. Typical product line systems and configurations have been tested, for compliance. Rev. 3 6/24/08 2500 West Union Hills Drive, Phoenix, AZ 85027, USA OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 2. Specifications 2.4. Honeywell European (CE) Declaration of Conformity (DoC) European Declaration of Conformity statements Language Statement Česky (Czech): Honeywell tímto prohlašuje, že tento Multinode je ve shodě se základními požadavky a dalšími příslušnými ustanoveními směrnice 1999/5/ES. Dansk (Danish): Undertegnede Honeywell erklærer herved, at følgende udstyr Multinode overholder de væsentlige krav og øvrige relevante krav i direktiv 1999/5/EF. Deutsch (German): Hiermit erklärt Honeywell, dass sich das Gerät Multinode in Übereinstimmung mit den grundlegenden Anforderungen und den übrigen einschlägigen Bestimmungen der Richtlinie 1999/5/EG befindet. Eesti (Estonian): Käesolevaga kinnitab Honeywell seadme Multinode vastavust direktiivi 1999/5/EÜ põhinõuetele ja nimetatud direktiivist tulenevatele teistele asjakohastele sätetele. English Hereby, Honeywell, declares that this Multinode is in compliance with the essential requirements and other relevant provisions of Directive 1999/5/EC. Español (Spanish): Por medio de la presente Honeywell declara que el Multinode cumple con los requisitos esenciales y cualesquiera otras disposiciones aplicables o exigibles de la Directiva 1999/5/CE. Ελληνική (Greek): ΜΕ ΤΗΝ ΠΑΡΟΥΣΑ Honeywell ∆ΗΛΩΝΕΙ ΟΤΙ Multinode ΣΥΜΜΟΡΦΩΝΕΤΑΙ ΠΡΟΣ ΤΙΣ ΟΥΣΙΩ∆ΕΙΣ ΑΠΑΙΤΗΣΕΙΣ ΚΑΙ ΤΙΣ ΛΟΙΠΕΣ ΣΧΕΤΙΚΕΣ ∆ΙΑΤΑΞΕΙΣ ΤΗΣ Ο∆ΗΓΙΑΣ 1999/5/ΕΚ. Français (French): Par la présente Honeywell déclare que l'appareil Multinode est conforme aux exigences essentielles et aux autres dispositions pertinentes de la directive 1999/5/CE. Italiano (Italian): Con la presente Honeywell dichiara che questo Multinode è conforme ai requisiti essenziali ed alle altre disposizioni pertinenti stabilite dalla direttiva 1999/5/CE. Latviski (Latvian): Ar šo Honeywell deklarē, ka Multinode atbilst Direktīvas 1999/5/EK būtiskajām prasībām un citiem ar to saistītajiem noteikumiem. Lietuvių (Lithuanian): Šiuo Honeywell deklaruoja, kad šis Multinode atitinka esminius reikalavimus ir kitas 1999/5/EB Direktyvos nuostatas. Nederlands (Dutch): Hierbij verklaart Honeywell dat het toestel Multinode in overeenstemming is met de essentiële eisen en de andere relevante bepalingen van richtlijn 1999/5/EG. Malti (Maltese): Hawnhekk, Honeywell, jiddikjara li dan Multinode jikkonforma malħtiġijiet essenzjali u ma provvedimenti oħrajn relevanti li hemm fidDirrettiva 1999/5/EC. Magyar (Hungarian): Alulírott, Honeywell nyilatkozom, hogy a Multinode megfelel a vonatkozó alapvetõ követelményeknek és az 1999/5/EC irányelv egyéb elõírásainak. Polski (Polish): Niniejszym Honeywell oświadcza, że Multinode jest zgodny z zasadniczymi wymogami oraz pozostałymi stosownymi OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 2. Specifications 2.4. Honeywell European (CE) Declaration of Conformity (DoC) Language Statement postanowieniami Dyrektywy 1999/5/EC. Português (Portuguese): Honeywell declara que este Multinode está conforme com os requisitos essenciais e outras disposições da Directiva 1999/5/CE. Slovensko (Slovenian): Honeywell izjavlja, da je ta Multinode v skladu z bistvenimi zahtevami in ostalimi relevantnimi določili direktive 1999/5/ES. Slovensky (Slovak): Honeywell týmto vyhlasuje, že Multinode spĺňa základné požiadavky a všetky príslušné ustanovenia Smernice 1999/5/ES. Suomi (Finnish): Honeywell vakuuttaa täten että Multinode tyyppinen laite on direktiivin 1999/5/EY oleellisten vaatimusten ja sitä koskevien direktiivin muiden ehtojen mukainen. Svenska (Swedish): Härmed intygar Honeywell att denna Multinode står I överensstämmelse med de väsentliga egenskapskrav och övriga relevanta bestämmelser som framgår av direktiv 1999/5/EG. Íslenska (Icelandic): Hér með lýsir Honeywell yfir því að Multinode er í samræmi við grunnkröfur og aðrar kröfur, sem gerðar eru í tilskipun 1999/5/EC. Norsk (Norwegian): Honeywell erklærer herved at utstyret Multinode er i samsvar med de grunnleggende krav og øvrige relevante krav i direktiv 1999/5/EF. For more information about the R&TTE Directive The following website contains additional information about the Radio and Telecommunications Terminal Equipment (R&TTE) directive: http://ec.europa.eu/enterprise/rtte/faq.htm Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 2. Specifications 2.4. Honeywell European (CE) Declaration of Conformity (DoC) Authentication Device Install the Authentication Device application on any PDA having 10 • Windows Mobile version 4.2+ • infrared port. OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 3. Preparation 3.1. Installation 3. Preparation 3.1 Installation Refer to the XYR 6000 Transmitter Quick Start Guide (document 34-XY-25-21) for installation, mounting and wiring of your XYR 6000 transmitter. 3.2 Configuration The XYR 6000 Transmitter contains the electronics interface compatible for connecting to the OneWireless network. An operator uses the Wireless Builder application to configure blocks and to change operating parameters. These changes are written to the transmitter when it is authenticated by a security key. 3.3 Connecting to network Use Authentication Device to connect your transmitter to the OneWireless network. See page 30. 3.4 Calibrating the transmitter Methods of calibration Calibration can be done at the transmitter or at Wireless Builder’s Method Manager. For all calibration methods at the transmitter, Wireless Builder must be used to unlock and take the channel out of service. For the STTW401, calibration applies only to the thermocouple/millivolt input, not to the discrete inputs. • User calibration - Calibrates the channel to the low and high range values for the selected channel’s input type. • Restore calibration - Calibration constants for the selected channel are restored to factory defaults. • Clear calibration - Factory or user calibration constants for the selected channel are cleared. Additional calibration commands are available in Wireless Builder. Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 11 3. Preparation 3.4. Calibrating the transmitter User calibration Calibrates the channel to the default low and high range values for the channel’s input type. Refer to the table for the input type desired. Table 1 User calibration: T/C and RTD Step Action In Wireless Builder, set the transmitter’s Write Lock to UNLOCKED. In Wireless Builder, set the transmitter’s channel to OOS (Out of Service). Loosen the M3 locking set screw on the transmitter’s battery end-cap (opposite end from display). Unscrew and remove the end cap. See page 14 for terminal connections. If your PV is connected to the channel, disconnect it first, then connect a calibrator source to the channel’s terminals. It is not necessary to disconnect any jumper wires on the terminals. At the transmitter display, verify that the channel’s number and PV value are not displayed. Use Authentication Device’s Device Local Configuration buttons to navigate to the transmitter’s CAL menu. If the transmitter is locked a LOCKED message will be displayed. Go to step 1. If CAL menu is passcode protected, enter the passcode. If the channel is not out of service a WRONG MODE message will be displayed. Go to step 2. Select the channel for calibration (CH 1, CH 2 or CH 3). Select USER CAL. Follow displayed instructions. • Using your calibration source, apply the low calibration value indicated on display. For TC apply cold junction compensated mV value. For RTD apply equivalent resistance value. • Press Enter to accept the value. Display will say BUSY, then CAL LO COMPLETE. • Press Enter to continue. • Apply the high calibration input value indicated on display. • Press Enter to accept the value. Display will say BUSY, then SUCCESS. Otherwise, the display will show one of the calibration error messages listed in Table 3. • Press Enter to return to PV display. 12 Repeat steps 1 through 6 for each channel you want to calibrate. Reverse steps 3 and 4. When ready, in Wireless Builder return the transmitter’s channel to service and set Write Lock to LOCKED. OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 3. Preparation 3.4. Calibrating the transmitter Table 2 User calibration: ohms (STTW400 only) Step Action In Wireless Builder, set the transmitter’s Write Lock to UNLOCKED. In Wireless Builder, set the transmitter’s channel to OOS (Out of Service). Wait 20 seconds. In Wireless Builder, do a Warm Start. Loosen the M3 locking set screw on the transmitter’s battery end-cap (opposite end from display). Unscrew and remove the end cap. See page 14 for terminal connections. If your PV is connected to the channel, disconnect it first. At the transmitter display, verify the channel’s number’ and PV value are not displayed. Use Authentication Device’s Device Local Configuration buttons to navigate to the transmitter’s CAL menu. If the transmitter is locked a LOCKED message will be displayed. Go to step 1. If CAL menu is passcode protected, enter the passcode. If the channel is not out of service a WRONG MODE message will be displayed. Go to step 2. You will need two 0-ohm wires (that is, two short wires) and one 1000-ohm resistor. The wires must be the same gauge as the resistor leads. • Select the channel for calibration (CH 1, CH 2). Select USER CAL. Display will say APPLY 0 OHMS. • For CH 1: Connect one short wire across terminals 1 and 2. Connect the other short wire across terminals 2 and 3. • For CH 2: Connect one short wire across terminals 4 and 5. Connect the other short wire across terminals 5 and 6. • Press Enter to accept the low value. Display will say BUSY, then CAL LO COMPLETE. • Press Enter to continue. Display will say APPLY 1000 OHMS. • For CH 1: Replace the short wire across terminals 1 and 2 with the 1000 ohm resistor. • For CH 2: Replace the short wire across terminals 4 and 5 with the 1000 ohm resistor. • Press Enter to accept the high value. Display will say BUSY, then SUCCESS. Otherwise, the display will show one of the calibration error messages listed in Table 3. • Press Enter to return to PV display. Repeat steps 1 through 7 for each channel you want to calibrate. Optional check (skip this step if desired): • With the calibration wiring still connected, use Wireless Builder to put the channel in AUTO mode. • Verify the PV for that channel is reading 1000 ohms. • Replace 1000 ohm wire with 0 ohm wire. Verify the PV for that channel is reading 0 ohms. Rev. 3 6/24/08 10 Remove calibration wiring and re-connect sensor wires. 11 When ready, in Wireless Builder return the transmitter’s channel to service and set Write Lock to LOCKED. OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 13 3. Preparation 3.4. Calibrating the transmitter Wiring for calibration ATTENTION When two wires will be installed in one screw terminal, both wires must be the same size (gauge) and construction (solid or stranded). TWO THERMOCOUPLE CONNECTION SINGLE THERMOCOUPLE CONNECTION T/C or mV (PV1) TB1 T/C or mV (PV1) TB1 TB2 T/C or mV (PV2) TB2 3 JUMPER WIRES 4 JUMPER WIRES T/C MAY BE GROUNDED TYPE ONLY ONE T/C MAY BE GROUNDED TYPE THREE THERMOCOUPLE CONNECTION TB1 TB2 T/C or mV (PV1) T/C or mV (PV2) T/C or mV SINGLE THREE -WIRE RTD RTD or Ohms (PV1) TB1 TB2 (PV3) 2 JUMPER WIRES ONLY ONE T/C MAY BE GROUNDED TYPE SINGLE THREE -WIRE RTD & T/C TWO THREE-WIRE RTD’s TB1 RTD or Ohms (PV1) TB2 RTD or Ohms (PV2) RTD or Ohms (PV1) TB1 TB2 T/C or mV (PV2) T/C MUST BE FLOATING (UNGROUNDED) TYPE SINGLE TWO-WIRE RTD T/C & SINGLE THREE -WIRE RTD T/C or mV (PV1) TB1 TB2 RTD or Ohms (PV2) T/C MUST BE FLOATING (UNGROUNDED) TYPE 14 RTD or Ohms (PV1) TB1 TB2 JUMPER WIRE OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 3. Preparation 3.4. Calibrating the transmitter T/C & SINGLE TWO-WIRE RTD TWO TWO-WIRE RTD’s RTD or Ohms (PV1) TB1 TB2 RTD or Ohms (PV2) T/C or mV (PV1) JUMPER WIRES RTD or Ohms (PV1) TB2 TB2 RTD or Ohms (PV2) T/C MUST BE FLOATING (UNGROUNDED) TYPE STTW 401 T/C AND 2 DI CONNECTION SINGLE TWO-WIRE RTD & T/C TB1 TB1 T/C or mV TB1 T/C (PV2) (PV1) TB2 S2 S1 T/C MUST BE FLOATING (UNGROUNDED) TYPE Rev. 3 6/24/08 2 JUMPER WIRES OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 15 3. Preparation 3.4. Calibrating the transmitter Table 3 Calibration error messages Message CALIBRATION_FAIL Meaning 1. Calibration gain is greater than 5%. 2. Calibration offset is greater than 5% of sensor span. (TC span is in mV. RTD span is in Ohms.) BAD_TRIM_POINT CAL_POINT_HI is greater than sensor high range value OR CAL_POINT_LO is less than greater than sensor low range value. BAD_USER_CALIBRATION CAL_SOURCE is user and user calibration constants contain invalid values. BAD_FACTORY_CALIBRATION 1. CAL_SOURCE is factory and factory calibration constants do not contain valid values. 2. CAL_RESTORE command was issued but factory calibration constants do not contain valid values. 