Olympus Medical Systems EC-1 Low Power Transmitter User Manual GT1629 FCC MAJ 1467 C1
Olympus Medical Systems Corp. Low Power Transmitter GT1629 FCC MAJ 1467 C1
Contents
- 1. Users manual Part 1
 - 2. Users manual Part 2
 - 3. Users manual Part 3
 - 4. Users manual Part 4
 - 5. Users manual part 5
 
Users manual part 5

210
Chapter 6  Capsule Endoscope Image Observation
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
The folders and the data associated with the selected 
examination (image data, case data, and exported images) 
will be removed during the deletion process.
3. Click the [OK] button on the examination data deletion screen (see Figure 
6.93). The selected examination data will be deleted. A progress screen is 
displayed while the examination data is being deleted.
Figure 6.93
4. When the deletion of examination data is complete, you will be returned to 
the examination list screen.
OK button

Chapter 6  Capsule Endoscope Image Observation
211
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Writing examination and thumbnail data onto a DVD
1. Select the examination data you wish to write onto a DVD and click the 
[DVD] button on the examination list screen (see Figure 6.94). The DVD 
writing screen is displayed.
Figure 6.94
You can select multiple examination data simultaneously. 
The thumbnail data associated with the selected examination 
data will be written automatically.
2. Use the drive selection box to select the destination drive to write the data to 
(see Figure 6.95). The amount of available disk space on the selected drive 
is displayed.
Figure 6.95
• The folder for the examination data will be created 
automatically during the writing process.
DVD button
Drive selection box

212
Chapter 6  Capsule Endoscope Image Observation
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
• The following DVD formats are supported:
• Though you use DVD-RW or DVD+RV, you cannot rewrite or 
delete data.
• While it is possible to copy the data to an USB HDD, the 
safety of data on the USB HDD cannot be assured, so it is 
advised that the user first back up the examination data to a 
DVD.
3. The selected examination data which you wish to write is displayed (see 
Figure 6.96).
Figure 6.96
4. Click the [Option] button to display the DVD writing options screen (see 
Figure 6.97).
Figure 6.97
5. Select the type of files to be written together with the examination data. Click 
the [Picture] button to write still images, and the [Movie] button to write video 
files (see Figure 6.98).
DVD : DVD-R, DVD+R
Select examination 
data
Option button

Chapter 6  Capsule Endoscope Image Observation
213
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Figure 6.98
All items are selected by default.
6. Click the [OK] button to save the settings and exit the DVD writing options 
screen (see Figure 6.99).
Figure 6.99
Picture button
Movie button
OK button

214
Chapter 6  Capsule Endoscope Image Observation
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
7. Check the “Password” checkbox to set a password (see Figure 6.100).
Figure 6.100
If you specify a password when writing the examination data, 
you will be required to enter the password to open the written 
examination data.
8. Check the “Make anonymous” checkbox to write the examination data with 
the individual information removed (see Figure 6.101).
Figure 6.101
• Individual information removed include “Patient name” and 
“Image showing individual information”.
• To remove “Image showing individual information”, you will 
need to set the hiding feature. For more information, refer to 
“Hiding images” on page 216.
Password checkbox
Make anonymous 
checkbox

Chapter 6  Capsule Endoscope Image Observation
215
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
9. Click the [OK] button on the DVD writing screen (see Figure 6.102). The 
selected examination data will be written. The progress screen is displayed 
while the examination data is being written.
Figure 6.102
• To stop the writing process, click the [Cancel] button on the 
progress screen.
• If there is insufficient storage space on the writing 
destination, an error message will be displayed. In that case, 
refer to Chapter 8, “Troubleshooting” on page 233.
• The DVD drive will open after writing the data and close 
automatically to verify the written data. Please wait until the 
verification is completed.
10. When writing of examination data is complete, the examination list screen 
will be displayed.
Finalize the report before you write the report onto the DVD.
OK button

216
Chapter 6  Capsule Endoscope Image Observation
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Hiding images 
You can specify images that contain the facial portrait of a patient to be hidden 
when copying or exporting examination data.
To hide images, the “Make anonymous” checkbox must be 
checked when copying or writing the examination data.
• Specify the first image and the last image of the display 
range. Images between the first and last images will be 
included, and all other images will be excluded (hidden). 
• By default, the first and last images of each examination are 
used as the first and last images of the display range.
1. On the main screen, open the examination data which contains the images 
to be hidden (“Opening examination data and thumbnail data from the 
examination list screen” on page 148).
2. Display your desired first image in the image display area (6.3 “Image 
observation” on page 155).
3. Right-click on the image and select “First” from “Display range” in the 
context menu (see Figure 6.103).
Figure 6.103
4. Display your desired last image in the image display area (6.3 “Image 
observation” on page 155).
“First”

Chapter 6  Capsule Endoscope Image Observation
217
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
5. Right-click on the image and select “Last” from “Display range” in the 
context menu (see Figure 6.104).
Figure 6.104
6. To reset the display range, right-click on an image and select “Reset” from 
“Display range” in the context menu (see Figure 6.105).
Figure 6.105
“Last”
“Reset”

218
Chapter 6  Capsule Endoscope Image Observation
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
6.6 Exporting image data 
You can export image data as still images or videos. Export settings can be 
configured on the export screen.
Export screen
Figure 6.106
1. Movie list
Displays the list of movie files to be created.
2. Picture list
Displays the list of picture files to be created.
3. Back button
Returns to the main screen.
4. Export list
Displays the list of movie files and picture files to be exported.
5. Drive selection box
Selects the drive to which the movie files and picture files are exported.
6. Format selection area
Selects the picture format used for exporting the picture files.
6. Format 
selection area
4. Export list
3. Back button
2. Picture list
1. Movie list
5. Drive 
selection box

Chapter 6  Capsule Endoscope Image Observation
219
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Movie type
Movie
The images before and after the selected thumbnail change into a movie. 
The number of images is defined in the movie settings.
Combined movie
The movies created for each thumbnail are combined. The number of 
images is defined in the movie settings.
Clipped movie
The images between the first and last of the selected thumbnails change 
into a movie.
Movie
Combined m ovie
Clipped m ovie
Thum bnail R a n g e  o f  th e  c re a te d  m o v ie
Movie
Combined movie
Clipped movie
Thumbnail Range of the created movie

