Radiometer analytical TTL090 Titration Workstation User Manual Usersguide gb

Radiometer analytical S.A.S. Titration Workstation Usersguide gb

Contents

Manual TIM9xx

Download: Radiometer analytical TTL090 Titration Workstation User Manual Usersguide gb
Mirror Download [FCC.gov]Radiometer analytical TTL090 Titration Workstation User Manual Usersguide gb
Document ID698108
Application IDlrvhwxYTpUB2axlF8OG2Eg==
Document DescriptionManual TIM9xx
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize59.07kB (738422 bits)
Date Submitted2006-08-28 00:00:00
Date Available2006-08-28 00:00:00
Creation Date2006-08-02 12:48:58
Producing SoftwareAcrobat Distiller 5.0.5 (Windows)
Document Lastmod2006-08-02 14:50:04
Document TitleUsersguide_gb
Document CreatorAdobe PageMaker 7.0
Document Author: HWimmer

TIM960, TIM965 and TIM980
Titration Workstations
User’s Guide
D21T078
Contents
D21T078 • Printed by Radiometer Analytical SAS • 2006-07B
Contents
Contents
General information ............................................................... 5
1. Introduction ........................................................................ 7
Read me! ...........................................................................................
Working in Supervisor mode ..............................................................
Working in Routine mode ...................................................................
Reagent handling using RFID technology ...........................................
10
11
2. Getting started .................................................................... 15
Keyboard ............................................................................................
Basic principles .................................................................................
Main window ......................................................................................
Reagents window ...............................................................................
Electrodes window .............................................................................
Cell window ........................................................................................
System configuration .........................................................................
15
16
18
19
21
23
25
3. Programming guidelines ................................................... 33
Create an electrode ............................................................................
Edit electrode screen .........................................................................
Programming electrode parameters ....................................................
Create a reagent ................................................................................
Edit reagent screen ............................................................................
Programming reagent parameters ......................................................
Create a method ................................................................................
Edit method screen ............................................................................
Programming method parameters ......................................................
Contents
34
37
39
40
41
43
44
45
47
4. Running analyses ............................................................... 49
Run electrode calibration ....................................................................
Run reagent calibration .......................................................................
Run method .......................................................................................
Run direct measurement ....................................................................
Run manual dosing ............................................................................
50
50
51
51
52
5. Viewing data ........................................................................ 53
How to access? .................................................................................
Sample results ...................................................................................
Reagent calibration results .................................................................
Electrode calibration results ...............................................................
Global variables .................................................................................
53
54
55
56
57
6. Printing data ....................................................................... 59
Manual printouts ................................................................................ 60
Automatic printouts ............................................................................ 62
7. Connection of peripherals ................................................. 63
8. Maintenance ........................................................................ 65
Cleaning ............................................................................................. 65
Transporting the instrument ................................................................ 65
Servicing ............................................................................................ 65
International Standards ......................................................... 66
Contents
General infor
mation
information
Safety Information
Please read this entire manual before unpacking, setting up, or operating
this equipment. Pay attention to all danger and caution statements.
Failure to do so could result in serious injury to the operator or damage
to the equipment. To ensure that the protection provided by this
equipment is not impaired, do not use and do not install this equipment in
any manner other than that specified in this manual.
Precautionary Labels
Read all labels and tags attached to the instrument. Personal injury or
damage to the instrument could occur if not observed.
This symbol, if noted on the instrument, references the
instruction manual for operation and/or safety information.
Electrical and electronical equipment marked with this
symbol may not be disposed of in European public
disposal systems after 13 August of 2005. In conformity
with European local and national regulations (EU
Directive 2002/96/EC), European electrical equipment
users must now return old or end-of life equipment to the
Producer for disposal at no charge to the user.
Note: For equipment supplied or produced by
"Radiometer Analytical", please contact www.hachlange.com and select your country for instructions on how
to return your equipment for proper disposal."
This symbol, when noted on the product, identifies the
location of a fuse or current limiting device.
Introduction
Warning !
The TitraLab system has been developed to meet the requirements of
volumetric titration applications. It is therefore aimed at experienced
users who have the knowledge required to operate the instrument and
implement the security instructions enclosed. Please remember that the
TitraLab system must not, under any circumstances, be used to perform
tests on living beings.
We accept no responsibility for using the TitraLab system and its
peripheral devices under conditions that are not specified in this
Reference Manual and its associated User’s Guide (part no. D21T078).
Compliance with FCC rules, part 15
Information to the user
NOTE: This equipment has been tested and found to comply with the
limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instruction, may cause harmful
interference to radio communications. However, there is no guarantee
that interference will not occur in a particular installation. If this
equipment does cause harmful interference to radio or television
reception which can be determined by turning the equipment off and on,
the user is encouraged to try to correct interference by one or more of
the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on circuit different from that to
which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
Warning!!
Changes and modifications that are not expressly approved by the party
responsible for compliance could void the user's authority to operate the
equipment.
Chapter 1
1. Intr
oduction
Introduction
Read me!
The interface of the Titration Manager has been specially designed to
clearly guide you through every step of the programming and running of
the analyses, whether you are a supervisor or a routine user.
An important part of this interface is to check and control the presence of
different elements necessary to run the defined application.
Thanks to a RFID tag attached to each burette stand, the Titration
Manager is able to know at any moment which reagent is installed on
which burette stands. Handling burettes and reagents is easier and more
secure. See "Titrant management using RFID tags" at the end of this
chapter.
Working in Supervisor mode
The Supervisor has access to all parameter edition screens for creation
and edition purposes.
When programming in “SUPERVISOR” mode, it is recommended to work
in stages. These stages should be carried out in the order described
below:
1. Define your electrode(s)
Identify electrodes (including temperature sensors) to be used for the
analysis.
Electrodes can be created from the following lists, Catalogue, Other or
Copy from. When creating the electrode, define if electrode calibration is
required (or not), if yes specify the “periodicity” of the calibrations and the
pH standards to be used. Refer to chapter 3.
2. Define reagent
Identify reagents to be used for the analysis.
Reagents can be created from the following lists, Catalogue, Other, Copy
from. When creating the reagent, define if reagent calibration is required
(or titre entered manually), if yes specify the “periodicity” of the calibrations and the calibration method. Refer to chapter 3.
If you are to perform a calibration, make sure that the electrode(s)
used for the calibration are the same as those used in the method.
Introduction
3. Create a new method or Edit a pre-programmed one
Create the measurement or titration method to be used for the analyses.
Enter the parameters required to calculate the results, refer to chapter 3.
When you have finished programming, select the method or preprogrammed application, refer to Short-Form Reminder no. 3.
