Medtronic MiniMed 3160 Personal Pump Communicator User Manual mp6025069 011 1

Medtronic MiniMed, Inc. Personal Pump Communicator mp6025069 011 1

Contents

Users Manual 2

Medtronic MiniMed 2007D Implantable Insulin Pump SystemPhysician’s Manual
© 2003, Medtronic MiniMed. All rights reserved.Medtronic MiniMed™ is a trademark of Medtronic MiniMedDual Wave™ is a trademark of Medtronic MiniMedSquare Wave™ is a trademark of Medtronic MiniMedSteri-strip® is a registered mark from 3MAventis® is a registered mark from Aventis PharmaceuticalGenapol® is a registered mark from Aventis PharmaceuticalLuer Lok® is a registered mark from BD and Co.This device is protected under one or more of the following U.S. Patents:U.S., international, and foreign patent applications are pending.4,731,051 5,217,442 5,527,307 6,427,0884,776,842 5,257,971 5,559,828 6,537,2685,167,633 5,460,618 5,797,733 6,562,0015,176,644 5,466,218 5,915,929 6,564,1055,197,322 5,514,103 6,283,943 6,571,1286025069-011 7/04 REF MMT-31600976
iTable of contentsCHAPTER 1 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Introduction  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Implantable Insulin Pump  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Insulin medication  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Side Port Catheter  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Personal Pump Communicator (PPC)  . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7CHAPTER 2 Indications and contraindications  . . . . . . . . . . 9Indications for use  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Contraindications for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Possible adverse effects  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10CHAPTER 3 Personal Pump Communicator (PPC)  . . . . . . 11Introduction  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11PPC icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11PPC buttons  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Communicating with the Pump  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Install/Replace the main battery   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Part 1: PPC/Pump system initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15Initialize the PPC  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15Set the time and date  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Set alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Set maximum bolus, basal rate and time display format . . . . . . . . . . 19Lock maximum bolus/basal, enter personal ID and password, stop Pump  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
iiProgram a basal rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Part 2: Additional PPC programming features   . . . . . . . . . . . . . . . . . . . . . . . 23Main menu  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Program a bolus  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Set a Normal bolus with the Variable Bolus feature turned off  . . . . . . . . 24Set a Normal bolus with the Variable Bolus feature turned on  . . . . . . . . 25Set a Square Wave Bolus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Set a Dual Wave Bolus   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Review bolus history  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Suspend mode  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29Programming a basal rate   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Programming basal delivery pattern  . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Setting basal rate profiles in each delivery pattern  . . . . . . . . . . . . . . . . . 31Program a temporary basal rate  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32Set a temporary basal rate  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32Stop a temporary basal rate  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Personal events   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34History  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Setup Pump  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Time and Date  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Auto off  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Alarms  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38Self test   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38Basal delivery pattern   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Initialize PPC to Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Setup II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Exit Setup menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Setup II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40Audio Bolus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40Activating the Audio Bolus feature  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
iiiSet an Audio Bolus from the main menu bolus screen  . . . . . . . . . . . . . . 41Variable Bolus  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41Maximum Bolus   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42Maximum Basal Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42Time display format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42Personal Events  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42Pump Setup  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42Exit Setup Menu   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42Supervisor mode   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43Refill   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43Priming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44Diagnostic rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Initialize to factory defaults  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47Stop Pump  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49Supervisor password  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49Exit supervisor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49Personal Pump Communicator messages   . . . . . . . . . . . . . . . . . . . . . . . . 50Technical history codes  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51Technical history PPC codes  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51Clinical history Pump codes   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52CHAPTER 4 Pump implantation  . . . . . . . . . . . . . . . . . . . . . 53Preprogramming and pre-testing the Pump  . . . . . . . . . . . . . . . . . . . . . . . . . . 53Registration card  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53Supplies and solutions  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53Supplies  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53Solutions   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54Emptying and filling the Pump  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54Remove shipping fluid from the Pump  . . . . . . . . . . . . . . . . . . . . . . . . . . 55Rinse the Pump with insulin (IN1)   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
ivFill the Pump with insulin  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58Measure stroke volume  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59Prepare the Side Port Catheter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60Performing the surgical procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64Pre-operative evaluation   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64Formation of the pump pocket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64Catheter placement   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66Post-operative management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67Post-operative hospitalization  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67Post-operative x-rays  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67CHAPTER 5 Pump refill procedure . . . . . . . . . . . . . . . . . . . 69Introduction  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69Supplies and solutions  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70Supplies: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70Solutions:  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70Prepare for pump refill . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71Perform the refill procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72Fill out the refill form   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72Label syringes  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72Prepare the refill syringe for emptying the Pump  . . . . . . . . . . . . . . . . . . 72Prepare the refill syringe for filling the Pump  . . . . . . . . . . . . . . . . . . . . . 73Empty the Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73Refill the Pump  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76Calculate extracted and refill amounts . . . . . . . . . . . . . . . . . . . . . . . . 77Calculate refill accuracy   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77CHAPTER 6 Explanting the Pump System . . . . . . . . . . . . . 79Explant considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79Returning devices/components to Medtronic MiniMed   . . . . . . . . . . . . . . . . 79
vCHAPTER 7 Warnings and precautions  . . . . . . . . . . . . . . . 81Warnings  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81Electrotherapy  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81Diagnostic ultrasound   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82Ultrasound therapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82Diagnostic radiation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82Elevated anti-insulin antibodies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82Environmental conditions  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83Sterilization  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85Emergencies and the use of conventional insulin supplies  . . . . . . . . . . . 85PPC reliability requirements  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85Maximum dosages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85Electrical and magnetic fields  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86CHAPTER 8 Adverse reactions . . . . . . . . . . . . . . . . . . . . . . 87Adverse reactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87Prevention   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87CHAPTER 9 System alarms and messages  . . . . . . . . . . . . 89Pump alarms   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90Alarm feedback  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90Pump low battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90Depleted pump battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90System error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91Pump self test fail   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91PPC low battery  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91PPC alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92Low reservoir . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92Empty reservoir  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92Telemetry communication error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
viInitialize alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93PPC not initialized  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93Battery replacement  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94Initialize to factory defaults  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94Pump stopped . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94Pump suspended   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95Auto off  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95Hourly maximum exceeded . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95Pump alarm table  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96CHAPTER 10 Troubleshooting Pump System under-delivery . . . . . . . . . . . . . . . . . . . . . . . . . 97Diagnostic procedures  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97Under-delivery caused by backflow   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97Under-delivery caused by catheter occlusion   . . . . . . . . . . . . . . . . . . . . . . . . 98CHAPTER 11 Technical specifications . . . . . . . . . . . . . . . . . 99Implantable Insulin Pump MMT-2007D  . . . . . . . . . . . . . . . . . . . . . . . . . 99Personal Pump Communicator (PPC) model MMT-3160 . . . . . . . . . . . 100Side Port Catheter  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101FCC compliance   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101APPENDIX A Label information symbol dictionary . . . . . . . . . . 103Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104Other information   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104APPENDIX B Implant worksheet . . . . . . . . . . . . . . . . . . . . . . . 105Implant worksheet form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105APPENDIX C Refill form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107Pump refill data  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107APPENDIX D Precautions and general procedures  . . . . . . . . . 109Special note and precautions  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
viiGeneral procedures   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110Locating the pump fill port and the side port . . . . . . . . . . . . . . . . . . . . . 111Accessing the pump inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112Venting the Medtronic MiniMed refill syringe (optional) . . . . . . . . . . . 113APPENDIX E Pump rinse procedure  . . . . . . . . . . . . . . . . . . . .  115Supplies and solutions  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115Preparing for the procedure  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117Prepare syringes for emptying the Pump . . . . . . . . . . . . . . . . . . . . . . . . 117Prepare syringes for filling the Pump   . . . . . . . . . . . . . . . . . . . . . . . . . . 117Program minimal basal rate  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118Remove insulin from the Pump and fill with NaOH  . . . . . . . . . . . . . . . . . . 118Equilibrate and pull NaOH through system . . . . . . . . . . . . . . . . . . . . . . . . . 120Remove NaOH and fill with rinse buffer . . . . . . . . . . . . . . . . . . . . . . . . 121Equilibrate and pull rinse buffer through system . . . . . . . . . . . . . . . . . . 122Remove rinse buffer and fill with insulin   . . . . . . . . . . . . . . . . . . . . . . . 123Equilibrate and pull insulin through system   . . . . . . . . . . . . . . . . . . . . . 125Remove guide needles and record fill amount . . . . . . . . . . . . . . . . . . . . 126Program new basal rate  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126APPENDIX F Side Port Catheter flush procedure  . . . . . . . . . . 127Supplies and solutions  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127Preparing for the procedure  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128Record patient’s blood glucose  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128Prepare syringes for emptying the Pump . . . . . . . . . . . . . . . . . . . . . . . . 129Prepare syringes for filling the Pump   . . . . . . . . . . . . . . . . . . . . . . . . . . 129Prepare syringe for flushing the Side Port Catheter   . . . . . . . . . . . . . . . 129Flushing the Side Port Catheter  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130Program minimal basal rate  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130Remove insulin and fill with rinse buffer  . . . . . . . . . . . . . . . . . . . . . . . 130Equilibrate and pull rinse buffer through system . . . . . . . . . . . . . . . . . . 131
viiiFlush side port catheter  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132Remove rinse buffer and fill with insulin   . . . . . . . . . . . . . . . . . . . . . . . 133Equilibrate and pull insulin through system   . . . . . . . . . . . . . . . . . . . . . 134Remove guide needles and record refill amount  . . . . . . . . . . . . . . . . . . 135Program new basal rate  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135Remove rinse buffer from catheter   . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136APPENDIX G Stroke volume measurement  . . . . . . . . . . . . . . . 137Supplies and solutions  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137Preparing for the procedure  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138Record patient’s blood glucose  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138Measuring stroke volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138Record patient’s blood glucose  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
ixList of figuresFigure 1: Implantable Insulin Pump and Personal Pump Communicator (PPC) . . . . . . . . . . . . . . . 1Figure 2: The Implantable Insulin Pump  . . . . . . . . . . . . . . . . . . . . 3Figure 3: Interior of the Implantable Insulin Pump  . . . . . . . . . . . . 4Figure 4: The Intraperitoneal Catheter and Side Port . . . . . . . . . . . 6Figure 5: Personal Pump Communicator (PPC) . . . . . . . . . . . . . . . 7Figure 6: Filling the hub of the refill needle   . . . . . . . . . . . . . . . . 55Figure 7: Venting the syringe head space   . . . . . . . . . . . . . . . . . . 57Figure 8: Testing Pump stroke volume with a pipette  . . . . . . . . . 60Figure 9: Tubing and retainer removal . . . . . . . . . . . . . . . . . . . . . 61Figure 10: Proper attachment of the Side Port Catheter to Pump   . 62Figure 11: Example of Pump placement  . . . . . . . . . . . . . . . . . . . . 65Figure 12: Filling the hub of the refill needle   . . . . . . . . . . . . . . . . 72Figure 13: Operation of the Pump inlet valve  . . . . . . . . . . . . . . . . 74Figure 14: Venting the Medtronic MiniMed refill syringe   . . . . . . 76Figure 15: Template and placement . . . . . . . . . . . . . . . . . . . . . . . 111Figure 16: Inlet valve  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112Figure 17: Venting the refill syringe  . . . . . . . . . . . . . . . . . . . . . . 113Figure 18: Stroke volume measurement setup  . . . . . . . . . . . . . . . 138Figure 19: Stroke volume measurement . . . . . . . . . . . . . . . . . . . . 139
x
1CHAPTER 1 DescriptionIntroductionThe Medtronic MiniMed 2007D Implantable Insulin Pump System, shown inFigure 1, brings together sophisticated new technologies to provide continuousintraperitoneal insulin therapy for patients with Insulin Dependent DiabetesMellitus (IDDM). The development of the Medtronic MiniMed 2007DImplantable Insulin Pump System is the result of years of cooperative researchand development between Medtronic MiniMed and: • The Johns Hopkins University, Applied Physics Laboratory• U.S. National Aeronautics and Space Administration, Goddard SpaceFlight Center. U.S. National Institutes of HealthFigure 1: Implantable Insulin Pump and Personal Pump Communicator (PPC)
Description2This manual is intended for use by the physician, surgeon, nurse specialist and allother members of the healthcare team who care for patients with the MedtronicMiniMed 2007D Implantable Insulin Pump System.The Medtronic MiniMed 2007D Implantable Insulin Pump System can be usedwith special U-400 insulin formulations specifically labeled for use with theMedtronic MiniMed Implantable Insulin Pump System.The system consists of three major components:• Implantable Insulin Pump• Side Port Catheter• Personal Pump Communicator (PPC)Each of these components, as well as system safety features, are discussed indetail in the following sections.
Description 3Implantable Insulin PumpThe Implantable Insulin Pump (Pump) is a round disc, 8.1 cm (3.2 inches) indiameter, 2.0 cm (0.8 inches) thick. The Pump weighs 131 grams (4.6 ounces)when empty. The outside case of the Pump is made of titanium. Titanium is abiocompatible metal used in many types of implantable medical devices. Atangential Side Port Catheter is attached to the Pump prior to implant, using alocking connector (see Figure 2).Figure 2: The Implantable Insulin PumpThe Implantable Insulin Pump is an advanced insulin infusion device withsophisticated microelectronics. It delivers a special insulin medication, using apulsatile solenoid pumping mechanism that is hermetically sealed inside thebiocompatible titanium case. Insulin delivery rates and profiles are programmedusing an external device, the Personal Pump Communicator (PPC). Specificinformation on the Implantable Insulin Pump features is outlined in the followingsections.
Description4The Pump has six major components. These components are outlined below: • medication reservoir• pumping mechanism• microelectronics• antenna• battery • tone transducerOther components of the Pump include the inlet valve, fill port, septum,cyclopentane gas and 20µm filter. Figure 3 shows the interior components of thePump.Figure 3: Interior of the Implantable Insulin PumpThe Medication Reservoir stores approximately 15 ml or 6,000 units of aspecial U-400 insulin. Depending on an individual’s insulin requirements, themedication reservoir is refilled once in approximately every two to three months.The medication reservoir is maintained at a negative pressure (vacuum) at alltimes to allow for safe and reliable filling. This vacuum prevents any risk ofinsulin leakage in the event of a breach in the Pump case or reservoir. Thereservoir is refilled with a special needle (Medtronic MiniMed Refill NeedleMMT-4102). The fill port has a 20 micron filter to prevent particulate materialfrom entering the Pump and a redundant septum and valve configuration toprevent entry of body fluids.
Description 5The Pumping Mechanism is a solenoid-operated, hermetically-welded pulsatilesystem. The pumping mechanism is designed to seal automatically to preventleakage both into and out of the reservoir under physiologic temperatures andpressures. The mechanism is designed to provide an insulin delivery accuracy ofwithin 10% from its labeled stroke volume. Individual Pumps are calibrated toone of seventeen different stroke volumes, ranging from 0.42 µL to 0.58 µL perstroke, in increments of 0.01 µL.The Microelectronics act as the brain of the Pump. The microelectronics containtwo microprocessors which monitor and control all pump-stroke activity. Allcommands delivered from the PPC via RF telemetry to the Pump are thenacknowledged back to the PPC. The Pump has a large memory which storesPump specifications and programming history.The Antenna receives radiowaves from the PPC and delivers PPC programmingcommands to the Pump microelectronics.The Battery is a custom-made lithium carbon mono-fluoride power cell, whichsupplies energy to the pumping mechanism and microelectronics. It is similar tobatteries used in pacemakers and is designed to provide 6 to 10 years of service,depending on the infusion rate (refer to pump specifications).The Tone Transducer emits beeps to indicate certain alarm conditions. Thesebeeps are designed to be audible through the skin and can be set with the PPC toone of two volumes. The Pump can also be programmed to emit beeps that signala programmed change in the medication delivery rate.Radio-Opaque Identification is featured in the Implantable Insulin Pump. Inthe event of an emergency, the name of the manufacturer and the Pump modelnumber can be identified with an x-ray.Insulin medicationOnly specially formulated U-400 insulins that are specifically labeled for usewith the Medtronic MiniMed Implantable Insulin Pump System can be used inthe model 2007D Pump.These special U-400 insulins are supplied in 10mL vials.
Description6Side Port CatheterThe Side Port Catheter (Catheter) transports insulin from the Pump into anindividual’s peritoneum where it is absorbed. The Catheter is made ofpolyethylene-lined silicone rubber, which is biocompatible with subcutaneousand intraperitoneal tissues and supports the stability of the special insulin. TheCatheter is designed with two perpendicular sections: a proximal subcutaneoussection which attaches tangentially to the Pump with a locking connector, and adistal section which is placed in the peritoneum (see Figure 4). To enable post-implant localization, a radioopaque stripe runs the length of the Catheter. The Catheter Side Port is intended to provide access to the Catheter and Pumpoutlet, in order to perform the non-surgical interventions and diagnosticprocedures described in Appendices E, F and G. The side port allows for theintroduction of a needle and small syringe to clear Catheter obstructions usingpressures up to 100 psi. It also allows for the introduction of a needle to verifyPump stroke volume.Figure 4: The Intraperitoneal Catheter and Side Port
Description 7Personal Pump Communicator (PPC)The Personal Pump Communicator (PPC) is the hand-held component of theMedtronic MiniMed 2007D Implantable Pump System (see Figure 5). The PPC allows the physician and patient to communicate with the Pump bytransmitting radio frequency messages when the PPC is held within 3” of thePump. Additionally, the PPC stores important programming information in itsmemory.Figure 5: Personal Pump Communicator (PPC) The PPC has been designed so the physician and patient can:• Program basal infusion rates (up to 48 basal rates per day, 3 differentpatterns)• Deliver or suspend an immediate, Square Wave, or Dual Wave Bolus ofinsulin• Deliver or cancel a temporary basal rate • Review the insulin delivery history• Enter personal events (meal, snack, exercise)Independent of the programming function, the PPC is able to receive and recordcertain programming data from the Pump. The recorded information is accessibleand can be displayed on the screen.
Description8
9CHAPTER 2 Indications and contraindicationsIndications for useThe Medtronic MiniMed 2007D Implantable Insulin Pump System is indicatedfor intraperitoneal administration of exogenous insulin in patients with diabetesmellitus.The model 2007D Pump can only be used with special U-400 Insulinformulations specifically labeled for use with the Medtronic MiniMedImplantable Insulin Pump Infusion System.Contraindications for use The Medtronic MiniMed 2007D Implantable Insulin Pump System iscontraindicated in patients who:• are unwilling or unable to monitor their blood glucose level at least fourtimes per day.• are unwilling or unable to make programming modifications to the Pumpbased on glucose level readings.• are unable or unwilling to administer insulin by other means, if necessary.• are unable or unwilling to comply with the guidance and advice of theirtreating physician and other healthcare providers. • reside at or travel (other than by pressurized commercial aircraft) atelevations above 8,000 feet.• have other medical or mental conditions which may place the patient at risk. • are unwilling or unable to return for routine insulin refills according to theirdosage requirements (approximately once every 90 days).• present or have a history of sensitivity to titanium alloy, polysulfone orsilicone materials used in the implanted components of the system.