220
Chapter 6  Capsule Endoscope Image Observation
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Movie settings
The number of images used to form a movie is set on the movie settings screen.
1. Select “Settings” from the “Tools” menu on the main screen.
2. Click the “Clips” tab. The movie settings screen is displayed (see Figure 
6.107).
Figure 6.107
3. Set the number of images to be added before and after the selected 
thumbnail (see Figure 6.108).
Figure 6.108
You can specify a number between 1 and 999.
4. Click the [OK] button on the movie settings screen.
Clips tab

Chapter 6  Capsule Endoscope Image Observation
221
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Creating pictures / movies / combined movies / clipped 
movies
The created pictures and movies are stored on the 
workstation. To export a picture or movie file, refer to 
“Exporting picture / movie files” on page 225.
1. On the main screen, open the examination data (see “Opening thumbnail 
data from the menu” on page 152).
2. Create thumbnails you wish to include in your export (see “Creating 
thumbnails” on page 163).
3. Select the thumbnails you wish to include in your export (see Figure 6.109).
To select multiple thumbnails, click the thumbnails while 
holding down the [Shift] key on the keyboard.
Figure 6.109

222
Chapter 6  Capsule Endoscope Image Observation
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
4. To create pictures, right-click on one of the selected thumbnails and select 
“Export” and then “Picture” from the context menu (see Figure 6.110).
Figure 6.110
The exported file name is determined automatically. The data 
is exported to the D drive as generic image files.
5. To create movies, right-click on one of the selected thumbnails and select 
“Movie” from the context menu (see Figure 6.111).
Figure 6.111
• The exported file name is determined automatically.
• The file is exported in AVI format.
Picture menu
Movie menu

Chapter 6  Capsule Endoscope Image Observation
223
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
6. To create a combined movie, right-click on one of the selected thumbnails 
and select “Combined movie” from the context menu (see Figure 6.112).
The images for the selected thumbnails are combined into a 
movie.
Figure 6.112
• The “Combined movie” context menu item can be selected 
only when multiple thumbnails have been selected.
• The exported file name is determined automatically. The data 
is exported as generic image files.
• The file is exported in AVI format.
• If the number of images that form the combined video 
exceeds 10000, the video will be divided into multiple files.
Combined movie 
menu

224
Chapter 6  Capsule Endoscope Image Observation
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
7. To create a clipped movie, right-click on one of the selected thumbnails and 
select “Clipped movie” from the context menu (see Figure 6.113).
All images between the first and the last selected thumbnails 
are combined into a movie.
Figure 6.113
• The “Combined movie” context menu item can be selected 
only when multiple thumbnails have been selected.
• The export file name is determined automatically.
• The file is exported in AVI format.
• If the number of images that form the combined video 
exceeds 10000, the video will be divided into multiple files.
Clipped movie 
menu

Chapter 6  Capsule Endoscope Image Observation
225
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Exporting picture / movie files
1. Click the [Export] button on the main screen (see Figure 6.114). The export 
screen is displayed.
Figure 6.114
2. Use the drive selection box to select the destination drive for the export (see 
Figure 6.115). The amount of available disk space on the selected drive is 
displayed.
Figure 6.115
Export button

226
Chapter 6  Capsule Endoscope Image Observation
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
3. Select the picture and/or movie files you wish to export (see Figure 6.116).
Figure 6.116
4. Click the [ ] button to add the picture and/or movie files you wish to export 
to the export list from the picture list or the movie list (see Figure 6.117).
Figure 6.117
To clear a picture or movie file from the export list select the 
file you wish to clear and click the [Clear] button.
 button

Chapter 6  Capsule Endoscope Image Observation
227
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
5. Select “Copy” or “Delete”.
Figure 6.118
6. You can copy the picture files in one of the following four formats (see Figure 
6.119).
•       BMP (bitmap)
•       JPEG
•       TIFF
•       PNG
Figure 6.119
Delete
Copy
Format 
selection area

228
Chapter 6  Capsule Endoscope Image Observation
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
7. Click the [OK] button on the export screen (see Figure 6.120). The selected 
files will be exported. The progress screen is displayed while the files are 
being exported.
Figure 6.120
• To stop the export process, click the [Cancel] button on the 
progress screen.
• If there is insufficient storage space on the export destination, 
an error message will be displayed. In this case, refer to 
Chapter 8, “Troubleshooting” on page 233.
8. When exporting is complete, you will be returned to the export screen.
The picture or movie files is exported to the following 
location.
(∗) Words in parentheses vary according to the exported 
data.
OK button

Chapter 6  Capsule Endoscope Image Observation
229
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
6.7 Browsing the image observation operation 
history
Operation history for the image observation is automatically recorded.
1. From the “Tools” menu on the main screen, select “Diagnosis log” and then 
“View” (see Figure 6.121). The diagnosis log screen is displayed.
Figure 6.121
• Date, time, and user name for each operation are recorded in 
the log.
• To export a diagnosis log, select “Diagnosis log” and then 
“Export” from the “Tools” menu on the main screen.
• The log can be exported in text format.

