Nokia Solutions and Networks T5CT1 Cellular CDMA base station User Manual IHET5CT1 GLI3 Manual 5 of 6
Nokia Solutions and Networks Cellular CDMA base station IHET5CT1 GLI3 Manual 5 of 6
Contents
- 1. IHET5CT1 GLI2 Manual 1 of 6
- 2. IHET5CT1 GLI2 Manual 2 of 6
- 3. IHET5CT1 GLI2 Manual 3 of 6
- 4. IHET5CT1 GLI2 Manual 4 of 6
- 5. IHET5CT1 GLI2 Manual 5 of 6
- 6. IHET5CT1 GLI2 Manual 6 of 6
- 7. IHET5CT1 GLI3 Manual 1 of 6
- 8. IHET5CT1 GLI3 Manual 4 of 6
- 9. IHET5CT1 GLI3 Manual 5 of 6
- 10. IHET5CT1 GLI3 Manual 6 of 6
IHET5CT1 GLI3 Manual 5 of 6
Test Equipment Setup 68P09255A57-2 Agilent E4406A/E4432B Test Equipment Interconnection To provide proper operation during testing when both units are required, the 10 MHz reference signal from the E4406A transmitter test set must be provided to the E4432B signal generator. Connect a BNC (M)-BNC (M) cable from the E4406A 10 MHz OUT (SWITCHED) connector to the E4432B 10MHz IN connector as shown in Figure F-5. Figure F-5: Agilent 10 MHz Reference Connections E4432B 10 MHz IN TO GPIB BOX E4406A 10 MHz OUT (SWITCHED) Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY F-11 Test Equipment Setup 68P09255A57-2 Calibrating Test Cable Setup using HP PCS Interface (HP83236) Table F-9 covers the procedure to calibrate the test equipment using the HP8921 Cellular Communications Analyzer equipped with the HP83236 PCS Interface. NOTE Table:note. Note 10pt Helvetica This calibration method must be executed with great care. Some losses are measured close to the minimum limit of the power meter sensor (-30 dBm). Prerequisites Ensure the following prerequisites have been met before proceeding: Test equipment to be calibrated has been connected correctly for cable calibration. Test equipment has been selected and calibrated. Table F-9: Calibrating Test Cable Setup (using the HP PCS Interface) Step Action NOTE Verify that GPIB controller is turned off. Insert HP83236 Manual Control System card into memory card slot. Press the Preset pushbutton. Under Screen Controls, press the TESTS pushbutton to display the TESTS (Main Menu) screen. Position the cursor at Select Procedure Location and select it. In the Choices selection box, select CARD. Position the cursor at Select Procedure Filename and select it. In the Choices selection box, select MANUAL. Position the cursor at RUN TEST and select it. HP must be in Control Mode Select YES. If using HP 83236A: Set channel number=: - Position cursor at Channel Number and select it. - Enter the chan# using the numeric keypad; press [Enter] and the screen will go blank. - When the screen reappears, the chan# will be displayed on the channel number line. Set RF Generator level: - Position the cursor at RF Generator Level and select it. - Enter -10 using the numeric keypad; press [Enter] and the screen will go blank. - When the screen reappears, the value -10 dBm will be displayed on the RF Generator Level line. . . . continued on next page F-12 If using HP 83236B: Set channel frequency: - Position cursor at Frequency Band and press Enter. - Select User Defined Frequency. - Go Back to Previous Menu. - Position the cursor to 83236 generator frequency and enter actual RX frequency. - Position the cursor to 83236 analyzer frequency and enter actual TX frequency. SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Test Equipment Setup 68P09255A57-2 Table F-9: Calibrating Test Cable Setup (using the HP PCS Interface) Step Action Set the user fixed Attenuation Setting to 0 dBm: - Position cursor at Analyzer Attenuation and select it - Position cursor at User Fixed Atten Settings and select it. - Enter 0 (zero) using the numeric keypad and press [Enter]. 10 Select Back to Previous Menu. 11 Record the HP83236 Generator Frequency Level: Record the HP83236B Generator Frequency Level: - Position cursor at Show Frequency and Level Details and select it. - Under HP83236 Frequencies and Levels, record the Generator Level. - Under HP83236B Frequencies and Levels, record the Generator Frequency Level (1850 - 1910 MHz). - Position cursor at Prev Menu and select it. 12 Click on Pause for Manual Measurement. 13 Connect the power sensor directly to the RF OUT ONLY port of the PCS Interface. 14 On the HP8921A, under To Screen, select CDMA GEN. 15 Move the cursor to the Amplitude field and click on the Amplitude value. 16 Increase the Amplitude value until the power meter reads 0 dBm ±0.2 dB. NOTE The Amplitude value can be increased coarsely until 0 dBM is reached; then fine tune the amplitude by adjusting the Increment Set to 0.1 dBm and targeting in on 0 dBm. 17 Disconnect the power sensor from the RF OUT ONLY port of the PCS Interface. * IMPORTANT The Power Meter sensor’s lower limit is -30 dBm. Thus, only components having losses ≤30 dB should be measured using this method. For further accuracy, always re-zero the power meter before connecting the power sensor to the component being calibrated. After connecting the power sensor to the component, record the calibrated loss immediately. 18 Disconnect all components in the test setup and calibrate each one separately by connecting each component, one-at-a-time, between the RF OUT ONLY PORT and the power sensor. Record the calibrated loss value displayed on the power meter. Example: (A) Test Cable(s) (B) 20 dB Attenuator = (B) Directional Coupler = -1.4 dB -20.1 dB -29.8 dB 19 After all components are calibrated, reassemble all components together and calculate the total test setup loss by adding up all the individual losses: Example: Total test setup loss = -1.4 -29.8 -20.1 = -51.3 dB. This calculated value will be used in the next series of tests. 20 Under Screen Controls press the TESTS button to display the TESTS (Main Menu) screen. 21 Select Continue (K2). 22 Select RF Generator Level and set to -119 dBm. . . . continued on next page Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY F-13 Test Equipment Setup 68P09255A57-2 Table F-9: Calibrating Test Cable Setup (using the HP PCS Interface) Step Action 23 Click on Pause for Manual Measurement. 