SD Biosensor 01GM24 NFC Blood Glucose Monitor User Manual GM24 GlucoNavii Mentor NFC F BW

SD Biosensor, Inc. NFC Blood Glucose Monitor GM24 GlucoNavii Mentor NFC F BW

Mentor manual

Mentor NFCCMentorNFC
ISD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMDear SD GlucoNavii® Mentor NFC  System Owner;Thank you for choosing SD GlucoNavii® Mentor NFC  Blood Glucose Monitoring System. Your new SD GlucoNavii® Mentor NFC  Blood Glucose Monitoring System is an important tool that can help you better manage your diabetes. Important steps for using the System are inside this guide. Please read it carefully. If you have questions, we are here to help.Please contact SD Biosensor, Inc.Tel : +82-31-300-0400Fax : +82-31-300-0499website : www.sdbiosensor.comWe oer assistance 24 hours a day, 365 days a year in many languages. You can also visit www.sdbiosensor.com for diabetes management tools and product demonstrations.Please refer to the instructions with following symbols in this User Instruction Guide.To identify conditions or practices that could result in damage to equipment or other property.To provide an additional useful information. Before using any product to test your blood glucose, read all instructions and practice the test. Do all quality control checks as directed and consult with a diabetes health care professional. These recommendations apply to all blood glucose monitoring systems and are supported by the American Association of Diabetes Educators, the American Diabetes Association, the U.S. Food and Drug Administration, and the Advanced Medical Technology Association.CAUTIONNOTENOTEnuGtGumj|GYWXZW[X]ZUGGGX YWXZTW[TX]GGG㝘䟸G\a\_a[W
2SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMTABLE OF CONTENTSChapter 1 : Understanding Your New System    1.  Before You Start Testing                      2.  Special Information for Healthcare Providers and Caregivers                                  3.  Indication for Use 4.  Product Description and the Principle of the use5.  The Complete SD GlucoNavii® Mentor NFC Blood Glucose Monitoring System6.  SD GlucoNavii® Mentor NFC  Meter7.  SD GlucoNavii® Mentor NFC  Test Strips8.  SD GlucoNavii® Mentor NFC  Accessories 9.  Changing the Battery10.  Meter Set up11.  Using SD GlucoNavii® Mentor NFC  Test StripsChapter 2 : Control Solution Test1.  Performing a Control Solution Test           2.  Troubleshooting Control Solution TestChapter 3 : Testing Your Blood Glucose1.  Getting a Drop of Blood2.  Performing a Blood Glucose Test3.  Understanding Test Results44666781213141626TABLE OF CONTENTS28293234343638nuGtGumj|GYWXZW[X]ZUGGGY YWXZTW[TX]GGG㝘䟸G\a\_a[W
3SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMTABLE OF CONTENTSChapter 4 : Using NFC function1.  Information on NFC function2.  Information on Caution for NFC3.  Using NFC function4.  How to use GlucoNavii App5.  How to use GlucoNavii DMS for PCChapter 5 : Using the Meter Memory1.  Searching Test Results2.  Downloading results to a computerChapter 6 : Maintenance and Troubleshooting1.  Performing a SD Glucose check strip2.  Cleaning the meter3.  Maintenance, Testing and Transportation4.  Cleaning the Lancing Device5.  Screen Messages and Troubleshooting6.  Warnings, Precautions and Limitations Chapter 7 : Product Technical Information1.  System SpecicationsAnnex 1 :  Information for Healthcare ProfessionalsAnnex 2 :  SymbolAnnex 3 :  References404040414243454650515152535455626464666769nuGtGumj|GYWXZW[X]ZUGGGZ YWXZTW[TX]GGG㝘䟸G\a\_a[X
4Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMCHAPTER 1 : Understanding Your New SystemThe SD GlucoNavii® Mentor NFC  Blood Glucose Monitoring System1. Before You Start TestingAbout the meter and test strips t Carefully  read  and  follow  the  instructions  in  the  User Instruction  Guide and Package Inserts  for the test strips and  SD Glucose control  solution. It  is  very  important  to follow  the  instructions  in  order  to  prevent  an  incorrect result or improper treatment. t The meter, test strips, and SD Glucose control solution are only for use outside the body (in vitro). t Your  new  meter  is  designed  for  testing  fresh  capillary whole  blood  samples  (for  example,  blood  from  your ngertip, palm, upper arm, or forearm). t Only  use  SD  GlucoNavii®  Mentor  test  strips.  Other  test strips will give inaccurate results or E-1 error message.t Do not use the SD GlucoNavii® Mentor NFC blood glucose monitoring  system  for  testing  of  serum  or  plasma  or arterial blood.t Inspect the container of test strips before using them for the rst time. If you see any damage to the container cap or if anything prevents the cap from closing properly, do not  use  test  strips.  Contact  SD  Biosensor,  Inc.  Damaged test strips can cause inaccurate results, which could lead to improper treatment. About your new meter t Set the beep, date, time, hypo warning, post-meal alarm and alarm on your meter before you begin testing. t SD GlucoNavii® Mentor NFC system has been found to be accurate at altitudes up to 12,388 feet. (3,776 meters) nuGtGumj|GYWXZW[X]ZUGGG[ YWXZTW[TX]GGG㝘䟸G\a\_a[X
5Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMt Keep the meter and testing supplies away from small children. t The battery door, test strips, lancets, protective disks, and control solution cap are choking hazards. t Do not eat the test strips. t Do not swallow or inject SD Glucose control solution, or use control solutions for any purpose. CAUTIONImportant Informationt Dehydration: Severe dehydration resulting from excessive water loss may  cause false low  results. If you believe  you are suering from severe dehydration, consult a healthcare professional immediately. t Low glucose results: If  your test result is lower than  70 mg/dL or is shown as Lo, it may mean hypoglycemia(low blood  glucose).  This  may  require  immediate  treatment according to your healthcare professional’s recommenda-tions. Although this result could be due to a test error, it is safer to treat rst, and then repeat the test. t High glucose results: If your test result is greater than 180 mg/dL or is shown as HI, it may mean hyperglycemia(high blood glucose). If you do not have symptoms, rst repeat the test. Your healthcare professional can work with you to  decide  what  actions,  if  any,  you  should  take  if  you continue to get  results higher than  180 mg/dL or  if you have symptoms. t Repeated  unexpected  results:  If  you  continue  to  get unexpected  results,  check  your  system  with  control solution. See Checking the System with Control Solution . If you are experiencing symptoms that are not consistent with  your  blood  glucose  results  and  you  have  followed all  instructions  in  this  User  Instruction  Guide,  call  your nuGtGumj|GYWXZW[X]ZUGGG\ YWXZTW[TX]GGG㝘䟸G\a\_a[X
6Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMhealthcare professional. Never ignore symptoms or make signicant  changes  to  your  diabetes  control  program without speaking to your healthcare professional. t Consult your physician to determine if it is appropriate for your child to be taught how to use the meter system or any other medical products.2.     Special Information for Healthcare Providers and Caregiverst  Do  not  use  this  device  to  measure  blood  glucose  in people  who  are  experiencing  cardiovascular  collapse (severe shock) or decreased peripheral blood ow. t Hematocrit: A hematocrit (percentage of your blood that is red blood cells) that is over 60% or below 20% can cause false results. 3.  Indication for use (Purpose of the devices)Your new SD GlucoNavii® Mentor NFC meter and accessories work together to measure the amount of glucose (sugar) in your blood.  Your  SD  GlucoNavii®  Mentor  blood  glucose  monitoring system  is  indicated  for  monitoring  glucose  in  fresh capillary  whole  blood  or  fresh  venous  blood.  Testing is  done  outside  the  body.  (in  vitro  diagnostic  use)  This system is indicated for home (over-the-counter or OTC) by person with diabetes, or in clinical settings by healthcare professionals,  as  an  aid  to  monitor  the  eectiveness  of diabetes control. This system should not be used for the diagnosis of diabetes.The SD GlucoNavii® Mentor NFC blood glucose monitoring system is suitable for self-testing.4.  Product Description and the Principle of the useSD  GlucoNavii®  Mentor  test  strip  is  designed  with  an electrode  that  measures  glucose  levels.  Glucose  in  the blood  sample  mixes  with  reagent  on  the  test  strip  that cause a small electric current. The amount of current that nuGtGumj|GYWXZW[X]ZUGGG] YWXZTW[TX]GGG㝘䟸G\a\_a[X
7Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMis created depends on how much glucose is in the blood.SD GlucoNavii® Mentor NFC meter measures the current that  is  created  and  converts  the  measurement  to  the amount of glucose that is in the blood.The blood glucose result is displayed on the meter’s LCD display.By  touching  a  drop  of  blood  to  the  tip  of  the  SD GlucoNavii® Mentor test strip, the strip’s reaction chamber automatically  draws  the  blood  into  the  strip  through capillary action. When the chamber is full, SD GlucoNavii® Mentor  NFC  meter  start  to  measure  the  blood  glucose level.  It  is  a  simple  and  practical  system  for  the  daily monitoring of your blood glucose level.5.  The Complete  GlucoNavii® NFC  BGMSThe system includes: t SD GlucoNavii® Mentor NFC  Meter t SD Glucose Check strip  t 3V battery type CR2032t User Instruction Guide t Quick Guidet Carrying Case t Lancing devicet Lancet<Option>t SD GlucoNavii® Mentor Test strips t SD GlucoNavii® Mentor Package Insertt SD Glucose Control Solutiont SD Glucose Control Solution Package Insertt Self-test Diary t GlucoNavii Application for Android phonet GlucoNavii DMS for PCt NFC Reader/WriternuGtGumj|GYWXZW[X]ZUGGG^ YWXZTW[TX]GGG㝘䟸G\a\_a[X
8Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMDisplayShows blood glucose result, messages, and glucose result stored in memory.MentorNFC6. SD GlucoNavii® Mentor NFC  MeterArrow ButtonsUse for meter setup and review of memory moving to the right and left.ON/OFF ButtonPress to turn meter ON or OFF.Test Strip SlotInsert test strip here.Battery CoverRemove cover to change battery.Location of NFC functionTap a smartphone or SD NFC R/W for communication.Data portCominucate test results with a personal computer. (If you have software)nuGtGumj|GYWXZW[X]ZUGGG_ YWXZTW[TX]GGG㝘䟸G\a\_a[X
9Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMIndicates Measurement timeIndicates during the meter settingIndicates beep settingWarns when thebattery is low or must be replacedIndicates if environmentaltemperature is exceeded during testingIndicates post-meal or pre-mealTesting dateIndicates a test result stored in memoryUnit of test result Test result and nfc messageDisplaynuGtGumj|GYWXZW[X]ZUGGG` YWXZTW[TX]GGG㝘䟸G\a\_a[X
10Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMIndicates the average result Test stripTell you when to apply the sample and indicates if you select whole blood for blood reference typeIndicates a controlsolution test resultIndicates alarm settingIndicates hypo warningIndicates S/W communicationnuGtGumj|GYWXZW[X]ZUGGGXW YWXZTW[TX]GGG㝘䟸G\a\_a[Y
11Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM1) Strip Stand-by DisplayAfter  turning  on  the  meter,  the  below  will  be  displayed automatically  to show  the  test  strip  icon  ashing. In  this Strip Stand-by Display, it’s able to enter setting mode of the meter or search the test results.2) Blood Stand-by DisplayIf a test strip is inserted into the meter in either Strip Stand-by  Display  or  turning  o,  the  below  will  be  displayed automatically  to  show  the  appropriate  test  result  after applying  blood sample.  And  if you  pull  o  the test  strip, it will return to Strip Stand-by Display. In this Blood Stand-by Display, the ON/OFF button doesn’t act. But if you need to check the meter or the test strip, you may use a control solution. Then the control solution icon in the left-side of test  strip  icon  will  be  displayed  by  pressing  left  button during 3 seconds.nuGtGumj|GYWXZW[X]ZUGGGXX YWXZTW[TX]GGG㝘䟸G\a\_a[Y
12Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM7.   SD GlucoNavii® Mentor  Test Strip SD GlucoNavii® Mentor NFC System measures the amount of glucose in whole blood. Blood is applied in the Yellow Window (TOP EDGE) of SD GlucoNavii® Mentor  Test Strip and is  automatically drawn into the  reaction cell where the reaction takes place.  (Top Edge)nuGtGumj|GYWXZW[X]ZUGGGXY YWXZTW[TX]GGG㝘䟸G\a\_a[Y
13Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMSD GlucoNavii® control solutionNFC Reader/Writer 8.  SD GlucoNavii® Mentor NFC  Accessories 12Lancing DeviceCap & Comfort dial with puncture depth selectionArmingKnobReleaseButtonCheck Strip  Lancet SD Glucose check strip ElectrodeArmingKnobnuGtGumj|GYWXZW[X]ZUGGGXZ YWXZTW[TX]GGG㝘䟸G\a\_a[Z
14Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM9. Changing the BatteryInserting and replacing the BatteryYour  meter  is  shipped  with  one  3V  battery  type  CR2032 that needs to be inserted before testing. The battery that comes with your meter can be found in the mesh pocket of your  carrying  case. Battery  life will  vary  depending on usage, so  always  keep a  spare  on hand. The  meter saves battery power by automatically turning o after 1 minute without inserting a test strip or 3 minutes with a test strip, from non-use. If the meter does automatically shut o, all tests in memory are saved.STEP-1:  Push  the  recessed  plastic  tab  of  the  battery compartment forward to ip and open the battery door.nuGtGumj|GYWXZW[X]ZUGGGX[ YWXZTW[TX]GGG㝘䟸G\a\_a[Z
15Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMSTEP-2:  Insert the 3V battery (type CR2032) into the compartment with “+” side facing you.   STEP-3:  Snap battery cover back in place.   STEP-4:  Push the ON/OFF button or insert a strip to start testing.After inserting or replacing the battery, conrm that the time and date are set correctly. If they are not, use the ON/OFF and left/right buttons to reset the meter before testing. See the “10. Meter Set up”. NOTEnuGtGumj|GYWXZW[X]ZUGGGX\ YWXZTW[TX]GGG㝘䟸G\a\_a[Z
16Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM10. Meter Set upSTEP-1 : Setting the Audible Beep1.  In  Strip  Stand-by  Display,  if  you  press  the  ON/OFF button  during  3  seconds, the  display  for  setting  the beep will appear, the rst step of the setting mode.2.  Set  the  beep  mode  on  or  o  b  y  pressing  either the  left  or  the  right  button  and  then  selecting  the preferred  feature  by pressing  the  ON/OFF button.  If you select the beep on feature, a ‘beep’ sound is made at the same time; otherwise, if you select the beep o feature, no sound is made. [ During 3 sec. ][ Left or right button ]MentMentor[ ON/OFF button ]nuGtGumj|GYWXZW[X]ZUGGGX] YWXZTW[TX]GGG㝘䟸G\a\_a[[
17Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMSTEP-2: Setting the Hypo warning  1.  After  beep  setting,  the  display  for  setting  the  hypo warning will appear, the second of setting mode.2.  Set the hypo warning mode ‘o’ or the result you want to select among 60, 70, 80 mg/dL(3.3, 3.9, 4.4 mmol/L) by pressing either the left or the right button and then selecting the preferred feature by pressing the ON/OFF button.t You can set the meter to let you know when your result indicates a possible low blood glucose (hypoglycemia). You can also select what blood glucose level you want this indicator to have 60, 70, 80 mg/dL(3.3, 3.9, 4.4 mmol/L).t If your results are lower than selected hypo result, the candy symbol will appear on LCD with a ‘beep’ sound. It is very important to manage your hypoglycemia.NOTEnuGtGumj|GYWXZW[X]ZUGGGX^ YWXZTW[TX]GGG㝘䟸G\a\_a[\
18Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMSTEP3: Setting the date and time[ Date Setting ]1.  The  third  step  of  setting  mode  is  the  Date  &  Time setting. After setting the hypo warning, the display for setting Date  & Time  will appear, the  third step  of the setting  mode. Set  the  correct year  by  pressing either the left or the right button and then select the correct year by pressing the ON/OFF button. Your new meter comes with a preset time and date. You may need to change the time to your time zone. Having the right time and date in your meter is important if you use the meter memory. It also helps your healthcare team interpret your results.NOTEnuGtGumj|GYWXZW[X]ZUGGGX_ YWXZTW[TX]GGG㝘䟸G\a\_a[\
19Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM[ Left or right button ]MentMentor[ ON/OFF button ]2.  Next will appear the setting display for month and day format. The  meter  can  display  the  month  and  day  in either  a  Month-Day (m-d)  format or  a  Day-Month  (d-m) format. Set the preferred format on the display by pressing either the left or the right button and select by pressing the ON/OFF button.3.  Set the correct month or day on the display by pressing either the left or the right button and select by pressing the ON/OFF button.nuGtGumj|GYWXZW[X]ZUGGGX` YWXZTW[TX]GGG㝘䟸G\a\_a[]
20Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM[ Time Setting ]1.  Next will appear the display for setting the 12 or 24 Hour clock format. The meter can display the time in either the  12h  format  or  the  24h  format.  Set  the  preferred format on the display by pressing either the left or the right button and select by pressing the ON/OFF button. 2.  Next will appear the setting display for time format. Set the correct hour and minute on the display by pressing either the left or the right button and select the correct time by pressing the ON/OFF button.[ Left or right button ]MentMentor[ ON/OFF button ]nuGtGumj|GYWXZW[X]ZUGGGYW YWXZTW[TX]GGG㝘䟸G\a\_a[^
21Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMSTEP4: Post-meal alarmYou  can  use  the  meter’s  post-meal  alarm  function  to remind you to test your blood glucose after meal.1.  After day  and time  setting, the  display for setting  the post-meal alarm will appear the fourth of setting mode.2.  Set  the  post-meal alarm  mode ‘2h’  or ‘o’  by  pressing either the  left or  the right button and  then select  the preferred feature by pressing the ON/OFF button.  [ Left or right button ]MentMentor[ ON/OFF button ]nuGtGumj|GYWXZW[X]ZUGGGYX YWXZTW[TX]GGG㝘䟸G\a\_a[_
22Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMt If you select the post-meal alarm ‘2h’ feature and test with pre-meal mark, the ‘clock symbol’ will appear on result display and the ‘beep’ sound will be made in 2 hours to remind you to test your blood glucose after meal for 1mimute. t If you perform the pre-meal test while the post-meal alarm setting is on, then the post-meal mark will appear automatically on your LCD when you test within following period: from 30min to 130min after your pre-meal test. t If you mark the new test result with a pre-meal mark, the old alarm setting will be ignored and only the new setting will sound in 2 hours.NOTEnuGtGumj|GYWXZW[X]ZUGGGYY YWXZTW[TX]GGG㝘䟸G\a\_a[_
23Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMSTEP5: Setting the alarmYou can use the meter’s alarm function to remind you to test your blood glucose.1.  After Post-meal alarm setting the display for setting the alarm will appear, the fth of setting mode.  2.  Set the rst alarm on or o by pressing either the left or the right button and then select the preferred feature by pressing the ON/OFF button.[ Left or right button ]MentMentor[ ON/OFF button ]t If you select the alarm o feature, next will appear the Strip Stand-by Display. t If you select the alarm on feature, you can set the alarm up to four times a day at any time you want.NOTEnuGtGumj|GYWXZW[X]ZUGGGYZ YWXZTW[TX]GGG㝘䟸G\a\_a[`
24Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM3.  If you select the alarm  on feature in rst alarm mode, the clock will blink. Set the correct time and minute you want to set an alarm on the display by pressing either the left or the right button and then select the preferred feature by pressing the ON/OFF button.4.  If  you nish setting the rst alarm, next will appear the second alarm setting mode. Set the alarm with the same way as above. [2, 3]5.  You can set the third and fourth alarm mode with the same way as above. [2, 3][ Left or right button ]MentMentor[ ON/OFF button ]If you select the alarm o feature in rst (also second, third and fourth) alarm mode, next will appear the Strip Stand-by Display.NOTEnuGtGumj|GYWXZW[X]ZUGGGY[ YWXZTW[TX]GGG㝘䟸G\a\_a\X
25Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMIf you nish alarm setting, next appear at the strip-stand by display.NOTEnuGtGumj|GYWXZW[X]ZUGGGY\ YWXZTW[TX]GGG㝘䟸G\a\_a\Z
26Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM11.  Using SD GlucoNavii® Mentor NFC  Test Strips[ Important Test Strip Information ]t SD  GlucoNavii®  Mentor  NFC  blood  glucose  test  strip should be used with SD GlucoNavii® Mentor NFC meter. Using  other  glucose  test  strip  can  cause  inaccurate  the result or ‘E-1’ error message.t After removing a test strip from the container, replace the container cap immediately and close it tightly. t Use the test strip within three minutes after you take it out of the container. t Store test strip containers in a cool, dry place at 2-32°C(36-90°F).  Keep  away  from  direct  sunlight  and  heat.  Do  not refrigerate test strips. t Do  not  expose  strips  to  heat,  moisture  or  humidity. Temperatures  outside  the  required  range,  as  well  as moisture and  humidity  (e.g. bathroom,  kitchen,  laundry room, car, or garage) can damage your test strips and lead to inaccurate results. t Store test strips  in their  original container  only to avoid damage or  contamination. Do  not transfer test  strips to any other storage device, and do not store outside of their original container.  t Do not use test strips from any container that is damaged or left open to air. t Write the opening date on the container label when you rst open it. Discard remaining SD GlucoNavii® Mentor Test Strips after the discard date. (6 months after rst opening from the container.) t SD GlucoNavii® Mentor Test Strips are for single use only. Never  reuse  a  test  strip  that  has  had  either  blood  or control solution applied to it. nuGtGumj|GYWXZW[X]ZUGGGY] YWXZTW[TX]GGG㝘䟸G\a\_a\Z