If a sample changer is to be used, define the sample changer in the
Configuration menu before selecting Working mode = SAC
sequence in the main window.
4. Check icons
The following icons indicate the exact state of your working system.
Sunny icon:
Everything is OK. Run the method or sequence.
Sunny icons are required to run the method.
Cloudy icon:
Action required within 12 or 24 hours. For example calibration and/or reagent bottle exchange.
Stormy icon:
Electrode/reagent calibration date elapsed or electrode(s)/
reagents(s) not installed.
Question mark:
It is a programming error, reagent and/or electrode is/are
not defined in the selected method. Revise the method
programming.
Calculator icon:
Reprocessing mode (Working mode = Reprocessing) is set
on the instrument.
Refer to Reagent and Electrode windows.
Chapter 1
When Cloudy/Stormy/Question mark icons appear, press 1 to run
the “Check” command.
Depending on the icons displayed, the Titration Manager will
automatically guide you through the steps necessary to run the
analysis.
a. Connect/install the electrode(s)
Connect/install electrodes and temperature sensors, Refer to
Short Form Reminder no. 1.
b. Install reagents(s)
If the burette is not installed, the Titration Manager will
automatically guide you to the Install burette procedure.
Refer to Short Form Reminder no. 2 and "Reagent
management using RFID tags" at the end of this chapter.
c. Calibrate electrode(s)
Refer to chapter 4 and Short Form Reminder no. 4.
d. Calibrate reagent(s) or Enter titre
Run the reagent calibration (refer to chapter 4 and Short Form
Reminder no. 5) or enter the titre manually.
e. Run the method or the sequence, when Sunny icons are
displayed,
Refer to Short Form Reminder no. 6 and 7.
Introduction
Working in Routine mode
In “ROUTINE” mode you are guided at every step, thanks to the cleartext messages and the icons present on the large graphic display.
A Routine operator has access to all the displays for checking purposes.
Running methods
When working in “ROUTINE” mode, it is necessary to install your
titration system according to the selected method or sequence, prior to
running a method or sequence.
1. Select the method or sequence
Refer to chapter 4.
2. Check icons and run the method or sequence
Refer to page 8.
10
Chapter 1
Reagent handling using RFID technology
What’s RFID means and how does it work?
Radio Frequency IDentification (RFID) is a generic term that is used to
describe a system that transmits the identity (in the form of a unique serial
number) of an object or person wirelessly, using radio waves. It's grouped
under the broad category of automatic identification technologies.
It’s the combination of a “tag” and a “reader”. A typical RFID tag consists
of a microchip attached to a radio antenna mounted on a substrate. The
chip can store data. The reader is used to retrieve the data stored on an
RFID tag. A typical reader is a device that has one or more antennas that
emit radio waves and receive signals back from the tag. The reader then
passes the information in digital form to a computer system.
In our assembly an RFID tag is attached inside the sleeve of the burette
stand. This tag is clearly visible.
RFID tag placed on the burette stand
The RFID reader is mounted inside the burette tap block at the back of
the titrator and located with a small RFID graphic icon.When the burette
is installed, the tag and the reader are face to face.
RFID reader mounted inside the burette tap block
Introduction
11
Which data contains a burette stand RFID tag?
All identification or location information concerning the burette stand and
its associated reagent (if a reagent is installed on that stand). The tag also
supports the reagent GLP data.
For the burette stand:
burette ID and burette volume: this information is present in the tag
as a factory setting and cannot be erased and changed,
burette first installation date.
For the reagent:
reagent identification parameters entered by the user while creating
the reagent: Reagent ID, target titer, expiry date, batch number and
calibration method parameters (if a calibration is requested),
reagent last calibration date and results.
The Burette ID, Burette volume, Reagent ID and Reagent target titer are
read on starting the following user actions:
burette installation, replacement and removal,
reagent installation, replacement (reagent of 2 different IDs), bottle
exchange and reagent removal.
At the end of these actions, new data is saved in the RFID tag of the
burette stand.
12
Chapter 1
Benefits of using RFID technology on burette stands
The Titration Manager knows at any moment which reagent is installed
on which burette stands.
Handling burettes and reagents is easier.
For example, no need to enter data on burette installation. The
Titration Manager reads automatically the burette and install
reagent identification data (burette ID and volume),
RFID tag prevents the operator from using wrong reagents.
On installing a reagent, the Titration Manager detects automatically
if a reagent is already installed or not. If a reagent with a different ID
is already installed, a reagent replacement is initiated. If a reagent
with a same ID is already installed, no installation is performed and
a reagent check data screen is displayed instead.
On removing a burette, the Titration Manager detects automatically
if a reagent is already installed or not. If a reagent is installed, the
Titration Manager removes the reagent before removing the
burette.
RFID tag supports the complete reagent GLP data.
By this way, you ever use updated reagents data and traceability of
result is ensured. There is no possible permutation of identification
data from one reagent to another.
Introduction
13
14
Chapter 1
2. Getting star
ted
started
Keyboard
1.
2.
3.
4.
5.
6.
7.
Alphanumeric keypad to enter data and parameters on the same
principle as mobile telephones. These keys can also be used for
quick access to the different menus; refer to the display on the
following page.
Del: deletes the character on which the cursor is positioned.
Operator may end the analysis before the max. titration time or the
max. volume have been reached, if he considers his analysis
finished. Calculations are performed.
Esc: returns to the previous screen. During analysis, this key allows
you to enter the method, electrode and titrant menu.
Print: prints the data concerning the screen displayed.
Stop: stops an analysis or a burette function. Press this key for 3
seconds in the Main window to gain access to the setup
parameters.
RIGHT, LEFT, UP, DOWN arrow keys are used to move to different
options within the menus.
✓: confirms a data entry, a message or a function asked for by the
user.
Getting started
15
Basic principles
When the instrument is switched on, the Main window is displayed.
When the instrument is switched on for the first time the screen will be
as follows:
Title bar
Selected line
Menu options.
Press the
corresponding number
for quick access to the
different menus. For
example, press 4 to
enter the Method library
Method tab
Reagent tab
Electrode tab
Cell tab
If required, you can adjust the contrast of the display by:
•
pressing 0 to increase the brightness,
•
pressing 7 to decrease the brightness.
The title bar in the menu indicates the instrument name (TIM960 for
example) and the actual time. You will be shown how to personalise the
name and adjust the time further on in the manual.
The RIGHT and LEFT arrow keys allow you to move from one tab to the
other and enter the Reagent, Electrode and Cell menus.
•
Work your way through the 4 tabs, then back to the Main window.
The UP and DOWN keys allow you to select a line. To enter an option,
select the line, and press ✓. You can also press the corresponding
numerical key.
16
Chapter 2
1.