24 Verify the HP8921A Communication Analyzer/83203A CDMA interface setup is as follows (fields not indicated remain at default): Verify the GPIB (HP-IB) address: under To Screen, select More select IO CONFIG Set HP-IB Adrs to 18 set Mode to Talk&Lstn Verify the HP8921A is displaying frequency (instead of RF channel) - Press the blue [SHIFT] button, then press the Screen Control [DUPLEX] button; this switches to the CONFIG (CONFIGURE) screen. - Use the cursor control to set RF Display to Freq 25 Refer toChapter 3 for assistance in setting the cable loss values into the LMF. F-14 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Test Equipment Setup 68P09255A57-2 Figure F-6: Cable CalibrationUsing HP8921 with PCS Interface MEMORY CARD SLOT POWER SENSOR (A) (A) POWER SENSOR (B) (B) 20 dB / 20 WATT ATTENUATOR POWER SENSOR (C) POWER SENSOR (C) 150 W NON-RADIA TING RF LOAD Aug 2002 30 dB DIRECTIONAL COUPLER FW00292 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY F-15 Test Equipment Setup 68P09255A57-2 Calibrating Test Cable Setup using Advantest R3465 NOTE Be sure the GPIB Interface is OFF for this procedure. Advantest R3465 Manual Test setup and calibration must be performed at both the TX and RX frequencies. Table F-10: Procedure for Calibrating Test Cable Setup Using Advantest R3465 Step Action * IMPORTANT - This procedure can only be performed after test equipment has been allowed to warm-up and stabilize for a minimum of 60 minutes. Press the SHIFT and the PRESET keys located below the display Press the ADVANCE key in the MEASUREMENT area of the control panel. Select the CDMA Sig CRT menu key Select the Setup CRT menu key Using the vernier knob and the cursor keys set the following parameters NOTE Fields not listed remain at default Generator Mode: SIGNAL Link: FORWARD Level Unit: dBm CalCorrection: ON Level Offset: OFF Select the return CRT menu key Press FREQ key in the ENTRY area Set the frequency to the desired value using the keypad entry keys Verify that the Mod CRT menu key is highlighting OFF; if not, press the Mod key to toggle it OFF. 10 Verify that the Output CRT menu key is highlighting OFF; if not, press the Output key to toggle it OFF. 11 Press the LEVEL key in the ENTRY area. 12 Set the LEVEL to 0 dBm using the key pad entry keys. 13 Zero power meter. Next connect the power sensor directly to the “RF OUT” port on the R3561L CDMA Test Source Unit. 14 Press the Output CRT menu key to toggle Output to ON. 15 Record the power meter reading ________________________ . . . continued on next page F-16 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Test Equipment Setup 68P09255A57-2 Table F-10: Procedure for Calibrating Test Cable Setup Using Advantest R3465 Step 16 Action Disconnect the power meter sensor from the R3561L RF OUT jack. * IMPORTANT The Power Meter sensor’s lower limit is -30 dBm. Thus, only components having losses < 30 dB should be measured using this method. For best accuracy, always re-zer o the power meter before connecting the power sensor to the component being calibrated. Then, after connecting the power sensor to the component, record the calibrated loss immediately. 17 Disconnect all components in the the test setup and calibrate each one separately. Connect each component one-at-a-time between the “RF OUT” port and the power sensor (see Figure F-7, “Setups A, B, and C”). Record the calibrated loss value displayed on the power meter for each connection. Example: (A) 1st Test Cable = -0.5 dB (B) 2nd Test Cable = -1.4 dB (C) 20 dB Attenuator = -20.1 dB (D) 30 dB Directional Coupler = -29.8 dB 18 Press the Output CRT menu key to toggle Output OFF. 19 Calculate the total test setup loss by adding up all the individual losses: Example: Total test setup loss = 0.5 + 1.4 + 20.1 + 29.8 = 51.8 dB This calculated value will be used in the next series of tests. 20 Press the FREQ key in the ENTRY area 21 Using the keypad entry keys, set the test frequency to the RX frequency 22 Repeat steps 9 through 19 for the RX frequency. 23 Refer to Chapter 3 for assistance in setting the cable loss values into the LMF. Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY F-17 Test Equipment Setup 68P09255A57-2 Figure F-7: Cable Calibration using Advantest R3465 RF OUT POWER SENSOR (A) & (B) (C) POWER SENSOR 20 DB / 2 WATT ATTENUATOR POWER SENSOR (C) POWER SENSOR (D) 100 W NON-RADIA TING RF LOAD F-18 FW00320 30 DB DIRECTIONAL COUPLER SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Test Equipment Setup 68P09255A57-2 Calibrating HP 437 Power Meter Precise transmit output power calibration measurements are made using a bolometer-type broadband power meter with a sensitive power sensor. Follow the steps outlined in Table F-11 to enter information unique to the power sensor before calibrating the test setup. Refer to Figure F-8 as required. NOTE Table:note. Note 10pt Helvetica This procedure must be done in conjunction with the automated calibration to enter power sensor specific calibration values. Figure F-8: Power Meter Detail CONNECT POWER SENSOR TO POWER REFERENCE WHEN CALIBRATING UNIT. POWER REFERENCE IS ENABLED USING THE SHIFT KEYS SHIFT (BLUE) PUSHBUTTON ACCESSES FUNCTION AND DATA ENTRY KEYS IDENTIFIED WITH LIGHT BLUE TEXT ON THE FRONT PANEL ABOVE THE BUTTONS CONNECT POWER SENSOR WITH POWER METER TURNED OFF FW00308 Table F-11: Power Meter Calibration Procedure Step Action ! CAUTION Do not connect/disconnect the power meter sensor cable with ac power applied to the meter. Disconnection could result in destruction of the sensing element or mis-calibration. - Make sure the power meter AC LINE pushbutton is OFF. - Connect the power sensor cable to the SENSOR input. Set the AC LINE pushbutton to ON. NOTE The calibration should be performed only after the power meter and sensor have been allowed to warm-up and stabilize for a minimum of 60 minutes. Aug 2002 Perform the following to set or verify the GPIB address: - To enter the SPECIAL data entry function, press [SHIFT] then [PRESET] . - Use the [ ] or [ ] button to select HP-IB ADRS; then press [ENTER]. - Use the [ ] or [ ] button to select HP-IB ADRS 13; then press [ENTER]. - To EXIT the SPECIAL data entry function press [SHIFT] then [ENTER]. . . . continued on next page SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY F-19 Test Equipment Setup 68P09255A57-2 Table F-11: Power Meter Calibration Procedure Step Action Perform the following to set or verify the correct power sensor model: - Press [SHIFT] then [ ] to select SENSOR. - Identify the power sensor model number from the sensor label. Use the [ ] or [ ] button to select the appropriate model; then press [ENTER]. NOTE Be sure the PWR REF (power reference) output is OFF (observe that the triangular indicator is NOT displayed as shown in Step 7). If on, press [SHIFT] then [ ] to turn it off. Press [ZERO] . Display will show “Zeroing ******.” Wait for process to complete. Connect the power sensor to the POWER REF output. To turn on the PWR REF, perform the following: - Press [SHIFT] then [ ]. - Verify that the triangular indicator (below) appears in the display above “PWR REF”. Perform the following to set the REF CF %: - Press ([SHIFT] then [ZERO] ) for CAL. - Enter the sensor’s REF CF % from the sensor’s decal using the arrow keys and press [ENTER]. (The power meter will display ”CAL *****” for a few seconds.) NOTE If the REF CAL FACTOR (REF CF) is not shown on the power sensor, assume it to be 100%. Perform the following to set the CAL FAC %: - Press [SHIFT] then [FREQ] for CAL FAC. - On the sensor’s decal, locate an approximate calibration percentage factor (CF%) at 2 GHz. Enter the sensor’s calibration % (CF%) using the arrow keys and press [ENTER]. When complete, the power meter will typically display 0.05 dBm. (Any reading between 0.00 and 0.10 is normal.) 