230
Chapter 7  Care Storage and Disposal
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Chapter 7 Care Storage and Disposal
7.1 Care
• After wiping with a place of moistened gauze, dry the capsule 
endoscope system thoroughly before using it again. If it is 
used while still wet, there is the risk of electrical shock.
• When cleaning the capsule endoscope system, always wear 
appropriate personal protection equipment. Blood, mucous 
and other potentially infectious material adhering the video 
system center could pose and infection control risk.
• Do not soak the system components in water, or sterilize 
them using autoclave or gas. Doing so will damage the 
system.
• Do not wipe the external surfaces with hard or abrasive 
wiping material. Doing so will scratch the surface.
After using the system, immediately perform the following cleaning procedure.
1. Remove dust, dirt, and other stains on the surface of the antenna pads and 
the antenna lead set using a piece of gauze moistened with a neutral 
detergent, then wipe the surfaces using a piece of gauze moistened with 
70% ethyl or isopropyl alcohol.
2. Remove dust, dirt, and other stains on the surface of the following 
equipment using a piece of gauze moistened with a neutral detergent, then 
wipe the surfaces using a piece of gauze moistened with 70% ethyl or 
isopropyl alcohol. After doing so, dry them thoroughly.
Do not soak the equipment in any liquid.
It may damage the equipment.
• Recorder unit • Recorder unit cradle • Real time viewer
• Real time viewer cable • Recorder unit cradle cable • Battery charger
• Recorder unit harness • Capsule activator

Chapter 7  Care Storage and Disposal
231
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
3. If the recorder unit harness has become stained, first clean off the stain 
using a piece of gauze moistened with a neutral detergent, then wipe off 
with a piece of cloth moistened with water. Alternatively, the recorder unit 
harness can be machine-washed.
Do not iron or dry clean the recorder unit harness to shrink 
slightly.
Note that machine-washing may cause the recorder unit 
harness to shrink slightly.
4. If the recorder unit harness becomes wet, be sure to dry it completely.
7.2 Storage
• Store the unused capsule endoscopes in a dark place, at 
temperature from 0°C/32°F to 25°C/77°F. Avoid storing the 
capsule endoscopes in a refrigerator or a freezer, as it may 
cause condensation.
• Keep water and any other liquid away from the system 
components. It may cause equipment damage.
• Do not store the system in a location exposed to direct 
sunlight, UV rays, X-rays, radio activity, or strong 
electromagnetic radiation (e.g. near microwave therapy 
equipment, shortwave therapy equipment, MRI equipment, 
radio, or mobile phones). They may damage the system.
• Do not store the system in a location affected by high 
temperature, high humidity, barometric pressure, dust, salt or 
ozone. They may damage the system.
• Store the system in a location immune to vibrations and 
impacts. They may damage the system.
After cleaning the equipment according to the procedure given in Section 7.1, 
“Care”, be sure to dry it off thoroughly. When the equipment is dry, place them 
into their designated positions in the carrying cases for storage.
Store the carrying cases in a clean and dry location.
For details of the storage environment, refer to “Operating/storage environment” 
on page 240 in the Appendix.

232
Chapter 7  Care Storage and Disposal
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
7.3 Disposal
• When disposing of any item, follow all applicable national and local laws and 
guidelines.
• The capsule endoscope is discharged naturally.

Chapter 8  Troubleshooting
233
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Chapter 8 Troubleshooting
If, during the inspections described in Chapter 5, “Capsule Endoscope 
Procedures” on page 96, the system or any of its components appears to have 
problems, refer to 8.1 “Troubleshooting” on page 233 to correct the problem.
If the problem persists, stop using the system, and refer to 8.2 “Returning the 
system for repair” on page 239 to correct the situation. If the problem cannot be 
resolved by these countermeasures, contact Olympus.
Never use the system if any abnormality is suspected. 
Damage to or irregularity in the system not only causes 
malfunctioning, but may also cause injury to the patient’s 
body cavity.
8.1 Troubleshooting
The following table lists the possible causes of and countermeasures for troubles 
that may occur due to equipment setting errors or deterioration of consumables. 
Also refer to “Recorder unit error messages” on page 235 and “Real time viewer 
error messages” on page 236 on pages 235 to 236.
Troubles or failures other than those listed in the following table require repair. 
As repair performed by persons who are not qualified by Olympus could cause 
patient or user injury and/or equipment damage, be sure to contact Olympus for 
repair (refer to Section 8.2).
Irregularity Possible Cause Solution
Capsule 
endoscope’s LED 
fails to illuminate.
The power is not turned ON. Use the capsule activator to turn 
the power ON.
The capsule endoscope is 
broken.
Use a new capsule endoscope.
Recorder unit’s 
power fails to come 
ON.
The battery pack is missing. Insert a charged battery pack.
The battery pack is low. Recharge the battery pack.
Recorder unit’s 
power lamp fails to 
illuminate.
The recorder unit is broken. Try turning on the recorder unit 
power again. If problem persists, 
contact Olympus.
Real time viewer’s 
power fails to come 
ON.
The battery pack is missing. Insert a charged battery pack.
The battery pack is low. Recharge the battery pack.

234
Chapter 8  Troubleshooting
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
The real time 
viewer’s power 
lamp fails to 
illuminate.
The real time viewer is 
broken.
Try turning on the viewer power 
again. If the problem persists, 
contact Olympus.
Images from the 
capsule 
endoscope are not 
displayed by the 
real time viewer. 
(When using the 
real time viewer 
alone.)
The real time viewer is 
broken.
Contact Olympus.
The capsule endoscope is 
broken.
Use a new capsule endoscope.
The real time viewer is too far 
from the capsule endoscope.
Reduce the distance between the 
real time viewer and the capsule 
endoscope.
Images from the 
capsule 
endoscope are not 
displayed by the 
real time viewer. 
(When connected 
to the recorder 
unit.)
The antenna lead set cable is 
broken.
Contact Olympus.
The real time viewer cable is 
not connected.
Connect the real time viewer 
cable properly.
The antenna lead set is not 
connected to the recorder 
unit.
Connect the antenna lead set to 
the recorder unit properly.
The antenna lead set is too 
far from the capsule 
endoscope.
Reduce the distance between the 
antenna and the capsule 
endoscope.
The recorder unit 
fails to turn OFF.
The power switch is not held 
down long enough.
Hold down the power switch for 2 
seconds or more.
The battery 
charger’s charge 
lamp blinks in 
yellow. (1.5 second 
intervals)
The battery pack cannot be 
recharged because its 
temperature is below 0°C 
(32°F).
Charge the battery at or above 
0°C (32°F).
The battery 
charger’s charge 
lamp blinks in 
yellow. (0.5 second 
intervals)
The battery pack or the 
battery charger is broken.
Use another battery pack or 
battery charger.
Irregularity Possible Cause Solution

