Fukuda Denshi Co DS7100 Patient Monitor User Manual DS71v1 1 FA 10 Maintenance 012

Fukuda Denshi Co Ltd Patient Monitor DS71v1 1 FA 10 Maintenance 012

Manual 9

Download: Fukuda Denshi Co DS7100 Patient Monitor User Manual DS71v1 1 FA 10 Maintenance 012
Mirror Download [FCC.gov]Fukuda Denshi Co DS7100 Patient Monitor User Manual DS71v1 1 FA 10 Maintenance 012
Document ID380411
Application IDwOVnFJBjOncRtHtUy8lSFQ==
Document DescriptionManual 9
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize127.96kB (1599460 bits)
Date Submitted2003-12-10 00:00:00
Date Available2003-12-10 00:00:00
Creation Date2003-12-06 12:02:47
Producing SoftwareAcrobat PDFWriter 4.0 Windows NT
Document Lastmod2003-12-08 09:45:12
Document TitleDS71v1-1_FA_10_Maintenance_012.PDF
Document CreatorMicrosoft Word - DS71v1-1_FA_10_Maintenance_012.doc
Document Author: ezashi

Chapter 10
Maintenance
●Daily Check ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・7
●Periodic Check ・・・・・・・・・・・・・・・・・・・・・・・・・・・・7
Time-Change Components ・・・・・・・・・・・・・・・・・・・・ 7
− Troubleshooting − ・・・・・・・・・・・・・・・・・・・・・・・・・・・ 9
ECG ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 9
Respiration ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 11
Invasive Blood Pressure・・・・・・・・・・・・・・・・・・・・・・ 11
SpO2 ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 12
Non-Invasive Blood Pressure ・・・・・・・・・・・・・・・・・ 13
Temperature・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 14
CO2 Concentration・・・・・・・・・・・・・・・・・・・・・・・・・・・ 15
Recorder ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 16
Telemetry ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 16
General ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 17
Battery ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 18
10−1
10
Maintenance
After Use / Display Panel ・・・・・ 2
− Handling −
Handling After Use ・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2
Handling the Display Panel ・・・・・・・・・・・・・・・・・・・・ 2
Device / Recording Paper ・・・・・・・・ 3
− Storage −
Storing the Device ・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3
Storing the Recording Paper ・・・・・・・・・・・・・・・・・・・ 3
Display Panel and Housing ・・・・・・ 4
− Cleaning −
Cleaning the Display Panel ・・・・・・・・・・・・・・・・・・・・ 4
Cleaning the Housing ・・・・・・・・・・・・・・・・・・・・・・・・・ 5
Disinfecting the Blood Pressure Transducers ・・・・・ 5
Disinfecting the Temperature Probes ・・・・・・・・・・・・ 5
Disinfecting the CO2 Filter Line ・・・・・・・・・・・・・・・・・ 5
Cleaning and Disinfecting the SpO2 Transducer・・・ 5
Cleaning and Disinfecting the NIBP Cuff ・・・・・・・・・ 5
Cleaning and Disinfecting the ECG lead Cable ・・・ 5
Handling the Battery Pack・・・・・ 6
− Battery −
Handling the Battery ・・・・・・・・・・・・・・・・・・・・・・・・・・ 6
Storing the Battery・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 6
− Maintenance Check − Daily and Periodic Check ・ 7
About the Maintenance Check ・・・・・・・・・・・・・・・・・ 7
− Handling −
After Use / Display Panel
This section describes precautions for handling the equipment.
Handling After Use
?Do not apply excessive force when disconnecting the cables. Always pull on the connector
housing and not on the cable.
?Clean the unit, accessories, and cables, and keep them together in one place for next use.
?Always check for adequate supply of disposable accessories such as ECG electrodes. If any
shortage, contact our service representative and supply as necessary.
Handling the Display Panel
?The display panel utilizes exclusive fluorescent light for the backlight.
As this fluorescent light tube has product life cycle, it needs to be replaced periodically. If the
display becomes dark, scintillates, or does not light, contact your nearest service representative.
?The LCD used for the display panel utilizes highly accurate picture elements of pixels over
99.99%, but there may be an absence (less than 0.01%) or constant lighting of pixels.
10−2
− Storage −
Device / Recording Paper
This section describes about the storage of the device and recording paper.
Storing the Device
?Store in a place where the device will not be exposed to splashing water.
?Store in a place where the device will not be adversely affected by atmospheric pressure,
temperature, humidity, ventilation, sunlight, dust or atmosphere containing salt or sulfur.
?Store in a level area where the device is not exposed to vibration and shock (including during
transportation).
?The following environmental conditions should be observed when storing the device.
Storage Temperature : −10∼60?C
Storage Humidity
: 10∼95% (at 60?C)
Storing the Recording Paper
The DS-7100 system utilizes heat sensitive recording paper. If placed in a high temperature for
long period of time, the print may become indistinct, and unable to read. When storing, follow the
precautions below.
?Store in a place where light is shut off and avoid direct sunlight.
?Do not leave the paper in a high temperature (50 °C or 122 °F or above).
?Do not store the paper in polyvinyl chloride bag.
?Do not expose the paper to alcohol, hydrochloric acid, or ester ketone.
?Avoid using adhesive agents other than water based glue.
10
Storage
10−3
− Cleaning −
Display Panel and Housing
This chapter explains about the cleaning of the device and sensors.
Cleaning the Display Panel
Since this device incorporates a touch screen, finger prints and other stains are likely to appear on
the display panel.
Follow the procedure below to clean the display panel.
1. Press the
Reference
Key Lock
key for more than 2 seconds.
The Key Lock key needs to be preprogrammed as user key.
Refer to ” 8. System Configuration Ward Setup” for user key setup.
2. Clean the touch panel.
While the “Touch key OFF" message is displayed, the touch panel key will be deactivated.
If “LEAD OFF” or other message is displayed, the key lock message will not be displayed.
3. Wipe the touch panel using cleaning cloth.
4. Press again the Key Lock key for more than 2 seconds.
The message will disappear and the keys will be activate again.
? Do not clean the touch panel using strong acid.
? A special coating is applied to the surface of the touch panel. Wipe the
CAUTION
surface with the soft cleaning cloth provided as optional accessory or
with commercially available eyeglass cleaning cloth.
10−4
Cleaning the Housing
Clean the housing using tightly squeezed gauze or an absorbent cotton cloth dampened with
alcohol or a neutral cleanser.
? Clean the equipment frequently so stains can be removed easily.
? To prevent injury, it is recommended to wear gloves when cleaning the
equipment.
? Do not allow liquids or cleaning solution to enter the monitor or
connectors.
? Do not use organic solvents, thinner, toluene and benzene to avoid
damaging the resin case.
? Do not polish the housing with abrasive or chemical cleaner.
C A U T I O N ? When sterilizing the entire room using a spray solution, pay close
attention not to have liquids get into the monitor or connectors.
? Use only neutral detergent to clean the housing. Do not use chemical
cloth, scrub brush, abrasive, polishing powder, hot water, volatile
solvent and chemicals (cleanser, thinner, benzine, benzol, and
synthetic detergent for house and furniture), or sharp-edged tools.
The surface resin coating may be damaged, resulting in discoloration,
scratches, and other problems.
Disinfecting the Blood Pressure Transducers
Disinfect the blood pressure transducers according to the manufacturer’s guidelines.
Disinfecting the Temperature Probes
10
Disinfect the Temperature Probes according to the manufacturer’s guidelines.
Cleaning
Disinfecting the CO2 Filter Line
Disinfect the Filter Line/Capno Line according to the manufacturer’s guidelines.
Cleaning and Disinfecting the SpO2 Transducer
?Do not soak the sensor in water or antiseptic solution.
?Wipe the DURASENSOR with disinfectant such as 70% alcohol. Do not disinfect by applying
radioactive rays, steam, or ethylene oxide.
?OXISENSOR is a disposable sensor. Do not reuse or attempt resterilization.
Cleaning and Disinfecting the NIBP Cuff
?Do not soak the sensor in water or antiseptic solution.