10 To turn off the PWR REF, perform the following: - Press [SHIFT] then [ ]. - Disconnect the power sensor from the POWER REF output. F-20 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Calibrating Gigatronics 8541C power meter 68P09255A57-2 Calibrating Gigatronics 8541C power meter Precise transmit output power calibration measurements are made using a bolometer-type broadband power meter with a sensitive power sensor. Follow the steps in Table F-12 to enter information unique to the power sensor. Table F-12: Calibrate Gigatronics 8541C Power Meter Step Action ! CAUTION Do not connect/disconnect the power meter sensor cable with AC power applied to the meter. Disconnection could result in destruction of the sensing element or miscalibration. NOTE Allow the power meter and sensor to warm up and stabilize for a minimum of 60 minutes before performing the calibration procedure. Make sure the power meter POWER pushbutton is OFF. Connect the power sensor cable to the SENSOR input. Set the POWER pushbutton to ON. Verify the Power GPIB mode and address: Press MENU. Use the Use the arrow key to select GPIB, and press ENTER. Use the Press arrow key to select CONFIG MENU, and press ENTER. arrow keys to set MODE to 8541C. and use the arrow keys as required to set ADDRESS to 13. Press ENTER. Connect the power sensor to the CALIBRATOR output connector. Press ZERO. Wait for the process to complete. Sensor factory calibration data is read to power meter during this process. Disconnect the power sensor from the CALIBRATOR output. Figure F-9: Gigatronics 8541C Power Meter Detail CONNECT POWER SENSOR TO CALIBRATOR POWER REFERENCE WHEN CALIBRATING/ZEROING UNIT CONNECT POWER SENSOR WITH POWER METER TURNED OFF FRONT View AC POWER GPIB CONNECTION REAR View FW00564 Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY F-21 Calibrating Gigatronics 8541C power meter 68P09255A57-2 Notes F-22 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Appendix G Power Calibration Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY G-1 Calibrating Output Power 68P09255A57-2 Calibrating Output Power Power Calibration This procedure is a guide to expanding your system with multiple carriers while the system remains in service. This procedure also allows you to perform on site maintenance (replace defective boards and recalibrate) while the remainder of the site stays in service. Motorola recommends that you perform this procedure during a maintenance window. This procedure cannot be performed on BTSs with 4-to-1 combiners. The procedure can only be performed on one side of the BTS at one time. That is, LPAs 1, 2 ,3, 7, 8, 9 (feed antennas 1, 2, 3) can be calibrated while LPAs 6, 7, 8, 10, 11, 12 (feed antennas 4, 5, 6) remain in service and vice versa. Equipment Warm up NOTE Calibration of the communications test set (or equivalent test equipment) must be performed at the site before calibrating the overall test set. Calibrate the test equipment after it has been allowed to warm-up and stabilize for a minimum of 60 minutes. CAUTION If any piece of test equipment (i.e., test cable, RF adapter) has been replaced, re-calibration must be performed. Failure to do so could introduce measurement errors, causing incorrect measurements and degradation to system performance. Power Delta Calibration Introduction The In-service calibration procedure has several differences from a normal calibration procedure. One of these is the use of a spectrum analyzer instead of a power meter to measure power. Power meters are broadband measurement devices and cannot be used to measure power during In-service Calibration since other carriers are operating. A spectrum analyzer can be used because it measures power at a given frequency. However, measuring power using a spectrum analyzer is less accurate than using a power meter. Therefore, you must compensate for the difference (delta) between the power meter and the spectrum analyzer. G-2 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Calibrating Output Power 68P09255A57-2 HP8921A Power Delta Calibration Use the HP8921A Spectrum Analyzer to measure power during In-Service Calibration for 800 MHz systems. After the offset value has been calculated, add it to the TX cable loss value. Follow the procedure in Table G-1 to perform the HP8921A Power Delta Calibration procedure. NOTE This procedure requires two HP8921As. Table G-1: HP8921A Power Delta Calibration Procedure Step Action * IMPORTANT Perform this procedure after test equipment has been allowed to warm-up and stabilize for a minimum of 60 minutes. Connect a short RF cable between the HP8921A Duplex Out port and the HP437B power sensor (see Figure G-1). Set the HP8921A signal source as follows: - Measure mode to CDMA Generator - Frequency to the CDMA Calibration target frequency - CW RF Path to IQ - Output Port to Dupl - Data Source to Random - Amplitude to 0 dBm Measure and record the power value reading on the HP437B Power Meter. Record the Power Meter reading as result A ________________________. Turn off the source HP8921A signal output, and disconnect the HP437B. NOTE Leave the settings on the source HP8921A for convenience in the following steps. Connect the short RF cable between the source HP8921A Duplex Out port and the measuring HP8921A RF-IN port (see Figure G-2). Ensure that the source HP8921A settings are the same as in Step 2. Set the measuring HP8921A as follows: - Measure mode to CDMA Anl - Frequency to the CDMA calibration target frequency - Input Attenuation to 0 dB - Input port to RF-IN - Gain to Auto - Analyzer Direction to Fwd Turn on the source HP8921A signal output. 10 Measure and record the channel power reading on the measuring HP8921A as result B ________________________. 11 Turn off the source HP8921A signal output and disconnect the equipment. . . . continued on next page Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY G-3 Calibrating Output Power 68P09255A57-2 Table G-1: HP8921A Power Delta Calibration Procedure Step 12 Action Compute the delta between HP437B and HP8921A using the following formula: Delta = A - B Example: Delta = -0.70 dBm - (-1.25 dBm) = 0.55 dBm Example: Delta = 0.26 dBm - 0.55 dBm = -0.29 dBm These examples are included to show the mathematics and do not represent actual readings. NOTE Add this delta value to the TX Cable Loss value during In-Service Calibration. Figure G-1: Delta Calibration Setup - HP8921A to HP437B HP 8921A HP437B SENSOR Power Sensor DUPLEX OUT Short RF Cable FW00801 Figure G-2: Delta Calibration Setup - HP8921A to HP8921A Measurement HP8921A Source HP8921A DUPLEX OUT RF IN/OUT Short RF Cable FW00802 Advantest R3465 Power Delta Calibration Follow the procedure in Table G-2 to perform the Advantest 3465 Power Delta Calibration procedure. G-4 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Calibrating Output Power 68P09255A57-2 Table G-2: Advantest Power Delta Calibration Procedure Step Action * IMPORTANT Perform this procedure after test equipment has been allowed to warm-up and stabilize for a minimum of 60 minutes. On the Advantest R3465: Press the SHIFT and the PRESET keys located below the CRT display. Press the ADVANCE key in the Measurement area of the control panel. Press the CDMA Sig CRT menu key. Press the FREQ key in the Entry area of the control panel. Set the frequency to the desired value using the keypad entry keys. Press the LEVEL key in the Entry area of the control panel. Set the LEVEL to 0 dBm using the keypad entry keys. Verify the Mod CRT menu key is highlighting OFF, if not press the Mod key to toggle it OFF. Verify the Output CRT menu key is highlighting OFF, if not press the Output key to toggle it OFF. On the HP 437 Power Meter: 10 Zero the Power Meter prior to connecting the power sensor to the RF cable from the signal generator. * IMPORTANT For best accuracy, always re-zero the power meter before connecting the power sensor to the component being calibrated. 