Intuitive Surgical CHB01 RFID TRANSCEIVER 3D-HD CAMERA HEAD User Manual da Vinci Si

Intuitive Surgical, Inc. RFID TRANSCEIVER 3D-HD CAMERA HEAD da Vinci Si

User Manual Part 3

Download: Intuitive Surgical CHB01 RFID TRANSCEIVER 3D-HD CAMERA HEAD User Manual da Vinci Si
Mirror Download [FCC.gov]Intuitive Surgical CHB01 RFID TRANSCEIVER 3D-HD CAMERA HEAD User Manual da Vinci Si
Document ID2607926
Application IDzRpybRLK0a/UrE0oKN152Q==
Document DescriptionUser Manual Part 3
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize349.79kB (4372377 bits)
Date Submitted2015-05-08 00:00:00
Date Available2015-05-08 00:00:00
Creation Date2015-04-29 15:41:39
Producing SoftwareAcrobat Distiller 11.0 (Windows)
Document Lastmod2015-04-29 15:44:03
Document Titleda Vinci Si User Manual.book
Document CreatorFrameMaker 11.0.2
Document Author: cynthiag

Appendix A: Error Handling
After Disabling an Instrument Arm
After you disable an instrument arm, you still can use the arm clutch and port clutch buttons
to move the arm out of the way.
In the event of an error specific to the ICB, the system presents the option to disable the ICB on
the touchscreen and touchpad. Once the ICB has been disabled, it can not be re-enabled until
the next power cycle. Intuitive Surgical designed this feature to allow a user to complete a
procedure without use of the EndoWrist instruments that employ the ICB, such as the Vessel
Sealer.
Non-Recoverable Faults
If a fault is non-recoverable, the system must be restarted. The following message is displayed:
Non-recoverable fault: XXXX
Restart System to continue.
Restarting the System During a Procedure
If a non-recoverable fault occurs during a procedure, you must completely remove all
instruments from the system. The endoscope does not need to be removed. Follow these
steps to restart the system:
1. Completely remove all instruments from the system. The endoscope does not need to be
removed. If an instrument is grasping tissue, follow the grip release instructions in
Chapter 9, Grip Release, on page 9-13.
WARNING: If it is not clinically possible to remove an instrument, closely monitor the
instrument arm during restart to ensure that no motion occurs.
2. Power off the system: Press the Power button on any system component.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Disabling the Instrument Control Box (ICB)
The system takes several seconds to shut down. When complete, all system Power buttons will be lit amber, indicating standby mode, and readiness for restart.
3. Restart the system: Press the Power button on any system component.
4. After the system has restarted successfully, then the instruments can be reinserted.
Note: During system restart, video is temporarily unavailable at the Surgeon Console
viewer and touchscreen monitor.
Note: If the fault cannot be cleared by a system restart, call Intuitive Surgical Technical
Support.
Emergency Stop
Press the red Emergency Stop button should it be necessary to stop system operation at any
time. The Emergency Stop button will cease robotic control of the instruments and
endoscope. The instruments and endoscope will stay in their last commanded position.
If the instrument grips are closed when the Emergency Stop button is pressed, the grips will
remain closed. However, the gripping force of the instrument may decrease.
Pressing Emergency Stop initiates a recoverable fault, which you can override by pressing
Recover on the touchscreen or touchpad. The Emergency Stop button illuminates when
pressed and remains illuminated until the fault is recovered.
da Vinci® Si™
A-3
A-4
Appendix A: Error Handling
Figure A.3 EPO button on rear of Patient Cart
The Emergency Power Off (EPO) button is on the back of the Patient Cart. Press this button to
completely remove power to the Patient Cart. The system classifies this a non-recoverable
fault. The system must be restarted.
Battery Backup
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
EPO (Emergency Power Off)
Should the Patient Cart be unplugged, the system will generate a recoverable fault that must
be addressed to continue a procedure. System operation will be allowed to continue on
reserve power, but with basic functionality only.
Note: Battery backup is only intended for safe removal of the system components from
the patient and is not intended for continuing the procedure.
Battery Low Condition
If there is insufficient battery backup power on the Patient Cart, cart drive is disabled and the
user will have to wait for the battery backup to charge.
• To move the cart manually, move the shift switches on the base of the cart to the neutral
position. When finished moving the cart, be sure to move the shift switches back to the
drive position.
Note: The backup battery is not user-serviceable, and must be replaced by authorized
personnel only. Contact Intuitive Surgical Technical Support for details.
Note: The Patient Cart battery should be adequately charged. If not, an error message
appears on the monitors. You can override the error if the Patient Cart is plugged into AC
power.
da Vinci® Si™
Appendix A: Error Handling
A.3 Conversion to Open Surgery
If a situation arises where a conversion to open surgery is required, perform the following
steps to remove the system from the patient:
1. Remove the instruments and endoscope from the patient. Note the following:
a. In case of system failure while the instrument is grasping tissue, the grips can be
manually opened by following the grip release instructions, see Grip Release on page
9-13 (Chapter 9, Patient Cart Use).
WARNING: Do not perform grip release on a non-faulted system without first pressing
the Emergency Stop button. Failure to observe this warning may result in unintended
instrument motion or damage to the grip release mechanism.
WARNING: Rotating the grip release tool too far and/or in the incorrect direction can
cause unintended instrument motion or damage to the grip release mechanism.
2. Disconnect the cannulae from the instrument and camera arms.
3. Move the instrument and camera arms away from the patient.
Note: If the system is in a fault state while converting to open surgery, the Patient Cart
will still allow use of the port clutch buttons. If the system loses all power, the arms and
setup joints may be overpowered to move the arms as necessary.
_________________________________End of section______________________________
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Note: Whenever possible, use Surgeon Console control to release the instrument grips.
da Vinci® Si™
A-5
B-1
Appendix B: da Vinci Si-e Surgical System
Appendix B: da Vinci Si-e Surgical System
This appendix provides detailed information and specifications for the da Vinci Si-e Surgical
System, an upgradable configuration of the da Vinci Si System, visibly distinguished by a 3-arm
Patient Cart. The da Vinci Si-e System is designed to be upgradable anytime to a full-featured
da Vinci Si System (single or dual console) – by Intuitive Surgical technicians. This section
describes the characteristics that distinguish it from the da Vinci Si System.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
B.1 System Component Compatibility
The da Vinci Si-e System uses the same Surgeon Console, which is interchangeable with any
da Vinci Si System. In contrast, the 3-arm Patient Cart and the Vision Cart of the da Vinci Si-e
System are not interchangeable with the 4-arm Patient Cart and Vision Cart of a da Vinci Si
System; the specific da Vinci Si-e System components must be used together for the
da Vinci Si-e System to work. The system software recognizes when you connect an
incompatible combination of Patient Cart and Vision Cart, notifies you on screen and prevents
use of the disallowed combination.
Note: The da Vinci Si-e System does not support dual console surgery.
da Vinci® Si™
Appendix B: da Vinci Si-e Surgical System
The da Vinci Si-e System supports use of external monitors in high definition or standard
definition, by means of the standard video out connectors on the back of the Core, the
Surgeon Console, and the Camera Control Unit (CCU). The table below describes the available
video output options on the back of the Core. These are not user-configurable: you cannot
select the video output format of the Video Out bay Aux connectors (back of the Core). The
da Vinci Si-e System selects the appropriate output format based on the device connected to
the Aux connector. See section 4.5 Video and Audio Connections and section H.5 Video Patch
Panels for more details.
Table B-1 Si-e Video Connections
Component
Connector Output
Format
Video Out bay Aux, back of DVI (analog and digital)
Core
Resolution
Overlay
Composite (analog)
S-Video (analog)
XGA, SXGA, WXGA+ or
720p, automatically
configured
NTSC or PALa
NTSC or PALa
Surgeon’s view
Surgeon’s view
Surgeon’s view
SD-SDI (digital)
NTSC or PALa
Surgeon’s view
a. NTSC or PAL is standard definition and is determined by country.
The Video Out bay Aux supports only one video format at a time.
Note: Video outputs make available only the surgeon’s view overlays. No external
monitor used with the da Vinci Si-e System can support the touchscreen overlays nor
functionality of the da Vinci Si System.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Use of Third-Party Monitors
Note: If the system has OnSite installed, the OnSite status icons will be present on the
Vision Cart monitor even though the touchscreen function is not available. All other
OnSite features are supported on the Si-e System.
da Vinci® Si™
B-2
B-3
Appendix B: da Vinci Si-e Surgical System
B.2 da Vinci Si-e Differences
Users of the da Vinci Si-e System should note the following differences in features and
behavior compared to the da Vinci Si System.
The da Vinci Si-e System has only two instrument arms, as reflected on the touchpad display:
Figure B.1 Two instrument arms appear on touchpad
Audio System
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Two Instrument Arms
Since the monitor includes a microphone and speakers, it provides support for two-way audio
communication between the surgeon and patient-side assistant. For the da Vinci Si-e System,
the volume control slider for the Vision Cart speakers is found on the Audio tab of the
touchpad, to the right of the Surgeon Console speaker control. Note that there is no
microphone mute button; to mute the microphone, drag the slider all the way to the left, as
shown.
da Vinci® Si™
Vision Cart
speaker volume
Surgeon Console
speaker volume
Figure B.2 Speaker volume control is on touchpad Audio tab
TilePro Not Available
TilePro (multi-image) mode is not available with the da Vinci Si-e System, and thus the option is
not present on the Display Preferences screen of the touchpad.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Appendix B: da Vinci Si-e Surgical System
Figure B.3 TilePro not present
da Vinci® Si™
B-4
B-5
Appendix B: da Vinci Si-e Surgical System
Figure B.4 TilePro QuickClick option not present
Telestration Not Available
Since telestration is done on the touchscreen, telestration is not possible with the da Vinci Si-e
System. However, note that the selected Display Eye option on the Display Preferences screen
(Figure B.3) does still determine whether the surgeon’s left (L) or right (R) eye image from the
stereo viewer passes out of the video connectors of the Core’s Video Out bay Aux.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Furthermore, the QuickClick option for TilePro activation is not offered on the touchpad
Control Preferences screen.
da Vinci® Si™
Appendix B: da Vinci Si-e Surgical System
Camera / Scope Setup via Touchpad Only
Figure B.5 Camera / Scope Setup on the touchpad
Follow these steps to calibrate the endoscope assembly from the touchpad of a da Vinci Si-e
System:
1. The sterile person should insert the endoscope tip fully inside the endoscope alignment
target, using the proper hole, which depends on the tip angle, so that the target
crosshairs are visible on the center of the stereo viewer.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
On the Si-e System, no touchscreen dictates that camera / scope setup must be done via the
touchpad. This circumstance also requires two people to perform calibration: one sterile
person to handle the endoscope and a second non-sterile person to work the touchpad at the
Surgeon Console.
Note: For 3D calibration to be successful, the crosshairs must be well centered on screen
and the target must be kept as still as possible on the endoscope.
2. The non-sterile person at the touchpad: From the Video tab, go to Camera / Scope
Setup and then touch the 3D Calibration button. The button name changes to Finish
Calibration and “Adjust as necessary” appears next to it, and the button and all arrow
buttons flash to prompt your input.
da Vinci® Si™
B-6
Appendix B: da Vinci Si-e Surgical System
Figure B.6 3D Calibration in progress
3. Touch the arrows on the touchpad to move the green crosshairs until aligned with the
magenta crosshairs, as seen in the stereo viewer.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
B-7
4. To save the calibration setting and exit calibration mode, touch Finish Calibration.
3D Calibration and Camera Head Button Functionality
For the da Vinci Si-e System, the camera head buttons do not support 3D calibration. Without a
touchscreen, you must perform 3D calibration from the touchpad, as described above. The
Vision Setup button, in particular, supports no functionality at all; when you press it the
system gives an error beep, but does not display a message. Nothing happens except the error
beep. The Focus In and Out arrow buttons still support focusing of the surgical image from
the camera head, and the Lamp On/Off button still functions.