Chapter 8  Troubleshooting
235
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Recorder unit error messages
Error code on Recorder 
Unit’s Display Panel Possible Cause Solution
System error 1.  Restart the system.
2.  If the error persists, contact 
Olympus.
RTC error 1. Connect the recorder unit to 
the workstation, and perform 
initialization.
2. If error persists after 
initialization, contact Olympus.
Internal memory 
error
Contact Olympus.
Battery pack error Replace with a new battery pack.
Internal memory full Connect the recorder unit to the 
workstation, and download the 
data from the Recorder Unit.
Uncharged battery 
pack
Recharge the battery pack, or 
replace it with a charged battery 
pack.
Patient Information 
not registered
Connect the recorder unit to the 
workstation, and perform 
initialization.
Low battery pack Recharge the battery pack, or 
replace it with a charged battery 
pack.
Less than 8 hours of 
storage space left in 
internal memory
Connect the recorder unit to the 
workstation, and download the 
data from the Recorder Unit.

236
Chapter 8  Troubleshooting
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
* The life span of a battery pack is approximately 1 year from the date of 
purchase, or 320 recharging cycles, whichever is first. When the battery 
pack approaches its end-of-life, the recorder unit will prompt for 
replacement. Order a new battery pack and replace.
Real time viewer error messages
Workstation error messages
Time to replace 
battery pack.*
Order a new battery pack from 
Oympus, and replace it. If you 
continue to use the battery pack 
for a prolonged period with these 
warning icons are displayed, the 
low battery pack icon (  on 
page 235) may blink in 
continuously, and may not allow 
the examination to be performed.
Antenna lead set 
disconnected
Connect the antenna lead set to 
the recorder unit.
Error Code on Real Time 
Viewer’s Display Panel Possible Cause Solution
Uncharged battery 
pack
Recharge the battery pack, or 
replace it with a charged battery 
pack.
E000 System error 1.  Restart the system.
2.  If the error persists, contact 
Olympus.
Message Solutions
System error occurred. (E∗∗∗)
Contact service center.
Contact Olympus.
System is unstable. 
Shutdown and restart workstation.
Shut down and restart workstation.
User ID and/or password is incorrect.
Enter correct user ID/password (case sensitive).
Enter correct user ID/password 
(case sensitive) with the right 
uppercase/lowercase combination.
Invalid value entered in date and/or time. Enter correct value.
Error code on Recorder 
Unit’s Display Panel Possible Cause Solution

Chapter 8  Troubleshooting
237
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Possible recorder failure. Contact Olympus.
Error occurred during communication with 
recorder.
Check connection between 
workstation and recorder 
Invalid value entered in date of birth. Enter correct value.
Invalid value entered in height. Enter correct value.
Invalid value entered in weight. Enter correct value.
Invalid value entered in waist size. Enter correct value.
Possible recorder failure.
If initialization fails, send recorder for repair.
Try again. Contact Olympus if the 
initialization fails again.
Cannot download because there is not enough 
disk space available.
Back up examination data to the 
workstation to free up space for 
download.
Low disk space on workstation for saving 
download data.
Back up examination data from the 
workstation to free up space for 
download.
If the recorder can not be recognized, turn off the 
recorder and remove it from the cradle.
Then, connect the recorder to the cradle again.
Connect the recorder to the cradle 
again.
Unusable file. Contact Olympus.
Cannot open file because password is incorrect. Enter correct password.
Cannot open thumbnail data because 
examination data and thumbnail data are for 
different examinations.
Open the appropriate findings file 
that matches the examination data.
Cannot check out data because destination does 
not have sufficient disk space.
Create enough disk space to check 
out.
Examination data contains multiple capsule IDs.
Before starting the observation, check the images 
before and after the capsule ID change to 
determine whether to use this examination data 
in the diagnosis.
Before starting the observation, 
check the images before and after 
the capsule ID change to determine 
whether to use this examination 
data in the diagnosis.
Cannot check in data because destination does 
not have sufficient disk space.
Create sufficient disk space to 
check in.
Cannot copy data because destination does not 
have sufficient disk space.
Create sufficient disk space to copy 
data.
Cannot write data because destination does not 
have sufficient disk space.
Create sufficient disk space to write 
data.
Cannot import data because workstation does not 
have sufficient disk space.
Create sufficient disk space to import data.
Create sufficient disk space to 
import data.
Cannot open history because report file does not 
exist.
Contact Olympus.
Message Solutions

238
Chapter 8  Troubleshooting
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
In the case that you have lost your password, contact 
Olympus.
The old password you typed is incorrect.
Please retype the old password.
Retype the old password.
This disk is write-protected.
Remove the write protection, or use another disk.
Remove the write protection, or use 
another disk.
One of the hard drives is corrupt. 
Contact Service Center for repairs.
Contact Olympus.
The DVD type is not supported. Use a DVD of the type specified in 
the manual.
Printer power is off. Turn on the printer.
Cannot display print preview because printer 
driver is not installed.
Install the printer driver.
Invalid data. Contact Olympus.
The system cannot find the file specified. Contact Olympus.
The system cannot find the path specified. Contact Olympus.
The specified module could not be found. Contact Olympus.
A file name cannot contain any of following 
charactors :
¥ / : * ? " < > |
Contact Olympus.
The media is write protected. Remove the write protection or use 
another media.
The process cannot access the file because it is 
being used by another process.
Contact Olympus.
The specified file is read only. Contact Olympus.
The system cannot open the file. Contact Olympus.
Access is denied. Contact Olympus.
The process cannot access the file because 
another process has locked a portion of the file.
Contact Olympus.
The directory or file cannot be created. Contact Olympus.
The file name is too long. Shorten the file name and retry.
There is not sufficient disk space in workstation. Contact Olympus.
The process cannot access the file. Contact Olympus.
Message Solutions

Chapter 8  Troubleshooting
239
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
8.2 Returning the system for repair
Olympus is not liable for any injury or damage occurring as a 
result of repairs attempted by non-Olympus personnel.
Before sending the system for repair, contact Olympus. With the system, include 
a description of malfunction or damage and the name and telephone number of 
the individual at your location who is most familiar with the problem.
To purchase accessories and consumables, contact 
Olympus.