Indications and contraindications10Possible adverse effectsThe model 2007D is essentially identical to the model 2007C Pump except forthe change in gas needed to maintain negative pressure in the reservoir.Evaluation of components used in the Medtronic MiniMed 2007C system hasspanned a period of approximately 4 years and involved approximately 380patients from both the U.S. and Europe. Although clinically relevant over-delivery of insulin did not occur during the 4 year evaluation period, there is apotential for such an occurrence.The following are specific adverse effects which should be understood by thephysician and explained to the patient. These do not include all adverse effectswhich can occur with surgery in general or with the use of this device, but areimportant considerations, particularly in the treatment of diabetic patients. Thegeneral surgical risks, as well as operative site cosmetic risks, should beexplained to the patient prior to surgery.• Abdominal Pain • Foreign Body Reaction• Inflammation at Refill Site • Skin Erosion• Infection • Pocket Lymph Edema• Necrosis • Hyperglycemia• Hypoglycemia • Ketoacidosis
11CHAPTER 3 Personal Pump Communicator (PPC)IntroductionThe Personal Pump Communicator (PPC) has a comprehensive set of program-ming features to control the Implantable Insulin Pump. The PPC cannot be usedby the patient until it has been initialized by the healthcare professional. Thischapter of the manual is divided in two parts:The first part (Part 1) will describe the PPC/PUMP system initialization processthat will be performed the day prior to implant.The second part (Part 2) will describe how to use the additional features that thehealthcare professional or patients can activate.PPC iconsAfter initialization, the PPC Main Screen displays the time (12hr. or 24hr. for-mat), month, day and a variety of icons. The type and purpose of these icons areas follows:Bell Icon:Displayed when the PPC receives a telemetry message from the Pump indicating that the Pump has detected an alarm condition, when a PPC error is detected and when the Pump is Suspended or Stopped.Reservoir Level Indicator Icon:The reservoir icon is composed of 4 segments that indicate how full the Pump reservoir is, based on the history of Pump delivery.Insulin Delivery Icon: The PPC simulates spinning the delivery icon when insulin delivery is in progress by displaying alternating patterns, the pattern changes every 4 seconds. When the Pump is delivering a bolus, the pattern will show three delivery segments. When the Pump is delivering a basal rate, the pattern will show one delivery segment. When the Pump is not delivering, all four segments will be displayed.
Personal Pump Communicator (PPC)12Certain features of the PPC such as programming and dosing limits can be setonly by the healthcare professional in a password-protected mode called theSupervisor Mode. Information pertaining to initializing the PPC and entering thePPC Supervisor Mode is not included in the Patient User Manual.PPC buttonsCommunicating with the PumpPlace the PPC near the Pump when the screen displays,“PPC COMMUNICATING.” The word "COMMUNICATING" will blink asindication of successful communication. If a communication link between thePPC and Pump is not established, a “TELEMETRY COMM ERROR 3” messagewill appear. The PPC will beep six times once every minute until the screen isacknowledged by pressing SEL and ACT. The screen will then display “PPCCOMMUNICATING” again.After a communication is established between the PPC and Pump and programinformation is successfully transferred to the Pump, the PPC will beep once andreturn to the Time/Date screen.• Always press the PPC buttons slowly and firmly. Wait until the screenchanges before pressing the button again. • A flashing value on the screen means that the value is activated, and can bechanged by pressing the arrow buttons.SEL(Select) The SEL button steps through each of the displays and menus.ACT(Activate) The ACT button activates programming changes in the Pump, new information to be entered into the PPC memory, and alarms to be turned off. As a safety check, ACT must be pressed to complete any programming changes. A single beep is heard after activating a change.or (Up or Down Arrows)The  or  arrows allow changes in the screen settings. Pressing  once will find the next highest setting, and pressing  once will find the next lowest setting. Holding down either button will rapidly scroll through the list of preset values. Desired values can then be programmed by pressing ACT.Sound Icon(Up Arrow) When the Audio Bolus feature is turned on, pressing  allows programming an Audio Bolus.Light Icon(Down Arrow) From the main operating screen, pressing  once will turn on the backlight. The backlight allows the Pump to be programmed in the dark. The backlight will turn off automatically after four seconds after the last button press.
Personal Pump Communicator (PPC) 13• The PPC cannot be turned off. Once the battery has been installed, the PPCis on and remains on until the battery is removed.• Certain types of Radio Frequency (RF) generating equipment could affectPPC communication with the Pump. If you are experiencingcommunication difficulties, change locations.• The time and date settings must be correct to ensure appropriate calculationof insulin delivery and display of daily totals and activity history.Install/Replace the main batteryThe battery used to power the PPC is a 1.5V AA alkaline. The life of the batteryis approximately 4 weeks during normal usage conditions. • Locate the battery door on the back of the PPC.• Slide the locking bar to the left.• Push the middle part of the PPC box (under the battery door) and lift bygently pulling up the battery door to unlatch.• Remove the old battery, noting the polarity. The screen will be blank.• Position the new battery so the + and - markings on the battery match thepolarity diagram in the battery compartment.• Close the battery door.• Slide the locking bar to the right.• The PPC screen will reappear within 30seconds:1. The PPC will beep 6 times, and after a fewseconds, the screen will display“CHECK PUMP STATUS”.PPCPUMP8 102 100 __ __ __08:26CHECKPUMP STATUSJAN O2
Personal Pump Communicator (PPC)142. Press SEL then ACT, and place the PPC near thePump.3. Wait a few seconds for the communication tocomplete. NOTE: When the PPC displays “PPCLOW BATTERY”, the message can be cleared, andprogramming continued. There should be sufficientenergy in the battery to communicate with the Pump a fewmore times, but the battery should be changed as soon aspossible.NOTE: If while programming the PPC, the screen goes blank, thePPC beeps six times and then the “CHECK PUMPSTATUS” message appears, the battery needs to bereplaced.PPCCOMMUNICATING
Personal Pump Communicator (PPC) 15Part 1: PPC/Pump system initializationThe Implantable Insulin Pump arrives from Medtronic MiniMed with preset fac-tory default values. During the initialization process these preset values aredownloaded into the PPC memory. The preset values can then be changed by thehealthcare professional, allowing the system to be personalized for each patient.The factory default values are as follows:Initialize the PPCWhen the healthcare professional receives a new PPC it must be “married” to thePump. Following are the basic steps used to initialize a Pump System the dayprior to implant: 1. The PPC is delivered without a battery in place.After installing a new battery, the PPC will beepsix times and the screen will identify the PPCsoftware used (see Chapter 3, Install/Replace theMain Battery).2. The screen now changes to“PPC NOT INITIALIZED”, and the PPC willbeep six times once every minute until theinitialization process is started. Press SEL andthen ACT, then quickly place the PPC over thePump.Bolus delivery type: Normal Locked set maximums: OffMaximum bolus: 25.0 U Password: YIQ8Audio bolus: Off Personal Events status: OFFAudio feedback: Disabled Personal ID: 000 (32 characters)Auto Off duration: Off PPC alarm type: HighBasal delivery pattern: A Refill amount: 25 gMaximum basal rate: 35.0 U/H Time format: 12 hoursInsulin concentration: 400 U/ml Variable Bolus status: OffPUMPPPC 8 102 100— — —PPC NOT INITIALIZED
Personal Pump Communicator (PPC)163. When a communication link has been established,the screen will read, “PPC COMMUNICATING”,and then will change to the next screen.4. “NO” is blinking. Check to make sure the serialnumber displayed on the screen matches thePump serial number. Press either  or  once tochange “NO” to “YES” and then press ACT.Place the PPC over the Pump.5. The screen again reads “PPCCOMMUNICATING”, and the PPC will beep 3times at the end of the programming sequence.During this process, the PPC receives all of thefactory preset values contained in the Pumpmemory.6. The screen will read “PUMP SUSPENDED”. ThePump and PPC are now “married”.7. Press SEL, then ACT and place the PPC near thePump.8. The screen now reads “PUMP INITIALIZED”.9. Press SEL and then ACT again, and the PPC willdisplay the Time/Date screen.NOTE: When the alarm type is set to“VIBRATE” the beeps from thePPC during normal programmingwill be low volume.PPCCOMMUNICATINGINITIALIZING "NO"PPC to PUMP• • • • • • • • • • • • • • • • • • • • • • • • • • • 47568COMMUNICATINGPPCPUMP SUSPENDEDPPC COMMUNICATINGPUMP INITIALIZED11:16 JAN O2
Personal Pump Communicator (PPC) 17Set the time and dateThe time and date settings must be correct to ensure accurate calculation of insu-lin delivery, daily totals, and the proper display of insulin activity history.1. Press SEL until the “SETUP PUMP” screen isdisplayed, then press ACT two times. The firsttwo digits of the time (hours) will be flashing.Use the  and  buttons to select the correcthour, then press ACT. The last two digits of thetime (minutes) will be flashing. Use the  and buttons to select the correct minute, then pressACT. Repeat the programming process to enterinformation for the year, month and day.2. After completing the programming process, thenplace the PPC near the Pump. The PPC willdisplay “PPC COMMUNICATING” whiletransferring the time and date information to thePump. The PPC will then move to the nextscreen, “AUTO-OFF.” Skip the “AUTO-OFF”screen by pressing SEL once to reach the nextscreen, “ALARMS”.08:32SETTIME-DATE 2004Jan 02PPCCOMMUNICATING
Personal Pump Communicator (PPC)18Set alarmsAlarms alert the user in the event the PPC or Pump recognizes an insulin deliveryproblem. The Alarm Feedback screen must always be in the “ON” position.1. Press ACT to enter the “ALARMS” menu.2. The PPC has three alarm options, two audibletones (Low/High) and a vibrate mode. Press the and  buttons to select the desired alarm, thenpress ACT.3. The screen will now display “SET ALARMFEEDBACK”. This setting should always be“ON”. Press ACT.4. Place the PPC near the Pump. When thecommunication is completed, the PPC screenwill change to “SELF TEST” and then to theTime/Date screen.NOTE: When the alarm type is set to “vibrate” the beeps from thePPC during normal programming will be low volume.ALARMSSETPPCALARM TYPELOW/HIGH/VIBRATESETALARMFEEDBACKON/OFFPPCCOMMUNICATINGSELF TEST
Personal Pump Communicator (PPC) 19Set maximum bolus, basal rate and time display formatThis programming is performed in the “SETUP II” menus. These screens allowhealthcare professionals to limit the maximum amount of insulin a patient candeliver, either when taking a bolus or setting a new basal rate. Access to the“SETUP II” menus is through the “SETUP PUMP” screen.1. Press SEL until the “SETUP PUMP” screen isdisplayed. Press ACT and press SEL to reachthe “SETUP II” screen. Press ACT and thenSEL to reach the “MAX BOLUS” screen.2. Press  ACT and the maximum bolus amount(units) will start flashing. Press the  and buttons to change the maximum allowable bolus(0.2 to 25.0 units) and then press ACT again.3. Place the PPC near the Pump and complete thecommunication process. The PPC screen willautomatically change to the “MAX BASAL”screen.4. Press ACT and the screen will change to “SETMAX BASAL RATE”. The maximum basalamount will start flashing.  Press the  and buttons to change the maximum allowable basalrate (0.2 to 35.0 units/hour) and then press ACTagain.5. Place the PPC near the Pump and complete thecommunication process. The PPC screen willautomatically change to “TIME FORMAT.”6. Press ACT and the screen will change to “SETTIME FORMAT.” Press the  and  buttons toselect either a 12 hour (AM/PM) or 24 hour(military time) format, and then press ACT.MAX BOLUS_ _ UMAX BOLUSSET_ _ UPPCCOMMUNICATINGMAX BASAL RATESET0.2U/HPPCCOMMUNICATINGSETTIME FORMAT 12/24 HOUR
Personal Pump Communicator (PPC)207. Place the PPC near the Pump and complete thecommunication process. The PPC screen willreturn to the “PERSONAL EVENTS” screen.Allow the PPC to time out and return to theTime/Date screen.NOTE: Adding screens to the main menu, such as “PERSONALEVENTS” increases the number of SEL button pressesrequired to reach “SETUP PUMP.”Lock maximum bolus/basal, enter personal ID and password, stop PumpTo access the Supervisor Mode press SEL until the “SETUP PUMP” screen isdisplayed. Then press and hold down the  and  buttons simultaneously untilthe “ENTER SUPERVISOR PASSWORD” screen appears.Patients should not be given the Supervisor Mode password, to avoid the accidental pro-gramming of a large priming bolus (99.8 U) or diagnostic insulin rate.1. The first zero will be flashing. Press the  and buttons to select the first digit, then press ACT.The screen advances to the second zero. Pressthe  and  buttons to select the second digit,then press ACT. Repeat for the last two digits.The factory pre-set password is YIQ8.2. Entry into the Supervisor Mode is indicated bythe screen “PUMP REFILL.”3. Press SEL until the “SET MAXIMUMS" screenis displayed, and then ACT to reach “SETMAXIMUMS”. Press the  and  buttons to select “ON” if thepatient is not given access to this feature, or“OFF” if the patient is permitted access. PressACT again.PPCCOMMUNICATING!WARNINGENTER SUPERVISOR PASSWORD0000PUMPREFILLSETMAXIMUMS  OFF/ON
Personal Pump Communicator (PPC) 214. Place the PPC near the Pump and complete thecommunication process. The PPC screen willautomatically advance to the “PERSONAL ID”screen. Press ACT.5. The first of the 32 possible ID locations isflashing. Enter the patient ID (alpha-numeric) bypressing the  and  buttons and then ACTafter each entry. Continue to press ACT,activating each “0” until the screen changes.6. Place the PPC near the Pump and complete thecommunication process.7. Press  SEL until the “SET SUPERVISORPASSWORD” screen is displayed. Then pressACT.8. The screen now reads, “SET SUPERVISORPASSWORD”. Press ACT. Use the  and buttons to enter a new supervisor password(alphanumeric), pressing ACT after each entry. 9. Record the password in the patient’s chart andimplant form.PPCCOMMUNICATINGSET PERSONAL ID• • • • • • • • • • • • • • • • • • • • • • • • • 20KOLO5PPCCOMMUNICATINGSET SUPERVISORPASSWORDYIQ8SETSUPERVISORPASSWORD0000
Personal Pump Communicator (PPC)22Program a basal rate1. From the Time/Date screen, press SEL until the“BASAL RATE” screen is displayed. Presetdelivery pattern “A”, a basal rate of 0.2 U/H, andthe word “NOW” is flashing. Press ACT. 2. A “1” now appears to the right of the “A”indicating that this programming will effect the1st basal change within the “A” pattern, (thereare 3 patterns available [A, B, C] and 48 basalchanges possible within each pattern). Theflashing 0.2 U/H indicates the value can bechanged. Use the  and  buttons to change thevalue and then press ACT. NOTE: 00:00 indicates a start time of MIDNIGHT in 24hr.display mode. In 12hr. display mode, the screen indicatesthe start time as 12:00am.3. The screen now displays “SET TIME”, and atime of 00:30 or 12:30 am (24 or 12 hourrespectively) and a “2.” If one basal is all that will be used, press ACTtwo times. If more than one basal rate is to beprogrammed, enter a start time and amount of the new basal rate for that timeperiod, then press ACT and enter the new basal rate. The user can enter anew basal rate at 30 minute intervals, up to 48 basal rates.4. Place the PPC near the Pump and complete thecommunication process.5. The PPC will briefly display the calculated totalbasal dose for 24 hours, based on the values andtimes entered in the Basal Rate programmingscreen. In this example, the total basal dose is4.8 U/day.08:26 OCT 12 BASAL RATE: A  NOW 100:00 0.2U/H BASAL RATE A00:30 _ _ U/H s 2PPCCOMMUNICATING24 HOUR TOTAL4.8U
Personal Pump Communicator (PPC) 23Part 2: Additional PPC programming featuresMain menuThis second part will describe how to program the additional features that thepatient or the healthcare professional can activate.Program a bolusA properly initialized PPC is now ready to program a bolus.The PPC/Pump allows you to set and deliver a bolus of insulin whenever needed.The PPC has several special features which allow you to customize the program-ming and delivery of boluses.• Normal Bolus and Audio Bolus• Square Wave Bolus• Dual Wave BolusNOTE: To use the Variable Bolus programming options, (e.g.square, dual), this option needs to be programmed “ON”in the SETUP II menu. If it is not “ON” only the defaultbolus, “Normal bolus”, will be available.