_________________________________End of section______________________________
da Vinci® Si™
Appendix C: Illuminator Information
C Appendix C: Illuminator Information
C.1 General Safety Precautions
Before operating, read all safety instructions. See Endoscopic Procedure Precautions on page
10 for additional safety instructions regarding use of the Illuminator. The Illuminator is a
source of high electrical voltage, intense light and heat. When used properly and with normal
precautions, it is a safe and effective light source.
CAUTION:
Third party light guides may not withstand light output levels of this light source.
Do not operate the light source without lamp module in place.
Disconnect power supply cord before servicing to avoid electric shock.
To reduce risk of electric shock, do not remove cover. Refer servicing to Intuitive
Surgical personnel.
CAUTION HOT. Do not remove lamp immediately after operation. Allow lamp to cool 5
minutes with fans running before removing power to the Illuminator.
The end of the light guide may be hot.
Keep cooling vents free from obstructions.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
This appendix provides detailed information and specifications for the integrated Illuminator,
also known as the Y1903 Xenon Fiber-Optic Light Source.
To prevent overheating, replace only with the same type and rating of lamp module.
Read instructions before replacing lamp module. (See 12.3 Illuminator Lamp Module
Replacement on page 12-2.)
The following label appears on the side of the lamp module above the removal handle.
Note: It may be necessary for the reader to be as close as 6 in (15 cm) from the label to
read this information.
Caution: Hot
Figure C.1 Lamp module label
Observe the caution statement on the label: “CAUTION: High-pressure lamp may explode if
improperly handled. Refer servicing to qualified service personnel.”
The label provides space to indicate the “SERIAL NO.” and “MODEL NO.” of the lamp module.
“LIGHT OUTPUT
” indicates that the lamp light emits from the side indicated by the
arrow.
da Vinci® Si™
C-1
C-2
Appendix C: Illuminator Information
Refer to Figure 12.2 on page 12-4 to see an image of the lamp module replacement label
affixed to the top of the lamp module.
Figure C.2 Illuminator front features
1. Optic Adapter: Accepts Olympus™ fiber-optic light cables.
2. LED Indicator: Shows the lamp status. Amber: Lamp off; Blue: Lamp on; Blue blinking: No
scope selected or detected.
3. Lamp On/Off Switch: Toggles the lamp on/off once the system has been powered up.
This button switch on the front panel is symbolized by an incandescent lamp. When
pushed, the blue “OFF” flashes on the display until lamp ignition occurs.
4. Display Window: Displays light output level from 0-100 in 10% increments when lamp is
on, and will read OFF when lamp is off.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
C.2 Illuminator Features
5. Lamp Hours: Displays number of usage hours on the lamp module. To read the lamp
hours, press the decrease (–) and increase (+) buttons simultaneously, and read the
number displayed on the display window. You may also view lamp hours by selecting
Utilities > Inventory Management on the touchscreen or touchpad, as described in
sections 7.2 and 10.3.
6. Intensity Control (– +): Control buttons to increase or decrease light output levels in
10% increments.
da Vinci® Si™
7
12
10
11
13
Figure C.3 Illuminator rear features
7. RS232 I/O Serial Port: 9-pin D Sub-Miniature interface for RS232 control features.
Labeled “Illuminator Control.”
8. Input Power Module: Consists of the Power On/Off switch, fuse drawer, and AC input
power receptacle.
9. Power On/Off Switch: The Power On/Off Switch is located on the back panel. When
switched ON, the system is energized and initiates standby mode, while the LED indicator
(on front) illuminates amber. In addition, the cooling fans run, and the display reads OFF.
When switched OFF, the system is de-energized, the LED indicator (on front) is not
illuminated, and the display is dark. Energizing the system does not automatically turn
the lamp on.
10. Fuse Drawer: The fuse drawer is located on the back panel beside the AC input power
receptacle. The fuse drawer contains two 6.3 amps main fuses.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Appendix C: Illuminator Information
11. AC Input Power Receptacle: The AC input power receptacle, located on the back panel,
is a three-prong receptacle that accepts a detachable AC power cord.
12. Unit Identifier Label
13. Light Source Label: Shown below.
da Vinci® Si™
C-3
C-4
Appendix C: Illuminator Information
Figure C.4 Light source label example
14. Lamp Module Access Drawer: Allows service technician access to the lamp module for
replacement. (See 12.3 Illuminator Lamp Module Replacement on page 12-2.) By pushing
in on the drawer, the latch mechanism will release, and the drawer will slide forward. To
close the drawer, push it in until the latch catches.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Note: It may be necessary for the reader to be as close as 6 in (15 cm) from the label to
read this information.
14. Lamp module
access drawer
Figure C.5 Lamp module access drawer open
da Vinci® Si™
Appendix C: Illuminator Information
C.3 Basic Troubleshooting
Table C-1 Basic Troubleshooting
Possible Problems
Remedy
•
•
No power to Illuminator
•
•
Vision Cart not connected or not powered on.
Fuse is blown.
Internal power supply not operating.
AC input power receptacle unplugged.
• Connect and power on system.
• Replace fuse.
• Contact Intuitive Surgical Technical
Support.
•
•
No light emits from unit
•
•
Lamp module access drawer open.
Lamp burned out.
Internal power supply not operating.
Fiber-optic cable not connected.
• Close drawer.
• Replace lamp module.
• Contact Intuitive Surgical Technical
Support.
• Connect fiber-optic cable correctly.
Lamp flickers or dims
Field of view is dim
Lamp is getting old.
Incorrect settings.
Replace lamp module.
Adjust Brightness controls
• Obstructed air intake leads to overheating, causes • Allow unit to cool 10 minutes.
Illuminator turns off
Remove obstructions.
thermal switch to trip.
after a few minutes of
• Fan not running; overheating causes thermal
• Contact Intuitive Surgical Technical
operation.
switch to trip.
Support.
C.4 Fuse Replacement
1. Switch power off on the back of the Illuminator and remove the power cord from the
back of the Illuminator.
2. Check for blown fuses by removing the fuse cover, located next to the three-prong power
receptacle. Carefully pull out the cover using a flat blade screwdriver (medium size) or
equivalent, as shown in Figure C.6.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Symptom
Figure C.6 Remove fuse cover
3. Replace blown fuse(s) with the same size and rating: 6.3A time delay: T6.3A fuses, size
5x20mm.
da Vinci® Si™
C-5
C-6
Appendix C: Illuminator Information
4. Re-install the fuse cover.
5. Reconnect the power cord
6. On the back of the Illuminator, turn the power switch on. The Illuminator should be
operative again.
C.5 Specifications of Y1903 Light Source
Note: The specifications in this section apply to the Y1903 Illuminator only and not the
da Vinci Si System.
Table C-2 Y1903 Light Source Specifications
Category
Specification
Comments
Electrical Input
100 - 240 VAC, 50/60Hz universal, 6.0A input
Located on rear panel, dual fuses
IEC320, 6’ (1.83m), configured for locale
Input Voltage
AC Power Connector
Line Cord
Performance and Features
• 2450 Lumens nominal initial output using
Olympus™ fiber.
• Spectral output 386 - 733nm nominal
• <10% instability p-p through 6 mm glass rod @
0-100Hz
• Beam profile to have “smooth” distribution with no
shadows or sharp peaks
Light Output
All light output specifications
refer to “system only”
performance. Light output via
optical fibers or other optical
components may vary.
Automatic shutdown in the event of overheating
Over-temperature
Protection
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Contact Intuitive Surgical Technical Support if the unit fails to operate properly again.
Unit to become fully operational <3 minutes
(target) after thermal shutdown and all
Overheat Recovery / Auto obstructions to air flow removed at
Cool
environmental temperature of <22 °C (72 °F)
Fiber-Optic Connection
Safety Feature
Fans will remain on in the event
of thermal shutdown when
power is on. Lamp must be
switched on by using Lamp
On/Off switch located on front
panel.
• Lamp will not ignite unless a fiber-optic cable is
Fans will remain on in the
fully inserted into the active port on the turret
absence of a fiber-optic cable
• Lamp power is cut or blocked if fiber-optic cable is inserted into active port when
removed from active port to prevent accidental
power is on
light leakage
Lamp Power Supply
Lamp Module
Lamp Module Replacement
Lamp Life
PS300-12 type
Cermax VQ (300 Watt)
By easy access to lamp module via drawer. No
tools required.
• 1000 hours to 50% of initial output specification
measured through 6 mm glass rod
• >1000 hours at a minimum output of 1225 Lumens
da Vinci® Si™
Intuitive Surgical PN 950093
Lamp replacement drawer
“interlocked” for safety. Lamp
power will be cut when drawer
is opened
Appendix C: Illuminator Information
Table C-2 Y1903 Light Source Specifications
Category
Specification
Comments
User Interface / Control
In this system user manual
Located on front panel
Lamp On/Off Switch
User controlled lamp on/off. Fan
operation independent of lamp
status.
Olympus™ port to fit Olympus fiber
Fiber-Optic Adapter
• Controlled by membrane buttons on front panel
• (+) and (–) buttons for relative intensity increments
Light Attenuation Shutter
• Relative level of illumination indicated by a digital
numeric display (blue numerals)
Displays number of elapsed hours of lamp
operation when you press (+) and (–) buttons at
same time
Lamp Hour Counter
Mechanical & Environmental
Dimensions
Weight
Touch Temperature
Ground Bound
Sterilization
Operating Temperature
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
User instructions
Storage Temperature
Operating & Storage
Humidity
Operating Pressure
Audible Noise
Shipping, Shock &
Vibration
Cooling
Height 5.5" (Without Feet) x Width 15.5" x Depth Designed for modular
≤17” (without front bezel) (14 x 39.4 x 43.2cm)
expansion
≤28 lbs.
Per UL60601 -1
Per UL60601 –1
The light source may be wiped-down with
hospital approved disinfectants (for example,
10% bleach + 90% water solution) applied with a
damp cloth (must not be wet)
6 °C to 35 °C
–20 °C to 75 °C
10 - 80% relative humidity, non-condensing
1 Atmosphere
≤ 40dB
per ISTA 3A
Vents to direct airflow toward the back of the
unit
Regulatory Approvals
•
•
•
•
Compliance to standards •
•
•
•
•
IEC 60601-1:1988+A1:1991+A2:1995+A1.3:1997
UL 60601-1:2003
EN 60601-1:1990+A1:1993+A2:1995+A1.3:1997
EN 60601-1-2:2001
CAN / CSA C22.2 No. 601.1/M90(R1997),B/98,S1-94
ANSI/AAMI ES60601-1: 2005
CAN/CSA-22.2 No. 60601-1 (2008)
EN 60601-1-2: 2007
CE mark
da Vinci® Si™
C-7
C-8
Appendix C: Illuminator Information
C.6 Classification of the Y1903 Light Source
• Class I: The light source relies on connection to the protective earth conductor to prevent
shock hazards.
Note: The da Vinci Si camera head provides isolation in accordance with a CF applied
part and is acceptable for use with the Y1903 Illuminator.
• Provides no protection against ingress of liquids.
• Mode of Operation: Suitable for continuous operation.
• Not suitable for use in the presence of a flammable anaesthetic mixture with air or with
oxygen or with nitrous oxide.