27Understanding Your New SystemSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMt Avoid getting dirt, food or liquids on the test strip. With clean, dry hands, you may touch the test strip anywhere on its surface. t Do not bend, cut, or alter SD GlucoNavii® Mentor Test Strip in any way. Do not swallow test strips. The test strip container may contain drying agents that are harmful if inhaled or swallowed and may cause skin or eye irritation.CAUTIONnuGtGumj|GYWXZW[X]ZUGGGY^ YWXZTW[TX]GGG㝘䟸G\a\_a\Z
Control Solution Test28SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMCHAPTER 2 : Control Solution TestWhy you do control solution test ; t SD  Glucose  control  solution  is  used  to  check  that  the meter and the test strips are working together as a system and that you are performing the test correctly. t It is very important that you do this simple check routinely to make sure you get an accurate result. When you do control solution test ;t You open a new box test strips. t You left the test strip container open or you think your test strips have been damaged. t Your test strips were stored in extreme temperatures and/or humidity. t You want to check the meter and test strips. t You dropped the meter. t Your test result does not agree with how you feel. t You want to check if you are testing correctly. Before you begin ; t Use only SD Glucose control solution. t Check  the  expiration  date  on  the  control  solution container. Record the opening date on the container label. Do Not use after expiration or discard date (date opened plus three months), whichever comes rst. t Control solution, meter and test strips should be at room temperature 18-30°C (64-86°F) before testing with control solution. nuGtGumj|GYWXZW[X]ZUGGGY_ YWXZTW[TX]GGG㝘䟸G\a\_a\Z
Control Solution Test29SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMt Shake  the  container,  discard  the  rst  drop  of  control solution, and wipe o the tip to ensure a proper sample and an accurate result. t Store  SD  Glucose  control  solution  tightly  closed  at temperatures between 8-30°C (46-86°F). Do Not refrigerate. 1.  Performing  Control Solution TestYou need the meter, a test strip, and control solution Level M or Level H. The control level is printed on the test strip label.For  more  information  how  to  obtain  SD  Glucose  control solution, call at +82-31-300-0400.t Do not swallow SD Glucose control solution; it is not for human consumption. t Do not apply SD Glucose control solution to the skin or eyes as it may cause irritation. CAUTIONA set of Level M and H control solutions is available for purchase. To order control solutions, talk to your pharmacist or medical surgical supply dealers. Your meter is designed to recognize the dierence between Control Solution and blood. The meter automatically stores the test results using a control solution, letting your review them. But the meter does not include them in averages.NOTEnuGtGumj|GYWXZW[X]ZUGGGY` YWXZTW[TX]GGG㝘䟸G\a\_a\Z
Control Solution Test30SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMSTEP-1:1)  Remove  a  new  test  strip  from  container.  Be  sure  to tightly replace container cap after removing test strip. 2) Insert a test strip (yellow window printed  ‘Mentor’ facing up) into test strip slot. The meter turns on automatically.STEP-2:1)  Press the left button for 3 seconds to check the testing system using a control solution in Blood Stand-by Display. If you don’t want a control solution test, press the left button again.2)   Shake  the  control  solution container  and  discard  the rst  drop  of  solution.  Gently  squeeze  the  container to form one small drop. Bring the drop to the edge of the strip, and allow the strip to automatically draw the control solution into the yellow window. When control solution is  applied to the test strip, the  meter counts down from 5 to 1 second on the display. Tightly replace the cap on control solution. MentorNFC[ Left button ]MentornuGtGumj|GYWXZW[X]ZUGGGZW YWXZTW[TX]GGG㝘䟸G\a\_a\[
Control Solution Test31SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM3)  The control solution result appears on the screen in just 5 seconds. 4)  Compare control solution result with the range printed on the test strip container.   If the results are not within the control range printed on the test strip container, then the meter and strips may not be  working  properly. Repeat  the control solution test.5)  Remove  the  used  test  strip  for  control  solution  from the meter and discard it.MentorNFCNOTE The control solution range printed on the test strip container is for SD GlucoNavii® Control Solution only. It is not a recommended range for your blood glucose level. [This is an example. Refer to the ranges on your test strip container.]Control RangeLevel M Level H90-140 mg/dL 170-240 mg/dL5.0-7.8 mmol/L 9.4-13.3 mmol/Lmg/dLmmol/L[Example]nuGtGumj|GYWXZW[X]ZUGGGZX YWXZTW[TX]GGG㝘䟸G\a\_a\[
Control Solution Test32SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMCheck ActionDid you do the test in control solution mode? Did you see “control solution container icon” on the screen with the result? If not, do the test again. Insert a test strip; Press the left button for 3 seconds in Blood Stand-by Display to display a control solution container icon. Have the test strips and/or control solution expired? Make sure that test strips and control solutions are not past expiration date. This date is shown the container/bottle. Make sure the expiration date of both a test strip and a control solution.(test strip : 6 months, control solution : 3months)Were control solutions at room temperature (18-30°C, 64-86°F) when used? If not, retest with new bottle of control solution, or warm up/cool down to room temperature.(18-30°C, 64-86°F) 2. Troubleshooting Control Solution Test nuGtGumj|GYWXZW[X]ZUGGGZY YWXZTW[TX]GGG㝘䟸G\a\_a\[
Control Solution Test33SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMCheck ActionDid you insert test strip rmly into meter? Make sure test strip is inserted into the test strip slot until it will go on further.Did you follow the procedure correctly? Read again “Chapter 2 : Control Solution Test” (pages 28-33) and retest. Were test strips stored correctly? (at 2-32°C, 36-90°F) Was the bottle cap replaced immediately after removing a test strip? If not, retest with a new container of test strips. Is the meter damaged? Does it show an error message?If yes, contact SD Biosensor, Inc. at +82-31-300-0400.Is the control result outside the acceptable range(printed on your test strip container)?Repeat the test. If you get the same results, do not use your meter and test strip until you solve the problem. If you still have problem, call SD Biosensor, Inc. at +82-31-300-0400.nuGtGumj|GYWXZW[X]ZUGGGZZ YWXZTW[TX]GGG㝘䟸G\a\_a\[
34Testing Your Blood Sugar 34SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMCHAPTER 3 : Testing Your Blood Glucose1.  Getting a Drop of Blood 1)  Wash your  hands in  warm, soapy  water.  Rinse  well and dry completely. Warming ngers can increase blood ow. 2)  Turn the lancet insert cap counterclockwise to remove it, insert the lancet into the lancing device holder and push down rmly until it is fully seated. Twist the lancet protective disk until it separates from the lancet. 3)  Replace the cap  and turn it  clockwise, until it  is snug. Adjust  the  puncture  depth  setting  by  turning  the comfort dial. The dial has 1 to 5 steps, and the higher the  step  number,  the  stronger  the  blood  sampling pressure on the puncture site.   The  comport  tip  oers  5  dierent  levels  of  skin penetration. t 1-2 : for soft or thin skin t 3 : for average skin t 4-5 : for thick or callused skin121212nuGtGumj|GYWXZW[X]ZUGGGZ[ YWXZTW[TX]GGG㝘䟸G\a\_a\\