Press 4 or select the line Method library and press ✓ to enter the
Method library screen. Press Esc to return to the Main window.
2.
Select the line Working mode and press ✓. The following
options are available:
Method: to run a single method.
Sequence: to create or run a sequence of methods. Beakers are
manually changed between two method runs.
SAC Method: to run a single method to be performed using a
sample changer.
SAC Sequence: to create or run a sequence of methods to be
performed using a sample changer.
Reprocessing: to reprocess the last curve with different parameters
from the original method and thus recalculate a result.
If you select a Reprocessing working mode, the Method, Reagent
and Electrode tabs display a calculator icon as shown below:
Note: the working mode selected will have no effect on the type of
method you wish to create.
Do not forget to define the sample changer in the Configuration
menu before selecting SAC Method or SAC Sequence.
Press ✓ to confirm your selection.
Getting started
17
Main window
When the instrument is switched on the Main window is displayed.
1.
Title bar: indicates the instruments name and the current time.
2.
Working mode: Select the way in which you want to work. Choose
from Method, SAC Method, Sequence, SAC Sequence,
Reprocessing; refer to previous page.
3.
Check or Run: check or run the selected method /sequence. The
method can be run when 2 sunny icons are displayed in the
Reagents and Electrodes tabs. If 2 cloudy/stormy icons are
displayed, activate the "Check" command. The Titration
Manager will automatically guide you through the necessary
operations required to solve the problem(s).
4.
Sequence/sample stack: if SAC Sequence or Sac Method
has been selected in step 2 above. Program the sample stack.
Select or edit the method sequence.
5.
Select method: select method to be used for the analyses.
6.
Method library: supervisor use only: create, edit, reset and delete
methods to correspond to your specific needs.
Or Display methods - routine use only : display the main
parameters of the selected method.
7.
GLP-Archives: access GLP tables and visualise the stored method
sample results and global variables.
8.
Method tab: Animated icon indicates when a method/sequence is
running.
18
Chapter 2
Reagents window
Use the RIGHT arrow key to move to this window.
1.
Title bar: indicates the name of the window.
2.
Calibrate/Enter titre: determine the concentration of the titrant
(titre) by running a calibration or a calibration sequence or by
entering the titre manually.
3.
Install reagents: install or replace reagents in a method or sequence.
4.
Check reagents: verify the identification parameters of the
reagents used in the working method or sequence.
5.
Reagent library- supervisor use only: create, edit, reset and delete
reagents to correspond to your specific needs.
6.
Manual dosing: delivers a reagent at a given speed and measures
a signal at a connected electrode.
7.
GLP-Archives: access GLP tables and visualise the stored reagent
calibration results.
8.
Burette functions: fill, empty, flush, rinse and burette service
position. Global flush of all installed burettes simultaneously.
9.
Reagent tab: Reagent status icon indicates the state of the reagent
system. Five types of icons can be displayed.
Getting started
19
Sunny icon. The reagent calibration or manual entry of the
titre has been performed for all the reagents present in the
system. Everything is just right!
Cloudy icon. The reagent calibration of one of the reagents in
the system should be performed within 12 or 24 hours.
The expiry date of one of the reagents in the system will
elapse in less than one week.
Stormy icon. The reagent calibration or the expiry date of
one of the reagents in the system has elapsed.
If acceptance limits have been set for the calibration: at
least one calibration result lies outside the programmed
acceptance limits.
At least one of the reagents present in the system has not
been installed.
Check the sequence or method, (press 1 in the Main
window). The instrument prompts you to do the necessary
operations. You are guided step by step.
Question mark. There is a problem in the editing of the
reagent system. You need to be in Supervisor mode to solve
the problem. Check the sequence/method
parameters of the reagent or the electrode.
Check the sequence or method, (press 1 in the Main
window). The instrument indicates the possible errors and
prompts you to correct them until ? disappears.
Calculator icon. Reprocessing mode (Working mode =
Reprocessing) is set on the instrument. This icon is
displayed irrespective of the status of the reagent system
used.
20
Chapter 2
Electrodes window
Use the RIGHT arrow key to move to this window.
1.
Title bar: indicates the name of the window.
2.
Calibrate electrode: run a calibration or a calibration sequence
using the installed electrodes.
3.
Install electrodes: connect, disconnect or replace electrodes in a
method or sequence.
4.
Check electrodes: check the identification parameters of the
electrodes used in the method or sequence.
5.
Electrode library - Supervisor use only: create, edit, reset and
delete electrodes stored in the instrument.
6.
Display measurement: displays mV and/or pH and/or temperature
at a connected electrode of the electrode system.
7.
GLP-Archives: access GLP tables and visualise the stored
electrode calibration results.
8.
Electrode tab: Electrode status icon indicates the state of the
electrode system. Five types of icons can be displayed.
Getting started
21
Sunny icon. The calibration has been performed on all the
electrodes present in the system. Everything is just right!
Cloudy icon. The electrode calibration of one of the
electrodes present in the system should be performed within
12 or 24 hours.
Stormy icon. The calibration date has elapsed for one of the
electrodes present in the system.
If acceptance limits have been set for the calibration: at
least one calibration result lies outside the programmed
acceptance limits.
At least one of the electrodes present in the system has not
been installed.
Check the sequence or method, (press 1 in the Main
window). The instrument prompts you to do the necessary
operations. You are guided step by step.
Question mark. There is a problem in the editing of the
electrode system. You need to be in Supervisor mode to
solve the problem. Check the sequence/method
parameters of the reagent or the electrode.
Check the sequence or method, (press 1 in the Main
window). The instrument indicates the possible errors and
prompts you to correct them until ? disappears.
Calculator icon. Reprocessing mode (Working mode =
Reprocessing) is set on the instrument. This icon is
displayed irrespective of the status of the electrode system
used.
22
Chapter 2
Cell window
Use the RIGHT arrow key to move to this window.
1.
Title bar: indicates the name of the window.
2.
Instrument: to select the stirrer of the Titration Manager or an
ABU62 connected to the Titration Manager.
3.
Internal stirring: command stirrer On/Off.
4.
Speed: select the internal stirring speed, from 100 to 1100 rpm by
steps of 50 rpm.
5.
Fill cell (TIM980 only): to fill the Karl-Fischer cell with solvent.
6.
Empty cell (TIM980 only): to empty the Karl-Fischer cell.
7.
Cell tab: animated icon indicates when the magnetic stirrer or
propeller is operating.
Getting started
23
Apply internal stirring
1.
For Instrument, select the stirrer of the Titration Manager or the
stirrer of an ABU62 connected to the Titration Manager.
2.
Select Internal stirring = ON,
3.
Press ✓ in the field Speed and select a stirring speed.
Apply external stirring
1.