C.7 Electromagnetic Compatibility
The Y1903 has been tested and found to comply with the limits for medical devices to the IEC
60601-1-2:2001(E). These limits are designed to provide reasonable protection against
harmful interference in a typical medical installation. The Y1903 generates, uses and can
radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to other devices in the vicinity. However, there is
no guarantee that interference will not occur in a particular installation. It can be determined
if this equipment causes interference by turning the power to the light source off and on. The
user is encouraged to try to alleviate interference problems by one or more of the following
measures:
• Re-orient or relocate the receiving device
• Increase the separation between the equipment.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
• Type BF: The Y1903 light source is classified as a BF equipment. The optic adapter is
grounded and only BF or CF applied parts should be used with the Y1903.
• Connect the equipment into an outlet on a separate electrical circuit from that of other
devices.
da Vinci® Si™
Appendix C: Illuminator Information
Warnings
• The Y1903 should not be used adjacent to or stacked with other equipment. However, if
adjacent or stacked use is necessary, the Y1903 should be observed to verify normal
operation in the configuration in which it is used.
Manufacturer’s declaration – electromagnetic immunity
The Y1903 is intended for use in the electromagnetic environment specified below. The customer or the user
of the Y1903 should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
• AC power cords other than those provided with the instrument may result in increased
emissions or decreased immunity.
Compliance Level
±6 kV contact
±8 kV air
Complies
±2 kV for power supply
lines ±1 kV for input/
output lines
Complies
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Complies
<5% UT (>95% dip in UT) Complies
for 0.5 cycle
40% UT (60% dip in UT) for
5 cycles
Voltage dips, short
70% UT (30% dip in UT) for
interruptions and voltage
25 cycles
variations on power supply
<5% UT (>95% dip in UT)
input lines
for 5 sec.
IEC 61000-4-11
3 A/m
Complies
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
Note: UT is the AC mains voltage before application of the test level.
da Vinci® Si™
Electromagnetic
Environment Guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the Y1903 requires
continued operation
during power mains
interruptions, it is
recommended that the
Y1903 be powered from
an uninterruptible power
supply or battery.
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment.
C-9
C-10
Appendix C: Illuminator Information
Manufacturer’s Declaration – Electromagnetic Emissions
The Y1903 is intended for use in the electromagnetic environment specified below. The customer or the user
of the Y1903 should assure that it is used in such an environment.
Compliance
Group 1
RF emissions
CISPR 11
Class B
RF emissions
CISPR 11
Harmonic emissions IEC 61000-3-2 Class B
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Complies
Electromagnetic environment guidance
The Y1903 uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
The Y1903 is suitable for use in all establishments,
including domestic and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
_________________________________End of section______________________________
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Emissions test
da Vinci® Si™
Appendix D: VisionBoom™ Use Instructions
D Appendix D: VisionBoom™ Use
This appendix provides instructions to use the da Vinci Si Surgical System installed in the
VisionBoom configuration. Integrators seeking installation instructions should refer to the
VisionBoom™ Installation Guide (PN 550539).
Note: This appendix provides only those instructions specific to the VisionBoom
configuration. Refer to relevant portions of this manual for instructions to use the
surgical system.
The da Vinci Si VisionBoom configuration eliminates clutter and improves efficiency in
operating room (OR) surgical environments by replacing the da Vinci Vision Cart, and its
associated cords and cables, with a convenient ceiling-mounted system.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Instructions
Figure D.1 Recommended side by side (left) and stacked (right) VisionBoom configurations
Note: The VisionBoom upgrade supports dual console surgery.
The ceiling-mounted equipment boom is the primary platform for OR integration and,
depending on the model selected, the typical boom can provide the support and space to
position most necessary clinical devices. The equipment boom is not a product sold by ISI but
by manufacturers such as Berchtold™, Steris™, Skytron™, etc., to name just a few vendors that
sell such equipment.
da Vinci® Si™
D-1
D-2
Appendix D: VisionBoom™ Use Instructions
D.1 General Notes and Cautions
• Air flow sufficient to support adequate cooling of da Vinci vision components is
critical to their proper function. The entire da Vinci System is designed to
undergo an automatic, controlled power-down sequence in case a component
or subsystem overheats while in normal operating mode, thereby preventing
system damage. (See Chapter 5 Startup for details.) To avoid overheating, do not
place anything on or near any da Vinci vision component on the boom,
especially if it might impede air flow. Do not route cables behind the Illuminator
on the boom shelf, to avoid blocking air flow behind it.
• ISI recommends that the boom be oriented during surgery so that air flow from
the components is directed away from the sterile field.
• ISI recommends that da Vinci vision components be left permanently in the
configuration in which they are transferred to the boom by ISI service
personnel. Rearrangement of da Vinci vision components could result in a
configuration that does not support adequate cooling or otherwise may result
in an increased risk of damage to or improper function of the components.
• da Vinci vision components are not designed to support external loads, and
therefore ISI discourages placement of any equipment on top of da Vinci vision
components mounted on the boom shelf.
CAUTION: To avoid overloading circuits, do not connect ancillary devices such as
insufflators or energy devices in common circuits with any system component,
particularly not with the vision components because they have large power
requirements. Ancillary devices must be connected to boom outlets on separate circuits
from all system components.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Note:
CAUTION: After a few minutes of use, the rear of da Vinci vision components may
become hot, particularly the Illuminator. Avoid touching the rear of these units during
use and for 10 minutes after use while components cool with internal fan operation.
D.2 da Vinci Si System Connections
In a VisionBoom configuration, Surgeon Consoles and the Patient Cart connect to the Core via
fiber interface wall plates. These wall plates connect via cabling inside the wall that terminates
in the fiber boom plates, which connect via their blue cables (1 m) directly to the back of the
Core. Surgeon Consoles and the Patient Cart can connect to any da Vinci wall plate in the
room. The Core recognizes the units connected to its fiber input ports automatically. Figure
D.2 illustrates how the system connections are made.
da Vinci® Si™
Appendix D: VisionBoom™ Use Instructions
Wall plate
Fiber cable
da Vinci
Patient Cart
Boom plates
Fiber cable
Wall plate
Equipment boom
with Core
Fiber cable
Fiber conduit through
walls/ceiling
Figure D.2 Fiber cables and conduits connect Core to Patient Cart and Surgeon Consoles
Figure D.3 shows where to find the fiber optic cable connectors on the rear of the Patient Cart
and Surgeon Console.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
da Vinci
Surgeon Console
Figure D.3 Fiber connectors on Patient Cart and Surgeon Console
Connecting the Fiber Cables
Note: The connections on the back of the Core (in Step 4: Connect Core, page 5) generally
are made only once and left connected unless the Core is removed from the boom.
da Vinci® Si™
D-3
D-4
Appendix D: VisionBoom™ Use Instructions
Follow these steps to connect the Patient Cart and one or two Surgeon Consoles to the Core in
the VisionBoom configuration.
Before connecting the blue da Vinci Si fiber cables, pull to remove the protective cap at each
end of the cable (Figure D.4). Note the position of the red alignment mark on the uncapped
cable end, which you must align with a similar mark on the fiber cable port for successful
insertion.
Red alignment mark
End cap on system cable
Uncapped end of cable
Figure D.4 Remove the cable end cap
Step 2: Connect Patient Cart
Connect a blue fiber cable (20 m) to the back of the Patient Cart (Figure D.5) and to the desired
fiber interface wall plate (Figure D.6). When lit solid blue, the LED above the fiber port
indicates a good connection to the Core. (It will not light blue until you complete the
connection from the boom plate to the Core – see Step 4: Connect Core, page 5.)
Blue light shows
good connection
when complete to
Core
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Step 1: Remove Cap
Observe red mark and align
Flip up receptacle cover & insert
Figure D.5 Connect fiber cable to Patient Cart
da Vinci® Si™
Align red marks and
push in to connect
Figure D.6 Connect other end of fiber cable to fiber wall plate
Step 3: Connect Surgeon Consoles
For each Surgeon Console you will use, connect a fiber cable to the fiber connector on the
back of the console (Figure D.7) and to an available fiber wall plate (as in Figure D.6). Again, a
blue LED indicates a good connection to the Core when connections in next step are
complete.
Blue light shows
good connection
when complete to
Core
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Appendix D: VisionBoom™ Use Instructions
Figure D.7 Connect fiber cables to each Surgeon Console
Step 4: Connect Core
Connect each 1 m (3’-3”) cable being used from its boom plate to the back of the Core to
complete the connections for each Surgeon Console and the Patient Cart (Figure D.8).
Blue LEDs
show good
connections
Figure D.8 Connect the boom plate fiber cables to the Core on the boom
da Vinci® Si™
D-5
D-6
Appendix D: VisionBoom™ Use Instructions
Figure D.9 Bypassing wall cabling
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Note: If, after connecting all cables as shown in steps 1 through 4, you still do not have a
good connection (no blue light) connect the long (20 m) blue fiber cables directly to the
core. These blue fiber cables are of sufficient length (20 m) to bypass the wall cabling
and connect directly to the core from the surgeon console and patient cart, as illustrated
in Figure D.9 below.
da Vinci® Si™
Appendix D: VisionBoom™ Use Instructions
D.3 Optional Core Connections
This section describes the additional, optional connections you can make between the Core
and third party devices.
Perform this step for each video connection you wish to make between the da Vinci Si System
and external monitors, recorders or other third party devices.
1. Connect each video output or TilePro input on the back of the Core to the desired
interface plate, monitor, or third party device. Figure D.10 shows a typical connection
from the DVI output to a DVI boom interface plate that supports connection to an OR
video switching system.
Core DVI output
DVI boom plate
Figure D.10 Example of DVI connection from Core to boom plate
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Core Video Connections
Electrosurgical Unit (ESU) Connections
Note: Refer to 4.4 Auxiliary Device Connections, page 4-9, for detailed instructions.
To connect one or more electrosurgical units (ESU), perform this step:
1. Connect the appropriate energy activation cable between any of the Energy receptacles
on the back of the Core and the appropriate connectors on the ESU.
Core Energy connection
Sample ESU connections
Figure D.11 Example of ESU connection to Core
da Vinci® Si™
D-7
D-8
Appendix D: VisionBoom™ Use Instructions
D.4 Camera Head and Cable Storage
Coil camera cables
Hang cables over holster
As when finished
Figure D.12 Camera and cable storage using the boom-mounted holster
D.5 Touchscreen Positioning
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
The camera holster is installed on the boom to provide a convenient location for storing the
camera head (without endoscope attached) when not in use. Figure D.12 illustrates how to
coil the cables and store the camera head.
The da Vinci Si touchscreen mounted on the boom can be positioned to either side of the
boom or directly off the front (Figure D.13). Position it according to the needs of the surgical
staff. Unless a sterile monitor drape is used, a sterile assistant requires a change of surgical
gloves after touching the touchscreen; alternatively, non-sterile surgical staff may operate an
undraped touchscreen.
Figure D.13 You can position touchscreen
Note: Refer to 7.4 Working with the Touchscreen Vision Controls, page 7-15, for
instructions to use the touchscreen.
da Vinci® Si™
Appendix D: VisionBoom™ Use Instructions
Similar to the positioning of the Vision Cart in relationship to the patient, the vision boom
needs to be positioned to a location that is convenient to the surgical staff to have access to
the equipment. The vision boom positioning also must take into consideration the location of
the third da Vinci instrument arm during the specific procedure performed. The boom must be
placed within reach of the 5.75 m (18’-6”) camera cable attached to the front of the Core.
VisionBoom
components
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
D.6 Boom Positioning
Figure D.14 Typical boom positioning
_________________________________End of section______________________________
da Vinci® Si™
D-9
E-1
Appendix E: OnSite™ for da Vinci® Surgical System
Appendix E: OnSite™ for da Vinci® Surgical
System
E.1 General Information
The following appendix is applicable only if your da Vinci Si System has da Vinci OnSite
enabled.
Contact Information
For Customer Service and Reporting of Complaints or Adverse Events
Use the following information for customer service, including ordering, reporting complaints
or adverse events, and general information regarding Intuitive Surgical or our products and
services.