11 Connect the RF cable from the R3561L CDMA Test Source Unit RF OUT port to the power sensor, refer to Figure G-3. 12 Press the Output CRT menu key to toggle the Output to ON. 13 Record the Power Meter reading as result A ________________________. 14 Press the Output CRT menu key to toggle the Output to OFF. 15 Connect the RF cable from the R3561L CDMA Test Source Unit RF OUT port to the Spectrum Analyzer INPUT Port, refer to Figure G-4. 16 Press the Output CRT menu key to change the Output to ON. 17 Press the CW key in the Measurement area of the control panel. 18 Press the LEVEL key in the Entry area of the control panel. 19 Set the REF LEVEL to 10 dBm using the keypad entry keys. 20 Press the dB/div CRT menu key. 21 Press the 10 dB/div CRT menu key. 22 Press the FREQ key in Entry area of the control panel. 23 Set the frequency to the desired value using the keypad entry keys. 24 Press the more 1/2 CRT menu key. . . . continued on next page Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY G-5 Calibrating Output Power 68P09255A57-2 Table G-2: Advantest Power Delta Calibration Procedure Step Action 25 Press the Preselector CRT menu key to highlight 3.0G. 26 Press the FORMAT key in the Display Control area of the control panel. 27 Press the TRACE CRT menu key. 28 Press the AVG A CRT menu key. 29 Set AVG to 20 using keypad entry keys. 30 Press the return CRT menu key. 31 Press the SPAN key in the Entry area of the control panel. 32 Press the Zero Span CRT menu key. 33 Press the BW key in the Entry area of the control panel. 34 Press the RBW CRT menu key to highlight MNL. using keypad entry keys enter 30 kHz. 35 Set RBW to 30 kHz using keypad entry keys. 36 Press the VBW CRT menu key to highlight MNL. 37 Set VBW to 1 MHz using keypad entry keys. 38 Press the Marker ON key in the Display Control area of the control panel. 39 Record the Marker Level reading as result B ________________________. 40 Calculate the Power Calibration Delta value. The delta value is the power meter measurement minus the Advantest measurement. Delta = A - B Example: Delta = -0.70 dBm - (-1.25 dBm) = 0.55 dBm Example: Delta = 0.26 dBm - 0.55 dBm = -0.29 dBm These examples are included to show the mathematics and do not represent actual readings. NOTE Add this delta value to the TX Cable Loss value during In-Service Calibration. Figure G-3: Delta Calibration Setup - R3561L to HP437B Advantest R3561L RF OUT Power Sensor HP437B Short RF Cable SENSOR FW00803 G-6 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Calibrating Output Power 68P09255A57-2 Figure G-4: Delta Calibration Setup - R3561L to R3465 RF OUT R3561L Short RF Cable R3465 INPUT FW00804 HP8935 Power Delta Calibration Follow the procedure in Table G-3 to perform the HP8935 Power Delta Calibration procedure. Table G-3: HP8935 Power Delta Calibration Procedure Step Action * IMPORTANT Perform this procedure after test equipment has been allowed to warm-up and stabilize for a minimum of 60 minutes. Connect a short RF cable between the HP8935 Duplex Out port and the HP437B power sensor (see Figure G-5). Set the HP8935 signal source as follows: - Measure mode to CDMA Gen - Frequency to the CDMA Calibration target frequency - CW RF Path to IQ - Output Port to Dupl - Data Source to Random - Amplitude to 0 dBm Measure and record the power value reading on the HP437B Power Meter. Record the Power Meter reading as result A ________________________. Turn off the source HP8935 signal output, and disconnect the HP437B. NOTE Leave the settings on the source HP8935 for convenience in the following steps. Connect the short RF cable between the source HP8935 Duplex Out port and the RF-IN/OUT port (see Figure G-6). Ensure that the source HP8935 settings are the same as in Step 2. Set the measuring HP8935 as follows: - Measure mode to CDMA Anl - Frequency to the CDMA calibration target frequency - Input Attenuation to 0 dB - Input port to RF-IN - Gain to Auto - Anl Dir to Fwd . . . continued on next page Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY G-7 Calibrating Output Power 68P09255A57-2 Table G-3: HP8935 Power Delta Calibration Procedure Step Action Turn on the source HP8935 signal output. 10 Set the Chn Pwr Cal to Calibrate and select to calibrate. 11 Measure and record the channel power reading on the measuring HP8935 as result B ________________________. 12 Turn off the source HP8935 signal output and disconnect the equipment. 13 Calculate the Power Calibration Delta value. The delta value is the power meter measurement minus the Advantest measurement. Delta = A - B Example: Delta = -0.70 dBm - (-1.25 dBm) = 0.55 dBm Example: Delta = 0.26 dBm - 0.55 dBm = -0.29 dBm These examples are included to show the mathematics and do not represent actual readings. NOTE Add this delta value to the TX Cable Loss value during In-Service Calibration. Figure G-5: Delta Calibration Setup - HP8935 to HP437B Hewlett-Packard Model HP 8935 ÁÁ ÁÁ ÁÁ ÁÁ HP437B SENSOR Power Sensor DUPLEX OUT Short RF Cable FW00805 Figure G-6: Delta Calibration Setup - HP8935 to HP8935 Hewlett-Packard Model HP 8935 DUPLEX OUT RF IN/OUT Short RF Cable G-8 FW00806 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Calibrating Output Power 68P09255A57-2 Agilent E4406A Power Delta Calibration The Agilent E4406A transmitter tester and E4432B signal generator test equipment combination can be used for CDMA 2000 as well as IS-95A/B operation modes. The power delta calibration is performed on the E4406A, but the E4432B is required to generate the reference signal used to calculate the power delta offset. After the offset value has been calculated, add it to the TX cable loss value in the LMF. Follow the procedure in Table G-4 to perform the Agilent E4406A Power Delta Calibration procedure. Table G-4: Agilent E4406A Power Delta Calibration Procedure Step Action * IMPORTANT Perform this procedure after test equipment has been allowed to warm-up and