?Wipe the NIBP cuff with disinfectant such as 70% alcohol. Do not disinfect by applying
radioactive rays, steam, or ethylene oxide.
Cleaning and Disinfecting the ECG lead Cable
?Do not soak the sensor in water or antiseptic solution.
?Wipe the ECG lead cable with disinfectant such as 70% alcohol. Do not disinfect by applying
radioactive rays, steam, or ethylene oxide.
10−5
− Battery −
Handling the Battery Pack
This section describes about the handling and storage of the battery pack.
Handling the Battery
?The battery pack can be continually used for more than 300 times (or about 1 year) under normal
temperature, but the continuous use will degrade the battery and shortens the usable time.
?When the battery operation time becomes short even after it is fully charged, the battery pack
needs to be replaced.
?When the charge time of the battery pack becomes short, the battery pack needs to be replaced.
?When the battery pack level becomes low, charge the battery well in advance for the next use.
Storing the Battery
To take advantage of the characteristic of battery pack, pay attention to the following when storing.
Storage Temperature and Humidity
?Store in an environment specified below without corrosive gas.
Storage Period
Storage
Temperature
Storage
Humidity
Within 30days
30 days∼90 days
90 days∼1year
−20∼60?C
−20∼45?C
−20∼35?C
65? 20%
?Do not store in an environment outside the specified temperature range or excessive high
humidity. This may result in leakage caused by expansion/contraction inside the battery pack,
or rusting of the metal part.
Long-term Storage
?If left installed in the monitor for long period of time, the electrolyte may leak, or inactivate the
battery which degrades the capacity recovery after storage. Therefore, always remove the
battery from the monitor when storing for long period of time. Contact our service
representative when removing the battery.
10−6
− Maintenance Check −
Daily and Periodic Check
This section explains the daily check and periodic check items of the device.
About the Maintenance Check
Periodic inspection must be performed. When reusing the device which was left unused for a
while, always check that the device operates properly and safely before use.
To ensure safety, reliability, and high performance, a “Daily Check” and “Periodic Inspection” must
be performed. We are not liable for any accident arising from lack of maintenance.
? Do not open the housing of this device.
C A U T I O N ? Avoid alcohol or other liquids from getting into the equipment.
●Daily Check
Perform daily inspection using the “Daily Check List” on the next page.
●Periodic Check
Periodic inspection of medical electronic equipment is mandatory to prevent failures and accidents
and to ensure safety and reliability.
Periodic maintenance may be performed by each medical institution or by a third party by
concluding a “Maintenance Contract”.
For more details, contact your nearest service representative.
10
Time-Change Components
EtCO2 Unit
Replacing Period: 7,000 hours (approx. 10 months of
continuous use)
LCD Unit, Inverter Unit
Replacing Period: 50,000 hours or 6 years
NIBP Unit
Replacing Period: 100,000 times of use or 6 years
Short Term Backup Battery
Replacing Period: 4 years∼6 years according to the used
frequency
Long Term Backup Battery
Replacing Period: 6 years
Battery Pack
Replacing Period: 1 year or 300 times of charging /
discharging.
C A U T I O N The time-change components must be replaced at specified period.
10−7
Maintenance Check
To ensure reliability of safety, function, and performance of this device, the time-change
components must be replaced periodically. When replacing, contact our service representative.
Daily Check List
No.
Inspected Date
Inspected by
Location
Device Type
Serial No.
Date of Purchase
Item
Appearance
Installation
Functions
Details
Criteria
Judgement
Visually check the exterior for
scratches, cracks, deformation,
and rust.
No abnormality should be
found.
□OK / □NG
Check whether the unit is
installed on a level surface.
The installation area must be
level and free from vibration
and shock.
□OK / □NG
The environmental condition
(ex. temperature, humidity) of
Check whether the unit is
the installed place should be
installed in a place susceptible to
as specified.
adverse environment.
The unit should not be
subjected to splashing water.
□OK / □NG
The home display appears,
and the lamp located at the
right side of the display panel
lights.
□OK / □NG
The date and time should be
correct.
□OK / □NG
Turn ON the monitor, and check
whether it operates normally.
Cables
Visually check all cables for any
damage.
No damage should be found.
□OK / □NG
Periodic Inspection
Check the date of previous
periodic inspection.
Should be within 1 year.
□OK / □NG
Should be within 6 months.
□OK / □NG
CO2 Calibration
(DS-7141)
Comment
10−8
Check the date of previous
calibration date.
Previous Date
Day
Year
Month
− Troubleshooting −
This section explains the troubleshooting for each case.
ECG
The “LEAD OFF” message is displayed.
Cause : The electrode is detached, or is not making good electrical contact with the skin.
Solution : ・Check if the electrodes are properly attached.
・Replace the electrode, or check the lead cable.
The “ECG failed” message is displayed.
Cause 1 : The ECG amplitude is 0.25mV or below for the waveform size of ×1, ×1/2, ×1/4,
and 0.150mV or below for the waveform size of ×2, ×4.
Solution : Change the electrode attachment site, or select the lead with higher QRS amplitude.
Note
: Using 4-electrode or 5-electrode instead of 3-electrode allows more accurate QRS
detection.
Cause 2 : The electrode contact is poor.
Electrical blanket or other noise source is near the patient.
Solution : Attach the electrode firmly.
・ Replace the lead cable if defective.
・ If any noise source is near the patient, locate it away from the patient as much as
possible.
Cause 2 : EMG is interfering.
Solution : ・Change the electrode site to a location where EMG will less likely to interfere.
・Select ESIS mode for the filter mode.
Note
: Selecting a ESIS mode for the filter mode will decrease the QRS amplitude and may
result in not counting the heart rate.
The “Check electrode” message is displayed.
Cause : The electrode contact with the skin is poor. There is substantial contact resistance
between the electrodes.
Solution : Replace all the electrodes.
Use the electrodes of the same type.
The “ECG unit error” message is displayed.
Cause : A communication error with the ECG measuring unit exists.
Solution : The breakage of wire or failure of the ECG unit can be considered.
Contact our service representative.
10−9
10
Troubleshooting
ECG waveform contains noise.
The “Artifact” message is displayed.
Cause 1 : The electrode contact is poor.
Electrical blanket or other noise source is near the patient.
Solution : Attach the electrodes firmly.
・ Replace the lead cable if defective.
・ If any noise source is near the patient, locate it away from the patient as much as
possible.
The measured data is displayed as “×××”.
Cause : The heart rate is outside the measurement range.
Solution : ・Check the electrode application.
・Replace the electrode, or check the lead cable.
Heart rate is not counted. Heart rate is low.
Cause : The ECG waveform amplitude is below the QRS detection level (0.3mV).
Solution : Change the electrode site, or select a lead with higher QRS amplitude.
Note
: Using 4-electrode or 5-electrode instead of 3-electrode allows more accurate QRS
detection.
Also, if large amount of noise is interfering, the noise may be erroneously detected as
QRS. It is recommended to change the electrode site and increase the ECG
amplitude.
Heart rate is not counted, and “LEAD OFF” message is displayed.
Cause : The electrode of the displayed lead type is detached, or is not making good electrical
contact with the skin.
Solution : ・Check the electrode application.
・Replace the electrode, or check the lead cable.
Artificial pacemaker is not displayed.
Cause : On the admit / discharge menu, Not used
Solution : Select Used for the pacemaker use.
is selected for the pacemaker use.
The “Pacemaker error” message is displayed.
Cause
: The pacemaker pulse is detected 16 pulses or more per second.
Solution 1 : Attach the electrodes firmly.
・ Replace the lead cable if defective.
・ If any noise source is near the patient, locate it away from the patient as much as
possible.
Solution 2 : If the patient is not wearing a pacemaker, set to Not used for the pacemaker use
in the patient admit/discharge menu.
The “ECG not connected” message is displayed.
Cause
: When the ECG relay cable is disconnected during ECG monitoring, this message will
be displayed.
Solution 1 : To cease monitoring, press the Alarm Silence key to clear the message and
silence the alarm.
Solution 2 : To continue monitoring, plug in the ECG relay cable. This will clear the message and
silence the alarm.