Connect the Stirring Propeller, part no. 847-731, to the Titration
Manager or ABU62 Propeller socket. Line 3 is automatically
replaced by External stirring.
2.
For Instrument, select the stirrer of the Titration Manager or the
stirrer of an ABU62 connected to the Titration Manager.
3.
Select External stirring = ON.
4.
Adjust stirring by turning the stirring propeller knob. You can consult
the corresponding table between the position (1 to 9) and the stirring
speed by pressing ✓ in the field Speed setting.
Fill the Karl-Fischer cell (TIM980 only)
1.
For Instrument, select TIM.
2.
Press 1 (Fill cell).
3.
Press the TIM980 filling cell button
then press ✓.
Empty the Karl-Fischer cell (TIM980 only)
1.
For Instrument, select TIM.
2.
Press 2 (Empty cell).
3.
Press the TIM980 emptying cell button
24
Chapter 2
then press ✓.
System configuration
Proceed as follows to configurate your workstation
1.
Press Stop for 3 seconds in the Main window to enter the Setup
menu.
2.
Supervisor code:
Entering a Supervisor code enables you to differentiate
between the 2 operator modes: Routine and Supervisor:
In Routine mode, the user is able to select and run methods. In
Supervisor mode, the user can create, edit, select and run
methods. A Supervisor code is also used to protect your
parameters from any unwanted changes.
•
3.
Continue without entering a Supervisor code.
Press 1 to enter the Configuration menu.
Use the UP and
DOWN arrow
keys to select
the parameter.
Horizontal
scroll bar
The position of the bar indicates the first screen in the
Configuration menu. Use the RIGHT arrow key
to move to the next screen.
Getting started
25
4.
Press the RIGHT arrow key.
The position of the bar indicates
the last screen in the
Configuration menu.
5.
Press the UP arrow key.
6.
Press the LEFT arrow key to return to the first screen in the
Configuration menu.
26
Chapter 2
Choosing the language
Choose your language for displays and printouts in the following way:
1.
Use the UP and DOWN arrow keys to select the Language line.
2.
Press ✓ to change a parameter as indicated at the bottom of the
screen.
3.
Select the language.
Esc allows you to leave the screen without changing the language.
Getting started
27
Setting the date and time
The current date and time are entered in the following displays:
1.
Select Time.
Enter the hours (from 00 to 23).
2.
RIGHT arrow key.
Enter the minutes (from 00 to 59).
3.
RIGHT arrow key.
Enter the seconds (from 00 to 59).
The LEFT arrow key allows you to return to the previous screen to
modify an entered value.
4.
28
Press ✓ to confirm (as indicated on the screen).
Chapter 2
5.
Select Date.
Enter the day (from 00 to 31).
6.
Press RIGHT arrow key.
Use the UP/DOWN keys to select the
month.
7.
Press RIGHT arrow key.
Enter the year (from 2000 to 2069).
8.
Press ✓.
Getting started
29
Customising the workstation
You can assign a name to your workstation, which will be permanently
displayed in the title bar of the Main window. Typing the name will
allow you to get used to using the instruments keypad.
1.
Before leaving the Configuration menu:
•
Select PC keyboard,
•
select English (US).
This allows you to use a QWERTY keyboard.
2.
Press Esc to return to the SETUP menu.
3.
Press 3 (Customise).
Customise screen
displayed for a
TIM960.
4.
Press ✓ to select the Station parameter.
5.
To replace "TIM960" by "Chem.lab-1", proceed as follows:
•
30
Press 7 until the letter "C" appears, then release the key. The
cursor moves to the next position.
Chapter 2
7
ABC
•
Press 9 until the letter "h" appears.
•
Continue until you have entered (em.lab-1).
•
To correct a typing error, proceed as follows:
Press the LEFT arrow key to position the cursor on the letter
"e". Press 7 to enter the letter "a".
The letter "e" has been inserted between the letter "a" and the
letter "b".
Press Del to delete the "e".
• Press ✓ to confirm the entry.
Getting started
31
6.
You can also use a PC keyboard to enter alphanumeric characters.
•
Connect the PC keyboard to the 6-pin plug situated on the left
hand side of the instrument.
•
Select the line Radiometer Analytical. This line is used
to enter information concerning the workplace, user(s) name(s),
location, address etc.
•
Enter the text using the PC keyboard (maximum of 32 characters
can be used).
If the characters shown on the display do not correspond to the
ones typed on the keyboard, redefine your keyboard type. To do
this, press Esc then 1 and select PC keyboard.
•
Press ✓ to confirm.
You have now finished the getting started section. Press Esc to leave the
Customise screen, then press 5 to quit the SETUP menu.
32
Chapter 2
3. Pr
ogramming guidelines
Programming
Only the Supervisor is allowed to program the menus, Electrode, Reagent,
and Method.
To select the Supervisor mode, press Stop for 3 seconds in the Main
window and enter the Supervisor code. Press 5 (Exit) and select
Return in mode = Supervisor.
✶ IMPORTANT ✶
For first time users, it is recommended to program the instrument as
follows:
1.
Create the electrode(s) to be used to perform the measurements.
2.
Create the reagent(s) to be used during the titration.
Then finally,
3.
Create the method(s), which will consequently use the electrode(s)
and reagent(s) created in the first two steps of programming.
Once you have finished programming, make sure that NO question
marks "?" are displayed in the Reagent and Electrode tabs!
Programming guidelines
33
Create an electrode
1.
Press 4 Electrode Library, then 1 New electrode.
2.
Press ✓ in the Function field and select the function of the
electrode, refer to the table below:
Electrode type
Single pH
Combined pH w/wo temp. sensor
Single metal/redox
Combined metal/redox w/wo temp. sensor
Single ISE
Combined ISE w/o temp. sensor
Reference single
Temperature sensor
Ground metal
Double metal
Select function
pH
pH
mV (i=0)
mV (i=0)
mV (i=0)
mV (i=0)
Reference
T°C
Ground
mV (i > 0)
3.
Press ✓ in the ID field.
4.
Select Other in the From field.
The option From = Catalogue allows you to create an
electrode from a list of Radiometer Analytical electrodes.
5.
Enter the identification (ID) of the electrode, (up to 16 characters).
6.
Press 1 twice.
34
Chapter 3
7.
8.
Select the electrode Type, for electrodes with pH or mV functions:
•
Single pH - pH electrode (no reference part),
•
Combined pH - pH electrode with a reference part (with or
without temperature sensor),
•
Single metal/redox - mV (i=0) function electrode with a reference part,
•
Combined metal/redox - mV (i=0) function electrode with a
reference part (with or without temperature sensor),
•
Single ISE - mV (i=0) function ion selective electrode with no
reference part,
•
Combined ISE - mV (i=0) function ion selective electrode with a
reference part (with or without temperature sensor).
