Powerwave Technologies 5JS0056 Multi Carrier RF Power Amplifier User Manual 273680
Powerwave Technologies Inc Multi Carrier RF Power Amplifier 273680
Contents
Manual 5

 G3S-800-140-031 Installation & Service Manual 
 Copyright Powerwave Technologies, Inc., September 2001. All rights reserved 
044-05095 Rev. A 5-1  September 2001 
Section 5  Maintenance 
5-1  Introduction 
This section contains periodic maintenance and performance test procedures for the Multicarrier 
Cellular Amplifier.  It also contains a list of test equipment required to perform the identified tasks. 
NOTE 
Check your sales order and equipment warranty before attempting to service or repair 
the unit.  Do not break the seals on equipment under warranty or the warranty will be 
null and void.  Do not return equipment for warranty or repair service until proper 
shipping instructions are received from the factory.  
5-2  Periodic Maintenance 
Periodic maintenance requirements are listed in table 5-1. Table 5-1 also lists the intervals at 
which the tasks should be performed. 
WARNING 
Wear proper eye protection to avoid eye injury when using compressed air. 
Table 5-1   Periodic Maintenance 
Task Interval  Action 
Cleaning   
Air Vents  30 Days  Inspect and clean per paragraph 5-4 
Inspection   
Cables and Connec-
tors 
12 Months  Inspect signal and power cables for 
frayed insulation.  Check RF connectors 
to be sure that they are tight. 
Performance Tests 12 Months  Perform annual test per paragraph 5-5. 
5-3  Test Equipment Required For Test 
Test equipment required to test the amplifier system is listed in table 5-2.  Equivalent test equip-
ment may be substituted for any item, keeping in mind that a thermistor type power meter is re-
quired. 
NOTE 
All RF test equipment must be calibrated to 0.05 dB resolution.  Any deviation from 
the nominal attenuation must be accounted for and factored into all output readings. 

 G3S-800-140-031 Installation & Service Manual 
 Copyright Powerwave Technologies, Inc., September 2001. All rights reserved 
044-05095 Rev. A 5-2  September 2001 
Table 5-2   Test Equipment Required 
Nomenclature Manufacturer Model 
Signal Generator  RDL  IMD-801D-03A 
30 dB Attenuator, 500 Watt  Weinschel Corp.  53-30-34 
20 dB Attenuator, 20 Watt 
(2 each) 
Tenuline  
Spectrum Analyzer  H.P.  8560E 
Coax Directional Coupler  H.P.  778D 
Power Meter/Sensor  H.P.  437B/8481A 
Network Analyzer  H.P.  8753C 
Current Probe     
5-4  Cleaning Air Inlets/Outlets 
The air inlets and outlets should be cleaned every 30 days.  If the equipment is operated in a se-
vere dust environment, they should be cleaned more often as necessary.  Turn off DC power 
source before removing fans.  If dust and dirt are allowed to accumulate, the cooling efficiency 
may be diminished.  Using either compressed air or a brush with soft bristles, loosen and remove 
accumulated dust and dirt from the air inlet panels. 
5-5  Performance Test 
Performance testing should be conducted every 12 months to ensure that the amplifier system 
meets the operational specifications listed in table 5-3.  Also verify system performance after any 
amplifier module is replaced in the field.  The test equipment required to perform the testing is 
listed in table 5-2, and the test setup is shown in figure 5-1. 
NOTE 
The frequencies used in this test are typical for an amplifier with a 18 MHz band 
from 851 MHz to 869 MHz.  Select evenly spaced F1, F2, F3, and F4 frequencies 
that cover the instantaneous bandwidth of your system. 
5-5.1   Amplifier System Performance Test 
This test is applicable to the G3S-800-140-031 amplifier modules. To perform the test, proceed 
as follows: 
1.  Connect test equipment to the amplifier as shown in figure 5-1. 
NOTE 
Do not apply any RF signals at this time. 
Turn on signal generator and set frequency F1 to 854 MHz, F2 to 857 MHz, F3 to 863 MHz, 
and F4 to 866 MHz.  Adjust each signal generator output so that the sum power output from 
all four signal generators equals -6 dBm at the input. 