stabilize for a minimum of 60 minutes. After it is warmed up and stabilized, calibrate the test equipment as described in the “Test Set Calibration” section of the Optimization/Calibration chapter in the SC 4812ET Optimization/ATP manual. Zero the Power Meter prior to connecting the power sensor to the RF cable from the signal generator. * IMPORTANT For best accuracy, always re-zero the power meter before connecting the power sensor to the component being calibrated. Connect a short RF cable from the E4432B RF OUTPUT connector the HP437 power meter power sensor (see Figure G-7). Set the E4432B signal generator as follows: - Press Preset to exit any modes for which the signal generator is configured. - Press Frequency and enter the frequency of the channel to be calibrated on the numeric keypad. - Using the soft keys to the right of the screen, select the frequency range to be measured; for example MHz. - Press Amplitude and, using the numeric keypad, set signal amplitude to 0 (zero). - Using the soft keys, set the measurement type to dBm. On the E4432B, press RF On/Off to toggle the RF output to RF ON. - Note that the RF On/Off status in the screen display changes. Measure and record the value reading on the HP437 power meter as result A____________________. On the E4432B, press RF On/Off to toggle the RF output to RF OFF. - Note that the RF On/Off status in the screen display changes. Disconnect the short RF cable from the HP437 power meter power sensor, and connect it to the RF INPUT connector on the E4406A transmitter tester (see Figure G-8). . . . continued on next page Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY G-9 Calibrating Output Power 68P09255A57-2 Table G-4: Agilent E4406A Power Delta Calibration Procedure Step Action * IMPORTANT Do not change the frequency and amplitude settings on the E4432B when performing the following steps. Set the E4406A as follows: - Press Preset to exit any modes for which the transmitter tester is configured - Press MODE and, using the soft keys to the right of the screen, select cdmaOne - Press MEASURE and, using the soft keys, select spectrum - Press Frequency and, using the soft keys, select Center Frequency - Enter the frequency of the channel to be calibrated using the numeric keypad - Using the soft keys, select the frequency range to be measured; for example, MHz - Press Input/Output and, using the soft keys, select Input Atten - Using the numeric keypad, set Input Atten to 0 (zero) and, using the soft keys, select dB - Using the soft keys, select External Atten and then select Mobile - Using the numeric keypad, set Mobile to 0 (zero) and, using the soft keys, select dB - Using the soft keys, select Base - Using the numeric keypad, set Base to 0 (zero) and, using the soft keys, select dB - Press MEASURE and, using the soft keys, select Channel Power On the E4432B signal generator, press RF On/Off to toggle the RF output to RF ON. - Note that the RF On/Off status in the screen display changes. 10 Read the measured Channel Power from the E4406A screen display and record it as result B____________________. 11 On the E4432B, press RF On/Off to toggle the RF output to RF OFF. - Note that the RF On/Off status in the screen display changes. 12 Calculate the Power Calibration Delta value. The delta value is the power meter measurement minus the Agilent measurement. Delta = A - B Example: Delta = -0.70 dBm - (-1.25 dBm) = 0.55 dBm Example: Delta = 0.26 dBm - 0.55 dBm = -0.29 dBm These examples are included to show the mathematics and do not represent actual readings. NOTE Add this delta value to the TX Cable Loss value during In-Service Calibration. G-10 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Calibrating Output Power 68P09255A57-2 Figure G-7: Delta Calibration Setup - Agilent E4432B to HP437 AGILENT E4432B AND E4406A HP437B SENSOR RF OUTPUT Power Sensor Short RF Cable FW00858 Figure G-8: Delta Calibration Setup - Agilent E4432B to Agilent E4406A AGILENT E4432B AND E4406A RF OUTPUT Short RF Cable RF INPUT FW00859 In-Service Calibration NOTE This feature does NOT have fault tolerance at this time. The system has no safe-guards to stop you from doing something that will take the BTS out of service. If possible, perform this procedure during a maintenance window. Follow the procedures in this section precisely, otherwise the entire BTS will most likely go OUT OF SERVICE. At the CBSC, only perform operations on expansion hardware when it is in the OOS_MANUAL state. The operator must be trained in the LMF operation prior to performing this procedure. Prerequisites Expansion hardware has been added in the CBSC database, and the CDF file has been generated. The expansion devices have been inserted into the C-CCP cage and are in the OOS_MANUAL state at the CBSC. The site specific cdf (with the expansion hardware) and cal files have been loaded onto the LMF. The LMF has the same code and dds files as the CBSC to download. Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY G-11 Calibrating Output Power NOTE 68P09255A57-2 Do not download code or data to any cards other than those you are working on. Downloading code or data to other cards will take the site OUT OF SERVICE. The code file version numbers must match the version numbers on the other cards in the frame. If the numbers do not match, the site may go OUT OF SERVICE. The BTS-#.cdf, CBSC-#.cdf, and CAL files for this BTS must have come from the CBSC. Test equipment has been configured per Figure G-9 or Figure G-10. An RFDS (or at a minimum a directional coupler), whose loss is already known, must be in line to perform the in-service calibration. Test equipment has been calibrated after 1 hour warm up. A short RF cable and two BNC-N adapters are available to perform Cable Calibration. The Power Delta Calibration has been performed (see Table G-1, Table G-2, or Table G-3). G-12 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Calibrating Output Power 68P09255A57-2 Figure G-9: Optimization/ATP Test Setup Using RFDS TEST SETS Optimization/ATP SET UP Hewlett-Packard Model HP 8935 SYNC MONITOR EVEN SEC TICK PULSE REFERENCE FROM CSM BOARD DUPLEX OUT FREQ MONITOR 19.6608 MHZ CLOCK REFERENCE FROM CSM BOARD ÁÁ Á Á ÁÁ ÁÁ HP-IB TO GPIB BOX NOTE: IF BTS RX/TX SIGNALS ARE DUPLEXED: BOTH THE TX AND RX TEST CABLES CONNECT TO THE DUPLEXED ANTENNA GROUP. 