Personal Pump Communicator (PPC)24Set a Normal bolus with the Variable Bolus feature turned off1. From the Time/Date screen, press SEL.The “BOLUS” screen is displayed, with the timeand date flashing.2. Press  ACT and the “SET BOLUS” screenappears. The dashes under “IMM” are flashing.Press the  and  buttons to enter an immediatebolus amount.3. Press  ACT and the “CONFIRM” screen isdisplayed, with the screen flashing. Confirm thebolus amount by pressing ACT again.4. Place the PPC near the Pump and complete thecommunication process.5. When the bolus programming is complete, thePPC will beep once and then briefly display theamount of insulin currently delivered.The Pump will beep at each of the first fivestrokes (if audio feedback is ON). The PPCbeeps and at the end of the bolus. Threesegments of the insulin delivery icon will bedisplayed and spinning slowly during the bolusdelivery. By pressing SEL you can read theamount of insulin delivered.08:13BOLUS PROGIMM10U 10UJas2--EXTSET BOLUSIMM_ _ UCONFIRM2.6UIMMPPCCOMMUNICATING08:13  BOLUS 0.0UJan  02
Personal Pump Communicator (PPC) 25Set a Normal bolus with the Variable Bolus feature turned on1. From the Time/Date screen press SEL until the“BOLUS” screen is displayed. The last bolusvalue programmed and the Time and Date willbe flashing.2. Press ACT and the “SET BOLUS TYPE” screenappears. If “NORMAL” is not flashing, use the and  buttons to select “NORMAL.” PressACT.3. The “SET BOLUS” screen appears, with dashesunder “IMM” flashing. Use the  and  buttonsto enter an immediate bolus amount.4. Press  ACT and the “CONFIRM” screen isdisplayed, with the screen flashing. Confirm thebolus amount by pressing ACT again.5. Place the PPC near the Pump and complete thecommunication process.6. When the bolus programming is complete, thePPC will beep once and then briefly display theamount of insulin currently delivered.The Pump will beep at each of the first fivestrokes (if audio feedback is ON). The PPCbeeps at the end of the bolus. Three segments ofthe insulin delivery icon will be displayed andspinning slowly during the bolus delivery. Bypressing SEL you can read the amount of insulindelivered.08:13BOLUSIMM_ _ U _ _ UJan 02EXTSETNORMALBOLUS TYPESET BOLUSIMM_ _ UCONFIRM IMM2.6UPPCCOMMUNICATING08:23BOLUS 5.0UJAN 04
Personal Pump Communicator (PPC)26Set a Square Wave BolusA Square Wave Bolus of insulin is delivered evenly over a preset period of time,from 30 minutes to 4 hours. A Square Wave Bolus may be desirable when eatinglong meals such as at banquets or receptions, high fat meals, or to compensate forgastroparesis. During a Square Wave Bolus, the programmed basal rate is deliv-ered simultaneously to the bolus.To access this feature you must first turn the Variable Bolus feature “ON” in the“SETUP II” menu.1. From the Time/Date screen, press SEL. The“BOLUS” screen is displayed, showing the lastbolus programmed with the time and dateflashing. 2. Press ACT and the “SET BOLUS TYPE” screenappears. Press the  and  buttons to select“SQUARE”. Press ACT.3. The “BOLUS” screen appears, with dashes under“EXT” flashing. Use the  and  buttons toenter an extended bolus amount. Press ACT.4. Blinking dashes will now appear under the bolusamount entered. Use the  and  buttons toenter a time duration for the Square Wave Bolus,in one-half hour increments from 30 minutes to 4hours. 5. Press ACT and the “CONFIRM BOLUS” screenis displayed, with the screen flashing. Confirmthe Square Wave Bolus by pressing ACT again.6. Place the PPC near the Pump and complete thecommunication process.08:13BOLUS PROGIMM10U _ _U--Jan 02EXTSETSQUAREBOLUS TYPESET BOLUSIMM_ _U 4.0UEXTSET BOLUSIMM2:00EXT_ _U 4.0UCONFIRM BOLUS    IMM              EXT2:00_ _U 4.0UPPCCOMMUNICATING
Personal Pump Communicator (PPC) 277. When the bolus programming is complete, thePPC will beep once and then briefly display theamount of insulin currently delivered.The Pump will beep at each of the first fivestrokes (if audio feedback is ON). The PPCbeeps at the end of the bolus. Three segments ofthe insulin delivery icon will be displayed andspinning slowly during the bolus delivery. Bypressing SEL you can read the amount of insulindelivered.Set a Dual Wave BolusThe Dual Wave Bolus programs a Normal Bolus immediately followed by aSquare Wave Bolus.To access this feature you must first turn the Variable Bolus feature ON in theSETUP II menu.1. From the Time/Date screen, press SEL. The“BOLUS” screen is displayed, showing the lastbolus programmed with the time and dateflashing.2. Press ACT and the “SET BOLUS TYPE” screenappears. Press the  and  buttons to select“DUAL.” Press ACT.3. The “SET BOLUS” screen appears, with dashesunder “IMM” flashing. Use the  and  buttonsto enter the immediate portion of the Dual WaveBolus. Press ACT.4. The “SET BOLUS” screen now shows dashesflashing under “EXT.” Use the  and  buttons to enter the extendedportion of the Dual Wave Bolus. Press ACT.16:06                   JAN 03BOLUS 0.0U08:13BOLUS_ _IMM_ _UJan 02_ _UEXTSETDUALBOLUS TYPESET BOLUSIMM2.0USET BOLUS    IMM              EXT2.0U 2.0U
Personal Pump Communicator (PPC)285. Blinking dashes will now appear under the bolusamount entered.Use the  and  buttons to enter a time durationfor the Square Wave Bolus, in one-half hourincrements from 30 minutes to four hours.6. Press ACT and the “CONFIRM BOLUS” screenis displayed, with the screen flashing. Confirmthe Dual Wave Bolus by pressing ACT again.7. Place the PPC over the pump and complete thecommunication process.8. When the bolus programming is complete, thePPC will beep once and then briefly display theamount of insulin currently delivered.The Pump will beep at each of the first fivestrokes (if audio feedback is ON). The PPCbeeps at the end of the bolus. Three segments ofthe insulin delivery icon will be displayed andspinning slowly during the bolus delivery. Bypressing  SEL, you can read the amount ofinsulin delivered.Review bolus historyTo review the type, amount, time and day of your last 512 insulin boluses.1. From the Time/Date screen, press SEL. The“BOLUS” screen is displayed, showing the lastbolus programmed. Use the  and  buttons to display previousboluses, along with the time and day each boluswas delivered.SET BOLUS    IMM              EXT2:002.0UCONFIRM BOLUSIMM2:00EXT2.0U 2.0UPPCCOMMUNICATING16:06                   JAN 03    BOLUS 2.0U08:23                  Jan 02BOLUSIMM                   EXT_ _UPROG_ _U_ _
Personal Pump Communicator (PPC) 29Suspend modeThe Suspend Pump mode allows the user to cancel a bolus delivery, while stilldelivering a basal rate of 0.2 U/hr.1. From the Time/Date screen, press SEL until the“SUSPEND PUMP” screen is displayed. PressACT. The screen will display a flashing“SUSPEND PUMP” message. Press ACT again.2. Place the PPC near the Pump and complete thecommunication process.3. When the communication is complete, the Pumpwill beep 3 times and the PPC screen will changeto “PUMP SUSPENDED”. All four segments ofthe insulin delivery icon are shown. The PPCwill beep every half-hour as long as the Pumpremains suspended.NOTE: To restart the Pump, press SEL. The “PUMPSUSPENDED” screen will begin flashing. Then pressACT.SUSPEND PUMPPPCCOMMUNICATING08:13PUMP SUSPENDEDOCT 12
Personal Pump Communicator (PPC)30Programming a basal rateBasic basal rate programming was described earlier in this chapter. This sectiondescribes additional basal rate options.Programming basal delivery patternThe PPC allows three basal delivery patterns. One such basal pattern could beused for a working day, another for a weekend day, etc. Each of the basal deliverypatterns is a set of up to 48 basal rates, one for each half-hour of the day. PatternA is the factory pre-set. To access profiles B or C you must enter the“SETUP PUMP” screens. 1. Press SEL until the “SETUP PUMP” screen isdisplayed. Press ACT. Press SEL again to accessthe basal rate profile screen, “DELIVERYPATTERN”.2. Press ACT and the screen will change to “SETDELIVERY PATTERN”. Use the  and  keysselect the pattern preferred; A, B, or C.  Eachpattern can contain up to 48 different basal rates.Press ACT after choosing a pattern.3. Place the PPC near the Pump and allow thecommunication to complete.NOTE: When the PPC times out, press SEL until the“BASAL RATE” screen is displayed. The basal patternselected in "SETUP PUMP" will now appear to the rightof “BASAL RATE” A, B, or C.DELIVERYPATTERN ASETDELIVERY                A,B,CPATTERNPPCCOMMUNICATING
Personal Pump Communicator (PPC) 31Setting basal rate profiles in each delivery patternEach of the basal delivery patterns is a set of up to 48 basal rates, one for eachhalf-hour of the day.1. Press SEL until the “BASAL RATE” screen isdisplayed. A basal pattern is selected (forexample Pattern A). Press ACT.2. A “SET RATE” and “1” is now displayed to theright of the “A” indicating this programming willset the 1st basal rate within the “A” profile. The“0.2U/H” is now flashing, indicating the valuecan be changed. Use the  and  arrow keys toprogram a new value, for example, “0.4 U/H”,and then press ACT.NOTE: 00:00 indicates a start time of MIDNIGHT in 24hr.display mode. 12:00am indicates a start time ofMIDNIGHT in 12hr. display mode.3. The screen now displays “SET TIME”, with atime of “00:30” or “12:30am” flashing (24 or 12hour respectively) and a “2.” This screen allowsthe second basal rate to be set. Enter a start timefor the 2nd basal rate within the “A” profile, forexample “04:30.” Press ACT. (Example: a second basal rate of 0.4U/Hstarting at 04:30.)4. This screen changes to “SET RATE” again,indicating the 2nd basal rate can now beprogrammed. Use the  and  arrow buttons toenter a new rate, for example “0.2U/H”, and thenpress ACT.5. A “3” now appears on the screen with “SETTIME.” Follow the same procedure previouslydescribed and program a new profile. If noadditional profiles are needed change theflashing time to dashes (by pressing ) and pressACT. BASAL RATE: A   NOW00:00 0.2U/H BASAL RATE: A SET RATE 100:00 0.2U/H SET TIME04:30 0.4U/H BASAL RATE: A 2                                    SET RATE BASAL RATE: 204:30 0.4U/H                                    SET TIME05:00 BASAL RATE: 30.2U/H
Personal Pump Communicator (PPC)326. The screen will indicate“PPC COMMUNICATING.” Place the PPC nearthe Pump and complete the communicationprocess.7. The screen will briefly display the calculated 24hour basal dose based on the basal rateprogramming. In this example a total of “8.4U”will be delivered.To set multiple basal profiles in the other patterns(A,B,C), select the pattern in "SETUP II" menuand follow the same procedure.Program a temporary basal rateA Temporary Basal Rate is often used when a brief change in basal delivery isrequired, for example during exercise.Set a temporary basal rate1. From the time and date display press SEL until the“TEMP BASAL” screen is displayed.2. Press  ACT and the “SET DURATION” screenappears. The time duration of the Temporary BasalRate is displayed as flashing dashes. Press the and   buttons to enter a time duration, in 30minute increments from 30 minutes to 24 hours.3. Press ACT and “SET AMOUNT” screen appears.The amount of the Temporary Basal Rate is nowflashing. Press the  and  buttons to enter adelivery rate. Press ACT again.PPCCOMMUNICATING24 HOUR TOTAL8.4U TEMP BASAL _ _  _ _ U/HTEMP BASAL_ _U/H00:30SET DURATIONTEMP BASAL1.5U/H00:30SET AMOUNT
Personal Pump Communicator (PPC) 334. Place the PPC near the Pump and complete thecommunication process. 5. When the communication is complete, the Pumpwill beep once. The PPC screen will briefly showthe “TEMP BASAL” screen before returning tothe Time/Date screen.NOTE: When the Pump is delivering a Temporary Basal rate, thefirst screen displayed when SEL is pressed is “TEMPBASAL.” The PPC will also beep every 30 minutes toalert the user that a Temporary Basal rate is currentlyactive.Stop a temporary basal rate1. From the Time/Date screen press SEL until the“TEMP BASAL” screen is displayed. Press ACTand the “SET DURATION” appears, with the timeduration flashing. Press  once until it resets todashes. Then press ACT.2. Place the PPC near the Pump and complete thecommunication process.3. When the communication is complete, the PPCwill briefly return to the “TEMP BASAL” screenwith the amount dashes flashing. Allow the PPCto return to the Time/Date screen.PPCCOMMUNICATING TEMP BASAL08:13 OCT 02TEMP BASAL_ _                1.5U/HSET DURATIONPPCCOMMUNICATINGTEMP BASAL_ _ SET AMOUNT_ _U/H
Personal Pump Communicator (PPC)34Personal eventsThis feature allows the user to enter event codes into the PPC memory, andrecord the time and date of entry. Preset event codes are: 1 = meal, 2 = snack, 3 =sick and 4 = exercise. In addition, other event codes A, B and C can be entered torecord other important events. These other event codes should be documentedprior to their use. To access the “EVENTS” screen in the main menu, “ON” must be activated inthe “SETUP II” menu, “PERSONAL EVENTS” screen.1. From the Time/Date screen, press SEL until“SETUP PUMP” is displayed, then press ACT.Press SEL until “SETUP II” is displayed, thenpress  ACT. Press SEL until “PERSONALEVENTS” screen is displayed. 2. Press ACT and “ON” or “OFF” begins flashing.3. Use the  and  buttons to select “ON”, thenpress ACT again. The “PERSONAL EVENTS”screen will now appear on the main menu. Allowthe PPC to return to the Time/Date screen.4. To set an event: From the Time/Date screen pressSEL until the “EVENT” screen is displayed thenpress  ACT. The screen changes to “SETEVENT” with the word “MEAL” flashing. Usethe  and  buttons to select the desired event.5. Press  ACT and the current time will appearflashing. Use the  and  buttons to enter thetime the event occurred. Press ACT when theproper time is displayed. Then allow the screenreturn to Time/Date.NOTE: Only historic or current event times can be entered.NOTE: If the Personal Events feature is turned “OFF” in“SETUP II”, events cannot be entered into the PPC. OFFPERSONALEVENTSEVENTS PERSONAL ON/OFFSETPERSONALEVENTS ON/OFFSETEVENTMEALSET TIME EVENTMEAL07:32Am
Personal Pump Communicator (PPC) 35HistoryHistorical Pump data, such as insulin medication remaining, amount of bolus andbasal delivery since the last refill, etc., can be accessed and read on the PPC.1. From the Time/Date screen press SEL until the“HISTORY” screen is displayed. Press ACT andthe “READ PUMP DATA” screen will appearflashing. Press ACT again.2. Place the PPC near the Pump and complete thecommunication process. The PPC will acquiredata from the Pump.3. The screen will change to “MED REMAINING”,indicating the estimated amount of insulinmedication remaining in the Pump. Record thisnumber if required.4. Press SEL and the screen will read “INSULINTOTAL.” Delivered amounts of basal and bolusinsulin medication are displayed for the dateflashing on the screen. Use the  and  buttonsto review other daily totals.5. Press  SEL and the screen will change to“CLINICAL HISTORY PPC”. Use the  and buttons to review other PPC events.6. Press  SEL and the screen will change to“CLINICAL HISTORY PUMP”. Use the  and buttons to review Pump events.READ PUMP DATA00000000000000000000000000010057PPCCOMMUNICATINGMED REMAINING2263U             Jan 02INSULIN TOTALBASAL        BOLUS 9U 22U6:16CLINICAL HISTORY11PPCJan 026:26CLINICAL HISTORY14 PUMPJan 02
Personal Pump Communicator (PPC)367. Press  SEL and the screen will change to“EST PUMP BATT”. This screen indicates thePump battery status, during no-load (batterypower is not used) and load (extendedcommunication sequence) conditions. The Pumpis set to alarm for low battery when the loaded(LD) voltage is at or below 2.5 volts.8. Press  SEL and the screen will read“EXIT HISTORY.” Press ACT. The PPC willreturn to the Time/Date screen.Jan 02EST PUMP BATTERY 2.9VNO LD          LD2.7V    EXIT HISTORY
Personal Pump Communicator (PPC) 37Setup PumpThe “SETUP PUMP” screen permits access to the primary Setup menus for thePump. Setup features discussed previously in this chapter are referenced here.Other Setup features not previously discussed are presented here. Press SEL untilthe “SETUP PUMP” screen is displayed, then press ACT to access the SETUPmenus.Time and DateThis feature is previously described in this Chapter. Auto offAuto Off is a safety feature, reminding the user to update insulin medicationdelivery programming in the Pump. An alarm can be set after a period of pro-gramming inactivity, from one to 16 hours. At the onset of the alarm the Pumpwill automatically be placed in SUSPEND mode.1. From the “SETUP PUMP” screen, press ACT.Press  SEL to reach the “AUTO OFF” screen,then press ACT. The screen will display “AUTOOFF” with flashing dashes/time. Use the  and buttons to select the number of hours beforean “AUTO OFF” alarm occurs.2. Press ACT. In this example, a time duration of10 hours was selected. The PPC will alarm if theuser did not attempt to program the Pump duringthe previous 10 hours and be placed in"SUSPEND" mode.3. Place the PPC near the Pump and complete thecommunication process.The PPC will beep onceand return to the Time/Date screen.AUTO OFF_ _ HRSSETAUTO OFF10 HRSPPCCOMMUNICATING
Personal Pump Communicator (PPC)38AlarmsThis feature is previously described in this Chapter. Self test1. Self Test allows the user to perform a diagnostictest of the Pump and PPC operating system.Messages are relayed between the PPC andPump. From the “SETUP PUMP” screen, pressACT. Press SEL until the “SELF TEST” screenis displayed. Press ACT.2. Place the PPC near the Pump and complete thecommunication process. 3. Verify that the following events occur:• A series of tones will be heard from the Pump (4 beeps).• An alarm tone will be heard from the PPC.• The backlight on the PPC will turn on.• The PPC screen will briefly activate all possible display icons, numbers,etc.• The PPC will vibrate. NOTE: If the above test results do not occur or the screendisplays irregular characters, please notifyMedtronic MiniMed. If the PPC displays the MedtronicMiniMed logo and software version, the PPC hasrestarted due to a low battery. Replace the batteryimmediately.If all electronics “pass” the Self Test, the screen will automatically change to“PPC PASSED/PUMP PASSED.” After several seconds, the screen will return tothe Time/Date screen.SELF TESTPPCCOMMUNICATING
Personal Pump Communicator (PPC) 39Basal delivery patternThis feature is previously described in this Chapter. Initialize PPC to PumpFrom the “SETUP PUMP” screen press ACT until “INITIALIZE PPC TOPUMP” is displayed. This option “marries” the PPC to the Pump, and it is used toinitialize a new PPC. For initialization programming follow the steps describedin, Part 1: PPC/Pump System initialization of this Chapter.Setup II“SETUP II” permits access to an additional group of Setup menus for the Pump.From the “SETUP PUMP” screen, press ACT until the “SETUP II” screen is dis-played. "SETUP II" features are presented in the next section.Exit Setup menu1. From the "SETUP PUMP” screen, press SELuntil the “EXIT SETUP MENU” screen isdisplayed. Press ACT. The PPC will return to theTime/Date screen.EXIT SETUP MENU
Personal Pump Communicator (PPC)40Setup II Audio BolusThe Audio Bolus feature allows the user to deliver a bolus without looking at thePPC screen. There are two Audio Bolus increments, 0.4U and 0.8U. With eachpress of the  button, the PPC will beep one time (0.4U setting) or two times(0.8U setting), depending on the delivery amount selected per button press.Activating the Audio Bolus feature1. From the “SETUP II” screen, press ACT. The“AUDIO BOLUS" screen is shown.2. Press  ACT, and the screen changes to “SETAUDIO BOLUS” with “OFF/ON” flashing. Usethe  and  buttons to choose “ON” to activatethe Audio Bolus feature or “OFF” to disable it.Then press ACT.3. If “ON” was chosen, the “SET STEP AUDIOBOLUS” appears with a step value of 0.4U or0.8U flashing on the screen. Use the  and buttons to choose between a step rate of 0.4U or0.8U. Press ACT. The Audio Bolus feature isnow activated.  NOTE: If “OFF” is chosen, pressing the  button will have noeffect.AUDIO BOLUSON 0.4USETONAUDIO BOLUS0.4USET STEPAUDIO BOLUS0.4U
Personal Pump Communicator (PPC) 41Set an Audio Bolus from the main menu bolus screen1. From the Time/Date screen press the  button.The PPC beeps either one or two times,depending upon the audio bolus step amount set.The user counts the number of beeps todetermine how much insulin was programmed.2. In this example the PPC was programmed for a0.8U increment, and the PPC beeps two times forthe 0.8U bolus. Press ACT and the audiosequence repeats to confirm the bolus amount.Press  ACT twice and the Pump will beprogrammed. The PPC beeps once when theprogramming is completed. Place the PPC on the Pump to complete thecommunication process.NOTE: While an audio bolus is being delivered, the  buttonwill not function.Variable BolusTo access the Variable Bolus feature enter “SETUP II” and press SEL one time.1. The screen will display “VARIABLE BOLUS”. Press ACT.2. The screen will now display “SET VARIABLEBOLUS” and “ON or OFF” will be flashing. Usethe   and  buttons to select “ON” or “OFF”,then press ACT.3. Place the PPC near the Pump to completecommunication. If “ON” was chosen theVariable Bolus option is now available in theMain Menu/Bolus screen.NOTE: If “OFF” is chosen, the VariableBolus option (Square, Dual) will not be available in the(Main Menu) Bolus screen.SET BOLUS 0.8UIMMPPCCOMMUNICATINGVARIABLE BOLUSONSET BOLUSPPCCOMMUNICATING
Personal Pump Communicator (PPC)42Maximum BolusThis feature is previously described in this Chapter. Maximum Basal RateThis feature is previously described in this Chapter.Time display formatThis feature is previously described in this Chapter.Personal EventsThis feature is previously described in this Chapter.Pump SetupThis screen allows the user to return to the setup mode "TIME/DATE" screen bypressing ACT.Exit Setup MenuPress SEL 7 times in the SETUP II mode. Pressing ACT will return the user tothe Main Menu "TIME/DATE" screen.