35Testing Your Blood Sugar 35SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM4)  After cocking the lancing device back, hold the lancing device rmly against the side of nger and then press the release button.1212t A lancet should only be used once. Do not share used lancets with another person. To prevent possible infection, a used lancet should not be touched by another person. t Used lancets in the regular trash can be dangerous. We recommend that you throw out the used lancets in sharps containers or test strip vials. Please be sure to save the cap so that the lancets cannot spill out of the container into the trash.CAUTIONnuGtGumj|GYWXZW[X]ZUGGGZ\ YWXZTW[TX]GGG㝘䟸G\a\_a\]
36Testing Your Blood Sugar 36SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM2. Performing a Blood Glucose Test1)  Remove a new test strip from container. Be sure to tightly replace container cap after removing test strip.2)  Make sure the meter turn o. And then, insert the test strip into the meter. 3)  When the  blood drop symbol ashes (Blood Stand-by Display), you are ready to perform a test. 4)  Let your arm hang down at your side to allow blood to ow to your ngertips. Grasp your nger just below the joint closest to the ngertip. 5)  Obtain  a  drop  of  blood  sample  using  the  lancet  and lancing device. 6)  Hold  your  nger  to  the  tip  of  the strip  until  the  yellow window is completely lled with blood. Do not place the blood drop on top of the strip. Always wash hands with warm, soapy water. Rinse and dry them before testing. NOTEMentorNFCnuGtGumj|GYWXZW[X]ZUGGGZ] YWXZTW[TX]GGG㝘䟸G\a\_a\]
37Testing Your Blood Sugar 37SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM7)  The  blood  will  be  drawn  into  the  strip  automatically.  If beeper is turned on, meter will beep to let you know the test is beginning.8)  When blood is applied to the strip, start the test.9)  You can set the pre-meal or post-meal test if you want. The symbol indicating pre-meal(       ) or post-meal(      ) appears on the screen. Choose what you want by pressing the left or right button. If you press the On/O button, no select any features.  10)  After applying the blood, the test result will appear for 5 seconds.11)  When  the  test  is  done,  pull  out  the  used  test  strip  and discard it. t Remove the inserted lancet from the lancing device and dispose the used lancet with care. t Always use fresh capillary or venous whole blood for tests.t If you want to use NFC function, refer to ‘Chapter 4. Using the NFC function’ on page  40~44 . NOTEnuGtGumj|GYWXZW[X]ZUGGGZ^ YWXZTW[TX]GGG㝘䟸G\a\_a\]
38Testing Your Blood Sugar 38SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM3. Understanding Test ResultsYour test results1.  After 5 seconds testing time from applying a blood into strip, you will receive a normal result, 10mg/dL to 600mg/dL.2.  If your blood glucose is above 600 mg/dL, you will receive a “HI” and is below 10 mg/dL, you will receive “Lo”. In these cases, repeat the test with new test strip. If this message show again, contact your healthcare professional immediately. If you perform the pre-meal test while the post-meal alarm setting is on, then the post-meal mark will appear automatically on your LCD when you test within following period: from 30min to 130min after your Pre-meal test. CAUTIONnuGtGumj|GYWXZW[X]ZUGGGZ_ YWXZTW[TX]GGG㝘䟸G\a\_a\]
39Testing Your Blood Sugar 39SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMNormal Blood Glucose Readings The normal fasting blood glucose range for an adult without diabetes is 74 - 106 mg/dL.2 Two hours after meals, the blood glucose range for an adult without diabetes is less than 140 mg/dL.1 t 'BTUJOHUPNHE-t IPVSTBGUFSNFBMTNHE-What This Means For You Frequent blood glucose testing is the best means to track how well you are doing with your diabetes management. It helps you track the eects of medications, diet, exercise, and stress management. Blood glucose test results can also tell you if your diabetes is changing. This may alert you to adjust your treatment plan. Always consult your healthcare professional before making any adjustments. Frequency of Testing Work with your healthcare professional to decide when and how often to test. This will depend on such things as age, type of diabetes, and medications. It is important to make testing part of your daily routine. nuGtGumj|GYWXZW[X]ZUGGGZ` YWXZTW[TX]GGG㝘䟸G\a\_a\]
4040Using NFC function 40SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMChapter 4: Using NFC function1. Information on NFC functionNFC(Near eld communication) is a set of standards for smartphones and  similar  devices  to  establish  radio  communication  with  each other  by  touching  them  together  or  bringing  them  into  close proximit.This  function  is  comfortable  for  test  results  back-up  at  PC  or smartphones.  So,  you  can  easily  monitor  your  blood  glucose results.SD  GlucoNavii®  Mentor  NFC    follows  “Felica”  standard  which include ISO/IEC 18092.2. Information on Caution for NFCt Check  your  smatphone  or  PC  OS  version.    For  PC,  It  is available for Windows XP, Windows  VISTA,  Windows 7 and need to NFC Reader/Writer purchased separately.t  For smartphone, It is available for only smartphones with Android OS v.2.3.1 or higher. Check the list in www.nfcworld. com/nfc-data/android/.t Before using this function, check the meter turn o. If turn the meter on, you cannot use NFC function.t Check the location of chip for NFC in your smartphone. It is dierent by manufacturer. For more information, refer to manufacturer website of your smartphone.t SD  NFC  software  for  PC  is  able  to  download  at  www.sdbiosensor.com . t If the distance between the meter and a smartphone(or NFC Reader/Writer) is  far(about over 2cm/0.06ft), NFC function is not operated.nuGtGumj|GYWXZW[X]ZUGGG[W YWXZTW[TX]GGG㝘䟸G\a\_a\]
4141Using NFC function41SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM3. Using NFC function1)  Check the location of NFC at your smartphone.2)  If  you  tap  SD  GlucoNavii®  Mentor  NFC    to  your smartphone,  two  kinds  of  action  are  happened.  First, start  GlucoNavii  App.,  seconds,  upload  the  glucose results stored in SD GlucoNavii® Mentor NFC meter to smartphone. 3)  After nishing the communication, check that the test results are transferred well in your smartphone or PC.More information on App or software for PC, refer to each manual. NOTEnuGtGumj|GYWXZW[X]ZUGGG[X YWXZTW[TX]GGG㝘䟸G\a\_a\^
4242Using NFC function 42SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM4. How to use GlucoNavii App1)  Download & Installt Search  ‘GlucoNavii’  in  Google  Play  at  your smartphone. You  can  search  other  keyword  that are ‘SD Biosensor’.t Install GlucoNavii App. in your smartphone.t Check the NFC function at your smartphone. If it is the ‘O’ setting, turn NFC function ‘On’.2)  Communicate with SD GlucoNavii® Mentor NFC meter t If  you  tap  SD  GlucoNavii®  Mentor  NFC  meter  to your  smart-phone,  two  kinds  of  action  will  be happened. First, if  App. is  not turning  it on, start App.  Seconds,  if  App.  is  already  running,  please, wait a minute to upload test results. 3)  Utilize GlucoNavii App       (1) Start App.       (2) If you need to upload new data, just tap SD GlucoNavii                   NFC meter to your smartphone.Please remember the points of contact for both devices.NOTENFC enabled smartphone with Android OS v.2.3.1 or higher. Check the list in http://www.nfcworld.com/nfc-data/android/.NOTEnuGtGumj|GYWXZW[X]ZUGGG[Y YWXZTW[TX]GGG㝘䟸G\a\_a\^