20 DB PAD (FOR 1.7/1.9 GHZ) 10 DB PAD RX (FOR 800 MHZ) TEST CABLE COMMUNICATIONS TEST SET EXT REF IN DUPLEX OUT ANTENNA TX TEST CABLE RF IN/OUT 20 DB PAD (FOR 1.7/1.9 GHZ) 10 DB PAD (FOR 800 MHZ) RFDS DUPLEXER DIRECTIONAL COUPLER RF IN/OUT RX TEST CABLE EVEN SECOND/ SYNC IN IEEE 488 GPIB BUS FWD COUPLED PORT GPIB CABLE TX TEST CABLE RX ANTENNA PORT TX ANTENNA PORT BTS FREQ MONITOR DIP SWITCH SETTINGS S MODE DATA FORMAT BAUD RATE ON SYNC MONITOR GPIB ADRS CSM G MODE RS232-GPIB INTERFACE BOX LAN RS232 NULL MODEM CABLE LAN 10BASET/ 10BASE2 CONVERTER UNIVERSAL TWISTED PAIR (UTP) CABLE (RJ45 CONNECTORS) CDMA LMF INTERNAL PCMCIA ETHERNET CARD REF FW00759 Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY G-13 Calibrating Output Power 68P09255A57-2 Figure G-10: IS-95 A/B/C Optimization/ATP Test Setup Using RFDS TEST SETS Optimization/ATP SET UP Advantest R3267 (Top) and R3562 (Bottom) NOTE: IF BTS RX/TX SIGNALS ARE DUPLEXED: BOTH THE TX AND RX TEST CABLES CONNECT TO THE DUPLEXED ANTENNA GROUP. RX TEST CABLE TO EXT TRIG ON REAR OF SPECTRUM ANALYZER ANTENNA INPUT 50 OHM 20 DB PAD (FOR 1.7/1.9 GHZ) 10 DB PAD (FOR 800 MHZ) MOD TIME BASE IN (EXT REF IN) RF OUT 50 OHM NOTE: SIGNAL GENERATOR EXT REF IN TX TEST CABLE BNC “T” EXT TRIG IN RFDS DUPLEXER DIRECTIONAL COUPLER COMMUNICATIONS TEST SET RF OUT 50 OHMS EVEN SECOND/ SYNC IN RF INPUT 50 OHMS IEEE 488 GPIB BUS FWD COUPLED PORT SYNC MONITOR EVEN SEC TICK PULSE REFERENCE FROM CSM BOARD FREQ MONITOR 19.6608 MHZ CLOCK REFERENCE FROM CSM BOARD SYNTHE REF IN ON REAR OF SIGNAL GENERATOR IS CONNECTED TO 10 MHZ OUT ON REAR OF SPECTRUM ANALYZER BNC “T” GPIB CABLE TX CABLE RX CABLE Agilent E4432B (Top) and E4406A (Bottom) RX ANTENNA PORT TX ANTENNA PORT RF OUT 50 OHMS DIP SWITCH SETTINGS BTS 19.6608 MHZ CLOCK FREQ MONITOR RF INPUT 50 OHMS TO TRIGGER IN ON REAR OF TRANSMITTER TESTER TO PATTERN TRIG IN ON REAR OF SIGNAL GENERATOR GPIB ADRS SYNC MONITOR EVEN SEC TICK PULSE REFERENCE FROM CSM BOARD LAN RS232 NULL MODEM CABLE LAN 10BASET/ 10BASE2 CONVERTER UNIVERSAL TWISTED PAIR (UTP) CABLE (RJ45 CONNECTORS) G MODE RS232-GPIB INTERFACE BOX CDMA LMF FREQ MONITOR 19.6608 MHZ CLOCK REFERENCE FROM CSM BOARD BNC “T” G-14 TO 10 MHZ IN (EXT REF IN) ON REAR OF TRANSMITTER TESTER ON SYNC MONITOR CSM S MODE DATA FORMAT BAUD RATE INTERNAL PCMCIA ETHERNET CARD REF. FW00759 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Calibrating Output Power 68P09255A57-2 Follow the procedure in Table G-5 to perform the In-Service Calibration. Table G-5: In-Service Calibration Step Action * IMPORTANT Perform this procedure after test equipment has been allowed to warm-up and stabilize for a minimum of 60 minutes. Set up the LMF for In-Service Calibration: - Start the LMF by double-clicking the LMF icon on the Windows desktop. - Click Tools>Options from the menu bar at the login screen. - Check the applicable spectrum analyzer check box and the signal generator check box on the Test Equipment tab. Ensure that the GPIB address is 18 for the CDMA analyzer and 1 for the signal generator. - Uncheck any other other equipment that is selected. - Click the Apply button. - Select the BTS Options tab in the LMF Option window. - Check the In-Service Calibration check box. - Click the Apply button. - Click the Dismiss button to close the LMF Option window. Login to the target BTS: - Select the target BTS icon. - Click the Login button at the login screen. Measure the Cable Loss using the Cable Calibration function: - Click Util>Cable Calibration from the menu bar at the main window. - Set the desired channel(s) and select TX and RX CABLE CAL at the cable calibration pop up window. - Click the OK button to perform cable calibration. - Follow the on-screen instructions to complete the cable loss measurement. NOTE - The measured value is input automatically to the cable loss file. - To view the cable loss file, click Util>Examine>Cable Loss>TX or RX. Add the spectrum analyzer power delta to the Cable Loss. - To view the cable loss file, click Util>Examine>Cable Loss>TX or RX. - Add the value computed in Table G-1, Table G-2, or Table G-3 to the TX Cable Loss. NOTE Be sure to include the sign of the value. The following examples are included to show the mathematics and do not represent actual readings: - Example: 5.65 dBm + 0.55 dBm = 6.20 dBm - Example: 5.65 dBm + (-0.29 dBm) = 5.36 dBm - Example: -5.65 dBm + 0.55 dBm = -5.10 dBm - Example: -5.65 dBm + (-0.29 dBm) = -5.94 dBm . . . continued on next page Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY G-15 Calibrating Output Power 68P09255A57-2 Table G-5: In-Service Calibration Step Action Input the Coupler Loss for the TX and RX tests: - Click Util>Edit>Coupler Loss>TX or RX from the menu bar at the main window. - Input the appropriate coupler loss for the target carrier(s) by referring to the information taken at the time of BTS installation. - Click the Save button. - Click the Dismiss button to close the window. - To view the coupler loss file, click Util>Examine>Coupler Loss>TX or RX. Have the CBSC operator put the redundant BBX OOS_MANUAL. ! CAUTION Be sure to download OOS devices only. Loading in-service devices takes them OUT OF SERVICE and can result in dropped calls. The code file version numbers must match the version numbers on the other cards in the frame. If the numbers do not match, the site may go OUT OF SERVICE. NOTE Be sure to include the redundant BBX in steps 9, 10, and 11. Download code and data to the target devices: - Click Tools>Update NextLoad>CDMA to set the code version that will be downloaded. - Select the BTS(s) you need, check the appropriate code version in the pop up window, and click the Save button to close. - Select the target BBX(s) on the C-CCP cage picture. - Click Device>Download>Code/Data to start downloading code. - Select the target BBX(s) on the C-CCP cage picture. - Click Device>Download>Data to start downloading data. . . . continued on next page G-16 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Calibrating Output Power 68P09255A57-2 Table G-5: In-Service Calibration Step Action ! CAUTION Perform the In-service Calibration procedure on OOS devices only. Select the desired test: - Select the target BBX(s) on the C-CCP cage picture. - Click Tests>[desired test] from the menu bar at the main window. - Select the target carrier and confirm the channel number in the pop up window. - Leave the Verify BLO check box checked. - From the Test Pattern pick list, select a test pattern. - Click the OK button to start calibration. - Follow the on-screen instructions, except, do not connect to the BTS antenna port, connect to the directional coupler (fwd) port associated with the on screen prompt antenna port. NOTE Selecting Pilot (default) performs tests using a pilot signal only. Selecting Standard performs tests using pilot, synch, paging and six traffic channels. This requires an MCC to be selected. Selecting CDFPilot performs tests using the CDF value for pilot gain and IS-97 gain values for all the other channels included in the Standard pattern setting (paging, synch, and six traffic). Using this pattern setting requires the selection of both a BBX and at least one MCC. Selecting CDF performs tests using pilot, synch, paging and six traffic channels, however, the gain for the channel elements is specified in the CDF file. Save the result and download the BLO data to the target BBX(s): - Click the Save Result button on the result screen. The window closes automatically. 