Fukuda Denshi Co DS7100 Patient Monitor User Manual DS71v1 1 FA 03 Vital 005

Fukuda Denshi Co Ltd Patient Monitor DS71v1 1 FA 03 Vital 005

User manual 4

Download: Fukuda Denshi Co DS7100 Patient Monitor User Manual DS71v1 1 FA 03 Vital 005
Mirror Download [FCC.gov]Fukuda Denshi Co DS7100 Patient Monitor User Manual DS71v1 1 FA 03 Vital 005
Document ID380406
Application IDwOVnFJBjOncRtHtUy8lSFQ==
Document DescriptionUser manual 4
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize171.05kB (2138085 bits)
Date Submitted2003-12-10 00:00:00
Date Available2003-12-10 00:00:00
Creation Date2003-12-06 11:47:25
Producing SoftwareAcrobat PDFWriter 4.0 Windows NT
Document Lastmod2003-12-06 15:27:14
Document TitleDS71v1-1_FA_03_Vital_005.PDF
Document CreatorMicrosoft Word - DS71v1-1_FA_03_Vital_005.doc
Document Author: ezashi

Chapter 3
Vital Application
This chapter describes the procedure for vital application, etc.
●QRS Classification ・・・・・・・・・・・・・・・・・・・・・・・・・5
●Arrhythmia Type ・・・・・・・・・・・・・・・・・・・・・・・・・・・5
Filter Selection ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 6
●Filter Mode Setup ・・・・・・・・・・・・・・・・・・・・・・・・・・6
●Procedure for Filter Mode Selection・・・・・・・・・・・7
●AC Filter ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・7
Lead Cable Types ・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 7
− Respiration (Impedance Measurement) − ・・・・・・・ 8
− To Measure the SpO2 − ・・・・・・・・・・・・・・・・・・・・・・ 9
●Functional and Fractional Saturation ・・・・・・・・ 12
●Measured Versus Calculated Saturation ・・・・・ 12
− To Measure the NIBP− ・・・・・・・・・・・・・・・・・・・・・・ 14
Procedure for Periodic Measurement・・・・・・・・・・・ 16
− To Measure the BP − ・・・・・・・・・・・・・・・・・・・・・・・ 17
−To Measure the CO2 Concentration − (DS-7141) 21
Patient Application and Display・・・・・・・・・・・・・・・・ 21
Procedure for Calibration (Every 6 Months) ・・・・・ 22
− To Measure the Temperature − ・・・・・・・・・・・・・・ 24
3−1
Vital Application
− To Acquire ECG Waveform − ・・・・・・・・・・・・・・・・・ 2
Before Attaching the Electrodes ・・・・・・・・・・・・・・・・ 2
Electrode Placement ・・・・・・・・・・・・・・・・・・・・・・・・・・ 3
Connection to the Patient Monitor・・・・・・・・・・・・・・・ 4
About the Arrhythmia Analysis ・・・・・・・・・・・・・・・・・ 5
− To Acquire ECG Waveform −
Before turning ON the power
1. Check the grounding.
2. Check the recording paper.
3. Turn ON the power.
The magazine will be released by pressing the
magazine release button.
Properly ground using the
3-way AC plug.
The grounding is required to
prevent AC noise.
Open the recorder magazine, and
check that there is enough
amount of paper installed.
Turn ON the power and
check for appropriate
display.
Before Attaching the Electrodes
1. Clean the electrode sites with an alcohol swab or other skin preparation. If necessary,
shave the electrode sites to remove excessive hair.
2. Peel off the backing of disposable electrode.
Pay attention not to touch the electrode jelly.
3−2
Electrode Placement
There are 3-electrode, 4-electrode, 5-electrode application depending on the cable type.
Using the 4-electrode or 5-electrode application allows simultaneous monitoring of 2 ECG
waveforms, and high accuracy of arrhythmia analysis can be attained.
Also, the displayed lead type can be changed.
For 3-electrode lead (1 waveform monitoring)
Lead Type
Ⅰ
Color
White
Black
LL
Red
Ⅱ
Ⅲ
Electrode Site
On the right infraclavicular fossa
On the left infraclavicular fossa
On the left midclavicular line, near the
supracrestal line.