In the U.S.
In Europe:
Intuitive Surgical, Inc.
1266 Kifer Road
Sunnyvale, CA 94086 USA
Toll free: 1.800.876.1310
Direct: 408.523.2100
Fax: 408.523.2377
Intuitive Surgical Sàrl
1, chemin des Mûriers,
1170 Aubonne, Switzerland
Toll free: +800.0821.2020
Direct: +41.21.821.2020
Fax: +41.21.821.2021
For Technical Support
If the system requires maintenance or service, please call our Technical Support line. In the
U.S., call 1-800-876-1310, where phones are staffed 24 hours a day, seven days a week. In
Europe, call +41.21.821.2020.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Manufacturer
Intuitive Surgical, Inc.
1266 Kifer Road
Sunnyvale, CA 94086 USA
www.intuitivesurgical.com
Intuitive Surgical Sàrl
1, chemin des Mûriers,
1170 Aubonne Switzerland
General Precautions, Warnings, and Contraindications
Note: All da Vinci Surgical System users must follow all instructions for use supplied with
the system, its components, instruments, and accessories. This includes the following
documents: Instruments and Accessories User Manual (PN 550675), Reprocessing
Instructions (PN 550875), and any instructions for use (IFUs) provided with instruments
or accessories.
WARNING: Be sure to read and understand all information, particularly the caution and
warning information, found in the applicable user manuals before using these products.
Failure to properly follow all instructions, including those in the da Vinci Surgical System
user manual, and instructions supplied with accessory devices such as generators, may
lead to injury and result in improper functioning of the device.
da Vinci® Si™
Appendix E: OnSite™ for da Vinci® Surgical System
CAUTION: Leakage current from interconnected electrical equipment may exceed safe
levels. To maintain the safety of patients and users, interconnect only with devices in
compliance with IEC 60601-1-1. It is the user’s responsibility to ensure that any
interconnected equipment not supplied by Intuitive Surgical maintains compliance with
IEC 60601-1-1.
CAUTION: Ethernet networks (both wired and wireless) are subject to losses of
connectivity that could disrupt use of OnSite or make data unreliable when it is received
at a remote location. Such disruptions, if they occur, have no effect on the performance
or functionality of the da Vinci Surgical System.
E.2 Indications for Use – OnSite
OnSite for da Vinci Surgical Systems is an accessory indicated for use by trained
Intuitive Surgical Field Service personnel to: (1) obtain system information for the purpose of
diagnosing faults, (2) remotely enable/disable features including configuration updates
through either a wired or wireless Ethernet connection between the da Vinci Surgical System
and the hospital’s Internet Protocol (IP) infrastructure.
E.3 Network Connections
OnSite requires a wired RJ45 Ethernet 10bT/100bT and/or wireless 802.11 network connection
with Internet access where the da Vinci Surgical System will be used.
E.4 Transmitter Module Label
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
CAUTION: OnSite components may be installed and serviced only by Intuitive Surgical
personnel. Do not attempt to install or service equipment without Intuitive Surgical
personnel.
When the optional wireless bridge is installed, the following Federal Communications
Commission (FCC) identification label will be affixed to the Surgeon Console.
Contains Transmitter Module FCC ID: SWX-NS2
Figure 1 Transmitter Module Label
da Vinci® Si™
E-2
E-3
Appendix E: OnSite™ for da Vinci® Surgical System
E.5 Introduction
1. Automated log retrieval, where da Vinci Surgical System uploads logs to an Intuitive
Surgical server when idle
2. Remote system status monitoring
3. Remote diagnostics and servicing
4. Remote configuration changes
5. Enable/disable device features
The monitoring capability enables a faster response time from the da Vinci dVSTAT™ (da Vinci
Surgery Technical Assistance Team) for problem resolution, real time diagnosis, and increased
diagnostic accuracy.
To implement remote service capabilities, the da Vinci Surgical System must have access to the
Internet. OnSite is designed to accomplish this using existing hospital networks.
E.6 OnSite System Requirements and Connections
The remote servicing features of OnSite are designed to be highly secure and to function
transparently. The da Vinci Surgical System communicates with an Intuitive Surgical server via
outgoing network connections to enable Intuitive Surgical service personnel to remotely
monitor and service the system while in use.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
OnSite provides connectivity that enables Intuitive Surgical service personnel to remotely
service the da Vinci Surgical System pre-operatively and intra-operatively. It enables the
following capabilities.
da Vinci® Si™
Appendix E: OnSite™ for da Vinci® Surgical System
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
In summary, OnSite consists of three major components, namely the da Vinci Surgical System
with installed networking components, the Intuitive Surgical server, and the remote user
(Intuitive Surgical Field Service personnel). The block diagram below illustrates the OnSite
networking infrastructure.
Figure 2 OnSite Networking Infrastructure
Note: To take advantage of the full potential of OnSite, the system must remain
connected to the network.
da Vinci® Si™
E-4
E-5
Appendix E: OnSite™ for da Vinci® Surgical System
Wired Network Connection
Figure 3 Network Security Device
To establish a wired connection:
Connect the da Vinci Ethernet connection to the hospital network (wall plate) using a CAT5e
industrial style network cable.
Ethernet to hospital network (wall)
Connects da Vinci System to network
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Intuitive Surgical field service personnel install a network security device inside the da Vinci
Surgical System, along with necessary cables and panels to enable a wired network
connection for OnSite.
Figure 4 Network Cable Connections (da Vinci Si)
da Vinci® Si™
Appendix E: OnSite™ for da Vinci® Surgical System
Optional Wireless Connection
Figure 5 Network Security Device with Wireless Bridge
Note: External connections are not required for the da Vinci Si Surgical System.
5.7 Disabling All Network Connectivity
If there is a need to disable all network connectivity for the da Vinci Si System, open the back of
the Vision Cart and disconnect the RJ-45 (Ethernet) connector at bottom center of the Core,
indicated below.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
There is an optional wireless connection available using the Network Security device and a
Wireless Bridge. See Section E.10 Wireless Connectivity Option and section E.13 OnSite
Appendix C: Wireless Bridge Data for details on wireless connectivity.
Disconnect to disable networking
Figure 6 RJ-45 Connector – Core (da Vinci Si)
Note: This action disables all network connectivity for the da Vinci Si System, but it
does not power off the wired or wireless networking equipment.
Note: To re-establish network connectivity, you must re-connect the indicated RJ-45
connector on the back of the Core.
da Vinci® Si™
E-6
E-7
Appendix E: OnSite™ for da Vinci® Surgical System
E.8 Automatic Status and System Log Retrieval
OnSite provides real-time system status monitoring and post-procedure upload of system
logs, for the support team to service the da Vinci System. When Intuitive Surgical field service
personnel enable the OnSite functionality, the da Vinci Surgical System can:
A. Provide status updates – typically every 10 seconds but can be configured for
different intervals
B. Upload all system logs to the Intuitive Surgical server after each procedure
2. Connect to field service diagnostic applications running on a remote laptop
E.9 OnSite Servicing and Diagnostics
OnSite enables remote servicing using current diagnostic applications that Intuitive Surgical
Field Service personnel normally use when the technician visits on site. When physically
present, the technician troubleshoots the system using a local connection between the laptop
and the da Vinci System hardware. OnSite enables the technician to troubleshoot remotely,
using the same set of diagnostic tools. Through a remote OnSite connection, the technician
can interact with the system in either Normal Mode or Maintenance Mode.
Normal Mode
In Normal Mode, OnSite can only enable remote monitoring of system status. This allows
dVSTAT to passively monitor information transmitted, with no ability to perform any activity
that impacts the movement or performance of the surgical system.
In Normal Mode, dVSTAT can:
• Receive system logs
• Check the condition of system switches and buttons
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
1. Connect to an Intuitive Surgical server for these purposes:
• Verify surgical instrument functionality.
Normal Mode – OnSite Mode Indications
While in Normal Mode, the da Vinci Si System indicates the status of the network connection.
da Vinci Network Offline
da Vinci Network Online
OnSite Session In Progress
Figure 7 OnSite Connection Status Indicators
Note: Once the Ethernet cable is connected, it can take up to two minutes to detect the
da Vinci network and update the status on the touchscreen.
Maintenance Mode
Note: Intuitive Surgical personnel can use Maintenance Mode only when they request it
and are granted verbal permission by OR staff present with the da Vinci Surgical System.
Maintenance Mode is a state where Intuitive Surgical technical support personnel can connect
remotely to the da Vinci System to perform diagnostic and troubleshooting operations.
da Vinci® Si™
Appendix E: OnSite™ for da Vinci® Surgical System
When in Maintenance Mode, the da Vinci Surgical System is not for human use.
Intuitive Surgical technician requests for this service requires facility staff to place the system in
Maintenance Mode at an agreed-upon time.
Putting the System into Maintenance Mode
• All system components are connected to AC power
• Surgeon Console and Patient Cart system cables are connected to the Core
When the system is in Standby Mode, the power buttons on the Surgeon Console, Vision Cart,
and Patient Cart are lit amber. When an Intuitive technician connects to the system, they have
the option to power on the system in Maintenance Mode.
During an OnSite session in Maintenance Mode, the system displays:
Maintenance Mode - Not for Human Use
Remote Session in Progress
System Servicing/ Diagnostics
An example of an OnSite servicing capability that requires assistance and feedback from
hospital personnel is when remotely testing the control and motion of the manipulators and
robotic arms. Refer to the following illustration.
Setup Joint
Telescoping Axis
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
To put the system into Standby Mode, make sure the following conditions are met:
Vertical range
of motion
Figure 8 Setup Joint and Instrument Arm
The following list shows several diagnostic capabilities that an Intuitive Surgical field service
technician can execute remotely when connected in Maintenance Mode via OnSite:
1. Test joints, internal sensors, and positioning potentiometers
2. Check condition of system switches and buttons
3. Check synchronization of system configuration
4. Modify system configurations
5. Perform arm motion and other diagnostic tests
da Vinci® Si™
E-8
E-9
Appendix E: OnSite™ for da Vinci® Surgical System
6. Verify surgical instrument functionality
7. Check usage hour meter data
Wireless Overview
A wireless bridge is installed on the da Vinci Surgical System to enable the Wireless
Connectivity Option. A hospital-supplied Wireless Access Point with Internet access is required
to establish wireless connectivity.
Wireless Network Requirements
Intuitive Surgical field service personnel will install and configure the Wireless Connectivity
Option. Below are the details of a suitable wireless network to support da Vinci wireless
applications.
Specifications
• The Wireless Connectivity Option utilizes the IEEE 802.11 wireless standard using either
802.11b or 802.11g at 2.4 GHz Industrial, Scientific, and Medical (ISM) band.
• The Wireless Bridge operates as a client to the hospital-supplied Wireless Access Point,
transmitting data back and forth between the hospital network and da Vinci applications.
• The Wireless Access Point must be located within 75 feet of the da Vinci Surgical System.
Security
• Wireless Network Infrastructure
•
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
E.10 Wireless Connectivity Option
Intuitive has tested the Wireless Connectivity Option in WPA2 pre-shared key mode
with AES encryption, and recommends that the Wireless Connectivity Option is
integrated into the hospital network using this security configuration.
• OnSite Software Application
•
The OnSite Software Application uses a Secure Socket Layer (SSL) session based on
unique certificates on the da Vinci System and the OnSite server.
•
Data being transmitted from the da Vinci Surgical System to the server is 128-bit
encrypted.