Personal Pump Communicator (PPC) 43Supervisor modeTo access Supervisor Mode (by a health care professional only) press SEL 6times in the "SETUP PUMP" screen, then press and hold the  and  buttonssimultaneously. Hold them down until the screen changes the "ENTER SUPER-VISOR PASSWORD" screen.RefillThe refill procedure will be described in Chapter 5 “Pump Refill Procedure.” Thefollowing screens are used at the end of this procedure, and allow you to enter thenew refill volume and to calculate the refill accuracy. 1. The "PUMP REFILL" screen appears when youaccess the Supervisor Mode. Press ACT.2. The PPC prompts for the residual amount ofinsulin removed from the Pump. Use the  and buttons to enter the weight (grams) ofextracted insulin.3. Press  ACT and the PPC prompts for the refillinsulin amount. Use the  and  buttons toenter the weight (grams) of insulin refilled in thePump.4. Press  ACT. Confirm that the insulinconcentration defaults to U-400. Press ACTagain.5. Place the PPC near the Pump to complete theprocess.PUMP REFILLSET EXTRACTED AMOUNT---- ---- gSET REFILL AMOUNT---- ---- gSET INSULIN                    U400CONCENTRATIONPPCCOMMUNICATING
Personal Pump Communicator (PPC)446. The calculated accuracy will be displayed briefly.Record this value on the refill form.PrimingA priming bolus is not intended as a therapeutic bolus. It should only be used when the Pump is not implanted.The Priming Bolus is used during the preparation of the Pump for implantation.This feature “Primes” the Pump piston chamber, outlet port and Catheter. ThePump will pulse approximately 500 times during this function. To access this fea-ture, the Pump must be in “PUMP STOPPED” mode. 1. From the “PUMP STOPPED” screen, press SEL.The screen will begin flashing.2. Press the  and  buttons at the same time untilthe “ENTER SUPERVISOR PASSWORD”screen appears.3. Re-enter the supervisor password, pressing ACTafter each letter or number is entered. The factorypreset password is: YIQ8. If the password hasbeen changed, enter the new password. When theSupervisor Password is successfully entered, thescreen will change to “PUMP REFILL”. PressSEL until the “PRIMING” screen appears.4. Press ACT and “NO” appears flashing. Press the and  buttons to change to “YES”, then pressACT.CALCULATEDACCURACY                         _ _ %!WARNING08:13        Oct 12PUMP STOPPEDENTER SUPERVISORPASSWORDYIQ8PRIMINGPRIMING                        NO99.8U
Personal Pump Communicator (PPC) 455. The word “CONFIRM” now appears. Press ACTto activate the factory preset priming bolus of99.8U.6. Place the PPC over the Pump and complete thecommunication process.7. The progress of the Priming Bolus can beverified by pressing SEL. The Priming Bolustakes approximately 10 minutes to complete.NOTE: The PPC will alarm every minute during the PrimingBolus. The Priming Bolus can be terminated at any timeby using the “SUSPEND PUMP” feature.                                       CONFIRMPRIMING                        YES99.8UPPCCOMMUNICATING08:13       OCT 12     PRIMING                  0.0U
Personal Pump Communicator (PPC)46Diagnostic rateA diagnostic rate is intended for use only when the Pump is filled with rinse buffer. The Diagnostic Rate feature is used to help “diagnose” Pump delivery problems.Fill the Pump with rinse buffer before using this feature. To access this feature,the Pump must be in the “PUMP STOPPED” mode. 1. From the “PUMP STOPPED” screen, press SEL.The screen will begin flashing.2. Press the  and  buttons at the same time untilthe “SET SUPERVISOR PASSWORD” screenappears. Re-enter the supervisor password,pressing  ACT after each letter or number isentered. The factory preset password is: YIQ8.3. When the Supervisor Password is successfullyentered, the screen will change to “PUMPREFILL”. Press SEL until the “DIAGNOSTICRATE” screen appears.4. Press ACT and the word “SET” appears flashing.Use the  and  buttons to enter a DiagnosticRate (units/hour).5. Press  ACT and the word “SET” changes to“DELIVER”. “NO” also appears flashing. Usethe  and  buttons change “NO” to “YES”,then press ACT. !WARNING08:13          OCT 12   PUMP STOPPED                        0000SETSUPERVISORPASSWORD _ _ U/HDIAGNOSTIC RATE 10 U/HSETDIAGNOSTIC RATE 10 U/HDELIVERDIAGNOSTIC RATE                          NO
Personal Pump Communicator (PPC) 476. The entire screen will be flashing. Press ACTagain to confirm the Diagnostic Delivery Rate.7. Place the PPC near the Pump and complete thecommunication process.8. The words “DIAGNOSTIC RATE” appearbriefly on the screen.The PPC then returns to theTime/Date screen.NOTE: Press SEL and the PPC will indicate“DIAGNOSTIC RATE.”NOTE: Diagnostic Rate Delivery can be terminated at any timeby using the “SUSPEND PUMP” feature.Initialize to factory defaultsThis feature resets all programmed parameters in the Pump to preset factorydefault values. To access this feature, the Pump must be in the “PUMPSTOPPED” mode. 1. From the “PUMP STOPPED” screen, press SEL.The screen will begin flashing. 2. Press the  and  buttons at the same time untilthe “ENTER SUPERVISOR PASSWORD”screen appears. Re-enter the SupervisorPassword, pressing ACT after each letter ornumber is entered. The factory preset passwordis: YIQ8. 10 U/HCONFIRMDIAGNOSTIC RATEPPCCOMMUNICATING08:13          OCT 12   DIAGNOSTIC RATE08:13          oct 12   PUMP STOPPED     ENTERPASSWORDSUPERVISOR 0000
Personal Pump Communicator (PPC)483. When the Supervisor Password is successfullyentered, the screen will change to “PUMPREFILL”. Press SEL until the “INITIALIZE TOFACTORY DEFAULTS” screen appears.4. Press  ACT and the word “CONFIRM” willappear. Press ACT again to activate.5. Place the PPC over the Pump and complete thecommunication process.6. The PPC will now read “PUMP SUSPENDED.”Press SEL then ACT.7. Place the PPC over the Pump and complete thecommunication process.8. The PPC now displays the “PUMP RESET”screen, and beeps six times every minute untilthe Pump is restarted. Press  SEL then ACT. The PPC returns to theTime/Date screen. The Pump may now bereprogrammed.INITIALIZE DEFAULTSTO FACTORYINITIALIZE DEFAULTSTO FACTORYCONFIRMPPCCOMMUNICATING08:13                 OCT 02PUMP SUSPENDEDPPCCOMMUNICATING08:13                 OCT 02PUMP RESET
Personal Pump Communicator (PPC) 49Stop PumpThis screen allows the Physician to stop the operation of the Pump. 1. From the “PUMP REFILL” screen in theSupervisor Menu, press SEL until the “STOPPUMP” screen appears.2. Press ACT and the word “CONFIRM” appearshighlighted with the entire screen flashing. PressACT again to confirm.3. Place the PPC near the Pump and allow thecommunication to complete.4. The PPC will beep four times to confirm the“STOP PUMP” command. To restart the Pump,press SEL and “PUMP STOPPED” will start toflash. Press ACT. Place the PPC near the Pump.When the communication is complete, the PPCreturns to the Time/Date screen.Supervisor passwordIf the password is lost, access the Supervisor Menu as follows:• Program the time to midnight (12:00 am or 00:00) and the date toJanuary 01.• Enter the password 0000 within one minute. If the password is notentered within one minute the time will be reset to midnight again.• Record the new password in the patient’s chart.Exit supervisorTo exit the Supervisor Programming Menu, press SEL until the “EXIT SUPER-VISOR” screen appears, then press ACT.STOP PUMP STOP PUMP CONFIRMPPCCOMMUNICATING08:13                 OCT 02PUMP STOPPED
Personal Pump Communicator (PPC)50Personal Pump Communicator messagesDisplay Screen Message Message Meaning or Action RequiredAUTO OFF PUMP SUSPENDED Auto Off time interval has elapsed. Pump operation is suspended.BOLUS 0.0U A bolus has been programmed and is being delivered.CHECK PUMP STATUS The PPC battery has been replaced. The PPC needs to check the Pump status.LOW RESERVOIR The PPC has recognized the Pump has 800 units (2 ml) or less insulin remaining in its reservoir. Schedule a Pump refill as soon as possible. Allowing the reservoir to completely empty may damage the Pump.EMPTY RESERVOIR  The PPC has recognized the Pump has 400 units (1 ml) or less insulin remaining in its reservoir. Schedule a Pump refill as soon as possible. Allowing the reservoir to completely empty may damage the Pump.HOURLY MAX EXCEEDED You attempt to deliver more than 2.5 times the bolus maximum in one hour. To clear the message, press SEL and ACT. You may exceed this limit by programming another bolus within 10 minutes.PPC LOW BATTERY The PPC battery is low. The PPC can be used for at least 24 hours. Replace the AA battery immediatelyPPC NOT INITIALIZED The PPC has not been initialized to a Pump.PUMP LOW BATTERY The Pump battery energy is low but still functioning. Schedule a replacement as soon as possible.PUMP INITIALIZED The PPC has been “married” to a Pump. Press SEL and ACT.PRIMING XX U A Priming Bolus has been programmed and is being delivered.PUMP RESET The Pump was reprogrammed to the preset factory values.PUMP SELF TEST FAIL A Pump malfunction was detected during a Self Test. PUMP STOPPED The Pump is in stop mode.PUMP STOPPED1 or 2 or 3 or 4 or 5 or 6 When the Pump recognizes a system malfunction, it automatically stops and insulin delivery ceases. PUMP SUSPENDED The Pump is in suspend modePUMP ERROR0 or 1 User attempts to initialize the PPC to a Pump that is not compatible with it. Clear by pressing SEL and ACT. Verify the personal ID of the Pump responding corresponds to the personal ID entered into the PPC.PUMP ERROR40 or 41 The Pump has invalid data. The PPC will not initialize to Pump.
Personal Pump Communicator (PPC) 51Technical history codesIn the “HISTORY” menu, the user can access clinical history from the PPC andthe Pump. Each number code corresponds to a particular event. The followingtable lists each of the number codes with their corresponding events. Technical history PPC codesTELEMETRY COMM ERROR 3The PPC and Pump are not communicating. Reposition the PPC over the Pump, then press SEL and ACT. If the error message persists, contact Medtronic MiniMed.TELEMETRY COMM ERROR 20 Move the PPC away from any other Pump and perform the request again.Display Screen Message Message Meaning or Action RequiredClinical Event Code Clinical Event CodeVersion Error 1 Time Out 2No Response 3 Retry Packet 4No Synchronization 5 Bad CRC 6Invalid transmission 7 RX Overflow 8Invalid Op code 9 TX Underflow 10Unknown IRQ Vector 12Bolus Total Error 13 NMI Occurred 14Bad Duplicates 15 Suspend Alarm 16Stop Pump Alarm 17 Diagnostic Rate Alarm 18Prime Pump Alarm 19 Exclusion List Full 20Max Clock Stealer 21 Min Clock Stealer 22Non Initialized PPC 23 Pump Self Test Error 24Pump Initialized 25 Pump Reset to Defaults 26Hourly Maximum 28
Personal Pump Communicator (PPC)52Clinical history Pump codesRefill Divide Error 29 Refill Invalid Calculation 30Divide by 0 31 EEPROM Error 32Bad EEPROM 33 Main Battery Low 34Temporary Basal Rate 37 Check Pump Status 38Invalid Concentration 40Invalid Stroke Volume 41 Battery Removed 42Clinical Event Code Clinical Event CodeIP Communication Error 1 Charge Time Too Long 2Post-Fire Voltage Reading too High 3 Over-Delivery Error 4Under-Delivery Error 5 Dead Battery 6Auto Off Interval Exceeded 7 Low Reservoir 8Empty Reservoir 9 Low Battery 10Normal Delivery Mode 11 Stop Delivery Mode 12Suspend Delivery Mode 13 Diagnostic Rate delivery mode 14Priming Bolus Delivery Mode 15 Insulin Concentration Change 16Clinical Event Code Clinical Event Code
53CHAPTER 4 Pump implantationPreprogramming and pre-testing the PumpEvery Implantable Insulin Pump is fully tested before shipment and is accompa-nied by a package insert indicating the measured stroke volume and Pump type.Prior to shipping, each Pump is filled with sterile rinse buffer to assure sterility. Programming of the Pump may be performed a day before Pump implantation.The Implantable Insulin Pump may be programmed through the outer tray of thesterile package. First, the PPC needs to be “married” to the Pump, then the Pumpfunction is verified by performing the initialization procedure described in Chap-ter 3, PPC/Pump System Initialization.Registration cardTo ensure proper patient identification and device serial number tracking, theRegistration Card that accompanies each Pump must be completed and promptlyreturned to Medtronic MiniMed.Supplies and solutionsBefore preparing the Pump, be sure to read Appendix D, Precautions and Gen-eral Procedures. Assemble the necessary materials prior to starting the proce-dure:Supplies• One (1) 100 µL sterile pipette (available non sterile from Medtronic MiniMed) MMT-4104• One (1) sterile scalpel blade• One (1) scale (0.01g resolution)• One (1) sterile beaker (or sterile barrier towel) for weighing• One (1) sterile bag for PPC
Pump implantation54• One (1) Pump System: PPC, Pump and Side Port Catheter• Sterile markers and Steri-Strips®• Three (3) refill Kits - MMT-4105:• 1 kit to remove shipping fluid from Pump• 1 kit to fill the Pump with U-400 INSULIN (to rinse the pump)• 1 kit to fill the Pump with U-400 INSULIN (final insulin fill)• One (1) 3 mL syringe (fluid barrier)• Three (3) Medtronic MiniMed refill needles, MMT-4102• Three (3) 18 gauge needles• One (1) Implant Worksheet• One (1) Back-up Pump System: PPC, Pump, Port CatheterSolutions• One (1) bottle of sterile water, room temperature• One (1) vial 10 mL Rinse Buffer (RB), MMT-2008• Four (4) vials 10 mL U-400 INSULIN, stored at room temperature for aminimum of ten hours prior to surgery.Emptying and filling the PumpTo prepare the Pump for implantation, the Pump must be emptied and then filledtwice and tested with insulin. These procedures are performed in the operatingroom prior to surgery. As this is a surgery, all supplies and required equipmentshould be prepared in accordance with the institution’s approved sterile proce-dures. When you remove the Pump from the sterile box, do not remove the plastic tubing placed at the Pump outlet. Trim the distal part with a scalpel blade. This tubing will be used for the “Stroke Volume Mea-surement” procedure.A sterile field is established in the operating room to prepare the RINSEBUFFER (RB) syringe and remove the shipping fluid from the Pump. Documentthe Pump, Catheter and PPC serial numbers on the Implant Worksheet.!WARNING
Pump implantation 55Remove shipping fluid from the Pump1. Using aseptic technique, open the Pump sterile package. Do not remove theplastic tubing from the Pump outlet. Trim the distal part of the tubing withthe scalpel blade.2. Remove the refill syringe from the refill kit package.3. Firmly attach the two-way stopcock to the refill syringe and attach an 18gauge needle to the stopcock.4. Draw approximately 4 mL of RINSE BUFFER into the refill syringe.5. Fill the hub of the refill needle with RINSEBUFFER and firmly attach it to thestopcock, prime the needle and close thestopcock. (See Figure 6.)6. Retract the plunger until it locks into place.This should be no further than 55 mL. Donot go beyond the vent hole. Press the lockinto the plunger groove to be sure it is firmlysecured.7. Fill one 3 mL syringe with RINSE BUFFER(RB) and attach a needle.8. Fill the Pump fill port using the 3 mLsyringe. Any time a refill needle is to beinserted into the Pump fill port, a fluidbarrier must be present to prevent air fromentering the Pump reservoir.9. With the stopcock still closed, insert the RBsyringe into the Pump fill port. 10. Open the stopcock and allow the syringevacuum to empty the Pump of shippingfluid. Allow 30 seconds after the fluid levelappears to have stopped rising in the refillsyringe, to assure the Pump is completelyemptied. Close the stopcock and remove thesyringe.Figure 6: Filling the hub of the refill needle
Pump implantation56Rinse the Pump with insulin (IN1)1. Remove the second refill syringe from the refill kit package.2. Firmly attach the two-way stopcock to the refill syringe and attach an 18gauge needle to the stopcock. Use the sterile marker to label the syringe,“IN1”.3. Draw 20 mL of INSULIN (two vials) into the refill syringe.4. Remove the 18 gauge needle and expel all air bubbles in the syringe.5. Close the stopcock.6. Retract the plunger until it locks into place. This should be no further than 55mL. Do not go beyond the vent hole. Press the lock into the plunger groove tobe sure it is firmly secured.7. Shake vigorously for a minimum of 30 seconds to degas the INSULIN.8. Point the syringe tip upward and slowly open the stopcock to vent thesyringe.9. Release the locking ring on the refill syringe.10. Expel air in the syringe and carefully observe to ensure no air bubbles remainin the syringe. If air bubbles are noted, repeat steps 5-10.11. Fill the hub of the refill needle with INSULIN and attach it to the stopcock.12. Prime the needle and close the stopcock.13. Refill the fill port with the 3 mL RINSE BUFFER syringe, as needed, tomaintain the fluid barrier.14. Use the “IN1” syringe containing the degassed INSULIN to fill the Pump.OPTIONALUse these steps only if the plunger is not moving forward while filling the reser-voir.1. With the refill needle pointing down, vent the syringe head space by pullingback firmly on the plunger until the second sealing ring on the black rubbercap passes beyond the vent hole (see Figure 7).
Pump implantation 57Figure 7: Venting the syringe head space2. With the stopcock closed, insert the refill needle into the Pump fill port.3. Open the stopcock. Maintain downward pressure on the barrel of the syringeto ensure the inlet valve of the Pump remains open.Never push on the refill syringe plunger to fill the Pump. When the refill needle is properly seated in the fill port of the Pump, the vacuum in the reser-voir will draw the insulin from the syringe.4. Allow the Pump vacuum to draw the INSULIN into the reservoir. When theINSULIN stops moving, the Pump is filled. Close the stopcock and removethe syringe.5. Prepare the IN1 syringe for aspiration. Remove any air from the syringe,prime the needle, close the stopcock and obtain a vacuum by pulling back onthe plunger until it locks. Press the lock into the plunger groove to be sure itis firmly secured. A fluid barrier of at least 2 mL should remain in thesyringe.!WARNING
Pump implantation586. Insert the refill needle back into the Pump fill port with the stopcock closed.7. Open the stopcock and empty the Pump. Allow 30 seconds after theINSULIN level appears to have stopped rising in the refill syringe to assurethe Pump is completely emptied.8. Close the stopcock and remove the syringe.Fill the Pump with insulin1. Remove another refill syringe (IN2) from a refill kit package.2. Firmly attach the two-way stopcock to the refill syringe and attach an 18gauge needle to the stopcock. Label the syringe IN2.3. Draw 20 mL of INSULIN (two vials) into the IN2 syringe.4. Remove the 18 gauge needle and expel all air bubbles in the syringe.5. Close the stopcock.6. Retract the plunger until it locks into place. This should be no further than 55mL. Do not go beyond the vent hole. Press the lock into the plunger groove tobe sure it is firmly secured.7. Shake vigorously for a minimum of 30 seconds to degas the INSULIN.8. Point the syringe tip upward and slowly open the stopcock to vent thesyringe.9. Release the locking ring on the refill syringe.10. Expel the air in the syringe and carefully observe it to ensure no air bubblesremain in the syringe. If air bubbles are noted, repeat steps 5 - 10.11. Fill the hub of the refill needle with INSULIN and attach it to the stopcock.12. Prime the needle and close the stopcock.13. Place a sterile beaker or sterile barrier towel on the scale and tare it (positionbalance at 0.)14. Weigh the refill syringe and record the weight on the Implant Worksheet.15. Insert the refill syringe into the Pump fill port, open the stopcock and allowthe Pump vacuum to draw INSULIN into the reservoir until the fluid levelstops moving. Close the stopcock.
Pump implantation 5916. Remove the IN2 syringe and prepare for aspiration without expelling anyfluid. Obtain a vacuum by pulling back on the plunger until it locks. Press thelock into the plunger groove to be sure it is firmly secured.17. Reenter the Pump fill port with the stopcock closed.18. Open the stopcock and withdraw 2 mL in order to activate the negativepressure safety feature of the Pump. Close the stopcock and remove thesyringe.19. Weigh the IN2 syringe to calculate the Pump refill amount. Subtract theweight after the fill and 2 mL withdrawal from the weight before the fill andrecord the difference on the Implant Worksheet.Measure stroke volume1. Place the PPC in the sterile bag. Enter the Supervisor Mode. The screendisplays “PUMP REFILL”. Press ACT until the “SET REFILL AMOUNT”screen appears. When filling the Pump for the first time, set the“EXTRACTED AMOUNT” to zero. Otherwise, enter the amount of insulinextracted. Then use the  and  buttons to enter the refill amount from theImplant Worksheet. Program the Pump.2. Again, enter the Supervisor Mode and program “STOP PUMP”. From thescreen “PUMP STOPPED”, press SEL. “PUMP STOPPED” will beginflashing. Press both the  and  buttons at the same time, until the screenchanges to the password screen. Reenter the password and the screen willchange to “PUMP REFILL.” Press SEL to reach the “PRIMING” screen.“NO” will be flashing. Use the  and  buttons to select “YES,” then pressACT. When the word “CONFIRM” is highlighted, press ACT again. Placethe PPC near the Pump to complete the communication process. Record thetime the priming bolus was initiated on the implant worksheet.3. Listen for five beeps and observe that fluid is being pumped from tubingattached to the Pump outlet.4. Attach the 18 gauge needle (from the 3 ml syringe filled with RINSEBUFFER) to the plastic tubing from the Pump outlet.5. Remove the 3 ml syringe and fill the needle hub.