When a combined electrode has been defined, the instrument will
prompt you to specify if it has a built-in temperature sensor or not.
When a single electrode has been defined, the instrument will
prompt you to define a reference electrode in the Edit
electrode screen.
9.
When a combined or a reference electrode has been defined,
enter the potential (in mV) of the reference element versus the
Standard Hydrogen Electrode (SHE).
This parameter enables to run calibration with automatic buffer
recognition when working with different reference electrode systems.
On the next page, see the table giving the potential at 25°C versus
the SHE of a few "reference elements/filling solution" couples.
10. When a combined pH or a single pH electrodes has been defined,
enter the internal pH of the electrode (pH int).
11. Press 1 to confirm.
12. Edit the electrode parameters (see page 37).
Programming guidelines
35
Reference element and filling
solution
Potential
versus SHE
Radiometer Analytical
Electrodes
Hg/Hg2Cl2 - Saturated KCl
SCE: Saturated Calomel Electrode
+244 mV
pHC4000, pHC4001, pHC4006,
XC601, REF401, REF421,
REF451, XR100, XR110,
XR130, XR150, MC408Pt
Hg/Hg2Cl2 - 1M LiCl
+280 mV
REF921
Hg/Hg2SO4 - Saturated K2SO4
+651 mV
REF601, REF621, XR200,
XR230, MC602Pt, MC6091Ag
Hg/Hg2SO4 - 1 M H2SO4
+616 mV
Hg/HgO - 0.1 M KOH
+174 mV
XR400, XR430, XR440
Ag/AgCl - Saturated KCl
+199 mV
XR300, XR820, XC100, XC111,
XC120, XC161, XC200, XC250,
Ag/AgCl - 3 M KCl
+208 mV
pHC3001, pHC3005, pHC3006,
pHC3011, pHC3081, pHC3185,
REF321, REF361, MC3051Pt,
ISEC301F
Ag/AgCl - 1 M KCl
+235 mV
Ag/AgCl - 0.6 M KCl (sea water)
+250 mV
Red Rod - Saturated KCl
+199 mV
36
Chapter 3
pHC2001, pHC2002, pHC2003,
pHC2005, pHC2011, pHC2015,
pHC2051, pHC2085, pHC2401,
pHC2441, pHC2501, pHC2601,
pHC2701, REF200, REF201,
REF251, REF261, MC2095Sb,
MC201Au-8
Edit electrode screen
1.
Press 4 Electrode library.
2.
Press ✓ and select the electrode to be edited from the list.
3.
Press 2 Edit electrode.
4.
Edit the necessary parameters.
a.
Electrode address: indicates which socket the electrode is
connected to:
TIM / E1, E2, Ref, Temp, GND or Pt-Pt (socket of the Titration
Manager),
ABU1 / E1, E2, Ref, Temp, GND, Pt-Pt (socket of an ABU62
connected to the Local socket of the Titration Manager),
ABU2 / E1, E2, Ref, Temp, GND, Pt-Pt (socket of an ABU62
connected to the Slave socket of another ABU62 connected to the
Local socket of the Titration Manager).
Programming guidelines
37
b.
Select whether the electrode should be calibrated or not
(Calibration request = Yes or No).
If Calibration request = No, the electrode parameter
edition is completed.
c.
Enter the Periodicity.
Indicates the maximum period of time between two calibrations. If
the period of time is exceeded, pH measurements can no longer be
performed using this pH electrode and the electrode must be
calibrated.
d.
Enter the Number of cycles.
Enter the number of times the calibration should be repeated. This
is the number of beakers to be prepared for a given standard. For
each standard, the result will be calculated from the mean measured
value.
e.
Enter the Number of buffers (1 to 5) to be used for the
calibration.
f.
Specify if a temperature sensor is in use.
If a temperature sensor is in use, select Probe.
If the temperature is entered manually via the numeric keypad,
select Entered.
If the calibration is to be performed at 25°C, select Fixed 25°C.
If you have selected Temperature = Probe, a temperature
sensor must have been selected in the Parameters screen.
38
Chapter 3
Programming electrode parameters
1.
If Calibrate request = Yes, enter the electrode definition
parameters and go to the last Edit electrode screen.
2.
Press 1:
To enter Calibration parameters: general parameters
used during the calibration,
3.
Press 2:
To enter Calibration solutions parameters concerning
the standards used for the calibration,
4.
Press 3:
To enter the Results parameters concerning acceptation criteria
that you can set on the results.
5.
Press 4:
To enter Printouts parameters defining the calibration report to
be printed.
Programming guidelines
39
Create a reagent
1.
Press 4 Reagent library then 1 New reagent.
2.
Press ✓ in the New reagent screen.
3.
Select From = Other.
The option From = Catalogue allows you to create a reagent
from a list of commonly used reagents.
4.
Press ✓ in the ID field and enter the reagent name (up to 16
characters).
5.
Enter the "approximate" titre of the reagent (5 characters) in the
Target titre field.
6.
Enter the units (mM, M, mN, N) indicated on the reagent bottle. Use
the following: mM = mmol/l, M = mol/l, mN = meq/l, N = eq/l or
mg/ml (Karl-Fischer reagents only).
Note: If molar units are selected (mM or M), it is necessary to enter
the stoichiometric coefficients for the chemical reaction in the
Results screen.
For info: once the units have been confirmed, they are added to the
reagent ID and target titre.
7.
Press 1 twice to create the reagent.
8.
Edit the reagent parameters (see next page).
40
Chapter 3
Edit reagent screen
1.
Press 4 Reagent library.
2.
Press ✓ and select the reagent to be edited from the list.
3.
Press 2 Edit reagent.
4.
Edit the necessary parameters.
a.
Modify the units (mM, M, mN, N) indicated on the reagent bottle. Use
the following: mM = mmol/l, M = mol/l, mN = meq/l, N = eq/l or
mg/ml (Karl-Fischer reagents only).
b.
Address:
Indicates the location of the burette containing the reagent:
TIM/1 or TIM/2 (position 1 or 2 on the Titration Manager),
ABU1/1 or ABU1/2 (position 1 or 2 of an ABU62 connected to the
Local socket of the Titration Manager),
Programming guidelines
41
.
ABU2/1 or ABU2/2 (position 1 or 2 of an ABU62 connected to the
Slave socket of another ABU62 connected to the Local socket of
the Titration Manager).
c.
Select Titre = Enter or Calibrate,
Select Enter if you already know the titrant concentration and
do not wish to perform a calibration.
Select Calibrate, if you wish to perform a calibration. Enter the
calibration parameters; see below.
d.