Quality of Service
• Wireless Bridge
•
Maximum latency of 50 ms between the Wireless Bridge and the hospital-supplied
Wireless Access Point
•
Wireless Channel that has 20% or less utilization
• Overall Network
•
Maximum end-to-end packet loss of less than 10%
•
Network latency should not exceed 300 ms
Once successfully installed and configured, Intuitive Surgical field service personnel conduct
an end-to-end functional test to ensure that OnSite functions as expected.
da Vinci® Si™
Appendix E: OnSite™ for da Vinci® Surgical System
It is possible that the wireless network conditions might degrade over time or experience
periods of disturbance; da Vinci applications have been designed to be robust to typical
network disturbances, but if an issue persists, contact Technical Support for assistance to
resolve the issue.
Note: Intuitive Surgical recommends that an active wired port be available when using
the Wireless Connectivity Option. The configuration for the Wireless Connectivity
Option provides a wired backup that the router will automatically activate when
plugged in. Refer to E.11 OnSite Appendix A: IT Requirements for details on how to
establish a wired connection.
Note: It is important to note that if the wireless network is modified or updated after the
Wireless Connectivity Option is installed, its suitability to support the wireless
applications should be re-assessed. In particular, contact Technical Support if any of the
following changes are planned or have occurred.
• If the Wireless Access Point or da Vinci Surgical System is moved from the location where
it resided during installation
• If the Wireless Access Point is replaced with a new make or model
Wireless Coexistence
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Note: After installation, Intuitive Surgical recommends that the hospital routinely
monitor to ensure that the Wireless Channel does not exceed 20% utilization, and the
latency between the Wireless Access Point and the Wireless Bridge does not exceed 50
ms. If either exceeds the specified levels, contact Intuitive Surgical Technical Support.
Wireless coexistence with other devices that transmit in the 2.4 GHz range is a concern since it
can impact the reliability of the wireless link. This section summarizes testing conducted by
Intuitive Surgical in an environment with other wireless devices representative of a typical
Operating Room to demonstrate that the Wireless Connectivity Option functioned as
expected. The test setup represented the worst case da Vinci Surgical System setup, and the
position of the common wireless devices was defined to ensure that they were located near
the Wireless Connectivity Option or the Wireless Access Point, and the path between the
transmitter and receiver for most paired devices passed through the signal path between the
Wireless Connectivity Option and the Wireless Access Point. Testing was conducted with a
wireless network that satisfied the characteristics identified in Wireless Overview, page 9.
The Wireless Access Point used during the testing was the Cisco Aironet 1240AG Series. The
Aironet 1240AG Series was configured to operate as a typical Access Point, and therefore
Wireless Access Points from other vendors should result in the same performance. Note that
the characteristics for a suitable wireless network are summarized in Wireless Overview, and
Intuitive Surgical field service personnel will confirm the wireless network is functioning as
expected after installation. A complete list of the common wireless devices used during the
testing (along with details on position, orientation, and type of data transmission) is
summarized in the table in Common Wireless Devices Tested, page 11.
da Vinci® Si™
E-10
Appendix E: OnSite™ for da Vinci® Surgical System
Note: If different types of wireless devices will be used in the Operating Room, or if the
wireless devices are used in different locations than what is described below, then
Intuitive recommends that performance is tested with the wireless devices active, before
use. If you encounter issues using the Wireless Connectivity Option in the presence of
other wireless devices in the Operating Room, contact Intuitive Surgical Technical
Support.
Note: The wireless coexistence testing conducted by Intuitive does not cover use in the
presence of MRI or diathermy machines. The Wireless Connectivity Option should not be
used in the vicinity of these devices.
Common Wireless Devices Tested
Common Wireless Devices
Disturbance Details
IOGear Model: GUW2015V (receiver)
GUWA200 (transmitter)
Wireless Monitor
3.1 GHz to 4.8 GHz Certified wireless USB
RF Modulation: QPSK/DCM; Data Rate: 480
Mbps
2 iPhone4 (3G and 2.4 GHz wireless)
Samsung (2.4GHz wireless)
Smart Phone/Device
Test Setup
Transmitter attached to a desktop PC
located 50 inches away from the wireless
bridge, and receiver attached to a
monitor on the boom.
Desktop PC oriented so the transmitter
has clear line of sight to the receiver
attached to the monitor, and PC
streaming 720p video.
One iPhone4 sitting on the arm rest of
the Surgeon Console paired with a
Bluetooth headset with a phone call in
progress. The iPhone is also connected to
the WAP. The second iPhone4 is paired
with the Bluetooth speaker.
Samsung phone in the room 72 inches
away from a paired Bluetooth headset
worn by someone at the patient side.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
E-11
Laptops with wireless
802.11 b; 2.4 GHz
Two Dell laptops connected to the WAP
on the same channel as the Wireless
Bridge, with one laptop streaming a
video over the network from YouTube.
The laptops are approximately 90 inches
away from the Wireless Bridge.
Wireless keyboard and mouse
Microsoft Wireless Desktop – Keyboard and
Mouse 7000:
2.4 GHz range(2,400 – 2,483.5 MHz) FCC IDs
C3K1345, C3K1142 and C3K1123
Wireless keyboard and mouse interfaced
with one of the desktop computers, and
physically sitting on top of the Vision
Cart, 55 inches apart.
Bluetooth keyboard
Microsoft Bluetooth Mobile Keyboard 6000
2.4 GHz range(2,400 – 2,483.5 MHz) FCC ID
C3K1390
Keyboard interfaced with one of the
desktop computers, and physically sitting
on top of the Vision Cart, 55 inches apart.
2.402-2.480 GHz range
Jawbone headset paired with the
iPhone4, worn by the surgeon at the
Surgeon Console and used during the
phone call.
Bluetooth headset #1
da Vinci® Si™
Appendix E: OnSite™ for da Vinci® Surgical System
Disturbance Details
Test Setup
Bluetooth headset #2
2.402-2.480 GHz range
Bluetooth headset paired with a
Samsung phone, located on the
operating room bed, 72 inches apart,
with a call active.
Bluetooth Speaker
Creative D100 Wireless; FCC ID IBAMF8090
2.402-2.480 GHz range
iPhone4 paired with the Bluetooth
speaker playing music located on the
operating room bed, 72 inches apart.
Cordless Telephone
Uniden 2.4 GHz Amplified Cordless Phone
System (Clarity-4205)
Phone base is on the desk, and the phone
is on the other side of the room, with the
base and phone continuously
communicating.
In the coexistence test, the microwave
oven is placed in the signal path between
the Wireless Bridge and the WAP, 20 feet
away from the WAP. The Wireless Bridge
and WAP are operating at maximum
distance in this test.
MagicChef Model MCD11E3W
Output Frequency 2450 MHz; FCC ID
C5F7NF1AMO100N
Microwave Oven
In the isolated test with the microwave
oven, it is placed in the signal path
between the Wireless Bridge and the
WAP at a distance where no impact is
observed, and then the Wireless bridge is
moved closer until the connection is
dropped. Wired connection is then
established.
In both test cases, the microwave oven is
oriented such that the seams in the door
are pointing toward the Wireless Bridge
and the Wireless Access Point.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Common Wireless Devices
Reader: TagMaster LR-3 Pro (PN 154400)
2.435 to 2.465 GHz range
RFID tags
ID-Tags: TagMaster S1255 MarkTag and
S1240 MarkTag MeM
2.435 to 2.465 GHz range
Covidien (formerly ValleyLab) Force FX
(GSTElectro02)
390 kHz
Electrosurgical Unit
da Vinci® Si™
RFID was tested by placing the Reader
and the ID-tags on each side of the signal
path, between the Wireless Bridge and
the Wireless Access Point. In the
coexistence test case, the tags and reader
were 30 inches apart. During the isolated
test with the RFID setup, they were 36
inches apart in the worst case
configuration.
Located in the Vision Cart, which is
placed as close to the Surgeon Console as
possible.
E-12
E-13
Appendix E: OnSite™ for da Vinci® Surgical System
Devices Known to Interfere
• Intuitive Surgical recommends keeping microwave ovens (1000 Watt) at least 25 feet from
the Wireless Bridge or Wireless Access Point, especially if it is located in the signal path.
Higher wattage microwaves should be placed at larger distances.
• If a microwave oven causes interference, use the wired backup to correct the problem.
7.6 meters
or 25 feet
7.6 meters
or 25 feet
Signal
Path
Wireless Access Point
Wireless Bridge
Figure E.1 Placement boundaries for microwave ovens
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Microwave Oven
Testing conducted by Intuitive Surgical determined that microwave ovens can disrupt wireless
communication in certain configurations:
RFID Reader (2.4 GHz)
Testing conducted by Intuitive Surgical determined that RFID readers operating in frequency
hopping mode, or configured to operate at a frequency that overlaps the channel being used
by the wireless bridge and the WAP, will cause minor network disturbances.
To eliminate the interference, Intuitive Surgical recommends the following:
• The RFID reader not operate in frequency hopping mode if it is being used in the same
room as the Wireless Connectivity Option.
• A separation of a least 4 MHz exists between the frequency range of the channel being
used by the wireless bridge/WAP and the operating frequency of the RFID reader. For
example, a wireless bridge/WAP operating on channel 10 spans 2.446 – 2.468 GHz;
therefore, to avoid disturbances from the RFID reader, its operating frequency should be
less than or equal to 2.442 GHz or greater than or equal to 2.472 GHz.
Valid RFID Reader
Operating Frequencies
Valid RFID Reader
4 MHz Operating Frequencies
4 MHz
Wireless Connectivity
Option Frequency Range
2.442
GHz
2.446
GHz
2.468
GHz
2.472
GHz
Figure 6 Valid RFID Reader Operating Frequencies
da Vinci® Si™
Appendix E: OnSite™ for da Vinci® Surgical System
Addressing Wireless Connectivity Problems
If you encounter connectivity problems while using the Wireless Connectivity Option, Intuitive
recommends you do the following:
• Determine if a device transmitting in the 2.4 GHz range is in the room, and if so, disable
the device to see if it resolves the connectivity problems.
• If you experience several disconnections, and the above step did not resolve the issue, or
if the interfering device must be used, then establish a wired network connection with
the da Vinci Surgical System (see Wired Network Connection for more information).
E.11 OnSite Appendix A: IT Requirements
Internet Access
The network security device that will be integrated into the da Vinci Surgical System requires
Internet access to contact servers at Intuitive Surgical.
Intuitive Surgical requires a wired RJ45 Ethernet 10bT/100bT network drop and/or a wireless
802.11 network with Internet access in the OR where the facility’s da Vinci Surgical System is
used. If your da Vinci Surgical System is used in multiple locations, then Intuitive Surgical
requests that be made available in each location.
OnSite is compatible with both DHCP and static networking addresses.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Note that RFID devices can operate outside the frequency range of what was included in the
testing summarized above. If RFID devices operating outside the range shown above exist in
the operating room, Intuitive Surgical recommends that performance is tested with the RFID
device active, before use.
Proxy Server
OnSite is compatible with most proxy servers. In some instances proxy authentication maybe
required to be by-passed.
Firewall
OnSite requires outbound port 443 open.
Network Topology
OnSite requires a minimum amount of bandwidth to post log files (generally less than 1 MB
per day).
da Vinci® Si™
E-14
E-15
Appendix E: OnSite™ for da Vinci® Surgical System
E.12 OnSite Appendix B: Electromagnetic Compatibility
The essential performance for da Vinci Wireless Connectivity during EMC testing was defined
as follows during any of the required tests:
• Video quality exceeded pre-defined metric demonstrating that the video quality was not
impacted
• Audio script test passed demonstrating that the audio link was not impacted
• No changes in programmable parameters
• No resets to factory defaults
• No change in operating mode
• No false alarms
• No initiation of any unintended operation
• No cessation or interruption of any intended operation
Exception: For Voltage Dips and Interrupts, acceptance criteria is no component failures and
is restorable to the pre-test state with operator intervention. For Radiated Immunity in the
band 2.0 - 2.5GHz, acceptance criteria is no component failures and is restorable to the
pre-test state with operator intervention, and restorable during test with a hard-wired
connection.