240
Appendix
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Appendix
Operating/storage environment
Capsule endoscope
*1 Storage outside the storage environment range may cause condensation 
and/or deterioration of the battery.
*2 no condensation (inside the capsule endoscope)
Antenna lead set
* no condensation
Recorder unit
* no condensation
Operating Environment Storage Environment
Ambient temperature 10 to 40°C (50 to 104°F) 0 to 25°C (32 to 77°F) *1
Relative humidity N/A 30 to 85% *2
Barometric pressure 700 to 1060 hPa
 (10.2 to 15.4 psia)
700 to 1060 hPa
 (10.2 to 15.4 psia)
Operating Environment Storage Environment
Ambient temperature 0 to 50°C (32 to 122°F) –20 to +70°C (–4 to +158°F)
Relative humidity 30 to 90% *10 to 95% *
Barometric pressure 700 to 1060 hPa 
(10.2 to 15.4 psia)
700 to 1060 hPa 
(10.2 to 15.4 psia)
Operating Environment Storage Environment
Ambient temperature 0 to 50°C (32 to 122°F) –20 to +70°C (–4 to +158°F)
Relative humidity 30 to 90% *10 to 95% *
Barometric pressure 700 to 1060 hPa 
(10.2 to 15.4 psia)
700 to 1060 hPa 
(10.2 to 15.4 psia)

Appendix
241
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Real time viewer
* no condensation
Recorder unit cradle
* no condensation
Battery charger
* no condensation
Operating Environment Storage Environment
Ambient temperature 10 to 40°C (50 to 104°F) –20 to +50°C (–4 to +122°F)
Relative humidity 30 to 85% *10 to 95% *
Barometric pressure 700 to 1060 hPa 
(10.2 to 15.4 psia)
700 to 1060 hPa 
(10.2 to 15.4 psia)
Operating Environment Storage Environment
Ambient temperature 10 to 40°C (50 to 104°F) –20 to +70°C (–4 to +158°F)
Relative humidity 30 to 85% *10 to 95% *
Barometric pressure 700 to 1060 hPa 
(10.2 to 15.4 psia)
700 to 1060 hPa 
(10.2 to 15.4 psia)
Operating Environment Storage Environment
Ambient temperature 0 to 40°C (32 to 104°F) –20 to +50°C (–4 to +122°F)
Relative humidity 30 to 85% *10 to 95% *
Barometric pressure 700 to 1060 hPa 
(10.2 to 15.4 psia)
700 to 1060 hPa 
(10.2 to 15.4 psia)

242
Appendix
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Battery pack
* no condensation
Recorder unit harness
* no condensation
Antenna lead cover
* no condensation
Operating Environment Storage Environment
Ambient 
temperature
0 to 50°C (32 to 122°F)
Recharging: 
0 to 40°C (32 to 104°F)
Discharging: 
0 to +50°C (–4 to +140°F)
0 to 50°C
(–32 to 122°F)
Avoid prolonged storage at 
temperatures exceeding 
40°C/104°F.
Relative humidity 30 to 90% *10 to 95% *
Barometric pressure 700 to 1060 hPa 
(10.2 to 15.4 psia)
700 to 1060 hPa 
(10.2 to 15.4 psia)
Operating Environment Storage Environment
Ambient temperature 0 to 50°C (32 to 122°F) –20 to +70°C (–4 to +158°F)
Relative humidity 30 to 90% *10 to 95% *
Barometric pressure 700 to 1060 hPa 
(10.2 to 15.4 psia)
700 to 1060 hPa 
(10.2 to 15.4 psia)
Operating Environment Storage Environment
Ambient temperature 10 to 45°C (50 to 113°F) Up to 27°C (80.6°F)
Relative humidity 30 to 90% *40 to 60% *
Barometric pressure 700 to 1060 hPa 
(10.2 to 15.4 psia)
700 to 1060 hPa 
(10.2 to 15.4 psia)

Appendix
243
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Workstation hardware
Printer
LCD monitor
Operating Environment Storage Environment
Ambient temperature 5 to 35°C (41 to 95°F) –40 to +60°C (–40 to +140°F)
Relative humidity 8 to 85% 8 to 90% *
Barometric pressure 700 to 1060 hPa 
(10.2 to 15.4 psia)
700 to 1060 hPa 
(10.2 to 15.4 psia)
Operating Environment Storage Environment
Ambient temperature 15 to 35°C (59 to 95°F) –40 to +70°C (–40 to +158°F)
Relative humidity 20 to 80% 20 to 80%
Barometric pressure 700 to 1060 hPa 
(10.2 to 15.4 psia)
700 to 1060 hPa 
(10.2 to 15.4 psia)
Operating Environment
Ambient temperature 15 to 35°C (59 to 95°F)
Relative humidity 20 to 80%
Barometric pressure 700 to 1060 hPa 
(10.2 to 15.4 psia)

244
Appendix
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Specifications
System chart
UP
UP
Antenna lead set
Capsule endoscope
Real time viewer
Recorder unit
Recorder unit cradle
Real time viewer cable
Receiver unit cradle cable
Workstation
Battery charger

Appendix
245
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
The system (components within the bold dotted line) is IEC 60601-1 certified in a 
configuration that includes the capsule endoscope, antenna lead set, recorder 
unit, real time viewer, and real time viewer cable. The recorder unit cradle, 
recorder unit cradle cable, battery charger, and workstation are IEC 60950 
certified.
• In order to provide the intended functionality, the system 
emits RF energy while in operation. This may affect electrical 
devices in the vicinity. The patient should keep a distance 
from such devices.
• Testing has shown only minimal effect on other electrical 
equipment while the system is operating. However, when 
inspecting or using the capsule endoscope, do so in a 
location such as a shielded room, where there is no risk that 
the system will interfere with the operation of other electrical 
equipment.
• The system conducts weak radio communication. Images 
may be lost or distorted when operating in an environment 
affected by electromagnetic transmissions (such as that of 
portable RF communication devices). Avoid use in such 
environments.