Select the type of method for Mode. Depending on your Titration
Manager, you have the choice between End Point, Monotonic IP,
Dynamic IP, Continuous IP or Karl-Fischer.
e.
Select the measurement mode for the calibration (pH, mV or mV at
imposed current).
f.
Enter the Periodicity.
Indicates the maximum period of time between two calibrations. If
the period of time has elapsed, calibrations can no longer be performed using this titrant.
g.
Enter the Number of tests.
This is the number of times the calibration method will be repeated,
i.e. the number of beakers to prepare for the calibration.
h.
Specify if a temperature sensor is in use.
If a temperature sensor is in use, select Probe.
If the temperature is entered manually via the numeric keypad,
select Entered.
If the calibration is to be performed at 25°C, select Fixed 25°C.
If you have selected Temperature = Probe, a temperature
sensor must be selected in the Calibration parameters
screen.
i.
42
If you want a message to be displayed upon starting the calibration,
select Notification = Yes then enter the message (3 lines
of 32 characters maximum).
Chapter 3
Programming reagent parameters
1.
If Titre = Calibrate, enter the reagent definition parameters
and go to the last Edit method screen.
2.
Press 1:
To enter Calibration parameters: general parameters
used during the calibration.
3.
Press 2:
To enter Standard parameters concerning the standard solution
used for the calibration.
4.
Press 3:
To enter the Results parameters concerning acceptation criteria
that you can set on the results and stoichiometric coefficients for
the chemical reaction (if required).
5.
Press 4:
To enter Printouts parameters defining the calibration report to
be printed.
Programming guidelines
43
Create a method
1.
Press ✓ and select Method.
2.
Press 4 Method library then 1 New method.
3.
Press ✓ in the ID field.
4.
Enter a method name (up to 16 characters).
5.
Press 1.
6.
Edit the method parameters (see the next page).
44
Chapter 3
Edit method screen
1.
Press 4 Method library.
2.
Press ✓, select the method to be edited from the list then press 2
Edit method.
3.
Edit the necessary parameters.
a.
Select the method mode. The modes available depend on the
Titration Manager used:
•
Measurement (pH, mV or mV at imposed current)
•
End point (titration to a preset mV or pH end point)
•
Coupled (methods chained in a same beaker)
•
Continuous IP (Inflection point detection with continuous addition
of the titrant)
•
Monotonic IP (Inflection point detection with incremental addition
of the titrant, increments of equal size)
Programming guidelines
45
•
Dynamic IP (Inflection point detection with incremental addition
of the titrant, increments of varying size)
•
Karl-Fischer
•
Addition (1 to 3 additions of reagent. No measurements. Additions
are performed simultaneously or successively)
b.
Select the measurement mode: pH, mV or mV at imposed current.
c.
Select the way the temperature is to be measured.
Probe: temperature sensor is required during measurements.
Entered: temperature is to be entered manually using the
keypad.
• Fixed 25°C: measurements are assumed to be performed at
25°C.
Enter the Number of tests, i.e. the number of times the
method is to be repeated.
•
•
d.
e.
If you want a message to be displayed upon starting the method,
select Notification = Yes then enter the message (3
lines of 32 characters maximum).
f.
Select the way the cell grounding of the measuring electrode will be
performed:
•
Reference: reference electrode connected to the Ref socket
on the titration system.
•
Metal: metal electrode connected to the GND socket on the
titration system. For titrations in very resistant medium.
•
Other: an electrode which does not belong to the electrode
system.
g.
Blank: A blank or blank sample is usually a solvent used to dissolve
or dilute the sample. This blank may contain traces of the species
you are analysing. Select Yes for Blank when you want to
subtract the volume of titrant used to titrate the solvent from the
volume found for the dissolved or diluted sample.
h.
QC sample: the concentration of a QC sample is known accurately
and the composition is close as possible to that of the samples to
be analysed. QC measurements enable you to perform quality
control tests on the method used. Refer to the Reference Manual for
more information concerning this option. Select Yes for QC
sample to perform measurements on QC samples.
46
Chapter 3
Programming method parameters
1.
Go to the last Edit method screen.
2.
Enter the parameters. The parameters displayed will all depend on
the measurement type and mode selected.
3.
Press 1:
To enter the Method parameters: general parameters
concerning the electrode and reagent (when required) used by the
method and the titration parameters.
4.
Press 2:
To enter the Sample parameters: concerning the preparation of
the sample to be analysed.
5.
Press 3:
To enter the Results parameters defining the results, the
acceptation criteria that you can set on the results and stoichiometric coefficients for the chemical reaction (if required).
6.
Press 4:
To enter the Printouts: parameters defining the analysis report
to be printed.
7.
Press 5:
To enter the QC data (only available when QC sample = Yes):
quality control parameters required when using QC samples.
Programming guidelines
47
48
Chapter 3
4. Running analyses
✶ IMPORTANT ✶
Before pressing the Run key, check the following points listed below:
1.
Select the method by pressing 3 in the main window.
2.
If Question marks "?" are present in the electrode and/or reagent
tabs. Programming error due to missing electrode/reagent
parameters. Contact your supervisor.
3.
Connect/Check electrode(s). Refer to Short-Form Reminder no. 1.
When installing/checking an electrode system, do not forget to
connect all the installed electrodes to the Titration Manager or
ABU62 sockets and to install these electrodes on the Titration
Manager (Location = TIM) or the SAC80 or SAC90 Sample Changer
(Location = SAC).
It is important to immerse, in the same measuring beaker, all the
installed electrodes having the same location.
For example: mV at imposed current and pH measurements.
If your installed electrode system comprises a double platinum
electrode with a combined pH electrode, the electrodes must remain
permanently immersed in the same beaker, in order to measure the
pH correctly. This is because it is the double platinum electrode that
provides the connection to the ground and not the reference
electrode.
This so-called " 3-electrode or differential measurement
configuration" will stop the double platinum electrode altering the
measurement at the pH electrode.
When using this configuration for an electrode calibration, the double
platinum electrode and the pH electrode must always be immersed in
the same beaker.
Important: Never remove an electrode or an electrode system
without running the Installation/checking procedure first. The
instrument needs to know its configuration at any time.
For more information, refer to the Reference Manual of your titration
system, keyword index: "Electrode connection - Important".
Running analyses
49
4.
Connect/Check reagent(s). Refer to Short-Form Reminder no. 2.
5.
Run the analysis when Sunny icons are visible in the electrode and
reagent tabs.
6.
Run an electrode calibration when a Stormy icon appears in the
electrode tab.
7.
Run a reagent calibration when a Stormy icon appears in the
titrant tab.
8.
If you are unable to display the "Run" command due to the presence
of Cloudy/Stormy icons in the Reagent and Electrode windows,
activate the "Check" command. The Titration Manager will
automatically guide you through the necessary operations required
to solve the problem(s).