4343Using NFC function43SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM5. How to use GlucoNavii DMS for PC1)  Download  the  software  at  www.sdbiosensor.com.  And then, start the software.2)  Turn the meter o. If not, you cannot start NFC function.3)  Tap  the  meter  with  NFC  Reader/Writer  plugged  with  PC. And then, start the communication.4)  After  nishing  the  communication,  you  can  look  the  log book(test results) or graph.If  you  have  any  question  or  queries  about  GlucoNavii App.or GlucoNavii DMS for PC, please email to (SD_Dev@sdbiosensor.com). More information on App or software for PC, refer to each manual. NOTEEuropean Union Directive Conformance StatementHereby, SD Biosensor declares that SD GlucoNavii® Mentor NFC blood glucose meter  is in compliance with the essential requirements and other relevant provisions of Directive 1999/5/EC.  You can view your product’s Declaration of Conformity(DoC) to Derictives 1995/5/EC(R&TTE) Directive at www.sdbiosensor.com.NOTEWARNING: Changes or modications made to this equipment not expressly approved by SD Biosensor, Inc. for compliance could void the user’s authority to operate the equipment.CAUTIONnuGtGumj|GYWXZW[X]ZUGGG[Z YWXZTW[TX]GGG㝘䟸G\a\_a\^
4444Using NFC function 44SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMThis equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation.If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment o and on, the user is encouraged to try to correct the interference by one or more of the following measures:t Reorient or relocate the receiving antenna.t Increase the separation between the equipment and receiver.t Connect the equipment into an outlet on a circuit dierent.t from that to which the receiver is connected.t Consult the dealer or an experienced radio/TV technician for help.NOTEnuGtGumj|GYWXZW[X]ZUGGG[[ YWXZTW[TX]GGG㝘䟸G\a\_a\^
4545Using the Meter Memory45SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMCHAPTER 5 : Using the Meter MemoryThe meter automatically stores about 300 glucose results, letting your review them in order from the most recent to the oldest. If you have set the time/date feature, the time and date of the results are also displayed. If the memory is full and a new result is added, the meter deletes the oldest result. The meter also calculates three kinds of 7, 14 and 30-day averages of test results stored in memory, 1)normal, 2)pre-meal and 3)post-meal state averages. You do not need to set  the  time and  date  for the  meter  to  give  you average calculations.  HI/Lo  result  (results  outside  of  the  meter’s reading range) and results with control solution symbol are not included in averages. nuGtGumj|GYWXZW[X]ZUGGG[\ YWXZTW[TX]GGG㝘䟸G\a\_a\^
4646Using the Meter Memory46SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM1. Searching Test Results1)  In Strip Stand-by Display, press the left arrow button to review in sequence from the most recent test result to the last test results stored in memory.[Normal result][Pre-meal result][Post-meal result][ Left button ]MentornuGtGumj|GYWXZW[X]ZUGGG[] YWXZTW[TX]GGG㝘䟸G\a\_a\_
4747Using the Meter Memory47SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM[Control Solution result]2)  If there aren’t the stored test results, the following display appears for 1 second, and then the meter will display Strip Stand-by Display automatically. nuGtGumj|GYWXZW[X]ZUGGG[^ YWXZTW[TX]GGG㝘䟸G\a\_a\_
4848Using the Meter Memory48SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM3)  In Strip Stand-by Display, press the right arrow button to review three kinds of 7,14 and 30 day averages of test  result  stored  in  memory  in  sequence.  (normal, pre-meal and post-meal state) You can also review the number of results at each average in the right bottom of the LCD window. If you press the right arrow button once  more after  displaying the  30  day  average(with post-meal  mark),  the  7-day  average  result  appear again. [Post-meal average][ Normal average ][Pre-meal average]nuGtGumj|GYWXZW[X]ZUGGG[_ YWXZTW[TX]GGG㝘䟸G\a\_a\_
4949Using the Meter Memory49SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM4)  If there aren’t any stored 7, 14 and 30-day average of test results, following display will appear on the LCD.You cannot search the stored test results and average of results in the meter, if a test strip is inserted in the meter, Blood Stand-by Display. After removing the test strip from the meter, you can search the test results and average of results stored in memory by pressing the left or the right button.NOTEnuGtGumj|GYWXZW[X]ZUGGG[` YWXZTW[TX]GGG㝘䟸G\a\_a\_
5050Using the Meter Memory50SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM2. Downloading results to a computerYou can use your meter with GlucoNavii DMS to store your records and to help you spot patterns for planning meals, exercise, and medication. GlucoNavii DMS puts information downloaded from the meter into charts, diagrams and graphs.1)  Obtain the GlucoNavii DMS and NFC Reader/Writer.2)  Install the software on a personal computer. Please refer to Software Product Manual.t While the meter is connected to the PC, it is unable to perform a blood glucose test.t For downloading GlucoNavii DMS and Software Product manual(both are free of charge), please visit  www.sdbiosensor.com.t For ordering NFC Reader/Writer, please contact our representative.t For more information,please refer to Software Product Manual. If the cable port got ESD while downloading data, time delay, a few seconds, may be happened. After a while, the meter retransmit data automatically. NOTEnuGtGumj|GYWXZW[X]ZUGGG\W YWXZTW[TX]GGG㝘䟸G\a\_a\_
515151Maintenance and Troubleshooting51SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMCHAPTER 6 : Maintenance and Troubleshooting 1. Performing Check strip TestWhen you do the check strip test?t When you want to easily check the performance of the meter. t Before using your meter for the rst time. t Whenever your result does not agree with the level you feel.t If you have repeated a test and the blood glucose result is still lower or higher than expected. How to Use the check strip1.  Insert SD Glucose Check strip (facing up ‘Check strip’ printed in arming knob) into test strip slot. The meter turns on automatically. 2.  If the check strip is inserted properly, the meter will start the check.Check Strip test does not replace Control Solution test.CAUTIONMentorNFCnuGtGumj|GYWXZW[X]ZUGGG\X YWXZTW[TX]GGG㝘䟸G\a\_a\_
525252Maintenance and Troubleshooting52SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM3.  The check result appears on the screen in 5 seconds. ‘OK’ message appears on the screen if there isn’t any problem for the meter.  If not, ‘EEE’ message appears on the screen.2. Cleaning the meterCaring for SD GlucoNavii® Mentor NFC meter is easy. Just keep  it  free  of  dust.  If  you  need  to  clean  it,  follow  these guideline carefully to help  you get  the best performance possible. To  prevent  malfunction  of  the  meter,  keep  the  test  strip port  free  of  blood,  moisture,  dirt,  or  dust.  Use  a  lint-free cloth  dampened  with  water  to  clean  meter.  Thoroughly wring out  cloth before use. Do  not use  an abrasive cloth or  antiseptic  solution,  as  these  may  damage  the  display screen.[  ‘OK’ Message ][  ‘EEE’ Message ]nuGtGumj|GYWXZW[X]ZUGGG\Y YWXZTW[TX]GGG㝘䟸G\a\_a\`