10−10
Respiration
The “CVA detected” message is displayed.
Cause : Heartbeat is interfering and superimposed on the respiration waveform.
Solution : Place the electrode as shown below where the heartbeat will be less likely to interfere.
Red
White
Black
“0” is displayed for respiration rate, or apnea alarm is generated.
Cause : The respiration waveform amplitude is below the detection level (0.2O).
Solution : Change the electrode site.
The respiration waveform and respiration rate is not displayed.
Cause 1 : The ECG relay cable designed for electrosurgical knife is used.
Solution : The impedance respiration can not be measured if the cable designed for
electrosurgical knife is used. Use the standard ECG relay cable if not using the
electrosurgical knife.
The measured data is displayed as “×××”.
Cause : The respiration rate is outside the measurement range.
Solution : ・Check the electrode application.
・Replace the electrode, or check the lead cable.
Invasive Blood Pressure
The “BP1 Transducer OFF”, “BP2 Transducer OFF” message is displayed.
Cause : The transducer for BP1 or BP2 is not connected.
Solution : Connect the transducer.
The “BP1 not zero balanced”, “BP2 not zero balanced” message is displayed.
Cause : The BP zero balance has not been performed since the power is turned ON.
Solution : Open the three-way cock of the transducer to air and perform zero balance.
The measured data is displayed as “ー ー ー”.
Cause : The BP zero balance has not been performed since the power is turned ON.
Solution : Open the three-way cock of the transducer to air and perform zero balance.
10−11
10
Troubleshooting
Cause 2 : The impedance respiration measurement is ceased.
Solution : Turn ON the impedance respiration measurement on the admit / discharge menu or
RESP configuration menu.
Note
: If the pacemaker with the minute ventilation measuring function is used, turn OFF the
impedance respiration measurement. Otherwise, both the pacemaker and the monitor
will not be able to perform accurate measurement.
BP value and waveform are not displayed properly.
Cause : Blood pressure line has not been zero balanced.
Solution : Open the three-way cock of the transducer to air and perform zero balance.
The measured data is displayed as “×××”.
Cause : The BP value is outside the measurement range.
Solution : Perform zero balance again.
The “BP not connected” message is displayed.
Cause
: When the BP interface cable or 2ch BP conversion cable is disconnected during BP
monitoring, this message will be displayed.
Solution 1 : To cease monitoring, press the Alarm Silence key to clear the message and
silence the alarm.
Solution 2 : To continue monitoring, plug in the BP interface cable or 2ch BP conversion cable.
This will clear the message and silence the alarm.
The “Incorr. BP cable” message is displayed.
Cause : The cable other than 2ch BP conversion cable is plugged in to the BP connector.
Solution : Use the 2ch BP conversion cable.
SpO2
The “Check SpO2 sensor” message is displayed.
Cause
: Sensor is detached from the patient.
Solution 1 : Check if the sensor part is properly attached to the patient.
Solution 2 : Check if the light emitting part and light receiving part of the sensor LED is aligned.
The “Pulse search” message is displayed.
Cause : The amplitude of the pulse waveform is low, or the sensor is not positioned correctly.
Solution : Check if the light emitting part and light receiving part of the sensor LED is aligned.
The “No pulse detect” message is displayed.
Cause : The amplitude of the pulse waveform is low, or the sensor is not positioned correctly.
Solution : Check if the light emitting part and light receiving part of the sensor LED is aligned.
The “Motion Artifact” message is displayed.
Cause : There is excessive body motion of the patient.
Solution : Change the sensor position where the body motion will have less effect.
The pulse waveform is not displayed, or interrupted
Situation : “Check SpO2 sensor” is displayed.
Cause 1 : The amplitude of the pulse waveform is low, or the sensor is not positioned correctly.
Solution : Check if the light emitting part and light receiving part of the sensor LED is aligned.
Cause 2 : Sensor is defective.
Solution : Replace the sensor.
Cause 3 : SpO2 sensor is not firmly connected to the SpO2 input connector.
Solution : Make sure the SpO2 sensor is securely connected.
Cause 4 : Sensor is exposed to light.
Solution : Place a black or dark cloth over the sensor to avoid direct sunlight. Also when not
used, avoid placing the sensor in light or unplug the sensor from the connector.
10−12
The SpO2 measurement is unstable.
Cause : There is excessive body motion of the patient which disables correct measurement.
Solution : 1. Have the patient lie still as much as possible.
2. Relocate the sensor, or change the sensor to which the body motion will have less
influence.
The “SpO2 unit error” message is displayed.
Cause 1 : There is a failure of communication with the SpO2 measurement unit.
Solution : Breaking of wire or SpO2 unit failure can be considered.
Contact our service representative.
Cause 2 : Sensor is defective.
Solution : Replace the sensor.
The “SpO2 sensor fault” message is displayed.
Cause : Sensor is defective.
Solution : Replace the sensor.
The “SpO2 not connected” message is displayed.
Cause
: When the SpO2 relay cable is disconnected during SpO2 monitoring, this message will
be displayed.
Solution 1 : To cease monitoring, press the Alarm Silence key to clear the message and
silence the alarm.
Solution 2 : To continue monitoring, plug in the SpO2 relay cable. This will clear the message
and silence the alarm.
Non-Invasive Blood Pressure
10
Troubleshooting
The cuff is not inflated although the pump is operating.
Cause1 : The air hose is not firmly connected, and the air is leaking.
Solution : Check if the air hose is properly connected.
Cause 2 : The cuff size is not corresponded to the selected patient type.
Solution : Check if the cuff size is corresponded to the selected patient type.
The monitor repeats the measurement, or “ー ー ー” is displayed for the numeric data.
Cause 1 : The measurement accuracy is not reliable due to body motion artifact.
Solution : Have the patient stay still as much as possible during the measurement.
Cause 2 : The pulse is too small to acquire reliable measurement accuracy.
Solution : Check if the cuff application is proper, and if the cuff size is corresponded to the
selected patient type.
The “Check NIBP hose” message is displayed.
Cause : The applied pressure to the cuff has exceeded the maximum limit. The measurement
time has exceeded the maximum limit.
Solution : Check if the cuff application is proper, if the cuff size is corresponded to the selected
patient type, or if the air hose is not bent. After checking the above, perform the
measurement again.
If the same message is displayed again, a failure of the equipment can be considered.
Cease the measurement, and contact our service representative.
10−13
The “NIBP unit error” message is displayed.
Cause : The zero balancing before the measurement has failed, and measurement could not be
started.
Solution : The body movement or other artifact may cause zero balance failure. During the
measurement, have the patient stay still as much as possible.
If the same message is displayed again, the failure of the equipment can be considered.
Cease the measurement, and contact our service representative.
The time of measurement disappears and the numeric data is displayed as “ー ー ー”.
Cause : The NIBP data will be erased when the preprogrammed NIBP erase time has elapsed.
Solution : Select the appropriate time for NIBP data erase time from 10min, 30min, 60min, 24hrs
which best fits the monitoring purpose.
Temperature
The “Wrong Temp Probe” message is displayed.
Cause 1 : The YSI-700 is used.
Solution : Use the YSI-400 temperature probe for measurement. The YSI-700 can not be used
with the DS-7100 series.
Cause 2 : There is a contact failure of the temperature probe.
Solution : Check if the temperature probe is properly inserted.
The numeric data is displayed as “×××”.
Cause : The temperature measurement is outside the measurement range.
Solution : Check if the temperature probe is properly inserted.
The “TEMP not connected” message is displayed.
Cause
: When the temperature sensor is disconnected during temperature monitoring, this
message will be displayed.
Solution 1 : To cease monitoring, press the Alarm Silence key to clear the message and
silence the alarm.
Solution 2 : To continue monitoring, plug in the temperature sensor. This will clear the message
and silence the alarm.
The “TEMP auto check” message is displayed. The numeric data is displayed as “ー ー ー”.
Cause : The temperature is calibrated once every hour on this monitor. During calibration, the
numeric data will be displayed as “ー ー ー”.