Run electrode calibration
Refer to Short-Form Reminder no. 4.
1.
Select the method.
2.
Install the electrode group.
3.
Enter the Electrodes window.
4.
Select 1 Calibrate electrode. Press ✓ to select the
electrode to be calibrated from the proposed list.
5.
Press 1 to start the calibration.
Run reagent calibration
Refer to Short-Form Reminder no. 5.
1.
Select the method.
2.
Install the reagent to be calibrated.
3.
Install the electrode group to be used to perform the calibration.
4.
Enter the Reagents window.
5.
Select 1 Calibrate/Enter titre. Press ✓ and select the
reagent from the proposed list.
6.
Press 1 to start the calibration.
50 Chapter 4
Run method
Refer to Short-Form Reminder no. 6.
1.
Select the method in the Method library.
2.
Install the electrode then the reagent groups to be used to perform
the analysis.
3.
Display the Main window.
4.
Press 1 to start the analysis.
Run direct measurement
Refer to Short-Form Reminder no. 8.
1.
Select the method.
2.
Install the electrode group to be used to perform the measurement.
3.
Prepare the sample solution and place on the sample stand.
4.
Display the Electrodes window.
5.
Press 5 to start the measurement.
6.
Select the electrode in the list of connected electrodes. You can
stop or start stirring by pressing 1.
Running analyses
51
Run manual dosing
Refer to Short-Form Reminder no. 10.
1.
Select the method.
2.
Install the electrode group to be used to perform the measurement
(if measurements are to be done).
3.
Install the reagent group to be used to perform the manual dosing.
4.
Prepare the sample solution and place on the sample stand.
5.
Display the Reagents window and press 5.
6.
Select the reagent in the list of installed reagents and select the
electrode in the list of connected electrodes (if measurements are to
be done).
7.
Enter an increment size, increment speed and maximum volume to
be dosed.
8.
Select the type of measurement to be performed (pH, mV, mV (i>0)
or no measurement).
9.
Press 1 to start.
52 Chapter 4
5. V
iewing data
Viewing
How to access?
The GLP-Archives (Good Laboratory Practice) command is available in
each window:
•
Main: press 5 to access,
•
Reagents: press 6 to access,
Electrodes: press 6 to access.
The GLP-Archives window displays the method, reagent or
electrode ID with the expiry date (reagent) and the installation date
(electrode).
Press ✓ and enter the result
number from where you want to
start visualising the results.
Example: result no. 9 out of 26
results stored for the method
selected.
Press 1 to consult.
The following pages, give a
summary of the information
available in the Main,
Electrodes and
Reagents windows.
Press 2 to consult the global variables.
Refer to the last page of this chapter.
Viewing data
53
Sample results
The last 200 sample results are saved in the archives.
Once you have selected the method and the result (see previous page),
the following data are displayed:
Result no. 9 / 26
Result no. 10 / 26
26 is the
number
of results
saved for
the
selected
method.
Scroll the results.
The results can be: Test results (Rx) or Average ± standard deviation
calculated on several tests of the same sample. "TIM" (or "SAC") means
that a TIM (or SAC) method has been run. "CPL, TIM" (or "CPL, SAC")
means that the method run belongs to a coupled method.
Mean and standard deviation
calculated on three R1 (and R2)
accepted test results.
54
Chapter 5
R1 and R2 = result of test
no. 3 (test accepted).
Reagent calibration results
The last 100 reagent calibration results are saved in the archives.
Once you have selected the reagent then the result (see first page of this
chapter), the following data are displayed:
Result no. 1 / 28
Result no. 2/ 28
28 is the
number
of results
saved for
the
selected
reagent.
Scroll the results.
The results can be: reagent titre entered by the operator, test performed
on a standard, mean calculated on several tests performed using the
same standard or the current calibration data. "TIM" (or "SAC") means
that a TIM (or SAC) method has been run.
Results of the current
calibration
Mean and standard deviation calculated on 4 tests performed and
accepted on the same standard.
Viewing data
55
Electrode calibration results
The last 100 electrode calibration results are saved in the archives.
Once you have selected the electrode then the result (see first page of
this chapter), the following data are displayed (for a pH calibration):
Result no. 14 / 19
Result no. 13 / 19
19 is the
number of
results
saved for
the
selected
electrode.
Scroll the results.
The results can be: cycle performed on a standard, mean calculated on
several cycles performed on the same standard or current calibration
data. "TIM" (or "SAC") means that a TIM (or SAC) method has been run.
Mean calculated on 3 cycles
performed and accepted. Each
cycle has been run using 2 buffer
solutions.
56
Chapter 5
Result of cycle no. 3 (cycle
accepted). This cycle has been
run using 2 buffer solutions.
Global variables
To view the G1 to G10 global variables:
•
Enter the Method/sequence window.
•
Press 5 (GLP-Archives).
•
Press 2 (Global variable).
Global variable G1
Global variable G2
Scroll the G1 to G10 global variables.
Viewing data
57
58
Chapter 5
6. Printing data
✶ IMPORTANT ✶
Before printing, it is necessary to perform the following:
1.
Enter the Setup window: press Stop 3 seconds in the Main
window.
2.
Select the Printer
Press 1.
In Configuration, select Printer = 80 columns and
Format = Listing or Page by Page.
3.
Enter User ID (if required)
In Configuration, select User ID = Yes. You will be
prompted to enter a user ID at the start of a run method. This ID will
appear on the printouts.
4.
Customise the printout (if required)
In Configuration, press 3.
In Customise, enter the name of your workstation (max. 4 lines
of 32 characters). This information will appear as a header at the
start of the printout.
5.
For automatic printout - select a condensed or detailed printout
In the Printouts screen of the Edit method/reagent/electrode,
select Detailed = High to obtain a full detailed printout.
Select Low for condensed printout.
Printing data
59
Manual printouts
Method library
Press Print, in the Main window to give you an overview of the methods
available in the method list.
Reagent library
Press Print, in the Reagent window to give you an overview of the
reagents available in the reagent list.
Electrode library
Press Print, in the Electrode window to give you an overview of
the electrodes available in the electrode list.
60
Chapter 6
Edit method data
In the Main window, press 4 Method library then Print to give you
an overview of the parameters of the current programmed method. These
are the parameters entered in the Edit method screen (press 2 to
access).
Edit reagent data
Press 4 Reagent library in the Reagent window then Print to
give you an overview of the parameters of the current programmed
reagent. These are the parameters entered in the Edit reagent
screen (press 2 to access).
Edit electrode data
Press 4 Electrode library in the Electrode window then
Print to give you an overview of the parameters of the current
programmed electrode. These are the parameters entered in the Edit
electrode screen (press 2 to access).