Pump implantation60Figure 8: Testing Pump stroke volume with a pipette6. Insert the pipette into the hub. A meniscus of fluid will be present in thepipette. This should be done on a flat surface with the pipette/tubingassembly horizontal (see Figure 8).7. Note the exact location of the fluid meniscus in the pipette.8. Measure the stroke volume by noting the volume displaced by a minimum of10 pulses. Record the total volume delivered and the number of pulses on theImplant Worksheet. Divide the total volume delivered by the total number ofpulses used to deliver that volume. The volume per stroke must be between0.40 and 0.60 µL. Record this value on the Implant Worksheet.Prepare the Side Port CatheterThe Side Port Catheter (“Catheter”) is attached to the Pump before implantation;however, it must be primed with INSULIN and inspected before being attachedto the Pump.Never insert a needle into the connector end of the Catheter for priming or for testing. There is a preci-sion sealing surface in the connector that may be damaged by the insertion of any needle.1. Keep the Catheter in its package until the Pump is prepared and strokevolume has been verified. Remove the Catheter from its sterile package anddocument the serial number on the Implant Worksheet.2. Using the last insulin syringe (IN2) from the Pump preparation procedure,enter the Catheter port cone and push a small volume of insulin out of theconnector end of the Catheter. This primes the proximal portion of theCatheter and cleans any debris from the connector outlet. Remove thesyringe.!WARNING
Pump implantation 613. (See Figure 9.) Remove the plastic tubing and tubing retainer from the Pumpoutlet. Remove the tubing retainer by rotating it one-quarter turn clockwise todisengage it from the connector flange. Then lift and remove the tubing andretainer together from the Pump outlet. Dispose of the tubing and retainer. Figure 9: Tubing and retainer removalView ARotate tubing retainerView BPull tubing retaineroff pumpView CTubing and retainerremoved
Pump implantation624. Inspect the outlet of the Pump for debris. Debris, especially fibers, left in thefluid path will compromise the function of the Catheter check valve. Rinsethe outlet tube of the Pump with water from a syringe, if necessary. Avoidwiping the Pump with gauze or drapes that will leave fibrous debris. 5. Pull the locking bar out and gently push the Catheter straight down onto thePump outlet. Never twist the connector.6. Only after the Catheter is seated, squeeze the connector (see Figure 10, item1), firmly against the Pump between the thumb and forefinger of one hand.Then push the locking bar (item 2) into place with the thumb of the otherhand.Figure 10: Proper attachment of the Side Port Catheter to Pump7. Verify the Catheter is securely attached to the Pump. Look into the connectorand observe the position of the locking bar with respect to the Pumpconnector outlet. Check the side port connector hub for abnormalities (e.g.,cracks, misalignment.)8. Using the IN2 refill syringe, enter the Catheter side port cone and push asmall amount of INSULIN out the Catheter tip. Verify that no insulin flow isobserved at the Pump/Catheter interface. Remove and reseat the Catheter ifflow is noted. Remove the syringe.9. Confirm insulin delivery from the tip of the Catheter and note it on theImplant Worksheet. The INSULIN will form a dome on the tip of theCatheter. Pulsatile delivery will not be obvious. If the priming bolus hasalready ended, program an immediate bolus (about 20 units) to confirmdelivery. (This bolus may be canceled after delivery is verified.)
Pump implantation 6310. At the end of the bolus, rinse the outside of the Pump thoroughly with sterilewater to remove traces of INSULIN. The Catheter tip should not be placed inthe water. Protect the Pump with sterile drapes until the surgeon is ready toimplant it. Verify that the priming bolus is finished or canceled prior to theimplantation.The Pump and Catheter are now ready for implantation.
Pump implantation64Performing the surgical procedurePre-operative evaluationThe Pump implant depth should not exceed 5 cm (2 inches) beneath the surfaceof the skin. If the Pump is implanted too deeply in the subcutaneous tissue it maybe difficult to access the refill port.Formation of the pump pocketPump implantation can be performed either under local or general anesthesia.After satisfactory anesthesia, a pre-selected abdominal site is prepped and drapedin the customary surgical method.A transverse or longitudinal incision is made and deepened through the skin andsubcutaneous tissue. Care should be taken in choosing the location of the incisionso the suture line is next to but not over the Pump inlet fill port or the Catheterside port. A subcutaneous blunt dissection is performed to create a pocket largeenough to accommodate the Pump. Meticulous hemostasis should be establishedwith electrocautery. The Pump is secured to the fascia of the abdominal musculature using the threesuture tabs provided. Securing the Pump with all three suture tabs is recom-mended, as it will prevent migration, rotation or inversion in the pocket and pos-sible dislodgment of the Catheter.Figure 11 indicates a possible Pump placement and corresponding incision sites.
Pump implantation 65Figure 11: Example of Pump placement
Pump implantation66Catheter placementThe Catheter is not designed to be tunneled subcutaneously. The Catheter inser-tion should be at the extreme edges of the Pump pocket.At a site not directly beneath the Pump, the layers of abdominal musculature aredivided to expose the peritoneum. Concentric purse string sutures are placed andthe peritoneum is exposed with a 1 cm incision. The distal tip of the Catheter isplaced in the peritoneal space. The flange of the Catheter is sutured firmly to thefascia of the abdominal musculature. Sutures should only be placed on the rein-forced flange of the Catheter. Sutures must not be placed directly on the Catheteritself.Never trim or cut the tip of the Catheter.The implant site should be irrigated with an antibiotic solution. Interruptedabsorbable sutures should be used to approximate the subcutaneous tissues andthe skin closed with a running subcuticular absorbable suture. The wound isdressed using standard surgical technique. At the end of the surgery, place anabdominal binder over the implant site. This binder must be worn until the cap-sule has formed (about 1 month.) The binder is recommended because it willobtain a better cosmetic result and minimize the post operative swelling.!WARNING
Pump implantation 67Post-operative managementPost-operative hospitalizationAfter implantation, the surgical incision should be inspected for any signs ofunusual swelling, tenderness, pain, or drainage. Potential post-operative compli-cations include Pump pocket seroma, wound dehiscence, wound infection, andcatheter dislodgment.Length of post-operative hospitalization depends upon how rapidly the patientadjusts to the Pump and how quickly blood glucose levelsstabilize. During post-operative hospitalization, the Pump can be reprogrammedto accommodate the needs of the patient. As post-operative stress decreases,insulin requirements may decline. It may be necessary to make several changesin insulin delivery rates during the first few days following implantation.Patients must be thoroughly educated in all aspects of follow-up care with thePump. The Medtronic MiniMed Patient Manual should be used to supplementpatient education. Prior to discharge, patients should:• Fully understand how to use their PPC;• Know what to do in the event of Pump or PPC difficulties;• Have scheduled their first refill appointment;• Have set an appointment for their first follow-up visit;Post-operative x-raysThe Catheter has been designed with a radio-opaque stripe for identification afterimplantation. A lateral and anterior X-ray is recommended after implantation tolocate and document the position of the Catheter. This radiograph should be keptwith the patient’s records for future reference.
Pump implantation68
69CHAPTER 5 Pump refill procedureIntroductionInitial appointments for Pump Refills should be made with patients at the time ofimplantation. Subsequent appointments should be made in advance, normallyduring each refill procedure. The Pump stores approximately 6,000 units of insu-lin medication and typically requires a refill every two to three months. In orderto maintain the physical stability of the insulin, the Pump must be refilled at amaximum interval of no more than 90 days. The PPC will display warning mes-sages to alert the patient of calculated low or emptied reservoir volumes.The low reservoir message will appear when the PPC calculates that 800 (2 ml)or fewer units of insulin medication remain in the Pump. The empty reservoirmessage will appear when the PPC calculates that 400 or fewer units (1 ml) ofinsulin remain. The time to medication depletion depends upon the insulin deliv-ery rate of the Pump.The Pump refill procedure should always be sched-uled in advance of PPC “low reservoir” or “empty reservoir” messages.!WARNING
Pump refill procedure70Supplies and solutionsThe refill kit (MMT-4105) is intended exclusively for use with theMedtronic MiniMed 2007D Implantable Insulin Pump System. The refill kit con-sists of a specially designed syringe and stopcock for removing fluids from andplacing fluids into the Pump. Use only refill needles (MMT-4102) available fromMedtronic MiniMed to perform all Pump refill procedures. This 22 gauge needle,featuring a rounded, lubricated tip and a beveled side hole, has been speciallydeveloped to mate with the Pump fill port.Use of other needles may result in damage to the Pump septum in the fill port, and may allow body fluids to enter the Pump.Below is a list of the supplies and solutions necessary to perform the refill proce-dure:Supplies:• Steri-Strips® and Markers• The Patient’s PPC, Glucose Monitoring Equipment, Drapes• One (1) Scale (0.01 gram resolution)• One (1) 250 ml Sterile Beaker• One (1) Port Locating Template - MMT-4106• Two (2) Refill Kits - MMT-4105 - Three (3) Medtronic MiniMed MMT-4102 Refill Needles (Extra needles should be available for use.)• Four (4) 18 Gauge Regular Bevel NeedlesSolutions:• One (1) 10 ml Vial, Rinse Buffer Solution, MMT-2008 • Two (2) 10 ml Vials, U-400 Insulin, stored at room temperature for aminimum of 10 hours prior to surgery.!WARNING
Pump refill procedure 71Prepare for pump refillNOTE: Before beginning any refill, rinse, flush, or pressuremeasurement procedure, carefully read Appendix D,Precautions and General Procedures. Never push on the refill syringe plunger to fill the Pump. When the refill needle is properly seated in the Pump fill port, the vacuum in the Pump reser-voir will draw the insulin from the syringe into the reservoir.To prepare for emptying and refilling the Pump, perform the following steps:• Use the “HISTORY” feature of the PPC to determine the amount of insulinmedication remaining in the Pump. Make sure that ACT is pressed on the“READ PUMP DATA” screen to update PPC history from the Pump.Record this value on the refill worksheet (Appendix C).• Prepare a sterile field. The physician should scrub, mask and glove for therefill procedure. Refilling the Pump is an aseptic procedure.• (See Appendix D, Precautions and General Procedures.)• Aseptic skin preparation of the patient’s Pump refill site should beperformed using your institution’s standard operating procedures.!WARNING
Pump refill procedure72Perform the refill procedureFill out the refill formFill out the top part of the Refill Form (Appendix C) with the patient’s I.D., refilldate, insulin lot number, previous refill volumes, insulin medication remaining,and the name of the person performing the refill procedure. This information canbe helpful for diagnostic purposes.Label syringesUse a sterile marker and Steri-Strips® to label one refill syringe “RB” (rinsebuffer syringe) and the other “IN” (insulin syringe).Prepare the refill syringe for emptying the Pump1. Firmly attach a stopcock to the RB syringe andattach the 18 gauge needle to the stopcock.Check connections to ensure they are secure.Draw 5 ml of Rinse Buffer into the RB refillsyringe. 2. Expel all air bubbles from the refill syringe.Remove the 18 gauge needle from thestopcock.3. Fill the hub of the refill needle with RinseBuffer from the RB refill syringe and attach itto the stopcock. (See Figure 12) Prime the refillneedle completely. Close the stopcock.4. Obtain the weight and record it on the RefillForm. Set the syringe aside.Figure 12: Filling the hub of the refill needle
Pump refill procedure 73Prepare the refill syringe for filling the Pump1. Firmly attach a stopcock to the IN refill syringe and attach an 18 gaugeneedle to the stopcock. Draw 20 ml of U-400 insulin into the syringe.2. Expel all air from the refill syringe. Close the stopcock and remove the 18gauge needle.3. Retract the plunger until it locks into place. This should be no further than 55ml - do not go beyond the vent hole. Press the lock into the plunger groove tobe sure it is firmly secured.4. Shake vigorously for a minimum of 30 seconds.5. Point the syringe tip upward and slowly open the stopcock to allow air toenter the syringe.6. Release the locking ring on the refill syringe. With the syringe pointedupward, push on the plunger to expel the air and prime the stopcock.7. Open the stopcock and prime the refill needle. Attach the refill needle to thestopcock. Close the stopcock.8. Inspect the syringe to ensure there are no air bubbles. If air bubbles are noted,remove the refill needle and repeat steps 2 - 8.9. Weigh the IN syringe and record on the refill form.10. Set the syringe aside.Empty the Pump1. Obtain a vacuum in the RB syringe by pulling back on the plunger until itlocks. Press the lock into the plunger groove and be sure it is firmly secured.2. Locate the fill port by centering the template over the Pump. The center holein the template should be directly over the Pump fill port. Insert the 18 gaugeguide needle into the Pump fill port. Local anesthesia may be used prior toinsertion of the 18 gauge needle.3. Press the primed refill needle (attached to the RB syringe) into the Pump fillport, by sliding it through the 18 gauge guide needle. Pull back slightly(approximately 2 mm) on the guide needle to allow the refill needle to enterthe Pump fill port and actuate the inlet valve.(See Figure 13)
Pump refill procedure74Figure 13: Operation of the Pump inlet valve18G Guide NeedleRefill NeedleValveFilterSeptumSpring2mmStep 3AStep 3BStep 3CStep 3A: Locate the port using the 18G guideneedle slide the refill needle down theguide needle.Step 3B: Pull back approximately 2mm on theguide needle to enter the Pump.Step 3C: Apply firm pressure to move the valveback and allow the insulin to pass abovethe valve through the filter and into thereservoir.
Pump refill procedure 754. Actuate the valve by applying a gentle downward force on the refill needle.This depresses the valve approximately 0.5 mm.5. Open the stopcock and allow residual insulin from the Pump to be drawn intothe RB refill syringe. Maintain a downward pressure on the syringe to ensurethat the inlet valve in the Pump remains open. Wait 30 additional secondsafter the insulin level appears to have stopped rising in the RB refill syringeto ensure the Pump is completely emptied.6. When all residual insulin is withdrawn from the Pump, close the stopcock.Remove the RB refill syringe.NOTE: Do not remove the 18 gauge guide needle.7. Weigh the RB syringe and record on the refill form.
Pump refill procedure76Refill the Pump1. Use the IN refill syringe containing 20 ml of degassed insulin, at roomtemperature.2. Enter the Pump by passing the refill needle with the attached IN refill syringethrough the 18 gauge guide needle. Pull back slightly (approximately 2 mm)on the 18 gauge needle to allow the refill needle to enter the Pump andactuate the inlet valve (see Figure 13).3. Open the stopcock. Maintain downward pressure on the IN refill syringe toensure the inlet valve in the Pump remains open. Allow the Pump to draw ininsulin until the fluid level stops moving. Close the stopcock. 4.  Remove the IN syringe and prepare for aspiration without expelling any fluidby pulling back on the plunger until it locks. Re-enter the Pump, open thestopcock and withdraw 2 ml in order to activate the negative pressure safetyfeature of the Pump.OPTIONALUse this step only if theplunger is not movingduring the fill of thereservoir. With the refill needlepointing down, vent thesyringe head space bypulling back firmly on theplunger until the secondsealing ring on the blackrubber cap passes beyondthe vent hole (see Figure14).Figure 14: Venting the Medtronic MiniMed refill syringe
Pump refill procedure 775. Remove the IN syringe. Remove the 18 gauge needle and apply pressure tothe insertion site.6. Weigh the IN syringe and record on the Refill Form. Calculate extracted and refill amounts1. Calculate the extracted amount by using the calculation section of the refillform.2. Calculate the refill amount by using the calculation section of the refill form.3. Enter the extracted amount (amount withdrawn as calculated on the refillworksheet at line E) and the refill amount (as calculated on the refillworksheet at line G) into the PPC.Calculate refill accuracy The PPC automatically calculates the refill accuracy following the PPC refill pro-gramming. You can also follow the calculation section of the refill form to calcu-late the refill accuracy.Enter the refill accuracy value (in %) on the line provided on the refill form. It isimportant to record this value to evaluate system function.
Pump refill procedure78
79CHAPTER 6 Explanting the Pump System Explant considerationsWhen the battery in the Pump is depleted or if acceptable glycemic control can-not be achieved, it may become necessary to explant the Pump system.Prior to explantation, it may be possible to perform interventions that could cor-rect certain conditions. These procedures are described in Appendices E and F. If all appropriate interventions have been exhausted without an acceptable out-come, please contact Medtronic MiniMed prior to scheduling a Pump explanta-tion.Returning devices/components to Medtronic MiniMedExplanted Pumps and Catheters or other components should be returned toMedtronic MiniMed for evaluation. Please call Medtronic MiniMed and obtain aReturned Sales Order (RSO) number prior to each return.Explanted devices must be sealed in an appropriate biohazard container andpacked with a gauze pad soaked with sterile saline. The shipping containershould be water tight. Chemical and reliability analysis require that the devicenot dry out during transportation. Be certain to include required patient information as well as the RSO number,date, and reason for the explant and place all pertinent documentation in a watertight document package. Put the RSO number on the shipping label.
Explanting the Pump System80Please take the necessary precautions when shipping the Pump System via com-mercial carrier to avoid damage to the Pump. Please return explanted Pump Sys-tems to:United States:Medtronic MiniMed18000 Devonshire StreetNorthridge, CA 91325Telephone: 1-818-576-5555Europe:Medtronic EOCEarl Bakkenstraat 106422 PJ HeerlenNetherlandsTel 3 145 56 68 000or
81CHAPTER 7 Warnings and precautions WarningsThe Physician should be completely familiar with the function of the Pump,Catheter, and PPC prior to use. Patients should be provided a complete copy ofthe Patient Manual and have demonstrated the ability to program the PPC,recognize and respond to safety alarms, and take care of the device prior todischarge.Only special U-400 insulins specifically labeled for use with the MedtronicMiniMed Insulin Infusion Systems may be used in the Medtronic MiniMed2007D Implantable Insulin Pump. Use of other insulin types may cause damageto the Pump mechanism resulting in impaired insulin delivery or Pump failure.Any unauthorized changes or modifications made to any component of theMedtronic MiniMed 2007D Implantable Insulin Pump System may preventeffective use of that and other components.ElectrotherapyThe Medtronic MiniMed 2007D Implantable Insulin Pump System has beentested in close proximity with electrosurgical, electrocoagulation, and cardiacdefibrillation medical equipment. Typical use of this type of device has notaffected the Pump. However, patients should be instructed to test Pump function(e.g., self test) after such procedures to determine that the Pump and PPC areoperating properly. If the system is not performing correctly, contact MedtronicMiniMed.
Warnings and precautions82Diagnostic ultrasound The Medtronic MiniMed 2007D Implantable Insulin Pump System has beentested during diagnostic ultrasound procedures. These procedures have no effecton Pump performance. However, patients should be instructed to test the Pumpfunction (e.g., self test) after such procedures to determine that the Pump andPPC are operating properly. If the system is not performing correctly, contactMedtronic MiniMed.Ultrasound therapyThe Medtronic MiniMed 2007D Implantable Insulin Pump System should not beexposed to therapeutic ultrasound procedures such as lithotripsy. Exposure toultrasound therapy may damage the Pump, and not be immediately apparent.Diagnostic radiationThe Medtronic MiniMed 2007D Implantable Insulin Pump System has beentested during diagnostic radiation procedures, such as Computed Tomographyand X-ray. These procedures have no effect on Pump performance. However,patients should be instructed to test the Pump function (e.g., self test) after suchprocedures to determine that the Pump and PPC are operating properly. If thesystem is not performing correctly, contact Medtronic MiniMed.Elevated anti-insulin antibodiesThe results of clinical investigations suggest a small population of patients maydevelop anti-insulin antibodies when using the Medtronic MiniMed 2007DImplantable Pump System and U-400 insulin. Patients with multiple autoimmunedisorders may be more susceptible to developing high titers of anti-insulinantibodies, which in turn may cause symptoms. There was no correlationbetween length of exposure to the study drug and the resultant titer of anti-insulinantibodies. Elevated levels of anti-insulin antibodies alone have been proven notto interfere with diabetes management using continuous intraperitoneal insulininfusion with this system.