Solution : The calibration will complete in 10 seconds. If the calibration does not complete within
10 seconds, cease the measurement and contact our service representative.
The “TEMP unit check” message is displayed.
Cause : Error is detected during temperature calibration.
Solution : A unit failure can be considered. Cease the measurement and contact our service
representative.
10−14
CO2 Concentration
The “Check filter line” message is displayed.
Cause : The sampling tube is clogged.
Solution : Replace the sampling tube.
The “Self-diag CO2” message remains displayed.
Cause : An error has occurred to the self-check procedure at power ON.
Solution : The CO2 unit failure can be considered.
The “Initializing CO2” message does not disappear.
Cause : An error has occurred during the initialization at power ON.
Solution : The CO2 unit failure can be considered.
The “Check CO2 unit” message is displayed.
Cause 1 : The exhaust connector is clogged.
Solution : After checking the exhaust system and removing the clog, press the “Restart CO2” key
on the CO2 configuration menu.
Cause 2 : The sampling tube or nasal prong is clogged.
Solution : After checking the inhalation system and removing the clog, press the “Restart CO2”
key on the CO2 configuration menu.
Cause 3 : The CO2 unit needs to be replaced.
Solution : Contact our service representative.
The “CO2 unit error” message is displayed.
Cause : There is a communication error with the CO2 unit.
Solution : The break of wire or CO2 unit failure can be considered.
Contact our service representative.
10
Cause 2 : The calibration is not properly performed.
Solution : Perform CO2 calibration again.
The “CO2 not connected” message is displayed.
Cause
: When the filter line is disconnected during CO2 monitoring, this message will be
displayed.
Solution 1 : To cease monitoring, press the Alarm Silence key to clear the message and
silence the alarm.
Solution 2 : To continue monitoring, plug in the filter line. This will clear the message and silence
the alarm.
10−15
Troubleshooting
There is substantial measurement error.
Cause 1 : 20 minutes have not yet elapsed since the power is turned ON.
Solution : For 20 minutes from turning ON the power, there will be a substantial measurement
error.
Recorder
No recording is performed.
Situation : The “Paper Out” message is displayed on the upper left of the screen.
The “Paper Out” message is displayed on the Rec. START/STOP
key.
Cause : There is no recording paper in the recorder magazine.
Solution : Install a new pad of paper into the paper magazine.
Situation : The “Magazine Open” message is displayed.
Cause : The paper magazine is open.
Solution : Close the magazine.
Situation : The “Paper jammed” message is displayed.
Cause : The paper is jammed.
Solution : Open the magazine and install the paper correctly.
Situation : No message is displayed, but recording can not be performed.
Cause : The recording paper is not correctly installed. The front and backside of the paper is
set oppositely.
Solution : The “END” printed side of the paper should be facing down in the magazine.
The second waveform and third waveform are not recorded.
Situation : The second waveform and third waveform are not recorded for manual recording or
alarm recording.
Cause : The second waveform and third waveform are not set on the recording setup menu.
Solution : Set the second waveform and the third waveform on each recording setup menu.
The “Recorder error” message is displayed.
Cause : The thermal head temperature has increased.
Solution : A damage to the thermal head can be considered.
Contact our service representative.
Telemetry
The “Telemetry unit error” message is displayed.
Cause : There is a communication error with the telemetry transmission unit.
Solution : The breaking of wire or telemetry transmission unit failure can be considered.
Contact our service representative.
There is no reception at the telemetry center.
Cause : The channel ID or group ID is not corresponded with the telemetry receiver.
Solution : Set the correct channel ID and group ID.
The BP waveform of 100mmHg or above can not be properly received.
Cause : The BP waveform and scale is not corresponded.
Solution : When BP waveform is above 100mmHg, set the BP scale above 100mmHg.
10−16
General
Nothing is displayed but the main power indicator is lighted.
Cause : A system error has occurred.
Solution : Turn off the power, unplug the power cable, and contact our service representative.
The “Adjusting” message is displayed. Numbers are displayed large on the display.
Cause : This is the test mode. Stop using the device immediately.
Solution : Restart the system. The test mode will be cancelled.
If the same situation is observed again, contact our service representative.
Turn off the DIP switch No.1.
The data is initialized each time the power is turned ON.
Cause 1 : The internal switch is set to initialize.
Solution : The internal switch setting needs to be changed. Contact our service representative.
Set the rotary switch to 0.
Cause 2 : The battery for backup memory is depleted.
Solution : The battery needs to be replaced. Contact our service representative.
The display is not clear.
Cause 1 : The display brightness is not adjusted.
Solution : Due to the LCD display characteristic, the visible range is limited. Adjust to the
appropriate brightness.
Cause 2 : The monitor is set to the night mode.
Solution : Cancel the night mode.
Cause 2 : Incorrect IC card is inserted.
Solution : Turn off the DIP switch No.8.
The clock is often delayed.
Cause : The battery for the backup memory is depleted. Check if the time is delayed when the
power is turned off.
Solution : The battery needs to be replaced. Contact our service representative.
10−17
10
Troubleshooting
The system does not start although the power switch is turned ON.
Cause 1 : The power cable is not connected.
The battery is not charged.
Solution : Turn off the power and connect the power cable. If the battery is not charged, use the
power cable until the battery charging is complete.
Battery
The operation time is short although the battery is charged.
Cause 1 : The battery life has expired.
Solution : The battery pack is a consumable product. Replace it once a year.
Cause 2 : The ambient temperature is too high or too low.
Solution : For safety, the charging operation will be in a standby mode when the battery pack
temperature becomes excessively high or low.
The charging will automatically resume when appropriate temperature is reached.
Charge the battery in an ambient temperature of 10∼30?C.
The charge lamp on the patient monitor does not light.
Cause 1 : The AC power cable is disconnected.
Solution : Plug in the AC power cable.
The battery pack can be charged only during the AC operation.
Cause 2 : The battery pack is not installed.
Solution : The battery pack is optional. If a battery pack is required, contact our service
representative and install the battery pack.
Cause 3 : The battery life has expired.
Solution : The battery pack is a consumable product. Replace it once a year.
During the charging procedure, the charge lamp (orange) does not switch to charge
complete status (green) and extinguishes.
Cause 1 : The battery pack temperature is too high or too low.
Solution : For safety, the charging operation will be in a standby mode when the battery pack
temperature becomes excessively high or low.
The charging will automatically resume when appropriate temperature is reached.
Cause 2 : The breakdown of battery pack can be considered.
Solution : If the charging operation does not complete within the specified charging time, the
charging operation will cease for safety purpose.
Contact our service representative and replace the battery pack.
Cause 3 : The battery life has expired.
Solution : The battery pack is a consumable product. Replace it once a year.
The “Charge battery” message is displayed.
Cause : The AC power cable is disconnected.
Solution : Plug in the AC power cable.
The battery pack can be charged only during the AC operation.
10−18
Chapter 11
Technical Information
− Specification/Performance − ・・・・・・・・・・・・・・・・・・ 2
Specification ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2
Performance・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3
Default and Backup・・・・・・・・・・・ 5
− Setup Item −
Patient Admit / Discharge・・・・・・・・・・・・・・・・・・・・・・ 5
Alarm Setup ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 5
Parameter Setup ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 6
Review Function Setup・・・・・・・・・・・・・・・・・・・・・・・・ 7
System Configuration Setup ・・・・・・・・・・・・・・・・・・・ 8
Display Mode Setup ・・・・・・・・・・・・・・・・・・・・・・・・・ 12
Alarm Mode Setup・・・・・・・・・・・・・・・・・・・・・・・・・・・ 13
Pin Assignments・・・ 14
− External Connection −
Serial Connector Output Signal・・・・・・・・・・・・・・・・ 14
Status I/O Signal ・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 14
11
Technical Information
11−1
− Specification/Performance −
This section states the specification and performance of this equipment.