To Acquire ECG Waveform
Symbol
RA
LA
For 4-electrode lead (Simultaneous 2 waveforms monitoring)
Lead Type
Ⅰ
Symbol
RA
LA
Color
White
Black
LL
Red
RL
Green
Ⅱ
Ⅲ
aVR
aVL
aVF
Electrode Site
On the right infraclavicular fossa
On the left infraclavicular fossa
On the left midclavicular line, near the
supracrestal line.
On the right midclavicular line at the
same height as F.
For 5-electrode lead (Simultaneous 2 waveforms monitoring)
Lead Type
Ⅰ
Symbol
RA
LA
Color
White
Black
LL
Red
RL
Green
Brown
Ⅱ
Ⅲ
aVR
aVL
aVF
Electrode Site
On the right infraclavicular fossa
On the left infraclavicular fossa
On the left midclavicular line, near the
supracrestal line.
On the right midclavicular line at the
same height as F.
Chest Lead (V1∼V6)
3−3
Connection to the Patient Monitor
1. Connect the lead cable to the electrode.
Clip on the lead cable end to the electrode convex
part.
? The indication for continuous use of the electrode is about one day.
? Replace the electrode if the skin contact gets loosen due to perspiring,
etc.
C A U T I O N ? When an electrode is attached at the same location for a long time,
some patients may develop a skin irritation. Check the patient’s skin
condition periodically and change the electrode site as required.
2. Connect the lead cable to the relay cable.
3. Plug in the relay cable to the ECG input connector (green) of the patient monitor.
4. Verify that the ECG waveform is displayed on the monitor.
Adjust the waveform size and position. The
monitoring lead can be also changed.
Reference
3−4
Refer to “ 6. Parameter Setup
ECG” for size / lead setup.
About the Arrhythmia Analysis
Arrhythmia Analysis Flow
QRS
Judgement
VPC?
Learn
Waveform
Pattern Matching
VPC!
Arrhythmia
The arrhythmia detection algorithm learns the normal waveform of the patient and
compares the waveform (QRS pattern) and RR interval for each heartbeat to determine the
VPC. It compares the parameters such as QRS amplitude, QRS width, QRS polarity, RR
interval, and selects abnormal QRS. Then the QRS with suspected VPC is pattern
matched to distinguish the noise and VPC. This will finally determine the VPC and
generate the arrhythmia alarm.
To Acquire ECG Waveform
●QRS Classification
Each heartbeat will be classified to the following patterns according to the QRS judgement.
N (Normal)
V (VPC)
S (SVPC)
P (Pacing Beat)
F (Fusion Beat)
? (Undetermined Beat)
Normal QRS beat
Ventricular Extrasystole
Supraventricular Extrasystole
Pacing beat
Fusion beat of pacing and spontaneous beat
Learning arrhythmia, or beat not matching the pattern
●Arrhythmia Type
With the above QRS judgement, the following 12 types of arrhythmia alarm can be generated.
Type
Meaning
ASYSTOLE
Cardiac Arrest
VF
Ventricular Fibrillation
VT
Ventricular Tachycardia
SLOW_VT
TACHY
BRADY
Tachycardia
Bradycardia
RUN
Consecutive VPC
COUPLET
Couplet Ventricular
Extrasystole
PAUSE
BIGEMINY
TRIGEMINY
Ventricular Bigeminy
Ventricular Trigeminy
FREQUENT
Frequent VPC
Detection Criteria
Cardiac arrest is detected for more than
preprogrammed time.
A random, rapid electrical activity of the heart is
detected.
9 or more continuous ventricular beats are detected.
(HR: 140bpm / 120bpm or over)
9 or more continuous ventricular beats are detected.
(HR: under 140bpm / 120bpm)
HR is over the upper alarm limit.
HR is below the lower alarm limit.
Continuous VPC exceeding the preprogrammed
value is detected.
2 continuous beats of VPC is detected.
Cardiac arrest of 1.5 seconds and over is detected.
QRS pattern of V-N-V-N-V-N is detected.
QRS pattern of V-N-N-V-N-N is detected.
VPC exceeding the preprogrammed value is
detected within 1 minute.
3−5
Filter Selection
●Filter Mode Setup
The waveform frequency characteristic can be selected from Monitor Mode, ESIS Mode, or ST
Display Mode according to the monitoring purpose.