246
Appendix
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Capsule endoscope
Product Name Capsule endoscope
(OLYMPUS EC TYPE 1)
Optics Field of view (maximum) 145°
Depth of field 0 to 20 mm
Sampling Rate 2 fps
Power Supply Power source Internal battery
Voltage DC 2.7 to 3.1V
Current 1 – 30 mA
Battery Life 8 hours
Size Weight 3.8 g
Dimensions ø 11 mm (diameter) × 26 mm (length)
Classification as 
Medical Electrical 
Equipment
Type of protection 
against electric shock Internal power supply
Degree of protection 
against electric shock TYPE BF
Degree of protection 
against explosion
Use under combustible atmosphere 
prohibited.
Medical Device 
Directive
This device complies with the 
requirements of directive 93/42/EEC 
concerning medical devices.
Classification: Class IIa
This device complies with the EMC 
requirements of EN 60601-1-2 when 
used in combination with devices 
bearing CE marking either on the 
products or in its instructions.
Emission: Class B of EN 55011
R&TTE Directive This device complies with the 
requirements of directive 
1999/5/EEC.
Classification: Class II
EMC Applied standard;
IEC 60601-1-2:2001
IEC 60601-2-18:1996
This instrument complies with the 
standard listed in the left column only 
when connecting with the 
instruments listed in the system 
diagram on page 244.
CISPR 11 of emission:
Group 1, Class B

Appendix
247
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Year of 
Manufacturer The year of manufacture is indicated 
by the first 4 digits of the serial 
number.
* The year of manufacture indicates the serial number of MAJ-1469.
FCC ID :38QEC-1 This device complies with part 15 of 
the FCC Rules. Operation is subject 
to the following two conditions:
(1) This device may not cause 
harmful interference, and (2) this 
device must accept any interference 
received, including interference that 
may cause undesired operation.
IC :4763B-EC1 “Operation is subject to the following 
two conditions: (1) this device may 
not cause interference, and (2) this 
device must accept any interference, 
including interference that may cause 
undesired opertation of the device.”
Product Name Capsule endoscope
(OLYMPUS EC TYPE 1)
2005033

248
Appendix
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Recorder unit
Product Name Recorder Unit (OLYMPUS RE-1)
Power Supply Power source Battery pack (MAJ-1473)
Voltage DC 6.7V – 8.4V
Current 550 mA
Battery Life 8 hours or longer
Size Weight 515 g (including battery)
Dimensions 90 mm (W) × 175 mm (H) × 43 mm (D)
Classification as 
Medical Electrical 
Equipment
Type of protection 
against electric shock Internal power supply
Degree of protection 
against electric shock TYPE BF
Degree of protection 
against explosion
Use under combustible atmosphere 
prohibited.
Medical Device 
Directive
This device complies with the 
requirements of Directive 93/42/EEC 
concerning medical devices.
Classification: Class I
This device complies with the EMC 
requirements of EN 60601-1-2 when 
used in combination with devices 
bearing CE marking either on the 
products or in its instructions.
Emission: Class B of EN 55011
WEEE Directive In accordance with European Directive 
2002/96/EC on Waste Electrical and 
Electronic Equipment, this symbol 
indicates that the product must not be 
disposed of as unsorted municipal 
waste, but should be collected 
separately.
Refer to your local Olympus distributor 
for return and/or collection systems 
available in your country.
Year of 
manufacture The year of manufacture is indicated 
by the second digit of the serial 
number.
1512345

Appendix
249
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
* (with battery pack inserted)
Real time viewer
EMC Applied standard;
IEC 60601-1-2:2001
This instrument complies with the 
standard listed in the left column only 
when connecting with the instruments 
listed in the system diagram on 
page 244.
CISPR 11 of emission:
Group 1, Class B
Product Name Real Time Viewer (OLYMPUS VU-1)
Power Supply Power source Battery pack (MAJ-1473)
Voltage DC 6.7V – 8.4V
Current 750 mA
Battery Life 3 hours or longer
Size Weight 550 g (including battery)
Dimensions 90 mm (W) × 175 mm (H) × 43 mm (D)
Classification as 
Medical Electrical 
Equipment
Type of protection 
against electric shock Internal power supply
Degree of protection 
against electric shock TYPE BF
Degree of protection 
against explosion
Use under combustible atmosphere 
prohibited.
Medical Device 
Directive and
R&TTE Directive
This device complies with the 
requirements of Directive 93/42/EEC 
concerning medical devices.
Classification: Class I
This device complies with the EMC 
requirements of EN 60601-1-2 when 
used in combination with devices 
bearing CE marking either on the 
products or in its instructions.
Emission: Class B of EN 55011
This device complies with the 
requirements of directive 1995/5/EEC.
Classification: Class I
Product Name Recorder Unit (OLYMPUS RE-1)

250
Appendix
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
* (with battery pack inserted)
Recorder unit cradle
WEEE Directive In accordance with European Directive 
2002/96/EC on Waste Electrical and 
Electronic Equipment, this symbol 
indicates that the product must not be 
disposed of as unsorted municipal 
waste, but should be collected 
separately.
Refer to your local Olympus distributor 
for return and/or collection systems 
available in your country.
Year of 
manufacture The year of manufacture is indicated 
by the second digit of the serial 
number.
EMC Applied standard;
IEC 60601-1-2:2001
This instrument complies with the 
standard listed in the left column only 
when connecting with the instruments 
listed in the system diagram on 
page 244.
CISPR 11 of emission:
Group 1, Class B
Product Name Recorder Unit Cradle (MAJ-1484)
Power Supply Voltage AC 100 – 240V
Frequency 50 – 60Hz
Current 2A
Size Weight Main body:230 g     AC Adapter:275 g
Dimensions 128 mm (W) × 80.8 mm (H) × 114.5 mm (D)
Medical Device 
Directive
This device complies with the requirements 
of directive 93/42/EEC concerning medical 
devices.
Classification: Class I
This device complies with the EMC 
requirements of EN 55022 and EN 55024 
when used in combination with devices 
bearing CE marking either on the products 
or in its instructions.
Emission: Class B of EN 55022
Product Name Real Time Viewer (OLYMPUS VU-1)
1512345