Specification
Size
260 (W) ×197 (D) ×264 (H) mm (not including the protrusion)
Weight (not including the battery)
DS-7141
: 5.3kg
DS-7101L
: 5.1kg
DS-7101LT
: 5.2kg
Environmental Condition
Operating Temperature
Operating Humidity
Transport / Storage Temperature :
Transport / Storage Temperature :
Safety
General Standard
10∼40?C
30∼85% (without condensation)
−10∼60?C
10∼95% (at 60?C)
: EN60601-1:1990
(Medical electrical equipment – Part 1: General requirements for
safety)
Amendment A1 to EN 60601-1:1993
Amendment A2 to EN 60601-1:1995
EMC Standard
: IEC 60601-1-2:1993
(Medical electrical equipment−Part 1: General requirements for safety
−2.Collateral standard:
Electromagnetic compatibility −requirements and
tests. IEC 601-1-2:1993 )
Withstand Voltage
: Class I, Type CF
Power Requirements
Voltage
Frequency
Power
Consumption
11−2
AC100∼240V ? 10%
DC14V ? 10%
50/60Hz
82VA
60W
Performance
Display
Device
Control
Waveform Trace
Waveform Speed
Parameter
8.4 inch TFT Color LCD
Touch Screen Type
Stationary Trace
ECG / SpO2 / BP (6.25mm/s, 12.5mm/s, 25mm/s)
RESP / CO2 (6.25mm/s, 12.5mm/s, 25mm/ s)
: ECG, RESP, TEMP, SpO2 (Arterial Oxygen Saturation), BP1, BP2,
NIBP, CO2 concentration
Recording
Recording Method
Recording Speed
Recording Waveform
Recording Width
Operation
Touch Screen
: Eight-Wire Resistive Analog Touch Screen
ECG
Lead Type
Frequency Characteristic
Input Impedance
Max. Input Voltage
Polarization Voltage
Common Mode
Rejection Ratio
HR Measurement Range
Thermal Array Type
25mm/s
Max. 3 waveforms
50mm
Wired 3-electrode, 4-electrode, 5-electrode
40Hz / 15Hz
5M? or above
±10mV
±825mV or above
: 80 dB or above
: Adult
0, 12∼300bpm±3% or ±5bpm
Neonate 0, 30∼300bpm±3% or ±5bpm
Waveform Size Selection : 1/4, 1/2, 1, 2, 4
Defibrillation Proof
: Provided
Temperature
Method
Probe
Measurement Range
No. of Channel
: Thermistor Method
: only YSI-400 series
: 0∼50?C±0.2?C
: 1 channel
Specification / Performance
Respiration
Method
Frequency Characteristic
Current
Measurement Range
11
Impedance Method
1.5Hz (adult, child) / 2.5Hz (neonate)
100μA or lower
0, 4∼150Bpm±5Bpm
SpO2 (Arterial Oxygen Saturation)
Method
: 2 Wavelength Pulse Wave Method
Measurement Range
: 0%∼100%
Resolution
: 1%
Accuracy
: Adult
at 70∼100% ±2%
at 0∼69% not specified
Neonate at 70∼100% ±3%
at 0∼69% not specified
Measurement Range
: 20∼250bpm±3bpm
11−3
Blood Pressure
Transducer Sensitivity
Measurement Range
Frequency Characteristic
Accuracy
Zero Balance Range
Measurement Range
Channel
5μV / V / mmHg
−50∼300mmHg
DC∼6Hz / 8Hz / 12Hz / 40Hz
±2% of full scale or within ±1mmHg
within ±150mmHg
Adult
20∼300bpm±3%
Neonate 30∼300bpm±3%
: 2 channels
Non-Invasive Blood Pressure
Method
: Oscillometric Method
Measurement Range
: 10∼280mmHg
Resolution
: 1mmHg
Accuracy
: ±4mmHg
Measurement Range
: 40∼240bpm±5%
CO2 Concentration (DS-7141)
Method
: Infra-Red Solid-State Method, Microstream Method
Measurement Range
: 0∼99mmHg
0-20min
0∼38mmHg:±4mmHg
39∼99mmHg:±12%
20min. and up
0∼38mmHg:±2mmHg
39∼99mmHg:±5%
RR Measurement Range : 0∼150bpm
0∼40bpm:±1bpm
41∼70bpm:±2bpm
71∼100bpm:±3%
101∼150bpm:±5%
Flow Rate
: 50±7.5 ml/min
System Response Time : 2.9 seconds (Typical)
Delay Time
: 2.7 seconds (Typical)
Rise Time
: 190 msec (maximum)
Telemetry (DS-7141, DS-7101LT)
Transmission Freq.
: 608~614 MHz
Exact frequency depends on the destination.
RF Output Power
: -15dBm Standard, 0dBm MAX
Channel Spacing
: 12.5 kHz
Modulation Mode
: Digital, Frequency Shift Keying (FSK)
11−4
− Setup Item −
Default and Backup
This section lists selection, default setting, and backup status for each setup item.
Backup Item
“○” : Setup item will be retained even when the power is turned OFF.
“△” : Setup item will be retained even when the power is turned OFF. When discharging
procedure is performed, the value will be reset to initial setting.
The alarm setup will be reset to initial setting with the selected alarm mode.
“−” : Setup item will be reset to initial setting when the power is turned OFF.
Patient Admit / Discharge
Item
Patient Name
Sex
Age
Birth Date
ID
Patient Type
Pacemaker
Impedance Measurement
Filter Mode
Bed ID
Room/Bed ID
Room ID
Selection
Numeric, Alphabet, Symbol (16
characters)
Male, Female
0∼150 years or 0∼999 days
Birth Date (Year, Month, Day)
Numeric, Alphabet, Symbol (10
characters)
Adult, Child, Neonate
Used, Not used
ON, OFF
Monitor, ST Display, ESIS
0∼999
Numeric, Alphabet, Symbol
(4 characters)
Default
Backup
Blank
△
Undetermined
0 year
Blank
△
△
△
Blank
△
Adult
Not used
ON
Monitor
○
△
△
○
○
BED−
○
Alarm Setup
ST
Selection
Suspend, ON
ON, OFF 20−300bpm
ON, OFF 3−10 sec.
ON, OFF
ON, OFF
ON, OFF
ON, OFF 2−8 beats
ON, OFF
ON, OFF 1.5−5 sec.
ON, OFF
ON, OFF
ON, OFF 1−50 beats / min.
ON, OFF 20−300
ON, OFF 20−300
ON, OFF
ST1 ±2.0mV / ±20mm
ST2 ±2.0mV / ±20mm
BP1 (mmHg)
ON, OFF
0−300mmHg
BP1 (kPa)
ON, OFF
0−40.0kPa
Default
Suspend
ON 40−120
ON 5 sec.
ON
ON
ON
ON 3 beats
OFF
OFF 2 sec.
OFF
OFF
OFF, 10 beats
ON
ON
Backup
11
Setup Item
Item
System Alarm
HR
AYSTOLE
VF
VT
SLOW_VT
RUN
COUPLET
PAUSE
BIGEMINY
TRIGEMINY
FREQUENT
TACHY
BRADY
○/△
OFF
ON
SYS
DIA
MEAN
ON
SYS
DIA
MEAN
80−180
OFF−OFF
OFF−OFF
10.0−24.0
OFF−OFF
OFF−OFF
11−5
Item
Selection
BP2(mmHg)
ON, OFF
0−300mmHg
BP2(kPa)
ON, OFF
0−40.0kPa
RR
APNEA
SpO2
ON, OFF
ON, OFF
ON, OFF
5−150Bpm
5−20sec.
50−100%
NIBP (mmHg)
ON, OFF
10−300mmHg
NIBP (kPa)
ON, OFF
1.5−40.0kPa
TEMP
EtCO2 (mmHg)
EtCO2 (kPa)
EtCO2 (%)
InspCO2 (mmHg)
InspCO2 (kPa)
InspCO2 (%)
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
30−50?C
1−100mmHg
0.1−13.3kPa
0.1−13.3%
1−4mmHg
0.1−0.4kPa
0.1−0.4%
Default
ON
SYS
OFF−OFF
DIA
OFF−OFF
MEAN OFF−OFF
ON
SYS
OFF−OFF
DIA
OFF−OFF
MEAN OFF−OFF
ON 5−30
ON 15sec.