1. Monitor Mode Frequency Characteristic Adult / Pediatric:0.5∼40Hz Neonate:1.6∼40Hz
This is the standard mode for ECG monitoring. The upper frequency is set to 40Hz to reduce
artifact caused by EMG, etc.
2. ESIS Mode Frequency Characteristic Adult / Pediatric:1.6∼15Hz Neonate:1.6∼15Hz
By selecting this mode when using electrosurgical instrument, electrical noise can be largely
reduced. Do not select this mode unless using electrosurgical instrument.
3. ST Display Mode Frequency Characteristic Adult / Pediatric:0.05∼40Hz
Select this mode if ST measurement is the main purpose of ECG monitoring.
NOTE
If “Neonate” is selected as patient type, ST display mode can not be selected
When the filter setup is changed, a notch will appear on the ECG waveform due
to the change in frequency characteristic.
NOTE
Reference
3−6
Refer to ” 6. Parameter Setup
ECG” for details of filter mode.
●Procedure for Filter Mode Selection
1. Press the ECG parameter key and display the ECG setup menu.
2. Press the
Config. key.
3. Select the filter mode from 3 selections.
If the ECG waveform is interfered with AC noise, the AC filter cuts off the frequency component
(50Hz or 60Hz).
The AC filter is always set to ON.
Reference
Refer to “8. System Configuration Hospital Setup
AC Filter” for AC filter setup (50Hz or 60Hz).
Lead Cable Types
There are various combinations of lead cable connecting type and electrode material.
Contact our service representative for details and select the appropriate electrode.
【for 3-electrode】
【for 4-electrode】
【for 5-electrode】
ECG Relay Cable (defibrillation-proof)
ECG Relay Cable (electrosurgery-proof)
ECG Lead Cable
ECG Relay Cable (defibrillation-proof)
ECG Relay Cable (electrosurgery-proof)
ECG Relay Cable (defibrillation-proof)
ECG Relay Cable (electrosurgery-proof)
ECG Lead Cable
CI-700D-3
CI-700E-3
3380.0648.13
CI-700D-4
CI-700E-4
CI-700D-5
CI-700E-5
3380.0661.13
3−7
To Acquire ECG Waveform
●AC Filter
− Respiration (Impedance Measurement) −
1. Verify that the ECG waveform is properly acquired.
The respiration waveform is detected from lead
Ⅱof ECG mentioned in the previous section.
Therefore if stable ECG is acquired, the
respiration waveform can be acquired at the
same time.
2. Verify that the respiration waveform and respiration rate is displayed on the home display.
Adjust the waveform size, baseline position and
sweep speed.
Reference
3−8
Refer to “6. Parameter Setup Respiration” for scale / baseline setup.
Refer to ”8. System Configuration Sweep Speed” for sweep speed setup.
− To Measure the SpO2 −
1. Prepare an appropriate probe or sensor for the patient.
Sensor Types
Probe Type
(Reusable type, for adult finger)
DS-100A
For adult with weight of 40kg and over.
This is for temporary use. When continuously using for long period
of time, use the following single-use type.
To Measure the SpO2
Single Use Type
OXISENSORⅢ N-25 (for neonate toe)
For neonate with weight of 3kg and over.
OXISENSORⅢ I-20 (for pediatric toe)
For pediatric with weight of 3kg∼20kg
OXISENSORⅢ D-20 (for pediatric finger)
For pediatric or adult with weight of 10kg∼50kg
OXISENSORⅢ D-25 (for adult finger)
For adult with weight of 30kg and over.
MAX Fast (for adult forehead)
For adult with weight of 40kg and over.
With the use of new technology of NELLCOR®, OXIMAX, stable
monitoring during body motion / low perfusion is possible.
3−9
2. Connect the sensor to the patient monitor.
(1) Connect the SpO2 relay cable (DOC-10) to
the SpO2 connector on the patient monitor.
(2) Insert the sensor into the SpO2 relay cable
connector, and lock with the transparent
part.
3. Attach the sensor to the patient.
If the nail is rough, dirty, or manicured, accurate measurement will not be
C A U T I O N possible. Change the finger or clean the nail before attaching the probe
and sensor.