Warnings and precautions 83Environmental conditionsThe Medtronic MiniMed 2007D Implantable Insulin Pump System should not beexposed to extreme electrical or magnetic fields. Although it is not possible toinclude every environmental condition that may affect the Pump, most are listedbelow. Please contact Medtronic MiniMed when in doubt if certainenvironmental exposure will affect the pump.DO NOT EXPOSE THE MEDTRONIC MINIMED 2007D IMPLANTABLE INSULIN PUMP SYSTEM TO ANY OF THE FOLLOWING CONDITIONS:• Magnetic Resonance Imaging (MRI) Procedures.• Lithotripsy Treatment.• Large Rotating Magnetic Fields - Refers to large industrial rotatingmagnetic fields, such as those found in Industrial Power Plants.• Magnets Held Directly Over the Pump.• High Power Radio or Satellite Transmitting Towers.• Altitudes Above 2,400 meters (8,000 feet).• Therapeutic Ultrasound• Therapeutic RadiationThe Medtronic MiniMed 2007D Implantable Insulin Pump System is notdesigned for use at elevations above 2,400 meters (8,000 feet). Use of the PumpSystem at higher elevations may result in insulin over delivery, which may causepersonal injury or death. Therefore, patients who live at elevations above 2,400meters (8,000 feet) should not use the Pump. Patients who use the Pump and whoplan to travel to an elevation above 2,400 meters (8,000 feet) should first havetheir Pump reservoir emptied of insulin, and they must self-administer insulin byother means for the duration of the trip and until their Pump reservoir is refilledagain.(This warning does not apply to travel on commercial aircraft because normalcabin pressurization is adequate to ensure proper operation of the Pump.)• Diving Below 7.6 Meters (25 Feet).• Physical Damage to the Pump or Pump Pocket.!WARNING
Warnings and precautions84Physicians should instruct patients to avoid any sharp blows or pressure directlyat the Pump location. A direct hit by an object such as a baseball may damage thePump and/or injure the Pump pocket. Extreme pressure on the Pump may causethe sutures and the tissue pocket to be damaged. As a result, the Pump may movefrom its pocket and possibly dislodge from the Catheter.If the Pump has been damaged by such a blow, the patient should be instructed tocontact their physician. Physicians may order a pressure bandage or a reductionin the patient’s physical activity until the pump pocket reheals completely.• Exposure to Extreme Body TemperaturesExposure to elevated body temperatures in excess of 104°F (40°C) willcompromise the negative pressure reservoir safety feature of the MedtronicMiniMed 2007D Implantable Pump. Glucose levels should be monitored closelyif this occurs.SterilizationThe Pump, Catheter and the refill kit are sterilized with ethylene oxide (EtO).The Pump System is NOT Reusable.If the sterile package has been opened, damaged or tampered with DO NOT USEthe packaged device.NOTE: Do Not Re-Sterilize the Implantable Insulin PumpNOTE: Do Not Re-Sterilize the Side Port CatheterNOTE: Do Not Re-Sterilize Components of the Refill Kit.The Pipette (MMT-4104), may be resterilized using autoclave, EtO or accordingto hospital procedures.The Template (MMT-4106) may be resterilized using steam autoclave or EtO perhospital procedure. This template is made from polycarbonate plastic and willhave a limited life in steam autoclave cycle. The template will not withstand dryheat autoclave cycles.
Warnings and precautions 85PrecautionsEmergencies and the use of conventional insulin suppliesPhysicians should advise patients who are implanted with the MedtronicMiniMed 2007D Implantable Insulin Pump System how to deal with emergencyconditions such as hyperglycemia and hypoglycemia. Patients should alwayscarry conventional insulin supplies with them, including insulin and a means toinject it, in the event of impaired insulin delivery by the Pump System. Deliveryof insulin can become impaired due to a failure of a Pump and/or PPC, or aCatheter occlusion. Replacement of the Pump, PPC or Catheter may be required.Physicians should review the Pump replacement and Catheter clearingprocedures in this Physicians Manual PPC reliability requirementsThe PPC is a sensitive electronic device, and can incur physical damage. If thePPC is dropped or receives an impact patients should be instructed toimmediately perform a SELF TEST to check the displays for proper operation. Ifthe PPC does not display correctly, a replacement PPC is required. Patientsshould initiate alternative diabetes management until a replacement PPC isreceived.The PPC housing is not watertight and it may malfunction if immersed in water.“Condensing humidity” conditions such as steam rooms should also be avoided,because condensation can also damage the PPC’s microelectronics. If eithersituation occurs, contact Medtronic MiniMed immediately and arrange for therepair or replacement of the PPC. If the PPC is accidentally splashed, sprayed orimmersed, remove excess moisture with a soft towel and then place the PPC in awarm place to thoroughly dry. When dry, perform a “SELF TEST.” If the PPCdoes not display correctly, call Medtronic MiniMed to replace the device. Maximum dosagesThe physician can program specific limitations to insulin Basal Rates and Bolusamounts, as well as total daily insulin usage. These limitations provide somecontrol of patients’ ability to program their insulin regimens, and to avoidoverdosing.
Warnings and precautions86Electrical and magnetic fieldsCommon electrical and magnetic fields that do not affect the Pump includemicrowave ovens, satellite receiving dishes, common household appliances,security devices found in department stores and airports, standard medical X-rays, cellular phone, and radiowaves.
87CHAPTER 8 Adverse reactionsAdverse reactionsIn clinical studies, adverse reactions associated with the Medtronic MiniMed2007D Implantable Pump included hypoglycemia, diabetic ketoacidosis,hyperglycemia, skin erosion, infection, abnormal healing, elevated anti-insulinantibodies, intestinal obstruction, post-operative discomfort and pain. PumpSystem malfunctions in order of frequency and seriousness include, insulinaggregation resulting in insulin under-delivery, Catheter occlusion or tissueovergrowth, premature depletion of the Pump battery, and failure of Pumpelectronics. Adverse events associated with the use of U-400 insulin aredescribed in the package insert accompanying the insulin medication.PreventionThe majority of adverse reactions in patients using the Medtronic MiniMed2007D Implantable Pump System can be prevented by teaching patientsdependable blood glucose monitoring. The patient plays a significant role indiagnosing and correcting Pump System performance problems. Should Pumpperformance change, the patient would be able to detect a change in bloodglucose levels.Patients should be instructed to contact their physi-cian’s office, if they experience unresolvable difficul-ties with the Pump System. Patient visits and diagnostic procedures may be necessary to correct these conditions. !WARNING
Adverse reactions88
89CHAPTER 9 System alarms and messagesThe Medtronic MiniMed 2007D Implantable Insulin Pump System is equippedwith various alarms and messages that ensure the correct function of the system.The Implantable Insulin Pump alarm system will “beep” when certain conditionsoccur. The beeps are designed to be audible through the skin and alert the patientthat the Pump needs attention. For severe alarm conditions the Pump will alarm 4tones each minute for 10 minutes then, 4 double tones each minute for 10 min-utes and repeat the pattern. Upon hearing the alarm, the patient must communi-cate with the PPC, to determine the alarm condition. The alarm can be cleared bypressing SEL then ACT. The following descriptions explain the alarm conditionsthe system may encounter.The PPC has three types of alarms: audible, vibrate and visual alarms. The chap-ter will describe for each alarm condition, which screen message appears and thevibrate or audible alarm associated with it. Two audible or vibrate alarm types are used:Alarm Type 1:If the PPC is set to “vibrate”, the vibrator will be turned on for 3 seconds everyminute for 30 minutes. If the PPC is set to “audible”, the PPC will beep 6 times every minute for 30 min-utes. If the alarm is not cleared in 30 minutes, the PPC will beep 6 alternatingtones. The PPC will continue to do so every minute.Alarm Type 2:If the PPC is set to “vibrate”, the vibrator will be turned on for 3 seconds every30 minutes while the condition exists. If the PPC is set to “audible”, the PPC will beep 3 times every 30 minutes whilethe condition exists.Some alarms can be cleared by pressing SEL then ACT.
System alarms and messages90Pump alarmsAlarm feedbackThe Alarm Feedback function allows the user to verify the Pump and Pumpalarm are operating normally. Physicians can also use Alarm Feedback to mea-sure the time intervals between Pump strokes to verify accurate insulin delivery.When Alarm Feedback is programmed “YES”, the Pump will beep on each of thefirst five Pump strokes:• Following a change in the delivery regimen, for example when completinga meal Bolus and then changing to a Basal Rate, or at the start of a bolus.• After the Alarm Feedback function is programmed “YES.” AlarmFeedback will stay on until programmed back to “NO.”Pump low batteryThe Implantable Insulin Pump battery is designed to last approximately ten yearsduring conditions of normal use (see Chapter 11 for Technical Specifications).Battery life may vary somewhat depending upon a user’s insulin deliveryrequirements. When battery energy becomes low, a voltage sensor in the Pumpwill trigger the Pump Low Battery Alarm. A Pump Low Battery Alarm indicatesthere is approximately eight weeks of battery energy remaining. The Pump Low Battery alarm can be cleared and the Pump will continue to oper-ate normally. However, users should be instructed to report the alarm immedi-ately to their physician. A Pump replacement or resumption of conventionalinsulin therapy should then be scheduled. Depleted pump batteryWhen there is no longer sufficient battery energy to power the Pump, the PumpLow Battery Alarm will cease and insulin delivery will stop. Alternate insulintherapy must be initiated.PPC Display Type of Alarm Pump AlarmPUMP LOW BATTERY 1 In 24 hours if no PPC communication
System alarms and messages 91System errorThe Implantable Insulin Pump has a sophisticated self-monitoring system thatperiodically checks for circuit faults. If a fault should occur in the Pump electron-ics, insulin delivery will stop. Conventional insulin therapy must be initiatedimmediately. Notify Medtronic MiniMed immediately. Pump self test failIf during a “SELF TEST” the Pump presents a malfunction, the insulin deliverywill stop. Clear the message by pressing SEL then ACT. Notify Medtronic Min-iMed immediately. PPC low batteryIf the PPC main battery (AA 1.5 volt alkaline) energy is low, the following alarmdisplay appears each time a new function is programmed:Clear this message by pressing SEL and then ACT. While the battery shouldhave sufficient energy for a few additional programming commands, the batteryshould be changed as soon as possible. For instructions on changing the battery,(see Chapter 3, Personal Pump Communicator).NOTE: If while programming the PPC, the screen goes blank, thePPC beeps six times and then the “CHECK PUMPSTATUS” message appears, the battery needs to bereplaced. (See Chapter 3 for instructions, Install/Replacethe Main Battery).PPC Display Type of Alarm Pump AlarmPUMP STOPPED1 or 2 or 3 or 4 or 5 or 6 1 In 5 minutes if no PPC communicationPPC Display Type of AlarmPUMP SELF TEST FAIL 1PPC Display Type of AlarmPPCLOW BATTERY 1
System alarms and messages92PPC alarmsThe PPC offers a choice of two alarms, audible and vibrate. In addition, a screenmessage appears indicating the type of alarm condition that occurred. Low reservoirWhen the Pump calculates that less than 800 units (2ml) of insulin remains in itsreservoir, the following display will appear:Clear this message by pressing SEL and ACT. Users should be instructed toreport this alarm to their physician as soon as possible and schedule an appoint-ment for a Pump refill.Empty reservoirWhen the Pump calculates that less that 400 units (1ml) of insulin remain in itsreservoir, the following display will appear:Clear this message by pressing SEL and ACT, and then continue programming.Users should be instructed to report this alarm to their physician as soon as possi-ble and schedule an appointment for a Pump refill. It is important not to allow thePump to deplete its insulin supply as this may result in catheter blockage.Telemetry communication errorIf programming is interrupted after partial transmission of a command, the PPCwill display the following message on the display screen:Reposition the PPC near the Pump, and then press SEL and ACT. The PPC willattempt to resume communication with the Pump.PPC Display Type of Alarm Pump AlarmLOWRESERVOIR 1 In 24 hours if no PPC communicationPPC Display Type of Alarm Pump AlarmEMPTYRESERVOIR 1 In 24 hours if no PPC communicationPPC Display Type of AlarmTELEMETRY COMM ERROR3 1
System alarms and messages 93Initialize alarmAttempting to initialize a PPC to a Pump that is not compatible with it, results inone of the following messages. The physician should press SEL and ACT toclear the alarm, then verify the personal ID of the Pump is correct.Attempting to initialize a PPC to a Pump that contains invalid stroke volume orinsulin concentration information, results in one of the following messages to bedisplayed. Contact Medtronic MiniMed for instructions.Responding “NO” to the “INITIALIZE PPC TO PUMP" 8 consecutive times willresult in the following message being displayed. Step away from any other Pumpin the area and perform the request again.PPC not initializedIf the PPC is not initialized to a Pump, the following message is displayed:When successfully completing a PPC initialization, the following message is dis-played:PPC Display Type of AlarmPUMP ERROR0 or 1 1PPC Display Type of AlarmPUMP ERROR40 or 41 1PPC Display Type of AlarmTELEMETRY ERROR201PPC Display Type of AlarmPPCNOT INITIALIZED 1PPC Display Type of AlarmPUMP INITIALIZED 1
System alarms and messages94Battery replacementIf the PPC main battery (AA 1.5V alkaline) has been replaced or the PPC recog-nizes the “PUMP STATUS” needs to be checked, the following message is dis-played:The user should reposition the PPC near the Pump, and then press SEL and ACT.The PPC will communicate with the Pump.NOTE: If while programming the PPC, the screen goes blank, thePPC beeps six times and then the “CHECK PUMPSTATUS” message appears, the battery needs to bereplaced. (See Chapter 3 for instructions, Install/Replacethe Main Battery.)Initialize to factory defaultsWhen the Pump is reinitialized to its factory default settings, the following mes-sage is displayed:Pump stoppedIf the Pump is intentionally stopped, the following message is displayed:To restart the Pump, press SEL and ACT, and then place the PPC near the Pumpand allow the communication to complete. The Pump can only be stopped in theSupervisor Mode.PPC Display Type of AlarmCHECKPUMP STATUS 1PPC Display Type of AlarmPUMPRESET 1PPC Display Type of AlarmPUMPSTOPPED 2
System alarms and messages 95Pump suspendedIf the Pump operation has been suspended, the following message is displayed:During “SUSPEND PUMP”, the Pump will deliver a basal rate of 0.2 U/h. To restart insulin delivery programming, press SEL and ACT, then placethe PPC near the Pump and allow the communication to complete. Auto offIf the “AUTO OFF” time interval elapses in the Pump and the PPC recognizesthis condition, the following message is displayed:Press SEL and ACT, and then place the PPC near the Pump and allow the com-munication to complete. The PPC then communicates with the Pump to reset the“AUTO OFF” duration. If five more minutes elapse, the Pump will initiate the internal alarm sequence ofa beep every 15 seconds for 10 minutes, then double-beeps every 15 seconds for10 minutes, then repeating the pattern. The alarm is cleared by pressing SEL andACT.Hourly maximum exceededAttempting to deliver more than 2.5 times the pre-programmed bolus maximumin one hour causes the following message to be displayed:Press SEL and ACT, and then place the PPC near the Pump and allow the com-munication to complete. The patient may exceed the pre-programmed bolus limitby programming another bolus within 10 minutes.PPC Display Type of AlarmPUMPSUSPENDED 2PPC Display Type of AlarmAUTO OFF PUMPPUMP SUSPENDED 1PPC Display Type of AlarmHOURLY MAXEXCEEDED 1
System alarms and messages96Pump alarm tableAlarm Condition Type of Alarm Pump CommunicationsAUTO OFF 1 YESCHECK PUMPSTATUS 1YESEMPTYRESERVOIR 1YESHOURLY MAXEXCEEDED 1NOLOW RESERVOIR 1 YESPPC LOW BATTERY 1NOPPC NOT INITIALIZED 1YESPUMP INITIALIZED 1 NOPUMPLOW BATTERY 1YESPUMPRESET 1NOPUMP SELF TEST FAIL 1NOPUMP ERROR0 or 1 1NOPUMP ERROR40 or 41 1NOPUMP STOPPED1 or 2 or 3 or 4 or 5 or 6 1YESPUMP SUSPENDED 2 YESPUMP STOPPED 2 YESTELEMETRY COMM ERROR 1YESTELEMETRYCOMM ERROR201NO
97CHAPTER 10 Troubleshooting Pump System under-deliveryPotential under-delivery of insulin by the Medtronic MiniMed 2007D Implant-able Insulin Pump System may result in an increase in daily programmed insulinusage, difficulty maintaining euglycemia, occasional hyperglycemia, and prob-lems calculating refill accuracy. This chapter describes how to diagnose potentialPump System problems that may cause insulin under-delivery, and offers poten-tial Pump and Catheter solutions to correct for under-delivery.Diagnostic proceduresWhen refill procedures reveal the possibility of a Pump System under-deliveryproblem, diagnostic procedures must be performed to verify if there is a problemwith either the Pump or Catheter. The Stroke Volume Measurement Proceduretests Pump function, while the Pressure Measurement Procedure tests Catheterpatency. These diagnostic procedures should be performed according to the stepsoutlined in Appendix G.Under-delivery caused by backflowBackflow results in the inverted flow of insulin through the Pump System. Back-flow is caused by insulin deposits that compromise valve integrity, and allow thenegative reservoir pressure (vacuum) to pull insulin back into the reservoir. Tocompensate for this under-delivery, the user can program appropriate increases intheir basal rates and bolus amounts.Backflow conditions are characterized by increases in daily programmed insulinusage, difficulty in maintaining euglycemia, increasingly negative refill accuracyand sometimes hyperglycemia. Confirm a backflow condition by performing theStroke Volume Measurement Procedure. Then rinse the Pump System withNaOH solution to dissolve insulin deposits, following the Pump Rinse Procedureoutlined in Appendix E.
Troubleshooting Pump System under-delivery98Under-delivery caused by catheter occlusionUnder-delivery caused by Catheter occlusion can occur either abruptly or gradu-ally. The insulin usage and clinical symptoms are identical to those of Pumpunder-delivery. Confirm a Catheter occlusion condition by performing a PressureMeasurement Procedure. Then perform the following procedures to clear theocclusion:• First, flush the Catheter by using the Side Port Catheter Flush Procedureoutlined in Appendix F.• If the Flush Procedure is unsuccessful, replace the Catheter.Catheter replacement surgery should be performed in a manner similar to the ini-tial Pump System implantation. PPC initialization will not be necessary. How-ever, after the Catheter replacement, Pump function and delivery verificationsmust be performed.