ON 90−OFF
ON
SYS
80−180
DIA
OFF−OFF
MEAN OFF−OFF
ON
SYS
10.0−24.0
DIA
OFF−OFF
MEAN OFF−OFF
OFF OFF−OFF
ON 30−45mmHg
ON 4.0−6.0kPa
ON 4.0−6.0%
ON 3mmHg
ON 0.4kPa
ON 0.4%
Backup
○/△
The alarm setup will be retained even after the power is turned OFF.
If discharging procedure is performed, the alarm setup will be initialized with the
selected alarm mode.
NOTE
Parameter Setup
Item
ECG
RESP
11−6
Selection
Ⅰ, Ⅱ, Ⅲ, aVR, aVL, aVF,
Default
ECG1 Lead Ⅱ
Lead
ECG2 Lead V
ECG1 ×1
Waveform Size
×1/4, ×1/2, ×1, ×2, ×4
ECG2 ×1
Filter Selection
Monitor, ST Display, ESIS
Monitor
HR Source
Auto, ECG, SpO2, BP
Auto
Automatic Lead Switch ON, OFF
OFF
Pacemaker Pulse
ON, OFF
ON
HR Average
ON, OFF, Instant, Average
Average
HR Sync. Indicator
ON, OFF
ON
Auto, 10ms, 20ms, 40ms,
Pace Pulse Mask Time
Auto
OFF
ECG Drift Filter
ON, OFF
OFF
Waveform Size
×1/4, ×1/2, ×1, ×2, ×4 ×1
RR Sync. Indicator
ON, OFF
ON
CVA
ON, OFF
OFF
RR Source
Auto, Impedance, CO2
Auto
Impedance Meas.
ON, OFF
ON
Backup
○/△
△
○
○
○
○
○
○
○
○
△
○
△
○
○
SpO2
NIBP
Item
Waveform Size
SpO2 SEC Alarm
Ignore NIBP
Auto Mode
End Tone
Quick SYS List
PR
Quick SYS
1-Min. Auto
Mean
Selection
×1/4, ×1/2, ×1, ×2, ×4
OFF, 10, 25, 50, 100
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
3, 5, 10 min.
10, 20 min.
ON, OFF
20, 50, 75, 100, 150, 200,
250, 300mmHg
4, 8, 12, 16, 20, 24, 32,
40kPa
6, 8, 12, 40Hz
Default
×1
OFF
ON
OFF
ON
ON
OFF
10 min.
20 min.
OFF
12Hz
○
OFF
○
Filter
ON, OFF
20, 50, 75, 100, 150, 200,
250, 300mmHg
4, 8, 12, 16, 20, 24, 32,
40kPa
6, 8, 12, 40Hz
12Hz
○
Mean
ON, OFF
OFF
○
Meas. Unit
mmHg, kPa, %
mmHg
EtCO2 Average
10, 20, 30 sec., OFF
50, 100mmHg
4, 8, 10kPa
4, 8, 10%
10 sec.
50mmHg
4kPa
4%
Scale
BP1
Filter
Mean
Scale
BP2
CO2
Scale
Backup
△
○
○
△
○
○
○
○
○
○
150mmHg
○/△
20kPa
50mmHg
○/△
8kPa
○
○/△
○/△
11
Review Function Setup
Parameter
Duration
Graphic
Trend
Scale
Selection
HR, PR, VPC, ST1/ST2, RR, APNEA,
SpO2, BP, NIBP, TEMP,
EtCO2 / InspCO2, EVENT1, EVENT2
1, 2, 4, 8, 12, 24 hours
HR
: 100, 200, 300bpm
ST
: ±0.2, ±0.5, ±1.0,
±2.0mV
±2, ±5, ±10, ±20mm
VPC
: 20, 50, 100 beats
BP1
: 20, 50, 100, 150, 200,
300mmHg
4, 8, 16, 20, 24, 40kPa
BP2
: 20, 50, 100, 150, 200,
300mmHg
4, 8, 16, 20, 24, 40kPa
NIBP
: 100, 150, 200, 300mmHg
16, 20, 24, 40kPa
TEMP : 20-45, 30-40?C
SpO2
: 0-100, 50-100, 80-100%
RR
: 100, 200, 300bpm
APNEA : 15, 30 sec.
CO2
: 50, 100mmHg
4, 8, 10kPa
4, 8, 10%
Default
Backup
HR
○
4 hours
○
HR
ST
: 100bpm
: ±0.5mV
±5mm
VPC
: 20 beats
BP1
: 150mmHg
20kPa
BP2
: 50mmHg
8kPa
NIBP
: 150mmHg
20kPa
TEMP : 30-40?C
SpO2
: 80-100%
RR
: 50bpm
APNEA : 15 sec.
CO2
: 50mmHg
4.0kPa
4.0%
○
11−7
Setup Item
Item
Item
Tabular
Trend
OCRG
Selection
Backup
Duration
1, 5, 10, 15, 30, 60 min.
60 min.
○
Display Time
4, 8 min.
8 min.
○
Waveform
Impedance Resp., CO2
Impedance Resp.
○
Parameter
ECG1, ECG2
HR/PR/BPR :
ST
RR
APNEA
SpO2
BP1
BP2
NIBP
TEMP
CO2
ASYSTOLE :
VF
VT
SLOW_VT :
RUN
COUPLET :
PAUSE
BIGEMINY :
TRIGEMINY :
FREQUENT :
TACHY
BRADY
120ms
○
Meas. Point
ECG1, ECG2, BP, SpO2, RESP, CO2
HR/PR/BPR : ON, OFF
ST
: ON, OFF
RR
: ON, OFF
APNEA
: ON, OFF
SpO2
: ON, OFF
BP1
: ON, OFF
BP2
: ON, OFF
NIBP
: ON, OFF
TEMP
: ON, OFF
CO2
: ON, OFF
ASYSTOLE : ON, OFF
VF
: ON, OFF
VT
: ON, OFF
SLOW_VT : ON, OFF
RUN
: ON, OFF
COUPLET : ON, OFF
PAUSE
: ON, OFF
BIGEMINY : ON, OFF
TRIGEMINY : ON, OFF
FREQUENT : ON, OFF
TACHY
: ON, OFF
BRADY
: ON, OFF
0∼560ms
Ref. Point
0∼−240ms
−80ms
Alarm Factor
Recall
Arrhythmia
Factor
ST
Meas.
Default
NOTE
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
○
○
○
○
? The graphic trend data, tabular trend data will be retained even after the
power is turned OFF.
? The ST data, OCRG data, recall data will be retained until 5 minutes after the
power is turned OFF.
System Configuration Setup
●Volume Setup
Volume
11−8
Item
Pulse
Selection
16 levels
Default
Level 8 from left
Backup
○
Key
16 levels
Level 10 from left
○
Alarm
16 levels
Level 10 from left
○
Other Bed
16 levels
Level 10 from left
○
Others
16 levels
Level 8 from left
○
●Display Configuration
Item
No. of waveforms
No. of numeric
data
Displayed
Display
waveform
Configuration
Displayed numeric
data
Enlarged Display
Short Trend
NOTE
Selection
0∼6 waveforms
Default
3 waveforms
0∼7 numeric data
4 numeric data
ECG1, ECG2, BP, SpO2
RESP, CO2
HR, BP, NIBP, SpO2
RESP, RR, CO2, VPC/ST
ON, OFF
ON, OFF, Overlap
ECG1, SpO2, RESP
Backup
○/△
HR, NIBP, SpO2, RR
OFF
OFF
By selecting ON for backup at discharge, the setup item will be stored
even after discharge.
Selecting OFF will initialize the display configuration to the initial setting of
the selected display mode.
●System Configuration Menu
Item
ECG1, BP, RESP
○
Rec. Duration
Selection
ECG1, ECG2, BP, SpO2
RESP, CO2
24 sec., Continuous
24 sec.
○
Delay Time
None, 8sec., 16 sec.
8 sec.
○
ON, OFF, Center
ECG1, ECG2, BP, SpO2
Wave Select
RESP, CO2, Alarm Factor
Rec. Duration
12, 24 sec.