Appendix
251
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Battery charger
WEEE Directive In accordance with European Directive 
2002/96/EC on Waste Electrical and 
Electronic Equipment, this symbol indicates 
that the product must not be disposed of as 
unsorted municipal waste, but should be 
collected separately.
Refer to your local Olympus distributor for 
return and/or collection systems available in 
your country.
Year of 
manufacture The year of manufacture is indicated by the 
second digit of the serial number.
Product Name Battery Charger (MAJ-1476)
Power Supply Voltage AC 100 – 240V
Frequency 50 – 60Hz
Current AC 100V 53VA / AC 240V 70VA
Size Weight 330 g
Dimensions 110 mm (W) × 48.6 mm (H) × 150 mm (D)
Medical Device 
Directive
This device complies with the requirements 
of directive 93/42/EEC concerning medical 
devices.
Classification: Class I
This device complies with the EMC 
requirements of EN 55022 and EN 55024 
when used in combination with devices 
bearing CE marking either on the products 
or in its instructions.
Emission: Class B of EN 55022.
WEEE Directive In accordance with European Directive 
2002/96/EC on Waste Electrical and 
Electronic Equipment, this symbol 
indicates that the product must not be 
disposed of as unsorted municipal waste, 
but should be collected separately.
Refer to your local Olympus distributor for 
return and/or collection systems available 
in your country.
Product Name Recorder Unit Cradle (MAJ-1484)
1512345

252
Appendix
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Battery pack
Real time viewer cable
Product Name Battery pack (MAJ-1473)
Type Lithium-Ion storage cell
Capacity 3600mAh
Voltage 7.4V
Recharging 
Time Approx. 2 hours
Size Weight 180 g
Dimensions 72 mm (W) × 24.7 mm (H) × 60 mm (D)
Medical Device 
Directive
This device complies with the requirements 
of Directive 93/42/EEC concerning medical 
devices.
Classification: Class I
This device complies with the EMC 
requirements of EN 60601-1-2 when used 
in combination with devices bearing CE 
marking either on the products or in its 
instructions.
Emission: Class B of EN 55011.
In accordance with the EU Batteries 
Directive, the battery packs are marked 
with a recycle logo. Follow all applicable 
regulations in their handling. If you are 
unsure of the regulations for your country, 
contact Olympus.
Product Name Real Time Viewer cable (MAJ-1485)
Length 1.05 m

Appendix
253
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Antenna lead set
Antenna lead cover
Product Name Antenna Lead Set (MAJ-1474)
Size Weight 200 g
Dimensions 88 mm (W) × 68.7 mm (H) × 23 mm (D)
The number of 
antennas 8
Medical Device 
Directive
and
R&TTE Directive
This device complies with the requirements of 
Directive 93/42/EEC concerning medical 
devices.
Classification: Class I
This device complies with the EMC 
requirements of EN 60601-1-2 when used in 
combination with devices bearing CE marking 
either on the products or in its instructions.
Emission: Class B of EN 55011
This device complies with the requirements of 
directive 1995/5/EEC.
Classification: Class I
WEEE Directive In accordance with European Directive 
2002/96/EC on Waste Electrical and Electronic 
Equipment, this symbol indicates that the 
product must not be disposed of as unsorted 
municipal waste, but should be collected 
separately.
Refer to your local Olympus distributor for 
return and/or collection systems available in 
your country.
Product Name Antenna Lead Cover (MAJ-1470)
Size Dimensions 92 mm (W) × 66 mm (H)

254
Appendix
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Recorder unit harness
Workstation hardware
Product Name Recorder Unit Harness (MAJ-1475)
Size Weight  220 g
Dimensions 1770 mm (W) × 220 mm (H)
Product Name Workstation Hardware (MAJ-1479)
Power Supply AC 100 – 240V / 50 – 60 Hz / 400 Hz
Power 
Consumption 450W
Size Weight 12.1 kg
Dimensions 168 mm (W) × 450 mm (H) × 456 mm (D)
Function Initialize Transfer the patient information and 
examination data from a workstation to a 
recorder before the examination.
Download Transfer the recorded data from a recorder to 
a workstation after the examination.
Observation Display images downloaded from a recorder.
Making report Insert the images and add comments to make 
a report.
Exporting picture / 
movie
Make pictures and movies from the selected 
thumbnails and export them.
Red color detection Select images showing the suspected 
bleeding symptom from all the recorded 
images.
Medical Device 
Directive
This device complies with the requirements of 
directive 93/42/EEC concerning medical 
devices.
Classification: Class I
This device complies with the EMC 
requirements of EN 55022 and EN 55024 
when used in combination with devices 
bearing CE marking either on the products or 
in its instructions.
Emission: Class B of EN 55022

Appendix
255
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Printer
LCD monitor
Product name Printer (MAJ-1480)
Power consumption 75W
Size Weight 9kg
Dimensions 510 mm (W) x 210 mm (H) x 420 mm (D)
Product name LCD monitor (MAJ-1481)
Power consumption 25W
Size Weight 5.7 kg
Dimensions 351 mm (W) x 422 mm (H) x 210 mm (D)

256
Appendix
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
EMC information
Compliance information and suggested electromagnetic 
environment: electromagnetic emission
The system is designed for use under the following electromagnetic 
environment.
Emission Standard Applicability Notes
Radio Frequency 
Emission 
CISPR 11
Group 1 In order to provide the intended 
functionality, the system must emit RF 
energy while in operation. This may affect 
electrical devices in the vicinity.
Radio Frequency 
Emission 
CISPR 11
Class B The amount of RF emission by the system 
is very small, with little possibility of 
affecting nearby electrical equipment.
Harmonic Emissions
IEC 61000-3-2
Not applicable The system is not powered by a 
commercial power source.
Voltage 
Fluctuations/Flicker 
Emissions
IEC 61000-3-3
Not applicable The system is not powered by a 
commercial power source.