【Probe Type Sensor】
Light Emitting Part
(1) Attach the probe as shown on left.
The probe cable should be on the nail side.
Light Reception Part
(2) Adjust the sensor so that the light-emitting
part (on cable side) touches the root of the
nail, and close the probe.
3−10
(3) Press the probe lightly so that the finger and
the rubber cover are appressed.
This is to stabilize the probe, and to avoid
ambient light.
【Single-use Type】
(1) Clean the attachment site with alcohol, etc.
Light Emitting Element
Light Receiving Element
(3) Fixate the cable with surgical tape so that the sensor does not come off when a cable is pulled.
Attachment to the toe
Attachment to the finger
4. Verify that the SpO2 is displayed.
Press the HOME key on the lower part of
the display.
Verify that the SpO2 measurement and SpO2
waveform are displayed on the home display.
3−11
To Measure the SpO2
(2) Align the light emitting element and light
receiving element of the sensor with the
measuring site in between when attaching the
sensor to patient.
●Functional and Fractional Saturation
The DS-7100 measures functional SpO2 and may therefore produce measurements that differ
from devices measuring fractional SpO2. "Functional" SpO2 is the amount of oxygenated
hemoglobin expressed as a percentage of the total amount of hemoglobin capable of transporting
oxygen. By utilizing the light of two different wavelengths, the DS-7100 can analyze for both
oxygenated and deoxygenated hemoglobin, and consequently, can determine the functional SpO2.
The DS-7100 does not detect the presence of abnormal hemoglobin, such as carboxyhemoglobin
or methemoglobin.
●Measured Versus Calculated Saturation
When SpO2 is calculated from a blood gas measurement of the partial pressure of arterial oxygen
(PaO2), the calculated value may differ from the DS-7100 SpO2 measurement. This is because
the calculated SpO2 may not have been corrected for the effects of variables that shift the
relationship between PaO2 and SpO2 : temperature, pH, the partial pressure of carbon
dioxide(PaCO2), and the concentrations of 2, 3-DPG and fetal hemoglobin.
? When measuring the SpO2 of patient with high fever or peripheral
circulatory insufficiency, check the sensor attachment periodically and
change the attachment site. The temperature of attachment site will
rise 2∼3?C due to the sensor heat which may result in burn injury.
? For the following case, accurate measurement may not be possible.
WARNING
3−12
Patient with excessive abnormal hemoglobin (HbCO, MetHb)
Patient with the pigment injected to the blood
Patient receiving CPR treatment
When a sensor is applied to a limb with NIBP cuff, arterial catheter, or
intracatheter
? When measuring at site with venous pulse
? Patient with body motion
? Patient with small pulse
CAUTION
3−13
To Measure the SpO2
? If irritation such as skin reddening or skin fit appears with the sensor
use, change the attachment site or stop using the sensor.
? When fixating the sensor with a tape, do not wind the tape too strong.
At the same time, check the blood flow constantly so that congestion
is not generated at the peripheral.
? Even a short duration of attachment may inhibit the blood flow and
generate compression necrosis and burn injury.
? Change the sensor attachment site constantly (every 4 hours). As
the temperature of sensor attachment site normally rises 2∼3?C,
compression necrosis and burn injury may generate.
? As the skin of neonate / low birth weight infant is immature, change
the sensor attachment site more frequently depending on the
condition.
? Direct sunlight to the sensor area can cause a measurement error.
Place a black or dark cloth over the sensor.
? When not performing the measurement, unplug the relay cable and
sensor from the SpO2 connector. Otherwise, the measurement data
may be erroneously displayed by the ambient light.

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.3
Linearized                      : No
Create Date                     : 2003:12:06 11:47:25Z
Modify Date                     : 2003:12:06 15:27:14+09:00
Page Count                      : 13
Creation Date                   : 2003:12:06 11:47:25Z
Author                          : ezashi
Producer                        : Acrobat PDFWriter 4.0 Windows NT
Mod Date                        : 2003:12:06 15:27:14+09:00
Metadata Date                   : 2003:12:06 15:27:14+09:00
Title                           : DS71v1-1_FA_03_Vital_005.PDF
Creator                         : ezashi
EXIF Metadata provided by EXIF.tools
FCC ID Filing: DV8DS7100

Navigation menu