10 Logout from the BTS and close the LMF session: - Click BTS>Logout to close the BTS connection. - Close the LMF window. 11 Restore the new “bts-*.cal” file to the CBSC. 12 Enable the target device(s) from the CBSC. Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY G-17 Calibrating Output Power 68P09255A57-2 Notes G-18 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Appendix H Cable Interconnection Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY H-1 Intra-Cabinet Cabling 68P09255A57-2 Intra-Cabinet Cabling SC 4812ET Intra-Cabinet Cabling This appendix provides the identification and location of the cables connecting the components which make up the SC 4812ET RF cabinet. The number of cables and components incorporated in the RF cabinet will vary depending on the the manner in which the cabinet is equipped. For example, a 3 sector, 2 carrier system will require less components and less cables than a 6 sector 2 carrier system. Refer to Table H-1 and Figure H-1 through Figure H-19 for the cable you wish to research. NOTE In some cases cables with the same number are used to connect two different signalling paths. These cables are designated A & B and the point they connect to and from is also designated A & B. Ensure the correct cable (A or B) is connected to the correct designation (A or B) connector or plug. Table H-1: SC4812ET RF CABINET INTER-CONNECT CABLES DRDC, Combiner, Trunking Backplane Cables CABLE # FROM NOTE’s TO 3064795A05 TX CIO Figure H-6 Trunking BP 3064735A10 TX Trunking BP Figure H-5 Combiner Connector Pnl 3064735A07 TX Combiner See Figure H-1, Figure H-11, Figure H-12, Figure H-13, and Figure H-14 DRDC 3064735A11 TX Combiner (See Above) DRDC 3064735A12 TX Combiner (See Above) DRDC 3064795A07 TX CIO Figure H-11 Trunking BP 3086435H01 TX Combiner QDS PkZ Adptr 3086435H02 TX Combiner QDS PkZ Adptr 3086435H03 TX Combiner QDS PkZ Adptr 3086168H01 Power LPA PDA Figure H-5 Trunk BP C-CCP Cables CABLE # FROM NOTE’s TO 3064809A01 Power CCCP/PDA See Figure H-1 & Figure H-5 C-CCP Backplane 3064899A04 LAN I/O A in See Figure H-1 & Figure H-5 C-CCP LAN I/O A In C-CCP Cables (cont) CABLE # FROM NOTEs TO 3064899A04 LAN I/O B in See Figure H-1 & Figure H-5 C-CCP LAN I/O B In 3086033H03 GPS Surge Arrestor See Figure H-1 & Figure H-5 C-CCPBackplane 3064899A04 LAN I/O B in See Figure H-1 & Figure H-5 C-CCP LAN I/O B In . . . continued on next page H-2 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Intra-Cabinet Cabling 68P09255A57-2 Table H-1: SC4812ET RF CABINET INTER-CONNECT CABLES 3064899A03 C-CCP LAN I/O A Out See Figure H-1 & Figure H-5 LAN I/O A Out 3064899A03 C-CCP LAN I/O B Out See Figure H-1 & Figure H-5 LAN I/O B Out 3064899A07 Sync CSM See Figure H-1 I/O 3086000H02 Site I/O C-CCP See Figure H-9 J2 on ETIB 3086001H02 A SPAN I/O BlkHd See Figure H-1 & Figure H-5 C-CCP/A SPAN I/O 3086001H02 B SPAN I/O BlkHd See Figure H-1 & Figure H-5 C-CCP/B SPAN I/O 3086086H02 Alarm C-CCP See Figure H-1 & Figure H-9 J1 on ETIB 3086366H02 HSO/LFR See Figure H-1 & Figure H-5 C-CCP HSO/LFR 4886044H01 LBD See Figure H-1 & Figure H-5 C-CCP Backplane Span I/O Board Cables CABLE # FROM NOTEs TO 3086601H01 SPAN I/O Pblock See Figure H-1 & Figure H-10 CSU 3086601H02 SPAN CSU See Figure H-1 & Figure H-10 A & B SPAN I/O BRD 3086001H02 SPAN CSU See Figure H-1 & Figure H-10 A & B SPAN Connector on C-CCP Backplane RFDS Cables CABLE # FROM NOTE’s TO 3064238A17 RFDS/ASU-1 Figure H-16 & Figure H-17 DRDC 3064238A18 RFDS/ASU-2 Figure H-16 & Figure H-17 DRDC 3064238A19 RFDS/ASU-1 Figure H-16 & Figure H-17 DRDC 3064238A20 RFDS/ASU-2 Figure H-16 & Figure H-17 DRDC ETIB/LPAC Cables CABLE # FROM NOTE’s TO 3064794A03 LPAC See Figure H-8 C-CCP Bkpln 3064794A05 ETIB See Figure H-1 & Figure H-9 RFDS 3086433H04 RGPS I/O See Figure H-1 & Figure H-9 ETIB (15 position) 3086169H01 Power, Heat Exchanger, PDA See Figure H-1 & Figure H-9 ETIB, OPT, HX, EBA 3086500H01 Alm/Ctrl ETIB See Figure H-1 & Figure H-9 HX/LPA 3086505H01 DC Cable (DC Power Dist) See Figure H-1 & Figure H-19 EBA Blower Assembly 3086566H01 LPAC See Figure H-1 & Figure H-8 1 Cable to each LPA Bk Pln 3086569H01 Door Intrusion Alarm See Figure H-5 & Figure H-19 Door Switch 3086655H02 LPAC See Figure H-8, Figure H-5 & Figure H-19 ETIB Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY H-3 Intra-Cabinet Cabling 68P09255A57-2 Figure H-1: 4812ET RF Cabinet Internal FRU Locations (See Figure H-8, Figure H-9), and Figure H-10) EBA RFDS ETIB LPAC SPAN I/O C-CCP Shelf (See Figure H-2 and Figure H-5) (See Figure H-11, Figure H-12, Figure H-13, and Figure H-14) OPTIONAL AREA (See Figure H-10) Combiner Cage DRDC LPA’s CSU Modem DC Power Dist. Punch Block (back) LAN OUT IN DRDC 19 MHz 2 Sec LAN I/O Span I/O to C-CCP SYS Sync Alarm to ETIB To Span I/O (See Figure H-11) LPA Trunking Backplane (See Figure H-5 (See Figure H-4, and Figure H-10) Figure H-5, Figure H-6 and Figure H-8) FW00698 H-4 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Intra-Cabinet Cabling 68P09255A57-2 C-CCP Cables and Cable Connectors The C-CCP Shelf assembly consists of the C-CCP Shelf and the attached backplane with cables and connectors (see Figure H-2 and Figure H-3). The C-CCP shelf contains all of the CDMA unique functions within the SC 4812ET RF frame. The C-CCP shelf contains the following components: Broadband Transceiver (BBX) cards Multi-Channel CDMA (MCC) cards Combiner Input/Output (CIO) card Power Supply cards Group Line Interface (GLI3) cards Alarm Monitor Reporting (AMR) cards Clock Synchronization Modules (CSM) High Stability Oscillator/Low frequency Receiver (HSO/LFR) Multicoupler Preselector Cards (MPC) CDMA Clock Distribution (CCD) card Integrated Frame Modem (IFM) card Switch card C-CCP Fan Modules Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY H-5 Intra-Cabinet Cabling 68P09255A57-2 Figure H-2: C-CCP Shelf Cable Numbers and Connectors (To J1 connector on the ETIB) Cable # 3086086H02 (To SPAN A I/O connector on the Bulkhead) Cable # 3086001H02 ALARMS (To J2 connector on the ETIB) Cable # 3086000H02 (To SPAN B I/O connector on the Bulkhead) Cable # 3086001H02 SITE I/O SPAN B SPAN A Cable # 3086366H02 LAN I/O A Cable # 3064899A04 LAN I/O B Cable # 3064899A04 MPC/EMPC-1 BBX2-6 BBX2-13 BBX2-5 BBX2-4 BBX2-3 BBX2-2 BBX2-1 MCC24-6 MCC24-5 MCC24-4 MCC24-3 MCC24-2 GLI3-1 MCC24-1 AMR-1 PS-3 PS-2 PS-1 19 mm Filter Panel HSO/LFR MPC/EMPC-2 Switch Card BBX2-12 BBX2-1 1 BBX2-10 BBX2-9 BBX2-8 BBX2-7 MCC24-12 MCC24-1 1 MCC24-10 MCC24-9 MCC24-8 MCC24-7 GLI3-2 AMR-2 FILLER CCD-1 CCD-2 CSM-2 CSM-1 HSO/LFR CIO To LAN I/O connectors on the Bulkhead LPAC C-CCP Shelf FW00699 Cable # 3064794A03 (To the C-CCP connector on the LPAC Module) H-6 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Intra-Cabinet Cabling 68P09255A57-2 Figure H-3: C-CCP Backplane (To J1 connector on the ETIB) Cable # 3086086H02 (To SPAN A I/O connector on the Bulkhead) Cable # 3086001H02 (To J2 connector on the ETIB) Cable # 3086000H02 (To SPAN B I/O connector on the