HR (HR / PR / BPR)
Alarm Factor
Numeric Data, Arrhythmia
ASYSTOLE, VF, VT,
SLOW_VT, RUN,
COUPLET, PAUSE,
Arrhythmia Record
BIGEMINY, TRIGEMINY,
FREQUENT, TACHY,
BRADY
Periodic Record
ON, OFF, Center
ECG1, ECG2, BP, SpO2,
Wave Select
RESP, CO2
Rec. Duration
6, 12, 24 sec.
OFF
○
ECG1, Alarm Factor
○
12 sec.
HR (HR / PR / BPR)
Arrhythmia
○
Periodic Interval
Wave Select
Manual
Recording
Alarm Record
Alarm
Recording
Rec.
Operation
Sweep
Speed
Night Mode
Backup
○
11
ASYSTOLE, VF, VT,
SLOW_VT, RUN,
TACHY, BRADY
○
OFF
○
ECG1, BP, RESP
○
12 sec.
○
60 min.
○
Paper Feed to Top ON, OFF
OFF
○
Paper Feed to End ON, OFF
ON
○
5, 10, 15, 30, 60 min.
QRS Classification ON, OFF
ON
○
ECG, BP, SpO2
25, 12.5, 6.25mm/s
25mm/s
○
RESP, CO2
25, 12.5, 6.25mm/s
6.25mm/s
○
Mode
Manual, Auto
Manual
○
Start Time
00:00∼23:59
22:00
○
Complete Time
00:00∼23:59
Time Disp. Only, Slightly
Dark, Dark
No change, Quiet, Very
quiet, Silence
ON, OFF
7:00
○
Dark
○
Very quiet
○
OFF
○
Display
Volume
Alarm Pole
Setup Item
Periodic
Recording
Default
11−9
Item
Selection
Default
Backup
ECG
Green
○
ST
Green
○
VPC
White
○
PACE
White
○
BP1
Red
○
Cyan
○
Cyan
○
SpO2
Yellow
○
TEMP
Orange
○
RESP
White
○
CO2
ST2 (Trend)
Alarm Suspend
1, 3, 5 min.
Cyan
Orange
3 min.
Alarm Silence
1, 3, 5 min.
3 min.
Item
AC filter
Selection
50, 60Hz
Default
According to Dip_SW
Date
07/19,
Jul.19
○
Alarm Mute
ON, OFF
OFF
○
Home Key Function
Home / Enlarge, Home
Home
○
Night Mode Cancel
Any Key, Night Mode Key
Any Key
○
Asystole,VF,VT
ON,ON/OFF
ON
○
BP
mmHg, kPa
mmHg
○
TEMP
?C, ?F
mm, mV
0801~0879, 0900~0979,
1000~1079, 1100~1179,
1200~1279, 1300~1379
00∼63
?C
mV
○
1100
○
00
○
BP2
Color
NIBP
Alarm Setup
32 colors
○
Fixed
○
○
●Hospital Setup
Unit
ST
Channel
Telemeter
Group
Jul.19,
19.Jul
Backup
○
○
●Ward Setup
Item
Trend Clip
Selection
ON, OFF
ON
BP Record Scale
20, 40mm
40mm
○
HR Low Limit for VT
120, 140bpm
120bpm
○
Password
ON, OFF
OFF
○
Discharge Mode
Admit, Cease
Admit
○
Event Key
ON, OFF
ON
○
Mean Calculation
Waveform, Calculation
All Key
(excluding system config.)
All Key
(excluding pre-set)
All Key
(excluding ward setup)
Waveform
○
All Key
○
All Key
○
All Key
○
Menu
Key Mask
System Config.
Pre-Set Menu
11−10
Default
Backup
○
Item
User Key
Selection
Alarm Pole
Alarm Level
Alarm Pole
Selection
Rec. START/STOP,
Alarm Silence, Alarm, Key
Lock, NIBP List, Graphic
Trend, Tabular Trend,
Recall, OCRG, Freeze,
NIBP Auto Mode,
Size/Lead, HR Source, BP
Zero, Admit/Discharge,
Night Mode, Display
Config., Record,
Tone/Volume, Other Bed,
ST Display, Cease,
Enlarged Display
ON, OFF
Level1, Level 1 and 2,
Level 1, 2, and 3
Default
Backup
Alarm Silence, Rec.
START/STOP,
Size/Lead,
Admit/Discharge
(from left of display)
○
ON
○
Level1
○
Level 1:Pattern 1
Pattern Setup
Pattern 1∼10
Level 2:Pattern 10
○
Level 3:Pattern 4
NIBP Data Erase Time
10, 30, 60min, 24hourr
○
60min
●Monitor Setup
Selection
Normal, Power Save
Normal
Message Icon
ON, OFF
OFF
○
Parameter Key Frame
ON, OFF
ON
○
Wide AC Filter
ON, OFF
OFF
○
Check Discharge at Power ON
ON, OFF
ON
○
Backup at Discharge
ON, OFF
OFF
○
HR, RR
Positive Logic, Negative
Logic
OFF, APNEA, Level 1,
Level 1 and 2,
Level 1, 2 and 3
Positive Logic, Negative
Logic, Pulse
HR
○
Positive Logic
○
Level 1
○
Positive Logic
○
Sync. Sygnal
Output
Status
Output Setup
Alarm Output
Default
Backup
○
11
Setup Item
Item
Battery Operation
11−11
Display Mode Setup
Item
Mode Selection
Mode 1
Mode 2
Mode 3
Mode 4
Mode 5
11−12
Default
Backup
○
No. of Waveforms
3 Waveform
○
No. of Numeric Data
4 Numeric Data
○
Displayed Waveforms
ECG1, SpO2, RESP
○
Displayed Numeric Data
HR, NIBP, SpO2, RR
○
Enlarged Display
OFF
○
Short Trend
OFF
○
Comment
CONFIG. 1
○
No. of Waveforms
3 Waveforms
○
No. of Numeric Data
4 Numeric Data
○
Displayed Waveforms
ECG1, SpO2, RESP
○
Displayed Numeric Data
HR, NIBP, SpO2, RR
○
Enlarged Display
ON
○
Short Trend
OFF
○
Comment
CONFIG. 2
○
No. of Waveforms
4 Waveforms
○
No. of Numeric Data
6 Numeric Data
○
Displayed Waveforms
ECG1, BP1/2 (overlap), SpO2, RESP
○
Displayed Numeric Data
HR, BP1,BP2, NIBP, SpO2, TEMP, RR
○
Enlarged Display
OFF
○
Short Trend
OFF
○
Comment
CONFIG. 3
○
No. of Waveforms
4 Waveforms
○
No. of Numeric Data
6 Numeric Data
○
Displayed Waveforms
Cascade, BP1/2 (overlap), SpO2, RESP
○
Displayed Numeric Data
HR, BP1, BP2, NIBP, SpO2, TEMP, RR
○
Enlarged Display
OFF
○
Short Trend
OFF
○
Comment
CONFIG. 4
○
No. of Waveforms
6 Waveforms
○
No. of Numeric Data
7 Numeric Data
○
Displayed Waveforms
○
Enlarged Display
ECG1, BP1/2 (overlap),, SpO2, CO2
HR, BP1, BP2, NIBP, SpO2, TEMP/ RR,
CO2
OFF
Short Trend
OFF
○
Comment
CONFIG. 5
○
Displayed Numeric Data
○
○
Alarm Mode Setup
Item
Alarm Mode
HR
AYSTOLE
VF
VT
SLOW_VT
RUN
COUPLET
PAUSE
BIGEMINY
TRIGEMINY
FREQUENT
TACHY
BRADY
ST
BP1 (mmHg)
BP1 (kPa)
Alarm Mode 1∼5
BP2 (mmHg)
BP2 (kPa)
NIBP (mmHg)
NIBP (%)
EtCO2 (mmHg)
EtCO2 (kPa)
EtCO2 (%)
InspCO2 (mmHg)
InspCO2 (kPa)
InspCO2 (%)
Backup
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
11
○
Setup Item
APNEA
SpO2
Default
ON 40−120
ON 5 sec.