99CHAPTER 11 Technical specificationsImplantable Insulin Pump MMT-2007DDiameterThicknessReservoir Volume8.1 cm (3.2 inches)2.0 cm (0.8 inches)13 ml to 15 mlWeight - Empty 131 gm (4.6 ounces)Insulin - Concentration U-400 Stroke Volume 0.42 to 0.58 ul per stroke0.17 to 0.23 units per strokeBasal Rate 0.2 to 35.0 units per hour (U/h)Basal Patterns 3 patterns of up to 48 basal rates each.Meal Bolus  0.2 to 25.0 unitsBolus Duration Immediate, Square Wave (30 minutes to 4 hours), or Dual Wave. Audio Bolus Temporary Basal Rate 0.2 to 35.0 units per hour30 minute increment duration30 minutes up to 24 hours delayDiagnostic Delivery Rate 10 to 150 U/hPower Supply Lithium - Carbon Monofluoride BatteryAudio Alarms Low BatterySystem ErrorSafety Features Negative Pressure Reservoir with Passive FillingPump Shutdown and Alarm with System Error (unique code sequences)Materials in contact with tissue Titanium, Silicone Rubber
Technical specifications100Personal Pump Communicator (PPC) model MMT-3160HeightLength Width8.9 cm (3.5 inches)7.0 cm (2.8 inches)2.0 cm (0.8 inches)Weight  115 gm (4.0 ounces)Main Power Source 1.5 Volt Alkaline Battery Type AAMain Battery Life 4 weeksOperating Temperature 0ºC to 40ºC (32ºF to 104ºF)Storage Temperature 0ºC to 30ºC (32ºF to 86ºF)Messages SuspendedCheck Pump StatusCommunication ErrorEmpty ReservoirHourly Maximum ExceededLow ReservoirPPC Not initializedPPC Low BatteryPump Self Test FailPump ResetPump StoppedPump Version ErrorTelemetry Communication Error
Technical specifications 101Side Port Catheter The Implantable Pump System complies with European RF Regulations. Opera-tion is subject to the following two conditions: (1) This device may not causeharmful interference, and (2) this device must accept any interference received,including interference that may cause undesired operation.Any changes or modifications to the system not expressly approved byMedtronic MiniMed could void the user’s authority to operate the system.FCC complianceThis device complies with Part 15 of the FCC Rules. Operation is subject to thefollowing two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interferencethat may cause undesired operation.MMT- 4027Length       Proximal:      Distal:11.8 ± 1.3 cm (4.7 ± 0.5 inches)17.8 ± 0.7 cm (7.0 ± 0.3 inches)MMT- 4024Length Proximal:Distal:11.8 ± 1.3 cm (4.7 ± 0.5 inches)10.2 ± 0.4 cm (4.0 ± 0.2 inchesMaterial Polyethylene-lined Silicone RubberSideport Polysulfone, Silicone Septum
Technical specifications102
103APPENDIX A Label information symbol dictionaryDo Not Reuse This DevicePlease Read “Important Information”Sterilized by Ethylene Oxide Manufacture Date (Year-Month) Lot Number Expiration Date (Use By Date)(Year-Month)REF MMT-XXXX Reference / Record Number (reorder number)SN Device Serial NumberIEC Icon indicating equipment providing a particular degree of protection against electric shock (Type B)CE MarkingCE MarkingRadio Frequency (RF) is not harmonized0976
104PackagingThe icon on each label indicates the contents of the package. The number withthe icon is the quantity. Descriptions of the icons are in the table below.Other informationIcon DescriptionMedtronic MiniMed 2007D Implantable Insulin PumpMedtronic MiniMed 3160 Personal Pump Communicator (PPC)Medtronic MiniMed Side Port Catheter - MMT 4024 and MMT 4027Medtronic MiniMed 4105 Refill KitMedtronic MiniMed 4102 Refill NeedlesManufactured by: Medtronic MiniMedDistributed in the United States by: Medtronic MiniMedDistributed in Europe by: Medtronic EOC, NetherlandsCE Marking:Notified Body: GMED - FranceNotified Body Authorizations: 0459, 0976Year of Initial CE Authorization: 2000
105APPENDIX B Implant worksheetImplant worksheet formPlease refer to “Implant Worksheet” attached.
Medtronic MiniMed Implant Worksheet FormPrep Date:_________ Center:_________ Patient Code:___Patient name:_________________________________Pump Label:Catheter Label:PPC Label:Insulin Lot Number:_________Communicator-settings:Patient Communicator ID:________ Supervisor Code:__Max meal bolus:____ U   Max basal rate:_______U/hMaximums (locked/unlocked):__ Alarm feedback "ON":__Basal rate programmed with PPC:________U/hWeight of full "IN 2" syringe before filling the reservoir:_____________________g (1)Weight of full "IN 2" syringe after filling the reservoir and removing 2ml from the pump:__________________ g (2)Total amount placed in the reservoir(1-2):_________________ gVerification of alarm feedback:  yes / noTime delivery started, priming bolus:_____________Calculated stroke volume:____________________a) number of strokes delivered:________________b) total volume delivered:_________________c) calculated stroke volume (b/a):__________ ulVerification of the delivery from the catheter tip:  yes / noImplant date:_____________ Surgeon name:________________ Anesthesia:____________________ Pocket depth:_________cmCatheter fixation:________________ Pump fixation:__________Catheter type:_______________    Pump type:_______________Pump orientation:Complications:
107APPENDIX C Refill formPump refill dataPlease refer to “Refill Data Form” attached.
Medtronic MiniMed Refill WorksheetBefore the refill: U/384.6=_________ g (F)*Insulin remaining from PPC:_____________________ (F)Data:Weight of primed "RB" syringe:_____________ g (A)Weight of "RB" syringe after insulin withdrawn:______________________     g (B)Weight of filled "IN" syringe:_____________      g (C)Weight of "IN" after pump filled and 2mlwithdrawn:______________________________ g (D)Amount withdrawn                    Refill Amount(B)_______________ g         (C)_________________ g                    (A)_______________ g        _ (D)________________ g= _______________  g (E)    = _______________ g (G)*Corrected for density of insulinPatient name:___________________ Patient code:________Center:_____________________ Physician name:____________Insulin "IN" _______  Rinse Buffer "RB"  ________Previous refill amount:____________________   g (1)Residual amount withdrawn (E):______________   g (2)Actual amount used (line 1 - line 2):____________  g (3)Theoretical amount used:__________________     g (4)(line 1 - remaining dose from PPC (F)Difference between actual and theoreticalamount used (line 4 - line 3):___________________ g (5)Refill accuracy (line 5 / line 4, then x 100):________ %Usable units of insulin (gx384.6): _______________ UAverage daily insulin use:____________________  U/dEstimated refill period:___________________ DaysSchedule next refill visit:_______________                                           D / M / Y
109APPENDIX D Precautions and general procedures Special note and precautions Before beginning the implant, refill, stroke volume measurement, rinse/flush, orpressure measurement procedure, carefully read this Appendix and keep theinformation in mind as you perform each procedure.1. All procedures should be performed using ASEPTIC TECHNIQUE.2. Air in the Pump System has been shown to be a significant contributingfactor to aggregation of insulin. Proper degassing of all solutions that enterthe Pump is essential. Read the insert in the refill kit packaging about thedegassing procedure.3. When using the refill kit, never release the syringe-locking ring while there isa vacuum within the refill syringe. The plunger of the syringe will snap backforcefully and may rupture the syringe and eject the contents.4. It is important that the syringe needle is perpendicular to the fill port toprevent binding when entering or exiting.5. The Side Port Catheter attachment site contains up to 20 units of INSULIN.The distal portion of the Catheter and the tubing can contain up to 13 units ofINSULIN. However, the INSULIN in the Pump fluid pathway and the sideport receptacle can be cleared by following the procedure to pull RINSEBUFFER through the side port. During a flush procedure to push out acatheter tip blockage, the 13 units in the distal catheter will be delivered tothe patient. The 13 units can be managed by programming continuous bolusamounts prior to the procedure, or with the use of oral or IV glucose asneeded during the procedure. Individual patient response to INSULIN bolusamounts must be considered.6. Patient blood glucose monitoring must be performed during, and up to onehour after the procedure. Oral and intravenous glucose should be availablefor glycemic management.
1107. After the flush procedure, at least 13 units of RINSE BUFFER will remain inthe distal portion of the catheter. Programmed bolus amounts to remove theRINSE BUFFER may be completed before the patient leaves the clinic. Aprescribed INSULIN basal rate should be programmed when the system isclear of RINSE BUFFER.8. Never push down on the plunger to fill the Pump. When the MedtronicMiniMed refill needle is properly seated in the fill port, the vacuum in thereservoir will draw the fluid from the syringe.9. Each step of the procedure will list the syringes and appropriate fluids neededfor that step. It is suggested that the syringes be prepared and labeled prior tothe start of the procedure.10. The Side Port Catheter is intended to be accessed only during the implantprocedure, the combined rinse/flush procedure, the catheter flush procedure,and for diagnostic procedures. The catheter is not intended to provide accessto the peritoneum for bolus injections of fluids or for withdrawal of bodyfluids.General procedures Before beginning any of the procedures, be sure you are familiar with the generaltechniques for locating the Pump fill port and the Side Port Catheter, accessingthe Pump inlet, venting the refill syringe, and preparing the syringes.
111Locating the pump fill port and the side portFigure 15: Template and placement1. Locate the Pump, then aseptically prep and drape the Pump area. Use topicalor local anesthetic if desired.2. To locate the Pump fill port, located in the center of the Pump, align thesterile template (MMT-4106) over the Pump by palpating the Pumpcircumference. Then insert an 18 gauge needle into the central Pump fill portas shown in Figure 15 above.3. To locate the side port, palpate around the circumference of the Pump. It ishelpful to grasp the side port between the thumb and index finger whenaligning the template.4. Using the template as a guide, insert the second 18 gauge needle into the sideport as shown. The side port can be accessed by inserting the guide needlejust inside the outer edge of the side port connection. This outer edge israised around the outer aspect of the side port, in order to help guide theneedle into the port.Template (MMT-4106)Placement on Skin with Guide Needles in Position
112Accessing the pump inlet(See Figure 16.) To enter the Pump fill port or the side port, make sure the refillsyringe stopcock is closed, then pass the Medtronic MiniMed refill needle(MMT- 4102) through the 18 gauge guide needle. Pull back slightly (approxi-mately 2 mm) on the guide needle to allow the Medtronic MiniMed refill needleto enter the Pump fill port or the side port. The refill needle entry opens the inletvalve of the Pump fill port.After passing through the septum and seating in the valve, you can feel theincreased force, about 0.5 pounds, required to move the valve 0.5 millimeter toopen. Maintain this downward pressure in the pump fill port to ensure that theinlet valve remains open. In the case of the side port, a downward pressure is notrequired after the needle is completely inserted.Figure 16: Inlet valve
113Venting the Medtronic MiniMed refill syringe (optional)In order to fill the Pump, it may be necessary to vent the headspace in theMedtronic MiniMed refill syringe. Refill syringes do not have lubricant on thesyringe housing or on the plunger tip. Some syringes may be “sticky” and requirethe venting procedure described below to assist the vacuum in the Pump to over-come this condition.To vent, hold the syringe with the refill needle pointing down. Vent the syringeheadspace by pulling back firmly on the plunger until the second sealing ring onthe rubber cap passes beyond the vent hole as shown in Figure 17. Figure 17: Venting the refill syringeNOTE: Be sure that the plunger is retracted far enoughthat the vent hole is exposed. The syringe mustbe held with the needle facing down and usedwithin 10 minutes once it is vented. If this timelimit is exceeded, repeat the degassingprocedure and re-vent.
114
115APPENDIX E Pump rinse procedure The purpose of this procedure is to dissolve insulin deposits within the Pump res-ervoir, pumping mechanism, and the Side Port Catheter “port.”Supplies and solutions Prior to performing this procedure, assemble the necessary supplies and solutionsas outlined below:Supplies• Steri-Strips® and markers • Local anesthesia (if necessary)• Sharps container• Safety glasses•PPC• Oral or IV glucose• Glucose monitoring equipment•Drapes• One (1) scale - 0.01 gram resolution• One (1) sterile bag for PPC (intestinal or cassette bags work well)• One (1) Side port locating template, MMT-4106• Seven (7) Refill kits, MMT-4105• Seven (7) Medtronic MiniMed MMT-4102 refill needles. (Extra needlesshould be available to use as needed)• Eight (8) sharp 18 gauge regular bevel needles
116Solutions• 50 ml. sterile 0.1M NaOH (0.4 grams NaOH per 100 ml of Sterile water forinjection) MMT-9005• Seven (7) 10 ml vials, rinse buffer solution, MMT-2008• Five (5) 10 ml vials, U-400 Insulin The 0.1M of NaOH used in this procedure can cause permanent eye damage. Safety glasses must be worn during this procedure.!WARNING
117Preparing for the procedure NOTE: Before beginning any refill, flush, stroke volumemeasurement, carefully read Appendix D, Precautionsand General Procedures, and keep this in mind as youperform each procedure.In order to prepare for the Side Port Catheter rinse and flush procedure, a total ofseven syringes will need to be labeled and then prepared with different solutions.Table 1 defines the syringe numbers and corresponding solutions:Prepare syringes for emptying the PumpSyringe 1 is used for emptying the Pump. Label and prepare syringe 1 perTable 1.Prepare syringes for filling the PumpDuring the Catheter rinse/flush procedure, syringes 2, 3, 4, 5, 6, and 7 are usedfor filling the Pump. Each of these syringes needs to be labeled with its solutionand syringe number, per Table 1, and then filled and degassed.Table 1: Syringe # Syringe type Solution 1Refill 5 ml RINSE BUFFER2Refill 20 ml NaOH3Refill 30 ml NaOH4Refill 20 ml RINSE BUFFER5Refill 30 ml RINSE BUFFER6Refill 20 ml INSULIN7Refill 30 ml INSULIN
118Program minimal basal ratePlace the PPC in a sterile bag and program the Pump to “SUSPEND” mode, thebasal rate will be 0.2 U/h.Remove insulin from the Pump and fill with NaOHINSULIN is removed from the Pump and then the Pump is filled with NaOH.The following volumes are used:• Syringe 1 - 5 ml RINSE BUFFER• Syringe 2 - 20 ml NaOH• Syringe 3 - 30 ml NaOHFollow the steps below to perform this procedure:Syringe 1: Rinse Buffer - Degassed1. Prime the needle. Weigh the syringe and record the weight on the refill form(Line A).2. Close the stopcock. Obtain a vacuum by pulling back on the plunger until itlocks. Press the lock into the plunger groove to be sure it is firmly secured.3. Enter the Pump with the refill needle.4. Open the stopcock; withdraw the INSULIN. After the INSULIN appears tohave stopped rising in the syringe, wait an additional 30 seconds to makesure all of the INSULIN and air are removed. NOTE: Air in the Pump System has been shown to be a significantcontributing factor to aggregation of INSULIN. Properdegassing of all solutions that enter the Pump isessential.5. When 30 seconds have elapsed, close the stopcock, remove the syringe.Weigh the syringe and record the weight on the refill form. Discard theINSULIN in the syringe except for a 5 ml barrier. This syringe will be used topull NaOH through the side port. Attach and prime a new refill needle.
119The 0.1M NaOH used in this procedure can cause permanent eye damage. Safety glasses must be worn during this procedure. Syringe 2: NaOH - Degassed1. Prime the needle. Close the stopcock2. Enter the Pump with the refill needle.3. Open the stopcock and allow the Pump to fill completely with NaOH. Whenthe fluid level stops moving, the Pump is filled.4. Close the stopcock. Remove the syringe.5. Prepare syringe 2 for aspiration. Remove the air from the syringe. Closestopcock and obtain a vacuum by pulling back on the plunger until it locks.Press the lock into the plunger groove to be sure it is firmly secured.6. Re-enter the Pump.7. Open the stopcock and remove the NaOH. After the fluid level stops moving,wait an additional 30 seconds for all the NaOH to be removed.8. When the time has elapsed, close the stopcock and remove the syringe.Discard syringe 2.Syringe 3: NaOH - Degassed1. Prime the needle. Close the stopcock.2. Enter the Pump fill port with the refill needle.3. Open the stopcock and allow the Pump to fill completely with NaOH. Whenthe Pump is filled, approximately 10 ml will remain in the syringe. Leave thisNaOH in the syringe, and leave this syringe in place for the next step.!WARNING
120Equilibrate and pull NaOH through systemThe reservoir pressure in the Pump is equilibrated with outside ambient pressureand NaOH is pulled through the pumping mechanism. This procedure uses thefollowing volumes in syringes 3 and 1:• Syringe 3 with residual NaOH from the previous section.• Syringe 1 with 5 ml insulin/buffer mixture from previous section.Syringe 3: NaOH - Residual1. Verify that the stopcock on syringe 3 is open.Syringe 1: Insulin - Residual1. Prime the needle. Close the stopcock.2. Obtain a vacuum by pulling back on the plunger until it locks. Press the lockinto the plunger groove to be sure it is firmly secured.3. Enter the side port with the stopcock closed.4. Open the stopcock.5. Program and deliver a 4 unit bolus to open the valve in the Pump mechanism.6. Observe syringe 3. When the level of the NaOH approaches 2 ml, close thestopcock to prevent air from entering the Pump. Observe syringe 1. At least 1ml of NaOH must pass into syringe 1 in order to thoroughly clean the Pumpmechanism. If less than 1 ml enters, repeat the bolus from Step 5, somepumps may require 3 successive boluses to obtain 1 ml.NOTE: In order to completely dissolve deposits in the Pumpmechanism, it is important to maintain a vacuum withsyringe 1 and deliver the entire bolus.7. After the bolus, close the stopcock on syringe 1. Remove both syringes. Setaside syringe 3 for use in removing the NaOH from the Pump fill port in thenext step. Discard syringe 1.
121Remove NaOH and fill with rinse bufferThe NaOH is removed from the Pump and the system is filled with RINSEBUFFER. This procedure uses the following volumes in syringes 3, 4, and 5:• Syringe 3 with residual of at least 5 ml NaOH from previous step• Syringe 4 with 20 ml RINSE BUFFER• Syringe 5 with 30 ml RINSE BUFFERSyringe 3: NaOH - Residual1. Prepare syringe for aspiration. 2. Close the stopcock and obtain a vacuum by pulling back on the plunger untilit locks. Press the lock into the plunger groove to be sure it is firmlysecured.Enter the Pump with the refill needle.3. Open the stopcock and withdraw all the NaOH. After the NaOH appears tohave stopped rising in the syringe, wait an additional 30 seconds to makesure all the NaOH and air is removed.4. When the time has elapsed, close the stopcock, remove the syringe, anddiscard.Syringe 4: Rinse Buffer - Degassed1. Prime the needle. Close the stopcock.2. Enter the Pump with the refill needle and syringe. 3. Open the stopcock and allow the Pump to fill completely with RINSEBUFFER. When the fluid level stops moving, the Pump is filled.4. Close the stopcock. Remove the syringe. 5. Prepare syringe 4 for aspiration. Obtain a vacuum by pulling back on theplunger until it locks. Press the lock into the plunger groove to be sure it isfirmly secured.6. Re-enter the Pump.7. Open the stopcock and remove the RINSE BUFFER. After the fluid levelstops moving, wait an additional 30 seconds for all the RINSE BUFFER andair to be removed.