Printing data
61
Automatic printouts
The Print key is inactive during a titration, pH/mV measurement,
electrode or reagent calibration.
The results obtained during a "Run" are printed automatically.
Depending on the option selected for Detailed in the Edit method/
reagent/electrode - Printout window, you will obtain different types of
printouts.
Example of a printout for a titration
Header: information entered
in the Customise screen
of the Setup window.
Title of report:
entered in Printouts
screen during
method editing
Foot note: automatically appears
at the end of all printouts
Titration results: obtained at
the end of the analysis
Analysis ID: User ID and Sample
ID entered at the start of the
titration (if option selected during
editing)
Calibration data: of the electrode
and reagent used to perform the
titration
62
Chapter 6
7. Connection of peripherals
PC, cable A95X501
Printer, cables A95P201/
A95X506
ABU62, cable A95X501
Balance Mettler (A95Z201,
A95Z205, A95Z206)
Sartorius, (A95Z202, A95Z204)
Sample Changer, cable
A95A202
684R999N999
MADE IN FRANCE
RADIOMETER ANALYTICAL SAS
FCC ID UHC-TTL090
This device complies with part 15 of the FCC Rules. Operation is subject
to the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation
Conforms to
Conforms
UL STD 61010A-1
Certified
Cer
tified to
CSA STD C22.2 No. 1010.1
3020116
PROPELLER
SLAVE
VE
PC/PRINTER
LOCAL
BALANCE
SAC80/90
AUX.
LINE FUSE T1A L250V
12V
V OUT
IN
5V
V OUT
100-240Vac
50-60Hz 45VA
45
GND
TEMP
TTL
T 1A 250V
Pt-Pt
REF
E2
E1
Line cord, 230 V (A95S001),
115 V (A95S002)
Single metal electrodes
for grounding only
Temperature sensors
Combined pH electrodes
w/wo temperature sensor
Single pH electrodes
Double metal electrodes
Single or combined metal
electrodes (Redox
measurements)
Reference electrodes
Single or combined ion
selective electrodes
Connection of peripherals
63
64
Chapter 7
8. Maintenance
Cleaning
The Titration Manager requires minimum maintenance. The exterior
surface can be cleaned with tepid water and wiped dry with a soft cloth.
Never use another solvent unless you have first consulted your
Radiometer Analytical representative
Transpor
ting the instrument
ransporting
Always use the packaging supplied by the manufacturer.
IMPORTANT!
Remove the metal rod before transporting the instrument.
Never pick-up or carry the instrument by the metal rod.
Servicing
DO NOT ATTEMPT TO SERVICE THIS PRODUCT YOURSELF,
except as noted in the Reference Manual. For servicing, please contact
your Radiometer Analytical service representative at the address given
below:
RADIOMETER ANALYTICAL SAS
72, rue d'Alsace
69627 Villeurbanne CEDEX - France
Tel.: +33 (0) 4 78 03 38 38
Fax: +33 (0) 4 78 03 38 27
E-mail: radiometer@nalytical.com
or your local service representative:
_________________________________
_________________________________
_________________________________
_________________________________
_________________________________
General information
65
Inter
national Standar
ds
International
Standards
RADIOMETER ANALYTICAL SAS
72 rue d’Alsace 69627 Villeurbanne Cedex, France
E-mail : radiometer@nalytical.com Web : www.radiometer-analytical.com
Tél. : +33 (0)4 78 03 38 38 - Fax : +33 (0)4 78 68 88 12
CE Marking
DECLARATION OF CONFORMITY
We, Radiometer Analytical SAS,
declare under our sole responsibility that the Potentiometric Titration Workstations TIM9xx,
including TIM960, TIM965, TIM980 and ABU62
are in conformity with the provisions of:
Radio and Telecommunications equipment directive 99/5/EEC, dated 9/3/1999,
Low voltage directive 73/23/EEC, dated 19/2/73
EMC directive 89/336/EEC, dated 3/5/1989, 92/31/EEC, dated 28/4/1992,
CE harmonisation directive 93/68/EEC, dated 22/7/1993,
following standards and severity levels of:
EN 61010-1, 1995
EN 300330-1, 1999,
ETSI TR102 070, 2002, and EN 301 489, 2001,
CISPR16-1, 1999, and EN 61326-1, 1997,
EN 55022, 1998, class A,
EN 61000-3-2, 2000, class A and EN 61000-3-3, 1995,
EN 61000-4-2, 1995, level 2 (4kV) with contact discharges and level 3 (8kV) with air discharges,
EN 61000-4-3, 2002, level 2 (3V/m),
EN 61000-4-4, 1995, level 2 (1kV) on AC power line,
EN 61000-4-5, 1995, level 2
EN 61000-4-6, 1996, level 2 (3V),
EN 61000-4-11, 1994,
after tests and qualification of our products, regarding "Radio" and "EMC" directives, performed by:
AEMC Lab, Sassenage (France) independent laboratory, Cofrac accredited under the number 1-0805,
with a Radio qualification report no. R0603066R-E-C and a EMC qualification report no.
R0603066C-E-C, dated June 15, 2006 which attests the conformity of the TIM9xx workstations,
and tests and qualification of the product, regarding "Low Voltage" directive, attested by our company, according
to the self-assessment procedure (article 10.1), recommended for the CE marking, as these products are UL / CSA
certified.
Year of the first CE marking: 2006.
Villeurbanne, July 20, 2006
RADIOMETER ANALYTICAL SAS
Ref. Document : A090B09A
66
Chapter 8
Georges RIVOIRARD
Quality Manager
Page 1/1
The Titration Manager complies with the following standards:
UL 61010A - 1
CAN / CSA C22 2 N° 1010.1 - 92.
The Titration Manager TIM9xx complies with part 15 of the FCC Rules.
Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including
interference that may cause undesired operation.
See "Compliance with FCC rules, part 15 - Information to the user" on
page 5.
General information
67
68
Chapter 8

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : No
Create Date                     : 2006:08:02 12:48:58Z
Modify Date                     : 2006:08:02 14:50:04+02:00
Subject                         : Usersguide_gb
Page Count                      : 68
Creation Date                   : 2006:08:02 12:48:58Z
Mod Date                        : 2006:08:02 14:50:04+02:00
Producer                        : Acrobat Distiller 5.0.5 (Windows)
Author                          : HWimmer
Keywords                        : 
Metadata Date                   : 2006:08:02 14:50:04+02:00
Creator                         : HWimmer
Title                           : Usersguide_gb
Description                     : Usersguide_gb
Page Mode                       : UseOutlines
EXIF Metadata provided by EXIF.tools
FCC ID Filing: UHC-TTL090

Navigation menu