535353Maintenance and Troubleshooting53SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM3.  Maintenance, Testing and TransportationThe meter needs little or no maintenance with normal use. It automatically tests its own systems every time you turn it on and lets you know if something is wrong. If you drop the meter or think it is not giving accurate results, make sure that your test strips  and control solution  haven’t expired, and then run a control solution test.Precautions for Maintenance, Testing and Transportation1)   Metert Keep the test strip slot free of dust.t Protect the internal meter from humidity. t The carrying case is designed to let you store a variety of supplies  you may need  and helps to  protect your meter. t If you keep the meter with the battery inserted, then keep it in a low humidity environment. t Do  not  modify  the  meter,  such  as  resembling  or  remodeling, NFC function is not operated well. If any function is not perated, please contact SD Biosensor, Inc.2)   Test stript The test strip is sensitive to humidity, keep it in a dry and  cool  environment,  and  do  not  store  in  direct sunlight.t After pulling out the test strip from its container, close a container cap of the test strip immediately.t The test strip container closes tightly and can protect the  test  strips,  so  you  should  keep  the  unused  test strips in the container in which they came.nuGtGumj|GYWXZW[X]ZUGGG\Z YWXZTW[TX]GGG㝘䟸G\a\_a\`
545454Maintenance and Troubleshooting54SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM3)   Lancet and Lancing devicet The needle of lancet is sharp, keep the lancet away from children. t Keep the lancet and lancing device dry and do not store in direct sunlight, or high heat and humidity locations.t A lancet should not use for the other intended use except sampling blood.t A lancet is for single use only. Do not reuse.t A  lancet  provided  with  SD  GlucoNavii®  Mentor  NFC  Blood glucose meter is able to use every lancing devices manufactured  by  other  company. But  if  at  all  possible, use  the  lancing  device  manufactured  by  SD  Biosensor, Inc.t Before using, check a packaging condition, if there is any problem, you should not use it.t If a lancet protective disk is loosed or needle of a lancet is exposed, you should  not use it.t To  reduce  the  chance  of  infection  for  the  used  lancet, discard it.4)   Control solutiont Keep SD Glucose control solution in 8-30°C (46-86°F) environment.t Do not refrigerate or freeze.t Do  not  use  SD  Glucose  control  solution  that  has passed the expiration date.t SD Glucose control solution can be used for 3 months after opening the container. Write the opened date on the container when you rst opened.t No reconstitution or dilution is necessary.t Wipe the container tip clean and reseal the container tightly after each use.4. Cleaning the Lancing DeviceClean  the  outside  of  SD  lancing  device  regularly  with 70%  isopropyl  (rubbing) alcohol.  Do not place  the entire device  under  water.  Do  not  use  bleach.  At  least  once  a week, disinfect the cap after cleaning by placing it in 70% rubbing  alcohol  for  10  minutes.  Allow  the  cap  to  air-dry after disinfecting.nuGtGumj|GYWXZW[X]ZUGGG\[ YWXZTW[TX]GGG㝘䟸G\a\_a\`
555555Maintenance and Troubleshooting55SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM5.  Screen Messages and TroubleshootingMessage Descriptiont The meter turns on normally. [Strip Stand-by Display]t The meter is ready for you to insert a test strip.[Blood Stand-by Display]t The meter is ready for a drop of blood.nuGtGumj|GYWXZW[X]ZUGGG\\ YWXZTW[TX]GGG㝘䟸G\a\_a\`
565656Maintenance and Troubleshooting56SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMt The meter shows the test result.nuGtGumj|GYWXZW[X]ZUGGG\] YWXZTW[TX]GGG㝘䟸G\a\_a\`
575757Maintenance and Troubleshooting57SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM[Normal Results Average][Pre-meal Results Average][Post-meal Results Average]t The meter shows three kinds of average  of results.nuGtGumj|GYWXZW[X]ZUGGG\^ YWXZTW[TX]GGG㝘䟸G\a\_a\`
585858Maintenance and Troubleshooting58SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMt The meter shows the saved results of blood glucose.t The meter is ready for a drop of control solution. nuGtGumj|GYWXZW[X]ZUGGG\_ YWXZTW[TX]GGG㝘䟸G\a\`aWW
595959Maintenance and Troubleshooting59SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM[Low battery]t At this time, battery is getting low but you can still perform about 50 tests. Replace the battery soon. See Chapter 1 “9. Changing the Battery”.[Replace battery]t Battery power is low. Replace the battery immediately. See Chapter 1 “9. Changing the Battery”. If you press the ON/OFF button after discharging of the battery, the battery icon will ash and then after ten seconds the meter will turn o automatically.[HI message]t Blood glucose may be higher than the measuring range of the system. See chapter 3 “4. Understanding Test Results”.nuGtGumj|GYWXZW[X]ZUGGG\` YWXZTW[TX]GGG㝘䟸G\a\`aWW
606060Maintenance and Troubleshooting60SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM[Lo message]t Blood glucose may be lower than the measuring range of the system. See chapter 3 “4. Understanding Test Results”.[Internal Error Message for a meter]t Turn o a meter. Then turn on the meter again. If there is still error message, please contact SD Biosensor, Inc. TEL: +82-31-300-0400.  [Strip Error]t Defective test strip or the test strip is damaged or inserted improperly. Discard this test strip and test again using new test strip. See Chapter 3 “2. Performing a Blood Glucose Test”. nuGtGumj|GYWXZW[X]ZUGGG]W YWXZTW[TX]GGG㝘䟸G\a\`aWW
616161Maintenance and Troubleshooting61SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM[Blood Sample Error]t An insucient amount of blood was applied. Discard this test strip and test again using new test strip and a larger sample, making sure blood is placed to the narrow channel in the top edge of the test strip. See Chapter 3 “2. Performing a Blood Glucose Test”.[Temperature Error]t If the environmental temperature is above or below the operating range of a meter, a thermometer icon will appear on the display. Move to an area between 8-45°C (46-113°F), wait for 30 minutes, and perform a test. Do not articially heat or cool the meter. See Chapter 6. “Product Technical Information”.[Communication Error]t The communication between meter and computer is failed. Connect again between meter and PC.nuGtGumj|GYWXZW[X]ZUGGG]X YWXZTW[TX]GGG㝘䟸G\a\`aWW