The da Vinci Surgical System complies with IEC60601-1-2:2001, General Requirements for
safety – Collateral standard: Electromagnetic compatibility. Special precautions and
installation information for the da Vinci Surgical System for electromagnetic compatibility
(EMC) are provided in the following section.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
• No component failures
Use only Intuitive Surgical-branded interconnection cables and accessories. Performance of
cables or accessories other than those specified by Intuitive Surgical as replacement parts for
internal components cannot be guaranteed. Any resulting damage to the system will not be
covered under warranty.
Equipment in the operating room, including the da Vinci Surgical System and other portable
or mobile communications equipment, can produce Electromagnetic Interference (EMI),
which may affect the function of these devices. Such effects are prevented by use of
equipment with EMI characteristics proven below recognized limits, as identified in the below
tables.
In the event of suspected interference from other equipment, which prevents the proper
functioning of the da Vinci Surgical System, contact Intuitive Surgical and/or discontinue use of
the system until the problem can be remedied.
The following Tables contain the Manufacturer’s declaration and additional information
required by IEC60601-1-2:2001.
da Vinci® Si™
Note: This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in
a commercial environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
Table 1: Manufacturer’s Declaration – Electromagnetic Emissions
The da Vinci Surgical System is intended for use in the electromagnetic environment specified below. The customer or
the user of the da Vinci Surgical System should assure that it is used in such an environment.
Emissions Test
Compliance
RF emissions CISPR 11
Group 1
RF emissions CISPR 11
Class A
Harmonic emissions IEC 61000-3-2 Class A
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Complies
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Appendix E: OnSite™ for da Vinci® Surgical System
da Vinci® Si™
Electromagnetic Environment – Guidance
The da Vinci Surgical System uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic environment.
The da Vinci Surgical System is suitable for use in all
establishments, other than domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
E-16
E-17
Appendix E: OnSite™ for da Vinci® Surgical System
Table 2: Manufacturer’s Declaration – Electromagnetic Immunity
The da Vinci Surgical System is intended for use in the electromagnetic environment specified below. The customer or
the user of the da Vinci Surgical System should assure that it is used in such an environment.
IEC 60601 Test Level
Electrostatic discharge
(ESD) IEC 61000-4-2
±6 kV contact ±8 kV air
Compliance Level
±6 kV contact ±8 kV air
Electromagnetic
Environment Guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
±2 kV for power supply lines ±2 kV for power supply lines be that of a U.S. commercial
±1 kV for input/ output lines ±1 kV for input/ output lines or hospital environment with
highly reliable service.
Electrical fast
transient/burst IEC
61000-4-4
±1 kV differential mode ±2
kV common mode
Surge IEC 61000-4-5
<5% UT (>95% dip in UT) for
Voltage dips, short
0.5 cycle 40% UT (60% dip in
interruptions and voltage
UT) for 5 cycles 70% UT (30%
variations on power supply
dip in UT) for 25 cycles <5%
input lines IEC 61000-4-11
UT (>95% dip in UT) for 5 sec.
±1 kV differential mode ±2
kV common mode
Mains power quality should
be that of a U.S. commercial
or hospital environment with
highly reliable service.
<5% UT (>95% dip in UT) for
0.5 cycle 40% UT (60% dip in
UT) for 5 cycles 70% UT (30%
dip in UT) for 25 cycles <5%
UT (>95% dip in UT) for 5 sec.
Mains power quality should
be that of a U.S. commercial
or hospital environment with
highly reliable service. If the
user of the da Vinci Surgical
System requires continued
operation during power
mains interruptions, it is
recommended that the da
Vinci Surgical System be
powered from an
uninterruptedly power
supply or a battery.
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Immunity Test
Power frequency (50/60
Hz) magnetic field IEC
61000-4-8
3 A/m
Note: UT is the AC mains voltage before application of the test level.
da Vinci® Si™
Appendix E: OnSite™ for da Vinci® Surgical System
Table 3: Manufacturer’s Declaration – Electromagnetic Immunity
The da Vinci Surgical System is intended for use in the electromagnetic environment specified below. The customer or
the user of the da Vinci Surgical System should assure that it is used in such an environment.
Conducted RF IEC
61000-4-6 Radiated
RF IEC 61000-4-3
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Immunity test
IEC 60601 test
level
Compliance level
3 Vrms 150 kHz to
3 Vrms
80 MHz
3 V/m 80 MHz to 3V/m
2.5 GHz
Electromagnetic environment – guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the da Vinci
Surgical System, including cables, than the
recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2 P
d = 1.2 P 80 MHz to 800 MHz
d = 2.3 P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m). Field strengths from fixed RF
transmitters, as determined by an electromagnetic site
survey,a should be less than the compliance level in
each frequency range.b Interference may occur in the
vicinity of equipment marked with the following
symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the da Vinci Surgical System is used exceeds the
applicable RF compliance level above, the da Vinci Surgical System should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orientating or relocating the da
Vinci Surgical System.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
da Vinci® Si™
E-18
E-19
Appendix E: OnSite™ for da Vinci® Surgical System
The da Vinci Surgical System is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the da Vinci Surgical System can help prevent electromagnetic interferences
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the da Vinci Surgical System as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output power
of transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d = 1.2 P
80 MHz to 800 MHz
d = 1.2 P
800 MHz to 2.5 GHz
d = 2.3 P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz to 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
Table 5: Compliance Information
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Table 4: Recommended separation distances between portable and mobile RF
communications equipment and the da Vinci Surgical System
Network Router:
CAN/CSA-C22.2 No 60950-1-03
ANSI/UL Std No 60950-1 1st Ed.
Wireless Bridge: *
Wireless Data Rate
802.11b/g: 2.4-2.4835 GHz
802.11b/g: 2.4-2.4835 GHz
* See OnSite Appendix C: Wireless Bridge Data on page 20 for more information
da Vinci® Si™
Appendix E: OnSite™ for da Vinci® Surgical System
NS2L DATASHEET
SYSTEM INFORMATION
Processor Specs
Memory Information
Networking Interface
Atheros MIPS 4KC, 180MHz
16MB SDRAM, 4MB Flash
1 X 10/100 BASE-TX (Cat. 5, RJ-45) Ethernet Interface
REGULATORY / COMPLIANCE INFORMATION
FCC Part 15.247, IC RS210, CE
YES
Tolerance
+/-1dB
+/-1dB
+/-1dB
+/-1dB
6Mbps
9Mbps
12Mbps
18Mbps
24Mbps
36Mbps
48Mbps
54Mbps
+/-1dB
+/-1dB
+/-1dB
+/-1dB
+/-1dB
+/-1dB
+/-1dB
+/-1dB
802.11b
RADIO OPERATING FREQUENCY 2412-2462 MHz
TX SPECIFICATIONS
RX SPECIFICATIONS
DataRate
TX Power
Tolerance
DataRate
Sensitivity
1Mbps
20 dBm
+/-1dB
1Mbps
-95 dBm
2Mbps
20 dBm
+/-1dB
2Mbps
-94 dBm
5.5Mbps
20 dBm
+/-1dB
5.5Mbps
-93 dBm
11Mbps
20 dBm
+/-1dB
11Mbps
-90 dBm
20
20
20
20
20
18
16
15
dBm
dBm
dBm
dBm
dBm
dBm
dBm
dBm
+/-1dB
+/-1dB
+/-1dB
+/-1dB
+/-1dB
+/-1dB
+/-1dB
+/-1dB
802.11g OFDM
802.11g OFDM
802.11b
Wireless Approvals
RoHS Compliance
6Mbps
9Mbps
12Mbps
18Mbps
24Mbps
36Mbps
48Mbps
54Mbps
ADJUSTABLE CHANNEL SIZE SUPPORT
10MHZ
5MHz
-92
-91
-89
-88
-84
-81
-75
-72
dBm
dBm
dBm
dBm
dBm
dBm
dBm
dBm
20MHz
RANGE PERFORMANCE
Outdoor (BaseStation Antenna Dependent):
Over 5km
ANTENNA
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
E.13 OnSite Appendix C: Wireless Bridge Data
Gain
Polarization
Polarization Selection
Azimuth
8dBi (2400-2500MHz)
Multi-Polarized
Software Controlled
Surival Wind
3dB Beamwidth Elevation
3dB Beamwidth Azimuth
216 km /hr
60 degrees
60 degrees
Elevation
Enclosure Size
Weight
Enclosure Characteristics
Mounting Kit
Max Power Consumption
Power Supply
Power Method
Operating Temperature
Operating Humidity
Shock and Vibration
PHYSICAL / ELECTRICAL / ENVIRONMENTAL
16.3 cm. length x 3.1 cm. height x 8cm. width
0.18kg
Outdoor UV Stabalized Plastic
Pole Mounting Kit included
4 Watts
12V, 1A (12 Watts). Supply and injector included
Passive Power over Ethernet (pairs 4,5+; 7,8 return)
-20C to +70C
5 to 95% Condensing
ETSI300-019-1.4
SOFTWARE
visit www.ubnt.com/airos
495-499 Montague Expwy. Milpitas, CA 95035 CA 95035 T(408)-942-3085
F (408)-351-4973 http://www.ubnt.com
__________________________________End of Section_______________________________
da Vinci® Si™
E-20
F-1
Appendix F: 8.5 mm Endoscope for the da Vinci Si System
Appendix F: 8.5 mm Endoscope for the
da Vinci Si System
This section provides details specific to the 8.5 mm 3D endoscope system designed to be used
with the da Vinci Si Surgical System. It augments the information within this manual regarding
endoscopes, especially under Endoscopesin section 7.1 Vision System Overview. Users should
consider the following:
• Users should have a thorough understanding of the use of the 12 mm endoscope system
in conjunction with the da Vinci Si Surgical System before using the 8.5 mm endoscope
and components. This section contains important information about the differences
between the 8.5 mm and 12 mm endoscopes and components.
WARNING: Be sure to read and understand all caution and warning information found
in this manual before using this product.
Indications for Use
The Intuitive Surgical 8.5 mm Endoscopic System is intended for endoscopic viewing of
internal surgery sites during minimally invasive surgery in the peritoneal cavity, thoracic
cavity, and peritoneum. It is designed for use with the Intuitive Surgical da Vinci Si Instrument
Control System during laparoscopic and thoracoscopic surgical procedures.
F.1 Overview
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
The 8.5 mm endoscope provides a 3D view of the operative field when used with the da Vinci
Si High Definition Vision System. The small diameter of the 8.5 mm endoscope enables the da
Vinci Si Surgical System to be used with a smaller endoscope port. While the system is not
docked, you can use the 8.5 mm endoscope for laparoscopy through a da Vinci 8 mm
instrument cannula.
Figure F.1 Using the 8.5 mm endoscope manually
da Vinci® Si™
Appendix F: 8.5 mm Endoscope for the da Vinci Si System
The 8.5 mm endoscope uses a three-piece system concept (endoscope, adapter and camera).
The 8.5 mm endoscope is compatible with the High Definition cameras, illuminators and light
guides provided with the da Vinci Si Surgical System. The 8.5 mm endoscope is available in
straight (0°) and angled (30°) tip configurations.
• While the system is docked, the 8.5 mm endoscope requires use of the 8.5 mm
Endoscope Cannula or a validated third-party endoscope cannula. Refer to the list of
Validated 3rd Party Products for da Vinci® Surgical Systems (PN 871770).
• The 8.5 mm endoscope requires use of the 8.5 mm Alignment Target.
• The 8.5 mm endoscope is approximately 90 mm shorter than the 12 mm endoscope.
• The 8.5 mm endoscope tip is not heated.
• The 8.5 mm endoscope is more flexible than the 12 mm endoscope. Therefore, special
care in the setup and handling of the 8.5 mm endoscope is required.