1228. When the time has elapsed, close the stopcock and remove the syringe and setaside.Syringe 5: Rinse Buffer - Degassed1. Prime the needle. Close the stopcock.2. Enter the Pump fill port with the refill needle.3. Open the stopcock and allow the Pump to fill completely with RINSEBUFFER. When the fluid level stops moving, the Pump is filled.   When thepump is filled, approximately 10 ml of RINSE BUFFER will remain in thesyringe. Leave the RINSE BUFFER in the syringe and leave the syringe inplace for the next step.Equilibrate and pull rinse buffer through systemThe reservoir pressure in the Pump is equilibrated with outside ambient pressureand RINSE BUFFER is pulled through the fluid system. This procedure uses thefollowing volumes in syringes 4 and 5:• Syringe 4 with RINSE BUFFER. Expel all RINSE BUFFER except 5 ml.Change refill needle if desired.• Syringe 5 with residual RINSE BUFFER from previous step.Syringe 5: Rinse Buffer1. Verify that the stopcock on syringe 5 is open.Syringe 4: Rinse Buffer - Residual1. Prime the needle of syringe 4. Close the stopcock.2. Obtain a vacuum by pulling back on the plunger until it locks. Press the lockinto the plunger groove to be sure it is firmly secured.3. Enter the side port with the stopcock closed.4. Open the stopcock.5. To open the valve in the Pump mechanism, program and deliver a 4 unitbolus using the PPC.6. Observe syringe 4. When at least 1 ml of RINSE BUFFER has enteredsyringe 4, close both stopcocks and remove the syringes. If less than 1 mlenters, repeat the bolus from Step 6. Some Pumps may require 3 successive
123boluses to obtain 1 ml. Discard syringe 4 and set aside syringe 5 to be usedlater to remove RINSE BUFFER.Remove rinse buffer and fill with insulinThe RINSE BUFFER is removed from the Pump fluid system and the system isfilled with INSULIN. This procedure uses the following volumes in syringes 5,6, and 7:• Syringe 5 with residual RINSE BUFFER from the previous steps• Syringe 6 with 20 ml INSULIN• Syringe 7 with 30 ml INSULINSyringe 5: Rinse Buffer - Residual1. Prime the needle. Close the stopcock.2. Obtain a vacuum by pulling back on the plunger until it locks. Press the lockinto the plunger groove to be sure it is firmly secured.3. Enter the Pump with the refill needle and syringe.4. Open the stopcock and remove the RINSE BUFFER. After the fluid levelstops moving, wait an additional 30 seconds to make sure all the RINSEBUFFER and air are removed.5. When the time has elapsed, close the stopcock, remove the syringe, anddiscard. Syringe 6: Insulin - Degassed1. Prime the needle. Close the stopcock.2. Enter the Pump fill port with the refill needle and syringe.3. Open the stopcock and allow the Pump to fill completely with INSULIN.When the fluid level stops moving, the Pump is filled.4. Close the stopcock and remove the syringe. 5. Prepare syringe 6 for aspiration. Prime needle and close stopcock. Obtain avacuum by pulling back on the plunger until it locks. Press the lock into theplunger groove to be sure it is firmly secured.6. Re-enter the Pump.
1247. Open the stopcock, remove the INSULIN. After the fluid level stops moving,wait an additional 30 seconds for all the INSULIN and air to be removed.8. When the time has elapsed, close the stopcock and remove the syringe.Discard all but 5 ml of solution. Remove any air, prime needle and closestopcock. Set aside the syringe.Syringe 7: Insulin - Degassed1. Prime the needle on syringe 7. Weigh and record the combined weight ofsyringes 6 and 7 on the Refill Form.2. Close the stopcock. Enter the Pump fill port with the refill needle.3. Open the stopcock and allow the Pump to fill completely with INSULIN.When the fluid level stops moving, the pump is filled. When the Pump isfilled, approximately 10 ml of INSULIN will remain in the syringe. Leavethe syringe in place for the next step.
125Equilibrate and pull insulin through systemThe reservoir pressure in the Pump is equilibrated with outside ambient pressureand INSULIN is pulled through the fluid system. This procedure uses the follow-ing volumes in syringes 6 and 7:• Syringe 7 with residual INSULIN from previous step.• Syringe 6 with 5 ml solution. To perform this procedure with syringes 6 and7, follow the steps below:Syringe 7: Insulin - Residual1. Maintain syringe 7 with approximately 10 ml of residual INSULIN in thePump fill port. Verify that the stopcock on syringe 7 is open.Syringe 6: Insulin - Residual1. Prime the syringe 6 needle. Close the stopcock.2. Obtain a vacuum in syringe 6 by pulling back on the plunger until it locks.Press the lock into the plunger groove to be sure it is firmly secured.3. Enter the side port with syringe 6 - stopcock closed.4. Open the stopcock on syringe 6.5. Using the PPC, program and deliver a 4 unit bolus.6. Observe syringe 7. If the level of INSULIN approaches 2 ml, close thestopcock to prevent air from entering the Pump. Observe syringe 6. At least 1ml of INSULIN must pass into syringe 6 to assure removal of all the RINSEBUFFER from the side port. If less than 1 ml enters, repeat the bolus fromStep 6, some Pumps may require 3 successive boluses to obtain 1 ml.7. Close both stopcocks. Remove syringe 7 from the Pump. Set aside thesyringe to be weighed.8. With the stopcock closed, remove syringe 6 from the side port. Syringe 6 willhave a vacuum from the previous step. Enter the Pump fill port with syringe6 and open the stopcock to remove 2 ml. This will restore the negativepressure safety feature of the Pump.9. Close the stopcock, remove syringe 6 from the Pump, and set it aside to beweighed.
126Remove guide needles and record fill amountAt this point, the 18 gauge guide needles should be removed and the refillamount should be calculated and recorded. Follow the steps below:1. Remove the 18 gauge guide needles. Apply pressure to the needle insertionsites.2. Weigh syringes 6 and 7 and record on the refill form. Subtract the combinedweight from the weights obtained in step 14 of the section, “Remove RINSEBUFFER and Fill with Insulin.” The result is the new refill amount.3. Record the extracted and new refill amounts in the PPC.Program new basal rateUsing the patient’s PPC, cancel the SUSPEND mode, to return to the normalpatient’s basal rate and modify if needed.
127APPENDIX F Side Port Catheter flush procedure When delivery of insulin is impaired due to catheter tip obstruction, the Cathetermay be flushed using 5 - 10 ml of RINSE BUFFER, MMT-2008.Supplies and solutions Prior to performing this procedure, assemble the necessary supplies and solutionsas outlined below:Supplies• Steri-Strips® and markers • Local anesthesia• Sharps container•PPC• Oral or IV glucose• Glucose monitoring equipment•Drapes• One (1) scale - 0.01 gram resolution• One (1) sterile bag for PPC (intestinal or cassette bags work well)• One (1) Side port locating template, MMT-4106• Five (5) Refill kits, MMT-4105• Six (6) Medtronic MiniMed MMT-4102 refill needles. (Extra needlesshould be available to use as needed)• Eight (8) 18 gauge regular bevel needles• One (1) 10 or 20 ml Luer Lock® syringe• One (1) stopcock - 2 way
128Solutions• Seven (7) 10 ml vials, Rinse Buffer Solution, MMT-2008• Five (5) 10 ml vials, U-400 Insulin NOTE: One-handed flush with a 10-20 mL syringe only.Preparing for the procedure NOTE: Before beginning any refill, flush, stroke volumemeasurement, carefully read Appendix D, Precautionsand General Procedures, and keep this in mind as youperform each procedure.In order to prepare for the Side Port Catheter rinse and flush procedure, a total offive different refill syringes and one 10 or 20 ml syringe will need to be labeledand prepared with different solutions. Table 2 defines the syringe numbers andcorresponding solutions.Record patient’s blood glucoseRecord the patient’s blood glucose value at the start of the procedure. Monitorblood glucose every 30 minutes, or as needed.Table 2: Preparation and labeling of syringesSyringe # Syringe type Solution #1 Refill 5 ml RINSE BUFFER #2 Refill 20 ml RINSE BUFFER #3 Refill 30 ml RINSE BUFFER #4  10 ml  5-10 mlRINSE BUFFER #5 Refill 20 mlINSULIN #6 Refill 30 ml INSULIN
129Prepare syringes for emptying the PumpSyringe 1 is used for emptying the Pump. Prepare syringe 1 as described in the"Precautions and General Procedures" section in Appendix D of this manual.Label the syringe per table 2.After the syringe is prepared, weigh the syringe and enter the weight on the RefillForm.Prepare syringes for filling the PumpDuring the Catheter flush procedure, syringes 2, 3, 5, and 6 are used for fillingthe Pump. Each syringe should be labeled per Table 2, then filled and degassedfollowing the General Procedures section in Appendix D of this manual.Prepare syringe for flushing the Side Port CatheterDuring the Catheter flush procedure, syringe 4 is used for flushing the Side PortCatheter. This syringe needs to be labeled and then prepared by following thesteps below:1. Firmly attach a stopcock to the 10 or 20 ml Luer Lock connector of a 10 or 20ml syringe. Then attach an 18 gauge needle to the stopcock.NOTE: Use only a 10-20 ml Luer Lock syringe for thisprocedure. Smaller syringes may damage the Catheterand/or Pump, by allowing too much high pressure.2. Draw 5-10 mL of RINSE BUFFER into the syringe.3. Expel all air from the syringe.4. Close the stopcock and remove the 18 gauge needle from the syringe.5. Fill the hub of the Medtronic MiniMed refill needle with RINSE BUFFERfrom the syringe and attach it to the syringe.6. Prime the Medtronic MiniMed refill needle completely.7. Close the stopcock.8. Set the syringe aside.
130Flushing the Side Port Catheter Program minimal basal ratePlace the PPC in a sterile bag and program the pump to “SUSPEND” mode, thebasal rate will be 0.2 U/h.Remove insulin and fill with rinse bufferThe INSULIN is removed from the Pump and the system is filled with RINSEBUFFER. This procedure uses the following volumes in syringes 1, 2, and 3:• Syringe 1 with 5 ml RINSE BUFFER• Syringe 2 with 20 ml RINSE BUFFER• Syringe 3 with 30 ml RINSE BUFFERSyringe 1: Rinse Buffer1. Prime the needle and close the stopcock. Weigh the syringe.2. Obtain a vacuum by pulling back on the plunger until it locks. Press the lockinto the plunger groove and be sure it is firmly secured.3. Enter the Pump with the refill needle.4. Open the stopcock and withdraw the INSULIN. After the INSULIN appearsto have stopped rising in the syringe, wait an additional 30 seconds to makesure all of the INSULIN and air are removed. NOTE: Air in the Pump System has been shown to be asignificant agonist to aggregation of INSULIN. Properdegassing of all solutions that enter the Pump isessential.5. When the time has elapsed, close the stopcock, remove the syringe. Weighthe syringe and record the weight on the Refill Worksheet.Syringe 2: Rinse Buffer - Degassed1. Prime the needle and close the stopcock.2. Enter the Pump with the refill needle.
1313. Open the stopcock and allow the Pump to fill completely with RINSEBUFFER. When the fluid level stops moving, the Pump is filled.4. Close the stopcock and remove the syringe.5. Prepare syringe 2 for aspiration. Obtain a vacuum by pulling back on theplunger until it locks. Press the lock into the plunger groove and be sure it isfirmly secured.6. Re-enter the Pump.7. Open the stopcock and remove the RINSE BUFFER. After the fluid levelstops moving, wait an additional 30 seconds for all the RINSE BUFFER andair to be removed.8. When the time has elapsed, close the stopcock and remove the syringe.9. Prime the needle and close the stopcock.Syringe 3: Rinse Buffer - Degassed1. Enter the Pump fill port with the refill needle.2. Open the stopcock and allow the Pump to fill completely with RINSEBUFFER. When the fluid level stops moving, the Pump is filled. When thePump is filled, approximately 10 mL of RINSE BUFFER will remain in thesyringe. Leave the RINSE BUFFER in the syringe, and leave the syringe inplace for the next step.Equilibrate and pull rinse buffer through systemThe reservoir pressure in the Pump is equilibrated with ambient pressure andRINSE BUFFER is pulled through the fluid system. This procedure uses syringes2 and 3:• Syringe 2 with RINSE BUFFER. Expel all RINSE BUFFER except 5 mL.Change the refill needle if desired.• Syringe 3 with residual RINSE BUFFER from the previous section.Syringe 3: Rinse Buffer - Residual1. Maintain syringe 3 with approximately 10 mL of RINSE BUFFER in thePump fill port. Verify that the stopcock on syringe 3 is open.Syringe 2: Rinse Buffer - Residual
1321. Prime the needle. Close the stopcock.2. Obtain a vacuum by pulling back on the plunger until it locks. Press the lockinto the plunger groove to be sure it is firmly secured.3. Enter the side port with the stopcock closed.4. Open the stopcock.5. Program and deliver a 4 unit bolus to open the valve in the Pump mechanism.6. Observe syringe 3. When the level of the RINSE BUFFER approaches 2 mL,close the stopcock to prevent air from entering the Pump. Observe syringe 2.When at least 1 mL of RINSE BUFFER has entered syringe 2 close bothstopcocks and remove the syringes. If less than 1 ml enters, repeat the bolusfrom Step 6. Some Pumps may require 3 successive boluses to obtain 1 ml.Discard syringe 2 and set aside syringe 3 to be used later to remove theRINSE BUFFER.Flush side port catheterThe Side Port Catheter is flushed using syringe 4 which is completely filled withRINSE BUFFER. Approximately 13 units of INSULIN remain in the distal Side Port Catheter. This INSULIN will be delivered to the patient rapidly in the next three steps. Alternatively, the INSULIN may be removed prior to flushing by programming continuous bolus amounts. Closely monitor blood glucose dur-ing INSULIN delivery, and administer intravenous glucose or glucogen as needed.Syringe 4: Rinse Buffer1. Prime the needle. Close the stopcock.2. Enter the side port with the needle.3. When the needle is firmly positioned in the side port, open the stopcock andquickly push the plunger all the way down. This should take no longer thanone to two minutes.4. After flushing, close the stopcock, remove and discard the syringe.!WARNING
133NOTE: This is the only time a plunger should be manuallypushed down. Remove rinse buffer and fill with insulinThe RINSE BUFFER is removed from the Pump fluid system and the system isfilled with insulin. This procedure uses the following volumes in syringes 3, 5,and 6:• Syringe 3 with residual RINSE BUFFER from the previous step• Syringe 5 with 20 ml INSULIN• Syringe 6 with 30 ml INSULINSyringe 3: Rinse Buffer - Residual1. Prime the needle. Close the stopcock.2. Obtain a vacuum by pulling back on the plunger until it locks. Press the lockinto the plunger groove and be sure it is firmly secured.3. Enter the Pump with the refill needle and syringe.4. Open the stopcock and remove the rinse buffer. After the fluid level stopsmoving, wait an additional 30 seconds to make sure all of the RINSEBUFFER and air is removed.5. When the time has elapsed, close the stopcock, remove the syringe, anddiscard it.Syringe 5: Insulin - Degassed1. Prime the needle and close the stopcock.2. Enter the Pump fill port with the refill needle.3. Open the stopcock and allow the Pump to fill completely with INSULIN.When the fluid level stops moving, the Pump is filled.4. Close the stopcock and remove the syringe.5. Prepare syringe 6 for aspiration. Obtain a vacuum by pulling back on theplunger until it locks. Press the lock into the plunger groove to be sure it isfirmly secured.6. Re-enter the Pump.
1347. Open the stopcock and remove the INSULIN. After the fluid level stopsmoving, wait an additional 30 seconds for all INSULIN and air to beremoved.8. When the time has elapsed, close the stopcock and remove the syringe.Discard all but 5 mL of the solution. Remove the air and set aside thesyringe.Syringe 6: Insulin - Degassed1.  Weigh and record the combined weight of syringes 5 and 6 on the RefillWorksheet. Prime the needles.2.  Close the stopcocks. Enter the Pump fill port with the refill needle.3.  Open the stopcock and allow the Pump to fill completely with INSULIN.When the fluid level stops moving, the Pump is filled. Approximately 10 mLof INSULIN will remain in the syringe. Leave the syringe in place for thenext step.Equilibrate and pull insulin through systemThe reservoir pressure in the Pump is equilibrated with outside ambient pressureand INSULIN is pulled through the fluid system. This procedure uses the follow-ing volumes in syringes 5 and 6:• Syringe 6 with residual INSULIN from the previous step• Syringe 5 with 5 mL solution of Rinse Buffer.Syringe 6: Insulin - Residual1. Maintain syringe 6 with approximately 10 mL of residual insulin in the Pumpfill port. Verify that the stopcock on syringe 6 is open.Syringe 5: Rinse Buffer -Residual1. Prime the syringe 5 needle and close the stopcock.2. Obtain a vacuum in syringe 5 by pulling back on the plunger until it locks.Press the lock into the plunger groove and be sure it is firmly secured.3. Enter the side port with syringe 5 - stopcock closed.4. Open the stopcock on syringe 5.
1355. Press  SEL. When the PPC screen is flashing “PUMP SUSPEND” pressACT. The system is now in normal mode.6. Using the PPC, program and deliver a 4 unit bolus.7. Observe syringe 6. If the level of INSULIN approaches 2 mL, close thestopcock to prevent air from entering the Pump. Observe syringe 5. At least 1mL of INSULIN must pass into syringe 5 to be sure that all of the RINSEBUFFER is removed from the side port. If less than 1 ml enters, repeat thebolus from Step 7, some Pumps may require 3 successive boluses to obtain 1ml.8. Close both stopcocks. Remove syringe 6 from the Pump. Set the syringeaside to be weighed.9. With the stopcock closed, remove syringe 5 from the side port. It will have avacuum inside from the previous step. Enter the Pump fill port with syringe 5and open the stopcock and remove 2 ml. This will restore the negativepressure safety feature of the Pump.10. Close the stopcock, remove syringe 5 from the Pump, and set it aside to beweighed.11. Press  SEL until the “SUSPEND PUMP” screen is displayed. Then PressACT.Remove guide needles and record refill amountAt this point, the 18 gauge guide needles should be removed, and the refillamount should be calculated and recorded.1. Remove the 18 gauge guide needles and discard them. Apply pressure to theinsertion sites.2. Weigh syringes 5 and 6 and record the combined weight on the Refill Form.Subtract the combined weight from the weight obtained in step 14. RemoveRinse Buffer and Fill with Insulin. The result is the new refill amount.3. Record the new and extracted refill amount in the PPC refill worksheet.Program new basal rateUsing the patient’s PPC, cancel the “SUSPEND” mode and allow the Pump toreturn to the patient’s original basal rate.
136Remove rinse buffer from catheterApproximately 13 units of RINSE BUFFER remain in the distal Side Port Cathe-ter. Depending on blood glucose values, program the appropriate bolus amount toremove the RINSE BUFFER from the Catheter. Release the patient when bloodglucose levels are stable.
137APPENDIX G Stroke volume measurement If under-delivery of insulin is noted during a refill procedure or suspected due toblood glucose control, it may be useful to verify the proper stroke volume of thePump by accessing the side port.In this procedure, a pipette is attached to a stopcock and refill needle. This systemis inserted into the side port. The insulin pulses in the pipette are measured to cal-culate the pump stroke volume.Supplies and solutions Prior to performing this procedure, assemble the necessary supplies and solutionsas outlined below:Supplies • Local anesthesia• Sharps container•PPC• Oral or IV glucose• Glucose monitoring equipment•Drapes• 5 ml or 3 ml syringe• One (1) sterile bag for PPC (intestinal or cassette bags work well)• One (1) Side port locating template, MMT-4106• One (1) sterile 100 microliter pipette, MMT-4104• Two (2) 18 gauge needles• One (1) MiniMed MMT-4102 refill needle. (Extra needles should beavailable to use as needed)
138• One (1) sterile stopcock - 3 waySolutions • One (1) 10 mL vial, Rinse Buffer Solution, MMT-2008Preparing for the procedure NOTE: Before beginning any refill, flush, stroke volumemeasurement, carefully read Appendix D, Precautionsand General Procedures, and keep this in mind as youperform each procedure.Record patient’s blood glucoseRecord the patient’s blood glucose value at the start of the procedure. Monitorblood glucose every 30 minutes, or as needed.Measuring stroke volume 1. In a sterile field before the procedure; prepare the apparatus. Prime thestopcock, pipette, and needle.2. Turn the stopcock as shown in Figure 18. Press on the syringe to obtain aliquid meniscus in the pipette.Figure 18: Stroke volume measurement setup
1393. Insert the needle into the side port.4. Turn the stopcock as shown in Figure 19. 5. Program a high rate bolus (approximately 20 units) and measure the strokevolume over 20 pulses in the pipette.6. Cancel the bolus at the end of the measurement.7. Turn the stopcock back to the position shown in Figure 19 to close. Removethe apparatus.Figure 19: Stroke volume measurement
140Record patient’s blood glucoseIt is likely that some insulin is pumped out of the Catheter and into the patient inthis procedure. Monitor the patient’s blood glucose and release the patient onlyafter blood glucose values are stable.