ON
ON
ON
ON 3 beats
OFF
OFF 2 sec.
OFF
OFF
OFF 10 beats
ON
ON
OFF
ON
SYS
80−180
DIA
OFF−OFF
MEAN OFF−OFF
ON
SYS
10.0−24.0
DIA
OFF−OFF
MEAN OFF−OFF
OFF
SYS
OFF−OFF
DIA
OFF−OFF
MEAN OFF−OFF
OFF
SYS
OFF−OFF
DIA
OFF−OFF
MEAN OFF−OFF
ON 15 sec.
ON 90−OFF
ON
SYS
80−180
DIA
OFF−OFF
MEAN OFF−OFF
ON
SYS
10.0−24.0
DIA
OFF−OFF
MEAN OFF−OFF
ON 30‐45
ON 4.0‐6.0
ON 4.0‐6.0
ON 3
ON 0.4
ON 0.4
○
○
○
○
○
○
○
○
○
○
11−13
− External Connection −
Pin Assignments
This section explains the connector pin assignments.
Serial Connector Output Signal
No.
Signal Type
RESET
RSV
TxD
SG
RxD
+5V
Description
Port Reset
Reserved
Serial Transmit Data Output
Signal GND
Serial Receive Data Input
+5V
Signal Level
TTL Hi Level Reset
――
RS232C
RSV
Reserved
――
NC
No Connection
――
Description
QRS SYNC Output
Alarm Output +
(Isolation)
Signal Level
Logic TTL
RS232C
+5V power supply (150mA)
Status I/O Signal
No.
Signal Type
QRS SYNC
ALM_OUT+
RSV
Reserved
――
RSV
Reserved
――
RSV
Reserved
――
RSV
Reserved
――
+5V
+5V
+5V power supply (150mA)
ALM_OUT-
GND
Alarm Output −
(Isolation)
Power Supply Digital GND
Photo MOS Relay Contact
Photo MOS Relay Contact
――
※ As the serial connector and status I/O connector uses the same isolation power supply, the
total power supply capacity for +5V should be up to 200mA.
11−14
Chapter 12
Accessories
− Accessories −
・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2
− Optional Accessories − ・・・・・・・・・・・・・・・・・・・・・・ 3
ECG, Impedance Respiration Measurement・・・・・・ 3
Invasive Blood Pressure Measurement ・・・・・・・・・・ 3
Non-Invasive Blood Pressure Measurement ・・・・・・ 3
Temperature Measurement ・・・・・・・・・・・・・・・・・・・・ 4
SpO2 Measurement ・・・・・・・・・・・・・・・・・・・・・・・・・・・ 4
CO2 Concentration Measurement ・・・・・・・・・・・・・・・ 4
Others ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 4
12
Accessories
12−1
− Accessories −
This section lists the accessories for the DS-7100 system.
? Use only the accessories specified for this device. Otherwise, proper
function cannot be executed.
C A U T I O N ? For quality improvement, specifications are subject to change without
prior notice.
Power Cable
: CS-34 (3m)
ECG Lead Cable (5-lead)
: #3380.0661.13
ECG Relay Cable (5-lead) : CI-700D-5
2ch BP Conversion Cable : CJ-7546
NIBP Air Hose
: OA-7109B (3.5m)
Adult Cuff (Medium)
: CUF-7102A
SpO2 Interface Cable
: DOC-10
Recording Paper
: OP-124TE
This Operation Manual
12−2
− Optional Accessories −
The following products are available as optional accessories for the DS-7100 system. Purchase
them as required.
? Use only the accessories specified for this device. Otherwise, proper
function cannot be executed.
C A U T I O N ? For quality improvement, specifications are subject to change without
prior notice.
ECG, Impedance Respiration Measurement
Item
ECG Lead Cable
ECG Lead Cable
ECG Lead Cable
ECG Relay Cable
Model Type
3380.0648.13
3380.0661.13
500308800
CI-700D-3
ECG Relay Cable
CI-700E-3
3-electrode (electrosurgery-proof) ※
ECG Relay Cable
CI-700D-4
4-electrode (defibrillation-proof)
ECG Relay Cable
CI-700E-4
4-electrode (electrosurgery-proof) ※
ECG Relay Cable
CI-700D-5
5-electrode (defibrillation-proof)
ECG Relay Cable
CI-700E-5
5-electrode (electrosurgery-proof) ※
CAUTION
Q’ty
Note
3-electrode (hook type)
5-electrode (hook type)
4-electrode (hook type)
3-electrode (defibrillation-proof)
※ When using the electrosurgery-proof type ECG relay cable, respiration
measurement can not be performed.
Invasive Blood Pressure Measurement
Model Type
CJ-369
CJ-410
CJ-7546
Q’ty
Note
Q’ty
Note
12
Optional Accessories
Item
Interface Cable (for Gambro)
Interface Cable (for Becton Dichinson)
2ch BP Conversion Cable
Non-Invasive Blood Pressure Measurement
Item
Adult Cuff (Large)
Adult Cuff (Medium)
Adult Cuff (Small)
Pediatric Cuff
Infant Cuff
NIBP Air Hose (1.5m)
NIBP Air Hose (3.5m)
NIBP Extension Hose (1.5m)
NIBP Extension Hose (3.5m)
Model Type
CUF-7101
CUF-7102A
CUF-7103
CUF-7104
CUF-7105
OA-7109A
OA-7109B
OA-7110A
OA-7110B
BP Conversion Socket
CUFJ-MO1
for connection to neonate
cuff
12−3
Temperature Measurement
Item
Rectal Temperature Probe (for adult)
Rectal Temperature Probe (for pediatric)
Body Surface Temperature Probe
Probe Cover
Model Type
401J
402J
409J
70 14 616
Q’ty
10
Note
Q’ty
Note
SpO2 Measurement
Item
SpO2 DURASENSOR®
Model Type
DS-100A
SpO2 OXISENSOR® Ⅲ
D-25
24 per box
SpO2 OXISENSOR Ⅲ
D-20
24 per box
SpO2 OXISENSOR Ⅲ
I-20
24 per box
SpO2 OXISENSOR Ⅲ
N-25
24 per box
SpO2 OXISENSOR Ⅲ
R-15
24 per box
SpO2 Relay Cable
MAX-FAST
DOC-10
MAX-FAST
24 per box
®
®
®
®
CO2 Concentration Measurement
Item
Model Type
Q’ty
FilterLine H Set (Adult / Pediatric)
XS04624
FilterLine H Set (Infant / Neonate)
006324
Capno Line H (Adult)
008177
Capno Line H (Pediatric)
008178
CapnoLine H (Baby / Neonate)
008179
CapnoLine H / O2 (Adult)
008180
CapnoLine H / O2 (Pediatric)
008181
Calibration Gas Kit (5% CO2)
GR08081
Note
with Nafion, adapter
(25 per box)
with Nafion, adapter
(25 per box)
for nasal, with Nafion
(25 per box)
for nasal, with Nafion
(25 per box)
for nasal, with Nafion
(25 per box)
for nasal, with Nafion,
oxygen delivery
(25 per box)
for nasal, with Nafion,
oxygen delivery
(25 per box)
Others
Item
Ground Cable
Battery Pack
DS-7100 Mount Kit
Recording Paper
Cleaning Cloth
12−4
Model Type
CE-12
T4UR18650F-2-4644
OA-451
OP-124TE
OA-57
Q’ty
Note

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : No
Create Date                     : 2003:12:06 12:02:47Z
Modify Date                     : 2003:12:08 09:45:12+09:00
Page Count                      : 36
Creation Date                   : 2003:12:06 12:02:47Z
Author                          : ezashi
Producer                        : Acrobat PDFWriter 4.0 Windows NT
Mod Date                        : 2003:12:08 09:45:12+09:00
Metadata Date                   : 2003:12:08 09:45:12+09:00
Title                           : DS71v1-1_FA_10_Maintenance_012.PDF
Creator                         : ezashi
EXIF Metadata provided by EXIF.tools
FCC ID Filing: DV8DS7100

Navigation menu