• To prevent damage during reprocessing, we strongly recommend you place the
endoscope in a properly designed sterilization tray or case, like those we identify in the
list of Suggested 3rd Party Products for da Vinci® Surgical Systems (PN 871771)
• The 8.5 mm endoscope has lower resolution and less brightness than the 12 mm
endoscope.
CAUTION: The 8.5 mm endoscope should only be used in cases where the image quality
of the 12 mm endoscope is not required. A 12 mm endoscope should be available for use
if an increase in image quality is preferred.
F.2 Working with the 8.5 mm Endoscope
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Special considerations for the 8.5 mm endoscope include:
The 8.5 mm endoscope uses the same camera arm and camera head drapes as the 12 mm
endoscope. However, the 8.5 mm endoscope requires a specific alignment target and
endoscope cannula. The 8.5 mm endoscope, alignment target and cannulae are clearly
marked “8.5 mm” (see Figure F.2 for examples below).
Figure F.2 “8.5 mm” marking on alignment target and endoscope
The following table provides the compatible combinations of reusable cannula, cannula
mount and alignment target for use with the 8.5 mm endoscope on da Vinci Si Surgical
Systems. For a list of disposable endoscope cannulae validated by Intuitive Surgical, refer to
the list of Validated 3rd Party Products for da Vinci® Surgical Systems (PN 871770).
da Vinci® Si™
F-2
F-3
Appendix F: 8.5 mm Endoscope for the da Vinci Si System
Alignment Target
Endoscope Cannula Mount
Reusable Endoscope Cannula
371679
371521 (ETH/TAUT)
420260
Refer to appropriate sections of this user manual for general instructions regarding
endoscope alignment and setup, including connections to the other components of the
Vision System and to the camera arm of the Patient Cart. Refer to the Reprocessing
Instructions for compatible sterilization methods and parameters for the 8.5 mm endoscopes.
________________________________End of section________________________________
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Table F-1 Compatible combinations of reusable cannula
da Vinci® Si™
Appendix G: Symbols, Icons and Text Messages Reference
G Appendix G: Symbols, Icons and Text
G.1 Overview
This appendix provides a reference for symbols, LEDs (colored lights), icons and text messages
you may see on system components or monitors.
Symbols
Symbols appear on system components and serve these purposes:
• Identification of important system connections and functions
• Provide caution and warning information
LED Status Indicators
Indications of the status of the instrument and camera arms are provided by LEDs on the top
of the insertion axis of each instrument and camera arm. The meanings of the colors are as
follows:
LED Quick Reference
Not ready for surgeon control
blinking
Arm or port clutch
blinking
Guided tool change
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Messages Reference
Ready for surgeon control
Warning: Intervention required
See display for details
Error: System restart required
See display for details
blinking
System working.
Please wait.
Figure G.1 LED Quick Reference
Corresponding LED icons—graphics that reproduce the LED status—appear simultaneously
on the touchscreen and stereo viewer.
da Vinci® Si™
G-1
G-2
Appendix G: Symbols, Icons and Text Messages Reference
On-Screen Icons and Text Messages
Other console
instrument status area/icon
Camera indicators:
angle, rotation, zoom
Modal message
area
Critical message
area/icon
Arm 2
status:
on left,
active (larger)
Arm 1
status:
on right,
active (larger)
Arm 3
status:
on left
inactive (smaller,
below, gray)
Left master
message area/icon
Right master
message area/icon
Camera arm
message area/icon
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Icons and text messages are overlaid on the video displays to provide information regarding
the status of the system. The following figures illustrate arrangement of overlaid elements in
the stereo viewer and touchscreen displays. Note that many overlaid elements appear only
when needed, and others are usually or always present.
Inactive instrument
name (gray)
Active instrument arm
message areas/icons
Footswitch map
Active instrument
Inactive instrument
name
arm message area/icon
Figure G.2 Stereo viewer display (SmartPedal technology)
da Vinci® Si™
Appendix G: Symbols, Icons and Text Messages Reference
Critical message
area/icon
Modal message
area
Arm 1 data:
on right,
instrument name,
energized (bolt),
foot over enabled
pedal pair (blue)
Instrument arm
message areas,
icons
Menu pop up
area
Camera message
area, icon
Camera angle, zoom,
rotation indicator
Menu tabs area
Figure G.3 Touchscreen display
G.2 Symbols and Icons Reference Table
Table G-1 Symbols and Icons
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Arm 2 data:
on left, surgeon
name or console
control
Single or dual console
masters & instruments status
Symbol or Icon
Meaning
Where Found
Read the operating instructions.
System component labels
Type CF Applied Part
On camera head and instruments
Type BF Applied Part
On Illuminator.
Note: The da Vinci Si camera
head provides isolation in
accordance with a CF applied part
and is acceptable for use with the
Y1903 Illuminator.
Illuminator lamp module
or
Caution: Hot
da Vinci® Si™
G-3
G-4
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Symbol or Icon
Meaning
Vision Setup button
Inside (not visible to users) the
Camera Control Unit, Core, Vision
Cart, Patient Cart and Surgeon
Console
Camera head
Focus In, Focus Out buttons
Camera head
Lamp On/Off button
Camera head
Flush Port
On instruments adjacent to flush
port
30 degrees up, indicates which side On one side of endoscope base
of scope should be on same side as
camera head buttons to achieve
this scope orientation.
30 degrees down, indicates which On one side of endoscope base
side of scope should be on same
side as camera head buttons to
achieve this scope orientation.
Turn as indicated to unlock.
Camera head and endoscope
30º
30º
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Protective ground
Where Found
Turn as indicated to lock.
da Vinci® Si™
Camera head and endoscope
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Meaning
Where Found
Do not autoclave.
Endoscope and camera head
Fragile, handle with care
Endoscope
Ethernet Connection
Inside Surgeon Console service
panel and rear of Patient Cart
Alternating Current
On product labels containing rating
information on rear of Patient Cart,
Surgeon Console and Vision Cart
Rear of Surgeon Console, Patient
Cart, Vision Cart, Camera Control
Unit, Illuminator and Core.
Note: The terminal is not
required for operation. It is
provided for convenience to
allow for other equipment to
be at the same equalization
potential as the da Vinci
Surgical System.
Inside Surgeon Console service
panel and rear of Patient Cart
Autoclave
Equipotential Terminal
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Symbol or Icon
Serial Port Connection
Standby—found on Power buttons
of Vision Cart, Patient Cart and
Surgeon Console, lit amber when in
standby mode (connected to mains
but not powered on), blue when
powered on.
Off (power: disconnection from
mains)
Power buttons on Patient Cart,
Vision Cart, Surgeon Console and
Core
On (power: connection to mains)
Rear of Patient Cart, Vision Cart and
Surgeon Console, Illuminator,
Camera Control Unit and Core
Back of Core and back of Surgeon
Console
DVI video port
DVI
da Vinci® Si™
Rear of Patient Cart, Vision Cart and
Surgeon Console, Illuminator,
Camera Control Unit and Core
G-5
G-6
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Symbol or Icon
Composite video ports
S-Video port
Composite
S-Video
Core Video port
Illuminator control ports
Back of CCU
Core control ports
Back of CCU
Audio bay, green
Audio
Left
Right
Core
Video
Illuminator
Control
Core
Control
Video Control
Touch
Screen
Where Found
Back of Core
Back of Core and back of Surgeon
Console
Back of Core and back of Surgeon
Console
Back of Core and back of Surgeon
Console
Back of Core and back of Surgeon
Console
Back of Core and back of Surgeon
Console
Back of Core and back of CCU
SDI video port
SDI
Video control ports
Back of Core
Touchscreen audio connection port Back of Core
Audio
Touchscreen video connection
ports
Touch
Screen
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Meaning
Back of Core
Video
Touchscreen communication ports Back of Core
Touch
Screen
Com
Service connection ports
Headset connection port
Back of Core
Back of Core and back of Surgeon
Console
Audio line in port
Back of Core and back of Surgeon
Console
Audio line out port
Back of Core and back of Surgeon
Console
Video out bay, orange, labeled
Back of Core and back of Surgeon
either “Aux” for auxiliary (on Core); Console
L (left) and R (right) (on Surgeon
Console); or 1 and 2 (optional bays
on Core)
Video in bay, blue
Back of Core and back of Surgeon
Console
SERVICE
Headset
Line In
Line Out
Video Out
TilePro Input
da Vinci® Si™
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Symbol or Icon
FIBER
PUSH
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
BRAKE
Meaning
Where Found
Fiber cable (system cable)
receptacle
Indicates the brakes on the Surgeon
Console, applied by stepping down.
Indicates where to push the
Surgeon Console.
Master clutch
Next to fiber cable receptacles on
back of Core
Both sides of Surgeon Console, near
floor
Both sides of Surgeon Console, near
handles
Clutch pedal on footswitch panel
Instrument arm swap
Arm swap kick-plate (left side) on
footswitch panel
Do not step here.
On Patient Cart base
Tip hazard during transport. Stow
touchscreen and close rear door
before moving cart.
On label on rear door of Vision Cart
Do not move Surgeon Console from Back of Surgeon Console
the back.
Dispose of in accordance with local System labels
regulations—particularly applies to
electronic components.
da Vinci® Si™
G-7
G-8
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Meaning
Where Found
Fiber cable (system cable)
receptacle
Next to system cable receptacles on
back of Core, Surgeon Console and
Patient Cart
Caution: Laser hazard
On blue covers of system cable
receptacles on Core, Surgeon
Console and Patient Cart
Video output format of associated Back of Core, with connection bays
output ports is configurable via the
touchscreen. On Video Settings tab,
select Video Output button.
Energy activation cable connection Back of Core
port
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Symbol or Icon
da Vinci® Si™
Video out bay, left video channel;
video has no overlays. This is
component video, made up of Y
(green port), PR (red port) and PB
(blue port). A similar label appears
for the right video channel.
Surgeon head-in sensor
Back of Camera Control Unit (CCU)
Indicates forward and reverse for
the Patient Cart motor drive
Right tiller handle
Indicates range of speed for the
Patient Cart motor drive
Right tiller handle
Surgeon Console viewer
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Symbol or Icon
Meaning
Where Found
Shows use of throttle
Between the tiller handles
Explains N=Neutral and D=Drive
positions for the motor drive shift
switches. Includes text, “IF NO
CART POWER, USE NEUTRAL TO
MOVE CART”
Top of Patient Cart motor drive
tiller.
Camera Arm setup joint
“Sweet Spot” label: Its limits
indicate recommended distance
range of Camera Arm remote center
from Patient Cart tower
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Shows use of throttle enable switch Between the tiller handles
da Vinci® Si™
“Drive” position: Patient Cart drive
motor engaged
Patient Cart base near motor shift
switches
“Neutral” position: Patient Cart
drive motor disengaged
Patient Cart base near motor shift
switches
Brake release
Near upper port clutch button on
instrument and camera arms
G-9
G-10
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Meaning
Where Found
Pinch/Crush Hazard
On Patient Cart, below upper port
clutch button and at junction of
setup joint and top of column on
instrument and camera arms; on
Surgeon Console
Emergency Power Off
Rear of Patient Cart
Not used on IS3000 system but may
Interference may occur in the
vicinity of equipment marked with appear on other equipment in the
this symbol.
OR
Speaker connection port
Back of Core
Microphone connection port
Back of Core
Critical message area
The system is preparing to shut
down. A message indicating the
number of seconds until shut down
appears in the body text area.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Symbol or Icon
General information icon. Appears Critical message area
when the system is providing
information that is not fault-related. Camera arm message area
General warning / recoverable fault. Critical message area
Appears when the system detects a
recoverable fault somewhere
Touchpad popup dialog
within the system not associated
with a particular arm or master.
