INTRODUCTION SECTION 1  1 Manual Purpose  The WAM (Wireless Patient Cable) user manual explains how to: · Acquire and transmit 12-lead ECG signals to a receiving device · Acquire and transmit 12-lead Rhythm strip to a receiving device · Power the WAM On and Off · LED Messages /  Buzzer Beep Tone o Power On/Off o RF communication link status o Low Battery Warning o Lead check  Audience This manual is written for clinical professionals who are expected to have a working knowledge of medical procedures and terminology as required for monitoring and/or acquiring clinical data from cardiac patients.  Indications for Use The WAM cable is a non-invasive prescription device: · Indicated for use as a radiofrequency physiological signal transmitter that acquires and delivers RF transmission of real-time electrocardiographic data obtained during resting/physiologic electrocardiographic testing. · Indicated for use in a clinical setting, by qualified medical professionals, properly trained for acquiring ECG data and use of the system. The personnel must be experienced in cardiovascular problematic situations and emergency procedures or pathologies related to cardiac involvements.  It is not intended as a sole means of diagnosis. · Indicated for use in hospital and physicians office, patients of any age where ECG acquisition is done (e.g., patients in Coronary Care Units, Step-Down Units, Emergency Departments, clinics, physicians offices, sports medicine facilities or rehabilitation departments). · It is not designed for use in highly invasive environments, such as an operating theatre.  The cardiac data and analysis provided is reviewed, confirmed, and used by trained medical personnel in the diagnosis of patients with various cardiac rhythm patterns.  System Description  The WAM incorporates wireless electrocardiographic technology to achieve the real-time acquisition and RF transmission of simultaneous 12-lead ECG data with diagnostic quality.  Transmission to a Mortara receiver module allows the cardiac signals to be displayed on a monitoring device such as an electrocardiograph without the need for a direct connection to an electrocardiograph.    The WAM uses a single AA battery.
 SECTION 1  2 The following equipment is necessary to use the WAM: · One AA battery, 1.5V · Mortara receiver module in the electrocardiograph · ECG cable  WAM with ECG Cable   Figure 1-1, Front View   ECG Patient Cable    Power On/Off         LED Indicator         12 Lead ECG Acquisition Rhythm Print  Battery Compartment
 SECTION 1   3 Part Numbers  Description  Part Numbers Wireless Patient Cable  WAM-XXX-XXXXX WAM Battery Cap Assembly  8356-008-51 WAM Carrying Pouch with Belt   8485-025-50 COMBINER REPLACEMENT WAM LEADS 5 POSITION      Plastic part “combines” 5 leads into a groups for plugging into the WAM 9293-046-01  LEAD SET WAM 10 WIRE BANANA AHA       Full set of 10 wires, grouped into two sets of 5, with “combiners” 9293-046-50 LEAD SET WAM 10 WIRE BANANA IEC       ll set of 10 wires, grouped into two sets of 5, with “combiners” 9293-046-51  REPLACEMENT LEAD SET WAM LIMBS BANANA AHA      Indicated leads, no combiners  9293-046-52 REPLACEMENT LEAD SET WAM LIMBS BANANA IEC      Indicated leads, no combiners  9293-046-53  REPLACEMENT LEAD SET WAM V1-V3 BANANA AHA      Indicated leads, no combiners  9293-046-54 REPLACEMENT LEAD SET WAM C1-C3 BANANA IEC      Indicated leads, no combiners  9293-046-55  REPLACEMENT LEAD SET WAM V4-V6 BANANA AHA      Indicated leads, no combiners  9293-046-56  REPLACEMENT LEAD SET WAM C4-C6 BANANA IEC      Indicated leads, no combiners  9293-046-57 LEAD SET WAM 10 WIRE CLIPS AHA      Full set of 10 wires, grouped into two sets of 5, with “combiners” 9293-047-50  LEAD SET WAM 10 WIRE CLIPS IEC      Full set of 10 wires, grouped into two sets of 5, with “combiners” 9293-047-51 REPLACEMENT LEAD SET WAM LIMBS CLIPS AHA      Indicated leads, no combiners  9293-047-52  REPLACEMENT LEAD SET WAM LIMBS CLIPS IEC      Indicated leads, no combiners  9293-047-53  REPLACEMENT LEAD SET WAM V1-V3 CLIPS AHA      Indicated leads, no combiners  9293-047-54  REPLACEMENT LEAD SET WAM C1-C3 CLIPS IEC      Indicated leads, no combiners  9293-047-55  REPLACEMENT LEAD SET WAM V4-V6 CLIPS AHA      Indicated leads, no combiners  9293-047-56  REPLACEMENT LEAD SET WAM C4-C6 CLIPS IEC      Indicated leads, no combiners  9293-047-57  WAM User Manual (XXX=language)  9515-XXX-XX-XXX WAM Short-Form Instruction Card (XXX=language)  9503-XXX-XX-XXX Case of 24 Monitoring Hookup Kits: ·  10 monitoring electrodes ·  1 abrasive pad ·  1 gauze wipe ·  1 razor ·  2 alcohol prep pads 9294-009-51
 SECTION 1  4   Approved Battery Models  Description Manufacturer  Part Numbers Alkaline, AA-type, 1.5V  Various  Various Lithium, AA-type, 1.5V  Energizer  L91  NOTE:  Replace cells with only those listed above.  Use of other cells may present a risk of fire or explosion.  To order additional supplies, contact a Mortara Instrument customer service representative at:  Mortara Instrument, Inc. 7865 N. 86th Street Milwaukee, WI  53224 Tel:  1.888.MORTARA (667.8272) Fax:  414.354.4760 Internet:  http://www.mortara.com  Specifications  Feature  Specifications Instrument Type  12-lead ECG digital transmitter cable Input Channels  Continuous 12-lead signal acquisition and transmission ECG Leads Transmitted  I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5 and V6 Special Functions  Power On/Off, Lead impedance check, during hookup, lead fail, and battery notification RF Protocol  2400-2483 MHz interference immune proprietary frequency hopping scheme Water Resistance  Electronics protected from water damage; meets EN 60529 rating IPX1  Defibrillator Protection  Complies with AAMI standards and IEC 60601-2-25 Function Keys  On/Off, 12-Lead ECG acquisition, Rhythm acquisition Device Classification  Type CF (ECG), battery operated Weight  6.7 oz. (190 g) with battery Dimensions  4.45” x 4.25” x 1.1” (11.3 cm x 10.8 cm x 2.79 cm) Battery  1 AA alkaline, 24-hour typical life  NOTE:  Operating range performance is 3 meters or 10 feet with.

Navigation menu