 G3S-800-140-031 Installation & Service Manual 
 Copyright Powerwave Technologies, Inc., September 2001. All rights reserved 
044-05095 Rev. A 5-3  September 2001 
Unit Under Test
G3S-800-140-031
Plug-in
A
mplifier
 Module
+27 Vdc
Gnd RF In
RF
Out
20 dB
Directionl Coupler
30 dB
A
ttenuator
500 W
20 dB
A
ttenuator
20 W Power Meter
Sensor Head
8482A
20
dB
A
tt
en
ua
tor
20
W
Dir
ect
ion
l
Co
upl
er
Power Meter
Sensor Head
8482A
Network Analyzer
8753C
Signal
Generator
Filter /
Isotlator
10 dB
Variable
A
ttenuator
Spectrum Analyzer
8651E
Figure 5-1   Amplifier System Test Setup Diagram 
5-5.1.1   Amplifier IMD Test And Current Test 
2.  Adjust attenuator for an input signal at -10 dBm.  Turn on the amplifier by setting RF ON 
switch of amplifier. Adjust variable attenuator to set amplifier power output on power meter 
to 140 watts.  Measure IMD on spectrum analyzer.  IMD should be -60 dBc max.  Record 
test data in table 5-3.  Set RF ON switch to OFF. 
3.  With the amplifier module set at 140 watts power output, use the current probe (magnetic 
field type) and measure the dc current flow from the +27 Vdc power source.  Current should 
be 70 amps maximum.  Record test data in table 5-3. 
5-5.1.2   Gain Test 
4.  Disconnect spectrum analyzer from test setup, and connect the network analyzer. 
5.  Set network analyzer as follows: 
¾  Power output to -10 dBm. 
¾  Frequency start to 869 MHz. 
¾  Frequency stop to 894 MHz. 
¾  Normalize the network analyzer for gain and return loss. 
6.  Check the gain across the band from 869 MHz to 894 MHz.  Gain should be between 58 dB.  
Record test data in table 5-3. 
5-5.1.3   Harmonics Test 
7.  With the power set at 140 watts power output, use the spectrum analyzer and check the fre-
quency band from 851 MHz to 869 MHz for harmonics.  Harmonics should be 5 dBm maxi-
mum.  Record test data in table 5-3. 

 G3S-800-140-031 Installation & Service Manual 
 Copyright Powerwave Technologies, Inc., September 2001. All rights reserved 
044-05095 Rev. A 5-4  September 2001 
5-5.1.4   Spurious Test 
8.  With the power amplifier set at 140 watts power output, use the spectrum analyzer and 
check the frequency band from 851 MHz to 869 MHz for spurious signals.  Spurious signals 
should be -60 dBc maximum.  Record test data in table 5-3. 
5-5.1.5   Input Return Loss Test 
9.  Reset and turn on amplifier module.  Read and record the S11 return loss measurement on 
network analyzer.  Input return loss should be –16 dB maximum.  Record test data in table  
5-3. 

 G3S-800-140-031 Installation & Service Manual 
 Copyright Powerwave Technologies, Inc., September 2001. All rights reserved 
044-05095 Rev. A 5-5  September 2001 
Table 5-3   Multicarrier Cellular Amplifier Test Data Sheet 
DATE _________________________________ 
AMPLIFIER S/N _________________________ 
TEST CONDITIONS: 
Load and Source Impedance:  50 Ohms 
VSWR:  < 1.2:1 
Supply Voltage:  +27 Vdc ±1.0 Vdc 
TEST SPECIFICATION MIN MAX DATA 
4-TONE IMD 
Vcc = 27 Vdc 
PO = 140 W 
Freq.: 853, 857, 863, and 866 
MHz 
-60 dBc 
RF Gain 
Vcc = 27 Vdc 
PO = 140 W 
Freq. = 860 MHz 
57.5 dB 
58.5 dB 
Gain Flatness 
Vcc = 27 Vdc ±1 Vdc 
PO =140 W 
851-869 MHz Band 
-0.5 dB 
+0.5 dB 
Harmonics 
Vcc = 27 Vdc 
PO = 140 W 
851-869 MHz Band 
5 dBm 
Spurious 
Vcc = 27 Vdc 
PO =140 W 
851-869 MHz Band 
-60 dBc 
Input Return 
Loss 
Vcc = 27 Vdc 
PO = 140 W 
851-869 MHz Band 
-16 dB 
DC Power 
Vcc = 27 Vdc 
PO = 140 W 
4 Tones 
70 Amps 
PASS _________________________________  FAIL  ______________________________  
Tested by ______________________________ 

 G3S-800-140-031 Installation & Service Manual 
 Copyright Powerwave Technologies, Inc., September 2001. All rights reserved 
044-05095 Rev. A 5-6  September 2001 
5-6  Field Replaceable Parts And Modules 
The following parts and modules can be replaced in the field on site by a qualified technician with 
experience maintaining RF power amplifiers and similar equipment: 
1.  G3S-800-140-031 power amplifier modules 
2. Cooling fans 
5-6.1   G3S-800-140-031 Power Amplifier Module 
To replace a power amplifier module, proceed as follows: 
1.  Set both the RF ON On/Off/Reset switch and the power ON/OFF switch on the front panel of 
the amplifier module to OFF. 
2.  Loosen two screws that secure amplifier module to subrack. 
3.  Use handle on front of module, and with a steady even pressure, pull module out of subrack. 
CAUTION 
When removing the amplifier from the subrack, it is very important to support the 
amplifier such that the rear of the module does not suddenly drop when it disengages 
from the track.  A drop such as this could damage the module. 
5-6.2   Cooling Fans 
To replace a cooling fan, proceed as follows: 
1.  Remove amplifier module from subrack; see paragraph 5-6.1 preceding. 
2.  Loosen four snap fasteners that secure fan to amplifier module.  Disconnect fan power con-
nector from amplifier module. 
Install replacement in reverse order of steps 1 and 2 above.