Appendix
257
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Compliance information and suggested electromagnetic 
environment: electromagnetic immunity
Immunity
Test
IEC 60601-1-2
Testing Level
Compatibili
ty Level Suggested Environment
Electrostatic 
Discharge 
(ESD)
IEC 61000-4-2
Contact: ±2, ±4, ±6kV
Aerial: ±2, ±4, ±8kV
See left Material unlikely to 
produce static electricity, 
such as wood, concrete, 
or ceramic tiles, are 
recommended for the 
floor. If the floor is made of 
materials that are likely to 
produce static electricity, 
maintain the relative 
humidity above 30%.
Electrical Fast 
Transient / 
Burst
IEC 61000-4-4
Power line ±2kV
I/O line ±1kV
Not 
applicable
The system does not have 
a cable for this signal.
Surge
IEC 61000-4-5
Differential mode: 
±0.5, ±1kV
Common mode: 
±0.5, ±1, ±2kV
Not 
applicable
The system is not 
powered by a commercial 
power source.
Voltage Dips,
Short 
Interrutions and 
Voltage 
Variations on 
Power Supply 
Input Lines
IEC 61000-4-11
<5%UT (>95% dip in UT)
for 0.5 cycle
40%UT (60% dip in UT)
for 5 cycle
70%UT (30% dip in UT)
for 25 cycle
<5%UT (>95% dip in UT)
for 5 sec.
Not 
applicable
The system is not 
powered by a commercial 
power source.
Power 
Frequency 
(50/60Hz) 
Magnetic Field
IEC 61000-4-8
3A/m See left Power frequency 
magnetic field should be 
similar in characteristics to 
general locations within a 
standard commercial or 
medical environment.

258
Appendix
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Cautions and recommended electromagnetic environment 
regarding portable and mobile RF communications 
equipment such as a cellphone
Immunity
Test
IEC 60601-1-2
Testing Level
Compatibility 
Level
Recommended separation 
distance
Radiated 
Immunity
IEC 61000-4-3
3V/m
(80MHz - 
2.5GHz)
See left d=1.2√P  80MHz to 800MHz
d=1.2√P  80MHz to 2.5GHz
Conducted 
Immunity
IEC 61000-4-6
3Vrms Not applicable —
Keep portable RF communication devices away from all parts of the system (including 
the cables) by at least the recommended separation distance, computed for the 
transmission frequency.
Here, P is the maximum power export of the device, as specified in watts (W) by the 
manufacturer, and d is the recommended separation distance in meters (m).
The system satisfies the requirements for IEC 60601-1-2 compliance, but may 
experience electromagnetic interference in an electromagnetic environment exceeding 
the requirements.
The system may experience electromagnetic interference near high frequency 
operating devices, and other devices marked with the following symbols.

Appendix
259
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Recommended separation distance between system and 
portable RF communication equipment
The above table provides some calculations of the distance 
between the system and the transmitter. The values may 
vary depending on the architectural structure and the 
environment in which the system is used.
Separation Distance Based on Transmission Frequency 
(m)
Maximum Power 
Output of
Transmitter P(W)
0.15 MHz to 80 MHz
d = 1.2
80 MHz to 800 MHz
d = 1.2
800 MHz to 2.5 GHz
d = 2.3
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
11.21.22.3
10 3.8 3.8 7.3
100 12 12 23
PPP

260
Appendix
OLYMPUS CAPSULE ENDOSCOPE SYSTEM

©2005 OLYMPUS MEDICAL SYSTEMS CORP. All rights reserved. 
No part of this publication may be reproduced or distributed without the 
express written permission of OLYMPUS MEDICAL SYSTEMS CORP.
OLYMPUS is a registered trademark of OLYMPUS CORPORATION.
Microsoft, Windows, Windows, NT, PowerPoint, and MS-DOS are 
either registered trademarks or trademarks of Microsoft Corporation in 
the United States and/or other countries.

Manufactured by
2951 Ishikawa-cho, Hachioji-shi, Tokyo 192-8507, Japan
Fax: (0426)46-2429  Telephone: (0426)42-2111
(Premises/Goods delivery) Wendenstrasse 14-18, 20097 Hamburg, Germany
(Letters) Postfach 10 49 08, 20034 Hamburg, Germany  Telephone: (040)237730
Two Corporate Center Drive, PO Box 9058 Melville, N.Y.
11747-9058, U.S.A.
Fax: (631)844-5442  Telephone: (631)844-5000
KeyMed House, Stock Road, Southend-on-Sea, Essex SS2 5QH, United Kingdom
Fax: (01702)465677  Telephone: (01702)616333
491B, River Valley Road #12-01/04, Valley Point Office Tower, Singapore 248373
Fax: 6834-2438  Telephone: 6834-0010
Room 1406, E Tower, GongYuan No.6 Royal Palace, No.6 GongYuanXiJie, Jian Guo Men Nei,
DongCheng District, Beijing, 100005, China
Fax: (10)6518-0865  Telephone: (10)6518-8080
117071, Moscow, Malaya Kaluzhskaya 19, bld. 1, fl.2, Russia
Fax: (095)958-2277  Telephone: (095)958-2245
31 Gilby Road, Mount Waverley, VIC., 3149, Australia
Fax: (03)9543-1350  Telephone: (03)9265-5400
6100 Blue Lagoon Drive, Suite 390 Miami, FL 33126-2087, U.S.A.
Fax: (305)261-4421  Telephone: (305)266-2332
Distributed by
 Printed in Japan 20050124 *0000
GT1629  FCC
One Corporate Drive, Orangeburg, N.Y. 10962, U.S.A.
Fax: (845)398-9444  Telephone: (845)398-9400
8F, Hyundai Marines Bldg., 646-1, Yeoksam-Dong, Kangnam-Gu, Seoul 135-080 Korea
Fax: (02)6255-3499  Telephone: (02)1544-3200