626262Maintenance and Troubleshooting62SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM6.    Warnings, Precautions and Limitationst Never  make  signicant  changes  to  your  diabetes control program or ignore physical symptom without consulting with your healthcare professional. t Severe dehydration  (excessive water  loss)  may cause false low results. If you believe you are suering from dehydration,  consult  your  healthcare  professional right away. t Extremes  in  hematocrit  may  aect  test  results. Hematocrit levels less than 20% may cause falsely high readings.  Hematocrit  levels  greater  than  60%  may cause falsely low readings.  t Inaccurate results may occur in severely hypotensive individuals  or  patients  in  shock.  Inaccurate  low results  may  occur  for  individuals  experiencing  a hyperglycemic-hyperosmolar  state,  with  or  without ketosis. Critically ill patients should not be tested with blood glucose meters. t Normal  endogenous  (within  body)  natural  levels of  uric  acid,  ascorbic  acid  (vitamin  C),  bilirubin, triglycerides,  and  hemoglobin  do  not  interfere  with your blood glucose results obtained.t Interferences  :  The  following  compounds,  elevated levels of ascorbic acid, uric acid, acetaminophen, total bilirubin, triglycerides may aect results.Compound LimitationAscorbic acid > 4 mg/dLUric acid> 9 mg/dLAcetaminophen> 6 mg/dLTotal bilirubin> 40 mg/dLTriglycerides> 1026 mg/dLnuGtGumj|GYWXZW[X]ZUGGG]Y YWXZTW[TX]GGG㝘䟸G\a\`aWW
636363Maintenance and Troubleshooting63SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMt Do  not  use  during  or  soon  after  xylose  absorption testing. Xylose in the blood will cause an interference.t SD GlucoNavii® Mentor NFC  System is not designed to be a substitute for pathology laboratory equipment and should not be used for the diagnosis of diabetes. t Do not use SD GlucoNavii® Mentor NFC  meter to test neonates. It has not been validated for neonatal use. nuGtGumj|GYWXZW[X]ZUGGG]Z YWXZTW[TX]GGG㝘䟸G\a\`aWW
646464Product Technical Information64SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMCHAPTER 7 : Product Technical Information1. System SpecicationsResult Range  10 - 600 mg/dL,(0.6 - 33.3 mmol/L) Calibration Plasma-equivalentSample  Fresh capillary whole blood Sample Size  0.3 microliter Test Time  5 seconds Assay Method  Glucose Oxidase Biosensor ON/OFF Source  Battery type CR2032 Battery Life  Around 1,000 tests Glucose Unit  mg/dL , mmol/LDisplay LCD (Customized)Controls3 ButtonsSize50mm ×93 mm ×18 mmWeight50g (with battery)Automatic Shuto t 1 minutes after last user action without inserting test strip into the meter t 3 minutes after last user action when inserting test strip into the meter nuGtGumj|GYWXZW[X]ZUGGG][ YWXZTW[TX]GGG㝘䟸G\a\`aWW
656565Product Technical Information65SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEM- System- Test strip OperationTemperature 10 - 45°C   (48 - 113°F)Hematocrit 20~60%AltitudeUp to 12,388 feet. (3,776 meters) StorageTemperature 2°C – 32°C (36°F – 90°F)Memory 300 blood glucose tests Function-  Hypo warning : 60, 70, 80 mg/dL(3.3, 3.9, 4.4 mmol/L)- Pre-meal and post-meal mark- Alarm setting (up to 4 times)- Post-meal Alarm-  7-, 14- and 30-day Averages of the following results  1)Normal Results  2)Pre-meal Results  3)Post-meal Results- Automatic shutonuGtGumj|GYWXZW[X]ZUGGG]\ YWXZTW[TX]GGG㝘䟸G\a\`aWW
666666Product Technical Information66SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMAnnex 1 : Information for Healthcare Professionals Decisions  about  whether  to  recommend  alternative  site testing (AST) should  take into account the  motivation and knowledge  level  of  the  patient  and  his  or  her  ability  to understand  the  considerations  relative  to  diabetes  and AST.  If  you  are  considering  recommending  AST  for  your patients, you  need  to  understand  that  there  is  a  potential for a signicant dierence between ngertip and alternative site  blood  glucose  test  results.  The  dierence  in  capillary bed  concentration  and  blood  perfusion  throughout  the body can lead to  sample site-to-site dierences in  glucose results. These physiological eects vary between individuals and  can  vary  within  a  single  individual  based  upon  his  or her  behavior  and  relative  physical  condition.  Our  studies involving  AST  of  adults  with  diabetes  show  that  most persons will nd their glucose level changes more quickly in the ngers’ blood than the alternative sites’ blood. This is especially important when glucose levels are falling or rising rapidly. If your patient is used to making treatment decisions based  upon  ngertip  readings, he  or  she should consider  the  delay  or  lag-time,  aecting  the  reading obtained from an alternative site.Healthcare professionals  : Follow the  infection  control procedures appropriate for your facility.nuGtGumj|GYWXZW[X]ZUGGG]] YWXZTW[TX]GGG㝘䟸G\a\`aWW
676767Product Technical Information67SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMAnnex 2 : Symbol The  following  list  describes  all  symbols  used  on  SD GlucoNavii® Mentor NFC   Blood glucose monitoring  (BGM) system.Symbol DescriptionCaution, consult accompanyingdocumentsBattery type used in this meterCrossed out wheeled bin: To discard it separately from other household wasteConsult instructions for useIn-vitro diagnostic medical device:This system is intended to use outside the body (in vitro diagnostic use).Serial number for this meter.Date of manufacture :To indicate the date of manufacture for this meternuGtGumj|GYWXZW[X]ZUGGG]^ YWXZTW[TX]GGG㝘䟸G\a\`aWW
686868Product Technical Information68SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMSymbol DescriptionLot No. or Batch CodeTo indicate the lot number for thissystemExpiry date:This system should be used by thegiven date.To indicate the temperaturelimitations in which the transportpackage has to be kept and handledCatalogue number:To indicate the catalogue number for this systemTo indicate the manufacturer$POUBJOT4VóDJFOUGPSO5FTUTTo use for 6 months after rst opening the container.DO NOT REUSE:To warn the user of a piece ofequipment that it is for single useonly and that it must not therefore be used more than once6nuGtGumj|GYWXZW[X]ZUGGG]_ YWXZTW[TX]GGG㝘䟸G\a\`aWX
696969Product Technical Information69SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMAnnex 3 : References1.  American  Diabetes  Association,  Clinical  Practice Recommendation Guidelines  2003,  Diabetes care, Vol. 26. Supplement 1. p.22 2.  Stedman,  TL.  Stedman’s  Medical  Dictionary,  27th Edition, 1999, p. 2082. 3.  Ellen T. Chen, James H. Nichols, Show-Hong Duh, Glen Hortin,  MD:  Diabetes  Technology  &  Therapeutics, Performance  Evaluation  of  Blood  Glucose  Monitoring Devices, Oct 2003, Vol. 5, No. 5 : 749 -768nuGtGumj|GYWXZW[X]ZUGGG]` YWXZTW[TX]GGG㝘䟸G\a\`aWX
707070Product Technical Information70SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMReturn You must contact SD Biosensor Customer Service at +82-31-300-0400  before  returning  your  meter.  You  will  be instructed how to return the meter  to SD  Biosensor, Inc. Returned  meters  without  this  authorization  will  not  be accepted.This  device  complies  with  Part  15  of  the  FCC  Rules. Operation is subject to the  following two conditions: (1) this device may not cause harmful interference, and (2) this device  must  accept  any  interference  received,  including interference that may cause undesired operation.nuGtGumj|GYWXZW[X]ZUGGG^W YWXZTW[TX]GGG㝘䟸G\a\`aWX
717171Product Technical Information71SD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMManufactured by                       SD BIOSENSOR, Inc. C-4th&5th Floor Digital Empire Building 980-3, Yeongtong-dong, Yeongtong-gu, Suwon-si, Kyonggi-do, Korea EC-Representative                           nuGtGumj|GYWXZW[X]ZUGGG^X YWXZTW[TX]GGG㝘䟸G\a\`aWX
727272Product Technical Information72MEMOSD GlucoNavii® Mentor NFC BLOOD GLUCOSE MONITORING SYSTEMnuGtGumj|GYWXZW[X]ZUGGG^Y YWXZTW[TX]GGG㝘䟸G\a\`aWX
 FCC ID-RPJ01GM2401680123L21MVN1ENR0Issue date: 2013.0401GM24/01GC24 series