Bulkhead) Cable # 3086001H02 SYSTEM LED Cable # 4886044H01 HSO/LFR LAN IN A LAN IN B CCCP Power 3064809A01 Cable # 3086366H02 Cable # 3064899A04 Cable # 3064899A04 To LAN I/O connectors on the Bulkhead Cable # 3064794A03 (To the C-CCP connector on the LPAC Module) Cable # LAN OUT A 3064899A03 LAN OUT B Cable # 3064899A03 LPAC GPS Cable # 3086033H03 FW00700 Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY H-7 Intra-Cabinet Cabling 68P09255A57-2 RF Cabinet LPA Cables There can be a maximum of 16 LPAs in an RF cabinet. The connections shown are for one LPA backplane which controls four LPAs. The remaining LPAs are connected in the same manner. Refer to Figure H-4, through Figure H-8 for the cables connected to the LPAs in the 4812ET RF cabinet. Figure H-4: LPAs for the SC 4812ET EBA ETIB RFDS 5 RU RACK SPACE SC 4812ET BTS RF Cabinet Note: LPA Component door not shown for clarity Unpopulated LPA Shelf Cover External Blower Assembly (EBA) LPA Module (4-Each Cage) FW00173 H-8 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Intra-Cabinet Cabling 68P09255A57-2 Figure H-5: BTS Combiner to LPA Backplane Cables LPA Backplane EBA RFDS LPA-1 ETIB TX OUT1 5 RU RACK SPACE Cable 3064735A10 (3 each) TX IN 1 TX OUT2 LPA-2 TX IN2 TX IN 3 LPA-3 TX OUT3 SC 4812ET BTS RF Cabinet LPA-4 4B 4A 1B 1A S1 C1 S2 S3 S1 C2 LPA-1 S2 5B 5A 2B 2A S3 S1 C3 S2 LPA-2 S3 6B 6A 3B 3A S1 C4 S2 S3 FROM APPROPRIATE LPA LPA 1, LPA 2, LPA 3, LPA 4 LPA-3 LPA-4 BACK Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY FW00708-REF H-9 Intra-Cabinet Cabling 68P09255A57-2 Figure H-6: Combiner to LPA Backplane/LPA Backplane To CIO Board Cables LPA Backplane 1 EBA RFDS LPA-1 ETIB TX OUT1 5 RU RACK SPACE TX IN 1 TX OUT2 Cable 3064735A10 (3 each) LPA-2 RED TX IN2 ORANGE YELLOW TX IN 3 LPA-3 TX OUT3 SC 4812ET BTS RF Cabinet LPA-4 4B 4A 1B S1 1A C1 S2 TO J15 on CIO Board (See Figure H-7) Cable 3064795A05 S3 S1 C2 LPA-1 S2 5B 5A 2B 2A S3 S1 C3 TX OUT1 S2 TX IN 1 TX OUT2 S3 6B 6A 3B 3A S1 C4 LPA-2 GREEN VIOLET TX IN2 S2 S3 BLUE TX IN 3 LPA-3 TX OUT3 LPA-4 FW00711 LPA Backplane 2 H-10 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Intra-Cabinet Cabling 68P09255A57-2 Figure H-7: Components Located on CIO Card SC 4812ET RF Cabinet ETIB EBA RFDS 5 RU RACK SPACE RX EXP A J12 RX EXP B J13 TX BTS 1-6 J15 BTS 7-12 M/F 1-6 J14 TX FW00237 Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY H-11 Intra-Cabinet Cabling 68P09255A57-2 LPAC Cabling The LPAC module provides the communication interface from the ETIB and C-CCP to the LPA through the trunking backplane. The LPAC interface board is contained in a protective housing which is mounted on the RF cabinet frame behind the ETIB module. See Figure H-3, Figure H-8 and Figure H-9 for connecting cables and connector locations. The LPAC is located internally to the frame as shown in Figure H-1. Figure H-8: LPAC Interface Board Connectors and Attaching Cable Numbers LPAC INTERFACE BOARD FRONT LPA-1 ALARM RIBBON CABLE TO ETIB 3086655H02 RIBBON CABLE TO C-CCP BACKPLANE 3064794A03 TX OUT1 LPA-2 TX IN TX OUT2 TX IN2 TX IN 3 LPA 1A, 1B 1C, 1D RIBBON CABLE 3086566H01 LPA-3 TX OUT3 LPA 2A, 2B 2C, 2D RIBBON CABLE 3086566H01 LPA-4 LPA 3A, 3B 3C, 3D RIBBON CABLE 3086566H01 LPA 4A, 4B 4C, 4D RIBBON CABLE 3086566H01 LPA BACKPLANE 1, 2, 3, & 4 FW00702 Cable # 3086566H01 (4 Connections each Side) NOTE: The LPAC is Located Behind the ETIB Module H-12 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Intra-Cabinet Cabling 68P09255A57-2 ETIB Cables and Cable Connectors The ETIB module (see Figure H-9) provides the interface for the LPA’s through the LPAC, punchblock, heat exchanger and alarms to the C-CCP backplane. The ETIB interface board is contained in a protective housing which is mounted on the RF cabinet frame. The ETIB is located internally to the frame as shown in Figure H-1. Figure H-9: ETIB I/O Connectors and Attaching Cable Numbers Cable # 3064794A05 (See Figure H-16) Cable # 3086168H01 J6 DC POWER Cable # 3086655H02 (See Figure H-8) P7 RFDS P2 LPAC P9 ALARMS IN (From Bulkhead ) Cable # 3064534A08 (See Figure H-1) J1 To ALARMS OUT (C-CCP Backplane) (Reserved) SITE I/O (C-CCP Backplane) P8 OPTIONS RECEPTACLE DOOR INTRUSION ALARM Cable # 3086569h01 Cable # 3086000H02 See Figure 2-5 and Figure H-3 J2 J5 LFR/HSO J4 Cable # 3086433H04 (See Figure H-1) Cable # 3086500h01 HEAT EXCH J3 RGD/RGPS Cable # 3086366H02 See Figure 2-5 and Figure H-3 Aug 2002 Cable # 3086086H02 See Figure 2-5 and Figure H-3 FW00701 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY H-13 Intra-Cabinet Cabling 68P09255A57-2 SPAN I/O Cable Connection Diagram The SPAN I/O card provides the frame interface and secondary surge protection for the T1 lines. There are two span cards in an RF cabinet. SPAN I/O A supports spans A, C, and E. SPAN I/O B supports span B, D, and F. See Figure H-10 for SPAN cables and cable connections. The SPAN I/O is located internally to the frame as shown in Figure H-1. H-14 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002 Intra-Cabinet Cabling 68P09255A57-2 Figure H-10: SPAN I/O Cables and Connectors 20 Pair RGD Punchblock Board (RGPS) SPAN I/O (A & B) Interface Module (Located Behind the LPAC Module, See Figure H-1) RGD/RGPS 50 Pair Punch Block 1A 2A 3A 1B 2B 3B 4A 5A 6A 4B 5B 6B Microwave (Alarms/ Spans) RF Expansion Ports Power Input +27V RF GPS LAN 1A 2A 3A 1B 2B 3B IN OUT Remote ASU Power Input 27V Ret 4A 5A 6A 4B 5B 6B 19 MHz Spans Modem Alams Antenna’s 2 Sec GND Lugs 4812ET Rear Connector Panel SPAN I/O WIRING DIAGRAM SPAN I/O A SPAN I/O B C-CCP Backplane 3086001H02 Bulkhead SPAN I/O A Bulkhead SPAN I/O Connector 3086601H01 CSU 3086601H02 SPAN I/O B 3086001H02 To/From GLI To/From Network DTE To/From GLI To/From Network DATA PORT DCE SLOT 2 T1 DDS SLOT 1 T1 TERMINAL NETWORK DATA PORT T1 DDS T1 TERMINAL NETWORK CONTROL PORT GROUP ADDRESS SHELF ADDRESS CSU Back View Aug 2002 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY FW00703 H-15 Intra-Cabinet Cabling 68P09255A57-2 DRDC/TRDC Cables and Cable Connections The DRDC is a Duplexer, RX Filter, and Directional Coupler which provides the RF interface at the rear of the cabinet. The connections are the antenna connection (outside rear), transmit into the DRDC TX filter. Receive out of DRDC (RX filter), and Directional coupler. The TRDC is a TX filter/RX filter/Directional Coupler that is the same as the DRDC except the TRDC has two antenna outputs (TX only and RX only) The TRDC is not available in the 1.9 GHz band. See Figure H-11, Figure H-12, Figure H-13, and Figure H-14 for the cable diagram that fits the configuration of your BTS site. H-16 SC4812ET Optimization/ATP Manual Software Release R16.1.x.x PRELIMINARY Aug 2002
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.3 Linearized : No Modify Date : 2002:11:01 17:10:12-05:00 Create Date : 2002:11:01 17:09:56-05:00 Producer : Acrobat Distiller 4.05 for Windows Page Count : 46EXIF Metadata provided by EXIF.tools