16 BAD_SENSOR Sensor is bad or faulty input thermocouple. BAD_UNITS Units in CAL UNITS parameter are invalid or not supported by the sensor type. INTERNAL_ERROR An error occurred during calibration that prevents calibration from being completed successfully. SUCCESS_WITH_EXCESS The calibration succeeded but the calculated calibration values are greater that 5 percent beyond the normal calibration values. Typically this indicates that the applied calibration value was significantly different from the expected value or that the sensor is not within expected tolerances for the applied characterization. OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 3. Preparation 3.4. Calibrating the transmitter Restore calibration Step Action In Wireless Builder, set the transmitter’s Write Lock to Unlocked. In Wireless Builder, set the transmitter’s channel to OOS (Out of Service). At the transmitter display, verify the channel’s PV value is displayed. Use Authentication Device’s Device Local Configuration buttons to navigate to the transmitter’s CAL menu. If CAL menu is passcode protected, enter the passcode. If the channel is not out of service a WRONG MODE message will be displayed. Go to step 1. If the transmitter is locked a LOCKED message will be displayed. Go to step 2. Select the channel (CH 1, CH 2 or CH 3). Select CAL RSTR Press Enter to continue. Display will say BUSY, then SUCCESS. If unsuccessful the display will show BAD_FACTORY_CALIBRATION. Rev. 3 6/24/08 When ready, in Wireless Builder return the transmitter’s channel to service and set Write Lock to Locked. OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 17 3. Preparation 3.4. Calibrating the transmitter Clear calibration Step Action In Wireless Builder, set the transmitter’s Write Lock to Unlocked. In Wireless Builder, set the transmitter’s channel to OOS (Out of Service). At the transmitter display, verify the channel’s number and PV value is not displayed. Use Authentication Device’s Device Local Configuration buttons to navigate to the transmitter’s CAL menu. If CAL menu is passcode protected, enter the passcode. If the channel is not out of service a WRONG MODE message will be displayed. Go to step 1. If the transmitter is locked a LOCKED message will be displayed. Go to step 2. Select the channel (CH 1, CH 2 or CH 3). Select CAL CLR Press Enter to continue. Display will say BUSY, then SUCCESS. Repeat for each channel desired. 18 When ready, in Wireless Builder return the transmitter’s channel to service and set Write Lock to Locked. OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 4. Function blocks 4.1. Introduction 4. Function blocks 4.1 Introduction This section explains the construction and contents of the XYR 6000 Temperature/DI Transmitter Function Blocks. 4.2 Block description Block types Blocks are the key elements that make up the transmitter’s configuration. The blocks contain data (block objects and parameters) which define the application, such as the inputs and outputs, signal processing and connections to other applications. The XYR 6000 Transmitter contains the following block types. Block Type Function Device Contains parameters related to the overall field device rather than a specific input or output channel within it. A field device has exactly one device block. TB (Transducer) Contains parameters related to a specific process input or output channel in a measurement or actuation device. A TB defines a measurement sensor channel for an analog process variable represented by a floating-point value. There is one TB per sensor. Radio Contains parameters related to radio communication between the transmitter and the multimode(s). Block diagram Figure 2 shows the blocks of the XYR 6000 Transmitter. Transmitter Sensor Transducer Block (TB) Algorithm Device Block OUT Read/ Write Publish Read/ Write Communication Stack Figure 2 Block Diagram Each of these blocks contains parameters that are standard OneWireless transmitter defined parameters. The Transducer and Device blocks contain standard parameters common to all XYR 6000 transmitter models (that is, pressure, temperature, corrosion, HLAI, multi-DI) as well as temperature-specific parameters. The radio block contains parameters for communication with the wireless network. Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 19 4. Function blocks 4.3. Parameter details 4.3 Parameter details The transmitter itself displays a few basic parameters, such as tag, serial number, device revision, build, device address, WFN ID. For more information on parameters, refer to the following documents. 20 • OneWireless Wireless Builder User’s Guide • OneWireless Builder Parameter Reference OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 5. Operation 5.1. Overview 5. Operation 5.1 Overview Transmitter display modes The transmitter has the following display modes. • Test. Appears briefly after power-up to self-test the display. • Connection status. Appears when transmitter is not fully connected to the OneWireless network. See section 5.2. • PV display. Default mode of the transmitter displays the PV and any status messages. See section 5.3 on page 23. • Quick view of transmitter identification parameters. Displays read-only parameters then returns to PV display. See section 5.4 on page 27. • Menu. See section 5.5 on page 28. Authentication Device To navigate the transmitter displays and menus, hold the Authentication Device no more than 6” (15 cm) from the transmitter and aim the infrared beam at the transmitter display while tapping the Device Local Configuration buttons (Table 10). You can also use the PDA’s buttons. Authentication Device menus are described in section 5.6 starting on page 29. Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 21 5. Operation 5.2. Transmitter connection status 5.2 Transmitter connection status Table 4 Transmitter connection status Displayed status Definition What to do NO KEY Transmitter needs a key from the Authentication Device and is not transmitting. Transmit a key to the transmitter. See page 30. NOT CONN Transmitter is in between discovery attempts. If Transmitter does not make a connection within five minutes, do the following: • Check that Key is correct for the network you are trying to join. • Check that Multinode(s) in the local area are turned on and are already a secure part of the network. • Check if KeyServer is active. • Check the KeyServer Event Log to see if the Transmitter is actively trying to join. Errors in the Event Log show that the Transmitter is trying to join but that there are problems. Consult the OneWireless Wireless Builder documentation for troubleshooting errors. DISCOVER Transmitter has not made a connection to a Multinode and is in discovery (searching for a connection to a Multinode). Transmitter will automatically enter a power saving mode if it cannot make a connection and will retry later. Wait for connection. If Transmitter does not make a connection within five minutes, see NOT CONN in this table. SECURING Transmitter has connected with the network and is validating its key. Wait for connection. If Transmitter does not make a connection within five minutes, see NOT CONN in this table. CONNECTD For units with radio firmware build* 53 or higher: For units with radio firmware build* 53 or higher: No action required. Transmitter has validated the key and has made a secure connection with at least two Multinodes. Transmitter should appear in Wireless Builder as an uncommissioned device. For units with radio firmware build* 52: Transmitter will periodically look for a second Multinode in order to form a redundant connection to the network. If connected with only one Multinode Wireless Builder will display a Secondary Multinode Address of 0. For units with radio firmware build* 52: Transmitter has validated the key and has made a secure connection with at least one Multinode. Transmitter should appear in Wireless Builder as an uncommissioned device. NO REDUN Appears only on units with radio firmware build* 53 or higher. No redundancy, that is, Transmitter has connected with only one Multinode. No action required. The Transmitter will periodically look for a second Multinode in order to form a redundant connection to the network *Use the PDA to determine your radio firmware build number (page 21). 22 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 5. Operation 5.3. Transmitter PV display 5.3 Transmitter PV display The following information is displayed in sequence for the first channel, second channel, etc. then repeating continuously unless another display mode (see Transmitter display modes on page 21) is selected. Table 5 Transmitter PV Display Item displayed Example Details Channel number Channel number 1, 2, or 3 PV value 350 Latest PV value. PV engineering units DEG F See Table 6 on page 23. PV status BAD See Table 7 on page 24. If PV status is not displayed then the PV value is good. Device status LOW BATT See Table 8 on page 24. If multiple device status messages are in effect, they are displayed one message per channel until all messages have been displayed. If no device status is displayed then the device status is normal. Table 6 PV engineering units Engineering units Rev. 3 6/24/08 Description Deg C Degrees Celsius Deg F Degrees Fahrenheit mV Millivolts Ohm Ohms OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 23 5. Operation 5.3. Transmitter PV display Table 7 PV status PV status Cause – Action (blank) • PV is normal – no action required BAD • Possible calibration error – Clear calibration • Transducer Block can not execute due to internal firmware state – Attempt cold restart of device. • Transducer Block can not execute due to hardware fault – Replace sensor module. See page 36. • UNC Sensor failure – Check the following. − connection between sensor board and terminal block − field wiring terminations on the terminal block − field sensor • Warning: Input inaccurate due to uncertain input data integrity. • Warning: Input inaccurate due to input conversion limitations or resolution. • Warning: Input outside of characterized range. Value is estimated. Table 8 Device status Transmitter display Wireless Builder display Definition What to do OUT SVC OOS All channels are out of service. Restore mode to Auto in Wireless Builder. CAL ERR Calibration Error Calibration Data Invalid or could not be read. Use Cal Clear, Restore, or User Calibrate. CFG ERR Configuration Error Configuration Check Error. Database is corrupted. Cold start and reload configuration. LOW BAT Low Battery Battery Voltage Critically Low Replace batteries as soon as possible. See page 37. NO RADIO Radio Interprocessor Comm Error Radio Board is not accessible. Restart both the radio and sensor. If condition persists, replace sensor module. See page 36. CJ FAIL CJ Failure Cold junction failure Check connectors on Terminal Board and sensor module. Replace terminal board. See page 44. BAD RADIO SPI 24 Sensor Radio SPI Communication Failure Radio detected loss of communication with sensor board over the inter-processor communication link. Restart both the radio and sensor. If condition persists, replace sensor module. See page 36. OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 5. Operation 5.3. Transmitter PV display Transmitter display Wireless Builder display Definition What to do BAD RADIO EEPROM EEPROM SPI Communication Failure Radio EEPROM SPI Communication failure The radio will not be able to perform firmware upgrades but will operate normally using installed code. Replace sensor module. See page 36. RADIO WDT RESET WDT Reset Occurred Radio Watch Dog Timeout detected Restart both the radio and sensor. If condition persists, replace sensor module. See page 36. BAD RADIO Radio Circuitry Failure Radio circuitry has failed The radio processor detected error on internal radio circuitry. Replace sensor module. See page 36. The following status messages have multiple meanings. Refer to Wireless Builder Device Status for exact cause. E FAIL A/D Failure Diagnostics detected defect with Analog to Digital Converter. Replace sensor module. See page 36. E FAIL Electronics Failure Electronic Failure detected on Sensor Board. Could be caused by one of the status items marked by *. Restart both the radio and sensor. If condition persists, replace sensor module. See page 36. E FAIL* NVM Fault* Startup diagnostics detected defect in Sensor Non-Volatile Memory Replace sensor module. See page 36. E FAIL* Program Memory Fault* Startup diagnostics detected defect in Sensor Read Only Memory Replace sensor module. See page 36. E FAIL* RAM Fault* Startup diagnostics detected defect in Processor Random Access Memory Replace sensor module. See page 36. INP FAIL Input 1 Failure Input 1 error Check input 1 connection. Check input 1 configuration. Restart the sensor. If condition persists, replace sensor module. See page 36. INP FAIL Input 2 Failure Input 2 error Check input 2 connection. Check input 2 configuration. Restart the sensor. If condition persists, replace sensor module. See page 36. Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 25 5. Operation 5.3. Transmitter PV display Transmitter display INP FAIL Wireless Builder display Input 3 Failure Definition Input 3 error What to do Check input 3 connection. Check input 3 configuration. Restart the sensor. If condition persists, replace sensor module. See page 36. INP FAIL Input Failure Input Error Possible meter body sensor failure. INP FAIL A/D Failure Diagnostics detected defect with Analog to Digital Converter. Replace sensor module. See page 36. The following statuses are displayed only in Wireless Builder Device Status. blank* Device/Firmware Mismatch* Sensor Board Firmware Error. The software did not pass verification tests. Replace sensor module. See page 36. blank* Heap Memory Not Available* Heap Allocation Failure. Software detected heap shortage and some communication packets may have been dropped. Clear by warm restart of device. If condition persists contact Honeywell service. blank Input 1 T/C Warning Faulty thermocouple, resistance excessive Check input 1 connection and wiring. Replace thermocouple. blank Input 2 T/C Warning Faulty thermocouple, resistance excessive Check input 2 connection and wiring. Replace thermocouple. blank Input 2 Warning Indeterminate discrete state at input Check input 2 connection and wiring. Check input contacts. blank Input 3 T/C Warning Faulty thermocouple, resistance excessive Check input 3 connection and wiring. Replace thermocouple. blank Input 3 Warning Indeterminate discrete state at input Check input 3 connection and wiring. Check input contacts. blank* 26 Watchdog Timer Error* Sensor Watchdog Timeout. The processor was restarted due to unexpected operation. Clear by warm restart of device. If condition persists contact Honeywell service. OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 5. Operation 5.4. Transmitter quick view of parameters 5.4 Transmitter quick view of parameters If you press the up or down arrow key during the PV display, the following quick view parameters are shown sequentially, then the PV display resumes. Parameter Transmitter type Description HONEYWELL XYR 6000 TEMPERATURE or HONEYWELL XYR 6000 TEMPERATURE DI Rev. 3 6/24/08 Tag The name given to this transmitter Serial number Transmitter serial number. This is the WBSN on the transmitter’s nameplate. Do not confuse this with the other nameplate item marked “Serial.” Device revision This parameter changes whenever objects and parameters are added, deleted, or the data type or range changes. It does not change if the application firmware changes without affecting the device description. Build Sensor firmware number OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 27 5. Operation 5.5. Transmitter menu 5.5 Transmitter menu Menu tree At the PV display, press Enter to access the menus. To interact with the menus use the Device Local Configuration onscreen buttons (page 32) or the buttons on your PDA. Table 9 Menu tree Menu item Description CAL Calibration menu. May be password-protected. See Table 10 on page 32 for password number entry. CH-1 Channel 1. CH-2 Channel 2. CH-3 Channel 3. CAL RSTR Restores selected channel to factory calibration. USER CAL Lets you calibrate selected channel’s low and high points. CAL CLR Clear the factory and user calibration. EXIT Exits calibration menu and returns to PV display. RADIO Radio menu PRI RSSI 28 Primary receive signal strength. Read only. Signal strength 00 is too weak to connect to the network. Displayed Value Value dBm Rx Margin dB 00 < -86 < 10 01 -86 to -81 10 to 15 02 -80 to -75 16 to 21 03 -74 to -69 22 to 27 04 -68 to -63 28 to 33 05 -62 to -57 34 to 39 06 -56 to -51 40 to 45 07 -50 to -45 46 to 51 08 -44 to -11 52 to 85 09 ≥ -10 Saturation SEC RSSI Secondary receive signal strength. Same as PRI RSSI. Read only. WFN ID Wireless Field Network ID. Read only. DEV ADD Device address. Read only. TX POWER Radio transmit power. Read only. OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 5. Operation 5.6. Authentication device menus 5.6 Authentication device menus Overview Hold the Authentication Device no more than 6” (15 cm) from the transmitter and aim the infrared beam at the transmitter display while tapping on the screen command or button. Main menu The main menu is shown below. Details start on the next page. Figure 3 Main menu Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 29 5. Operation 5.6. Authentication device menus Security and Node Deployment Use this to: • receive new security keys, • transmit security keys for connecting the transmitter (or other nodes) to the OneWireless network, • clear all security keys from the PDA, • clear the transmitter’s key and reset its configuration to factory default (such as for decommissioning). Figure 4 Security and Node Deployment 30 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 5. Operation 5.6. Authentication device menus To connect your transmitter to the OneWireless network perform the following steps. Step Action If the PDA contains no keys, obtain new security keys from the PC application Key Server Manager. To do this, select Receive Security Keys. Keys can be received either through Infrared (by aiming PDA at the infrared dongle) or through an ActiveSync/USB connection. See Key Server Communication Method under Advanced options on page 35 for details. Important: The Comm Method settings must match in the PC’s Key Server Manager and in the Authentication Device (both must be set to Infrared or both to ActiveSync) in order for your PDA to receive security keys. See Key Server Communication Method under Advanced options on page 35 for details. When the Authentication Device has valid unexpired keys, aim it at the transmitter and transmit a key to the transmitter. The transmitter will validate the key and then use it to make a connection to the OneWireless Network. The Transmitter may continue to show the diagnostic message “NO KEY” for a brief time while it validates the key before showing the “DISCOVER” message. To verify your transmitter has been authenticated, see the Connection prompt on the Read Node Info screen (page 33). To decommission your transmitter from the OneWireless network, select Clear Key and Restart Node. This clears the transmitter’s key, network and security configurations, and resets the transmitter to its factory default settings. perform the following steps. Select Clear Keys from Handheld (under Advanced Options) when: • The PDA has keys from one system, but you have moved your Authentication Device to another system, or • you want to clear all keys so that you cannot deploy any more keys without going to the key server manager and getting more. For more details on keys, refer to Getting Started with Honeywell OneWireless Solutions. Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 31 5. Operation 5.6. Authentication device menus Device Local Configuration Use Device Local Configuration buttons (Table 10) to navigate the transmitter menus (Table 9) and to make selections and changes. You can also use the PDA buttons. Figure 5 Device Local Configuration screen Table 10 Buttons for Device Local Configuration Button Function • Enter the Menu Tree. • Enter submenu of the menu that is appearing on the screen. • Execute action. • Submit the entered number while doing number entry. • Read value of certain displayed parameters. • Go to the next menu in the same level. • View quick view parameters in Normal Display Sequence (PV Display). • During number entry, increment the digit or change +/- sign. • Go to the previous menu in the same level. • View quick view parameters in Normal Display Sequence (PV Display). • During number entry, decrement the digit or change +/- sign. • Go to the upper menu level. • When changing a number value, move cursor to the left/more significant digit, then wrap around to the least significant digit. 32 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 5. Operation 5.6. Authentication device menus Read Node Information Use this to read the transmitter’s information shown in Figure 6. Similar to quick view parameters on the transmitter display. (See page 27.) Figure 6 Read Node Information Table 11 Read Node Information Item Description Tag The name given to this transmitter Serial Transmitter serial number. This is the WBSN on the transmitter’s nameplate. Do not confuse this with the other nameplate item marked “Serial.” NwAddr Network Address of the device in hexadecimal. DevRev Device Revision. This parameter changes whenever objects and parameters are added, deleted, or their data type or range changes. It does not change if the application firmware changes without affecting the device description. Range: 0 to 65535. Build Sensor firmware and radio firmware build numbers. Radio Hardware radio type, FHSS or DSSS WFN ID: Wireless Field Network ID. Range: 0 to 255. Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 33 5. Operation 5.6. Authentication device menus Item FH Mode Description Frequency group or frequency channel selection used by the wireless network of the device. The value must match the value set in the gateway and interface nodes to allow communication between the device and the wireless network. Modes: US Channel #1 US Channel #6 US Channel #11 US Guard Bands EU Channel #1 EU Channel #7 EU Channel #13 EU Guard Bands US/EU Spec Div A US/EU Spec Div B US/EU Channel #3 US/EU Channel #10 Complete Spectrum Connection The first line displays one of the following connection states. No Security Key – No security key has been deployed to the device or multinode. The user must give a security key to the device or multinode before it will join the wireless sensor network. No Connection – A security key exists in the device or multinode, but no connection has been formed. The device or multinode is waiting to form a connection and will automatically retry shortly. Users may transmit a new security key in order to force the device or multinode to immediately retry to form a connection. Discovering – The device is attempting to form a connection to the wireless sensor network. The device is discovering multinodes and, if a multinode is found, will transition to the securing state. Securing – The device is attempting to form a connection to the wireless sensor network. The device has discovered one or two multinodes and is attempting to form a secure session. If successful, the device will transition to the connected state. Connected – A secure connection is formed with one or two multinodes. The second line contains detailed state information useful for problem reporting. 34 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 5. Operation 5.6. Authentication device menus Advanced Options Advanced options are non-typical configuration commands. Figure 7 Advanced Options Table 12 Advanced Options Item Description Key Server Communication Method Determines how the PDA will receive security keys from the PC’s Key Server Manager application. From the Comm Method menu select one of the following methods. ActiveSync – Select this to receive keys over a USB connection, such as while the PDA battery is being charged in its base. Infrared – Select this to receive keys over the infrared port. Important: The Comm Method settings match in the PC’s Key Server Manager and in the Authentication Device (both must be set to Infrared or both to ActiveSync) in order for your PDA to receive security keys. Read Tracelog Flag Not available for transmitters. Used with multinodes. Reads conditional tracelog flag value. Tracelog flags are used to enable and disable logging functionality used for field support. Write Tracelog Flag Not available for transmitters. Used with multinodes. Writes conditional tracelog flag value. Tracelog flags are used to enable and disable logging functionality used for field support . Select Infrared Communication Port Overrides the detected infrared communication port detected on your PDA. If infrared communication is not functioning, you can override the detected settings using this option. Read TX Power Level Reads the transmission power level of the transmitter radio. Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 35 6. Maintenance/Repair 6.1. Parts 6. Maintenance/Repair 6.1 Parts The following replacement parts may be ordered from Honeywell. Part number Qty. Description 50015866-501 ELECTRONICS MODULE ASSEMBLY aka SENSOR MODULE for Temperature 50015866-505 ELECTRONICS MODULE ASSEMBLY aka SENSOR MODULE for TemperatureInstrinsically Safe 50015866-509 ELECTRONICS MODULE ASSEMBLY aka SENSOR MODULE for TemperatureDSSS Intrinsically Safe 50015866-513 ELECTRONICS MODULE ASSEMBLY aka SENSOR MODULE for Temperature & DI-Instrinsically Safe 50015866-514 ELECTRONICS MODULE ASSEMBLY aka SENSOR MODULE for Temperature & DI-DSSS Instrinsically Safe 50015866-517 ELECTRONICS MODULE ASSEMBLY aka SENSOR MODULE for Temperature & DI 50015843-501 TERMINAL BOARD 50015623-501 CAP ASSEMBLY, BATTERY, ALUMINUM, DARK BEIGE 50016190-501 CAP ASSEMBLY, LCD, ALUMINUM, DARK BEIGE 50026009-501 CAP ASSEMBLY, BATTERY, STAINLESS STEEL 50026127-501 CAP ASSEMBLY, LCD, STAINLESS STEEL 50016229-501 ANTENNA ASSEMBLY, 2 dBi INTEGRAL RIGHT-ANGLE, ALUMINUM 50016229-502 ANTENNA ASSEMBLY, 2 dBi INTEGRAL RIGHT-ANGLE, STAINLESS STEEL 50020767-501 ANTENNA ASSEMBLY, 2 dBi INTEGRAL STRAIGHT, STAINLESS STEEL 50020767-502 ANTENNA ASSEMBLY, 2 dBi INTEGRAL STRAIGHT, ALUMINUM 50031714-501 ANTENNA ASSEMBLY, 4 dBi INTEGRAL STRAIGHT, STAINLESS STEEL 50031714-502 ANTENNA ASSEMBLY, 4 dBi INTEGRAL STRAIGHT, ALUMINUM 50031715-501 ANTENNA ASSEMBLY, 4 dBi INTEGRAL RIGHT-ANGLE, ALUMINUM 50031715-502 ANTENNA ASSEMBLY, 4 dBi INTEGRAL RIGHT-ANGLE, STAINLESS STEEL 50018414-001 REMOTE OMNI-DIRECTIONAL ANTENNA, 8 dBi 50018415-001 REMOTE DIRECTIONAL ANTENNA, 14 dBi 50016577-501 ANTENNA ADAPTER ASSEMBLY, REMOTE, TYPE TNC, STAINLESS STEEL 50016577-502 ANTENNA ADAPTER ASSEMBLY, REMOTE, TYPE TNC, ALUMINUM 50028364-501 ANTENNA ADAPTER ASSEMBLY, REMOTE, TYPE N, ALUMINUM 50028364-502 ANTENNA ADAPTER ASSEMBLY, REMOTE, TYPE N, STAINLESS STEEL 50018110-001 COAX CABLE ASSY, 1.0M ( 3.3 Ft) LONG, RP-TNC - N-MALE 50018110-002 COAX CABLE ASSY, 3.0M (10.0 Ft) LONG, RP-TNC - N-MALE 50018110-003 COAX CABLE ASSY, 10.0M (33.0 Ft) LONG, RP-TNC - N-MALE continued 36 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 6. Maintenance/Repair 6.2. Replacing batteries Part number Qty. Description 50018278-001 COAX CABLE ASSY, 1.0M ( 3.3 Ft) LONG, N-MALE - N-MALE 50018278-002 COAX CABLE ASSY, 3.0M (10.0 Ft) LONG, N-MALE - N-MALE 50018278-003 COAX CABLE ASSY, 10.0M (33.0 Ft) LONG, N-MALE - N-MALE 50018279-090 LIGHTNING SURGE ARRESTOR 50025288-001 BATTERY HOLDER, PRESSURE 50025288-002 BATTERY HOLDER, TEMPERATURE, HLAI, CORROSION 50026010-001 3.6V LITHIUM THIONYL CHLORIDE (Li-SOCI2) BATTERY 50026010-002 3.6V LITHIUM THIONYL CHLORIDE (Li-SOCI2) BATTERY 50026010-003 10 3.6V LITHIUM THIONYL CHLORIDE (Li-SOCI2) BATTERY 6.2 Replacing batteries When to replace When the transmitter displays a LO BATT message you have 2-4 weeks to replace both batteries before they expire. When batteries are removed or expired, all transmitter data is retained in the transmitter’s nonvolatile memory. Tools required • #1 Phillips Screwdriver or 1/8” Slotted Screwdriver • Torque Screwdriver • 1.5 mm hex key Procedure ATTENTION Batteries must be replaced only by a trained service technician. WARNINGS • Risk of death or serious injury by explosion. Do not open transmitter enclosure when an explosive gas atmosphere is present. • Batteries must not be changed in an explosive gas atmosphere. • The batteries used in this device may present a risk of fire or chemical burn if mistreated. Do not recharge, disassemble, heat above 100°C (212°F), or incinerate. SHOCK HAZARD Depending on your installation, transmitter input wiring sources may contain high voltage. Disconnect all power from transmitter input sources before accessing the batteries. Failure to do so could result in death or serious injury if the input terminals or wires are accidentally touched. Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 37 6. Maintenance/Repair 6.2. Replacing batteries Figure 8 Battery replacement Table 13 Battery replacement procedure Step Action ATTENTION You must replace both batteries. Both batteries must be the same model from the same manufacturer. Mixing old and new batteries or different manufacturers is not permitted. Use only the following 3.6V lithium thionyl chloride (Li-SOCl2) batteries (non-rechargeable), size D. No other batteries are approved for use in XYR 6000 Wireless Transmitters. 38 • Xeno Energy XL-205F • Eagle Picher PT-2300H • Tadiran TL-5930/s • Honeywell p/n 50026010-001 (Two 3.6V lithium thionyl chloride batteries) (1 transmitter) • Honeywell p/n 50026010-002 (Four 3.6V lithium thionyl chloride batteries) (2 transmitters) • Honeywell p/n 50026010-003 (Ten 3.6V lithium thionyl chloride batteries) (5 transmitters) Loosen the M3 locking set screw on the battery end-cap (opposite end from display). See item 1 in Figure 8. Unscrew and remove the end cap. Using thumb and forefinger, squeeze the battery connector at top and bottom to disengage the locking mechanism, then pull to disconnect. See item 2 in Figure 8. Loosen the two battery holder retaining screws (closest to the batteries). See item 3 in Figure 8. The screws are captive. Pull the battery holder out of the transmitter. Remove the old batteries from the battery holder. If needed, pry out the batteries by using a slotted screwdriver as a lever in the holder’s side slots. See item 4 in Figure 8. OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 6. Maintenance/Repair 6.3. Replacing sensor module Step Action Insert the new batteries using correct polarity shown on the battery holder. Insert the battery holder into the transmitter. Reattach the screws and tighten to 0,4 – 0,6 N-M (3.5 – 5.3 Lb-in). Re-connect battery connector. Honeywell recommends lubricating the end cap O-ring with a Silicone Grease such as Dow Corning #55 or equivalent before replacing the end cap. 6.3 Screw the end cap back on and tighten the M3 locking screw. Dispose of used battery promptly per local regulations or the battery manufacturer’s recommendations. Keep away from children. Do not disassemble and do not dispose of in fire. Replacing sensor module When to replace Various error messages can help you diagnose a faulty module. These are described elsewhere in this manual. Tools required • #1 Phillips Screwdriver or 1/8” Slotted Screwdriver • Torque Screwdriver • 1.5 mm hex key Procedure WARNING Risk of death or serious injury by explosion. Do not open transmitter enclosure when an explosive gas atmosphere is present. CAUTION Take precautions against electrostatic discharge to prevent damaging the sensor module. Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 39 6. Maintenance/Repair 6.3. Replacing sensor module Table 14 Sensor module replacement Step Action Honeywell recommends that the transmitter be removed from service and moved to a clean area before servicing. Loosen the M3 locking set screw on the display end-cap. See item 1 in Figure 9. Unscrew and remove the end cap. Loosen the two screws on the sensor module. See items 2 in Figure 9. Disconnect each connector on the sensor module. See items 3 in Figure 9. Install new sensor module. Be sure to orient sensor module in the proper viewing orientation before tightening two sensor compartment screws. Reverse steps 1-4. Torque screws to 0,4 – 0,6 N-M (3.5 – 5.3 Lb-in). Honeywell recommends lubricating the end cap O-ring with a Silicone Grease such as Dow Corning #55 or equivalent before replacing the end cap. Clean the transmitter’s nameplate and affix WBSN label over the existing WBSN (Wireless Builder Serial Number). Return transmitter to service. Figure 9 Sensor module removal and replacement 40 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 6. Maintenance/Repair 6.4. Replacing antenna 6.4 Replacing antenna Tools required • #1 Phillips Screwdriver or 1/8” Slotted Screwdriver • Torque Screwdriver • 1.5 mm hex key Procedure ATTENTION You must replace your antenna with the same type, that is, right-angle, straight, or remote. Changing to a different antenna type is not permitted by approval agencies. CAUTION Take precautions against electrostatic discharge to prevent damaging the sensor module. WARNING POTENTIAL ELECTROSTATIC CHARGING HAZARD The integrally mounted antenna shroud is made of Teflon® and has a surface resistance greater than 1Gohm per square. When the XYR 6000 transmitter is installed in potentially hazardous locations care should be taken not to electrostatically charge the surface of the antenna shroud by rubbing the surface with a cloth, or cleaning the surface with a solvent. If electrostatically charged, discharge of the antenna shroud to a person or a tool could possibly ignite a surrounding hazardous atmosphere. Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 41 6. Maintenance/Repair 6.4. Replacing antenna Table 15 Antenna replacement procedure Step 42 Action Honeywell recommends that the transmitter be removed from service and moved to a clean area before servicing. Loosen the M3 locking set screw on the display end-cap. See item 1 in Figure 10. Unscrew and remove the front end cap. Loosen the two screws on the sensor module. See items 2 in Figure 10. Remove the sensor module from the transmitter body and disconnect the antenna connector from CN2 connector on the sensor module. See item 3 in Figure 10. Loosen the locking set screw at the antenna base. Unscrew the antenna from the transmitter. Remove the antenna and its connector from the transmitter. See Figure 10. Feed the new antenna’s connector through the antenna hole to the front of the transmitter. Do not connect to sensor module yet. Lubricate with Silicone Grease such as Dow Corning #55. Screw new antenna into transmitter body until finger-tight, then back off 180 degrees to permit adjustment later. Attach antenna connector to CN2 connector on sensor module. See item 3 in Figure 10. Insert sensor module. Orient in the proper viewing orientation before tightening two sensor compartment screws. See items 2 in Figure 10. Torque screws to 0,4 – 0,6 N-M (3.5 – 5.3 Lbin). Replace the front end cap. Honeywell recommends lubricating the front end cap O-ring with a Silicone Grease such as Dow Corning #55 or equivalent before replacing the end cap. 10 Adjust antenna for best reception. Don’t rotate antenna more than 180 degrees either direction or you could twist and break the antenna wiring inside. Tighten the antenna locking set screw. OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 6. Maintenance/Repair 6.4. Replacing antenna Antenna connector Figure 10 Antenna replacement Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 43 6. Maintenance/Repair 6.5. Replacing terminal board 6.5 Replacing terminal board When to replace Various error messages can help you diagnose a faulty terminal board. These are described elsewhere in this manual. Tools required • #1 Phillips Screwdriver or 1/8” Slotted Screwdriver • Torque Screwdriver • 1.5 mm hex key Procedure WARNING Risk of death or serious injury by explosion. Do not open transmitter enclosure when an explosive gas atmosphere is present. Table 16 Terminal board replacement procedure Step Action Honeywell recommends that the transmitter be removed from service and moved to a clean area before servicing. SHOCK HAZARD Depending on your installation, transmitter input wiring sources may contain high voltage. Disconnect all power from transmitter input sources before accessing the terminal board. Failure to do so could result in death or serious injury. 44 Loosen the M3 locking set screw on the battery end-cap (opposite end from display). See item 1 in Figure 11. Unscrew and remove the end cap. Using thumb and forefinger, squeeze the battery connector at top and bottom to disengage the locking mechanism, then pull to disconnect. See item 2 in Figure 11. Loosen the two battery holder retaining screws (closest to the batteries). See item 3 in Figure 11. The screws are captive. Pull the battery holder out of the transmitter. OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08 6. Maintenance/Repair 6.5. Replacing terminal board Step Action Disconnect field wiring from terminal board and label it to ease reconnection. Remove and save the 3 screws that attach the terminal board to the housing. Take care because these screws are not captive. See item 4 in Figure 11. Remove terminal board by disconnecting cable from back of the terminal board. Do not pull on the wires or you could damage them. Instead, depress the latch while pulling on the connector. Attach connector to new terminal board. Observe correct polarity of the connector. Verify that the cable is latched to the terminal board. 10 Fasten terminal board with screws from step 7. 11 Re-connect field wiring. 12 Insert the battery holder into the transmitter. Reattach the screws and tighten to 0,4 – 0,6 N-M (3.5 – 5.3 Lb-in). Re-connect battery connector. Honeywell recommends lubricating the end cap O-ring with a Silicone Grease such as Dow Corning #55 or equivalent before replacing the end cap. 13 Screw the end cap back on and tighten the M3 locking screw. Figure 11 Terminal board replacement Rev. 3 6/24/08 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual 45 6. Maintenance/Repair 6.5. Replacing terminal board 46 OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual Rev. 3 6/24/08
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.3 Linearized : No Create Date : 2008:09:04 10:16:01-07:00 Modify Date : 2008:09:04 10:16:01-07:00 Page Count : 60 Creation Date : 2008:09:04 17:16:01Z Mod Date : 2008:09:04 17:16:01Z Producer : Acrobat Distiller 5.0.5 (Windows) Metadata Date : 2008:09:04 17:16:01ZEXIF Metadata provided by EXIF.tools