Touchpad error handling area
da Vinci® Si™
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Meaning
General critical warning /
non-recoverable fault. Appears
when the system detects a
non-recoverable fault somewhere
within the system not associated
with a particular arm or master.
Where Found
Critical message area
Touchpad error handling area
Can also appear by itself on the
touchscreen and touchpad when a
critical startup error has occurred
Scope not detected. Appears when Critical message area
the system does not detect an
endoscope and the user is
Touchpad popup dialog
attempting to go into following.
Guided tool change in progress
Instrument arm message area
The system has detected a problem Instrument arm message area
with the instrument. This can
appear when the instrument is
expired, when the instrument is
incompatible with the system,
when the system is not prepared to
have an instrument installed on it,
or when the system is having
difficulty communicating with the
instrument.
General informational icon related Instrument arm message area
to the instrument. This can appear
when the system is downloading
new instrument information from a
plug-and-play instrument.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Symbol or Icon
Camera arm message area
Camera arm is currently clutched
and is free to be moved by OR staff
around its remote center.
da Vinci® Si™
G-11
G-12
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Symbol or Icon
Meaning
Where Found
The system has detected a
non-recoverable fault on the left
master.
Left master message area
The system has detected a
non-recoverable fault on the right
master.
Right master message area
The system has detected a
non-recoverable fault on an
instrument arm.
Instrument arm message area
The system has detected a
non-recoverable fault on the
camera arm.
Camera arm message area
Move the right master grips to
match the instrument grips (i.e.,
“Follow on matching grip”).
Right master message area
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Instrument arm is currently
Instrument arm message area
clutched and is free to be moved by
OR staff around its remote center.
da Vinci® Si™
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Symbol or Icon
Meaning
Where Found
The system has detected a
recoverable fault or other
resolvable problem on the left
master.
Left master message area
The system has detected a
recoverable fault or other
resolvable problem on the right
master.
Right master message area
The system has detected a
recoverable fault or other
resolvable problem on an
instrument arm.
Instrument arm message area
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Move the left master grips to match Left master message area
the instrument grips (i.e., “Follow
on matching grip”).
The system has detected a
Camera arm message area
recoverable fault or other
resolvable problem on the camera Above footswitch map when
arm.
camera control pedal is pressed on
other console.
In dual console mode, appears
when camera control pedal is
pressed on other console,
explaining why instruments stop
moving and firing.
The instrument tip is still inside the Instrument arm message area
cannula. To continue, you must
clutch the instrument arm and
advance the tip into the body.
da Vinci® Si™
G-13
G-14
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Meaning
Where Found
The system has detected excessive Instrument arm message area
force on an instrument arm. This
normally means that some external
object is pushing on an arm.
This instrument arm is locked;
unlock it via the touchpad to use.
Instrument arm message area
Hit the arm swap pedal to use the
instrument in question.
Instrument arm message area
The system has detected a problem Instrument arm message area
with a cannula.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Symbol or Icon
Relax your grip on the left master so Left master message area
that it can self-align.
Relax your grip on the right master Right master message area
so that it can self-align
da Vinci® Si™
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Symbol or Icon
Meaning
Where Found
Roll the right master grip to
proceed.
Right master message area
Your instruments have been
Critical message area
reassigned. Tap ‘Arm Swap’ pedal to
acknowledge and continue.
Touchscreen instrument arm status
area
Dual console instrument status area
(touchpad and touchscreen)
The instrument arm is conducting
its self test.
Instrument arm message area
The camera arm is conducting its
self test.
Camera arm message area
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Roll the left master grip to proceed. Left master message area
The left master is conducting its self Left master message area
test.
da Vinci® Si™
G-15
G-16
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Meaning
Where Found
The right master is conducting its
self test.
Right master message area
Select the desired motion scaling
level.
Touchpad popup menu
The ergonomic settings are being
adjusted.
Critical message area
Setup arm has been moved
Instrument or camera arm message
unexpectedly; press one of the port area
clutch buttons to continue.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Symbol or Icon
Energy activation is currently
unavailable.
Energy may be unavailable because
the instrument installed is not an
energy instrument or because no
compatible ESU for the installed
instrument is detected.
Energy activation is currently
available.
Touchscreen instrument arm status
area
3D viewer instrument arm status
area
Touchscreen instrument arm status
area
3D viewer instrument arm status
area
da Vinci® Si™
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Meaning
Where Found
No scope is detected.
Touchscreen camera status area
Scope detected. This icon is
Touchpad central column on main
page (shown vertically)
Touchscreen camera status area
Touchpad central column on main
page (shown vertically)
accompanied by 30˚ , 30˚ , or 0˚.
Digital Zoom. This icon is
accompanied by one of the
following zoom levels:
•
•
•
•
Touchscreen camera status area.
Touchpad central column on main
page
Wide
Full
2x
4x
Indicates which manipulator is
currently associated with the
surgeon’s right hand.
Touchscreen instrument arm status
area
Touchpad instrument arm status
area
Indicates which manipulator is
currently associated with the
surgeon’s left hand.
Touchscreen instrument arm status
area
Touchpad instrument arm status
area
Indicates TilePro input 1.
Button on touchscreen display
Indicates TilePro input 2.
Button on touchscreen display
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Symbol or Icon
da Vinci® Si™
G-17
G-18
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Symbol or Icon
Meaning
Button on touchscreen display (for
selecting endoscopic camera view)
Camera control foot pedal
Surgeon Console
Touchscreen master status area
(shown when Surgeon Console is in
camera control)
Button on touchscreen display (for
selecting “surgeon’s TilePro view”)
Electronic brightness control (does Touchscreen and touchpad displays
not affect Illuminator light output).
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Camera
Where Found
Increase setting.
Touchscreen display buttons; can
be used as an alternative to sliders
Decrease setting.
Touchscreen display buttons; can
be used as an alternative to sliders
Close tab menu.
Touchscreen display; used to close
the tab menu
Mute microphone.
Touchscreen and touchpad
displays; used to mute the local
microphone (touchpad version
mutes the Surgeon Console
microphone and touchscreen
version mutes the Vision Cart
microphone)
Touchscreen and touchpad
displays; used to enable the local
microphone (touchpad version
enables the Surgeon Console
microphone and touchscreen
version enables the Vision Cart
microphone)
Enable microphone.
da Vinci® Si™
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Meaning
Where Found
Speaker volume
Touchscreen and touchpad
displays; labels the slider that
controls the local speaker volume
Erase telestration marks.
Touchscreen display
Instrument arm locked (when on
screen).
Instrument arm message area on
touchscreen.
Brake applied (when brake pedal is Touchpad instrument arm lock
depressed on Surgeon Console).
button
Adjust right.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Symbol or Icon
da Vinci® Si™
Above applied brakes on Surgeon
Console
Touchscreen / touchpad displays
for camera calibration
Adjust left.
Touchscreen / touchpad displays
for camera calibration
Adjust up.
Touchscreen / touchpad displays
for camera calibration
Adjust down.
Touchscreen / touchpad displays
for camera calibration
Secondary Energy Pedal
3D viewer footswitch map
G-19
G-20
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Meaning
Where Found
Secondary Energy Pedal (when
pressed)
3D viewer footswitch map
Primary Energy Pedal
3D viewer footswitch map
Primary Energy Pedal (when
pressed)
3D viewer footswitch map
Arm Swap Pedal
3D viewer footswitch map
Arm Swap Pedal (when pressed)
3D viewer footswitch map
Master Clutch Pedal
3D viewer footswitch map
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Symbol or Icon
da Vinci® Si™
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Symbol or Icon
Meaning
Where Found
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Master Clutch Pedal (when pressed) 3D viewer footswitch map
Camera Control Pedal
3D viewer footswitch map
Camera Control Pedal (when
pressed)
3D viewer footswitch map
Masters status
3D viewer footswitch map
Masters status: left finger clutch
activated
3D viewer footswitch map
Masters status: right finger clutch
activated)
3D viewer footswitch map
3D viewer footswitch map
Masters status during camera
control: indicates that roll-to-focus
is available
3D viewer footswitch map
Masters status during camera
control, when master roll is causing
camera to focus in the + direction
3D viewer footswitch map
Master controls status during
camera control, when master roll is
causing camera to focus in the –
direction)
da Vinci® Si™
G-21
G-22
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Symbol or Icon
Meaning
3D viewer (top center)
Touchscreen (lower right)
3D viewer (top center)
Camera rotation indicator when
Touchscreen (lower right)
angle is indeterminate; this can
happen when no scope is selected
or when scope is looking straight
up or straight down.
Touchscreen masters & instruments
Left and right master are not
associated with an instrument arm. status area (top center)
Left master is associated with an
instrument on arm 2 and right
master is associated with an
instrument on arm 1.
Cautery instrument on arm 1 with
right master is not energized for
some reason.
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Camera rotation indicator
Where Found
3D viewer (upper left) in dual
console mode
Touchscreen masters & instruments
status area (top center)
3D viewer (upper left) in dual
console mode
Touchscreen masters & instruments
status area (top center)
3D viewer (upper left) in dual
console mode
Energized instrument on arm 1 with Touchscreen masters & instruments
right master is locked.
status area (top center)
3D viewer (upper left) in dual
console mode
There is a problem with the cannula Touchscreen masters & instruments
status area (top center)
on arm 1 with the right master.
3D viewer (upper left) in dual
console mode
The instrument on arm 1 with right Touchscreen masters & instruments
status area (top center)
master is experiencing excessive
external force.
3D viewer (upper left) in dual
console mode
da Vinci® Si™
Appendix G: Symbols, Icons and Text Messages Reference
Table G-1 Symbols and Icons
Meaning
Where Found
System reports “No instrument
Touchscreen masters & instruments
installed” on instrument arm 1 with status area (top center)
right master.
3D viewer (upper left) in dual
console mode
Right master is finger clutched.
Touchscreen masters & instruments
status area (top center)
Recoverable problem with right
master.
Instrument arm 1 clutched.
3D viewer (upper left) in dual
console mode
Touchscreen masters & instruments
status area (top center)
3D viewer (upper left) in dual
console mode
Touchscreen masters & instruments
status area (top center)
da Vinci Network Offline.
3D viewer (upper left) in dual
console mode
Touchscreen lower right
da Vinci Network Online.
Touchscreen lower right
OnSite Session In Progress.
Touchscreen lower right
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
Symbol or Icon
da Vinci® Si™
G-23
G-24
Appendix G: Symbols, Icons and Text Messages Reference
Table G-2 contains a list of text messages that can appear on screen, sorted alphabetically. The
text messages can appear in one or several contexts and locations and are written so as to be
understood in each context on screen, and therefore not explained further. Variables are
shown in italics inside angled brackets, like this: . This table is provided as a
reference and to support translation into languages not supported in the system software.
Table G-2 Text Messages
Message
10/9/14
DRAFT/PRE-RELEASE/CONFIDENTIAL
G.3 Text Messages Reference Table
# A B
C P1 P2 P3 P4

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : No
Author                          : cynthiag
Create Date                     : 2015:04:29 15:41:39-07:00
Modify Date                     : 2015:04:29 15:44:03-07:00
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 4.2.1-c043 52.372728, 2009/01/18-15:08:04
Metadata Date                   : 2015:04:29 15:44:03-07:00
Creator Tool                    : FrameMaker 11.0.2
Format                          : application/pdf
Title                           : da Vinci Si User Manual.book
Creator                         : cynthiag
Document ID                     : uuid:6a216d88-fa85-4245-aef9-c15c3cdacef7
Instance ID                     : uuid:7f58aa17-2d45-45b3-b370-850733c4ef12
Producer                        : Acrobat Distiller 11.0 (Windows)
Page Count                      : 100
EXIF Metadata provided by EXIF.tools
FCC ID Filing: 2AAZF-CHB01

Navigation menu